• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall medtronic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall medtronicsee related information
Date Initiated by FirmMay 10, 2023
Date PostedJune 16, 2023
Recall Status1 Open3, Classified
Recall NumberZ-1819-2023
Recall Event ID 92328
PMA NumberP980016S713 
Product Classification Implantable cardioverter defibrillator (non-CRT) - Product Code LWS
ProductICD COBALT XT VR MRI DF4, Model Number DVPA2D4; Implantable Cardioverter Defibrillators
Code Information GTIN 00763000178468. Lot Serial numbers: RSD605765S, RSD605767S, RSD607880S, RSD607881S, RSD607886S, RSD607890S, RSD603551S, RSD603552S, RSD603553S, RSD603554S, RSD603555S, RSD600536S, RSD601541S, RSD602283S, RSD602524S, RSD602542S, RSD602924S, RSD603303S, RSD603396S, RSD607136S, RSD608177S, RSD608547S, RSD606272S, RSD607000S, RSD607711S, RSD607712S, RSD602036S, RSD602645S, RSD602659S, RSD603096S, RSD603670S, RSD604815S, RSD608193S, RSD605918S, RSD605919S, RSD605920S, RSD605921S, RSD605922S, RSD605923S, RSD605924S, RSD605926S, RSD605927S, RSD605929S, RSD605930S, RSD605931S, RSD605932S, RSD605933S, RSD604238S, RSD604294S, RSD604312S, RSD604313S, RSD604314S, RSD604614S, RSD604616S, RSD604710S, RSD604721S, RSD604725S, RSD605925S, RSD607788S, RSD607789S, RSD607790S, RSD607791S, RSD607792S, RSD607793S, RSD607794S, RSD607795S, RSD607797S, RSD607800S, RSD607801S, RSD607802S, RSD609083S, RSD609084S, RSD609085S, RSD605068S, RSD608525S, RSD609065S, RSD609066S, RSD609067S, RSD609069S, RSD609071S, RSD609072S, RSD609073S, RSD609075S, RSD609079S, RSD609080S, RSD609081S, RSD609082S, RSD601603S, RSD601960S, RSD602778S, RSD603181S, RSD603321S, RSD603672S, RSD603879S, RSD606554S, RSD608293S, RSD600145S, RSD601926S, RSD602728S, RSD603152S, RSD604065S, RSD607875S, RSD603750S, RSD606677S, RSD600100S, RSD600200S, RSD601455S, RSD601551S, RSD601552S, RSD601750S, RSD602233S, RSD608150S, RSD609043S, RSD608404S, RSD608422S, RSD609843S, RSD609845S, RSD609846S, RSD603825S, RSD603826S, RSD609842S, RSD610592S, RSD610593S, RSD608739S, RSD608718S, RSD606638S, RSD606653S, RSD608253S, RSD601449S, RSD601450S, RSD603168S, RSD610535S, RSD600188S, RSD605289S, RSD605802S, RSD606161S, RSD608659S, RSD608720S, RSD603329S, RSD604030S, RSD605997S, RSD605998S, RSD608060S, RSD605348S, RSD605787S, RSD605790S, RSD606704S, RSD608634S, RSD608795S, RSD608855S, RSD609276S, RSD609277S, RSD609278S, RSD600209S, RSD600214S, RSD600366S, RSD600475S, RSD600484S, RSD600485S, RSD600499S, RSD600513S, RSD600525S, RSD600531S, RSD602048S, RSD602376S, RSD602378S, RSD602600S, RSD603398S, RSD604018S, RSD604559S, RSD605190S, RSD610520S, RSD610522S, RSD607657S, RSD600316S, RSD611020S, RSD603301S, RSD608946S, RSD610202S, RSD601558S, RSD601934S, RSD605328S, RSD605330S, RSD605333S, RSD605771S, RSD606050S, RSD606273S, RSD606413S, RSD606708S, RSD606948S, RSD606950S, RSD607114S, RSD607319S, RSD607703S, RSD608135S, RSD608453S, RSD608458S, RSD608560S, RSD608561S, RSD608563S, RSD609935S, RSD610123S, RSD600348S, RSD600349S, RSD600554S, RSD600609S, RSD602325S, RSD602582S, RSD602587S, RSD603101S, RSD603127S, RSD603297S, RSD603309S, RSD603363S, RSD603386S, RSD604519S, RSD605221S, RSD609936S, RSD605349S, RSD605799S, RSD600498S, RSD600501S, RSD600526S, RSD601840S, RSD608211S, RSD610933S, RSD605143S, RSD605747S, RSD605759S, RSD605762S, RSD605960S, RSD606078S, RSD606179S, RSD606836S, RSD606837S, RSD606838S, RSD607819S, RSD607823S, RSD608173S, RSD608239S, RSD608278S, RSD608555S, RSD608617S, RSD608618S, RSD608619S, RSD608622S, RSD608919S, RSD609186S, RSD610047S, RSD610081S, RSD610126S, RSD603694S, RSD603695S, RSD603696S, RSD603698S, RSD603706S, RSD603707S, RSD603840S, RSD603846S, RSD603884S, RSD604095S, RSD604096S, RSD604097S, RSD604220S, RSD604577S, RSD604767S, RSD605144S, RSD610136S, RSD610470S, RSD605466S, RSD605725S, RSD606088S, RSD606090S, RSD606732S, RSD606733S, RSD606734S, RSD603662S, RSD607692S, RSD608434S, RSD608630S, RSD608728S, RSD602666S, RSD603643S, RSD604822S, RSD607310S, RSD610154S, RSD602490S, RSD604734S, RSD604748S, RSD605724S, RSD607667S, RSD607668S, RSD607669S, RSD608120S, RSD608121S, RSD604717S, RSD607834S, RSD608947S, RSD610041S, RSD605304S, RSD602534S, RSD604804S, RSD605138S, RSD600615S, RSD600103S, RSD600106S, RSD600138S, RSD600150S, RSD602411S, RSD602402S, RSD602430S, RSD601435S, RSD602346S, RSD602407S, RSD602450S, RSD602684S, RSD603008S, RSD607163S, RSD610016S, RSD608939S, RSD604595S, RSD604597S, RSD604598S, RSD604601S, RSD604603S, RSD604605S, RSD604606S, RSD604607S, RSD604608S, RSD604610S, RSD604611S, RSD604612S, RSD608938S, RSD609017S, RSD609018S, RSD609019S, RSD609020S, RSD609021S, RSD609022S, RSD607651S, RSD607654S, RSD607658S, RSD607660S, RSD607662S, RSD600216S, RSD602212S, RSD602603S, RSD602741S, RSD602742S, RSD602743S, RSD602746S, RSD603040S, RSD603939S, RSD604478S, RSD610015S, RSD605331S, RSD605447S, RSD605464S, RSD606053S, RSD607493S, RSD608136S, RSD610193S, RSD605225S, RSD608809S, RSD608878S, RSD603094S, RSD606072S, RSD606735S, RSD608266S, RSD608267S, RSD608895S, RSD609991S, RSD604327S, RSD606969S, RSD607868S, RSD608624S, RSD606976S, RSD602299S, RSD603345S, RSD603611S, RSD605726S, RSD605727S, RSD606425S, RSD606531S, RSD606668S, RSD606669S, RSD607069S, RSD607070S, RSD607339S, RSD607341S, RSD607503S, RSD607596S, RSD607916S, RSD608341S, RSD608607S, RSD608608S, RSD608752S, RSD608753S, RSD609056S, RSD609057S, RSD610227S, RSD610233S, RSD610534S, RSD610766S, RSD610769S, RSD605353S, RSD605354S, RSD605896S, RSD605897S, RSD605913S, RSD605914S, RSD606280S, RSD606281S, RSD606754S, RSD607125S, RSD607582S, RSD608861S, RSD600193S, RSD600419S, RSD600441S, RSD600446S, RSD600579S, RSD601268S, RSD601392S, RSD601801S, RSD601842S, RSD601964S, RSD602058S, RSD602059S, RSD602204S, RSD602239S, RSD602384S, RSD602406S, RSD602526S, RSD602552S, RSD602780S, RSD602875S, RSD602877S, RSD603421S, RSD603956S, RSD603977S, RSD603979S, RSD604044S, RSD604049S, RSD604110S, RSD604530S, RSD604757S, RSD604806S, RSD604893S, RSD604961S, RSD605191S, RSD611147S, RSD611148S, RSD608157S, RSD603622S, RSD608807S, RSD608639S, RSD610025S, RSD610162S, RSD608872S, RSD602497S, RSD604475S, RSD604477S, RSD605139S, RSD601935S, RSD602389S, RSD605323S, RSD605803S, RSD600239S, RSD601555S, RSD602326S, RSD602328S, RSD602380S, RSD602393S, RSD602456S, RSD602469S, RSD602618S, RSD603120S, RSD605758S, RSD605761S, RSD608913S, RSD608914S, RSD608995S, RSD608999S, RSD609270S, RSD605142S, RSD608996S, RSD608997S, RSD608998S, RSD609000S, RSD601877S, RSD608645S, RSD610033S, RSD600319S, RSD601434S, RSD601878S, RSD602282S, RSD602731S, RSD608867S, RSD602485S, RSD602493S, RSD602630S, RSD602706S, RSD602773S, RSD602930S, RSD602931S, RSD603001S, RSD603004S, RSD603215S, RSD603248S, RSD603251S, RSD603366S, RSD603367S, RSD602350S, RSD602479S, RSD602705S, RSD608488S, RSD602500S, RSD607532S, RSD603731S, RSD603768S, RSD608096S, RSD608097S, RSD608098S, RSD608099S, RSD609811S, RSD609813S, RSD609814S, RSD609810S, RSD609812S, RSD609815S, RSD605355S, RSD605356S, RSD606265S, RSD606266S, RSD606696S, RSD606729S, RSD606939S, RSD607107S, RSD607230S, RSD607231S, RSD607544S, RSD608141S, RSD608197S, RSD608430S, RSD608769S, RSD609914S, RSD610118S, RSD610293S, RSD600223S, RSD600231S, RSD600233S, RSD600250S, RSD600266S, RSD600270S, RSD600326S, RSD600330S, RSD600364S, RSD600422S, RSD600565S, RSD601792S, RSD601936S, RSD601937S, RSD602145S, RSD602166S, RSD602213S, RSD602297S, RSD602354S, RSD602686S, RSD602900S, RSD602993S, RSD603021S, RSD603858S, RSD603866S, RSD603872S, RSD603933S, RSD604008S, RSD604013S, RSD604208S, RSD604214S, RSD604759S, RSD604760S, RSD605192S, RSD607305S, RSD600386S, RSD600607S, RSD602675S, RSD602992S, RSD603898S, RSD605915S, RSD605935S, RSD606256S, RSD606258S, RSD601882S, RSD601884S, RSD601885S, RSD601886S, RSD601888S, RSD604502S, RSD604503S, RSD605157S, RSD605158S, RSD605934S, RSD609016S, RSD605368S, RSD605369S, RSD606782S, RSD606803S, RSD606840S, RSD606841S, RSD607292S, RSD607891S, RSD608080S, RSD608081S, RSD608082S, RSD608219S, RSD608337S, RSD608570S, RSD608783S, RSD608924S, RSD609007S, RSD610166S, RSD610591S, RSD605365S, RSD605902S, RSD606974S, RSD607833S, RSD608176S, RSD608754S, RSD602015S, RSD608831S, RSD607380S, RSD606233S, RSD606247S, RSD609704S, RSD609705S, RSD602458S, RSD602509S, RSD602713S, RSD604029S, RSD604537S, RSD607708S, RSD601793S, RSD602521S, RSD603070S, RSD603735S, RSD608586S, RSD602540S, RSD605774S, RSD606194S, RSD606270S, RSD606701S, RSD606703S, RSD606997S, RSD607110S, RSD607487S, RSD607554S, RSD607556S, RSD607557S, RSD607694S, RSD608203S, RSD608559S, RSD608730S, RSD608773S, RSD609925S, RSD609930S, RSD610097S, RSD610098S, RSD610181S, RSD610300S, RSD600259S, RSD602585S, RSD602606S, RSD602620S, RSD603610S, RSD603646S, RSD610475S, RSD608880S, RSD609052S, RSD600401S, RSD602545S, RSD603326S, RSD603652S, RSD601410S, RSD601412S, RSD609103S, RSD609135S, RSD609138S, RSD609145S, RSD608129S, RSD608647S, RSD600199S, RSD601363S, RSD601372S, RSD601904S, RSD603745S, RSD605322S, RSD606414S, RSD606415S, RSD606688S, RSD606736S, RSD607173S, RSD608190S, RSD608268S, RSD608602S, RSD600311S, RSD600563S, RSD602336S, RSD602397S, RSD602621S, RSD602772S, RSD603074S, RSD603975S, RSD605218S, RSD610568S, RSD610569S, RSD600222S, RSD600581S, RSD602120S, RSD602870S, RSD603207S, RSD607237S, RSD602661S, RSD604820S, RSD609209S, RSD609210S, RSD609212S, RSD603016S, RSD603250S, RSD607321S, RSD608921S, RSD611122S, RSD608859S, RSD610763S, RSD600370S, RSD608772S, RSD607061S, RSD607223S, RSD607311S, RSD608155S, RSD608243S, RSD608262S, RSD602539S, RSD603741S, RSD603751S, RSD603844S, RSD608128S, RSD610119S, RSD600489S, RSD610515S, RSD607829S, RSD600371S, RSD600460S, RSD601354S, RSD601828S, RSD602462S, RSD602682S, RSD603196S, RSD604021S, RSD605989S, RSD605990S, RSD609635S, RSD609636S, RSD608204S, RSD603876S, RSD605351S, RSD606900S, RSD608759S, RSD610223S, RSD600210S, RSD601389S, RSD603005S, RSD603039S, RSD604534S, RSD604780S, RSD606238S, RSD606489S, RSD606490S, RSD606657S, RSD606983S, RSD607459S, RSD607460S, RSD607811S, RSD607812S, RSD607813S, RSD608362S, RSD608609S, RSD608610S, RSD608611S, RSD608613S, RSD608614S, RSD608750S, RSD608751S, RSD609046S, RSD609047S, RSD610130S, RSD610131S, RSD610134S, RSD604121S, RSD604122S, RSD604123S, RSD604124S, RSD604126S, RSD604128S, RSD604130S, RSD604131S, RSD604132S, RSD604133S, RSD604134S, RSD604135S, RSD604138S, RSD604139S, RSD604141S, RSD604142S, RSD604147S, RSD604682S, RSD604684S, RSD604690S, RSD604691S, RSD604692S, RSD604693S, RSD604694S, RSD604695S, RSD604697S, RSD604699S, RSD604700S, RSD604702S, RSD604703S, RSD604704S, RSD604705S, RSD604707S, RSD604873S, RSD604875S, RSD604876S, RSD604877S, RSD608612S, RSD608615S, RSD610082S, RSD610129S, RSD610132S, RSD610133S, RSD606294S, RSD606297S, RSD606427S, RSD606973S, RSD607131S, RSD607222S, RSD607225S, RSD607570S, RSD607571S, RSD608183S, RSD600185S, RSD600215S, RSD600217S, RSD600261S, RSD600299S, RSD600332S, RSD600333S, RSD600338S, RSD600357S, RSD600360S, RSD600383S, RSD600414S, RSD600416S, RSD600434S, RSD600436S, RSD600451S, RSD600552S, RSD600559S, RSD600597S, RSD600606S, RSD601366S, RSD601440S, RSD601559S, RSD601596S, RSD601819S, RSD601859S, RSD601871S, RSD601924S, RSD601925S, RSD602480S, RSD602679S, RSD602918S, RSD603033S, RSD603064S, RSD603153S, RSD603255S, RSD603547S, RSD603868S, RSD604533S, RSD609191S, RSD609194S, RSD609197S, RSD609201S, RSD609192S, RSD609195S, RSD609196S, RSD609198S, RSD609199S, RSD605826S, RSD605827S, RSD606225S, RSD606437S, RSD606438S, RSD606690S, RSD606901S, RSD607027S, RSD607251S, RSD607258S, RSD607346S, RSD607362S, RSD607364S, RSD607756S, RSD608134S, RSD608188S, RSD608189S, RSD608625S, RSD608626S, RSD608688S, RSD608786S, RSD608832S, RSD600532S, RSD601814S, RSD601830S, RSD601833S, RSD601836S, RSD601906S, RSD602123S, RSD602134S, RSD602361S, RSD602451S, RSD602678S, RSD602880S, RSD603230S, RSD603233S, RSD603235S, RSD603236S, RSD603277S, RSD603278S, RSD603299S, RSD603385S, RSD603518S, RSD603587S, RSD603717S, RSD604514S, RSD605069S, RSD605231S, RSD601391S, RSD602236S, RSD604771S, RSD609968S, RSD601569S, RSD606747S, RSD603034S, RSD605292S, RSD605392S, RSD605816S, RSD606063S, RSD606146S, RSD606201S, RSD606441S, RSD606484S, RSD606485S, RSD606743S, RSD606818S, RSD607017S, RSD607091S, RSD607094S, RSD607594S, RSD607602S, RSD608793S, RSD600124S, RSD600184S, RSD600189S, RSD600191S, RSD600194S, RSD600452S, RSD601332S, RSD601334S, RSD601365S, RSD602760S, RSD602761S, RSD603035S, RSD603044S, RSD603404S, RSD603405S, RSD603407S, RSD603914S, RSD603915S, RSD604892S, RSD605891S, RSD607159S, RSD610125S, RSD608071S, RSD608072S, RSD608073S, RSD608074S, RSD608075S, RSD607160S, RSD608516S, RSD608545S, RSD609014S, RSD609015S, RSD610128S, RSD610231S, RSD610854S, RSD600262S, RSD600505S, RSD601983S, RSD602719S, RSD603043S, RSD603346S, RSD603690S, RSD603954S, RSD603955S, RSD610153S, RSD606475S, RSD609149S, RSD601368S, RSD602917S, RSD601853S, RSD609053S, RSD610443S, RSD610474S, RSD610476S, RSD610477S, RSD610952S, RSD610949S, RSD605950S, RSD605281S, RSD606816S, RSD603129S, RSD601844S, RSD607702S, RSD600608S, RSD601426S, RSD602063S, RSD602096S, RSD602588S, RSD602629S, RSD603012S, RSD603037S, RSD603331S, RSD603548S, RSD604517S, RSD603392S, RSD605302S, RSD605905S, RSD606040S, RSD606296S, RSD606914S, RSD607129S, RSD607130S, RSD607158S, RSD607312S, RSD607830S, RSD600324S, RSD600398S, RSD601820S, RSD601841S, RSD602251S, RSD602475S, RSD602755S, RSD602800S, RSD603154S, RSD603182S, RSD603887S, RSD604047S, RSD604741S, RSD605313S, RSD605314S, RSD605315S, RSD605316S, RSD605317S, RSD605885S, RSD605886S, RSD605887S, RSD605888S, RSD605889S, RSD603828S, RSD603842S, RSD603889S, RSD603894S, RSD603895S, RSD603896S, RSD604258S, RSD604259S, RSD604300S, RSD604301S, RSD604303S, RSD604334S, RSD604509S, RSD604510S, RSD604511S, RSD604512S, RSD604513S, RSD604569S, RSD604584S, RSD604585S, RSD604586S, RSD606491S, RSD606493S, RSD606507S, RSD606508S, RSD605909S, RSD602486S, RSD603086S, RSD601568S, RSD605312S, RSD606822S, RSD607055S, RSD607135S, RSD607352S, RSD608182S, RSD608643S, RSD608760S, RSD610032S, RSD610194S, RSD600329S, RSD600379S, RSD602288S, RSD602418S, RSD602494S, RSD602564S, RSD602566S, RSD602613S, RSD602866S, RSD602980S, RSD603253S, RSD603867S, RSD603869S, RSD603870S, RSD604026S, RSD604532S, RSD604772S, RSD604520S, RSD606674S, RSD605340S, RSD606048S, RSD606190S, RSD606191S, RSD606220S, RSD606697S, RSD606938S, RSD607105S, RSD607314S, RSD607317S, RSD607542S, RSD607547S, RSD604575S, RSD604819S, RSD604859S, RSD609281S, RSD609282S, RSD609284S, RSD609279S, RSD609280S, RSD609283S, RSD600218S, RSD600246S, RSD600315S, RSD600365S, RSD602691S, RSD602711S, RSD603191S, RSD603633S, RSD603684S, RSD608144S, RSD606678S, RSD603088S, RSD608427S, RSD608724S, RSD608725S, RSD602011S, RSD602013S, RSD602135S, RSD602136S, RSD602137S, RSD605417S, RSD608564S, RSD609223S, RSD610094S, RSD610452S, RSD609894S, RSD606726S, RSD607156S, RSD607681S, RSD608231S, RSD610273S, RSD600447S, RSD600604S, RSD600605S, RSD601378S, RSD601379S, RSD601380S, RSD602258S, RSD602460S, RSD602730S, RSD610157S, RSD605382S, RSD606057S, RSD606738S, RSD606758S, RSD606994S, RSD607111S, RSD607263S, RSD608275S, RSD601405S, RSD602285S, RSD603418S, RSD603679S, RSD606177S, RSD606615S, RSD610172S, RSD601852S, RSD602427S, RSD603186S, RSD603382S, RSD603637S, RSD603926S, RSD606418S, RSD606448S, RSD606477S, RSD606641S, RSD610524S, RSD608584S, RSD608963S, RSD610190S, RSD600284S, RSD600472S, RSD600582S, RSD600596S, RSD603065S, RSD603066S, RSD603730S, RSD610564S, RSD605805S, RSD602031S, RSD603383S, RSD603674S, RSD605115S, RSD606783S, RSD606784S, RSD606785S, RSD606786S, RSD606790S, RSD606793S, RSD606797S, RSD606798S, RSD606801S, RSD606805S, RSD606806S, RSD606807S, RSD606808S, RSD606833S, RSD606834S, RSD606843S, RSD608748S, RSD608749S, RSD608784S, RSD608798S, RSD609224S, RSD606788S, RSD606835S, RSD608714S, RSD609225S, RSD610870S, RSD610871S, RSD606021S, RSD606022S, RSD606023S, RSD606024S, RSD606026S, RSD606028S, RSD606032S, RSD606033S, RSD606035S, RSD609207S, RSD606479S, RSD608234S, RSD608779S, RSD608808S, RSD603080S, RSD606207S, RSD606766S, RSD608159S, RSD601800S, RSD602491S, RSD603172S, RSD603184S, RSD605815S, RSD608695S, RSD604560S, RSD604884S, RSD606084S, RSD606761S, RSD607599S, RSD605291S, RSD605809S, RSD606163S, RSD607770S, RSD608979S, RSD600267S, RSD601456S, RSD601457S, RSD602067S, RSD607761S, RSD610935S, RSD611144S, RSD607810S, RSD607837S, RSD607838S, RSD607840S, RSD608257S, RSD610055S, RSD610057S, RSD608869S, RSD601415S, RSD601421S, RSD601550S, RSD602287S, RSD603614S, RSD604019S, RSD606221S, RSD607723S, RSD608685S, RSD606913S, RSD607458S, RSD607572S, RSD600394S, RSD600448S, RSD601923S, RSD604761S, RSD606086S, RSD606626S, RSD609042S, RSD603656S, RSD602391S, RSD605811S, RSD605825S, RSD608192S, RSD608702S, RSD608788S, RSD605819S, RSD606202S, RSD607092S, RSD608576S, RSD601897S, RSD606410S, RSD607300S, RSD608269S, RSD610214S, RSD610243S, RSD601916S, RSD610551S, RSD605391S, RSD606442S, RSD606705S, RSD607766S, RSD607769S, RSD603389S, RSD604508S, RSD604959S, RSD605109S, RSD605189S, RSD600511S, RSD602665S, RSD610346S, RSD608172S, RSD607533S, RSD601542S, RSD602499S, RSD602570S, RSD602608S, RSD603015S, RSD603208S, RSD604230S, RSD604239S, RSD604861S, RSD604862S, RSD604864S, RSD606894S, RSD607995S, RSD603762S, RSD610961S, RSD610962S, RSD610964S, RSD603330S, RSD604090S, RSD610790S, RSD600478S, RSD600481S, RSD602699S, RSD605252S, RSD605743S, RSD605876S, RSD605877S, RSD606158S, RSD606290S, RSD608652S, RSD608936S, RSD600555S, RSD602319S, RSD602913S, RSD602984S, RSD603173S, RSD603736S, RSD604060S, RSD604062S, RSD604544S, RSD604856S, RSD611046S, RSD607161S, RSD607581S, RSD600507S, RSD603274S, RSD610211S, RSD602248S, RSD604046S, RSD605949S, RSD607782S, RSD610770S, RSD600387S, RSD600483S, RSD602447S, RSD602517S, RSD603399S, RSD607360S, RSD608207S, RSD608233S, RSD608600S, RSD608692S, RSD609901S, RSD610159S, RSD607133S, RSD601575S, RSD601586S, RSD602271S, RSD602925S, RSD603773S, RSD602215S, RSD603589S, RSD610628S, RSD603743S, RSD604196S, RSD603263S, RSD606424S, RSD606896S, RSD607809S, RSD602206S, RSD602217S, RSD602694S, RSD602937S, RSD603861S, RSD608948S, RSD604740S, RSD606143S, RSD606226S, RSD607592S, RSD602432S, RSD608866S, RSD610222S, RSD601413S, RSD601873S, RSD602092S, RSD602932S, RSD603549S, RSD603964S, RSD610853S, RSD606222S, RSD606634S, RSD603974S, RSD607099S, RSD610782S, RSD605244S, RSD605344S, RSD605352S, RSD605367S, RSD605390S, RSD605476S, RSD605754S, RSD605800S, RSD606097S, RSD606156S, RSD606160S, RSD606200S, RSD606443S, RSD606482S, RSD606640S, RSD606642S, RSD606762S, RSD607327S, RSD607330S, RSD607497S, RSD607498S, RSD607583S, RSD607595S, RSD607771S, RSD607779S, RSD608163S, RSD608305S, RSD608494S, RSD608499S, RSD608721S, RSD608746S, RSD608747S, RSD610528S, RSD610786S, RSD600092S, RSD600256S, RSD600275S, RSD600345S, RSD600482S, RSD600553S, RSD600585S, RSD602108S, RSD602320S, RSD602518S, RSD602595S, RSD602642S, RSD602717S, RSD602720S, RSD602737S, RSD603031S, RSD603324S, RSD603361S, RSD603377S, RSD603419S, RSD603638S, RSD603647S, RSD603710S, RSD603969S, RSD604528S, RSD604558S, RSD604578S, RSD604891S, RSD605114S, RSD602421S, RSD605385S, RSD606694S, RSD607296S, RSD607566S, RSD608640S, RSD610027S, RSD610497S, RSD610498S, RSD602457S, RSD602747S, RSD607524S, RSD606142S, RSD604943S, RSD601839S, RSD603234S, RSD610106S, RSD606778S, RSD607098S, RSD607283S, RSD607326S, RSD607584S, RSD607764S, RSD608214S, RSD608803S, RSD608974S, RSD608975S, RSD608976S, RSD610350S, RSD605780S, RSD600305S, RSD601958S, RSD602465S, RSD602886S, RSD602919S, RSD602942S, RSD603201S, RSD605016S, RSD600257S, RSD607104S, RSD607106S, RSD602461S, RSD602714S, RSD605741S, RSD605742S, RSD605744S, RSD607935S, RSD607936S, RSD607937S, RSD607939S, RSD604223S, RSD604224S, RSD604226S, RSD604869S, RSD604871S, RSD604872S, RSD600449S, RSD600454S, RSD600584S, RSD601447S, RSD601452S, RSD601556S, RSD601560S, RSD602084S, RSD602345S, RSD603661S, RSD610029S, RSD605939S, RSD607284S, RSD606199S, RSD606772S, RSD607097S, RSD607100S, RSD607598S, RSD607601S, RSD607762S, RSD607772S, RSD607773S, RSD608654S, RSD608658S, RSD608717S, RSD608719S, RSD608927S, RSD608928S, RSD608656S, RSD610526S, RSD610639S, RSD610747S, RSD608687S, RSD609895S, RSD608493S, RSD600437S, RSD600456S, RSD603219S, RSD603562S, RSD606157S, RSD606763S, RSD607279S, RSD607328S, RSD608079S, RSD608313S, RSD608585S, RSD608828S, RSD610114S, RSD610221S, RSD600307S, RSD603256S, RSD603601S, RSD604043S, RSD604587S, RSD611062S, RSD605775S, RSD606449S, RSD600263S, RSD601386S, RSD602303S, RSD602314S, RSD602626S, RSD602704S, RSD603855S, RSD604465S, RSD605159S, RSD603820S, RSD604556S, RSD604857S, RSD606504S, RSD607568S, RSD607715S, RSD608621S, RSD600292S, RSD601594S, RSD601893S, RSD602175S, RSD603414S, RSD603099S, RSD603292S, RSD603294S, RSD603355S, RSD605778S, RSD606094S, RSD606951S, RSD606999S, RSD607116S, RSD603880S, RSD603881S, RSD603936S, RSD608237S, RSD608681S, RSD607745S, RSD605361S, RSD605770S, RSD606274S, RSD602245S, RSD606683S, RSD609973S, RSD606295S, RSD606910S, RSD610539S, RSD606076S, RSD603246S, RSD608738S, RSD603957S, RSD604751S, RSD604753S, RSD607749S, RSD608704S, RSD609939S, RSD610103S, RSD610324S, RSD603026S, RSD603724S, RSD604811S, RSD604942S, RSD609213S, RSD609214S, RSD610137S, RSD607113S, RSD604033S, RSD605729S, RSD600286S, RSD600291S, RSD600298S, RSD600301S, RSD600393S, RSD600535S, RSD600598S, RSD602203S, RSD602434S, RSD602472S, RSD602610S, RSD602611S, RSD602700S, RSD602735S, RSD602756S, RSD602921S, RSD602933S, RSD603725S, RSD603848S, RSD603849S, RSD604068S, RSD604471S, RSD604522S, RSD604570S, RSD604889S, RSD606065S, RSD606219S, RSD606474S, RSD606942S, RSD607371S, RSD607685S, RSD609996S, RSD608577S, RSD610096S, RSD602176S, RSD602198S, RSD602658S, RSD603131S, RSD603149S, RSD603150S, RSD603612S, RSD604504S, RSD605387S, RSD606469S, RSD606728S, RSD608147S, RSD609227S, RSD602513S, RSD603176S, RSD604749S, RSD605951S, RSD606450S, RSD604874S, RSD606411S, RSD606451S, RSD606452S, RSD603774S, RSD603778S, RSD605419S, RSD605420S, RSD605422S, RSD605424S, RSD605425S, RSD605426S, RSD605441S, RSD605444S, RSD606245S, RSD606248S, RSD606791S, RSD607274S, RSD607291S, RSD607320S, RSD607519S, RSD608076S, RSD608077S, RSD608078S, RSD608595S, RSD608833S, RSD608954S, RSD604205S, RSD605141S, RSD608937S, RSD610224S, RSD610438S, RSD605377S, RSD605378S, RSD605903S, RSD606647S, RSD606842S, RSD608761S, RSD608822S, RSD608987S, RSD609883S, RSD610212S, RSD600213S, RSD602710S, RSD602908S, RSD603577S, RSD610638S, RSD606172S, RSD608215S, RSD608513S, RSD603617S, RSD609095S, RSD609096S, RSD609098S, RSD609099S, RSD609097S, RSD605409S, RSD606282S, RSD606613S, RSD606825S, RSD606922S, RSD606923S, RSD607123S, RSD607124S, RSD607127S, RSD607128S, RSD608100S, RSD608594S, RSD608881S, RSD600566S, RSD600572S, RSD602093S, RSD602097S, RSD602121S, RSD602197S, RSD602395S, RSD602614S, RSD602874S, RSD602889S, RSD603147S, RSD603381S, RSD603729S, RSD603968S, RSD604535S, RSD604776S, RSD605013S, RSD605102S, RSD605136S, RSD605195S, RSD603349S, RSD605363S, RSD605907S, RSD606920S, RSD608185S, RSD608868S, RSD610019S, RSD610576S, RSD601557S, RSD602723S, RSD602724S, RSD602725S, RSD603058S, RSD603334S, RSD603348S, RSD603949S, RSD604215S, RSD604536S, RSD604752S, RSD604789S, RSD610048S, RSD610836S, RSD610840S, RSD600377S, RSD601367S, RSD610548S, RSD603911S, RSD606926S, RSD607139S, RSD607140S, RSD607659S, RSD607784S, RSD607785S, RSD602797S, RSD607787S, RSD609023S, RSD609024S, RSD609025S, RSD609026S, RSD605910S, RSD605911S, RSD605912S, RSD602309S, RSD602975S, RSD602981S, RSD602982S, RSD602110S, RSD605880S, RSD606361S, RSD607901S, RSD607908S, RSD609236S, RSD609237S, RSD609238S, RSD605360S, RSD606755S, RSD606756S, RSD607120S, RSD602722S, RSD603701S, RSD600391S, RSD602511S, RSD608340S, RSD609721S, RSD604518S, RSD605036S, RSD606083S, RSD606259S, RSD606467S, RSD606614S, RSD607153S, RSD607670S, RSD610271S, RSD602182S, RSD602801S, RSD602804S, RSD602869S, RSD602907S, RSD603000S, RSD603145S, RSD603356S, RSD603641S, RSD603704S, RSD603728S, RSD603865S, RSD604066S, RSD604111S, RSD604113S, RSD604118S, RSD604885S, RSD608087S, RSD608646S, RSD608734S, RSD600198S, RSD601909S, RSD602481S, RSD610360S, RSD602423S, RSD607597S, RSD600558S, RSD602085S, RSD603970S, RSD606205S, RSD606481S, RSD608649S, RSD608844S, RSD608967S, RSD602453S, RSD603057S, RSD603103S, RSD603124S, RSD603283S, RSD604063S, RSD604064S, RSD606283S, RSD606918S, RSD610850S, RSD600203S, RSD600399S, RSD602094S, RSD603746S, RSD604779S, RSD605196S, RSD605394S, RSD606372S, RSD606749S, RSD607288S, RSD607525S, RSD608222S, RSD608589S, RSD608983S, RSD609882S, RSD610236S, RSD610453S, RSD608110S, RSD608112S, RSD608113S, RSD608114S, RSD608115S, RSD606183S, RSD606197S, RSD606496S, RSD606637S, RSD606902S, RSD607090S, RSD607591S, RSD607757S, RSD608834S, RSD610554S, RSD601987S, RSD602138S, RSD602195S, RSD602317S, RSD602318S, RSD602424S, RSD602425S, RSD602428S, RSD603030S, RSD603703S, RSD603719S, RSD603834S, RSD603857S, RSD604886S, RSD605219S, RSD610920S, RSD606223S, RSD607019S, RSD607359S, RSD608187S, RSD608274S, RSD609978S, RSD609829S, RSD609830S, RSD609831S, RSD604218S, RSD604225S, RSD606059S, RSD606173S, RSD606278S, RSD606727S, RSD606937S, RSD607299S, RSD607454S, RSD607736S, RSD608637S, RSD608929S, RSD608971S, RSD610018S, RSD610284S, RSD610295S, RSD610574S, RSD610666S, RSD600545S, RSD602781S, RSD603051S, RSD604525S, RSD604883S, RSD610665S, RSD610875S, RSD611026S, RSD606611S, RSD608737S, RSD608758S, RSD608863S, RSD600235S, RSD602355S, RSD605132S, RSD607064S, RSD607065S, RSD607066S, RSD607298S, RSD607395S, RSD607567S, RSD608578S, RSD608641S, RSD610026S, RSD610127S, RSD610307S, RSD610309S, RSD610509S, RSD610502S, RSD610513S, RSD610310S, RSD605783S, RSD606137S, RSD606725S, RSD607535S, RSD607678S, RSD608762S, RSD608764S, RSD610454S, RSD610455S, RSD601377S, RSD602281S, RSD602388S, RSD602410S, RSD603546S, RSD603875S, RSD604515S, RSD605179S, RSD606506S, RSD608549S, RSD608700S, RSD610461S, RSD610562S, RSD610884S, RSD607087S, RSD607754S, RSD607313S, RSD608744S, RSD608706S, RSD607345S, RSD607579S, RSD608597S, RSD601390S, RSD603962S, RSD606494S, RSD610939S, RSD608858S, RSD610523S, RSD610938S, RSD604775S, RSD607141S, RSD606151S, RSD610545S, RSD606056S, RSD606367S, RSD606741S, RSD600529S, RSD601931S, RSD602106S, RSD602905S, RSD603097S, RSD603619S, RSD603734S, RSD607289S, RSD605018S, RSD602733S, RSD602785S, RSD602149S, RSD604058S, RSD610043S, RSD610042S, RSD606144S, RSD606416S, RSD606689S, RSD608530S, RSD608596S, RSD600232S, RSD600344S, RSD601425S, RSD602533S, RSD603420S, RSD604016S, RSD605320S, RSD605375S, RSD605781S, RSD606263S, RSD600430S, RSD600539S, RSD601359S, RSD602091S, RSD602095S, RSD602692S, RSD604954S, RSD603100S, RSD603873S, RSD606810S, RSD606811S, RSD609100S, RSD609101S, RSD605480S, RSD606060S, RSD607539S, RSD600590S, RSD601863S, RSD601865S, RSD601891S, RSD602433S, RSD602512S, RSD602594S, RSD602752S, RSD602782S, RSD602788S, RSD603567S, RSD603618S, RSD603625S, RSD603629S, RSD603758S, RSD604860S, RSD605045S, RSD607108S, RSD607109S, RSD608771S, RSD610449S, RSD600586S, RSD600588S, RSD600589S, RSD602604S, RSD602687S, RSD602890S, RSD603045S, RSD603046S, RSD603265S, RSD603336S, RSD603342S, RSD603749S, RSD604202S, RSD611039S, RSD611056S, RSD604524S, RSD608528S, RSD603041S, RSD605199S, RSD606277S, RSD606630S, RSD607522S, RSD608224S, RSD606230S, RSD606433S, RSD606434S, RSD606681S, RSD607469S, RSD608132S, RSD608605S, RSD609694S, RSD609950S, RSD610092S, RSD610178S, RSD610232S, RSD602209S, RSD602634S, RSD603020S, RSD605358S, RSD608591S, RSD604802S, RSD606210S, RSD608092S, RSD610447S, RSD610876S, RSD600201S, RSD600435S, RSD601914S, RSD601985S, RSD603391S, RSD606176S, RSD607752S, RSD604927S, RSD605301S, RSD605410S, RSD606612S, RSD606967S, RSD607054S, RSD607309S, RSD608179S, RSD608272S, RSD603171S, RSD603360S, RSD603929S, RSD603948S, RSD604901S, RSD606011S, RSD605894S, RSD608873S, RSD610068S, RSD600224S, RSD602344S, RSD603068S, RSD605916S, RSD607747S, RSD607748S, RSD602220S, RSD601361S, RSD602471S, RSD606073S, RSD609956S, RSD606954S, RSD608935S, RSD603702S, RSD604554S, RSD604747S, RSD608705S, RSD601373S, RSD605167S, RSD606285S, RSD610610S, RSD610612S, RSD605942S, RSD605943S, RSD605944S, RSD605945S, RSD605946S, RSD605947S, RSD607623S, RSD607624S, RSD608167S, RSD608168S, RSD608169S, RSD608171S, RSD609045S, RSD610071S, RSD610072S, RSD610608S, RSD610613S, RSD605463S, RSD605794S, RSD606145S, RSD606181S, RSD607144S, RSD607342S, RSD607727S, RSD607734S, RSD608871S, RSD610030S, RSD610421S, RSD602504S, RSD603747S, RSD603748S, RSD604807S, RSD605177S, RSD607726S, RSD610422S, RSD610611S, RSD608223S, RSD600570S, RSD603335S, RSD610626S, RSD606930S, RSD606932S, RSD606933S, RSD606984S, RSD607729S, RSD607732S, RSD607820S, RSD607821S, RSD607822S, RSD606141S, RSD606262S, RSD606422S, RSD606423S, RSD606470S, RSD606676S, RSD607302S, RSD608552S, RSD608553S, RSD610087S, RSD610089S, RSD600443S, RSD601880S, RSD602146S, RSD602196S, RSD602262S, RSD602396S, RSD602972S, RSD602986S, RSD603077S, RSD603083S, RSD603658S, RSD603668S, RSD603671S, RSD603732S, RSD603937S, RSD604571S, RSD604956S, RSD605183S, RSD605234S, RSD610088S, RSD610090S, RSD610889S, RSD601336S, RSD605786S, RSD606070S, RSD606152S, RSD606631S, RSD606692S, RSD606693S, RSD606730S, RSD607067S, RSD607068S, RSD607145S, RSD607538S, RSD608505S, RSD608642S, RSD601591S, RSD602592S, RSD602624S, RSD602662S, RSD603055S, RSD603305S, RSD603306S, RSD603576S, RSD610872S, RSD611063S, RSD610623S, RSD607126S, RSD601564S, RSD601850S, RSD602617S, RSD605393S, RSD606648S, RSD603190S, RSD608918S, RSD604770S, RSD609055S, RSD602576S, RSD602579S, RSD603091S, RSD603093S, RSD604549S, RSD604552S, RSD605818S, RSD606731S, RSD606991S, RSD607562S, RSD610095S, RSD602051S, RSD602414S, RSD602416S, RSD602712S, RSD603287S, RSD603384S, RSD603395S, RSD603415S, RSD603913S, RSD605206S, RSD611149S, RSD607318S, RSD602404S, RSD601988S, RSD603054S, RSD606895S, RSD610363S, RSD610368S, RSD610386S, RSD610532S, RSD600346S, RSD602242S, RSD602941S, RSD603388S, RSD610530S, RSD610533S, RSD608648S, RSD600600S, RSD601576S, RSD602589S, RSD602935S, RSD607998S, RSD607999S, RSD608000S, RSD608001S, RSD608002S, RSD608003S, RSD608004S, RSD609233S, RSD609234S, RSD609235S, RSD602605S, RSD602669S, RSD602673S, RSD602676S, RSD602685S, RSD603678S, RSD603680S, RSD603683S, RSD603686S, RSD603687S, RSD603688S, RSD603689S, RSD603697S, RSD603699S, RSD603700S, RSD604227S, RSD604228S, RSD604235S, RSD604241S, RSD604244S, RSD604247S, RSD604251S, RSD609228S, RSD609229S, RSD609230S, RSD609231S, RSD609232S, RSD605764S, RSD605766S, RSD606629S, RSD607115S, RSD607264S, RSD607265S, RSD607268S, RSD607569S, RSD602256S, RSD602744S, RSD603535S, RSD607142S, RSD601956S, RSD602701S, RSD603243S, RSD603279S, RSD603522S, RSD604014S, RSD608804S, RSD606150S, RSD606180S, RSD604038S, RSD607350S, RSD607351S, RSD608904S, RSD608956S, RSD608957S, RSD602012S, RSD602358S, RSD602359S, RSD604297S, RSD604299S, RSD604719S, RSD604723S, RSD604726S, RSD605121S, RSD607349S, RSD608408S, RSD608581S, RSD608903S, RSD606644S, RSD606718S, RSD604059S, RSD600098S, RSD606706S, RSD602266S, RSD606287S, RSD605296S, RSD606169S, RSD606739S, RSD606799S, RSD607629S, RSD608796S, RSD610200S, RSD610343S, RSD600592S, RSD601335S, RSD602377S, RSD602381S, RSD602702S, RSD603558S, RSD603920S, RSD604958S, RSD610932S, RSD605793S, RSD605804S, RSD606149S, RSD606649S, RSD606804S, RSD606921S, RSD607286S, RSD608210S, RSD608333S, RSD608506S, RSD608636S, RSD608805S, RSD608949S, RSD600367S, RSD601563S, RSD602101S, RSD602394S, RSD603006S, RSD603315S, RSD603599S, RSD611019S, RSD610201S, RSD606074S, RSD608127S, RSD607778S, RSD610527S, RSD608777S, RSD608813S, RSD609089S, RSD607563S, RSD610863S, RSD608806S, RSD610014S, RSD606812S, RSD610519S, RSD607608S, RSD608277S, RSD608534S, RSD600450S, RSD601548S, RSD603756S, RSD605984S, RSD607621S, RSD607731S, RSD602583S, RSD611010S, RSD608933S, RSD608934S, RSD610174S, RSD610175S, RSD610315S, RSD608091S, RSD602155S, RSD603156S, RSD608265S, RSD610912S, RSD609027S, RSD607357S, RSD602904S, RSD604055S, RSD607393S, RSD610397S, RSD605188S, RSD608830S, RSD606665S, RSD607152S, RSD608689S, RSD608988S, RSD602887S, RSD606989S, RSD600186S, RSD600279S, RSD600280S, RSD601382S, RSD601384S, RSD601385S, RSD601387S, RSD601401S, RSD601566S, RSD601892S, RSD601921S, RSD601967S, RSD601968S, RSD601975S, RSD602227S, RSD602569S, RSD602695S, RSD603166S, RSD603249S, RSD603976S, RSD605892S, RSD605904S, RSD606286S, RSD606824S, RSD606831S, RSD606908S, RSD607118S, RSD607119S, RSD607574S, RSD607578S, RSD608153S, RSD608154S, RSD608156S, RSD608164S, RSD608165S, RSD608256S, RSD608258S, RSD608259S, RSD608527S, RSD608592S, RSD610031S, RSD610051S, RSD610052S, RSD610423S, RSD610424S, RSD610425S, RSD610062S, RSD610070S, RSD610802S, RSD610830S, RSD611017S, RSD609692S, RSD609693S, RSD600322S, RSD600350S, RSD600439S, RSD602315S, RSD603934S, RSD610913S, RSD606486S, RSD608247S, RSD608657S, RSD602052S, RSD603049S, RSD603623S, RSD603624S, RSD603841S, RSD605300S, RSD607575S, RSD607841S, RSD607921S, RSD608151S, RSD608546S, RSD610138S, RSD610140S, RSD603270S, RSD601847S, RSD602593S, RSD603347S, RSD603752S, RSD605740S, RSD605948S, RSD603897S, RSD604307S, RSD605288S, RSD605294S, RSD605295S, RSD605350S, RSD605380S, RSD605475S, RSD606148S, RSD606203S, RSD606980S, RSD607076S, RSD607275S, RSD608218S, RSD608635S, RSD608740S, RSD608945S, RSD600337S, RSD600362S, RSD600488S, RSD600490S, RSD600527S, RSD602231S, RSD602340S, RSD602351S, RSD602375S, RSD602635S, RSD603052S, RSD603082S, RSD603202S, RSD603286S, RSD603357S, RSD603378S, RSD603545S, RSD604040S, RSD608633S, RSD606061S, RSD607325S, RSD608213S, RSD608497S, RSD609929S, RSD608701S, RSD606147S, RSD606936S, RSD607520S, RSD605168S, RSD606184S, RSD605107S, RSD608216S, RSD606052S, RSD606492S, RSD606670S, RSD606671S, RSD607143S, RSD607536S, RSD608598S, RSD608782S, RSD610156S, RSD601381S, RSD602144S, RSD602520S, RSD602690S, RSD602738S, RSD602909S, RSD603161S, RSD603174S, RSD603254S, RSD603393S, RSD603550S, RSD603579S, RSD603833S, RSD604795S, RSD605054S, RSD600384S, RSD603204S, RSD603319S, RSD608984S, RSD601442S, RSD603133S, RSD608058S, RSD600594S, RSD601856S, RSD602250S, RSD607948S, RSD605374S, RSD605784S, RSD606471S, RSD606675S, RSD601797S, RSD602775S, RSD603332S, RSD603524S, RSD603613S, RSD604523S, RSD604946S, RSD604947S, RSD607210S, RSD607576S, RSD608573S, RSD608756S, RSD600317S, RSD602399S, RSD603042S, RSD603859S, RSD604531S, RSD610428S, RSD606264S, RSD606617S, RSD606679S, RSD606993S, RSD607077S, RSD608208S, RSD608693S, RSD608699S, RSD608763S, RSD609903S, RSD609904S, RSD610158S, RSD610456S, RSD610566S, RSD610567S, RSD601939S, RSD601981S, RSD602064S, RSD602199S, RSD602205S, RSD602292S, RSD602466S, RSD602971S, RSD602977S, RSD603285S, RSD603733S, RSD610660S, RSD606664S, RSD605782S, RSD606080S, RSD606081S, RSD606211S, RSD606616S, RSD606992S, RSD607680S, RSD608232S, RSD608264S, RSD608464S, RSD608489S, RSD608599S, RSD610457S, RSD610893S, RSD601358S, RSD601567S, RSD602385S, RSD602547S, RSD602732S, RSD602906S, RSD603195S, RSD603281S, RSD603343S, RSD603371S, RSD603608S, RSD603713S, RSD603935S, RSD604812S, RSD604888S, RSD610890S, RSD611075S, RSD608212S, RSD607628S, RSD607876S, RSD604199S, RSD604200S, RSD605151S, RSD605152S, RSD605155S, RSD609050S, RSD605788S, RSD605789S, RSD605791S, RSD606170S, RSD606171S, RSD606652S, RSD607045S, RSD607074S, RSD607075S, RSD607278S, RSD607605S, RSD607627S, RSD608797S, RSD608812S, RSD600240S, RSD600271S, RSD600318S, RSD600375S, RSD600376S, RSD600415S, RSD600420S, RSD600431S, RSD600432S, RSD600444S, RSD600455S, RSD600544S, RSD600546S, RSD600578S, RSD602117S, RSD602143S, RSD602291S, RSD602519S, RSD602759S, RSD602765S, RSD602771S, RSD603667S, RSD603685S, RSD603874S, RSD603922S, RSD603923S, RSD604039S, RSD604201S, RSD604526S, RSD604846S, RSD604848S, RSD605112S, RSD608800S, RSD610345S, RSD611130S, RSD603860S, RSD606066S, RSD606698S, RSD606941S, RSD607543S, RSD607546S, RSD608138S, RSD609907S, RSD605453S, RSD605776S, RSD606473S, RSD607374S, RSD607690S, RSD608766S, RSD608815S, RSD602141S, RSD602934S, RSD604024S, RSD604816S, RSD606154S, RSD607746S, RSD609599S, RSD610881S, RSD605334S, RSD600427S, RSD600519S, RSD602243S, RSD607942S, RSD607943S, RSD606209S, RSD607285S, RSD608511S, RSD608579S, RSD608791S, RSD608981S, RSD610076S, RSD610077S, RSD610205S, RSD611034S, RSD603280S, RSD603727S, RSD603739S, RSD603770S, RSD604080S, RSD604183S, RSD604221S, RSD604222S, RSD604260S, RSD604311S, RSD606839S, RSD600520S, RSD600517S, RSD600518S, RSD601861S, RSD610897S, RSD610458S, RSD605372S, RSD605683S, RSD605785S, RSD606292S, RSD606447S, RSD606468S, RSD606498S, RSD606627S, RSD606682S, RSD607001S, RSD607303S, RSD607361S, RSD607534S, RSD608145S, RSD608235S, RSD608554S, RSD608603S, RSD608604S, RSD608696S, RSD608697S, RSD608770S, RSD609960S, RSD609963S, RSD610091S, RSD610160S, RSD610161S, RSD610253S, RSD610255S, RSD610459S, RSD610571S, RSD610572S, RSD610573S, RSD610899S, RSD600205S, RSD600228S, RSD600229S, RSD600236S, RSD600254S, RSD600258S, RSD600260S, RSD600328S, RSD600335S, RSD600388S, RSD600423S, RSD600493S, RSD600509S, RSD600533S, RSD600580S, RSD600612S, RSD601572S, RSD601946S, RSD601963S, RSD601969S, RSD602200S, RSD602229S, RSD602241S, RSD602371S, RSD602398S, RSD602751S, RSD602768S, RSD602769S, RSD602770S, RSD602882S, RSD602884S, RSD602899S, RSD603958S, RSD604573S, RSD604890S, RSD604940S, RSD610901S, RSD610902S, RSD611071S, RSD606789S, RSD606814S, RSD610111S, RSD608518S, RSD602090S, RSD602597S, RSD602916S, RSD603002S, RSD608850S, RSD601855S, RSD602492S, RSD603218S, RSD607994S, RSD602073S, RSD609182S, RSD609184S, RSD605343S, RSD606487S, RSD606748S, RSD606768S, RSD606800S, RSD607101S, RSD608653S, RSD608715S, RSD608973S, RSD600373S, RSD600407S, RSD600408S, RSD600409S, RSD600440S, RSD600468S, RSD600470S, RSD600471S, RSD600528S, RSD600549S, RSD600599S, RSD601846S, RSD602098S, RSD602104S, RSD602107S, RSD602310S, RSD602321S, RSD602369S, RSD603692S, RSD607085S, RSD603159S, RSD605908S, RSD607162S, RSD604750S, RSD609816S, RSD610396S, RSD608139S, RSD606187S, RSD608732S, RSD610109S, RSD611007S, RSD608590S, RSD600494S, RSD602305S, RSD607023S, RSD607944S, RSD607945S, RSD609887S, RSD609888S, RSD606940S, RSD607541S, RSD607687S, RSD607691S, RSD608198S, RSD608722S, RSD608765S, RSD604023S, RSD608140S, RSD605341S, RSD606208S, RSD606440S, RSD606623S, RSD610209S, RSD610210S, RSD602043S, RSD602061S, RSD602111S, RSD602367S, RSD602454S, RSD602663S, RSD602784S, RSD602915S, RSD603007S, RSD603032S, RSD603337S, RSD603353S, RSD603400S, RSD603402S, RSD603636S, RSD603930S, RSD603932S, RSD604506S, RSD610529S, RSD605345S, RSD606162S, RSD606985S, RSD606986S, RSD607047S, RSD607051S, RSD607397S, RSD607760S, RSD608158S, RSD608227S, RSD608289S, RSD608587S, RSD608711S, RSD608843S, RSD608952S, RSD610021S, RSD610112S, RSD610587S, RSD600208S, RSD600290S, RSD600336S, RSD600404S, RSD600506S, RSD600530S, RSD601579S, RSD601858S, RSD601941S, RSD602035S, RSD602038S, RSD602065S, RSD602379S, RSD602468S, RSD602562S, RSD603158S, RSD603160S, RSD603209S, RSD603942S, RSD609204S, RSD609206S, RSD608230S, RSD608531S, RSD602607S, RSD604731S, RSD606196S, RSD606691S, RSD607149S, RSD607741S, RSD608801S, RSD608865S, RSD610877S, RSD605940S, RSD605941S, RSD607593S, RSD600202S, RSD600273S, RSD601573S, RSD602343S, RSD610182S, RSD610517S, RSD602357S, RSD601626S, RSD605823S, RSD606288S, RSD606903S, RSD607089S, RSD608199S, RSD608631S, RSD608962S, RSD610011S, RSD610189S, RSD600515S, RSD601587S, RSD601961S, RSD602062S, RSD602446S, RSD604840S, RSD604957S, RSD608835S, RSD610921S, RSD611124S, RSD611131S, RSD600137S, RSD606944S, RSD607006S, RSD607171S, RSD610101S, RSD600272S, RSD607733S, RSD606945S, RSD607742S, RSD610468S, RSD604231S, RSD607589S, RSD608778S, RSD608912S, RSD607220S, RSD607587S, RSD608152S, RSD608574S, RSD606417S, RSD602443S, RSD606049S, RSD606082S, RSD606261S, RSD610656S, RSD602201S, RSD602439S, RSD604928S, RSD608819S, RSD608900S, RSD601949S, RSD608978S, RSD608498S, RSD606820S, RSD607920S, RSD601577S, RSD602140S, RSD602347S, RSD604736S, RSD602363S, RSD602365S, RSD602878S, RSD607122S, RSD608852S, RSD608862S, RSD610075S, RSD610208S, RSD610074S, RSD607221S, RSD600312S, RSD602257S, RSD604791S, RSD605194S, RSD606911S, RSD607137S, RSD603862S, RSD603952S, RSD604960S, RSD607386S, RSD606430S, RSD602739S, RSD602766S, RSD603079S, RSD603128S, RSD603417S, RSD600320S, RSD600321S, RSD600325S, RSD605967S, RSD605969S, RSD605971S, RSD605972S, RSD605973S, RSD605975S, RSD605978S, RSD605980S, RSD605981S, RSD605982S, RSD605983S, RSD603580S, RSD603581S, RSD604001S, RSD604002S, RSD604003S, RSD604004S, RSD604005S, RSD604580S, RSD604582S, RSD604583S, RSD604708S, RSD604711S, RSD604712S, RSD605962S, RSD605963S, RSD605964S, RSD605965S, RSD605966S, RSD605968S, RSD605970S, RSD605974S, RSD605976S, RSD605977S, RSD605979S, RSD609594S, RSD609595S, RSD609596S, RSD609597S, RSD609598S, RSD602779S, RSD602865S, RSD601543S, RSD606830S, RSD606912S, RSD607854S, RSD610012S, RSD610141S, RSD601940S, RSD602879S, RSD602973S, RSD604007S, RSD610230S, RSD602056S, RSD602057S, RSD602060S, RSD609819S, RSD609886S, RSD606155S, RSD607294S, RSD607560S, RSD607132S, RSD608123S, RSD605347S, RSD605479S, RSD605879S, RSD608122S, RSD608161S, RSD608217S, RSD608966S, RSD600541S, RSD600543S, RSD600547S, RSD600548S, RSD601908S, RSD602109S, RSD602139S, RSD602202S, RSD602514S, RSD602726S, RSD602736S, RSD603060S, RSD603341S, RSD603634S, RSD603776S, RSD603944S, RSD604061S, RSD605954S, RSD606085S, RSD607250S, RSD608731S, RSD601887S, RSD608148S, RSD609932S, RSD602400S, RSD602591S, RSD603198S, RSD603575S, RSD608061S, RSD608062S, RSD608063S, RSD608064S, RSD608065S, RSD608066S, RSD609836S, RSD609837S, RSD609838S, RSD609839S, RSD606632S, RSD607239S, RSD608246S, RSD608638S, RSD608707S, RSD609877S, RSD609878S, RSD609880S, RSD610514S, RSD601438S, RSD609881S, RSD609723S, RSD609724S, RSD607262S, RSD608847S, RSD609051S, RSD610667S, RSD611080S, RSD607946S, RSD607947S, RSD604212S, RSD606666S, RSD606667S, RSD609271S, RSD609272S, RSD609274S, RSD600343S, RSD601972S, RSD608276S, RSD608875S, RSD605306S, RSD605898S, RSD605899S, RSD607180S, RSD607181S, RSD608678S, RSD608742S, RSD601978S, RSD602740S, RSD602764S, RSD602987S, RSD603422S, RSD603586S, RSD604542S, RSD604792S, RSD605200S, RSD605201S, RSD610079S, RSD606018S, RSD606019S, RSD607980S, RSD607981S, RSD609189S, RSD609190S, RSD605407S, RSD605451S, RSD605893S, RSD605895S, RSD606906S, RSD606907S, RSD607204S, RSD607455S, RSD607573S, RSD607655S, RSD607663S, RSD607664S, RSD608180S, RSD608181S, RSD608755S, RSD601903S, RSD602525S, RSD602871S, RSD602873S, RSD602988S, RSD603038S, RSD603089S, RSD603162S, RSD603411S, RSD603716S, RSD603723S, RSD603941S, RSD604476S, RSD604803S, RSD604920S, RSD605768S, RSD606054S, RSD606091S, RSD608273S, RSD608542S, RSD608774S, RSD600253S, RSD602087S, RSD602129S, RSD602192S, RSD602783S, RSD602929S, RSD603847S, RSD609041S, RSD605383S, RSD609994S, RSD602322S, RSD607737S, RSD601553S, RSD604841S, RSD607689S, RSD602426S, RSD610124S, RSD609817S, RSD609818S, RSD610597S, RSD605242S, RSD605928S, RSD606178S, RSD606628S, RSD606684S, RSD606724S, RSD608236S, RSD608698S, RSD608768S, RSD609971S, RSD600251S, RSD601953S, RSD601979S, RSD601980S, RSD602452S, RSD603009S, RSD603864S, RSD604042S, RSD604507S, RSD604679S, RSD604768S, RSD605178S, RSD603627S, RSD607719S, RSD610731S, RSD603408S, RSD605900S, RSD605906S, RSD605379S, RSD609176S, RSD609166S, RSD609174S, RSD606267S, RSD606269S, RSD606699S, RSD607483S, RSD608137S, RSD608195S, RSD608558S, RSD608726S, RSD601441S, RSD601578S, RSD601971S, RSD602180S, RSD602356S, RSD602501S, RSD602505S, RSD602716S, RSD602753S, RSD602881S, RSD602912S, RSD602920S, RSD602966S, RSD602995S, RSD603130S, RSD603146S, RSD603640S, RSD603927S, RSD603963S, RSD604818S, RSD605211S, RSD602683S, RSD607917S, RSD610020S, RSD601388S, RSD602249S, RSD602968S, RSD600480S, RSD608270S, RSD601938S, RSD602852S, RSD601369S, RSD603210S, RSD605446S, RSD605448S, RSD605449S, RSD605952S, RSD605953S, RSD605955S, RSD607548S, RSD607549S, RSD607551S, RSD607552S, RSD607553S, RSD607949S, RSD607950S, RSD607951S, RSD608897S, RSD608898S, RSD609059S, RSD609060S, RSD609061S, RSD609062S, RSD609064S, RSD610168S, RSD604252S, RSD604253S, RSD604302S, RSD604315S, RSD604474S, RSD604563S, RSD604564S, RSD604764S, RSD604765S, RSD608955S, RSD609063S, RSD610167S, RSD610436S, RSD610437S, RSD606428S, RSD606609S, RSD606826S, RSD606828S, RSD606829S, RSD600510S, RSD602734S, RSD602786S, RSD603067S, RSD603262S, RSD603267S, RSD603600S, RSD603606S, RSD603891S, RSD603899S, RSD604048S, RSD611105S, RSD611107S, RSD600512S, RSD600241S, RSD608226S, RSD608743S, RSD600230S, RSD602476S, RSD603061S, RSD608824S, RSD606212S, RSD606495S, RSD606817S, RSD608816S, RSD610450S, RSD603340S, RSD605695S, RSD607540S, RSD610215S, RSD604015S, RSD605127S, RSD606431S, RSD604457S, RSD608104S, RSD608105S, RSD608799S, RSD608802S, RSD605324S, RSD605792S, RSD606375S, RSD607240S, RSD607831S, RSD600410S, RSD600418S, RSD600453S, RSD600556S, RSD600583S, RSD601443S, RSD601445S, RSD602715S, RSD603050S, RSD603293S, RSD603352S, RSD604472S, RSD604521S, RSD605209S, RSD608709S, RSD607527S, RSD607730S, RSD606472S, RSD606962S, RSD607528S, RSD607728S, RSD608504S, RSD608571S, RSD608789S, RSD610312S, RSD610868S, RSD605797S, RSD605995S, RSD606406S, RSD606636S, RSD607323S, RSD600293S, RSD600313S, RSD600561S, RSD603711S, RSD600603S, RSD601829S, RSD601977S, RSD602304S, RSD602459S, RSD602883S, RSD603078S, RSD603170S, RSD603175S, RSD605305S, RSD602718S, RSD608094S, RSD608095S, RSD602301S, RSD602616S, RSD600611S, RSD602080S, RSD602374S, RSD602455S, RSD608194S, RSD610960S, RSD606740S, RSD608543S, RSD603359S, RSD606863S, RSD605878S, RSD606206S, RSD606645S, RSD606717S, RSD606987S, RSD606988S, RSD608160S, RSD608162S, RSD608870S, RSD610022S, RSD600308S, RSD600557S, RSD600567S, RSD603401S, RSD603691S, RSD604852S, RSD610024S, RSD608575S, RSD610192S, RSD601889S, RSD607803S, RSD608712S, RSD609275S, RSD605723S, RSD605883S, RSD605884S, RSD606509S, RSD607347S, RSD608347S, RSD602927S, RSD602928S, RSD603313S, RSD603317S, RSD603578S, RSD604088S, RSD604197S, RSD604308S, RSD604337S, RSD605140S, RSD605881S, RSD605882S, RSD607666S, RSD609215S, RSD609216S, RSD600369S, RSD605011S, RSD606821S, RSD608083S, RSD606435S, RSD600601S, RSD601364S, RSD601460S, RSD601911S, RSD601922S, RSD601973S, RSD603665S, RSD608086S, RSD608084S, RSD607531S, RSD607316S, RSD606965S, RSD605376S, RSD606488S, RSD606633S, RSD606635S, RSD606686S, RSD606687S, RSD607084S, RSD607172S, RSD608191S, RSD608538S, RSD608601S, RSD608891S, RSD600195S, RSD600226S, RSD600264S, RSD600309S, RSD600339S, RSD600340S, RSD600342S, RSD600400S, RSD600425S, RSD600457S, RSD600534S, RSD601370S, RSD601402S, RSD601458S, RSD601932S, RSD601950S, RSD601951S, RSD601974S, RSD601976S, RSD602081S, RSD602083S, RSD602207S, RSD602240S, RSD602246S, RSD602313S, RSD602408S, RSD602478S, RSD602489S, RSD602707S, RSD602798S, RSD602799S, RSD604025S, RSD604572S, RSD605207S, RSD605208S, RSD605820S, RSD605822S, RSD605985S, RSD605987S, RSD605988S, RSD606068S, RSD606257S, RSD606284S, RSD606654S, RSD606655S, RSD606656S, RSD606658S, RSD606659S, RSD606660S, RSD606661S, RSD606662S, RSD606663S, RSD607781S, RSD607816S, RSD607818S, RSD607869S, RSD607871S, RSD607872S, RSD608992S, RSD602329S, RSD602330S, RSD602331S, RSD602332S, RSD602334S, RSD602335S, RSD602337S, RSD602339S, RSD603364S, RSD603368S, RSD603369S, RSD603370S, RSD603372S, RSD603373S, RSD603374S, RSD603375S, RSD603376S, RSD603379S, RSD605145S, RSD605146S, RSD605147S, RSD605148S, RSD605149S, RSD605150S, RSD607874S, RSD608993S, RSD608994S, RSD609001S, RSD609002S, RSD609003S, RSD609004S, RSD610142S, RSD605450S, RSD607164S, RSD607661S, RSD608593S, RSD608874S, RSD602940S, RSD603351S, RSD603742S, RSD604793S, RSD610143S, RSD605336S, RSD605773S, RSD606193S, RSD606624S, RSD606996S, RSD606998S, RSD607233S, RSD609928S, RSD600514S, RSD600521S, RSD601948S, RSD602338S, RSD602688S, RSD602901S, RSD603048S, RSD603053S, RSD603056S, RSD603169S, RSD604831S, RSD605162S, RSD605212S, RSD609926S, RSD605338S, RSD605381S, RSD606087S, RSD608814S, RSD610186S, RSD604051S, RSD605335S, RSD605772S, RSD606192S, RSD606700S, RSD607696S, RSD608201S, RSD608202S, RSD608452S, RSD608727S, RSD608729S, RSD609927S, RSD600417S, RSD600467S, RSD601545S, RSD602294S, RSD602405S, RSD602622S, RSD602689S, RSD603028S, RSD603272S, RSD603295S, RSD603302S, RSD604576S, RSD604827S, RSD605812S, RSD607590S, RSD602429S, RSD602767S, RSD605051S, RSD608089S, RSD608090S, RSD606651S, RSD600374S, RSD606058S, RSD607751S, RSD609671S, RSD609672S, RSD609673S, RSD609674S, RSD609675S, RSD609676S, RSD609678S, RSD609679S, RSD609680S, RSD609682S, RSD609683S, RSD609684S, RSD609685S, RSD609686S, RSD609689S, RSD609691S, RSD610967S, RSD610970S, RSD610971S, RSD610972S, RSD610975S, RSD610976S, RSD610977S, RSD610979S, RSD610980S, RSD610981S, RSD610982S, RSD610984S, RSD606935S, RSD601581S, RSD606227S, RSD606721S, RSD607053S, RSD600249S, RSD600296S, RSD611169S, RSD610446S, RSD600487S, RSD600575S, RSD601860S, RSD601912S, RSD602596S, RSD603010S, RSD603013S, RSD603231S, RSD605256S, RSD600268S, RSD600269S, RSD603380S, RSD603616S, RSD603655S, RSD604010S, RSD604012S, RSD604045S, RSD605901S, RSD608572S, RSD608745S, RSD608757S, RSD600341S, RSD600403S, RSD600428S, RSD601459S, RSD601791S, RSD601870S, RSD602272S, RSD602619S, RSD603916S, RSD601451S, RSD601826S, RSD602306S, RSD608842S, RSD601876S, RSD606217S, RSD608776S, RSD603290S, RSD604089S, RSD604680S, RSD608093S, RSD608628S, RSD602333S, RSD601416S, RSD601429S, RSD601869S, RSD605389S, RSD605760S, RSD605801S, RSD605807S, RSD606164S, RSD606166S, RSD606167S, RSD606444S, RSD606744S, RSD606777S, RSD607277S, RSD607390S, RSD607765S, RSD607777S, RSD607780S, RSD608220S, RSD608716S, RSD608932S, RSD600015S, RSD600108S, RSD600237S, RSD600238S, RSD600245S, RSD600247S, RSD600389S, RSD600438S, RSD601868S, RSD602078S, RSD602709S, RSD603593S, RSD603910S, RSD604529S, RSD604579S, RSD605113S, RSD608502S, RSD608655S, RSD608713S, RSD609931S, RSD605779S, RSD601544S, RSD601818S, RSD608526S, RSD600523S, RSD601952S, RSD606745S, RSD605362S, RSD606823S, RSD600610S, RSD602970S, RSD603071S, RSD603413S, RSD600306S, RSD607287S, RSD607121S, RSD607306S, RSD607996S, RSD607997S, RSD608690S, RSD605769S, RSD606071S, RSD606716S, RSD603069S, RSD604032S, RSD607561S, RSD610582S, RSD608295S, RSD605470S, RSD603387S, RSD600283S, RSD605917S, RSD608205S, RSD602068S, RSD602069S, RSD609285S, RSD609286S, RSD609287S, RSD609288S, RSD609289S, RSD605701S, RSD605703S, RSD605716S, RSD606240S, RSD606241S, RSD606242S, RSD606244S, RSD607396S, RSD608922S, RSD600197S, RSD601338S, RSD601362S, RSD601910S, RSD601929S, RSD602113S, RSD602390S, RSD602727S, RSD602729S, RSD602758S, RSD603338S, RSD603631S, RSD603744S, RSD603965S, RSD603967S, RSD605017S, RSD605021S, RSD605023S, RSD605026S, RSD607036S, RSD605959S, RSD605961S, RSD607739S, RSD607750S, RSD607775S, RSD609365S, RSD603980S, RSD603981S, RSD603985S, RSD603993S, RSD603995S, RSD603997S, RSD603998S, RSD603999S, RSD604000S, RSD604538S, RSD604539S, RSD604540S, RSD604541S, RSD604567S, RSD604568S, RSD605958S, RSD607807S, RSD609290S, RSD609419S, RSD609533S, RSD609591S, RSD609592S, RSD609593S, RSD601957S, RSD601959S, RSD602382S, RSD602383S, RSD602386S, RSD602387S, RSD603259S, RSD603260S, RSD603651S, RSD603705S, RSD606646S, RSD608781S, RSD602911S, RSD608012S, RSD608965S, RSD606707S, RSD600334S, RSD602777S, RSD604551S, RSD602039S, RSD602368S, RSD603635S, RSD603878S, RSD609859S, RSD609860S, RSD609861S, RSD609862S, RSD609863S, RSD609865S, RSD609867S, RSD609868S, RSD609869S, RSD609870S, RSD609872S, RSD609874S, RSD610985S, RSD610990S, RSD610991S, RSD610992S, RSD610993S, RSD610994S, RSD611133S, RSD611134S, RSD611136S, RSD611138S, RSD611139S, RSD611140S, RSD611141S, RSD602054S, RSD608672S, RSD608673S, RSD609706S, RSD609709S, RSD609710S, RSD609711S, RSD609712S, RSD609713S, RSD609714S, RSD609715S, RSD609716S, RSD609717S, RSD609718S, RSD609719S, RSD608674S, RSD608675S, RSD608676S, RSD608677S, RSD602862S, RSD605999S, RSD606002S, RSD606006S, RSD606008S, RSD606009S, RSD606010S, RSD606882S, RSD606012S, RSD606013S, RSD606017S, RSD606874S, RSD606875S, RSD606876S, RSD606877S, RSD606878S, RSD606879S, RSD606880S, RSD606881S, RSD606883S, RSD606884S, RSD606885S, RSD606886S, RSD606887S, RSD606888S, RSD606889S, RSD606890S, RSD606891S, RSD606892S, RSD606893S, RSD607982S, RSD607983S, RSD607984S, RSD607985S, RSD607986S, RSD607987S, RSD607988S, RSD607989S, RSD607990S, RSD607991S, RSD607992S, RSD607993S, RSD609743S, RSD609744S, RSD609745S, RSD610034S, RSD610035S, RSD610036S, RSD610037S, RSD610039S, RSD610040S, RSD600252S, RSD600285S, RSD601400S, RSD601561S, RSD601570S, RSD601796S, RSD601872S, RSD603244S, RSD603523S, RSD609696S, RSD609697S, RSD609698S, RSD610542S, RSD610543S, RSD608911S, RSD608910S, RSD608896S, RSD609645S, RSD609646S, RSD602664S, RSD603298S, RSD608255S, RSD603165S, RSD603659S, RSD605828S, RSD605829S, RSD605830S, RSD605831S, RSD606036S, RSD606037S, RSD606039S, RSD606041S, RSD606042S, RSD606043S, RSD606047S, RSD603024S, RSD603320S, RSD603712S, RSD603715S, RSD603790S, RSD603829S, RSD603863S, RSD603883S, RSD604092S, RSD604093S, RSD604094S, RSD604879S, RSD604880S, RSD604881S, RSD604882S, RSD605124S, RSD605130S, RSD606038S, RSD606044S, RSD606046S, RSD607825S, RSD607826S, RSD607827S, RSD607828S, RSD608241S, RSD608242S, RSD608244S, RSD608245S, RSD608250S, RSD608989S, RSD608990S, RSD608991S, RSD610225S, RSD607060S, RSD609178S, RSD609179S, RSD609180S, RSD607940S, RSD607941S, RSD608174S, RSD608175S, RSD608254S, RSD608616S, RSD608684S, RSD608901S, RSD608959S, RSD609040S, RSD610073S, RSD610135S, RSD610173S, RSD610516S, RSD610536S, RSD610588S, RSD610606S, RSD611004S, RSD605359S, RSD606753S, RSD606898S, RSD607353S, RSD607355S, RSD607585S, RSD607586S, RSD607622S, RSD607653S, RSD607656S, RSD608178S, RSD610054S, RSD610469S, RSD601810S, RSD601812S, RSD602558S, RSD602599S, RSD602994S, RSD602999S, RSD603003S, RSD603085S, RSD603241S, RSD603318S, RSD603328S, RSD603410S, RSD603642S, RSD603945S, RSD603947S, RSD605012S, RSD611109S, RSD605991S, RSD605992S, RSD605993S, RSD605994S, RSD605996S, RSD606852S, RSD607401S, RSD607407S, RSD607412S, RSD608015S, RSD608019S, RSD608005S, RSD608006S, RSD608007S, RSD608008S, RSD608009S, RSD608010S, RSD608011S, RSD608013S, RSD608014S, RSD608016S, RSD608017S, RSD608018S, RSD608661S, RSD608662S, RSD608663S, RSD608664S, RSD608665S, RSD608666S, RSD608667S, RSD608668S, RSD608669S, RSD608670S, RSD608671S, RSD610958S, RSD610959S, RSD605728S, RSD607095S, RSD607322S, RSD606051S, RSD610122S, RSD606751S, RSD608923S, RSD600462S, RSD601849S, RSD603722S, RSD606099S, RSD606795S, RSD607052S, RSD608507S, RSD608849S, RSD603755S, RSD605327S, RSD609044S, RSD600304S, RSD600564S, RSD601574S, RSD602693S, RSD606069S, RSD606767S, RSD607102S, RSD607103S, RSD607558S, RSD601395S, RSD601437S, RSD601580S, RSD602996S, RSD606198S, RSD606289S, RSD606849S, RSD606904S, RSD606905S, RSD607007S, RSD607009S, RSD607495S, RSD607963S, RSD608200S, RSD608529S, RSD608703S, RSD610003S, RSD610005S, RSD610555S, RSD606625S, RSD608142S, RSD610945S, RSD608059S, RSD605319S, RSD606188S, RSD606723S, RSD607079S, RSD607301S, RSD608490S, RSD600591S, RSD600614S, RSD603252S, RSD603532S, RSD605088S, RSD605185S, RSD602721S, RSD609221S, RSD609222S, RSD606064S, RSD606650S, RSD606832S, RSD608982S, RSD609048S, RSD600204S, RSD600274S, RSD600358S, RSD600463S, RSD600464S, RSD600465S, RSD600466S, RSD601584S, RSD601585S, RSD601590S, RSD602269S, RSD602586S, RSD603327S, RSD603333S, RSD607082S, RSD607530S, RSD608130S, RSD608533S, RSD602473S, RSD606673S, RSD608124S, RSD605137S, RSD604878S, RSD601547S, RSD602119S, RSD602903S, RSD603203S, RSD603571S, RSD600310S, RSD606394S, RSD600474S, RSD602417S, RSD603362S, RSD606505S, RSD608238S, RSD605066S, RSD600206S, RSD601874S, RSD602370S, RSD602802S, RSD603676S, RSD610847S, RSD606432S, RSD606476S, RSD608775S, RSD600281S, RSD600568S, RSD602151S, RSD602327S, RSD602632S, RSD602997S, RSD603059S, RSD603394S, RSD603416S, RSD603628S, RSD606819S, RSD600479S, RSD606436S, RSD600359S, RSD604550S, RSD602902S, RSD607526S, RSD608583S, RSD602053S, RSD604545S, RSD603563S, RSD609890S, RSD606182S, RSD606946S, RSD607168S, RSD608787S, RSD609987S, RSD609989S, RSD610171S, RSD601933S, RSD603754S, RSD602474S, RSD606949S, RSD607112S, RSD608133S, RSD609203S, RSD605104S, RSD605105S, RSD600192S, RSD607080S, RSD607452S, RSD607523S, RSD607783S, RSD608229S, RSD608854S, RSD608985S, RSD600392S, RSD600395S, RSD600402S, RSD600426S, RSD600442S, RSD600508S, RSD601427S, RSD601966S, RSD602470S, RSD602876S, RSD603212S, RSD603365S, RSD603397S, RSD603559S, RSD604041S, RSD610557S, RSD610640S, RSD605763S, RSD606092S, RSD602312S, RSD603525S, RSD605321S, RSD605817S, RSD608790S, RSD603325S, RSD603615S, RSD603630S, RSD604037S, RSD605171S, RSD605173S, RSD608864S, RSD606466S, RSD606760S, RSD600469S, RSD605745S, RSD605956S, RSD605957S, RSD606970S, RSD607307S, RSD607824S, RSD608263S, RSD608423S, RSD608644S, RSD600207S, RSD600459S, RSD601430S, RSD601879S, RSD602296S, RSD602776S, RSD602969S, RSD602998S, RSD603019S, RSD603257S, RSD603925S, RSD611005S, RSD602014S, RSD602420S, RSD606079S, RSD606695S, RSD607011S, RSD603837S, RSD606186S, RSD603412S, RSD604574S, RSD606224S, RSD606963S, RSD607260S, RSD607588S, RSD607753S, RSD603940S, RSD607146S, RSD607604S, RSD610661S, RSD607369S, RSD607577S, RSD605308S, RSD608260S, RSD608544S, RSD610352S, RSD600476S, RSD601448S, RSD603247S, RSD604067S, RSD606672S, RSD606189S, RSD600353S, RSD602040S, RSD602150S, RSD604953S, RSD602142S, RSD603107S, RSD604031S, RSD608088S, RSD600524S, RSD602295S, RSD602609S, RSD606813S, RSD608228S, RSD610448S, RSD603753S, RSD607157S, RSD608735S, RSD602868S, RSD606787S, RSD602555S, RSD605245S, RSD605746S, RSD606168S, RSD606715S, RSD606764S, RSD607344S, RSD607952S, RSD601896S, RSD602680S, RSD602681S, RSD603626S, RSD603972S, RSD603973S, RSD604505S, RSD610099S, RSD608107S, RSD608108S, RSD608109S, RSD609821S, RSD609822S, RSD609823S, RSD609824S, RSD609827S, RSD609828S, RSD606062S, RSD606445S, RSD606802S, RSD607324S, RSD607343S, RSD602018S, RSD602105S, RSD603104S, RSD603125S, RSD603136S, RSD603322S, RSD603609S, RSD603924S, RSD605032S, RSD607796S, RSD603598S, RSD605384S, RSD602409S, RSD602872S, RSD603164S, RSD603314S, RSD610304S, RSD606174S, RSD606175S, RSD607147S, RSD607148S, RSD607356S, RSD607358S, RSD607603S, RSD607735S, RSD607738S, RSD607740S, RSD608708S, RSD608930S, RSD610873S, RSD610664S, RSD605342S, RSD605874S, RSD605875S, RSD606643S, RSD606765S, RSD607282S, RSD607580S, RSD607758S, RSD608225S, RSD608500S, RSD608660S, RSD608710S, RSD610220S, RSD610590S, RSD600148S, RSD601927S, RSD602010S, RSD602147S, RSD602172S, RSD602448S, RSD602703S, RSD603073S, RSD603105S, RSD603123S, RSD603126S, RSD603137S, RSD603151S, RSD603245S, RSD603909S, RSD603960S, RSD604499S, RSD604543S, RSD604854S, RSD603390S, RSD600503S, RSD601571S, RSD602867S, RSD605357S, RSD608184S, RSD608736S, RSD609010S, RSD610229S, RSD602280S, RSD604896S, RSD606003S, RSD600303S, RSD601376S, RSD602366S, RSD602412S, RSD602615S, RSD602922S, RSD603271S, RSD609820S, RSD606067S, RSD607228S, RSD607304S, RSD608271S, RSD602601S, RSD610954S, RSD607086S, RSD607606S, RSD607607S, RSD608209S, RSD602244S, RSD605169S, RSD605170S, RSD608686S, RSD610164S, RSD610165S, RSD610465S, RSD600351S, RSD601428S, RSD601822S, RSD601902S, RSD602372S, RSD602373S, RSD602858S, RSD603737S, RSD608261S, RSD600220S, RSD600294S, RSD600381S, RSD601399S, RSD601406S, RSD603871S, RSD604020S, RSD607081S, RSD601606S, RSD601945S, RSD602293S, RSD602527S, RSD603660S, RSD606497S, RSD606499S, RSD606500S, RSD606501S, RSD606503S, RSD606844S, RSD606845S, RSD606846S, RSD606847S, RSD606848S, RSD606850S, RSD606851S, RSD607331S, RSD607332S, RSD607333S, RSD607334S, RSD607335S, RSD607955S, RSD607966S, RSD607954S, RSD607956S, RSD607957S, RSD607958S, RSD607960S, RSD607961S, RSD607962S, RSD607964S, RSD607965S, RSD607967S, RSD607968S, RSD607969S, RSD607970S, RSD607971S, RSD607972S, RSD607973S, RSD607974S, RSD607975S, RSD607976S, RSD607977S, RSD607978S, RSD607979S, RSD609875S, RSD609876S, RSD610006S, RSD607088S, RSD609983S, RSD605203S, RSD609634S, RSD602627S, RSD602754S, RSD603339S, RSD604022S, RSD604027S, RSD610152S, RSD603063S, RSD608517S, RSD610078S, RSD605937S, RSD605938S, RSD606252S, RSD608116S, RSD608117S, RSD608118S, RSD608119S, RSD609094S, RSD605050S, RSD609092S, RSD609093S, RSD610104S, RSD610105S, RSD610578S, RSD605346S, RSD607032S, RSD607035S, RSD607625S, RSD607626S, RSD607892S, RSD607897S, RSD607899S, RSD608950S, RSD602910S, RSD602985S, RSD602989S, RSD602990S, RSD602991S, RSD603025S, RSD603573S, RSD608951S, RSD608103S, RSD608101S, RSD608102S, RSD606780S, RSD606075S, RSD606737S, RSD606853S, RSD609746S, RSD609747S, RSD609748S, RSD609749S, RSD609750S, RSD609751S, RSD609752S, RSD609753S, RSD609754S, RSD609755S, RSD609756S, RSD609757S, RSD609758S, RSD609759S, RSD609760S, RSD609761S, RSD609762S, RSD609763S, RSD609764S, RSD609765S, RSD609766S, RSD609768S, RSD609769S, RSD609770S, RSD609771S, RSD609773S, RSD609774S, RSD609775S, RSD609776S, RSD609777S, RSD609778S, RSD609779S, RSD609780S, RSD609781S, RSD609782S, RSD609783S, RSD609784S, RSD609785S, RSD609786S, RSD609787S, RSD609788S, RSD609789S, RSD609791S, RSD609792S, RSD609793S, RSD609794S, RSD609795S, RSD609797S, RSD609798S, RSD609799S, RSD609800S, RSD609801S, RSD609802S, RSD609803S, RSD609804S, RSD609806S, RSD609807S, RSD609808S, RSD609809S, RSD606685S, RSD606185S, RSD606899S, RSD607083S, RSD608540S, RSD610170S, RSD610316S, RSD610472S, RSD602817S, RSD604245S, RSD604945S, RSD606419S, RSD602464S, RSD603300S, RSD604555S, RSD610306S, RSD606750S, RSD600187S, RSD604028S, RSD604546S, RSD609847S, RSD609848S, RSD609850S, RSD609852S, RSD609853S, RSD609855S, RSD609856S, RSD609857S, RSD609667S, RSD609648S, RSD609651S, RSD609652S, RSD609653S, RSD609654S, RSD609655S, RSD609657S, RSD609658S, RSD609659S, RSD609660S, RSD609662S, RSD609663S, RSD609664S, RSD609665S, RSD609666S, RSD609668S, RSD609669S, RSD609670S, RSD610226S, RSD610234S, RSD610235S, RSD605413S, RSD605415S, RSD606255S, RSD607293S, RSD607518S, RSD608067S, RSD608068S, RSD608069S, RSD608070S, RSD601432S, RSD603122S, RSD603311S, RSD603409S, RSD605748S, RSD605749S, RSD605750S, RSD605751S, RSD605752S, RSD605753S, RSD605756S, RSD605757S, RSD606214S, RSD606216S, RSD606218S, RSD606306S, RSD606333S, RSD606478S, RSD606480S, RSD606618S, RSD606619S, RSD606620S, RSD606621S, RSD606622S, RSD607671S, RSD607672S, RSD607673S, RSD607675S, RSD607676S, RSD607677S, RSD607923S, RSD607925S, RSD607927S, RSD607929S, RSD607931S, RSD607933S, RSD608279S, RSD608280S, RSD608282S, RSD608283S, RSD608284S, RSD608836S, RSD608837S, RSD608839S, RSD608840S, RSD608841S, RSD608906S, RSD608907S, RSD609088S, RSD606341S, RSD608281S, RSD608838S, RSD608908S, RSD608909S, RSD608960S, RSD608961S, RSD609086S, RSD609087S, RSD610008S, RSD610009S, RSD610010S, RSD610614S, RSD610615S, RSD610617S, RSD610618S, RSD610619S, RSD610620S, RSD605047S, RSD605339S, RSD604825S, RSD607695S, RSD609005S, RSD611110S, RSD611111S, RSD611112S, RSD611113S, RSD611114S, RSD611115S, RSD608248S, RSD608249S, RSD608252S, RSD608926S, RSD609058S, RSD608251S, RSD608925S, RSD608969S, RSD608970S, RSD609637S, RSD609638S, RSD609639S, RSD609640S, RSD609641S, RSD609642S, RSD609643S, RSD610473S, RSD605175S, RSD610013S, RSD603189S, RSD606446S, RSD606093S, RSD606195S, RSD606429S, RSD606746S, RSD602757S, RSD605427S, RSD605428S, RSD605429S, RSD605430S, RSD605431S, RSD605432S, RSD605435S, RSD605436S, RSD605437S, RSD605439S, RSD605440S, RSD605482S, RSD605483S, RSD605485S, RSD605490S, RSD605493S, RSD605675S, RSD605676S, RSD605682S, RSD605730S, RSD605731S, RSD605732S, RSD605733S, RSD605734S, RSD605735S, RSD605736S, RSD605737S, RSD605738S, RSD605832S, RSD605833S, RSD605834S, RSD605835S, RSD605836S, RSD605837S, RSD605838S, RSD605839S, RSD605840S, RSD605841S, RSD605844S, RSD605845S, RSD605847S, RSD605848S, RSD605849S, RSD605850S, RSD605852S, RSD605854S, RSD605856S, RSD605859S, RSD605860S, RSD605861S, RSD605862S, RSD605864S, RSD605865S, RSD605867S, RSD605868S, RSD605869S, RSD605870S, RSD605871S, RSD605873S, RSD606101S, RSD606453S, RSD606454S, RSD606455S, RSD606456S, RSD606457S, RSD606458S, RSD606459S, RSD606460S, RSD606461S, RSD606462S, RSD606464S, RSD606757S, RSD606759S, RSD606769S, RSD606770S, RSD606771S, RSD606773S, RSD606774S, RSD606775S, RSD606779S, RSD606792S, RSD606794S, RSD606796S, RSD606854S, RSD606855S, RSD606856S, RSD606857S, RSD606859S, RSD606860S, RSD606861S, RSD606862S, RSD606864S, RSD606865S, RSD606866S, RSD606867S, RSD606868S, RSD606869S, RSD606870S, RSD606871S, RSD606872S, RSD606873S, RSD606924S, RSD607138S, RSD607246S, RSD607247S, RSD607252S, RSD607255S, RSD607256S, RSD607257S, RSD607400S, RSD607436S, RSD607444S, RSD607445S, RSD607450S, RSD607499S, RSD607501S, RSD607514S, RSD607515S, RSD607611S, RSD607616S, RSD607617S, RSD607618S, RSD607619S, RSD607620S, RSD607630S, RSD607631S, RSD607632S, RSD607633S, RSD607634S, RSD607636S, RSD607638S, RSD607649S, RSD608020S, RSD608021S, RSD608022S, RSD608023S, RSD608024S, RSD608027S, RSD608029S, RSD608030S, RSD608031S, RSD608032S, RSD608033S, RSD608034S, RSD608035S, RSD608036S, RSD608037S, RSD608040S, RSD608041S, RSD608042S, RSD608043S, RSD608044S, RSD608045S, RSD608046S, RSD608047S, RSD608048S, RSD608049S, RSD608050S, RSD608051S, RSD608055S, RSD608057S, RSD602041S, RSD602044S, RSD602047S, RSD602049S, RSD602050S, RSD602348S, RSD602349S, RSD602352S, RSD602623S, RSD602625S, RSD602628S, RSD602631S, RSD602633S, RSD602636S, RSD602637S, RSD602638S, RSD602640S, RSD602641S, RSD602643S, RSD602644S, RSD602646S, RSD602647S, RSD602648S, RSD602649S, RSD602650S, RSD602652S, RSD602653S, RSD602654S, RSD602655S, RSD602656S, RSD602657S, RSD602660S, RSD602667S, RSD602668S, RSD602670S, RSD602671S, RSD602674S, RSD602762S, RSD602891S, RSD602892S, RSD602893S, RSD602894S, RSD602895S, RSD602896S, RSD602897S, RSD602898S, RSD602926S, RSD602939S, RSD602944S, RSD602965S, RSD602967S, RSD602976S, RSD602978S, RSD602983S, RSD603011S, RSD603014S, RSD603017S, RSD603018S, RSD603029S, RSD603108S, RSD603110S, RSD603112S, RSD603114S, RSD603115S, RSD603116S, RSD603117S, RSD603118S, RSD603135S, RSD603138S, RSD603139S, RSD603140S, RSD603141S, RSD603142S, RSD603213S, RSD603214S, RSD603216S, RSD603217S, RSD603220S, RSD603221S, RSD603222S, RSD603223S, RSD603224S, RSD603225S, RSD603227S, RSD603239S, RSD603296S, RSD603304S, RSD603307S, RSD603308S, RSD603310S, RSD603632S, RSD603675S, RSD603759S, RSD603760S, RSD603761S, RSD603763S, RSD603764S, RSD603765S, RSD603794S, RSD603795S, RSD603796S, RSD603798S, RSD603800S, RSD603801S, RSD603803S, RSD603804S, RSD603807S, RSD603809S, RSD603810S, RSD603811S, RSD603813S, RSD603814S, RSD603815S, RSD603819S, RSD603821S, RSD603822S, RSD603823S, RSD603850S, RSD603851S, RSD603852S, RSD603853S, RSD603882S, RSD603885S, RSD603900S, RSD603901S, RSD603902S, RSD603903S, RSD603904S, RSD603905S, RSD603906S, RSD603908S, RSD603983S, RSD603986S, RSD603987S, RSD603988S, RSD603989S, RSD603990S, RSD603991S, RSD603992S, RSD603996S, RSD604009S, RSD604052S, RSD604053S, RSD604054S, RSD604056S, RSD604057S, RSD604069S, RSD604070S, RSD604072S, RSD604073S, RSD604075S, RSD604077S, RSD604078S, RSD604079S, RSD604081S, RSD604082S, RSD604083S, RSD604084S, RSD604086S, RSD604087S, RSD604098S, RSD604099S, RSD604100S, RSD604101S, RSD604103S, RSD604104S, RSD604106S, RSD604107S, RSD604108S, RSD604182S, RSD604184S, RSD604185S, RSD604187S, RSD604188S, RSD604189S, RSD604190S, RSD604191S, RSD604192S, RSD604193S, RSD604194S, RSD604195S, RSD604203S, RSD604242S, RSD604249S, RSD604263S, RSD604269S, RSD604270S, RSD604271S, RSD604276S, RSD604277S, RSD604278S, RSD604280S, RSD604281S, RSD604283S, RSD604286S, RSD604291S, RSD604293S, RSD604305S, RSD604320S, RSD604325S, RSD604328S, RSD604330S, RSD604331S, RSD604332S, RSD604333S, RSD604413S, RSD604450S, RSD604451S, RSD604453S, RSD604454S, RSD604456S, RSD604458S, RSD604459S, RSD604460S, RSD604461S, RSD604462S, RSD604463S, RSD604464S, RSD604468S, RSD604469S, RSD604470S, RSD604480S, RSD604481S, RSD604482S, RSD604483S, RSD604484S, RSD604485S, RSD604486S, RSD604487S, RSD604488S, RSD604489S, RSD604490S, RSD604491S, RSD604493S, RSD604494S, RSD604495S, RSD604496S, RSD604498S, RSD604500S, RSD604593S, RSD604718S, RSD604720S, RSD604722S, RSD604724S, RSD604727S, RSD604728S, RSD604730S, RSD604732S, RSD604733S, RSD604735S, RSD604737S, RSD604739S, RSD604743S, RSD604744S, RSD604773S, RSD604774S, RSD604788S, RSD604790S, RSD604798S, RSD604800S, RSD604801S, RSD605433S, RSD605739S, RSD605846S, RSD605851S, RSD605853S, RSD605855S, RSD605857S, RSD605858S, RSD605863S, RSD605866S, RSD605872S, RSD606463S, RSD607253S, RSD607517S, RSD607550S, RSD607609S, RSD607610S, RSD607612S, RSD607613S, RSD607614S, RSD607615S, RSD607637S, RSD607650S, RSD608025S, RSD608026S, RSD608028S, RSD608038S, RSD608039S, RSD608052S, RSD608053S, RSD608056S, RSD608884S, RSD608885S, RSD608886S, RSD608887S, RSD608888S, RSD608889S, RSD608890S, RSD608940S, RSD608941S, RSD608942S, RSD608943S, RSD608944S, RSD609028S, RSD609030S, RSD609031S, RSD609032S, RSD609033S, RSD609034S, RSD609035S, RSD609036S, RSD609037S, RSD609038S, RSD609239S, RSD609240S, RSD609241S, RSD609242S, RSD609243S, RSD609244S, RSD609248S, RSD609249S, RSD609250S, RSD609251S, RSD609253S, RSD609254S, RSD609255S, RSD609257S, RSD609258S, RSD609260S, RSD609261S, RSD609262S, RSD609263S, RSD609264S, RSD609265S, RSD609266S, RSD609267S, RSD609268S, RSD609269S, RSD610286S, RSD610287S, RSD610290S, RSD610997S, RSD610999S, RSD611000S, RSD611001S, RSD611002S, RSD611003S, RSD602099S, RSD602103S, RSD602639S, RSD603519S, RSD606291S, RSD608501S, RSD608650S, RSD600500S, RSD601866S, RSD602598S, RSD604833S, RSD604839S, RSD603098S, RSD603205S, RSD601862S, RSD608894S, RSD607716S, RSD608465S, RSD605395S, RSD605397S, RSD605398S, RSD605402S, RSD605403S, RSD606781S, RSD608551S, RSD603534S, RSD603597S, RSD603726S, RSD603780S, RSD603787S, RSD603886S, RSD604318S, RSD604319S, RSD604562S, RSD604566S, RSD604755S, RSD605399S, RSD605400S, RSD608682S, RSD608683S, RSD608899S, RSD610580S, RSD610581S, RSD606439S, RSD608580S, RSD600243S, RSD600361S, RSD600550S, RSD602403S, RSD603072S, RSD603912S, RSD609725S, RSD609726S, RSD609727S, RSD609728S, RSD609730S, RSD609731S, RSD609733S, RSD609735S, RSD609736S, RSD609737S, RSD609738S, RSD609739S, RSD609740S, RSD609741S, RSD609742S, RSD602974S, RSD605053S, RSD606639S, RSD607814S, RSD609090S, RSD609091S, RSD608905S, RSD608958S, RSD606031S, RSD606055S, RSD606420S, RSD606421S, RSD606719S, RSD606722S, RSD606934S, RSD607071S, RSD608376S, RSD610149S, RSD610150S, RSD600288S, RSD600289S, RSD600297S, RSD600323S, RSD600354S, RSD600385S, RSD600458S, RSD601333S, RSD601396S, RSD601397S, RSD601417S, RSD601418S, RSD601423S, RSD601466S, RSD601467S, RSD601477S, RSD601527S, RSD601582S, RSD601815S, RSD601838S, RSD601845S, RSD601848S, RSD602749S, RSD602794S, RSD602796S, RSD604921S, RSD604923S, RSD606165S, RSD604905S, RSD601337S, RSD601374S, RSD601404S, RSD601414S, RSD601485S, RSD602792S, RSD610148S, RSD602748S, RSD602750S, RSD604922S, RSD607073S, RSD610151S; GTIN 00763000178482, Lot Serial Numbers: RSD601019S, RSD604949S, RSD610833S, RSD605089S, RSD605370S, RSD606575S, RSD609132S, RSD609134S, RSD610496S, RSD607446S, RSD605241S, RSD605565S, RSD610837S, RSD605662S, RSD605106S, RSD610781S, RSD609142S, RSD608373S, RSD608419S, RSD608483S, RSD608424S, RSD607224S, RSD605279S, RSD607354S, RSD607934S, RSD605696S, RSD609107S, RSD609393S, RSD605515S, RSD609434S, RSD608143S, RSD609366S, RSD609369S, RSD609370S, RSD609427S, RSD609461S, RSD609330S, RSD609333S, RSD609341S, RSD609347S, RSD609354S, RSD609356S, RSD605080S, RSD607422S, RSD607857S, RSD608334S, RSD610741S, RSD609879S, RSD607024S, RSD607457S, RSD607513S, RSD611008S, RSD611175S, RSD610635S, RSD607423S, RSD605814S, RSD610256S, RSD610649S, RSD610677S, RSD607930S, RSD610713S, RSD606526S, RSD610487S, RSD610798S, RSD610791S, RSD610796S, RSD608339S, RSD607846S, RSD610752S, RSD607928S, RSD609570S, RSD600764S, RSD601649S, RSD604930S, RSD605237S, RSD600018S, RSD600646S, RSD601031S, RSD601816S, RSD601835S, RSD601857S, RSD601875S, RSD601990S, RSD602807S, RSD602959S, RSD603782S, RSD604746S, RSD604787S, RSD605239S, RSD605795S, RSD606229S, RSD606346S, RSD602822S, RSD602810S, RSD601711S, RSD601712S, RSD601713S, RSD609826S, RSD609832S, RSD609833S, RSD605599S, RSD605600S, RSD606327S, RSD610269S, RSD603477S, RSD607565S, RSD601743S, RSD600090S, RSD602553S, RSD603084S, RSD603087S, RSD603467S, RSD603843S, RSD604931S, RSD604977S, RSD605025S, RSD605055S, RSD605056S, RSD606527S, RSD606535S, RSD607192S, RSD607244S, RSD609372S, RSD609549S, RSD601204S, RSD601222S, RSD605559S, RSD605560S, RSD605561S, RSD605562S, RSD608457S, RSD608459S, RSD608460S, RSD608461S, RSD610278S, RSD610279S, RSD610280S, RSD610281S, RSD602183S, RSD603479S, RSD606321S, RSD607243S, RSD601664S, RSD603653S, RSD604978S, RSD610239S, RSD605228S, RSD607014S, RSD609316S, RSD610241S, RSD600062S, RSD602551S, RSD603602S, RSD604322S, RSD607018S, RSD606328S, RSD606329S, RSD607910S, RSD607903S, RSD609922S, RSD609941S, RSD602169S, RSD603788S, RSD604663S, RSD605652S, RSD607870S, RSD608977S, RSD608980S, RSD600746S, RSD600753S, RSD603291S, RSD603424S, RSD603429S, RSD603430S, RSD603480S, RSD603530S, RSD605601S, RSD606917S, RSD604916S, RSD606369S, RSD607046S, RSD609790S, RSD602805S, RSD603781S, RSD606510S, RSD606511S, RSD607259S, RSD604440S, RSD605386S, RSD604216S, RSD605072S, RSD605624S, RSD605810S, RSD606956S, RSD607463S, RSD607856S, RSD607858S, RSD609376S, RSD610697S, RSD610969S, RSD610973S, RSD611097S, RSD604411S, RSD605161S, RSD600755S, RSD600763S, RSD600771S, RSD600968S, RSD600970S, RSD601004S, RSD601005S, RSD601006S, RSD601233S, RSD601240S, RSD601265S, RSD601316S, RSD601723S, RSD601724S, RSD601789S, RSD601803S, RSD602026S, RSD602027S, RSD602467S, RSD602482S, RSD602961S, RSD603180S, RSD603508S, RSD603709S, RSD603792S, RSD603793S, RSD604036S, RSD604149S, RSD604341S, RSD604412S, RSD604674S, RSD604678S, RSD604685S, RSD604687S, RSD604911S, RSD604968S, RSD605059S, RSD605202S, RSD605226S, RSD605546S, RSD605628S, RSD605671S, RSD606016S, RSD606020S, RSD606347S, RSD606348S, RSD606349S, RSD606403S, RSD606518S, RSD606925S, RSD606928S, RSD607372S, RSD607464S, RSD607888S, RSD608294S, RSD608296S, RSD608297S, RSD608330S, RSD609556S, RSD609557S, RSD609921S, RSD610483S, RSD610675S, RSD611092S, RSD603498S, RSD603499S, RSD603521S, RSD601042S, RSD604434S, RSD604688S, RSD605594S, RSD606103S, RSD606580S, RSD607040S, RSD607134S, RSD607384S, RSD607385S, RSD607420S, RSD608470S, RSD609548S, RSD609560S, RSD609561S, RSD601025S, RSD601050S, RSD601100S, RSD601112S, RSD601289S, RSD601673S, RSD602960S, RSD603121S, RSD603456S, RSD603506S, RSD603544S, RSD603590S, RSD603605S, RSD603620S, RSD603639S, RSD603644S, RSD603783S, RSD603818S, RSD603835S, RSD603836S, RSD604152S, RSD604178S, RSD604295S, RSD604298S, RSD604338S, RSD604347S, RSD604358S, RSD604386S, RSD604389S, RSD604403S, RSD604441S, RSD604442S, RSD604448S, RSD604645S, RSD605164S, RSD605165S, RSD606528S, RSD606529S, RSD606947S, RSD606952S, RSD606953S, RSD606955S, RSD606966S, RSD607198S, RSD607199S, RSD607516S, RSD607521S, RSD607529S, RSD607682S, RSD607683S, RSD609312S, RSD609385S, RSD609386S, RSD609559S, RSD609841S, RSD609913S, RSD609943S, RSD610237S, RSD610700S, RSD603465S, RSD603513S, RSD600064S, RSD601342S, RSD601353S, RSD602835S, RSD602847S, RSD604963S, RSD600174S, RSD600175S, RSD600743S, RSD600766S, RSD600802S, RSD600805S, RSD600806S, RSD600807S, RSD600808S, RSD600809S, RSD601159S, RSD601209S, RSD601214S, RSD601678S, RSD601680S, RSD601895S, RSD601984S, RSD601986S, RSD602029S, RSD602843S, RSD602844S, RSD602845S, RSD603155S, RSD603177S, RSD603461S, RSD603481S, RSD603482S, RSD603507S, RSD603517S, RSD603779S, RSD604206S, RSD604211S, RSD604326S, RSD604348S, RSD604387S, RSD604432S, RSD604452S, RSD604643S, RSD605539S, RSD605545S, RSD605642S, RSD606135S, RSD606136S, RSD606138S, RSD606302S, RSD606303S, RSD606304S, RSD606587S, RSD608350S, RSD609308S, RSD609309S, RSD609310S, RSD609602S, RSD609614S, RSD609621S, RSD609656S, RSD610676S, RSD610695S, RSD610696S, RSD610718S, RSD610746S, RSD610750S, RSD608290S, RSD608291S, RSD608345S, RSD608364S, RSD608462S, RSD609949S, RSD609952S, RSD600022S, RSD600030S, RSD600061S, RSD600063S, RSD600094S, RSD600163S, RSD600647S, RSD600727S, RSD600765S, RSD600817S, RSD601287S, RSD601420S, RSD601507S, RSD601651S, RSD601657S, RSD601825S, RSD602554S, RSD603095S, RSD603119S, RSD603496S, RSD604158S, RSD604394S, RSD604897S, RSD604976S, RSD604979S, RSD605038S, RSD605052S, RSD605236S, RSD606335S, RSD606532S, RSD606586S, RSD607365S, RSD607368S, RSD607863S, RSD608318S, RSD608319S, RSD609373S, RSD604924S, RSD604932S, RSD604980S, RSD606539S, RSD607391S, RSD607718S, RSD609966S, RSD601752S, RSD601775S, RSD602006S, RSD602074S, RSD602958S, RSD603185S, RSD603539S, RSD603540S, RSD603541S, RSD604144S, RSD604796S, RSD604799S, RSD605172S, RSD605423S, RSD606827S, RSD606919S, RSD607155S, RSD607165S, RSD607471S, RSD607472S, RSD609382S, RSD609383S, RSD609552S, RSD609553S, RSD610678S, RSD610679S, RSD611171S, RSD611173S, RSD605842S, RSD601666S, RSD600042S, RSD600043S, RSD600047S, RSD600048S, RSD600168S, RSD601320S, RSD601483S, RSD601506S, RSD601681S, RSD601682S, RSD601720S, RSD601728S, RSD601749S, RSD601756S, RSD601900S, RSD601928S, RSD602189S, RSD602838S, RSD602850S, RSD604179S, RSD604813S, RSD604925S, RSD604965S, RSD605057S, RSD605229S, RSD605509S, RSD606313S, RSD606314S, RSD606316S, RSD602072S, RSD602837S, RSD603512S, RSD603568S, RSD604180S, RSD600095S, RSD600099S, RSD600181S, RSD600679S, RSD603718S, RSD603721S, RSD604159S, RSD604275S, RSD604809S, RSD604908S, RSD607033S, RSD607034S, RSD610694S, RSD602857S, RSD603188S, RSD603603S, RSD606368S, RSD607189S, RSD609944S, RSD610683S, RSD610702S, RSD602124S, RSD605533S, RSD600023S, RSD606315S, RSD600170S, RSD600801S, RSD600815S, RSD601318S, RSD601757S, RSD604952S, RSD605041S, RSD604170S, RSD606366S, RSD609311S, RSD611078S, RSD604964S, RSD605043S, RSD606318S, RSD607025S, RSD607041S, RSD607186S, RSD607559S, RSD607600S, RSD607853S, RSD609315S, RSD609885S, RSD600019S, RSD601261S, RSD601346S, RSD601737S, RSD601806S, RSD603423S, RSD603494S, RSD604156S, RSD604329S, RSD605626S, RSD605640S, RSD605777S, RSD606015S, RSD606105S, RSD608468S, RSD609835S, RSD609920S, RSD600046S, RSD601817S, RSD606045S, RSD606540S, RSD601398S, RSD601407S, RSD601408S, RSD601409S, RSD602025S, RSD602173S, RSD602846S, RSD601330S, RSD601677S, RSD601729S, RSD603529S, RSD603584S, RSD605679S, RSD606364S, RSD607026S, RSD608368S, RSD609942S, RSD610672S, RSD604236S, RSD606027S, RSD606029S, RSD606340S, RSD606359S, RSD606360S, RSD607376S, RSD607377S, RSD607378S, RSD607381S, RSD608478S, RSD608480S, RSD610720S, RSD610721S, RSD610722S, RSD610723S, RSD610744S, RSD610745S, RSD609990S, RSD609992S, RSD603978S, RSD604395S, RSD607431S, RSD608304S, RSD601501S, RSD602483S, RSD603486S, RSD603505S, RSD605160S, RSD605605S, RSD609546S, RSD607462S, RSD607866S, RSD607867S, RSD608332S, RSD609304S, RSD609305S, RSD601350S, RSD601351S, RSD601352S, RSD604756S, RSD604794S, RSD605602S, RSD607058S, RSD609122S, RSD609123S, RSD609306S, RSD609307S, RSD611081S, RSD607851S, RSD607852S, RSD609545S, RSD610986S, RSD603943S, RSD601475S, RSD604565S, RSD608300S, RSD608302S, RSD601263S, RSD602023S, RSD602032S, RSD602033S, RSD602034S, RSD602046S, RSD602066S, RSD605217S, RSD605220S, RSD605222S, RSD605223S, RSD605224S, RSD605210S, RSD606339S, RSD606358S, RSD601026S, RSD601043S, RSD601045S, RSD601047S, RSD604342S, RSD604431S, RSD605468S, RSD605523S, RSD607466S, RSD600044S, RSD601898S, RSD602177S, RSD602793S, RSD602954S, RSD603657S, RSD603666S, RSD603677S, RSD604384S, RSD605034S, RSD605598S, RSD606102S, RSD606582S, RSD608336S, RSD608477S, RSD610261S, RSD610262S, RSD610998S, RSD603693S, RSD601807S, RSD601808S, RSD601864S, RSD601867S, RSD603406S, RSD603604S, RSD604999S, RSD605621S, RSD606522S, RSD607191S, RSD607373S, RSD607847S, RSD609319S, RSD609959S, RSD611158S, RSD601230S, RSD601328S, RSD601329S, RSD601285S, RSD601915S, RSD601918S, RSD604155S, RSD604665S, RSD607021S, RSD607022S, RSD601632S, RSD601832S, RSD600945S, RSD600947S, RSD600949S, RSD600950S, RSD601484S, RSD601493S, RSD601494S, RSD601495S, RSD601496S, RSD601497S, RSD601498S, RSD601499S, RSD601500S, RSD603592S, RSD604344S, RSD604381S, RSD604382S, RSD604689S, RSD604715S, RSD604769S, RSD605547S, RSD605548S, RSD605645S, RSD605646S, RSD605649S, RSD605650S, RSD605651S, RSD609796S, RSD609834S, RSD610631S, RSD610942S, RSD606336S, RSD606927S, RSD604915S, RSD605033S, RSD607467S, RSD608292S, RSD608471S, RSD601099S, RSD603460S, RSD604151S, RSD606312S, RSD601944S, RSD601947S, RSD601955S, RSD601203S, RSD604950S, RSD603455S, RSD603483S, RSD604169S, RSD604990S, RSD604991S, RSD605550S, RSD605551S, RSD605632S, RSD605634S, RSD605635S, RSD605637S, RSD606357S, RSD610724S, RSD610725S, RSD610742S, RSD610743S, RSD610943S, RSD610946S, RSD600917S, RSD601254S, RSD601411S, RSD601804S, RSD603471S, RSD603473S, RSD610265S, RSD610266S, RSD600772S, RSD600803S, RSD600934S, RSD601241S, RSD601247S, RSD601249S, RSD601288S, RSD601760S, RSD601813S, RSD602181S, RSD603484S, RSD603890S, RSD604219S, RSD604853S, RSD604989S, RSD605008S, RSD606530S, RSD602112S, RSD602115S, RSD602116S, RSD601644S, RSD601751S, RSD601883S, RSD602531S, RSD603570S, RSD603772S, RSD604383S, RSD604455S, RSD607474S, RSD603893S, RSD607200S, RSD604895S, RSD609318S, RSD605486S, RSD605525S, RSD605526S, RSD610508S, RSD610510S, RSD610511S, RSD601424S, RSD601436S, RSD601439S, RSD601444S, RSD601761S, RSD601997S, RSD601999S, RSD602000S, RSD602030S, RSD602070S, RSD602102S, RSD604145S, RSD604366S, RSD604713S, RSD604907S, RSD606407S, RSD607382S, RSD608369S, RSD607388S, RSD607475S, RSD607652S, RSD600768S, RSD601962S, RSD603431S, RSD603569S, RSD603681S, RSD603682S, RSD604150S, RSD604375S, RSD604377S, RSD604378S, RSD604379S, RSD604729S, RSD606350S, RSD606356S, RSD607197S, RSD607416S, RSD607418S, RSD607906S, RSD609562S, RSD609563S, RSD609564S, RSD610726S, RSD610736S, RSD610947S, RSD610951S, RSD610974S, RSD610983S, RSD611157S, RSD604817S, RSD604823S, RSD604834S, RSD606030S, RSD606034S, RSD606077S, RSD606089S, RSD606095S, RSD604912S, RSD604371S, RSD606235S, RSD606237S, RSD601309S, RSD602082S, RSD601995S, RSD601996S, RSD602079S, RSD602088S, RSD602188S, RSD602948S, RSD602946S, RSD603463S, RSD603557S, RSD603654S, RSD604335S, RSD605001S, RSD605002S, RSD605003S, RSD605004S, RSD606231S, RSD606232S, RSD606234S, RSD606537S, RSD606542S, RSD606543S, RSD607013S, RSD607031S, RSD607366S, RSD607367S, RSD607724S, RSD607862S, RSD607893S, RSD608328S, RSD608366S, RSD609133S, RSD609565S, RSD609915S, RSD610317S, RSD610480S, RSD610481S, RSD610482S, RSD610485S, RSD610716S, RSD602957S, RSD604233S, RSD606298S, RSD607839S, RSD601837S, RSD601843S, RSD603109S, RSD604701S, RSD604903S, RSD604906S, RSD605214S, RSD606246S, RSD606326S, RSD607375S, RSD607476S, RSD607508S, RSD609954S, RSD610268S, RSD602001S, RSD604367S, RSD605508S, RSD610755S, RSD610756S, RSD610759S, RSD610926S, RSD610931S, RSD610937S, RSD602071S, RSD602075S, RSD602823S, RSD605227S, RSD605230S, RSD605232S, RSD605233S, RSD606405S, RSD607850S, RSD608346S, RSD608365S, RSD610017S, RSD610049S, RSD610698S, RSD610699S, RSD604926S, RSD604944S, RSD608303S, RSD601470S, RSD601472S, RSD603474S, RSD603503S, RSD604267S, RSD604349S, RSD605630S, RSD605755S, RSD607194S, RSD607195S, RSD608370S, RSD608454S, RSD609313S, RSD609314S, RSD610267S, RSD610521S, RSD610673S, RSD610955S, RSD601360S, RSD601744S, RSD603591S, RSD603775S, RSD603786S, RSD604784S, RSD605661S, RSD606250S, RSD606513S, RSD607242S, RSD602184S, RSD604385S, RSD605569S, RSD603785S, RSD605181S, RSD609558S, RSD606293S, RSD607470S, RSD601033S, RSD601809S, RSD604975S, RSD605639S, RSD607002S, RSD609690S, RSD610275S, RSD610277S, RSD601028S, RSD605073S, RSD605076S, RSD605213S, RSD606520S, RSD608320S, RSD604346S, RSD604354S, RSD604355S, RSD604357S, RSD607154S, RSD607510S, RSD607511S, RSD607817S, RSD607835S, RSD607896S, RSD608299S, RSD608335S, RSD608367S, RSD608514S, RSD609550S, RSD609568S, RSD610512S, RSD602270S, RSD602273S, RSD602274S, RSD607408S, RSD607409S, RSD607410S, RSD608348S, RSD608349S, RSD608474S, RSD608475S, RSD608476S, RSD609961S, RSD609962S, RSD610728S, RSD610729S, RSD610730S, RSD610734S, RSD610735S, RSD611057S, RSD611163S, RSD611165S, RSD611167S, RSD611168S, RSD604847S, RSD604865S, RSD605535S, RSD609945S, RSD609946S, RSD609947S, RSD602118S, RSD602122S, RSD606365S, RSD602187S, RSD604309S, RSD606960S, RSD607919S, RSD610753S, RSD601672S, RSD601687S, RSD602438S, RSD602544S, RSD603163S, RSD603183S, RSD603193S, RSD603276S, RSD603288S, RSD604557S, RSD605456S, RSD605524S, RSD606562S, RSD606570S, RSD606571S, RSD606958S, RSD606978S, RSD606979S, RSD607029S, RSD607488S, RSD608306S, RSD608359S, RSD608360S, RSD608455S, RSD609154S, RSD609303S, RSD603892S, RSD604105S, RSD604336S, RSD604339S, RSD605014S, RSD605643S, RSD605808S, RSD606344S, RSD607188S, RSD608329S, RSD602572S, RSD603273S, RSD604148S, RSD604380S, RSD604516S, RSD604866S, RSD605603S, RSD606465S, RSD606592S, RSD609567S, RSD611096S, RSD601269S, RSD601271S, RSD601321S, RSD601600S, RSD601645S, RSD601831S, RSD603199S, RSD603229S, RSD603464S, RSD603504S, RSD604345S, RSD605534S, RSD606025S, RSD606311S, RSD606524S, RSD606581S, RSD607037S, RSD607271S, RSD607383S, RSD607848S, RSD610274S, RSD610297S, RSD600773S, RSD601907S, RSD602556S, RSD604951S, RSD603466S, RSD604163S, RSD607864S, RSD607865S, RSD610630S, RSD610779S, RSD607900S, RSD600635S, RSD600636S, RSD601356S, RSD601748S, RSD601905S, RSD602164S, RSD603487S, RSD603594S, RSD603771S, RSD603918S, RSD604288S, RSD604304S, RSD604368S, RSD604435S, RSD605061S, RSD605204S, RSD606301S, RSD606590S, RSD606591S, RSD607016S, RSD607405S, RSD610238S, RSD601490S, RSD601491S, RSD603832S, RSD605629S, RSD606338S, RSD606929S, RSD610953S, RSD601262S, RSD601264S, RSD601635S, RSD603542S, RSD605058S, RSD605528S, RSD604373S, RSD600038S, RSD600041S, RSD600761S, RSD600769S, RSD600770S, RSD600928S, RSD601075S, RSD601108S, RSD601310S, RSD601311S, RSD601700S, RSD601701S, RSD602161S, RSD602162S, RSD602167S, RSD602168S, RSD603238S, RSD603428S, RSD603436S, RSD603444S, RSD603446S, RSD603449S, RSD603451S, RSD603453S, RSD603458S, RSD603459S, RSD603470S, RSD604186S, RSD604207S, RSD604209S, RSD604210S, RSD604213S, RSD604343S, RSD604706S, RSD604785S, RSD607049S, RSD607050S, RSD607056S, RSD607057S, RSD607167S, RSD607170S, RSD607175S, RSD607176S, RSD607178S, RSD607179S, RSD607182S, RSD607183S, RSD607184S, RSD607185S, RSD606334S, RSD607020S, RSD606107S, RSD603663S, RSD603664S, RSD603950S, RSD604405S, RSD604409S, RSD605243S, RSD606351S, RSD608298S, RSD610693S, RSD601762S, RSD601765S, RSD601930S, RSD601942S, RSD602007S, RSD602008S, RSD602009S, RSD602016S, RSD602017S, RSD602019S, RSD602020S, RSD602021S, RSD602022S, RSD602515S, RSD602516S, RSD602853S, RSD602854S, RSD602855S, RSD603134S, RSD603144S, RSD603148S, RSD603489S, RSD603526S, RSD603543S, RSD603572S, RSD607042S, RSD607043S, RSD607044S, RSD607392S, RSD607398S, RSD607399S, RSD607402S, RSD607404S, RSD609387S, RSD609388S, RSD609892S, RSD609893S, RSD610308S, RSD610311S, RSD610313S, RSD610684S, RSD610685S, RSD607004S, RSD607005S, RSD607010S, RSD604250S, RSD609124S, RSD602537S, RSD602842S, RSD603566S, RSD604365S, RSD604933S, RSD604996S, RSD606106S, RSD606108S, RSD606140S, RSD606401S, RSD606319S, RSD606536S, RSD608467S, RSD600779S, RSD600780S, RSD600781S, RSD600782S, RSD600783S, RSD600784S, RSD600785S, RSD600792S, RSD600793S, RSD600796S, RSD600797S, RSD600798S, RSD601300S, RSD601301S, RSD601302S, RSD601304S, RSD601463S, RSD601502S, RSD601505S, RSD601659S, RSD601745S, RSD602502S, RSD602503S, RSD602949S, RSD602952S, RSD603561S, RSD603839S, RSD604939S, RSD605064S, RSD605065S, RSD605083S, RSD605092S, RSD608354S, RSD609849S, RSD609997S, RSD600171S, RSD600759S, RSD600930S, RSD600931S, RSD601032S, RSD601208S, RSD601251S, RSD601252S, RSD601266S, RSD601422S, RSD602077S, RSD603427S, RSD603817S, RSD604272S, RSD604362S, RSD604917S, RSD604997S, RSD605166S, RSD606561S, RSD607932S, RSD608301S, RSD601205S, RSD601446S, RSD601746S, RSD602002S, RSD602840S, RSD602951S, RSD603441S, RSD603565S, RSD604248S, RSD604350S, RSD604356S, RSD604406S, RSD604414S, RSD604446S, RSD604955S, RSD604971S, RSD604974S, RSD605078S, RSD605079S, RSD605505S, RSD608456S, RSD609302S, RSD609384S, RSD609995S, RSD609999S, RSD602003S, RSD601315S, RSD602841S, RSD603232S, RSD603240S, RSD603261S, RSD603432S, RSD603497S, RSD603917S, RSD604143S, RSD604273S, RSD604351S, RSD604364S, RSD604418S, RSD604427S, RSD604698S, RSD604781S, RSD604918S, RSD604919S, RSD604981S, RSD604986S, RSD604988S, RSD605027S, RSD605516S, RSD606139S, RSD606307S, RSD606525S, RSD606552S, RSD606959S, RSD607028S, RSD607419S, RSD607684S, RSD607759S, RSD608326S, RSD609825S, RSD600031S, RSD602495S, RSD602498S, RSD603192S, RSD603269S, RSD603469S, RSD603583S, RSD603585S, RSD603907S, RSD604011S, RSD604112S, RSD604323S, RSD604843S, RSD604855S, RSD604899S, RSD606331S, RSD606332S, RSD606402S, RSD606541S, RSD607406S, RSD608338S, RSD608352S, RSD608353S, RSD609117S, RSD610507S, RSD601016S, RSD601344S, RSD601345S, RSD601741S, RSD602535S, RSD602560S, RSD602563S, RSD602574S, RSD602578S, RSD602787S, RSD602790S, RSD603194S, RSD603426S, RSD603607S, RSD603714S, RSD610298S, RSD600787S, RSD600788S, RSD600789S, RSD601758S, RSD601919S, RSD601920S, RSD601970S, RSD603510S, RSD603516S, RSD603537S, RSD603766S, RSD604994S, RSD601260S, RSD601668S, RSD601854S, RSD602160S, RSD602178S, RSD602179S, RSD602194S, RSD603527S, RSD603845S, RSD604296S, RSD604447S, RSD604868S, RSD604870S, RSD605006S, RSD606228S, RSD606558S, RSD603588S, RSD601992S, RSD602004S, RSD604902S, RSD605000S, RSD606317S, RSD606324S, RSD606325S, RSD606546S, RSD606559S, RSD601250S, RSD600087S, RSD600967S, RSD600016S, RSD600017S, RSD600176S, RSD600177S, RSD600178S, RSD600179S, RSD600180S, RSD600182S, RSD600183S, RSD600295S, RSD600628S, RSD600633S, RSD600639S, RSD600644S, RSD600748S, RSD600800S, RSD600810S, RSD600812S, RSD600813S, RSD600814S, RSD600914S, RSD600973S, RSD601201S, RSD601258S, RSD601290S, RSD601291S, RSD601292S, RSD601294S, RSD601298S, RSD601299S, RSD601306S, RSD601339S, RSD601419S, RSD601658S, RSD601690S, RSD601695S, RSD601715S, RSD601734S, RSD601736S, RSD601738S, RSD601778S, RSD601782S, RSD601788S, RSD601901S, RSD601943S, RSD601965S, RSD601989S, RSD601991S, RSD602005S, RSD602028S, RSD602170S, RSD602186S, RSD602210S, RSD602221S, RSD602488S, RSD602530S, RSD602532S, RSD602543S, RSD602546S, RSD602573S, RSD602803S, RSD602806S, RSD602836S, RSD602851S, RSD603434S, RSD603438S, RSD603440S, RSD603442S, RSD603447S, RSD603509S, RSD603511S, RSD603538S, RSD603649S, RSD603777S, RSD603797S, RSD603802S, RSD603805S, RSD603806S, RSD603808S, RSD603812S, RSD603816S, RSD604217S, RSD604229S, RSD604254S, RSD604264S, RSD604359S, RSD604360S, RSD604970S, RSD604972S, RSD604973S, RSD604982S, RSD604983S, RSD604984S, RSD604985S, RSD604992S, RSD605028S, RSD605029S, RSD605030S, RSD605180S, RSD605520S, RSD606305S, RSD606308S, RSD606309S, RSD606310S, RSD606553S, RSD606555S, RSD606556S, RSD606595S, RSD606596S, RSD606600S, RSD606961S, RSD606964S, RSD606971S, RSD606972S, RSD606975S, RSD606977S, RSD607166S, RSD607190S, RSD607504S, RSD607506S, RSD607688S, RSD607706S, RSD607832S, RSD607860S, RSD607885S, RSD607887S, RSD607913S, RSD608311S, RSD600172S, RSD600652S, RSD600713S, RSD600717S, RSD600720S, RSD600721S, RSD600722S, RSD600811S, RSD601278S, RSD601282S, RSD601283S, RSD601319S, RSD601331S, RSD601355S, RSD601691S, RSD601716S, RSD601982S, RSD601993S, RSD602165S, RSD602171S, RSD602191S, RSD602848S, RSD603111S, RSD603502S, RSD603556S, RSD603560S, RSD604352S, RSD604967S, RSD604969S, RSD604987S, RSD604993S, RSD604995S, RSD605031S, RSD601348S, RSD601349S, RSD601646S, RSD601667S, RSD601739S, RSD603187S, RSD603457S, RSD603514S, RSD603767S, RSD604415S, RSD604778S, RSD604849S, RSD605040S, RSD605519S, RSD606320S, RSD605163S, RSD607193S, RSD605037S, RSD605215S, RSD605813S, RSD606096S, RSD606098S, RSD606100S, RSD606584S, RSD606585S, RSD608309S, RSD608315S, RSD607924S, RSD606251S, RSD606550S, RSD606568S, RSD609980S, RSD601307S, RSD601242S, RSD601246S, RSD601725S, RSD601727S, RSD601994S, RSD601998S, RSD602163S, RSD603520S, RSD603528S, RSD603824S, RSD603830S, RSD603831S, RSD604265S, RSD604266S, RSD604898S, RSD604941S, RSD604948S, RSD606516S, RSD601625S, RSD602808S, RSD604388S, RSD606515S, RSD607003S, RSD607922S, RSD603574S, RSD603648S, RSD602809S, RSD601772S, RSD601627S, RSD607245S, RSD609681S, RSD601199S, RSD601597S, RSD604120S, RSD604129S, RSD604659S, RSD605689S, RSD602559S, RSD603178S, RSD603179S, RSD601899S, RSD607843S, RSD605506S, RSD606153S, RSD607169S, RSD608314S, RSD609375S, RSD609544S, RSD601127S, RSD601536S, RSD601549S, RSD601562S, RSD602829S, RSD602830S, RSD602861S, RSD603789S, RSD605257S, RSD601228S, RSD601235S, RSD605250S, RSD605265S, RSD601191S, RSD601215S, RSD603799S, RSD605254S, RSD601034S, RSD601322S, RSD604165S, RSD605412S, RSD608481S, RSD601072S, RSD601082S, RSD601126S, RSD601138S, RSD601151S, RSD605249S, RSD605418S, RSD605445S, RSD605471S, RSD606376S, RSD608450S, RSD601588S, RSD605286S, RSD607207S, RSD605282S, RSD601171S, RSD601705S, RSD601799S, RSD602863S, RSD604164S, RSD606393S, RSD607216S, RSD601143S, RSD601218S, RSD601092S, RSD606383S, RSD606384S, RSD607212S, RSD607213S, RSD607297S, RSD607308S, RSD601142S, RSD601115S, RSD601128S, RSD601186S, RSD601534S, RSD601634S, RSD601669S, RSD602316S, RSD605278S, RSD605280S, RSD607329S, RSD607438S, RSD608166S, RSD601080S, RSD605247S, RSD605128S, RSD606268S, RSD606275S, RSD605039S, RSD607205S, RSD601592S, RSD601599S, RSD603966S, RSD604117S, RSD604125S, RSD604430S, RSD604738S, RSD604762S, RSD605125S, RSD605529S, RSD605532S, RSD605542S, RSD605543S, RSD605552S, RSD605657S, RSD606127S, RSD606128S, RSD606130S, RSD606131S, RSD606253S, RSD606260S, RSD607201S, RSD607202S, RSD608492S, RSD604116S, RSD606120S, RSD606122S, RSD601136S, RSD604175S, RSD607447S, RSD608412S, RSD608495S, RSD607241S, RSD606104S, RSD606110S, RSD609443S, RSD605573S, RSD605656S, RSD606113S, RSD601117S, RSD601119S, RSD605264S, RSD605255S, RSD604786S, RSD606119S, RSD609114S, RSD604681S, RSD604683S, RSD605097S, RSD606004S, RSD605664S, RSD605706S, RSD605707S, RSD605708S, RSD605709S, RSD607235S, RSD607236S, RSD607238S, RSD608523S, RSD608535S, RSD608537S, RSD608539S, RSD608541S, RSD609321S, RSD609322S, RSD609323S, RSD609324S, RSD609325S, RSD609326S, RSD609327S, RSD606117S, RSD601482S, RSD601508S, RSD601517S, RSD601518S, RSD605131S, RSD605133S, RSD605604S, RSD605606S, RSD605607S, RSD605617S, RSD605618S, RSD605712S, RSD605713S, RSD605714S, RSD605715S, RSD606123S, RSD606124S, RSD606129S, RSD601593S, RSD601595S, RSD603838S, RSD603946S, RSD603951S, RSD604439S, RSD605721S, RSD601094S, RSD601095S, RSD601148S, RSD601538S, RSD601628S, RSD602860S, RSD605271S, RSD601654S, RSD602831S, RSD604172S, RSD604176S, RSD605269S, RSD604443S, RSD604444S, RSD604745S, RSD604828S, RSD604830S, RSD605122S, RSD605123S, RSD605619S, RSD605654S, RSD605685S, RSD605717S, RSD605718S, RSD606115S, RSD606116S, RSD606125S, RSD606126S, RSD607894S, RSD607918S, RSD608431S, RSD608469S, RSD608510S, RSD608532S, RSD608567S, RSD608569S, RSD609175S, RSD609187S, RSD609208S, RSD601509S, RSD605110S, RSD605684S, RSD607232S, RSD601141S, RSD601145S, RSD605293S, RSD605499S, RSD607439S, RSD604127S, RSD604136S, RSD604436S, RSD604437S, RSD604777S, RSD605116S, RSD605668S, RSD605669S, RSD605672S, RSD605680S, RSD605688S, RSD605697S, RSD605698S, RSD606109S, RSD606111S, RSD606132S, RSD606133S, RSD605093S, RSD608986S, RSD606112S, RSD605570S, RSD605686S, RSD609169S, RSD604393S, RSD605262S, RSD602953S, RSD603791S, RSD605284S, RSD605326S, RSD607441S, RSD601102S, RSD601144S, RSD601183S, RSD601189S, RSD601513S, RSD601546S, RSD601565S, RSD601589S, RSD601655S, RSD601786S, RSD601827S, RSD601834S, RSD602864S, RSD602945S, RSD605272S, RSD606385S, RSD606391S, RSD606392S, RSD606396S, RSD608391S, RSD601150S, RSD601129S, RSD601305S, RSD602825S, RSD601272S, RSD601610S, RSD601616S, RSD606602S, RSD601529S, RSD601802S, RSD606330S, RSD606603S, RSD601604S, RSD605240S, RSD601088S, RSD606588S, RSD607208S, RSD601602S, RSD601607S, RSD601611S, RSD601613S, RSD601514S, RSD601523S, RSD601524S, RSD604677S, RSD605096S, RSD605100S, RSD605103S, RSD605108S, RSD605118S, RSD605455S, RSD605658S, RSD605700S, RSD605719S, RSD606118S, RSD606299S, RSD607234S, RSD608496S, RSD608508S, RSD609070S, RSD604782S, RSD605690S, RSD605576S, RSD601598S, RSD601601S, RSD604433S, RSD604445S, RSD604783S, RSD601077S, RSD601539S, RSD601653S, RSD602832S, RSD602859S, RSD605307S, RSD607218S, RSD607219S, RSD607337S, RSD602228S, RSD601039S, RSD601052S, RSD601091S, RSD601227S, RSD601248S, RSD601313S, RSD602284S, RSD602849S, RSD605290S, RSD605299S, RSD607877S, RSD608425S, RSD601605S, RSD601608S, RSD605085S, RSD605095S, RSD605704S, RSD606254S, RSD601156S, RSD601181S, RSD605478S, RSD601476S, RSD601478S, RSD601480S, RSD601481S, RSD601516S, RSD601519S, RSD601520S, RSD601521S, RSD601522S, RSD601525S, RSD604160S, RSD604166S, RSD604675S, RSD605094S, RSD605099S, RSD605318S, RSD605443S, RSD605503S, RSD605579S, RSD605580S, RSD605660S, RSD605665S, RSD605666S, RSD605681S, RSD605692S, RSD605699S, RSD605702S, RSD605705S, RSD605710S, RSD605720S, RSD606271S, RSD601617S, RSD601674S, RSD602364S, RSD602826S, RSD601089S, RSD601213S, RSD601216S, RSD601747S, RSD605253S, RSD601054S, RSD601101S, RSD601170S, RSD601232S, RSD601639S, RSD601076S, RSD601087S, RSD601149S, RSD601155S, RSD601161S, RSD601165S, RSD601179S, RSD601198S, RSD601219S, RSD601735S, RSD601794S, RSD602814S, RSD602816S, RSD602818S, RSD602824S, RSD604171S, RSD605246S, RSD605248S, RSD605267S, RSD605469S, RSD605557S, RSD606379S, RSD607226S, RSD607489S, RSD608387S, RSD608411S, RSD601174S, RSD600754S, RSD601188S, RSD601244S, RSD601312S, RSD601504S, RSD601530S, RSD601535S, RSD601717S, RSD601811S, RSD602289S, RSD602811S, RSD605310S, RSD605325S, RSD606371S, RSD606373S, RSD606374S, RSD606399S, RSD607214S, RSD607437S, RSD608435S, RSD601147S, RSD601805S, RSD601063S, RSD601065S, RSD601066S, RSD601123S, RSD601132S, RSD605275S, RSD601106S, RSD601116S, RSD601533S, RSD601742S, RSD602819S, RSD604438S, RSD605258S, RSD605260S, RSD608406S, RSD601193S, RSD601531S, RSD601630S, RSD601296S, RSD606378S, RSD602230S, RSD603827S, RSD606381S, RSD601086S, RSD601163S, RSD601164S, RSD602226S, RSD606382S, RSD601512S, RSD601146S, RSD601122S, RSD606395S, RSD608448S, RSD601062S, RSD601124S, RSD608287S, RSD601158S, RSD601097S, RSD601098S, RSD602225S, RSD605277S, RSD601078S, RSD601243S, RSD602828S, RSD605311S, RSD606397S, RSD601486S, RSD605481S, RSD601192S, RSD608485S, RSD605274S, RSD606388S, RSD606579S, RSD601104S, RSD601621S, RSD608392S, RSD601064S, RSD608403S, RSD608429S, RSD601103S, RSD601184S, RSD602279S, RSD604173S, RSD605329S, RSD601220S, RSD605303S, RSD607878S, RSD601111S, RSD601114S, RSD601169S, RSD601270S, RSD601303S, RSD601314S, RSD601326S, RSD601528S, RSD601618S, RSD601624S, RSD602353S, RSD602431S, RSD604162S, RSD605458S, RSD605473S, RSD605502S, RSD608394S, RSD608443S, RSD608451S, RSD601631S, RSD604161S, RSD601280S, RSD606380S, RSD601176S, RSD601178S, RSD601622S, RSD602302S, RSD602401S, RSD601180S, RSD605488S, RSD601162S, RSD601157S, RSD601182S, RSD601173S, RSD601139S, RSD601167S, RSD601083S, RSD601160S, RSD601217S, RSD601554S, RSD601798S, RSD602827S, RSD606377S, RSD607248S, RSD601081S, RSD601152S, RSD601153S, RSD605530S, RSD604274S, RSD604287S, RSD604290S, RSD604937S, RSD605007S, RSD605042S, RSD606548S, RSD606551S, RSD606599S, RSD607492S, RSD607907S, RSD603953S, RSD604696S, RSD605513S, RSD605521S, RSD605694S, RSD601660S, RSD603439S, RSD604268S, RSD604363S, RSD605238S, RSD607798S, RSD600767S, RSD600818S, RSD605405S, RSD601685S, RSD605259S, RSD608482S, RSD603959S, RSD603961S, RSD604137S, RSD604140S, RSD605119S, RSD605120S, RSD605578S, RSD604119S, RSD601206S, RSD602548S, RSD602833S, RSD607048S, RSD607096S, RSD609299S, RSD608445S, RSD601069S, RSD604449S, RSD604676S, RSD605015S, RSD605035S, RSD605566S, RSD605588S, RSD605593S, RSD605595S, RSD605678S, RSD605711S, RSD605796S, RSD605798S, RSD605806S, RSD606005S, RSD606300S, RSD606337S, RSD607030S, RSD607038S, RSD607039S, RSD607855S, RSD608288S, RSD610270S, RSD610488S, RSD610505S, RSD610506S, RSD610751S, RSD611053S, RSD605154S, RSD604709S, RSD601683S, RSD602193S, RSD600794S, RSD600795S, RSD600821S, RSD601718S, RSD601719S, RSD602237S, RSD602238S, RSD602247S, RSD602252S, RSD604243S, RSD605176S, RSD606563S, RSD607914S, RSD609252S, RSD610500S, RSD610501S, RSD600091S, RSD600093S, RSD605693S, RSD601341S, RSD603645S, RSD602086S, RSD602089S, RSD607426S, RSD607432S, RSD601730S, RSD602223S, RSD601637S, RSD601130S, RSD608395S, RSD600733S, RSD601532S, RSD601619S, RSD605273S, RSD602571S, RSD604390S, RSD606943S, RSD601079S, RSD605287S, RSD603531S, RSD604673S, RSD607413S, RSD610491S, RSD605309S, RSD601323S, RSD601325S, RSD601327S, RSD601340S, RSD601694S, RSD602795S, RSD603200S, RSD603228S, RSD604372S, RSD604805S, RSD604808S, RSD604914S, RSD606370S, RSD601609S, RSD610259S, RSD610260S, RSD610717S, RSD609948S, RSD609953S, RSD602254S, RSD602255S, RSD602259S, RSD602260S, RSD602264S, RSD602265S, RSD602267S, RSD602268S, RSD602275S, RSD602276S, RSD602278S, RSD606276S, RSD607363S, RSD607714S, RSD607755S, RSD607776S, RSD607786S, RSD609418S, RSD609420S, RSD609528S, RSD602253S, RSD601721S, RSD601722S, RSD601726S, RSD601731S, RSD601732S, RSD601733S, RSD602235S, RSD605583S, RSD605587S, RSD607815S, RSD608414S, RSD608417S, RSD609188S, RSD609320S, RSD609466S, RSD609467S, RSD609468S, RSD609469S, RSD609470S, RSD607709S, RSD607710S, RSD609377S, RSD609378S, RSD609379S, RSD609380S, RSD609381S, RSD609395S, RSD609396S, RSD609405S, RSD609406S, RSD600959S, RSD600960S, RSD601014S, RSD601200S, RSD601286S, RSD604321S, RSD606604S, RSD606605S, RSD606608S, RSD606858S, RSD606995S, RSD607062S, RSD607063S, RSD607276S, RSD607280S, RSD607281S, RSD605235S, RSD605553S, RSD605622S, RSD605623S, RSD605625S, RSD605638S, RSD603197S, RSD603888S, RSD604672S, RSD605184S, RSD605186S, RSD605193S, RSD605197S, RSD606352S, RSD607269S, RSD607270S, RSD607414S, RSD607491S, RSD607496S, RSD607722S, RSD607859S, RSD608307S, RSD608310S, RSD608355S, RSD609292S, RSD609297S, RSD609298S, RSD609301S, RSD609734S, RSD609858S, RSD609866S, RSD609871S, RSD609899S, RSD609900S, RSD609902S, RSD609908S, RSD609982S, RSD609988S, RSD601121S, RSD601125S, RSD601224S, RSD601225S, RSD607229S, RSD605174S, RSD607502S, RSD608344S, RSD610252S, RSD600775S, RSD601676S, RSD601759S, RSD601894S, RSD602487S, RSD602549S, RSD602834S, RSD603075S, RSD603443S, RSD603515S, RSD603533S, RSD603582S, RSD604686S, RSD604714S, RSD604966S, RSD605504S, RSD606215S, RSD606342S, RSD606547S, RSD606549S, RSD606564S, RSD606565S, RSD606598S, RSD606916S, RSD607261S, RSD607266S, RSD607486S, RSD607494S, RSD607911S, RSD609296S, RSD600776S, RSD603266S, RSD604282S, RSD604936S, RSD607679S, RSD607909S, RSD607912S, RSD604416S, RSD602568S, RSD603536S, RSD600778S, RSD601709S, RSD604374S, RSD603211S, RSD606249S, RSD607665S, RSD608343S, RSD609873S, RSD610493S, RSD601851S, RSD605514S, RSD606159S, RSD608357S, RSD602185S, RSD605198S, RSD606239S, RSD606982S, RSD604938S, RSD602076S, RSD603445S, RSD603452S, RSD604423S, RSD606343S, RSD604370S, RSD605511S, RSD609295S, RSD604396S, RSD604998S, RSD605046S, RSD605070S, RSD605084S, RSD605153S, RSD605266S, RSD605276S, RSD605283S, RSD605285S, RSD605332S, RSD605401S, RSD605408S, RSD605416S, RSD605421S, RSD605461S, RSD605465S, RSD605474S, RSD605484S, RSD605495S, RSD605522S, RSD605536S, RSD605538S, RSD605554S, RSD605563S, RSD605567S, RSD605644S, RSD605647S, RSD605659S, RSD605667S, RSD605670S, RSD605821S, RSD605824S, RSD606114S, RSD606121S, RSD606134S, RSD606323S, RSD606355S, RSD606387S, RSD606390S, RSD606404S, RSD606409S, RSD606512S, RSD606519S, RSD606538S, RSD606544S, RSD606566S, RSD606594S, RSD607015S, RSD607187S, RSD607203S, RSD607211S, RSD607273S, RSD607348S, RSD607370S, RSD607379S, RSD607387S, RSD607421S, RSD607424S, RSD607425S, RSD607448S, RSD607461S, RSD607478S, RSD607490S, RSD607500S, RSD607509S, RSD607512S, RSD607545S, RSD607555S, RSD607686S, RSD607697S, RSD607698S, RSD607700S, RSD607704S, RSD607707S, RSD607713S, RSD607774S, RSD607842S, RSD607882S, RSD607884S, RSD607902S, RSD607904S, RSD607915S, RSD607926S, RSD607938S, RSD608170S, RSD608308S, RSD608312S, RSD608316S, RSD608317S, RSD608331S, RSD608358S, RSD608371S, RSD608384S, RSD608393S, RSD608405S, RSD608407S, RSD608409S, RSD608410S, RSD608415S, RSD608428S, RSD608436S, RSD608442S, RSD608444S, RSD608472S, RSD608473S, RSD608479S, RSD608509S, RSD608520S, RSD608522S, RSD608562S, RSD608566S, RSD608853S, RSD608876S, RSD609006S, RSD609008S, RSD609009S, RSD609029S, RSD609039S, RSD609068S, RSD609074S, RSD609076S, RSD609078S, RSD609105S, RSD609106S, RSD609108S, RSD609109S, RSD609110S, RSD609111S, RSD609112S, RSD609125S, RSD609126S, RSD609140S, RSD609147S, RSD609156S, RSD609157S, RSD609158S, RSD609161S, RSD609163S, RSD609168S, RSD609170S, RSD609171S, RSD609172S, RSD609177S, RSD609183S, RSD609185S, RSD609200S, RSD609202S, RSD609205S, RSD609211S, RSD609217S, RSD609218S, RSD609219S, RSD609226S, RSD609246S, RSD609256S, RSD609273S, RSD609317S, RSD609328S, RSD609329S, RSD609331S, RSD609332S, RSD609334S, RSD609335S, RSD609336S, RSD609337S, RSD609338S, RSD609339S, RSD609340S, RSD609342S, RSD609343S, RSD609344S, RSD609345S, RSD609346S, RSD609348S, RSD609349S, RSD609350S, RSD609351S, RSD609352S, RSD609353S, RSD609355S, RSD609357S, RSD609358S, RSD609359S, RSD609360S, RSD609361S, RSD609362S, RSD609363S, RSD609364S, RSD609367S, RSD609368S, RSD609371S, RSD609394S, RSD609397S, RSD609398S, RSD609399S, RSD609400S, RSD609401S, RSD609402S, RSD609404S, RSD609423S, RSD609424S, RSD609425S, RSD609426S, RSD609428S, RSD609429S, RSD609430S, RSD609431S, RSD609432S, RSD609433S, RSD609435S, RSD609436S, RSD609437S, RSD609438S, RSD609439S, RSD609440S, RSD609441S, RSD609442S, RSD609445S, RSD609446S, RSD609447S, RSD609448S, RSD609449S, RSD609450S, RSD609451S, RSD609452S, RSD609455S, RSD609458S, RSD609459S, RSD609460S, RSD609462S, RSD609463S, RSD609464S, RSD609465S, RSD609551S, RSD609554S, RSD609571S, RSD609572S, RSD609889S, RSD609951S, RSD609955S, RSD609970S, RSD609972S, RSD610242S, RSD610244S, RSD610245S, RSD610246S, RSD610248S, RSD610257S, RSD610285S, RSD610288S, RSD610305S, RSD610320S, RSD610331S, RSD610332S, RSD610333S, RSD610334S, RSD610478S, RSD610479S, RSD610484S, RSD610486S, RSD610492S, RSD610503S, RSD610504S, RSD610531S, RSD610552S, RSD610558S, RSD610560S, RSD610636S, RSD610643S, RSD610647S, RSD610650S, RSD610681S, RSD610686S, RSD610687S, RSD610688S, RSD610689S, RSD610691S, RSD610692S, RSD610701S, RSD610705S, RSD610707S, RSD610708S, RSD610712S, RSD610714S, RSD610740S, RSD610774S, RSD610776S, RSD610777S, RSD610788S, RSD610789S, RSD610795S, RSD610797S, RSD610801S, RSD610814S, RSD610816S, RSD610823S, RSD610824S, RSD610825S, RSD610834S, RSD610835S, RSD610838S, RSD610839S, RSD610843S, RSD610844S, RSD610845S, RSD610848S, RSD610849S, RSD610882S, RSD610919S, RSD610922S, RSD610934S, RSD610941S, RSD610963S, RSD610965S, RSD610966S, RSD610968S, RSD611013S, RSD611022S, RSD611054S, RSD611090S, RSD611095S, RSD611159S, RSD611160S, RSD611189S, RSD611191S, RSD611239S, RSD611247S, RSD603437S, RSD604279S, RSD604284S, RSD609993S, RSD610303S, RSD610762S, RSD610764S, RSD611194S, RSD611196S, RSD611198S, RSD611199S, RSD611202S, RSD611211S, RSD611213S, RSD611215S, RSD611241S, RSD611246S, RSD603621S, RSD604316S, RSD605691S, RSD606345S, RSD606353S, RSD607427S, RSD607721S, RSD608433S, RSD608447S, RSD608519S, RSD609173S, RSD609193S, RSD609294S, RSD609300S, RSD609444S, RSD609547S, RSD609566S, RSD610000S, RSD610250S, RSD610251S, RSD610282S, RSD610314S, RSD610330S, RSD610494S, RSD610495S, RSD610632S, RSD610682S, RSD610738S, RSD610784S, RSD610828S, RSD610995S, RSD611009S; GTIN 00763000380397, Lot Serial Numbers: RSD600738S, RSD601503S, RSD600649S, RSD605111S, RSD600735S, RSD601221S, RSD602436S, RSD601636S, RSD607763S, RSD601474S, RSD604654S, RSD605492S, RSD600675S, RSD600804S, RSD602148S, RSD600068S, RSD600066S, RSD600085S, RSD600757S, RSD601027S, RSD601275S, RSD604653S, RSD605575S, RSD601515S, RSD605090S, RSD604660S, RSD604842S, RSD603275S, RSD600670S, RSD604167S, RSD605438S, RSD600070S, RSD600079S, RSD600698S, RSD601194S, RSD601710S, RSD604255S, RSD605071S, RSD605616S, RSD600152S, RSD600637S, RSD602964S, RSD605592S, RSD600119S, RSD600073S, RSD600102S, RSD600131S, RSD600631S, RSD601702S, RSD601707S, RSD602125S, RSD604361S, RSD600706S, RSD604234S, RSD605077S, RSD602449S, RSD607254S, RSD600657S, RSD600673S, RSD601699S, RSD601768S, RSD605608S, RSD604642S, RSD604652S, RSD605126S, RSD605454S, RSD600642S, RSD601540S, RSD604837S, RSD600662S, RSD601755S, RSD604644S, RSD605462S, RSD601195S, RSD601703S, RSD604256S, RSD601267S, RSD601343S, RSD604391S, RSD604650S, RSD601048S, RSD601393S, RSD601781S, RSD602126S, RSD600648S, RSD601656S, RSD605497S, RSD605571S, RSD605648S, RSD601113S, RSD605010S, RSD602508S, RSD601090S, RSD600643S, RSD600110S, RSD600149S, RSD600651S, RSD600774S, RSD604826S, RSD600702S, RSD604401S, RSD604153S, RSD600122S, RSD600734S, RSD600740S, RSD603454S, RSD604146S, RSD600689S, RSD605663S, RSD600653S, RSD600704S, RSD601109S, RSD601281S, RSD600708S, RSD601118S, RSD602131S, RSD605086S, RSD600671S, RSD600799S, RSD600729S, RSD601084S, RSD604428S, RSD600147S, RSD600151S, RSD600694S, RSD600709S, RSD601023S, RSD603268S, RSD604561S, RSD605020S, RSD600162S, RSD607449S, RSD604426S, RSD601468S, RSD601675S, RSD602962S, RSD601046S, RSD605337S, RSD606607S, RSD601471S, RSD600677S, RSD600688S, RSD600086S, RSD601787S, RSD603092S, RSD600084S, RSD602506S, RSD604844S, RSD600692S, RSD601058S, RSD601697S, RSD604398S, RSD605584S, RSD601196S, RSD601708S, RSD604655S, RSD600730S, RSD600116S, RSD601704S, RSD602956S, RSD600117S, RSD600157S, RSD600632S, RSD600690S, RSD601648S, RSD601780S, RSD603472S, RSD606569S, RSD600082S, RSD600681S, RSD605597S, RSD605585S, RSD605609S, RSD608565S, RSD604204S, RSD605510S, RSD600723S, RSD601256S, RSD604592S, RSD601035S, RSD601479S, RSD604814S, RSD604392S, RSD600724S, RSD600173S, RSD600751S, RSD601238S, RSD604174S, RSD605117S, RSD605067S, RSD601049S, RSD600109S, RSD600166S, RSD600655S, RSD600719S, RSD601297S, RSD601492S, RSD604168S, RSD604410S, RSD605496S, RSD604657S, RSD600069S, RSD600104S, RSD600113S, RSD600160S, RSD600613S, RSD600674S, RSD600725S, RSD600752S, RSD600816S, RSD601038S, RSD601041S, RSD601465S, RSD601766S, RSD601776S, RSD603113S, RSD604262S, RSD604658S, RSD604845S, RSD605091S, RSD605373S, RSD605494S, RSD605498S, RSD600167S, RSD604402S, RSD605501S, RSD608536S, RSD600083S, RSD600111S, RSD601044S, RSD601223S, RSD601469S, RSD604829S, RSD601110S, RSD600710S, RSD600125S, RSD600714S, RSD601040S, RSD601688S, RSD605044S, RSD601698S, RSD600105S, RSD600678S, RSD600682S, RSD604285S, RSD604424S, RSD605611S, RSD600076S, RSD600155S, RSD600696S, RSD604425S, RSD600128S, RSD604408S, RSD606606S, RSD600742S, RSD600650S, RSD602133S, RSD603284S, RSD605457S, RSD604838S, RSD605019S, RSD602577S, RSD605610S, RSD609131S, RSD601693S, RSD600135S, RSD600749S, RSD601767S, RSD605572S, RSD605087S, RSD600050S, RSD601790S, RSD602159S, RSD605098S, RSD607453S, RSD600140S, RSD600747S, RSD600756S, RSD601510S, RSD601647S, RSD603450S, RSD601068S, RSD610846S, RSD603282S, RSD604641S, RSD604867S, RSD600141S, RSD601061S, RSD601671S, RSD604115S, RSD604646S, RSD604647S, RSD604648S, RSD600107S, RSD601037S, RSD601638S, RSD603435S, RSD604851S, RSD605129S, RSD605518S, RSD605517S, RSD601085S, RSD607465S, RSD600750S, RSD600144S, RSD600786S, RSD605673S, RSD600153S, RSD601777S, RSD601779S, RSD605537S, RSD607507S, RSD600101S, RSD600640S, RSD600676S, RSD600693S, RSD601783S, RSD604177S, RSD607768S, RSD601187S, RSD604821S, RSD605434S, RSD600054S, RSD603931S, RSD604863S, RSD600052S, RSD601024S, RSD600051S, RSD600056S, RSD605452S, RSD604109S, RSD600118S, RSD601120S, RSD601207S, RSD601623S, RSD604404S, RSD604832S, RSD605414S, RSD605459S, RSD600668S, RSD602507S, RSD604649S, RSD605411S, RSD600638S, RSD600665S, RSD603076S, RSD600072S, RSD600142S, RSD601383S, RSD608521S, RSD602442S, RSD604232S, RSD600126S, RSD600661S, RSD603264S, RSD604429S, RSD600687S, RSD600703S, RSD601253S, RSD600718S, RSD604417S, RSD602445S, RSD601093S, RSD604651S, RSD604835S, RSD601071S, RSD600053S, RSD607799S, RSD600658S, RSD604836S, RSD601464S, RSD601764S, RSD604664S, RSD605487S, RSD600660S, RSD601692S, RSD600699S, RSD600762S, RSD602950S, RSD603476S, RSD604034S, RSD605048S, RSD605074S, RSD605527S, RSD605589S, RSD605615S, RSD607705S, RSD600088S, RSD601029S, RSD600055S, RSD604237S, RSD605396S, RSD605491S, RSD601473S, RSD605101S, RSD605613S, RSD600627S, RSD601212S, RSD603468S, RSD605024S, RSD605075S, RSD602955S, RSD603921S, RSD604399S, RSD600731S, RSD603919S, RSD604419S, RSD602437S, RSD604289S, RSD605564S, RSD601231S, RSD600146S, RSD601051S, RSD604824S, RSD603495S, RSD600133S, RSD601057S, RSD607898S, RSD600626S, RSD604157S, RSD601234S, RSD605558S, RSD600663S, RSD600058S, RSD601526S, RSD604858S, RSD601488S, RSD605442S, RSD606557S, RSD607403S, RSD607456S, RSD605500S, RSD600680S, RSD601030S, RSD604420S, RSD600659S, RSD601055S, RSD601105S, RSD604198S, RSD606483S, RSD600077S, RSD604850S, RSD600726S, RSD604292S, RSD600683S, RSD609136S, RSD604240S, RSD604407S, RSD605009S, RSD605591S, RSD604421S, RSD600685S, RSD600791S, RSD603475S, RSD603206S, RSD605022S, RSD605049S, RSD600078S, RSD603143S, RSD601137S, RSD605596S, RSD609130S, RSD610737S
FEI Number 2182208
Recalling Firm/
Manufacturer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
8200 Coral Sea St Ne
Mounds View MN 55112-4391
For Additional Information ContactTechnical Services
800-929-4043
Manufacturer Reason
for Recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Quantity in Commerce9325 units
DistributionUS Nationwide - Worldwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA DatabasePMAs with Product Code = LWS
PMAs with Product Code = LWS
-
-