| | Class 2 Device Recall |  |
| Date Initiated by Firm | May 15, 2023 |
| Date Posted | August 03, 2023 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2299-2023 |
| Recall Event ID |
92399 |
| Product Classification |
Cardiac catheterization kit - Product Code OES
|
| Product | Custom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following:
a. MULTI-LUMEN CVC BUNDLE KIT, Model Number: ECVC475;
b. MULTI-LUMEN CVC INSERTION BUNDLE, Model Number: ECVC3900;
c. MULTIMED 16G 1L 20CM CVC STOCK+BUNDLE US, Model Number: STCVC20US;
d. MULTIMED 7F 2L 16CM STOCK + BUNDLE US, Model Number: STCVC21US;
e. MULTIMED 7F 3L 16CM CVC BUNDLE, Model Number: ECVC6585A;
f. MULTIMED 7F 3L 20CM PI, Model Number: ECVC7235A;
g. MULTIMED 7FR 16CM CVC STK + BUNDLE-US, Model Number: STCVC03US;
h. MULTIMED 7FR 3L 16CM CVC BUNDLE, Model Number: ECVC6655A;
i. MULTIMED 7FR 3L 20CM CVC BUNDLE-US, Model Number: ECVC7935;
j. MULTIMED 7FR 3L 20CM CVC STK + BUNDLE-US, Model Number: STCVC04US;
k. MULTIMED 7FR, 20CM, 3L CVC BUNDLE, Model Number: ECVC7670;
l. MULTIMED 8.5F 20CM QUAD CVC BUNDLE, Model Number: ECVC7090;
m. MULTIMED 8.5FR 16CM QUAD CVC BUNDLE, Model Number: ECVC6930;
n. MULTIMED 8.5FR 2L 16CM CVC STK+BUNDLE-US, Model Number: STCVC13US;
o. NEURO CATH PACK (NCLUA)227-LF, Model Number: DYNJ47646D, DYNJ47646F, DYNJ47646G, DYNJ47646I;
p. O.R. DOUBLE LUMEN INSERTION TRAY, Model Number: ECVC8010;
q. OLEAN GENERAL CVC BUNDLE, Model Number: ECVC4615;
r. OR ANGIO PACK, Model Number: DYNJ59184B;
s. OR ANGIOGRAM, Model Number: DYNJ909267;
t. OR ANGIOGRAPHY DRAPE, Model Number: DYNJ906297D;
u. OR ANGIOGRAPHY PACK, Model Numbers: DYNJ67245A, DYNJ67245B;
v. OR PERCUTANEOUS CATH PACK, Model Number: DYNJ63372A, DYNJ63372B, DYNJ63372C, DYNJ63372D;
w. OR VANTEX 7FR 3L 16CM CVC, Model Number: ECVC6665;
x. OR VASCULAR ANGIO PACK, Model Number: DYNJ69763, DYNJ69763A;
y. PACK CARDIAC CATH, Model Number: DYNJ65481;
z. PACK PERCUTANEOUS ANGIO CHRG, Model Number: DYNJ60112B;
aa. PEDIATRIC CATH ANGIO PACK, Model Number: DYNJ54755G, DYNJ54755I;
bb. PEDIATRIC CATH PACK-LF, Model Number: DYNJ0843244K;
cc. PERCUTANEOUS ANGIO PACK-LF, Model Number: DYNJ57645C;
dd. PERCUTANEOUS PACK, Model Number: DYNJ46035C, DYNJ46035D;
ee. PERMA CATH PACK, Model Number: DYNJ62007A;
ff. PI CVC BUNDLE, VANTEX 3L, 16CM, Model Number: ECVC3665;
gg. PI CVC BUNDLE, VANTEX 3L, 20CM, Model Number: ECVC6190;
hh. PI VANTEX 7FR 16CM 3L CVC BUNDLE-MAXPLUS, Model Number: ECVC4380;
ii. PI VANTEX 7FR 16CM 3L CVC-MicroClave, Model Number: ECVC5655;
jj. PI VANTEX 7FR 20CM 3L CVC-MicroClave, Model Number: ECVC5650;
kk. PI VANTEX 8.5FR 16CM 4L CVC-MicroClave, Model Number: ECVC5660;
ll. PI VANTEX 8.5FR 20CM 4L CVC-MicroClave, Model Number: ECVC5685;
mm. PICC LINE PACK-LF, Model Number: DYNJ0275614F;
nn. PK, RADIOLOGY, Model Number: DYNJ43894B;
oo. PORT A CATH PACK, Model Number: DYNJ59307F;
pp. PORT PACK, Model Number: DYNJ58162;
qq. PORTACATH PACK, Model Number: DYNJ51240A;
rr. QUAD LUMEN CVC BUNDLE 8.5F, 16CM, Model Number: ECVC3700;
ss. QUAD LUMEN CVC BUNDLE, 8.5F, 20CM, Model Number: ECVC4995;
tt. QUAD LUMEN CVC INSERTION BUNDLE, Model Numbers: ECVC5025, ECVC4135;
uu. QUAD LUMEN KIT, Model Number: ECVC5820;
vv. RADIOFREQUENCY ABLATION PACK, Model Number: DYNJ65913;
ww. RADIOLOGY PACK, Model Numbers: DYNJ41190, DYNJ43894A, DYNJ44087A, DYNJ44524F, DYNJ52806C;
xx. RADIOLOGY PACK 108402-LF, Model Number: DYNJVB1000;
yy. RR-ADULT ECMO PACK, Model Number: DYNJ69742, DYNJ69742B;
zz. SHEATH INTRODUCER BUNDLE, Model Number: ECVC2930, ECVC5835;
aaa. SK-ANGIO TRAY-LF, Model Number: DYNJ82030;
bbb. SK-INTERVENTIONAL TRAY-LF, Model Number: DYNJ82041;
ccc. SPECIAL PROCEDURE CATH LAB PK, Model Number: DYNJ68729;
ddd. SPECIAL PROCEDURE PACK, Model Number: DYNJ38314G;
eee. SPECIAL PROCEDURE PACK-LF, Model Number: DYNJ0416535AA, DYNJ0416535W, DYNJ0458661C;
fff. SPECIAL PROCEDURE PK-LF, Model Number: DYNJ0604508R;
ggg. SPECIAL PROCEDURES, Model Number: DYNJ23288C;
hhh. STANDARD ANGIO TRAY MV, Model Number: DYNJ67136A;
iii. STD, AVA ,3XI ,3L 8.5F KIC, Model Number: ECVC7465A;
jjj. STD, AVA, 3L, 9F, HF, KIC, Model Number: ECVC810 |
| Code Information |
All lot numbers for the following models and corresponding UPCs distributed from Medline between 04/01/2020 and 04/28/2023:
a. Model Number: ECVC475, UPC Number: 10653160333081.
b. Model Number: ECVC3900, UPC Number: 10653160334828.
c. Model Number: STCVC20US, UPC Number: 10653160339076.
d. Model Number: STCVC21US, UPC Number: 10653160339045.
e. Model Number: ECVC6585A, UPC Number: 10653160338826.
f. Model Number: ECVC7235A, UPC Number: 10653160354550.
g. Model Number: STCVC03US, UPC Number: 10653160339212.
h. Model Number: ECVC6655A, UPC Number:10653160339267.
i. Model Number: ECVC7935, UPC Number: 10653160353669.
j. Model Number: STCVC04US, UPC Number: 10653160339199.
k. Model Number: ECVC7670, UPC Number:10653160348054.
l. Model Number: ECVC7090, UPC Number: 10653160339243.
m. Model Number: ECVC6930, UPC Number:10653160339250.
n. Model Number: STCVC13US, UPC Number: 10653160339137.
o. Model Number: DYNJ47646D, UPC Number: 019348995047; Model Number: DYNJ47646F, UPC Number: 019532715551; Model Number: DYNJ47646G, UPC Number: 019532717502; Model Number: DYNJ47646I, UPC Number: 019532735712.
p. Model Number: ECVC8010, UPC Number: 10653160356356.
q. Model Number: ECVC4615, UPC Number: 10653160336075.
r. Model Number: DYNJ59184B, UPC Number: 019348994910.
s. Model Number: DYNJ909267, UPC Number: 019532731461.
t. Model Number: DYNJ906297D, UPC Number: 019532714128.
u. Model Number: DYNJ67245A, UPC Number: 019532702425; Model Number: DYNJ67245B, UPC Number: 019532729222.
v. Model Number: DYNJ63372A, UPC Number: 019348929205; Model Number: DYNJ63372B, UPC Number: 019348947044; Model Number: DYNJ63372C, UPC Number: 019532708614; Model Number: DYNJ63372D, UPC Number: 019532727047.
w. Model Number: ECVC6665, UPC Number: 10653160337843.
x. Model Number: DYNJ69763, UPC Number: 019348997737; Model Number: DYNJ69763A, UPC Number: 019532733001.
y. Model Number: DYNJ65481, UPC Number: 019348928046.
z. Model Number: DYNJ60112B, UPC Number: 019532711884.
aa. Model Number: DYNJ54755G, UPC Number: 019348998776; Model Number: DYNJ54755I, UPC Number: 019532702126.
bb. Model Number: DYNJ0843244K, UPC Number: 019532728859.
cc. Model Number: DYNJ57645C, UPC Number:019532728741.
dd. Model Number: DYNJ46035C, UPC Number: 019348934095; Model Number: DYNJ46035D, UPC Number: 019348949364.
ee. Model Number: DYNJ62007A, UPC Number: 019348987890.
ff. Model Number: ECVC3665, UPC Number: 10653160336044.
gg. Model Number: ECVC6190, UPC Number: 10653160337751.
hh. Model Number: ECVC4380, UPC Number: 10653160338147.
ii. Model Number: ECVC5655, UPC Number: 10653160337621.
jj. Model Number: ECVC5650, UPC Number: 10653160336457.
kk. Model Number: ECVC5660, UPC Number: 10653160337638.
ll. Model Number: ECVC5685, UPC Number: 10653160337652.
mm. Model Number: DYNJ0275614F, UPC Number: 088994217434.
nn. Model Number: DYNJ43894B, UPC Number: 019348981048.
oo. Model Number: DYNJ59307F, UPC Number: 019532705885.
pp. Model Number: DYNJ58162, UPC Number: 088994282933.
qq. Model Number: DYNJ51240A, UPC Number: 088994249352.
rr. Model Number: ECVC3700, UPC Number: 10653160332886.
ss. Model Number: ECVC4995, UPC Number: 10653160335283.
tt. Model Number: ECVC5025, UPC Number: 10653160333142; Model Number: ECVC4135, UPC Number: 10653160335900.
uu. Model Number: ECVC5820, UPC Number: 10653160338123.
vv. Model Number: DYNJ65913, UPC Number: 019348930317.
ww. Model Number: DYNJ41190, UPC Number: 088438989908; Model Number: DYNJ43894A, UPC Number: 088994213269; Model Number: DYNJ44087A, UPC Number: 088994256180; Model Number: DYNJ44524F, UPC Number: 019348946748; Model Number: DYNJ52806C, UPC Number: 019348929290.
xx. Model Number: DYNJVB1000, UPC Number: 019348980099.
yy. Model Number: DYNJ69742, UPC Number: 019348997490; Model Number: DYNJ69742B, UPC Number: 019532713688.
zz. Model Number: ECVC2930, UPC Number: 10653160335368; Model Number: ECVC5835, UPC Number: 10653160333364.
aaa. Model Number: DYNJ82030, UPC Number: 019532716237.
bbb. Model Number: DYNJ82041, UPC Number: 019532716221.
ccc. Model Number: DYNJ68729, UPC Number: 019348983654.
ddd. Model Number: DYNJ38314G, UPC Number: 019348960334.
eee. Model Number: DYNJ0416535AA, UPC Number: 019532735459; Model Number: DYNJ0416535W, UPC Number: 019348956435; Model Number: DYNJ0458661C, UPC Number: 088994255094.
fff. Model Number: DYNJ0604508R, UPC Number: 019348982844.
ggg. Model Number: DYNJ23288C, UPC Number: 019348995414.
hhh. Model Number: DYNJ67136A, UPC Number: 019348993567.
iii. Model Number: ECVC7465A, UPC Number: 10653160350088.
jjj. Model Number: ECVC8100, UPC Number: 10653160357841.
kkk. Model Number: DYNJ27613J, UPC Number: 019348956435; Model Number: DYNJ27613K, UPC Number: 019532713118.
lll. Model Number: DYNJ80925, UPC Number: 019532708511.
mmm. Model Number: DYNJ52425J, UPC Number: 019348994014.
nnn. Model Number: DYNJ83251, UPC Number: 019532724865.
ooo. Model Number: ECVC7310, UPC Number: 10653160345060; Model Number: ECVC7310A, UPC Number: 10653160352792.
ppp. Model Number: ECVC2965, UPC Number: 10653160338611; Model Number: ECVC3155, UPC Number: 10653160334132; Model Number: ECVC5475, UPC Number: 10653160338116.
qqq. Model Number: ECVC5015, UPC Number: 10653160338956.
rrr. Model Number: ECVC3395, UPC Number: 10653160338635.
sss. Model Number: ECVC5810, UPC Number: 10653160336419.
ttt. Model Number: ECVC255, UPC Number: 10653160331735; Model Number: ECVC5200, UPC Number: 10653160336150.
uuu. Model Number: ECVC825A, UPC Number: 10653160333548.
vvv. Model Number: ECVC7180, UPC Number: 10653160343165.
www. Model Number: ECVC7425, UPC Number: 10653160347699; Model Number: STCVC2040, UPC Number: 10653160336785; Model Number: STCVC2040A, UPC Number: 10653160345701.
xxx. Model Number: ECVC7290, UPC Number: 10653160344933.
yyy. Model Number: ECVC5355B, UPC Number: 10653160357742.
zzz. Model Number: ECVC6020A, UPC Number: 10653160338413.
aaaa. Model Number: ECVC1910, UPC Number:10653160334453; Model Number: ECVC5870A, UPC Number:10653160338383.
bbbb. Model Number: ECVC6055A, UPC Number: 10653160335849.
cccc. Model Number: ECVC6540, UPC Number: 10653160339397.
dddd. Model Number: ECVC3735A, UPC Number: 10653160334798.
eeee. Model Number: ECVC3355, UPC Number: 10653160333982; Model Number: ECVC3400, UPC Number: 10653160340591; Model Number: ECVC3460, UPC Number: 10653160332794; Model Number: ECVC3785, UPC Number: 10653160336006; Model Number: ECVC3810, UPC Number: 10653160336013; Model Number: ECVC4880, UPC Number: 10653160333111; Model Number: ECVC5020, UPC Number: 10653160334903; Model Number: ECVC5020A, UPC Number: 10653160358251; Model Number: ECVC5125, UPC Number: 10653160334934; Model Number: ECVC5270, UPC Number: 10653160336198; Model Number: ECVC6395, UPC Number: 10653160336372.
ffff. Model Number: ECVC1490, UPC Number: 10653160332183; Model Number: ECVC1490A, UPC Number: 10653160359036; Model Number: ECVC1495, UPC Number: 10653160334569; Model Number: ECVC1495A, UPC Number: 10653160358794.
gggg. Model Number: ECVC1710, UPC Number: 10653160332268.
hhhh. Model Number: ECVC5815, UPC Number: 10653160336402.
iiii. Model Number: DYNJ67990A, UPC Number: 019532726533.
jjjj.Model Number: DYNJ41562F, UPC Number: 019348964108.
kkkk. Model Number: DYNJ41562D, UPC Number: 019348958812.
llll. Model Number: ECVC1655, UPC Number: 10653160254935.
mmmm. Model Number: ECVC7170, UPC Number: 10653160333524.
nnnn. Model Number: ECVC6680, UPC Number: 10653160337867; Model Number: ECVC6680A, UPC Number: 10653160337164.
oooo. Model Number: ECVC6675, UPC Number: 10653160337850; Model Number: ECVC6675A, UPC Number: 10653160336617.
pppp. Model Number: ECVC7970, UPC Number: 10653160353980.
qqqq. Model Number: ECVC6815, UPC Number: 10653160335115.
rrrr. Model Number: ECVC6840, UPC Number: 10653160336761.
ssss. Model Number: ECVC6835, UPC Number: 10653160336778.
tttt. Model Number: ECVC7485, UPC Number: 10653160354802.
uuuu. Model Number: ECVC7615, UPC Number: 10653160348009; Model Number: ECVC7690, UPC Number: 10653160348153.
vvvv. Model Number: ECVC6630, UPC Number: 10653160337492.
wwww. Model Number: ECVC6690, UPC Number: 10653160336600.
xxxx. Model Number: ECVC8135, UPC Number: 10653160358558; Model Number: ECVC8160, UPC Number: 10653160358404.
yyyy. Model Number: ECVC7915, UPC Number: 10653160353355.
zzzz. Model Number: ECVC6550, UPC Number: 10653160335276; Model Number: ECVC6595, UPC Number: 10653160337485; Model Number: ECVC7115B, UPC Number: 10653160350279; Model Number: ECVC7275, UPC Number: 10653160342700; Model Number: ECVC7275A, UPC Number: 10653160344292; Model Number: ECVC8015, UPC Number: 10653160355366.
aaaaa. Model Number: ECVC7400, UPC Number: 10193489199956; Model Number: ECVC7960, UPC Number: 10653160353881; Model Number: ECVC8050, UPC Number: 10653160355601.
bbbbb. Model Number: ECVC7945, UPC Number: 10653160353966.
ccccc. Model Number: ECVC7625, UPC Number: 10653160356295.
ddddd. Model Number: ECVC7895, UPC Number: 10653160355328.
eeeee. Model Number: STCVC02US, UPC Number: 10653160336815; Model Number: STCVC02UST, UPC Number: 10653160358589.
fffff. Model Number: ECVC7405, UPC Number: 10653160347132.
ggggg. Model Number: ECVC7295, UPC Number: 10653160344650.
hhhhh. Model Number: ECVC6765, UPC Number: 10653160336839.
iiiii. Model Number: ECVC7495B, UPC Number: 10653160355748.
jjjjj. Model Number: ECVC6505, UPC Number: 10653160336327; Model Number: ECVC7300, UPC Number: 10653160344698.
kkkkk. Model Number: ECVC7560, UPC Number: 10653160347705; Model Number: ECVC7955, UPC Number: 10653160353867; Model Number: ECVC8020, UPC Number: 10653160355458; Model Number: ECVC8045, UPC Number: 10653160355595; Model Number: ECVC8085, UPC Number: 10653160356905.
lllll. Model Number: ECVC8025, UPC Number: 10653160355434.
mmmmm. Model Number: ECVC8040, UPC Number: 10653160355700.
nnnnn. Model Number: ECVC7565, UPC Number: 10653160347682.
ooooo. Model Number: ECVC7830A, UPC Number: 10653160355342.
ppppp. Model Number: STCVC01US, UPC Number: 10653160336846; Model Number: STCVC01UST, UPC Number: 10653160358565.
qqqqq. Model Number: ECVC7630, UPC Number: 10653160347750.
rrrrr. Model Number: ECVC6050A, UPC Number: 10653160335894; Model Number: ECVC6670, UPC Number: 10653160336624; Model Number: ECVC7660, UPC Number: 10653160347804.
sssss. Model Number: ECVC7385, UPC Number: 10653160346210; Model Number: ECVC7385A, UPC Number: 10653160347392; Model Number: ECVC7700, UPC Number: 10653160348375.
ttttt. Model Number: ECVC5595, UPC Number: 10653160336495.
uuuuu. Model Number: ECVC6800, UPC Number: 10653160335092.
vvvvv. Model Number: ECVC7785, UPC Number: 10653160349365.
wwwww. Model Number: ECVC6825A, UPC Number: 10653160336631.
xxxxx. Model Number: ECVC7505A, UPC Number: 10653160352853.
yyyyy. Model Number: ECVC7835A, UPC Number: 10653160355359.
zzzzz. Model Number: ECVC8080, UPC Number: 10653160356899.
aaaaaa. Model Number: STCVC06US, UPC Number: 10653160338314.
bbbbbb. Model Number: ECVC1320, UPC Number: 10653160335795.
cccccc. Model Number: ECVC6255, UPC Number: 10653160336525.
dddddd. Model Number: ECVC6260, UPC Number: 10653160336488.
eeeeee. Model Number: ECVC5080, UPC Number: 10653160337690.
ffffff. Model Number: ECVC6920, UPC Number: 10653160336716; Model Number: ECVC7000, UPC Number: 10193489196399.
gggggg. Model Number: ECVC6915, UPC Number: 10653160336723; Model Number: ECVC6975, UPC Number: 10653160336679.
hhhhhh. Model Number: ECVC4105, UPC Number: 10653160334835.
iiiiii. Model Number: ECVC6340A, UPC Number: 10653160355038.
jjjjjj. Model Number: DYNJ0429121K, UPC Number: 019532706950.
kkkkkk. Model Number: ECVC7330, UPC Number: 10653160345565. |
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
MEDLINE INDUSTRIES, LP - Northfield 3 Lakes Dr Northfield IL 60093-2753
|
| For Additional Information Contact | Karin Johnson 886-359-1704 |
Manufacturer Reason for Recall | Potential for the labeled sterile ultrasound gel component within packs to not meet sterility specifications. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Medline notified customers beginning on 05/15/2023, via mail and email with letter. Customers with Medline Probe Cover Kits were instructed to check inventory for affected kits, destroy any affected kits, and complete and return the response form. Distributors were instructed to notify their customers and have them document and destroy affected kits. Customers with the non-sterile and sterile procedural kits were instructed to quarantine kits, complete and return the response form and, once updated labels are received, over label and discard the affected ultrasound gel product within the kits. Distributors were instructed to notify customers of the recall and have customers return any affected product for return to Medline.
Medline identified additional units of Medline Probe Cover Kits and notified customers on 06/22/2023. Customers were instructed to quarantine kits, complete and return the response form and, once updated labels are received, over label and discard the affected ultrasound gel product within the kits. |
| Quantity in Commerce | 104,755 kits |
| Distribution | Worldwide distribution - US Nationwide and the countries of Canada, China, Costa Rica, India, Lebanon, Panama, Qatar, Saudi Arabia, Singapore, United Arab Emirates. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|