| | Class 1 Device Recall Teleflex |  |
| Date Initiated by Firm | May 25, 2023 |
| Date Posted | June 29, 2023 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1848-2023 |
| Recall Event ID |
92373 |
| 510(K)Number | K961837 |
| Product Classification |
Tube, tracheal (w/wo connector) - Product Code BTR
|
| Product | Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504565 |
| Code Information |
UDI/DI 4026704547609, Batch Numbers: 18FG05, 18FG13, 18FG25, 18GG20, 18HG07, 18HG12, 18HG19, 18HG32, 18HG38, 18IG06, 18IG09, 18IG15, 18JG04, 18JG21, 18JG34, 18KG35, 18LG39, 19AG01, 19AG24, 19AG27, 19AG37, 19CT32, 19CT41, 19CT58, 19CT65, 19DT06, 19DT20, 19DT33, 19ET32, 19ET54, 19ET60, 19FT24, 19FT41, 19GT28, 19GT38, 19HT10, 19HT66, 19IT14, 19IT25, 19JT18, 19JT29, 19JT42, 19JT57, 19KT08, 19KT41, 19KT55, 19LT32, 20AT09, 20AT14, 20AT25, 20AT53, 20BT27, 20CT05, 20CT12, 20CT36, 20CT44, 20DT09, 20ET04, 20ET16, 20GT14, KME20G0189, KME20G0316, KME20H2312, KME20H2585, KME20H2586, KME20H2587, KME20J0320, KME20J0570, KME20K0465, KME20K0939, KME20K1551, KME20K1628, KME20K2246, KME20L0488, KME20L1840, KME20M0022, KME20M0854, KME20M1102, KME20M1579, KME21K1128, KME21L0328, KME21L2340, KME21M2283, KME21M2446, KME21M2568, KME22A0675, KME22A2415, KME22A3028, KME22A3156, KME22A3313, KME22B0067, KME22B1218, KME22B1409, KME22C0499, KME22C0863, KME22C1073, KME22C1225, KME22C1655, KME22C1941, KME22D0747, KME22D0755, KME22F1616, KME22F1617, KME22G0632, KME22G1224, KME22G2467, KME22H1900, KME22H3348, KME22J1135, KME22J1136, KME22J1928, KME22J3352, KME22K2413, KME22K2414, KME22L0241, KME22L0242, KME22L1792 |
| FEI Number |
3005747797
|
Recalling Firm/ Manufacturer |
TELEFLEX LLC 3015 Carrington Mill Blvd Morrisville NC 27560-5437
|
Manufacturer Reason for Recall | Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Teleflex issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 05/25/2012 by FedEx. The notice explained the problem, the risk, and requested the following: Users should cease use and distribution of affected product and immediately quarantine the affected product. Distributors were directed to cease use, distribution, and quarantine the product, and to notify their customers.
Teleflex is seeking the return of the affected product.
For further information:
Contact: Customer Service, Telephone: 1-866-396-2111
FAX: 1-855-419-8507, Email: Recalls@teleflex.com |
| Quantity in Commerce | 92930 units |
| Distribution | US Nationwide distribution including Puerto Rico. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|