• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Heartware

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Heartwaresee related information
Date Initiated by FirmOctober 31, 2023
Date PostedDecember 02, 2023
Recall Status1 Open3, Classified
Recall NumberZ-0479-2024
Recall Event ID 93405
Product Classification Ventricular (assist) bypass - Product Code DSQ
ProductPUMP 1104 HVAD (HeartWare Ventricular Assist Device) IMPLANT KIT, Model Number 1104
Code Information GTIN none, Serial Numbers: HW10001, HW10002, HW10003, HW10004, HW10005, HW10006, HW10007, HW10008, HW10009, HW10010, HW10011, HW10012, HW10013, HW10014, HW10015, HW10016, HW10017, HW10018, HW10019, HW10020, HW10021, HW10022, HW10023, HW10024, HW10025, HW10026, HW10027, HW10028, HW10029, HW10030, HW10031, HW10032, HW10033, HW10034, HW10035, HW10036, HW10037, HW10038, HW10039, HW10040, HW10041, HW10042, HW10043, HW10044, HW10045, HW10046, HW10047, HW10048, HW10049, HW10050, HW10051, HW10052, HW10053, HW10055, HW10056, HW10057, HW10058, HW10059, HW10060, HW10061, HW10062, HW10063, HW10064, HW10065, HW10066, HW10067, HW10068, HW10070, HW10071, HW10072, HW10073, HW10074, HW10075, HW10076, HW10077, HW10079, HW10080, HW10081, HW10082, HW10083, HW10084, HW10085, HW10086, HW10087, HW10088, HW10089, HW10090, HW10091, HW10092, HW10093, HW10094, HW10095, HW10096, HW10097, HW10098, HW10099, HW10100, HW10101, HW10102, HW10103, HW10104, HW10105, HW10106, HW10107, HW10108, HW10109, HW10110, HW10111, HW10112, HW10113, HW10114, HW10115, HW10116, HW10117, HW10118, HW10119, HW10120, HW10121, HW10122, HW10123, HW10124, HW10125, HW10126, HW10127, HW10128, HW10129, HW10130, HW10131, HW10132, HW10133, HW10135, HW10136, HW10137, HW10138, HW10139, HW10140, HW10141, HW10142, HW10143, HW10144, HW10145, HW10146, HW10147, HW10148, HW10149, HW10150, HW10151, HW10152, HW10153, HW10154, HW10155, HW10156, HW10157, HW10158, HW10159, HW10160, HW10161, HW10162, HW10163, HW10164, HW10165, HW10166, HW10167, HW10168, HW10169, HW10170, HW10171, HW10172, HW10173, HW10174, HW10175, HW10176, HW10177, hw10178, HW10179, HW10180, HW10181, HW10182, HW10183, HW10184, HW10185, HW10186, HW10187, HW10188, HW10189, HW10190, HW10191, HW10192, HW10193, HW10194, HW10195, HW10196, HW10197, HW10198, HW10200, HW10201, HW10202, HW10203, HW10204, HW10205, HW10206, HW10207, HW10209, HW10210, HW10211, HW10212, HW10213, HW10214, HW10215, HW10216, HW10217, HW10218, HW10219, HW10220, HW10221, HW10222, hw10223, HW10224, HW10225, HW10226, HW10227, HW10228, HW10229, HW10230, HW10231, HW10232, HW10233, HW10234, HW10235, HW10236, HW10237, HW10238, HW10239, HW10240, HW10241, HW10242, HW10243, HW10244, HW10245, HW10246, HW10247, HW10248, HW10249, HW10250, HW10251, HW10252, HW10253, HW10254, HW10255, HW10256, HW10257, HW10258, HW10259, HW10260, HW10261, HW10262, HW10263, HW10264, HW10265, HW10266, HW10267, HW10268, HW10269, HW10270, HW10271, HW10272, HW10273, HW10274, HW10275, HW10276, HW10277, HW10278, HW10279, HW10280, HW10281, HW10282, HW10283, HW10284, HW10285, HW10286, HW10287, HW10288, HW10289, HW10290, HW10291, HW10292, HW10293, HW10294, HW10295, HW10296, HW10297, HW10299, HW10300, HW10301, HW10304, HW10305, HW10306, HW10307, HW10308, HW10309, HW10310, HW10311, HW10312, HW10313, HW10314, HW10316, HW10318, HW10319, HW10321, HW10322, HW10323, HW10324, HW10325, HW10326, HW10327, HW10328, HW10329, HW10330, HW10331, HW10332, HW10333, HW10334, HW10335, HW10336, HW10337, HW10338, HW10339, HW10340, HW10341, HW10342, HW10343, HW10344, HW10345, HW10346, HW10347, HW10348, HW10349, HW10350, HW10351, HW10352, HW10353, HW10354, HW10355, HW10356, HW10357, HW10358, HW10360, HW10361, HW10362, HW10363, HW10364, HW10365, HW10366, HW10367, HW10368, HW10369, HW10370, HW10371, HW10372, HW10373, HW10374, HW10375, HW10376, HW10377, HW10378, HW10379, HW10380, HW10381, HW10382, HW10383, HW10384, HW10385, HW10386, HW10387, HW10389, HW10390, HW10391, HW10392, HW10393, HW10394, HW10395, HW10397, HW10398, HW10399, HW10400, HW10401, HW10402, HW10403, HW10404, HW10405, HW10406, HW10407, HW10408, HW10409, HW10410, HW10411, HW10412, HW10413, HW10414, HW10415, HW10416, HW10417, HW10418, HW10419, HW10420, HW10421, HW10422, HW10423, HW10424, HW10425, HW10426, HW10427, HW10428, HW10429, HW10430, HW10431, HW10432, HW10433, HW10434, HW10435, HW10436, HW10437, HW10438, HW10439, HW10440, HW10441, HW10442, HW10443, HW10444, HW10445, HW10446, HW10447, HW10448, HW10449, HW10451, HW10452, HW10453, HW10454, HW10456, HW10457, HW10458, HW10459, HW10460, HW10461, HW10462, HW10463, HW10464, HW10465, HW10466, HW10467, HW10468, HW10469, HW10470, HW10471, HW10472, HW10473, HW10474, HW10475, HW10476, HW10477, HW10478, HW10479, HW10480, HW10481, HW10482, HW10483, HW10484, HW10485, HW10486, HW10487, HW10488, HW10489, HW10490, HW10491, HW10492, HW10493, HW10494, HW10495, HW10496, HW10497, HW10498, HW10499, HW10500, HW10513, HW10514, HW10515, HW10516, HW10517, HW10518, HW10519, HW10520, HW10521, HW10522, HW10523, HW10524, HW10526, HW10527, HW10528, HW10529, HW10530, HW10531, HW10532, HW10533, HW10534, HW10535, HW10536, HW10537, HW10538, HW10539, HW10540, HW10541, HW10542, HW10543, HW10544, HW10545, HW10546, HW10547, HW10548, HW10549, HW10550, HW10551, HW10552, HW10553, HW10554, HW10555, HW10556, HW10557, HW10558, HW10559, HW10560, HW10561, HW10562, HW10563, HW10564, HW10565, HW10566, HW10567, HW10568, HW10569, HW10570, HW10571, HW10572, HW10573, HW10574, HW10575, HW10577, HW10578, HW10579, HW10580, HW10581, HW10582, HW10583, HW10584, HW10585, HW10586, HW10587, HW10589, HW10590, HW10591, HW10592, HW10593, HW10594, HW10595, HW10596, HW10597, HW10598, HW10599, HW10600, HW10601, HW10602, HW10603, HW10604, HW10605, HW10606, HW10607, HW10608, HW10609, HW10610, HW10611, HW10612, HW10613, HW10614, HW10615, HW10616, HW10617, HW10618, HW10619, HW10620, HW10621, HW10622, HW10623, HW10624, HW10625, HW10626, HW10627, HW10628, HW10629, HW10630, HW10631, HW10632, HW10633, HW10635, HW10638, HW10639, HW10640, HW10641, HW10642, HW10643, HW10644, HW10645, HW10646, HW10647, HW10648, HW10650, HW10651, HW10652, HW10653, HW10654, HW10655, HW10656, HW10657, HW10658, HW10659, HW10660, HW10661, HW10662, HW10663, HW10664, HW10665, HW10666, HW10667, HW10668, HW10669, HW10670, HW10671, HW10672, HW10673, HW10674, HW10675, HW10676, HW10677, HW10678, HW10679, HW10680, HW10681, HW10682, HW10683, HW10684, HW10685, HW10688, HW10689, HW10690, HW10691, HW10692, HW10693, HW10694, HW10695, HW10696, HW10697, HW10698, HW10699, HW10700, HW10701, HW10702, HW10703, HW10704, HW10705, HW10706, HW10707, HW10708, HW10709, HW10710, HW10711, HW10712, HW10713, HW10714, HW10715, HW10716, HW10717, HW10718, HW10719, HW10720, HW10721, HW10722, HW10723, HW10724, HW10725, HW10726, HW10727, HW10728, HW10729, HW10730, HW10731, HW10732, HW10733, HW10734, HW10735, HW10736, HW10737, HW10738, HW10739, HW10740, HW10741, HW10742, hw10743, HW10744, HW10745, HW10746, HW10747, HW10748, HW10749, HW10750, HW10751, HW10752, HW10753, HW10754, HW10755, HW10756, HW10757, HW10758, HW10759, HW10760, HW10761, HW10762, HW10763, HW10764, HW10765, HW10766, HW10767, HW10768, HW10769, HW10770, HW10771, HW10772, HW10773, HW10774, HW10775, HW10776, HW10777, HW10778, HW10779, HW10780, HW10781, HW10782, HW10783, HW10784, HW10785, HW10786, HW10787, HW10788, HW10789, HW10790, HW10791, HW10792, HW10793, HW10794, HW10795, HW10796, HW10797, HW10798, HW10799, HW10800, HW10801, HW10802, HW10803, HW10804, HW10805, HW10806, HW10807, HW10808, HW10809, HW10810, HW10811, HW10812, HW10813, HW10814, HW10815, HW10816, HW10817, HW10818, HW10819, HW10820, HW10821, HW10822, HW10823, HW10824, HW10825, HW10826, HW10827, HW10828, HW10829, HW10830, HW10831, HW10832, HW10833, HW10834, HW10835, HW10836, HW10837, HW10838, HW10839, HW10840, HW10841, HW10842, HW10843, HW10844, HW10845, HW10846, HW10848, HW10849, HW10850, HW10851, HW10852, HW10853, HW10854, HW10855, HW10856, HW10857, HW10858, HW10859, HW10860, HW10861, HW10862, HW10863, HW10864, HW10865, HW10867, HW10868, HW10869, HW10870, HW10871, HW10872, HW10873, HW10874, HW10875, HW10876, HW10877, HW10879, HW10880, HW10881, HW10882, HW10883, HW10884, HW10885, HW10886, HW10887, HW10888, HW10889, HW10890, HW10891, HW10892, HW10893, HW10894, HW10895, HW10896, HW10897, HW10898, HW10899, HW10900, HW10901, HW10902, HW10903, HW10904, HW10905, HW10906, HW10907, HW10908, HW10909, HW10910, HW10911, HW10912, HW10913, HW10915, HW10916, HW10917, HW10918, HW10919, HW10920, hw10921, HW10922, HW10923, HW10924, HW10925, HW10926, HW10927, HW10928, HW10929, HW10930, HW10931, HW10932, HW10933, HW10934, HW10935, HW10936, HW10937, HW10938, HW10939, HW10940, HW10941, hw10942, HW10944, HW10945, HW10946, HW10947, HW10948, HW10949, HW10950, HW10951, HW10952, HW10953, HW10954, HW10955, HW10956, HW10957, HW10958, HW10959, HW10960, HW10961, HW10962, HW10963, HW10965, HW10966, HW10967, HW10968, HW10969, HW10970, HW10972, HW10973, HW10974, HW10975, HW10978, HW10979, HW10980, HW10981, HW10982, HW10983, HW10984, HW10985, HW10986, HW10987, HW10988, HW10989, HW10990, HW10992, HW10993, HW10994, HW10995, HW10997, HW10998, HW10999, HW11000, HW11001, HW11002, HW11003, HW11005, HW11006, HW11007, HW11008, HW11009, HW11015, HW11049, HW11055, HW11058, HW11068, HW11084, HW11136, HW11182, HW11200, HW11207, HW11208, HW11220, HW11249, HW11566, HW11570, HW11571, HW11839, HW11840, HW11841, HW11842, HW11844, HW11845, HW1319, HW1320, HW1321, HW1322, HW1323, HW1324, HW1325, HW1326, HW1327, HW1328, HW1329, HW1331, HW1332, HW1333, HW1334, HW1335, HW1336, HW1337, HW1338, HW1339, hw1340, HW1341, HW1342, HW1382, HW1383, HW1384, HW1385, HW1386, HW1388, HW1389, HW1392, HW1393, HW1394, HW1395, HW1843, HW1844, HW1850, HW1851, HW1853, HW1855, HW1858, HW1859, HW1864, HW1877, HW1878, HW1881, HW1888, HW1891, HW1892, HW1894, HW1898, HW1902, HW1903, HW1996, HW1997, HW20001, HW20002, HW20003, HW20004, HW20005, HW20006, HW20007, HW20008, HW20009, HW20010, HW20011, HW20012, HW20014, HW20015, HW20016, HW20017, HW20018, HW20019, HW20020, HW20021, HW20022, HW20023, HW20024, HW20025, HW20026, HW20027, HW20028, HW20029, HW20030, HW20031, HW20032, HW20033, HW20034, HW20035, HW20036, HW20037, HW20038, HW20039, HW20040, HW20041, HW20042, HW20043, HW20044, HW20045, HW20046, HW20047, HW20048, HW20049, HW20050, HW20051, HW20052, HW20053, HW20054, HW20055, HW20056, HW20057, HW20058, HW20059, HW20060, HW20061, HW20062, HW20064, HW20065, HW20066, HW20067, HW20068, HW20069, HW20070, HW20071, HW20072, HW20073, HW20074, HW20075, HW20076, HW20077, HW20078, HW20079, HW20080, HW20081, HW20082, HW20083, HW20084, HW20085, HW20086, HW20087, HW20088, HW20089, HW2009, HW20090, HW20091, HW20092, HW20093, HW20094, HW20095, HW20096, HW20097, HW20098, HW20099, HW20100, HW20101, HW20102, HW20103, HW20104, HW20105, HW20106, HW20107, HW20108, HW20109, HW20110, HW20111, HW20112, HW20113, HW20114, HW20115, HW20116, HW20117, HW20118, HW20119, HW20120, HW20121, HW20122, HW20123, HW20124, HW20125, HW20126, HW20127, HW20128, HW20129, HW20130, HW20131, HW20132, HW20133, HW20134, HW20135, HW20136, HW20137, HW20138, HW20139, HW20140, HW20141, HW20142, HW20143, HW20144, HW20145, HW20146, HW20148, HW20149, HW20150, HW20151, HW20152, HW20153, HW20154, HW20155, HW20156, HW20157, HW20158, HW20159, HW20160, HW20161, HW20162, HW20163, HW20164, HW20165, HW20166, HW20167, HW20169, HW20170, HW20171, HW20172, HW20173, HW20174, HW20175, HW20177, HW20178, HW20179, HW20180, HW20181, HW20182, HW20183, HW20184, HW20185, HW20186, HW20187, HW20188, HW20189, HW20190, HW20191, HW20192, HW20193, HW20194, HW20195, HW20197, HW20198, HW20199, HW20200, HW20201, HW20202, HW20203, HW20204, HW20205, HW20206, HW20207, HW20208, HW20209, HW20210, HW20211, HW20212, HW20213, HW20214, HW20215, HW20216, HW20217, HW20218, HW20219, HW20220, HW20221, HW20222, HW20223, HW20224, HW20225, HW20226, HW20227, HW20228, HW20229, HW20230, HW20231, HW20232, HW20233, HW20234, HW20235, HW20236, HW20237, HW20238, HW20239, HW20240, HW20241, HW20242, HW20243, HW20244, HW20245, HW20246, HW20247, HW20248, HW20249, HW20250, HW20251, HW20252, HW20253, HW20254, HW20255, HW20256, HW20257, HW20258, HW20259, HW20260, HW20261, HW20262, HW20263, HW20264, HW20265, HW20267, HW20268, HW20269, HW2027, HW20270, HW20271, HW20272, HW20273, HW20274, HW20275, HW20276, HW20277, HW20278, HW20279, HW20280, HW20281, HW20282, HW20283, HW20284, HW20285, HW20286, HW20287, HW20288, HW20289, HW20290, HW20292, HW20293, HW20294, HW20295, HW20296, HW20297, HW20298, HW20299, HW20300, HW20301, HW20302, HW20303, HW20304, HW20305, HW20306, HW20307, HW20308, HW20309, HW20310, HW20311, HW20312, HW20313, HW20314, HW20315, HW20316, HW20317, HW20318, HW20319, HW20320, HW20321, HW20322, HW20323, HW20324, HW20325, HW20326, HW20327, HW20328, HW20329, HW20330, HW20331, HW20332, HW20333, HW20334, HW20335, HW20336, HW20337, HW20338, HW20339, HW20340, HW20341, HW20342, HW20343, HW20344, HW20345, HW20346, HW20347, HW20348, HW20349, HW20350, HW20351, HW20352, HW20353, HW20354, HW20355, HW20356, HW20357, HW20358, HW20359, HW2036, HW20360, HW20361, HW20362, HW20363, HW20364, HW20365, HW20366, HW20367, HW20368, HW20369, HW2037, HW20370, HW20371, HW20372, HW20373, HW20374, HW20375, HW20376, HW20377, HW20378, HW20379, HW2038, HW20380, HW20381, HW20382, HW20383, HW20384, HW20385, HW20386, HW20387, HW20388, HW20389, HW2039, HW20390, HW20391, HW20392, HW20393, HW20394, HW20395, HW20396, HW20397, HW20398, HW20399, HW2040, HW20400, HW20401, HW20402, HW20403, HW20404, HW20405, HW20406, HW20407, HW20408, HW20409, HW2041, HW20410, HW20411, HW20412, HW20413, HW20414, HW20415, HW20416, HW20417, HW20418, HW20419, HW20420, HW20421, HW20422, HW20423, HW20424, HW20425, HW20426, HW20427, HW20428, HW20429, HW2043, HW20430, HW20431, HW20432, HW20433, HW20434, HW20435, HW20436, HW20437, HW20438, HW2044, HW20440, HW20441, HW20442, HW20443, HW20444, HW20445, HW20446, HW20447, HW20448, HW20449, HW2045, HW20450, HW20451, HW20452, HW20453, HW20454, HW20455, HW20456, HW20457, HW20458, HW20459, HW20460, HW20461, HW20462, HW20463, HW20464, HW20465, HW20466, HW20467, HW20468, HW20469, HW20470, HW20471, HW20472, HW20473, HW20474, HW20475, HW20477, HW20478, HW20479, HW20480, HW20481, HW20482, HW20483, HW20484, HW20485, HW20486, HW20487, HW20488, HW20489, HW20490, HW20491, HW20492, HW20493, HW20494, HW20495, HW20496, HW20497, HW20500, HW20501, HW20502, HW20503, HW20506, HW20507, HW20508, HW20509, HW20513, HW20514, HW20515, HW20518, HW20522, HW20524, HW20525, HW20527, HW20528, HW20529, HW20530, HW20531, HW20532, HW20533, HW20534, HW20535, HW20536, HW20537, HW20538, HW20539, HW20540, HW20541, HW20542, HW20543, HW20544, HW20545, HW20546, HW20547, HW20548, HW20549, HW20550, HW20551, HW20552, HW20553, HW20554, HW20555, HW20556, HW20557, HW20558, HW20559, HW20560, HW20561, HW20562, HW20563, HW20564, HW20565, HW20566, HW20567, HW20569, HW20571, HW20572, HW20574, HW20575, HW20576, HW20577, HW20578, HW20579, HW20580, HW20581, HW20582, HW20583, HW20584, HW20585, HW20586, HW20587, HW20588, HW20589, HW20590, HW20591, HW20592, HW20593, HW20594, HW20595, HW20596, HW20597, HW20598, HW20599, HW20600, HW20601, HW20602, HW20603, HW20604, HW20605, HW20606, HW20607, HW20608, HW20609, HW2061, HW20611, HW20612, HW20613, HW20614, HW20615, HW20616, HW20617, HW20618, HW20619, HW2062, HW20620, HW20621, HW20623, HW20624, HW20625, HW20626, HW20627, HW20628, HW20629, HW20630, HW20631, HW20632, HW20633, HW20634, HW20635, HW20636, HW20637, HW20638, HW20639, HW2064, HW20640, HW20641, HW20642, HW20643, HW20644, HW20645, HW20646, HW20648, HW20649, HW20650, HW20651, HW20652, HW20653, HW20654, HW20655, HW20656, HW20657, HW20658, HW20659, HW2066, HW20661, HW20663, HW20664, HW20665, HW20666, HW20667, HW20668, HW20669, HW2067, HW20670, HW20671, HW20672, HW20673, HW20674, HW20675, HW20676, HW20677, HW20678, HW20679, HW20680, HW20681, HW20682, HW20683, HW20684, HW20685, HW20686, HW20687, HW20688, HW20689, HW2069, HW20690, HW20691, HW20692, HW20694, HW20695, HW20696, HW20697, HW20698, HW20699, HW20700, HW20701, HW20702, HW20703, HW20704, HW20705, HW20706, HW20707, HW20708, HW20709, HW20710, HW20711, HW20712, HW20713, HW20714, HW20715, HW20716, HW20717, HW20718, HW20719, HW20720, HW20721, HW20722, HW20723, HW20725, HW20726, HW20727, HW20728, HW20729, HW20730, HW20731, HW20732, HW20733, HW20734, HW20735, HW20736, HW20737, HW20738, HW20739, HW20740, HW20741, HW20742, HW20743, HW20744, HW20745, HW20746, HW20747, HW20748, HW20749, HW20750, HW20751, HW20752, HW20753, HW20754, HW20755, HW20756, HW20757, HW20758, hW20759, HW20760, HW20761, HW20762, HW20763, HW20764, HW20765, HW20766, HW20767, HW20768, HW20769, HW20770, HW20771, HW20772, HW20773, HW20774, HW20775, HW20776, HW20777, HW20778, HW20779, HW20780, HW20781, HW20782, HW20783, HW20784, HW20785, HW20786, HW20787, HW20788, HW20789, HW2079, HW20790, HW20791, HW20792, HW20793, HW20794, HW20795, HW20796, HW20797, HW20798, HW20799, HW20800, HW20801, HW20802, HW20803, HW20804, HW20805, HW20806, HW20807, HW20808, HW20809, HW20811, HW20812, HW20813, HW20814, HW20815, HW20816, HW20817, HW20819, HW20820, HW20821, HW20822, HW20823, HW20824, HW20825, HW20826, HW20827, HW20828, HW20829, HW20830, HW20831, HW20832, HW20833, HW20834, HW20835, HW20836, HW20837, HW20838, HW20839, HW2084, HW20840, HW20841, HW20842, HW20843, HW20844, HW20845, HW20846, HW20847, HW20848, HW20849, HW2085, HW20850, HW20851, HW20852, HW20853, HW20854, HW20855, HW20856, HW20857, HW20858, HW20859, HW2086, HW20860, HW20861, HW20862, HW20863, HW20864, HW20865, HW20866, HW20867, HW20868, HW20869, HW20870, HW20871, HW20872, HW20873, HW20874, HW20875, HW20876, HW20877, HW20878, HW20879, HW20880, HW20881, HW20882, HW20883, HW20884, HW20885, HW20886, HW20887, HW20888, HW20889, HW2089, HW20890, HW20891, HW20892, HW20893, HW20894, HW20895, HW20897, HW20898, HW20899, HW20900, HW20901, HW20902, HW20903, HW20904, HW20905, HW20906, HW20907, HW20908, HW20909, HW20910, HW20911, HW20912, HW20913, HW20914, HW20915, HW20916, HW20917, HW20918, HW20919, HW2092, HW20920, HW20921, HW20922, HW20923, HW20924, HW20925, HW20926, HW20927, HW20928, HW20929, HW20930, HW20931, HW20932, HW20933, HW20934, HW20935, HW20936, HW20937, HW20938, HW20939, HW20940, HW20941, HW20942, HW20943, HW20944, HW20945, HW20946, HW20947, HW20948, HW20949, HW20950, HW20951, HW20952, HW20953, HW20954, HW20955, HW20956, HW20957, HW20958, HW20959, HW2096, HW20960, HW20961, HW20962, HW20963, HW20964, HW20965, HW20966, HW20967, HW20968, HW20969, HW2097, HW20970, HW20971, HW20972, HW20973, HW20974, HW20975, HW20976, HW20977, HW20978, HW20979, HW2098, HW20980, HW20981, HW20982, HW20983, HW20984, HW20985, HW20986, HW20987, HW20988, HW20989, HW2099, HW20990, HW20991, HW20992, HW20993, HW20994, HW20995, HW20996, HW20997, HW20998, HW20999, HW2100, HW21000, HW21001, HW21002, HW21003, HW21004, HW21005, HW21006, HW21007, HW21008, HW21009, HW2101, HW21010, HW21011, HW21012, HW21013, HW21014, HW21015, HW21016, HW21017, HW21018, HW21019, HW2102, HW21020, HW21021, HW21022, HW21023, HW21024, HW21025, HW21026, HW21027, HW21028, HW21029, HW2103, HW21030, HW21031, HW21032, HW21033, HW21034, HW21035, HW21036, HW21037, HW21038, HW21039, HW21040, HW21041, HW21042, HW21043, HW21044, HW21045, HW21046, HW21047, HW21048, HW21049, HW2105, HW21050, HW21051, HW21052, HW21053, HW21054, HW21055, HW21056, HW21057, HW21058, HW21060, HW21061, HW21062, HW21063, HW21064, HW21065, HW21066, HW21067, HW21068, HW21069, HW21070, HW21071, HW21072, HW21073, HW21074, HW21075, HW21076, HW21077, HW21078, HW21079, HW21080, HW21081, HW21082, HW21083, HW21084, HW21085, HW21086, HW21087, HW21088, HW21089, HW21090, HW21091, HW21092, HW21093, HW21094, HW21095, HW21096, HW21097, HW21098, HW21099, HW21100, HW21101, HW21102, HW21103, HW21104, HW21105, HW21106, HW21107, HW21108, HW21109, HW21110, HW21111, HW21112, HW21113, HW21114, HW21115, HW21116, HW21117, HW21118, HW21119, HW21120, HW21121, HW21122, HW21123, HW21124, HW21125, HW21126, HW21127, HW21128, HW21129, HW21130, HW21131, HW21132, HW21133, HW21134, HW21135, HW21136, HW21137, HW21138, HW21139, HW21140, HW21141, HW21142, HW21143, HW21144, HW21145, HW21146, HW21147, HW21148, HW21149, HW21150, HW21151, HW21152, HW21153, HW21154, HW21155, HW21156, HW21157, HW21158, HW21159, HW21160, HW21161, HW21162, HW21163, HW21164, HW21165, HW21166, HW21167, HW21168, HW21169, HW21170, HW21171, HW21172, HW21173, HW21174, HW21175, HW21176, HW21177, HW21178, HW21179, HW21180, HW21181, HW21182, HW21183, HW21184, HW21185, HW21186, HW21187, HW21188, HW21189, HW21190, HW21191, HW21192, HW21193, HW21194, HW21195, HW21196, HW21197, HW21198, HW21199, HW21200, HW21201, HW21202, HW21203, HW21204, HW21205, HW21206, HW21207, HW21208, HW21209, HW21210, HW21211, HW21212, HW21213, HW21214, HW21215, HW21216, HW21217, HW21218, HW21219, HW21220, HW21221, HW21222, HW21223, HW21224, HW21225, HW21226, HW21227, HW21228, HW21229, HW21230, HW21231, HW21232, HW21233, HW21234, HW21235, HW21236, HW21237, HW21238, HW21239, HW2124, HW21240, HW21241, HW21242, HW21243, HW21244, HW21245, HW21246, HW21247, HW21248, HW21249, HW2125, HW21250, HW21251, HW21252, HW21253, HW21254, HW21255, HW21256, HW21257, HW21258, HW21259, HW21260, HW21261, HW21262, HW21263, HW21264, HW21265, HW21266, HW21267, HW21268, HW21269, HW21270, HW21271, HW21272, HW21273, HW21274, HW21275, HW21276, HW21277, HW21278, HW21279, HW21280, HW21281, HW21282, HW21283, HW21284, HW21285, HW21286, HW21287, HW21288, HW21289, HW21290, HW21291, HW21292, HW21293, HW21294, HW21295, HW21296, HW21297, HW21298, HW21299, HW21300, HW21301, HW21302, HW21303, HW21304, HW21305, HW21306, HW21307, HW21308, HW21309, HW21310, HW21311, HW21312, HW21313, HW21314, HW21315, HW21316, HW21317, HW21318, HW21319, HW21320, HW21321, HW21322, HW21323, HW21324, HW21325, HW21326, HW21327, HW21328, HW21329, HW21330, HW21331, HW21332, HW21333, HW21334, HW21335, HW21336, HW21337, HW21338, HW21339, HW21340, HW21341, HW21342, HW21343, HW21344, HW21345, HW21346, HW21347, HW21348, HW21349, HW21350, HW21351, HW21352, HW21353, HW21354, HW21355, HW21356, HW21357, HW21358, HW21359, HW21360, HW21361, HW21362, HW21363, HW21364, HW21365, HW21366, HW21367, HW21368, HW21369, HW21371, HW21372, HW21373, HW21374, HW21375, HW21376, HW21377, HW21378, HW21379, HW21380, HW21381, HW21382, HW21383, HW21384, HW21385, HW21386, HW21387, HW21388, HW21390, HW21391, HW21392, HW21393, HW21394, HW21395, HW21396, HW21397, HW21398, HW21399, HW2140, HW21400, HW21402, HW21403, HW21404, HW21405, HW21406, HW21407, HW21408, HW21409, HW21410, HW21411, HW21412, HW21413, HW21414, HW21415, HW21416, HW21417, HW21418, HW21419, HW21420, HW21421, HW21422, HW21423, HW21424, HW21425, HW21426, HW21427, HW21428, HW21429, HW21430, HW21431, HW21432, HW21433, HW21434, HW21435, HW21436, HW21437, HW21438, HW21439, HW21440, HW21441, HW21442, HW21443, HW21444, HW21445, HW21446, HW21448, HW21449, HW21450, HW21452, HW21453, HW21454, HW21455, HW21457, HW21458, HW21459, HW21460, HW21461, HW21462, HW21463, HW21464, HW21465, HW21466, HW21467, HW21468, HW21469, HW21470, HW21471, HW21472, HW21473, HW21474, HW21475, HW21476, HW21477, HW21478, HW21479, HW21480, HW21481, HW21482, HW21483, HW21484, HW21485, HW21486, HW21487, HW21489, HW21490, HW21491, HW21492, HW21493, HW21494, HW21495, HW21496, HW21497, HW21498, HW21499, HW21500, HW21501, HW21502, HW21504, HW21505, HW21506, HW21508, HW21509, HW21511, HW21512, HW21513, HW21514, HW21515, HW21516, HW21517, HW21518, HW21519, HW21520, HW21521, HW21522, HW21523, HW21524, HW21525, HW21526, HW21531, HW21536, HW21537, HW21538, HW21539, HW21540, HW21545, HW21546, HW21547, HW21554, HW21555, HW21557, HW21558, HW21563, HW21569, HW21571, HW21574, HW21584, HW21588, HW21591, HW21593, HW21594, HW21595, HW21596, HW21597, HW21598, HW21600, HW21601, HW21602, HW21603, HW21604, HW21605, HW21606, HW21613, HW2162, HW21622, HW21623, HW21636, HW21637, HW21639, HW2164, HW21640, HW21641, HW21642, HW21643, HW21644, HW21645, HW21646, HW21647, HW21648, HW21649, HW2165, HW21650, HW21651, HW21652, HW21655, HW21656, HW21657, HW21658, HW21659, HW2166, HW21660, HW21661, HW21664, HW21665, HW21666, HW21667, HW21668, HW21669, HW2167, HW21671, HW21675, HW21676, HW21677, HW2168, HW21681, HW21682, HW21684, HW2169, HW2170, HW21710, HW2172, HW21727, HW21728, HW21729, HW2173, HW2174, HW21742, HW21746, HW21747, HW21748, HW21749, HW2175, HW21750, HW21751, HW21752, HW21753, HW21754, HW21755, HW21756, HW21757, HW21758, HW21759, HW2176, HW21760, HW21761, HW21762, HW21763, HW21764, HW21765, HW21767, HW21768, HW21769, HW21770, HW21771, HW21772, HW21773, HW21774, HW21775, HW21776, HW21777, HW21778, HW21779, HW21780, HW21781, HW21782.HW21783, HW21784, HW21785, HW21786, HW21787, HW21788, HW21789, HW21790, HW21791, HW21792, HW21793, HW21794, HW21795, HW21796, HW21797, HW21798, HW21799, HW21800, HW21801, HW21802, HW21804, HW21805, HW21806, HW21807, HW21808, HW21809, HW21810, HW21811, HW21812, HW21813, HW21814, HW21815, HW21816, HW21817, HW21819, HW21820, HW21821, HW21822, HW21823, HW21824, HW21825, HW21826, HW21827, HW21828, HW21829, HW21830, HW21831, HW21832, HW21833, HW21834, HW21835, HW21836, HW21837, HW21838, HW21839, HW21840, HW21841, HW21842, HW21843, HW21844, HW21845, HW21846, HW21848, HW21849, HW21850, HW21851, HW21852, HW21853, HW21854, HW21855, HW21856, HW21857, HW21858, HW21859, HW21860, HW21861, HW21862, HW21864, HW21865, HW21866, HW21867, HW21868, HW21869, HW21870, HW21871, HW21872, HW21873, HW21874, HW21875, HW21876, HW21877, HW21878, HW21879, HW21880, HW21881, HW21882, HW21883, HW21884, HW21885, HW21886, HW21887, HW21888, HW21889, HW21890, HW21891, HW21892, HW21893, HW21894, HW21895, HW21896, HW21897, HW21898, HW21899, HW2190, HW21900, HW21901, HW21902, HW21903, HW21904, HW21905, HW21906, HW21907, HW21908, HW21909, HW2191, HW21910, HW21911, HW21912, HW21913, HW21914, HW21915, HW21916, HW21917, HW21918, HW21919, HW21920, HW21921, HW21922, HW21923, HW21924, HW21925, HW21926, HW21927, HW21928, HW21929, HW2193, HW21930, HW21931, HW21932, HW21933, HW21934, HW21935, HW21936, HW21937, HW21938, HW21939, HW2194, HW21940, HW21941, HW2195, HW2196, HW2197, HW2198, HW2199, HW2200, HW2201, HW2202, HW2204, HW2213, HW22133, HW2214, HW22145, HW2215, HW2216, HW2217, HW2218, HW22180, HW22181, HW22182, HW22183, HW22185, HW22186, HW22187, HW22188, HW22189, HW2219, HW22190, HW22191, HW22192, HW22193, HW22194, HW22195, HW22196, HW22197, HW22198, HW22199, HW22200, HW2222, HW22221, HW22225, HW22227, HW22229, HW2223, HW22230, HW22231, HW22232, HW22233, HW22235, HW22236, HW22237, HW22238, HW22239, HW2224, HW22240, HW22241, HW22242, HW22243, HW22244, HW22245, HW22247, HW22248, HW22249, HW22251, HW22252, HW22253, HW22254, HW22255, HW22256, HW22257, HW22258, HW22259, HW2226, HW22260, HW2227, HW2228, HW22280, HW22294, HW22295, HW22296, HW22297, HW22298, HW22299, HW22300, HW22301, HW22302, HW22303, HW22304, HW22305, HW22306, HW22307, HW22308, HW22309, HW22310, HW22311, HW22312, HW22313, HW22314, HW22315, HW22316, HW22317, HW2232, HW22327, HW22328, HW22329, HW22330, HW22331, HW22332, HW22333, HW22334, HW22335, HW22336, HW22356, HW22361, HW22362, HW22363, HW22364, HW22365, HW22366, HW22367, HW22368, HW22369, HW2237, HW22370, HW22371, HW22372, HW22373, HW22374, HW22375, HW22376, HW22377, HW22378, HW22379, HW2238, HW22380, HW22381, HW22382, HW22383, HW22384, HW22385, HW22386, HW22387, HW22389, HW2239, HW22390, HW22391, HW22395, HW22396, HW22397, HW22398, HW22399, HW22400, HW22401, HW22402, HW22403, HW22404, HW22406, HW22407, HW22408, HW22410, HW22411, HW22412, HW22413, HW22414, HW22415, HW22416, HW22417, HW22418, HW22419, HW22420, HW22421, HW22422, HW22423, HW22424, HW22425, HW22427, HW22429, HW22430, HW22431, HW22432, HW22433, HW22434, HW22435, HW22436, HW22437, HW22438, HW22439, HW2244, HW22440, HW22441, HW22442, HW22443, HW22444, HW22445, HW22446, HW22447, HW22448, HW22451, HW2246, HW22471, HW2248, HW2249, HW22499, HW2250, HW22500, HW22501, HW22502, HW22503, HW22504, HW22505, HW22506, HW22507, HW22508, HW22509, HW2251, HW22510, HW22511, HW22512, HW22513, HW22515, HW22516, HW22517, HW22518, HW2252, HW22520, HW22521, HW22522, HW22523, HW22524, HW22525, HW22526, HW22527, HW22529, HW2253, HW22530, HW22532, HW22533, HW2254, HW22541, HW22542, HW22543, HW22544, HW22545, HW22548, HW22549, HW2255, HW22550, HW22573, HW22581, HW22582, HW22583, HW22586, HW22587, HW22588, HW22589, HW22590, HW22591, HW22592, HW22593, HW22595, HW22596, HW22597, HW22598, HW22599, HW22600, HW22602, HW22603, HW22604, HW22605, HW22627, HW22628, HW22630, HW22631, HW22633, HW22634, HW22635, HW22636, HW22638, HW22640, HW22641, HW22642, HW22645, HW22646, HW22647, HW22648, HW22649, HW22650, HW22651, HW22652, HW2267.HW22678, HW22679, HW22680, HW22681, HW22683, HW22684, HW22685, HW22686, HW22687, HW22688, HW22689, HW22690, HW22691, HW22692, HW22693, HW22694, HW22695, HW22696, HW22697, HW22698, HW22699, HW22700, HW22701, HW22702, HW22703, HW22705, HW22707, HW22708, HW22709, HW22710, HW22712, HW22713, HW22714, HW22718, HW22719, HW22721, HW22724, HW2273, HW2274, HW2275, HW2276, HW22761, HW22764, HW2277, HW22779, HW2278, HW22780, HW22781, HW22782, HW22783, HW22784, HW22785, HW22786, HW22788, HW22789, HW2279, HW22790, HW22791, HW22794, HW2280, HW2282, HW22825, HW22826, HW22827, HW22828, HW22829, HW22830, HW22832, HW22833, HW22834, HW22835, HW22836, HW22837, HW22838, HW22839, HW2284, HW22840, HW22841, HW22842, HW22843, HW22844, HW22845, HW22846, HW22847, HW22849, HW2285, HW22850, HW22851, HW22852, HW22853, HW22854, HW22855, HW22856, HW22857, HW22858, HW22859, HW22860, HW22861, HW22862, HW22863, HW22864, HW22865, HW22866, HW22867, HW22868, HW22869, HW22870, HW22871, HW22872, HW22886, HW22892, HW22899, HW22900, HW22901, HW22902, HW22903, HW22904, HW22905, HW22906, HW22907, HW22908, HW22909, HW22911, HW22912, HW22913, HW22914, HW22915, HW22916, HW22917, HW22918, HW22919, HW22920, HW22921, HW22922, HW22939, HW22940, HW22941, HW22942, HW22943, HW22944, HW22945, HW22946, HW22947, HW22948, HW22949, HW2295, HW22950, HW22952, HW22953, HW22954, HW22955, HW22956, HW22957, HW22958, HW22959, HW22960, HW22961, HW22962, HW22963, HW22964, HW22965, HW22966, HW22967, HW22968, HW2297, HW22970, HW22971, HW2299, HW23007, HW2301, HW23011, HW23012, HW23015, HW23016, HW23017, HW23018, HW23019, HW2302, HW23020, HW23021, HW23022, HW23023, HW23024, HW23025, HW23026, HW23027, HW23028, HW23029, HW2303, HW23030, HW23031, HW23032, HW23033, HW23034, HW23035, HW23036, HW23037, HW23038, HW23039, HW2304, HW23040, HW23041, HW23042, HW23043, HW23044, HW23045, HW23046, HW23047, HW23048, HW23049, HW2305, HW23050, HW23051, HW23053, HW23054, HW23055, HW23056, HW23057, HW23058, HW23059, HW23060, HW23061, HW23062, HW2307, HW2308, HW23081, HW2309, HW2310, HW23106FSCA, HW2311, HW23112FSCA, HW23124, HW23125, HW23127, HW23128, HW23129, HW2313, HW23130, HW23131, HW23132, HW23133, HW23134, HW23135, HW23136, HW23137, HW23138, HW2314, HW23140, HW23141, HW23142, HW23143, HW23144, HW23145, HW23146, HW23147, HW2315, HW2316, HW2317, HW23174, HW23180FSCA, HW23187, HW23188, HW23189, HW23190, HW23191, HW23192, HW23194, HW23195, HW23196, HW23197, HW23198, HW23201, HW23202, HW23203, HW23204, HW23205, HW23206, HW23208, HW23209, HW23210, HW23211, HW23212, HW23213, HW23214, HW23215, HW23216, HW23217, HW23218, HW23219, HW23220, HW2324, HW23253, HW23254, HW23255, HW23256, HW23257, HW23258, HW23259, HW2326, HW23260, HW23261, HW23262, HW23263, HW23264FSCA, HW23265, HW23266, HW23267, HW23268, HW23269, HW23270, HW23271, HW23272, HW23273, HW23274, HW23275, HW23276, HW23277, HW23278, HW23279, HW2328, HW23280, HW23281, HW23282, HW23283, HW23284, HW23285, HW23286, HW23287, HW23288, HW23289, HW2329, HW23298, HW23301, HW2331, HW23311, HW23317, HW2332, HW23322, HW23323, HW23324, HW23325, HW23326, HW23327, HW23328, HW23329, HW2333, HW23330, HW23331, HW23332, HW23333, HW23334, HW23336, HW23337, HW23338, HW23339, HW2334, HW23340, HW23341, HW23342, HW23343, HW23344, HW23345, HW23346, HW23347, HW23348, HW23349, HW2335, HW23350, HW23351, HW23352, HW23353, HW23354, HW23355, HW23356, HW23357, HW23358, HW23359, HW2336, HW23360, HW23361, HW23362, HW23363, HW23365, HW23366, HW23367, HW23368, HW23369, HW2337, HW23370, HW23371, HW23372, HW23373, HW23374, HW23375, HW23376, HW23377, HW23378, HW23379, HW2338, HW23380, HW23381, HW23382, HW23387, HW2340, HW23401, HW23402, HW23403, HW23404, HW23405, HW23406, HW2341, HW23415, HW2342, HW2343, HW23438, HW23439, HW2344, HW23440, HW23442, HW23443, HW23444, HW23445, HW23449, HW23450, HW23451, HW23452, HW23453, HW23454, HW23455, HW23456, HW23457, HW2346, HW2347, HW23475, HW2348, HW2350, HW23506, HW23507, HW23508, HW23509, HW23510, HW23511, HW23512, HW23513, HW23514, HW23515, HW23516, HW23517, HW23518, HW23519, HW2352, HW23520, HW23521, HW23522, HW23523, HW23524, HW23525, HW23526, HW23527, HW23539, HW23548, HW23549, HW23550, HW23551, HW23552, HW23553, HW2357, HW23573, HW23574, HW23575, HW23576, HW23577, HW23578, HW23579, HW2358, HW23580, HW23581, HW23582, HW23584, HW23585, HW23586, HW23587, HW23588, HW23589, HW23590, HW23591, HW23592, HW23593, HW23594, HW23595, HW23596, HW2360, HW23609, HW23610, HW23611, HW23612, HW23613, HW23614, HW23615, HW23617, HW23619, HW2364, HW23642, HW23643, HW23644, HW23645, HW23646, HW23647, HW23648, HW23649, HW2365, HW23650, HW23651, HW23652, HW23653, HW23654, HW23655, HW23656, HW23657, HW23658, HW23659, HW23660, HW23661, HW23662, HW23663, HW23664, HW23665, HW23666, HW23667, HW23668, HW23669, HW2367, HW23670, HW23671, HW23672, HW23673, HW23674, HW23679, HW23680, HW23681, HW23682, HW23683, HW23684, HW23687, HW23692, HW23695, HW23699, HW2370, HW2371, HW23712, HW23713, HW23714, HW23715, HW23716, HW23717, HW23718, HW23719, HW2372, HW23720, HW23721, HW23722, HW23724, HW2373, HW23733, HW23734, HW23735, HW23736, HW23738, HW23739, HW2374, HW23740, HW23741, HW23742, HW23743, HW23744, HW23745, HW23747, HW23748, HW23749, HW2375, HW23750, HW23751, HW23752, HW23753, HW23754, HW2376, HW2377, HW2378, HW2379, HW23793FSCA, HW2381, HW2382, HW23825, HW2383, HW23835, HW23837, HW23838, HW23839, HW2384, HW23840, HW23841, HW23842, HW23843, HW23844, HW2385, HW2386, HW23862, HW23863, HW23864, HW23865, HW23866, HW23867, HW23868, HW23869, HW2387, HW23870, HW23873, HW23874, HW23875, HW23876, HW23878, HW2389, HW2390, HW2391, HW23910, HW23911, HW23912, HW23913, HW23914, HW23915, HW23917, HW23918, HW23919, HW2392, HW23920, HW23921, HW23922, HW23923, HW23924, HW23925, HW23927, HW23928, HW23929, HW2393, HW23930, HW23931, HW23932, HW23934, HW23935, HW23936, HW23937, HW23938, HW2394, HW2395, HW23953, HW2396, HW23966, HW23967, HW23968, HW23969, HW2397, HW23970, HW23971, HW23972, HW23973, HW23974, HW23975, HW23976, HW23977, HW23978, HW23979, HW2398, HW23980, HW23981, HW23982, HW23983, HW23984, HW23985, HW23986, HW2399, HW2400, HW2401, HW24014, HW24015, HW24016, HW24017, HW24018, HW24019, HW2402, HW24020, HW24021, HW24022, HW24023, HW24024, HW24025, HW24026, HW24027, HW24028, HW24029, HW2403, HW24030, HW24031, HW24032, HW24033, HW24035, HW24037, HW24038, HW24039, HW2404, HW24040, HW24041, HW24042, HW24043, HW24046FSCA, HW2405, HW2406, HW2407, HW24074, HW24078, HW24079, HW2408, HW24080, HW24081, HW24082, HW24083, HW24084, HW24085, HW24086, HW24087, HW24088, HW24089, HW2409, HW24090, HW24091, HW24092, HW24093, HW24094, HW24096, HW24097, HW24098, HW24099, HW2410, HW24100, HW2411, HW2412, HW24122, HW24126FSCA, HW24135, HW24136, HW24137, HW24138, HW24139, HW2414, HW24140, HW24141, HW24142, HW24143, HW24145, HW24146, HW24147, HW24148, HW24149, HW24150, HW24151, HW24152, HW24153, HW24154, HW24155, HW24156, HW24157, HW24158, HW24159, HW2416, HW24160, HW24161, HW24162, HW24163, HW24164, HW24165, HW24166, HW24167, HW24168, HW2417, HW24170, HW24171, HW24173, HW24174, HW24175, HW24176, HW24177, HW24178, HW2418, HW24180, HW24181, HW24182, HW24184, HW24185, HW24186, HW24187, HW24188, HW24189, HW2419, HW24190, HW24191, HW24192, HW2420, HW24209FSCA, HW2421, HW24219, HW24220, HW24221, HW24222, HW24223, HW24224, HW24225, HW24226, HW24227, HW24228, HW24229, HW2423, HW24230, HW24231, HW24232, HW24233, HW24234, HW24235, HW24236, HW24237, HW24238, HW24239, HW2424, HW24240, HW24241, HW24242, HW24243, HW24259, HW2426, HW24265, HW24266, HW24267, HW24268, HW24269, HW24270, HW24271, HW24272, HW24273, HW24274, HW24275, HW24276, HW24277, HW24278, HW24279, HW24280, HW24281, HW24282, HW24283, HW24284, HW24285, HW24286, HW24287, HW24288, HW24289, HW2429, HW24290, HW24291, HW24292, HW24293, HW24295, HW24296, HW24297, HW24298, HW24299, HW24300, HW24301, HW24302, HW24303, HW24304, HW24305, HW24306, HW24307, HW2431, HW24319, HW2432, HW24320, HW24321, HW24322, HW24323, HW24324, HW24325, HW2433, HW24330FSCA, HW24334, HW2435, HW24354, HW24356, HW24357, HW24358, HW24359, HW24360, HW24361, HW24362, HW24363, HW24364, HW24365, HW24366, HW24367, HW24368, HW24369, HW2437, HW24370, HW24371, HW24372, HW24373, HW24374, HW24375, HW24376, HW24377, HW24378, HW24379, HW24380, HW24381, HW24382, HW24383, HW24384, HW24385, HW24386, HW24387, HW24388, HW24389, HW2439, HW24390, HW24391, HW24392, HW24393, HW24394, HW24395, HW24396, HW24397, HW24398, HW24399, HW2440, HW24401, HW24402, HW24403, HW24404, HW24405, HW24406, HW24407, HW24408, HW24409, HW2441, HW24410, HW24411, HW24412, HW24413, HW24414, HW24415, HW24416, HW24417, HW24418, HW24419, HW2442, HW24420, HW24422, HW24423, HW24425, HW24426, HW24427, HW24428, HW24429, HW24430, HW24431, HW24432, HW24433, HW24435, HW24436, HW24437, HW24438, HW24439, HW24440, HW24441, HW24442, HW24443, HW24444, HW24445, HW24446, HW24447, HW24448, HW24449, HW2445, HW24450, HW24451, HW24452, HW24453, HW24454, HW24455, HW24456, HW24457, HW24458, HW24459, HW24460, HW24461, HW24462, HW24463, HW24465, HW24466, HW24467, HW24468, HW24469, HW24470, HW24471, HW24472, HW24473, HW24474, HW24475, HW24476, HW24477, HW24478, HW24479, HW2448, HW24480, HW24481, HW24482, HW24483, HW24484, HW24485, HW24486, HW24487, HW24488, HW24489, HW2449, HW24490, HW24491, HW24492, HW24493, HW24494, HW2453, HW24538, HW24539, HW24540, HW24541, HW24542, HW24543, HW24544, HW24545, HW24546, HW24547, HW24548, HW24549, HW24550, HW24551, HW24552, HW24553, HW24554, HW24555, HW24556, HW24557, HW24558, HW24559, HW24560, HW24562, HW24563, HW24565, HW24566, HW24567, HW24568, HW24569, HW2457, HW24570, HW24571, HW24572, HW24573, HW24574, HW24575, HW24576, HW24577, HW24579, HW2458, HW24580, HW24581, HW24582, HW24583, HW24584, HW24585, HW24586, HW24587, HW24588, HW24589, HW2459, HW24590, HW24591, HW24592, HW24593, HW24598FSCA, HW2460, HW24604, HW24606FSCA, HW2461, HW24610, HW24616, HW24619, HW2462, HW24620, HW24621, HW24622, HW24623, HW24624, HW24625, HW24627, HW24628, HW24629, HW2463, HW24630, HW24632, HW24633, HW24634, HW24635, HW24636, HW24637, HW24638, HW24639, HW2464, HW24640, HW24641, HW24642, HW24643, HW24644, HW24645, HW24646, HW24647, HW24648, HW24649, HW2465, HW24650, HW24651, HW24652, HW24653, HW24655, HW24656, HW24657, HW24658, HW24659, HW24660, HW24661, HW24663, HW24664, HW24665, HW24666, HW24667, HW24668, HW24669, HW2467, HW24670, HW24671, HW24672, HW24673, HW24674, HW24675, HW24676, HW24677, HW24679, HW2468, HW24680, HW24681, HW24682, HW24684, HW2469, HW24691, HW24692, HW24693, HW24694, HW24695, HW24696, HW24697, HW24698, HW24699, HW24701, HW24702, HW24703, HW24704, HW24705, HW24706, HW24707, HW24708, HW24709, HW2471, HW24710, HW24712, HW24713, HW24714, HW24715, HW24717, HW24718, HW24719, HW2472, HW24720, HW24721, HW24722, HW24723, HW2474, HW24749FSCA, HW2475, HW24756, HW24757, HW24758, HW24759, HW24760, HW24761, HW24762, HW24774, HW24789, HW2479, HW24792, HW24793, HW24794FSCA, HW24795, HW24796, HW24797, HW24798, HW24799, HW2480, HW24800, HW24801, HW24803, HW24804, HW24805, HW24807, HW24808, HW24809, HW24810, HW24811, HW24812, HW2482, HW2483, HW24833, HW24834, HW24835, HW24836, HW24838, HW24839, HW2484, HW24840, HW24841, HW24842, HW24843, HW24844, HW24845, HW2485, HW2486, HW2487, HW24886, HW2489, HW24900, HW24904, HW24905, HW24906, HW24907, HW24908, HW24909, HW2491, HW24910, HW24911, HW24912, HW24913, HW24924, HW24925, HW24927, HW24928, HW24929, HW2493, HW24930, HW24931, HW24932, HW24933, HW24934, HW24935, HW24936, HW24937, HW24938, HW24939, HW2494, HW24940, HW24941, HW24942, HW2495, HW24958, HW2496, HW24963, HW24964, HW24965, HW24966, HW24967, HW24968, HW24969, HW24970, HW24971, HW24972, HW24973, HW24984, HW24985, HW24986, HW24988, HW24989, HW2499, HW24990, HW24991, HW24992, HW24994, HW24996, HW24997, HW24998, HW24999, HW2500, HW25001, HW2501, HW2505, HW25062, HW2507, HW25079, HW25080, HW25081, HW25082, HW25083, HW25084, HW25085, HW25086, HW25087, HW25088, HW25089, HW2509, HW25090, HW25094, HW2510, HW25101, HW25102, HW25103, HW25104, HW25105, HW25106, HW25107, HW25108, HW25109, HW2511, HW25110, HW25111, HW25112, HW25113, HW25114, HW25115, HW25116, HW2513, HW25131, HW25137, HW25138, HW25139, HW2514, HW25140, HW25141, HW25142, HW25143, HW25144, HW25145, HW25146, HW25147, HW25148, HW2515, HW25158, HW25159, HW2516, HW25160, HW25161, HW25162, HW25163, HW25165, HW25166, HW25167, HW25168, HW25173, HW2518, HW25185, HW2519, hw25193, HW25201, HW25209, HW25218, HW25221, HW25222, HW25225, HW25228, HW2523, HW25232, HW25233, HW25234, HW25235, HW25236, HW25237, HW25238, HW25243, HW25244, HW25245, HW25247, HW25248, HW25249, HW2525, HW25250, HW25252, HW25255, HW25257, HW25258, HW25259, HW2526, HW25260, HW25261, HW25266, HW25269, HW2527, HW25270, HW25271, HW25272, HW25274, HW2529, HW25293, HW25294, HW25295, HW25296, HW25297, HW25298, HW2530, HW25300, HW25301, HW25302, HW25303, HW25304, HW25305, HW25306, HW25307, HW25308, HW25309, HW2531, HW25310, HW25311, HW25312, HW25313, HW25317, HW25319, HW2533, HW25334, HW25335, HW25336, HW25337, HW25338, HW25339, HW25340, HW25341, HW25343, HW25344, HW25345, HW25346, HW25347, HW25348, HW25353, HW25354, HW25356, HW25357, HW25358, HW25359, HW2536, HW25360, HW25361, HW25362, HW25363, HW25364, HW25365, HW25366, HW25367, HW25368, HW25369, HW2537, HW25370, HW25371, HW25372, HW25373, HW25374, HW25375, HW25376, HW25378, HW25379, HW25380, HW25381, HW25382, HW25383, HW25384, HW25385, HW25386, HW25387, HW25388, HW25389, HW2539, HW25390, HW25391, HW25393, HW25394, HW25395, HW25396, HW25397, HW25398, HW25399, HW25400, HW25401, HW25402, HW25404, HW25405, HW25406, HW25407, HW25427, HW2543, HW25433, HW25439, HW2545, HW25452, HW25455, HW25464, HW25465, HW25466, HW25467, HW25468, HW25469, HW25470, HW25471, HW25472, HW25473, HW25474, HW25475, HW25476, HW25478, HW25483, HW25486, HW25487, HW25488, HW25489, HW25490, HW25492, HW2550, HW25516FSCA, HW25518, HW25519, HW25520, HW25521, HW25522, HW25523, HW25524, HW25525, HW25526, HW25528, HW25529, HW25530, HW25531, HW25533, HW25535, HW25536, HW25537, HW2554, HW25544FSCA, HW25545FSCA, HW25547, HW25549, HW25550, HW25552, HW25554, HW25555, HW25556, HW25557, HW25558, HW25559, HW25560, HW25561, HW25562, HW25563, HW25565, HW25566, HW25567, HW25581, HW25582, HW25584, HW25585, HW25586, HW25587, HW25588, HW25589, HW2559, HW25590, HW25591, HW25592, HW25593, HW25594, HW25595, HW25596, HW25599, HW2560, HW25601, HW25602, HW25603, HW25604, HW25605, HW25606, HW25607, HW25608, HW25609, HW2561, HW25610, HW25612, HW25613, HW25614, HW25617, HW25619, HW2562, HW25620, HW25621, HW25622, HW25624, HW25627, HW25628, HW25629, HW2563, HW25630, HW25631, HW25632, HW25633, HW25635, HW25636, HW25637, HW25638, HW25639, HW2564, HW25640, HW25641, HW25642, HW25657, HW2566, HW25661, hw25665, HW25666, HW25667, HW25668, HW25669, HW2567, HW25671, HW25672, HW25673, HW25674, HW25675, HW25676, HW25677, HW25678, HW2568, HW25680, HW25681, HW25682, HW25685, HW25686, HW25687, HW25688, HW25689, HW2569, HW25690, HW25691, HW25692, HW25694, HW25695, HW25697, HW25698, HW2570, HW25700, HW25703, HW25705, HW25709, HW2571, HW25712, HW2572, HW25727, HW25728, HW25729, HW2573, HW25730, HW25731, HW25732, HW25733, HW25734, HW25735, HW25736, HW25738, HW25739, HW2574, HW25740, HW25741, HW25742, HW25743, HW25744, HW25745, HW25746, HW25747, HW25748, HW25749, HW2575, HW25750, HW25751, HW25752, HW25753, HW25757FSCA, HW25759, HW25761, HW25764, HW25765, HW25766, HW25768, HW25769, HW2577, HW25770, HW25771, HW25772, HW25775, HW2578, HW25780, HW25788, HW25789, HW2579, HW25790, HW25791, HW25792, HW25793, HW25794, HW25795, HW25796, HW25797, HW25798, HW25799, HW2580, HW25800, HW25802, HW25803, HW25804, HW25805, HW25806, HW25807, HW25808, HW25809, HW25811, HW25812, HW25813, HW25814, HW25816, HW25817, HW25818, HW25819, HW2582, HW25820, HW25821, hw25822, HW25823, HW25825, HW25826, hw25828, HW25829, HW25830, HW25831, HW25832, HW25833, HW25834, HW25835, HW25836, HW25837, HW2584, HW25840, HW25841, HW25842, HW25843, HW25844, HW25845, HW25847, HW25849, HW25850, HW25851, HW25852, HW25853, HW25859, HW2586, HW25862, HW25864, HW25871, HW25875, HW25876, HW25877, HW25878, HW25879, HW2588, HW25880, HW25881, HW25882, HW25883, HW25884, HW25885, HW25886, HW25887, HW25888, HW25889, HW2589, HW25890, HW25891, HW25892, HW25893, HW25895, HW25896, HW25897, HW25898, HW25899, HW25900, HW25901, HW25902, HW25903, HW25904, HW25905, HW25906, HW25907, HW25908, HW25911, HW25918, HW25919, HW2592, HW25920, HW25921, HW25922, HW25924, HW25925, HW25926, HW25927, HW25929, HW2593, HW25930, HW25931, HW25932, HW25933, HW25934, HW25935, HW2594, HW2595, HW2597, HW25996, HW26002, HW26006, HW2601, HW26011, HW26015, HW2602, HW2603, HW26031, HW26032, HW26037, HW26040, HW26044, HW26045, HW26047, HW26048, HW26049, HW26050, HW26051, HW26053, HW26055, HW26058, HW26059, HW2606, HW2607, HW26070, HW26074, HW26076, HW26077, HW26078, HW26079, HW2608, HW26080, HW2609, HW26093, HW26094, HW26095, HW26096, HW26097, HW26098, HW26099, HW2610, HW26100, HW26101, HW26102, HW26103, HW26104, HW26105, HW26106, HW26108, HW26109, HW2611, HW26110, HW26111, HW26112, HW26113, HW26114, HW26115, HW26116, HW26119, HW2612, HW26120, HW26121, HW26122, HW26124, HW26125, HW26126, HW26128, HW2613, HW26138, HW2614, HW26140, HW26143, HW26145, HW26146, HW26147, HW26148, HW26149, HW2615, HW26150, HW26151, HW26153, HW26155, HW26156, HW26157, HW26158, HW26159, HW2616, HW26160, HW26162, HW26163, HW26164, HW26165, HW26166, HW26167, HW26168, HW26169, HW26170, HW26171, HW26172, HW26173, HW26180, HW2619, HW26196, HW26199, HW26200, HW26201, HW26202, HW26203, HW26204, HW26205, HW26206, HW26207, HW26208, HW26209, HW26210, HW26211, HW26212, HW26213, HW26214, HW26215, HW26216, HW26217, HW26218, HW26219, HW26220, HW26221, HW26222, HW26223, HW26224, HW26225, HW26226, HW26227, HW26228, HW26229, HW2623, HW26230, HW26231, HW26232, HW26233, HW26235, HW26236, HW26237, HW26238, HW26239, HW2624, HW26240, HW26241, HW26242, HW26243, HW26244, HW26245, HW26246, HW26248, HW26249, HW26250, HW26251, HW26252, HW26253, HW26254, HW26255, HW26256, HW26257, HW26258, HW26259, HW2626, HW26260, HW26261, HW26262, HW26263, HW26264, HW26265, HW26266, HW26267, HW26268, HW26269, HW26270, HW26271, HW26272, HW26273, HW26274, HW26275, HW26276, HW26277, HW26278, HW26279, HW2628, HW26280, HW26281, HW26282, HW26283, HW26284, HW26285, HW26286, HW26287, HW26288, HW26289, HW26290, HW26291, HW26292, HW26293, HW26294, HW26295, HW26296, HW26297, HW26298, HW26299, HW26301, HW26302, HW26303, HW26304, HW26305, HW26306, HW26307, HW26309, HW26310, HW2632, HW26329, HW26330, HW26331, HW26333, HW26334, HW26335, HW26336, HW2634, HW26366, HW2638, HW26381, HW26382, HW26383, HW26384, HW26385, HW26386, HW26387, HW26388, HW26389, HW2639, HW26390, HW26391, HW26392, HW26393, HW26394, HW26395, HW26396, HW26397, HW26398, HW26399, HW2640, HW2642, HW26421, HW26428, HW26436, HW26437, HW26438, HW26439, HW2644, HW26440, HW26441, HW26442, HW26443, HW26444, HW26445, HW26446, HW26447, HW26448, HW26449, HW2645, HW26450, HW26451, HW26458, HW26459, HW2647, HW26484, HW2649, HW26495, HW26498, HW26499, HW2650, HW26500, HW26501, HW26502, HW26504, HW26505, HW26506, HW26507, HW26509, HW2651, HW26510, HW26511, HW26512, HW26513, HW26514, HW26515, HW26517, HW26549, HW26550, HW26551, HW26552, HW26553, HW26554, HW26555, HW26556, HW26557, HW26558, HW26559, HW2656, HW26560, HW26562, HW26563, HW26565, HW26566, HW26567, HW26568, HW26569, HW2657, HW26570, HW26571, HW26572, HW26573, HW26574, HW26575, HW2658, HW2661, HW26612, HW26613, HW26614, HW26615, HW26616, HW26617, HW26618, HW26619, HW2662, HW26620, HW26621, HW26622, HW2663, HW2664, HW26647, HW26648, HW26649, HW2665, HW26650, HW26651, HW26652, HW26653, HW26655, HW26656, HW26657, HW26658, HW26659, HW2666, HW26660, HW26661, HW26662, HW26663, HW26664, HW26665, HW2668, HW26689, HW2669, HW26690, HW26691, HW26692, HW26696, HW26697, HW26698, HW26699, HW2670, HW26700, HW26701, HW26702, HW26703, HW26704, HW26706, HW26707, HW26708, HW26709, HW26715, HW2672, HW2673, HW2674, HW26740, HW26741, hw26742, HW26743, HW26744, HW26746, HW26747, HW26748, HW26749, HW26750, HW26751, HW26753, HW26754, HW26755, HW26762, HW26765, HW26766, HW26778, HW2678, HW26787, HW26788, HW26789, HW2679, HW26790, HW26791, HW26792, HW26793, HW26794, HW26795, HW26796, HW2680, HW2681, HW2682, HW26835, HW26836, HW26838, HW26839, HW2684, HW26841, HW26843, HW26844, HW26846, HW26847, HW26849, HW2685, HW26850, HW2686, HW2687, HW26871, HW2688, HW26888, HW2689, HW2692, HW2693, HW26935, HW26936, HW26938, HW26939, HW2694, HW26940, HW26941, HW2696, HW26964, HW26969, HW26971, HW26976, HW26977, HW26979, HW26980, HW26981, HW26982, HW26983, HW26984, HW26985, HW2699, HW2700, HW27005, HW2701, HW27012, HW27021, HW27024, HW27025, HW27026, HW27027, HW27028, HW27029, HW2703, HW27030, HW27031, HW27032, HW27033, HW27034, HW27035, HW27036, HW27037, HW27038, HW27039, HW27040, HW27041, HW27042, HW27043, HW27044, HW27045, HW27046, HW27047, HW27048, HW27049, HW27050, HW27051, HW27052, HW27053, HW27054, HW27055, HW27059, HW27060, HW27061, HW27062, HW27063, HW27064, HW27065, HW27066, HW27067, HW27068, HW27069, HW27070, HW27071, HW27072, HW27073, HW27074, HW27075, HW27076, HW27077, HW27078, HW27079, HW2708, HW27080, HW27088, HW2709, HW27093, HW27099, HW27101, HW27104, HW27105, HW27106, HW27107, HW27108, HW27109, HW2711, HW27110, HW27111, HW27112, HW27113, HW27114, HW27115, HW27116, HW27117, HW27118, HW2712, HW27120, HW27121, HW27123, HW27125, HW27126, HW27127, HW27128, HW27129, HW27130, HW27131, HW27132, HW27133, HW27135, HW27136, HW27137, HW27138, HW27139, HW27140, HW27142, HW27143, HW27144, HW27145, HW27146, HW27147, HW27148, HW27150, HW27151, HW27154, HW27159, HW2716, HW27161, HW27162, HW27166, HW27167, HW27168, HW27169, HW2717, HW27170, HW27171, HW27173, HW27174, HW27175, HW27176, HW27177, HW27178, HW27181, HW27182, HW27183, HW27184, HW27185, HW27186, HW27187, HW27188, HW27189, HW27191, HW27192, HW27193, HW27194, HW27195, HW27196, HW27197, HW27198, HW27199, HW2720, HW27200, HW27201, HW27202, HW27203, HW27204, HW27205, HW27206, HW27207, HW27208, HW2721, HW27210, HW27211, HW27212, HW27213, HW27214, HW27215, HW27216, HW27217, HW27219, HW27222, HW27223, HW27224, HW27226, HW27227, HW27230, HW27231, HW27232, HW27233, HW27234, HW27235, HW27236, HW27238, HW2724, HW2725, HW2728, HW2729, HW2730, HW27300, HW2731, HW27311, HW27318, HW2732, HW27329, HW2733, HW27337, HW2734, HW27340, HW2735, HW27350, HW27351, HW27354, HW27355, HW27356, HW27357, HW2736, HW27360, HW27361, HW27364, HW27367, HW2737, HW27375, HW2738, HW27383, HW27384, HW27387, HW27389, HW2739, HW27390, HW27391, HW27392, HW27393, HW27395, HW27396, HW27397, HW27398, HW27399, HW27400, HW27401, HW27402, HW27403, HW27404, HW27406, HW27407, HW27408, HW27409, HW27410, HW27411, HW27412, HW27413, HW27414, HW27415, HW27416, HW27417, HW27418, HW27419, HW2743, HW27434, HW2744, HW27442, HW27444, HW27445, HW27446, HW27448, HW27449, HW2745, HW27451, HW27453, HW27454, HW27458, HW27463, HW27468, HW2747, HW27476, HW27483, HW27487, HW27489, HW2749, HW27492, HW27509, HW2751, HW2752, HW27527, HW27528, HW27529, HW2753, HW27530, HW27531, HW27532, HW27533, HW27534, HW27535, HW27536, HW27537, HW27538, HW2754, HW27542, HW27548, HW2755, HW27553, HW2756, HW27568, HW2757, HW27570, HW27571, HW27573, HW27575, HW27576, HW27577, HW27578, HW27579, HW27580, HW27581, HW27582, HW27583, HW27584, HW27586, HW27587, HW27589, HW2759, HW27590, HW27591, HW27592, HW27593, HW27594, HW27595, HW27596, HW27598, HW27599, HW2760, HW27600, HW27601, HW27602, HW27604, HW27605, HW2761, HW27639, HW2764, HW27646, HW2765, HW27650, HW27651, HW27652, HW2766, HW27660, HW27661, HW27662, HW27663, HW27664, HW27665, HW27666, HW27667, HW27668, HW27669, HW2767, HW27670, HW27671, HW27672, HW27673, HW27674, HW27675, HW27676, HW27677, HW27678, HW27680, HW27681, HW27688, HW27690, HW27692, HW27693, HW27694, HW27695, HW27696, HW27698, HW27699, HW2770, HW27700, HW27701, HW27705, HW27707, HW27711, HW2772, HW27728, HW27729, HW27730, HW27731, HW27732, HW27733, HW27734, HW27735, HW27736, HW27737, HW27738, HW27739, HW2774, HW27740, HW27741, HW27742, HW27744, HW27745, HW27746, HW27747, HW27748, HW27759, HW2776, HW27763, HW27766, HW27767, HW2777, HW27775, HW27776, HW27777, HW27778, HW27779, HW2778, HW27780, HW27781, HW27782, HW27784, HW27785, HW27786, HW27787, HW27788, HW2779, HW2780, HW27806, HW27807, HW27808, HW2781, HW2782, HW27823, HW2783, HW27837, HW2784, HW27846, HW27847, HW27848, HW27849, HW2785, HW27850, HW27851, HW27852, HW27853, HW27854, HW27856, HW27858, HW27859, HW2786, HW27860, HW27861, HW27862, HW27863, HW27864, HW27865, HW27867, HW2787, HW27871, HW27872, HW27873, HW27874, HW27875, HW27876, HW27877, HW27879, HW2788, HW27880, HW27881, HW27882, HW27883, HW27884, HW27885, HW27886, HW27887, HW27888, HW2789, HW27899, HW2790, HW27900, HW27902, HW27904, HW27908, HW27909, HW2791, HW27910, HW27913, HW27915, HW27916, HW27923, HW2795, HW27955, HW27957, HW27958, HW27959, HW2796, HW27960, HW27961, HW27962, HW27963, HW27964, HW27965, HW27966, HW27967, HW27969, HW27970, HW27971, HW27972, HW27973, HW27974, HW27975, HW27976, HW27977, HW27978, HW27979, HW27982, HW27983, HW27984, HW27985, HW27986, HW27987, HW27988, HW27989, HW2799, HW27990, HW27991, HW27992, HW27994, HW27995, HW27996, HW27997, HW27998, HW27999, HW2800, HW28000, HW28001, HW28002, HW28004, HW2801, HW28010, HW28013, HW28014, HW28015, HW28022, HW28023, HW28024, HW28025, HW28026, HW2803, HW28031, HW28032, HW28033, HW28035, HW28036, HW28037, HW28038, HW28039, HW28041, HW28042, HW28043, HW28044, HW28045, HW28046, HW28047, HW28048, HW28050, HW28051, HW28053, HW28054, HW28055, HW28056, HW28057, HW28058, HW28059, HW2806, HW28060, HW28061, HW28062, HW28063, HW28064, HW28065, HW28066, HW28067, HW28068, HW28069, HW2807, HW28070, HW28071, HW28073, HW28075, HW28076, HW28077, HW28078, HW28079, HW2808, HW28080, HW28081, HW28082, HW28083, HW2809, HW2810, HW28106, HW2811, HW28125, HW28127, HW2813, HW28139, HW2814, HW28144, HW28145, HW28146, HW28147, HW28148, HW28149, HW28150, HW28151, HW28152, HW28153, HW28154, HW28155, HW28156, HW28157, HW28158, HW28159, HW2816, HW28160, HW28161, HW28162, HW28163, HW28164, HW28165, HW28166, HW28168, HW28169, HW2817, HW28170, HW28173, HW28175, HW28176, HW2818, HW28194, HW28195, HW28196, HW28197, HW28198, HW28199, HW2820, HW28200, HW28201, HW28202, HW28203, HW28204, HW28205, HW28206, HW28207, HW28208, HW28209, HW2821, HW28210, HW28211, HW28212, HW28214, HW28215, HW28216, HW28217, HW28218, HW2822, HW28220, HW28222, HW28223, HW28224, HW28227, HW28228, HW28229, HW2823, HW28230, HW28231, HW28232, HW28234, HW28236, HW28238, HW28239, HW2824, HW28247, HW2826, HW28266, HW2828, HW28286, HW2829, HW28290, HW28296, HW28299, HW2830, HW28307, HW28309, HW28310, HW28311, HW28313, HW28314, HW28316, HW28317, HW28319, HW28328, HW2833, HW28336, HW28337, HW28338, HW28339, HW2834, HW28340, HW28341, HW28342, HW28343, HW28344, HW28345, HW28346, HW28347, HW28348, HW28349, HW2835, HW28350, HW28351, HW28352, HW28353, HW2836, HW28370, HW28371, HW28372, HW28373, HW28374, HW28375, HW28377, HW28378, HW28379, HW28380, HW28383, HW28384, HW28385, HW28387, HW28411, HW28412, HW28413, HW28414, HW28415, HW28416, HW28417, HW28418, HW28419, HW2842, HW28420, HW28422, HW28423, HW2843, HW2844, HW28445, HW28446, HW28447, HW2845, HW2846, HW28469, HW28471, HW28472, HW28475, HW28477, HW28478, HW28479, HW28480, HW28482, HW28484, HW28485, HW28494, HW2851, HW2852, HW28536, HW28537, HW28538, HW28539, HW28540, HW28541, HW28542, HW28543, HW28544, HW28545, HW28547, HW28550, HW28551, HW28552, HW28553, HW28554, HW28555, HW28556, HW28557, HW28558, HW28560, HW28561, HW28562, HW28563, HW28564, HW28566, HW28567, HW28568, HW28569, HW28572, HW28574, HW28575, HW28576, HW28577, HW28578, HW28579, HW2858, HW28580, HW28581, HW28582, HW28583, HW28584, HW28587, HW28588, HW2859, HW28590, HW2860, HW2861, HW28616, HW28619, HW2862, HW28621, HW28622, HW28623, HW28624, HW28625, HW28626, HW28627, HW28628, HW28629, HW28630, HW28631, HW28632, HW28633, HW28636, HW28637, HW28638, HW28639, HW2864, HW28640, HW28660, HW28665, HW28666, HW28667, HW28668, HW28669, HW2867, HW28670, HW28673, HW28677, HW28678, HW28679, HW2868, HW28680, HW28681, HW28683, HW28701, HW28705, HW2872, HW28723, HW28737, HW2874, HW28740, HW28745, HW28747, HW28748, HW2875, HW28750, HW28751, HW28752, HW28753, HW28754, HW28757, HW28759, HW2876, HW28761, HW28764, HW28766, HW28767, HW2877, HW28770, HW28774, HW28776, HW2878, HW28780, HW28782, HW2879, HW28792, HW28796, HW28809, HW2881, HW28817, HW2882, HW28821, HW28824, HW28825, HW28826, HW28827, HW28828, HW28829, HW28831, HW28833, HW28834, HW28837, HW28838, HW28839, HW2884, HW28844, HW28845, HW28846, HW28847, HW28849, HW28850, HW28852, HW28853, HW28854, HW28855, HW28856, HW28858, HW2886, HW28860, HW28861, HW28862, HW28863, HW28864, HW28865, HW28866, HW28867, HW28868, HW2887, HW2888, HW28885, HW28886, HW28887, HW28888, HW2889, HW28890, HW28891, HW28892, HW28894, HW28895, HW28896, HW28897, HW2890, HW2891, HW2892, HW28921, HW28922, HW28923, HW28924, HW28925, HW28926, HW28927, HW28928, HW2893, HW28930, HW28931, HW28932, HW28933, HW28934, HW28935, HW2894, HW2895, HW28955, HW28958, HW28959, HW2896, HW28960, HW28962, HW28963, HW28964, HW28967, HW2897, HW28973, HW28977, HW2898, HW28981, HW28982, HW2900, HW29002, HW29006, HW29007, HW29008, HW29009, HW2901, HW29014, HW29015, HW29016, HW29017, HW29018, HW29019, HW2902, HW29020, HW29021, HW29022, HW29023, HW29024, HW29025, HW29028, HW29029, HW2903, HW2904, HW2905, HW2906, HW29067, HW29068, HW29069, HW2907, HW29070, HW29071, HW29072, HW29073, HW29074, HW29075, HW29076, HW2908, HW2909, HW29092, HW29093, HW29095, HW29097, HW29098, HW29099, HW2910, HW29100, HW29102, HW2913, HW29130, HW29131, HW29133, HW29134, HW29135, HW2914, HW2915, HW2916, HW2917, HW2918, HW2919, HW29192, HW2920, HW29208, HW2921, HW29214, HW29216, HW29217, HW29218, HW2922, HW29220, HW29222, HW29223, HW29224, HW29227, HW2923, HW29231, HW29233, HW29236, HW29237, HW29239, HW2924, HW29240, HW29241, HW29245, HW29247, HW2925, HW29254, HW2926, HW2927, HW2928, HW2929, HW29299, HW2930, HW29303, HW29305, HW29306, HW29307, HW29308, HW29312, HW29313, HW29314, HW2932, HW29321, HW2933, HW2934, HW29344, HW2935, HW29351, HW29352, HW29353, HW29354, HW29355, HW29356, HW29357, HW29359, HW2936, HW29360, HW29361, HW29363, HW29365, HW29366, HW29367, HW2937, HW29370, HW29371, HW2939, HW2940, HW29406, HW29415, HW2942, HW29421, HW29432, HW2944, HW29440, HW29442, HW29444, HW29446, HW2945, HW29454, HW29462, HW29463, HW29464, HW2947, HW2948, HW29482, HW2949, HW2950, HW2951, HW2952, HW2953, HW2954, HW2955, HW29570, HW29572, HW29573, HW29574, HW29575, HW29576, HW29581, HW29582, HW29585, HW29617, HW29618, HW2962, HW2970, HW2971, HW2972, HW2973, HW29734, HW2974, HW29746, HW29747, HW29748, HW29749, HW2975, HW29750, HW29752, HW29757, HW29758, HW29759, HW2976, HW29760, HW29762, HW29763, HW29765, HW29768, HW29769, HW2977, HW29770, HW29771, HW2978, HW29789, HW2979, HW29795, HW2980, HW2981, HW2982, HW2983, HW2984, HW29840, HW29841, HW29842, HW29843, HW29844, HW29846, HW2985, HW2986, HW2988, HW2991, HW2994, HW3011, HW30135, HW30136, HW30137, HW30139, HW30140, HW30141, HW30142, HW30143, HW30144, HW3020, HW30252, HW30253, HW30255, HW3026, HW3031, HW3032, HW30359, HW30361, HW30362, HW30389, HW30391, HW30394, HW30395, HW30396, HW30401, HW30405, HW3041, HW3042, HW3043, HW3044, HW3045, HW3046, HW3047, HW3048, HW3049, HW30497, HW3050, HW3051, HW30514, HW30515, HW30516, HW30517, HW30518, HW3052, HW30522, HW3053, HW3054, HW3055, HW3056, HW3057, HW3058, HW3059, HW3060, HW3062, HW3063, HW3064, HW3065, HW3066, HW30662, HW30663, HW30664, HW3067, HW3068, HW3069, HW3070, HW3071, HW30737, HW3077, HW30776, HW3079, HW30791, hw30793, hw30794, hw30795, hw30798, hw30799, hw30800, HW30807, HW30812, HW30817, HW30827, HW30835, HW3084, HW30860, HW30861, HW3089, HW3090, HW3091, HW3092, HW3094, HW3095, HW3097, HW3098, HW3099, HW3102, HW3105, HW3108, HW3109, HW3110, HW3112, HW3113, HW3114, HW3115, HW3116, HW3117, HW3119, HW3120, HW3122, HW3123, HW3125, HW3127, HW3128, HW3129, HW3130, HW3131, HW3132, HW3133, HW3134, HW3135, HW3136, HW3137, HW3138, HW3139, HW3140, HW3141, HW3142, HW3143, HW3146, HW3147, HW3148, HW3149, HW3150, HW3151, HW3152, HW3153, HW3155, HW3156, HW3157, HW3158, HW3159, HW3160, HW3161, HW3162, HW3163, HW3165, HW3166, HW3168, HW3169, HW3170, HW3172, HW3173, HW3174, HW3176, HW3177, HW3178, HW3179, HW3180, HW3181, HW3182, HW3184, HW3187, HW3188, HW3190, HW3191, HW3192, HW3193, HW3194, HW3195, HW3196, HW3197, HW3198, HW3199, HW3200, HW3201, HW3202, HW3203, HW3204, HW3205, HW3206, HW3207, HW3208, HW3209, HW3211, HW3212, HW3213, HW3214, HW3215, HW3216, HW3217, HW3218, HW3219, HW3220, HW3221, HW3222, HW3223, HW3225, HW3226, HW3227, HW3229, HW3230, HW3231, HW3232, HW3233, HW3234, HW3235, HW3236, HW3237, HW3238, HW3239, HW3240, HW3241, HW3242, HW3243, HW3244, HW3245, HW3246, HW3247, HW3248, HW3249, HW3250, HW3251, HW3252, HW3253, HW3254, HW3255, HW3256, HW3257, HW3258, HW3259, HW3260, HW3261, HW3262, HW3263, HW3264, HW3265, HW3266, HW3267, HW3268, HW3270, HW3272, HW3282, HW3286, HW3292, HW3302, HW3312, HW3314, HW3317, HW3331, HW3333, HW3345, HW3357, HW3358, HW3359, HW3367, HW3389, HW3390, HW3391, HW3393, HW3457, HW3467, HW3472, HW3520, HW3521, HW3533, HW3548, HW3549, HW3555, HW3559, HW3562, HW3664, HW3666, HW3670, HW3671, HW3677, HW3680, HW3689, HW3691, HW3704, HW3709, HW3714, HW3717, HW3720, HW3721, HW3723, HW3724, HW3725, HW3729, HW3731, HW3732, HW3733, HW3738, HW3743, HW3745, HW3747, HW3750, HW3753, HW3755, HW3759, HW3760, HW3761, HW3762, HW3764, HW3766, HW3819, HW3824, HW3826, HW3827, HW3830, HW3835, HW3837, HW3838, HW3839, HW3840, HW3978, HW4033, HW4039, HW4051, HW4101, HW4104, HW4106, HW4113, HW4132, HW4153, HW4159, HW4161, HW4166, HW4167, HW4174, HW4179, HW4189, HW4191, HW4200, HW4206, HW4213, HW4230, HW4239, HW4240, HW4244, HW4245, HW4248, HW4250, HW4252, HW4253, HW4256, HW4263, HW4264, HW4265, HW4266, HW4267, HW4269, HW4270, HW4271, HW4273, HW4274, HW4275, HW4276, HW4277, HW4278, HW4279, HW4283, HW4284, HW4286, HW4287, HW4288, HW4289, HW4291, HW4293, HW4295, HW4297, hw4300, HW4302, HW4305, HW4306, HW4311, HW4312, HW4324, HW4326, HW4359, HW4379, HW4405, HW4407, HW4422, HW4466, HW4468, HW4469, HW4471, HW24578FSCA, HW24626FSCA, HW24716FSCA, HW24926FSCA, HW25583FSCA, HW25699FSCA, HW25702FSCA, HW25706FSCA; GTIN 00888707000239, Serial Numbers: HW34737, HW30645, HW30877, HW30634, HW30633, HW26332; GTIN 00888707000581, Serial Numbers: HW29094, HW29225, HW29226, HW29242, HW29833, HW30499, HW30801, HW29519, HW29441, HW29515, HW29447, HW28548, HW29680, HW31998, HW31727, HW31442, HW31380, HW29430, HW31144, HW31376, HW29203, HW29425, HW30876, HW30638, HW31187, HW30706, HW30589, HW32416, HW29429, HW29207, HW29679, HW29517, HW29516, HW29520, HW29438, HW34017, HW29427, HW29439, HW29428, HW32023; GTIN 00888707000604, Serial Numbers: HW29580, HW29591, HW29470, HW29521; GTIN 00888707000628, Serial Numbers: HW29433; GTIN 00888707000635, Serial Numbers: HW29523, HW29471; GTIN 00888707000642, Serial Numbers: HW29524; GTIN 00888707000666, Serial Numbers: HW29473; GTIN 00888707000680, Serial Numbers: HW30711, HW29845, HW30635, HW29578, HW30639, HW29571, HW31150, HW31143, HW31145, HW30646, HW30734; GTIN 00888707001649, Serial Numbers: HW36168, HW35672, HW35094, HW35753, HW35694, HW34031, HW34327, HW36121, HW35486, HW35640, HW35750, HW35602, HW35719, HW32018, HW32022, HW34963, HW35465, HW35476, HW31415, HW35421, HW35372, HW35378, HW30862, HW31587, HW32015, HW35470, HW34836, HW35371, HW35157, HW35315, HW34936, HW35147, HW32000, HW32081, HW35366, HW35038, HW29684, HW32064, HW29653, HW32268, HW32050, HW32056, HW31423, HW30726, HW30949, HW30529, HW31417, HW32042, HW32001, HW30525, HW31505, HW31985, HW31982, HW31808, HW31806, HW31592, HW31590, HW31588, HW31589, HW31811, HW31812, HW31133, HW31546, HW31496, HW31422, HW30549, HW31032, HW31036, HW31305, HW31374, HW31130, HW31155, HW31120, HW31092, HW30528, HW30526, HW29537, HW30523, HW29505, HW29597, HW30353, HW30354, HW30355, HW29487, HW29967, HW35475, HW34296, HW29876, HW29235, HW34431, HW35659, HW34341, HW32245, HW35310, HW30145, HW30146, HW29659, HW34508, HW35328, HW35412, HW35658, HW35550, HW35547, HW30371, HW35664, HW35662, HW35559, HW35657, HW35311, HW35068, HW35085, HW35349, HW35290, HW32301, HW29985, HW29864, HW31511, HW34507, HW32255, HW34343, HW32021, HW31506, HW30222, HW29474, HW35564, HW34764, HW36165, HW36154, HW31048, HW34324, HW35554, HW35019, HW35546, HW34340, HW32014, HW30421, HW31801, HW30368, HW35665, HW35666, HW35555, HW35599, HW35556, HW35333, HW35065, HW35090, HW34692, HW30836, HW34180, HW30806, HW30805, HW30098, HW28741, HW29989, HW35560, HW35557, HW35558, HW35402, HW34906, HW31009, HW31014, HW31016, HW35667, HW35237, HW35250, HW35021, HW35041, HW34148, HW34420, HW32138, HW32147, HW31815, HW31891, HW31483, HW31510, HW31018, HW31058, HW31063, HW31070, HW35375, HW35033, HW35663, HW35035, HW32270, HW35365, HW35418, HW35309, HW35312, HW35313, HW32410, HW34745, HW35114, HW35196, HW35197, HW34029, HW31071, HW34132, HW32024, HW32265, HW35308, HW31424, HW30223, HW35411, HW31566, HW31513, HW31554, HW29991, HW30373, HW31394, HW32055, HW34994, HW34266, HW34268, HW31632, HW31512, HW31809, HW35485, HW34977, HW30840, HW31088, HW35373, HW35314, HW35598, HW31795, HW35405, HW35551, HW30100, HW35329, HW32186, HW35408, HW35553, HW35552, HW35423, HW32264, HW34941, HW35338, HW35409, HW34932, HW35404, HW35330, HW35370, HW34940, HW35318, HW35320, HW35210, HW31921, HW31973, HW34346, HW34641, HW34345, HW32420, HW34418, HW34992, HW34271, HW32394, HW32094, HW34269, HW34301, HW32112, HW34270, HW34348, HW32123, HW32560, HW32561, HW34094, HW31919, HW34093, HW31813, HW31507, HW29132, HW32559, HW32558, HW31421, HW31420, HW31639, HW31893, HW30738, HW31892, HW31516, HW31685, HW31515, HW31551, HW29232, HW31082, HW29705, HW30950, HW31517, HW30096, HW30939, HW29682, HW30374, HW31514, HW30740, HW30951, HW30952, HW28177, HW30839, HW29992, HW30838, HW30934, HW30741, HW30802, HW29563, HW30804, HW30743, HW30841, HW30837, HW30736, HW29362, HW29683, HW29993, HW29999, HW30352, HW30375, HW30803, HW30739, HW30524, HW30101, HW29996, HW27679, HW27682, HW29234, HW29994, HW29658, HW29875, HW29988, HW29566, HW29564, HW28968, HW35368; GTIN 00888707002028, Serial Numbers: HW30595, HW35644, HW35630, HW35677, HW34377, HW31686, HW34948, HW35626, HW31457, HW31847, HW34196, HW34025, HW32201, HW31550, HW31260, HW30596, HW30962, HW30973, HW30472, HW30118, HW30461, HW30116, HW30397, HW30365, HW29246, HW30154, HW29788, HW30380, HW32193, HW35032, HW35189, HW35563, HW35488, HW35504, HW32220, HW35691, HW35730, HW35489, HW35215, HW35731, HW34520, HW35539, HW35583, HW34925, HW31944, HW34902, HW35325, HW34336, HW35545, HW34914, HW34868, HW34990, HW34877, HW35326, HW35209, HW30961, HW31547, HW30829, HW35480, HW35541, HW34946, HW35631, HW35704, HW35487, HW34211, HW30039, HW35494, HW35340, HW35632, HW35136, HW35690, HW35478, HW35695, HW34329, HW31846, HW35502, HW35582, HW32279, HW34673, HW34904, HW35669, HW35702, HW35670, HW35633, HW35676, HW35360, HW35641, HW34643, HW35506, HW34373, HW35628, HW35701, HW32499, HW31810, HW35708, HW35707, HW34034, HW34035, HW35668, HW35696, HW35705, HW30484, HW35703, HW34605, HW32471, HW35692, HW35683, HW35698, HW35211, HW35688, HW35693, HW35625, HW35697, HW35161, HW35131, HW34551, HW35689, HW34924, HW35534, HW35686, HW35634, HW35684, HW35682, HW35321, HW35681, HW35673, HW34610, HW35124, HW35687, HW35680, HW35639, HW35646, HW35678, HW29602, HW34957, HW35627, HW35674, HW35367, HW35647, HW35637, HW35586, HW35588, HW32357, HW35012, HW35492, HW34100, HW34226, HW35284, HW35535, HW31552, HW34344, HW34428, HW34609, HW35543, HW35635, HW35636, HW35671, HW35134, HW32273, HW34979, HW30400, HW35213, HW35548, HW35578, HW35587, HW35495, HW35624, HW35584, HW32159, HW34051, HW34066, HW34372, HW34866, HW35501, HW31843, HW29534, HW34555, HW35123, HW35099, HW35508, HW34446, HW34590, HW35540, HW35542, HW35577, HW35580, HW35464, HW35477, HW31411, HW35334, HW35500, HW35324, HW35629, HW32516, HW29273, HW31192, HW34933, HW35191, HW34997, HW35337, HW35339, HW31571, HW34939, HW35133, HW35385, HW35361, HW35499, HW34106, HW35143, HW34810, HW31235, HW35332, HW32335, HW34371, HW35319, HW35493, HW35220, HW29700, HW34113, HW34169, HW35481, HW35317, HW35138, HW35341, HW35357, HW35122, HW31548, HW35204, HW34596, HW34805, HW34369, HW35482, HW34824, HW34928, HW35214, HW35305, HW34944, HW35205, HW32393, HW34934, HW30814, HW34588, HW34326, HW35212, HW34845, HW35483, HW35141, HW29548, HW34808, HW35479, HW35128, HW34043, HW35125, HW34938, HW32224, HW34806, HW34949, HW34547, HW34947, HW34427, HW29211, HW31254, HW31409, HW31410, HW31448, HW34840, HW34923, HW34752, HW34903, HW34192, HW29593, HW30651, HW34550, HW34842, HW32457, HW34910, HW34913, HW34048, HW34522, HW34523, HW34908, HW34204, HW34425, HW31458, HW34046, HW34126, HW34521, HW30863, HW31455, HW32285, HW34187, HW34379, HW34556, HW34768, HW34331, HW34045, HW34207, HW34495, HW34603, HW31858, HW32259, HW32292, HW31456, HW34193, HW34205, HW29549, HW32248, HW32290, HW32472, HW34096, HW34151, HW32504, HW34179, HW34519, HW30208, HW32162, HW34546, HW34328, HW34342, HW34429, HW31849, HW34317, HW31991, HW34319, HW34347, HW34368, HW34376, HW34432, HW29249, HW30498, HW32192, HW32223, HW31835, HW34378, HW31255, HW32218, HW34117, HW34375, HW32281, HW30083, HW31831, HW31838, HW32190, HW34041, HW34121, HW32260, HW34181, HW32179, HW34036, HW34042, HW34186, HW29522, HW32199, HW34209, HW31249, HW29011, HW31579, HW32209, HW32479, HW34110, HW31103, HW34197, HW31834, HW34111, HW32216, HW34104, HW31115, HW32143, HW32204, HW34109, HW31223, HW31836, HW34047, HW32534, HW34038, HW31947, HW29815, HW32508, HW32511, HW32386, HW32400, HW32427, HW30406, HW31383, HW32161, HW32315, HW31452, HW31543, HW32231, HW32214, HW32426, HW34037, HW34040, HW29567, HW32228, HW32423, HW30320, HW31388, HW30867, HW31883, HW34044, HW30342, HW32468, HW32470, HW32510, HW32229, HW29154, HW32141, HW32230, HW32258, HW32286, HW31981, HW31451, HW30601, HW31778, HW32465, HW31844, HW31845, HW32464, HW32261, HW31842, HW32194, HW32206, HW29426, HW31445, HW31128, HW31267, HW29667, HW31104, HW32222, HW31117, HW32197, HW32210, HW31749, HW32068, HW31979, HW32172, HW32215, HW32121, HW31731, HW31447, HW30153, HW31840, HW34001, HW32189, HW30390, HW31559, HW32156, HW31449, HW31839, HW32164, HW32167, HW31261, HW31855, HW31986, HW32116, HW31830, HW31987, HW31988, HW31989, HW29465, HW29472, HW31777, HW31876, HW31549, HW30768, HW30769, HW30475, HW31134, HW30358, HW30430, HW31697, HW30821, HW31564, HW31578, HW31832, HW31750, HW31262, HW30398, HW31453, HW31819, HW31076, HW31446, HW30871, HW31573, HW31399, HW31414, HW31544, HW31545, HW31541, HW30402, HW30953, HW31694, HW30363, HW30597, HW30604, HW31242, HW30966, HW31107, HW30082, HW31020, HW29612, HW30494, HW30865, HW30211, HW30150, HW31113, HW31454, HW31219, HW31412, HW30410, HW30974, HW31132, HW31232, HW30360, HW31119, HW29588, HW31101, HW31387, HW31111, HW31226, HW30963, HW31087, HW31402, HW31390, HW31398, HW31114, HW29909, HW31386, HW30107, HW30607, HW30957, HW29199, HW29252, HW30968, HW31258, HW30818, HW31251, HW30178, HW30970, HW29603, HW31228, HW30366, HW30809, HW31072, HW29686, HW30471, HW29219, HW29635, HW30964, HW30976, HW31015, HW30956, HW30959, HW31105, HW30393, HW30599, HW30600, HW30820, HW30868, HW30038, HW30157, HW30825, HW29780, HW30267, HW29268, HW29629, HW30491, HW30960, HW29551, HW29670, HW29896, HW29650, HW30111, HW30652, HW30967, HW30810, HW30656, HW29676, HW30449, HW30958, HW30654, HW30864, HW30813, HW30830, HW30870, HW29212, HW29764, HW30455, HW30811, HW30816, HW30193, HW30815, HW30384, HW30486, HW30822, HW30383, HW30606, HW29701, HW30207, HW30598, HW30954, HW30955, HW30105, HW30824, HW30404, HW30603, HW30869, HW30462, HW30608, HW30866, HW30473, HW30108, HW30210, HW30372, HW30399, HW30823, HW29727, HW30408, HW30227, HW30110, HW30808, HW30369, HW29965, HW30407, HW30483, HW29820, HW29995, HW30231, HW30602, HW29604, HW30104, HW30392, HW30485, HW29283, HW30387, HW29798, HW30088, HW30370, HW30385, HW29825, HW30212, HW30463, HW29927, HW30109, HW30319, HW30409, HW30080, HW30367, HW30224, HW29787, HW29790, HW30115, HW30149, HW30200, HW30376, HW29866, HW30364, HW29276, HW29717, HW29877, HW29911, HW29266, HW29282, HW29387, HW29541, HW29668, HW29669, HW29675, HW29791, HW30040, HW30195, HW30205, HW30209, HW30213, HW29781, HW29832, HW29873, HW29983, HW29990, HW30214, HW30151, HW30085, HW30087, HW29485, HW29526, HW29903, HW30152, HW30201, HW30084, HW30225, HW29850, HW29271, HW30081, HW30156, HW29284, HW29586, HW29587, HW29600, HW29666, HW29702, HW29976, HW30086, HW29286, HW29229, HW29594, HW29932, HW29867, HW30113, HW30106, HW30197, HW29539, HW30198, HW30202, HW30204, HW30199, HW30147, HW30148, HW30155, HW30196, HW30194, HW29583, HW29685, HW29543, HW30117, HW29542, HW29741, HW29894, HW30102, HW29660, HW30103, HW29547, HW29755, HW29904, HW30089, HW29674, HW29562, HW29671, HW29784, HW29776, HW30079, HW29922, HW29302, HW29309, HW29900, HW29579, HW29595, HW29887, HW29546, HW29665, HW29782, HW29687, HW29279, HW29535, HW29285, HW29661, HW29754, HW29540, HW29552, HW29532, HW29664, HW29503, HW29274, HW29533, HW29287, HW29290, HW29538, HW29502, HW29536, HW29293, HW29265, HW29592, HW29590, HW29589, HW29915, HW32459, HW32419, HW29672, HW29673, HW34202, HW31244, HW32458, HW30771, HW30972, HW30971, HW32509, HW29248, HW34330, HW29786, HW34002, HW30770, HW28848, HW30382, HW29949, HW29928, HW29329, HW30378, HW30226, HW30112, HW29923, HW29895, HW29333, HW29310; GTIN 00888707002141, Serial Numbers: HW29475, HW34848, HW35711, HW34849; GTIN 00888707002264, Serial Numbers: HW35773, HW35741; GTIN 00888707002387, Serial Numbers: HW34953, HW31075, HW34349, HW34274, HW29783, HW34272, HW30978, HW32028, HW30046, HW30844, HW30412, HW30047, HW30843, HW29847, HW34352, HW35605, HW35713, HW35344, HW35717, HW35607, HW35718, HW32153, HW35148, HW29918, HW35342, HW34359, HW35721, HW35723, HW32513, HW35327, HW34358, HW35080, HW32282, HW35604, HW35144, HW35714, HW29870, HW35606, HW35596, HW35145, HW35537, HW34366, HW31342, HW29827, HW35603, HW34436, HW34357, HW32027, HW30045, HW34092, HW31089, HW34374, HW34968, HW32563, HW29794, HW31903, HW34350, HW32514, HW34091, HW32562, HW30505, HW34351, HW34273, HW34125, HW32163, HW31860, HW32149, HW31926, HW32026, HW28851, HW31080, HW31557, HW34004, HW30842, HW31924, HW29793, HW31683, HW30765, HW31556, HW31555, HW31091, HW30846, HW31161, HW31348, HW34006, HW30845, HW31095, HW31077, HW29961, HW30468, HW30591, HW29677, HW30590, HW29839, HW30091, HW30763, HW30090, HW30764, HW28585, HW30411, HW29256, HW29476, HW29689, HW29997, HW29614, HW29688, HW28354; GTIN 00888707002394, Serial numbers: HW35700, HW34561, HW35699, HW35776, HW35679, HW35779, HW35706, HW35651, HW35336, HW35510, HW35397, HW35611, HW34582, HW35609, HW35649, HW35648, HW35538, HW34461, HW35509, HW35186, HW35566, HW35610, HW35484, HW35150, HW35386, HW34557, HW34950, HW34996, HW34023, HW35149, HW31993, HW34955, HW35016, HW34991, HW34022, HW34395, HW32517, HW32238, HW32200, HW32207, HW34751, HW32217, HW32515, HW32239, HW32240, HW30512, HW31583, HW31584, HW31582, HW31992, HW30506, HW30508, HW30655, HW30509, HW30510, HW30469, HW30511, HW31580, HW30414, HW29275, HW29916, HW30413, HW30003, HW30042, HW29690, HW29691, HW29568, HW29569; GTIN 00888707002400, Serial Numbers: HW35653, HW35573, HW35154, HW35034, HW34210, HW34012, HW29693, HW32257, HW29518, HW30092, HW28671, HW29477, HW30043; GTIN 00888707002417, Serial Numbers: HW35763, HW35761, HW35756, HW35613, HW35348, HW34421, HW34380, HW30831, HW30381, HW29478, HW29729, HW30345, HW30322, HW30501, HW30324, HW30093, HW29694, HW29963; GTIN 00888707002424, Serial Numbers: HW31851, HW31698, HW30650, HW30767, HW30766, HW29480, HW29696; GTIN 00888707002431, Serial Numbers: HW35733, HW35119, HW34615, HW34381, HW34024, HW32029, HW32180, HW31123, HW30979, HW30858, HW30229, HW30114, HW29921, HW29917; GTIN 00888707002448, Serial Numbers: HW35597, HW35589, HW35511, HW35398, HW34617, HW34095, HW32546, HW31780, HW31779; GTIN 00888707002455, Serial Numbers: HW35544, HW34637, HW34026, HW32198, HW32025, HW31852, HW31856; GTIN 00888707002462, Serial Numbers: HW31700, HW30832, HW30658, HW30503, HW32035, HW31586; GTIN 00888707002486, Serial Numbers: HW31569, HW31872, HW31873, HW31875, HW31877, HW32053, HW34015, HW34251, HW34252, HW35725, HW35734, HW35745, HW35746, HW35760, HW35762, HW35765, HW35685, HW35593, HW35600, HW35612, HW35645, HW35650, HW35652, HW35740, HW35742, HW35743, HW35744, HW35568, HW35615, HW35618, HW35737, HW35738, HW35747, HW34818, HW34960, HW35616, HW35617, HW35748, HW35751, HW35752, HW35525, HW35526, HW35527, HW35528, HW35565, HW35567, HW35569, HW35570, HW35614, HW35518, HW35519, HW35520, HW35521, HW35522, HW35523, HW35512, HW35513, HW35514, HW35516, HW35517, HW35300, HW35303, HW35304, HW35306, HW35307, HW35297, HW35298, HW34945, HW34954, HW35171, HW35228, HW35230, HW35278, HW35282, HW35283, HW34280, HW34656, HW34867, HW34873, HW34879, HW34880, HW34882, HW34885, HW34888, HW34890, HW34897, HW34898, HW34900, HW34905, HW34909, HW34911, HW34912, HW34762, HW34767, HW34769, HW34779, HW34853, HW34738, HW34739, HW34742, HW34748, HW34749, HW34750, HW34753, HW34757, HW34758, HW34761, HW34719, HW34443, HW34444, HW34452, HW34454, HW34619, HW34620, HW34621, HW34622, HW34623, HW34624, HW34629, HW34630, HW34631, HW34636, HW34442, HW34354, HW34355, HW34356, HW34361, HW34364, HW34433, HW34434, HW34437, HW34439, HW34440, HW32221, HW32236, HW32243, HW32250, HW32506, HW34122, HW34123, HW34124, HW34275, HW34277, HW32085, HW32109, HW32165, HW32166, HW32195, HW32205, HW32211, HW32009, HW32011, HW32038, HW32046, HW32062, HW32065, HW32071, HW31995, HW31996, HW31997, HW31999, HW32002, HW32008, HW32010, HW32013, HW32032, HW32034, HW31444, HW31673, HW31927, HW31932, HW31934, HW31939, HW31940, HW31941, HW32057, HW31814, HW31817, HW31818, HW31820, HW31824, HW31825, HW31826, HW31428, HW31430, HW31431, HW31432, HW31433, HW31434, HW31435, HW31436, HW31437, HW31438, HW31509, HW31518, HW31519, HW31425, HW31427, HW30856, HW30872, HW30980, HW30981, HW31100, HW31110, HW31112, HW31121, HW31137, HW31138, HW30850, HW30851, HW30852, HW30853, HW30854, HW30855, HW30847, HW30848, HW30849, HW30982, HW30983; GTIN 00888707002493, Serial Numbers: HW35574, HW35529, HW29698, HW29483, HW32242, HW29697, HW31178, HW31821, HW35350, HW32246, HW32519, HW32518, HW31520, HW35660, HW35661, HW35785, HW35252, HW35591, HW35530, HW35401, HW35346, HW35400, HW35343, HW35345, HW35158, HW34915, HW35155, HW34367, HW34365, HW31823, HW31827, HW32086, HW31183, HW31353, HW31358, HW31164, HW31171, HW31173, HW30984, HW31094, HW31096, HW30873, HW30415, HW29633, HW30239, HW30121, HW29869, HW29936, HW30044, HW29926, HW29800, HW29797, HW29796; GTIN 00888707002516, Serial Numbers: HW35474, HW34524, HW31162, HW32507, HW31147, HW31124, HW31146, HW29213, HW29970; GTIN 00888707002523, Serial Numbers: HW35786, HW32102, HW35490, HW35218, HW35619, HW35531, HW35217, HW35216, HW35049, HW35040, HW35043, HW34986, HW34028, HW34027, HW32520, HW32521, HW34020, HW30859, HW34018, HW32036, HW32040, HW32030, HW31859, HW31704, HW31703, HW31702, HW31166, HW31129, HW30985, HW30661, HW30660, HW30095, HW30504, HW30094, HW29975, HW29974, HW29972, HW29699, HW29615; GTIN 00888707002530, Serial Numbers: HW31108, HW35590, HW35533, HW35532, HW35000, HW35162, HW32182, HW34559, HW31928, HW32263, HW32253, HW31828, HW30797, HW31440, HW30986, HW30666, HW30879, HW30881, HW30022, HW30594, HW29251, HW29802; GTIN 00888707002547, Serial Numbers: HW35251, HW34513, HW32373, HW35275, HW32385, HW34517, HW34518, HW34516, HW32557, HW32118, HW31408, HW29561, HW34514, HW31393, HW30502, HW31158, HW34383, HW34382, HW31156, HW31416, HW34054, HW31406, HW31413, HW32269, HW34053, HW32309, HW31429, HW31382, HW34050, HW31937, HW31994, HW31141, HW31426, HW31136, HW31140, HW32554, HW34049, HW34003, HW32524, HW32523, HW34008, HW34007, HW32336, HW31791, HW31553, HW32347, HW32522, HW30783, HW30619, HW30632, HW30796, HW30819, HW30826, HW30593, HW29931, HW29601; GTIN 00888707002554, Serial Numbers: HW35710, HW35585, HW35780, HW34484, HW34625, HW35316, HW35709, HW35772, HW35777, HW35758, HW35771, HW35782, HW35052, HW35458, HW35728, HW35764, HW35536, HW35759, HW34362, HW35323, HW35767, HW34337, HW35623, HW35732, HW34693, HW35775, HW35057, HW35359, HW34846, HW35362, HW34594, HW35061, HW35067, HW35620, HW35622, HW34884, HW34838, HW34889, HW35363, HW34732, HW35364, HW34591, HW34682, HW34833, HW34876; GTIN 00888707002585, Serial Numbers: HW35355, HW35575, HW35353, HW35576, HW35497, HW31170, HW31168, HW31079, HW31081, HW35095, HW35164, HW35351, HW35496, HW30833, HW30834, HW30948; GTIN 00888707002608, Serial Numbers: HW37178, HW37177; GTIN 00888707003063, Serial Numbers: HW42161, HW42230, HW42238, HW42160, HW42235, HW43009, HW41927, HW41953, HW42678, HW42720, HW42724, HW42793, HW42650, HW42651, HW42829, HW42831, HW42158, HW42089, HW42237, HW42790, HW42019, HW42219, HW42797, HW42018, HW41903, HW42006, HW42008, HW41926, HW41950, HW42007, HW42004, HW41928, HW41924, HW41925, HW41929, HW41911; GTIN 00888707003070, Serial Numbers: HW42110, HW41994, HW41895, HW41897, HW41936, HW41894, HW42666, HW42155, HW42228, HW42157, HW42328, HW42500, HW42326, HW42329, HW42331, HW42332, HW42000, HW42116, HW42330, HW41997, HW42153, HW41986, HW42225, HW42148, HW41934, HW42291, HW41993, HW41900, HW42252, HW42229, HW42022, HW42233, HW41899, HW42114, HW42232, HW41892, HW42005, HW42009, HW42021, HW42026, HW42099, HW42101, HW42103, HW42104, HW42111, HW42223, HW42001, HW41977, HW41937, HW41938, HW41939, HW41940, HW41941, HW41971, HW41972, HW41973, HW41974, HW41975, HW41901, HW42915, HW41991, HW42002, HW42307, HW42333, HW41932; GTIN 00888707003100, Serial Numbers: HW42671, HW42901, HW42663, HW42660, HW42313, HW41967, HW41962, HW42672, HW41965, HW41964, HW41968, HW41966, HW41961; GTIN 00888707003117, Serial Numbers: HW42292, HW42024, HW42025, HW42339, HW41960, HW42023, HW42337, HW42665, HW42061, HW42647, HW42335, HW42130, HW42091, HW41958, HW41955, HW41957, HW41959; GTIN 00888707003209, Serial Numbers: HW41982, HW41983, HW41984, HW41988, HW41990, HW41915, HW41916, HW41918, HW41919, HW41945, HW41946, HW41947, HW41948, HW41978, HW41979, HW41981; GTIN 00888707003254, Serial Numbers: HW42213, HW41944, HW41943; GTIN 00888707004558, Serial Numbers: HW35655, HW35656, HW34785, HW34759; GTIN 00888707005449, Serial Numbers: HW42861, HW42855, HW42003, HW42785, HW42017, HW42648, HW42011, HW41923, HW41970; GTIN 00888707005456, Serial Numbers: HW42893, HW43008, HW42891, HW42898, HW42892; GTIN 00888707005517, Serial Numbers: HW42789, HW42900; GTIN 00888707005555, Serial Numbers: HW41912, HW41907; GTIN 00888707005579, Serial Numbers: HW41921; GTIN 00888707005586, Serial Numbers: HW41920, HW41905, HW42895, HW42786, HW42598, HW41910; GTIN 00888707005951, Serial Numbers: HW35581, HW35833, HW35503, HW35838, HW35797, HW35948, HW35841, HW35933, HW35897, HW35888, HW35788, HW35893, HW35887, HW35791, HW35790, HW35891, HW35840, HW35934, HW35892, HW35946, HW35947, HW35896, HW35900, HW35843, HW35961, HW35952, HW35889, HW35890, HW35894, HW35842, HW35839, HW35826, HW35823, HW35768, HW35798, HW35799, HW35825, HW35830, HW35524, HW35895, HW35925, HW35955, HW35953, HW35846, HW35954, HW35845, HW35836, HW35898, HW35835, HW35837, HW35796, HW35794, HW35945, HW35844, HW35930, HW35926, HW35928, HW35921, HW35739, HW35792, HW35789, HW35787, HW35905, HW35906, HW35907, HW35831, HW35950, HW35949, HW35720, HW35821, HW35279, HW35960, HW35715, HW35951, HW35834, HW35827, HW35824, HW35822, HW35675, HW35795; GTIN 00888707005968, Serial Numbers: HW35855, HW35860, HW35976, HW35816, HW35857, HW35994, HW35859, HW35815, HW35975, HW35929, HW35801, HW35991, HW35935, HW35931, HW35938, HW35814, HW35969, HW35996, HW35997, HW35851, HW35803, HW35978, HW35924, HW35806, HW35995, HW35939, HW35990, HW35989, HW35993, HW35867, HW35875, HW35936, HW35970, HW35971, HW35988, HW35863, HW35968, HW35972, HW35804, HW35847, HW35864, HW35923, HW35927, HW35937, HW35856, HW35870, HW35966, HW35967, HW35853, HW35854, HW35862, HW35852, HW35861, HW35858, HW35817, HW35848, HW35807, HW35802, HW35850, HW35812, HW35805, HW35820, HW35811, HW35809, HW35818, HW35813, HW35808, HW35810; GTIN 00888707006019, Serial Numbers: HW36352, HW36350, HW36146, HW36221, HW36353, HW36159, HW36223, HW36351, HW36155, HW36167, HW36222, HW35877, HW35873, HW36158, HW36157, HW36156, HW31983, HW36118, HW36145, HW36120, HW36119, HW34720, HW35377, HW36116, HW36058, HW36117, HW35879, HW36000, HW35871, HW36114, HW36115, HW36113, HW36060, HW36061, HW36062, HW36059, HW36057, HW36056, HW35880, HW35869, HW35998, HW35878, HW35866, HW34974, HW35849; GTIN 00888707006309, Serial Numbers: HW35941, HW35942, HW35999, HW35884, HW35885, HW35876, HW35886, HW35882; GTIN 00888707006316, Serial Numbers: HW35868, HW35962, HW35963, HW35754, HW35865, HW35940, HW35965, HW35964, HW35958, HW35774; GTIN 00888707006323, Serial Numbers: HW36694, HW36072, HW36080, HW36841, HW36971, HW37235, HW37237, HW36800, HW36795, HW36074, HW36851, HW36904, HW36774, HW36770, HW36796, HW36747, HW35064, HW36339, HW36772, HW36688, HW36749, HW36127, HW36124, HW36340, HW36125, HW35977, HW36126, HW35736, HW36076, HW36079, HW36081, HW36123, HW36073, HW36024, HW36034, HW36032, HW36033, HW36029, HW36025, HW36028, HW35781, HW36876, HW36872, HW36873, HW36641, HW36137, HW36699, HW36207, HW37201, HW36735, HW36451, HW37236, HW36984, HW36651, HW36296, HW36292, HW36208, HW36236, HW36244, HW35729, HW36045, HW37251, HW36238, HW35654, HW36188, HW35390, HW36761, HW36794, HW36533, HW36548, HW36549, HW37202, HW36865, HW36798, HW36740, HW36687, HW36649, HW36539, HW36605, HW36375, HW36452, HW35722, HW36297, HW36189, HW36082, HW34982, HW37305, HW36905, HW36826, HW36374, HW36149, HW36147, HW36206, HW36210, HW37219, HW37205, HW36970, HW36376, HW36934, HW37206, HW36745, HW37204, HW36871, HW36336, HW36861, HW36742, HW36975, HW36697, HW36285, HW36284, HW37239, HW36894, HW36537, HW36535, HW36449, HW36446, HW36985, HW36973, HW36822, HW36801, HW36204, HW36205, HW36065, HW36078, HW36064, HW36344, HW36346, HW36042, HW37250, HW37279, HW36907, HW36908, HW36550, HW36551, HW36606, HW36607, HW36366, HW36367, HW36342, HW36343, HW36190, HW36191, HW36044, HW36043, HW36283, HW36368, HW36929, HW36909, HW36911, HW36922, HW36923, HW36693, HW36639, HW36849, HW36863, HW36843, HW36862, HW36730, HW36377, HW36674, HW36675, HW36676, HW36534, HW36529, HW36532, HW36267, HW35621, HW35727, HW36148, HW36140, HW36141, HW36142, HW35389, HW35384, HW35388, HW35383, HW35380, HW35382, HW36848, HW36647, HW36743, HW36771, HW36458, HW36527, HW36538, HW36696, HW36686, HW37307, HW37171, HW37173, HW36645, HW36738, HW36364, HW36144, HW36729, HW36450, HW36739, HW36821, HW36799, HW35899, HW36443, HW36746, HW36608, HW36797, HW36741, HW36187, HW36454, HW36737, HW36203, HW36453, HW36290, HW36291, HW35381, HW36234, HW35642, HW35770, HW36143, HW36245, HW36646, HW36744, HW36150, HW36151, HW36731, HW36347, HW36269, HW36268, HW36239, HW36202, HW36242, HW36209, HW35468, HW37249, HW36525, HW36046, HW36293, HW36295, HW36286, HW36931, HW36874, HW36338, HW36459, HW36913, HW36886, HW36892, HW36271, HW36524, HW36439, HW36077, HW37331, HW36457, HW36341, HW36456, HW36521, HW36866, HW36444, HW36887, HW36925, HW36875, HW35379, HW36867, HW36378, HW36864, HW36869, HW36842, HW36844, HW36846, HW36845, HW36522, HW36528, HW36824, HW36825, HW36380, HW36153, HW36661, HW36128, HW36640, HW36440, HW36379, HW36520, HW36455, HW36275, HW36274, HW36445, HW35712, HW36235, HW36441, HW36272, HW36136, HW36135, HW36134, HW36133, HW35031, HW35391, HW35392, HW36027, HW35783, HW35755, HW36030; GTIN 00888707006330, Serial Numbers: HW36818, HW36131, HW36556, HW36406, HW36328, HW36071, HW36255, HW36258, HW37193, HW37196, HW37215, HW37195, HW37245, HW36827, HW36788, HW37366, HW37234, HW36680, HW36432, HW36813, HW37213, HW36685, HW37241, HW36599, HW36678, HW37242, HW36702, HW36681, HW36823, HW36997, HW36673, HW37276, HW36626, HW36936, HW37350, HW37369, HW36937, HW37365, HW36942, HW36683, HW36988, HW37216, HW36722, HW36724, HW36881, HW37362, HW37246, HW37273, HW37199, HW36752, HW37243, HW37002, HW37222, HW37223, HW37225, HW37226, HW37228, HW37230, HW37231, HW37232, HW37233, HW36513, HW36816, HW36713, HW36966, HW35562, HW37218, HW36982, HW36996, HW36944, HW37210, HW37214, HW37212, HW35571, HW36428, HW36703, HW37194, HW36940, HW36664, HW36969, HW37005, HW36999, HW37000, HW37184, HW37185, HW37186, HW37187, HW37188, HW37189, HW37190, HW36834, HW36006, HW36718, HW36753, HW36902, HW36669, HW36964, HW36919, HW36817, HW36561, HW36562, HW36705, HW36832, HW36357, HW36723, HW36882, HW36636, HW36354, HW36836, HW36655, HW36945, HW36877, HW36597, HW36398, HW36812, HW36899, HW36900, HW36891, HW36897, HW36898, HW36901, HW36719, HW36437, HW36670, HW36710, HW36407, HW36727, HW36726, HW36883, HW36557, HW36831, HW36750, HW36829, HW36830, HW36833, HW36835, HW36837, HW36839, HW36840, HW36853, HW36854, HW36855, HW36601, HW36708, HW36709, HW36725, HW36728, HW36785, HW36815, HW36511, HW36704, HW35829, HW36256, HW36789, HW36720, HW36325, HW36786, HW36755, HW36564, HW36754, HW36248, HW36624, HW36711, HW36194, HW36254, HW36660, HW36596, HW36559, HW36019, HW36405, HW36598, HW36600, HW36672, HW36510, HW36712, HW32505, HW36409, HW36595, HW36628, HW36630, HW36659, HW36514, HW36555, HW36560, HW36553, HW36192, HW36434, HW36530, HW36363, HW36637, HW36671, HW36505, HW36662, HW36633, HW36429, HW35726, HW36251, HW36253, HW36631, HW36326, HW36506, HW36603, HW36625, HW36233, HW36316, HW31699, HW36360, HW36430, HW36602, HW36334, HW36435, HW36509, HW36411, HW36500, HW36507, HW36563, HW36310, HW36438, HW36419, HW36503, HW36397, HW36404, HW36501, HW36504, HW36436, HW36502, HW36355, HW36403, HW36396, HW36412, HW36433, HW35769, HW36314, HW35784, HW36401, HW36408, HW36400, HW36356, HW36410, HW36299, HW36300, HW36333, HW31419, HW36199, HW36307, HW36335, HW36361, HW35766, HW36298, HW36309, HW36359, HW36177, HW36313, HW36362, HW36358, HW36138, HW36185, HW36320, HW36317, HW35872, HW36319, HW36201, HW36186, HW36005, HW36226, HW36324, HW36332, HW36227, HW36318, HW36321, HW36323, HW36225, HW36231, HW35749, HW36276, HW36303, HW36230, HW36232, HW36260, HW36070, HW36228, HW36259, HW36195, HW36007, HW36252, HW36229, HW36069, HW36224, HW36178, HW36176, HW36021, HW36200, HW36179, HW36184, HW36020, HW36193, HW36013, HW36175, HW36197, HW36198, HW36132, HW36180, HW35396, HW36010, HW36018, HW35393, HW36174, HW31733, HW36009, HW36008, HW37200, HW37367, HW36554, HW36654, HW36677, HW36515, HW36508, HW36399, HW36305; GTIN 00888707006347, Serial Numbers: HW37432, HW37434, HW37440, HW37447, HW37263, HW37358, HW37378, HW37381, HW37386, HW37387, HW36963, HW37006, HW37007, HW37008, HW37009, HW37010, HW37154, HW37155, HW37156, HW37157, HW37161, HW37162, HW37164, HW37165, HW37166, HW37168, HW37169, HW36808, HW36809, HW36810, HW36918, HW36946, HW36948, HW36949, HW36950, HW36951, HW36952, HW36953, HW36954, HW36955, HW36956, HW36959, HW36960, HW36961, HW36962, HW36757, HW36759, HW36760, HW36763, HW36764, HW36765, HW36766, HW36767, HW36768, HW36769, HW36775, HW36776, HW36777, HW36778, HW36779, HW36780, HW36781, HW36782, HW36783, HW36784, HW36790, HW36791, HW36792, HW36793, HW36804, HW36805, HW36565, HW36582, HW36583, HW36584, HW36585, HW36586, HW36612, HW36615, HW36616, HW36618, HW36619, HW36620, HW36621, HW36622, HW36623, HW36611, HW36566, HW36568, HW36569, HW36570, HW36571, HW36572, HW36573, HW36574, HW36575, HW36576, HW36577, HW36578, HW36579, HW36581, HW36040, HW36048, HW36049, HW36050, HW36051, HW36052, HW36053, HW36054, HW36055, HW36088, HW36089, HW36092, HW36093, HW36094, HW36095, HW36096, HW36097, HW36100, HW36101, HW36102, HW36103, HW36105, HW36106, HW36107, HW36108, HW36109, HW36110, HW36111, HW36112, HW36170, HW36540, HW36541, HW36542, HW36543, HW36544, HW35920, HW36001, HW36002, HW36003, HW36016, HW36035, HW36036, HW36037, HW36038, HW36039, HW35902, HW35903, HW35908, HW35909, HW35910, HW35911, HW35912, HW35913, HW35914, HW35915, HW35918; GTIN 00888707006392, Serial Numbers: HW36870, HW36516, HW37272, HW36979, HW36858, HW36860, HW36652, HW36916, HW36820, HW36658, HW36859, HW36803, HW36819, HW36219, HW36668, HW36590, HW36692, HW36700, HW36802, HW35986, HW36517, HW36656, HW36589, HW36218, HW36588, HW36587, HW36289, HW36418, HW36311, HW35987, HW36417, HW36217, HW36220, HW36215, HW36213, HW36214, HW36216, HW36086, HW36085, HW36066; GTIN 00888707006408, Serial Numbers: HW37349, HW36706, HW37254, HW36690, HW36814, HW36689, HW37253, HW36414, HW37252, HW36171, HW37159, HW37209, HW36994, HW37183, HW36266, HW36638, HW36349, HW37160, HW36992, HW36806, HW36993, HW36212, HW36707, HW36644, HW36172, HW36666, HW36665, HW36657, HW36261, HW36130, HW36265, HW36413, HW36594, HW36173, HW36348, HW36416, HW36415, HW36129, HW36263, HW36264, HW34292, HW36250, HW36247, HW36084, HW36087, HW36083; GTIN 00888707006446, Serial Numbers: HW35982, HW36494, HW37375, HW37347, HW37345, HW37339, HW37342, HW37340, HW37329, HW37335, HW37262, HW37261, HW37257, HW37255, HW36926, HW36977, HW36914, HW36957, HW36976, HW37256, HW36915, HW36924, HW36974, HW36958, HW36610, HW36493, HW37260, HW37301, HW36546, HW37258, HW36519, HW36609, HW36547, HW36495, HW36545, HW36474, HW36499, HW36473, HW36497, HW36496, HW35985, HW36492, HW36486, HW36487, HW36489, HW36482, HW36483, HW36484, HW36481, HW36472, HW36480, HW36464, HW36470, HW36471, HW36465, HW36388, HW36386, HW36469, HW36387, HW36462, HW36460, HW35979, HW35980, HW36385, HW35984, HW36381, HW36383, HW36287, HW36278, HW36163, HW36164, HW36382, HW36384, HW36162, HW36161, HW36160; GTIN 00888707006675, Serial Numbers: HW36139; GTIN 00888707006736, Serial Numbers: HW36041; GTIN 00888707006767, Serial Numbers: HW36281; GTIN 00888707007290, Serial Numbers: HW36593, HW36591, HW36122, HW36104, HW35919, HW35800, HW36427, HW36425, HW35922, HW36422, HW36421; GTIN 00888707009140, Serial Numbers: HW37437, HW36369, HW37462, HW37461, HW37453, HW37355, HW37363, HW37456, HW37454, HW37455, HW37452, HW36691, HW37379, HW37380, HW36448, HW37336, HW37318, HW37227, HW37348, HW37308, HW37356, HW37326, HW37323, HW37321, HW37317, HW37311, HW35916, HW37268, HW37270, HW37176, HW37220, HW37377, HW37269, HW37266, HW37175, HW37174
FEI Number 3007042319
Recalling Firm/
Manufacturer
Heartware, Inc.
14400 Nw 60th Ave
Miami Lakes FL 33014-2807
For Additional Information ContactMCS Customer Service
877-367-4823
Manufacturer Reason
for Recall
Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm may sound and the recommended troubleshooting actions and (2) instructions relating to the useful life of the HVAD system components
FDA Determined
Cause 2
Labeling design
ActionHeartware issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 10/31/2023 via letter (UPS). The notice explained the issue and informed them of the updates to the HVAD IFU and PM (patient manual).
Quantity in Commerce10287 units
DistributionWorldwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-