| Date Initiated by Firm | November 29, 2023 |
| Date Posted | December 20, 2023 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0549-2024 |
| Recall Event ID |
93474 |
| 510(K)Number | K183311 |
| Product Classification |
Pump, infusion - Product Code FRN
|
| Product | Ivenix Infusion System (IIS), Large Volume Pump LVP-0004 |
| Code Information |
UDI-DI 00811505030320
Serial numbers
2213800363
2213800343
2213800328
2213800222
2213800173
2212201120
2212201119
2212201118
2212201115
2212201113
2212201111
2212201100
2212201095
2212201093
2212201088
2212201076
2212201074
2212201061
2212201059
2212201046
2212201029
2212201024
2212201003
2212201002
2212200993
2212200990
2212200987
2212200986
2212200982
2212200978
2212200958
2212200952
2212200945
2212200938
2212200929
2212200914
2205900177
2205900176
2205900170
2205900167
2205900158
2205900152
2205900150
2205900143
2205900133
2205900132
2205900128
2205900126
2205900125
2205900113
2205900112
2205900110
2205900109
2205900108
2205900107
2205900106
2205900098
2205900096
2205900094
2205900091
2205900090
2205900074
2205900072
2205900071
2205900067
2205900062
2205900061
2201811620
2201811615
2201811614
2201811605
2118200011
2201811603
2201811600
2201811598
2201811597
2201811596
2201811594
2201811590
2201811585
2201811584
2201811582
2201811579
2201811577
2201811568
2201811567
2201811561
2201811560
2119300422
2201811554
2119300424
2201811550
2201811549
2201811548
2201811545
2201811542
2201811541
2119300431
2201811539
2201811533
2119300434
2119300435
2119300436
2119300437
2119300438
2201811532
2201811530
2201811529
2119300442
2201811527
2201811526
2201811524
2124300052
2201811520
2201811519
2201811517
2201811514
2201811510
2201811507
2201811503
2124300060
2201811499
2201811493
2201811488
2201811484
2201811482
2201811481
2201811479
2201811477
2201811476
2201811474
2201811471
2201811467
2201811460
2201811459
2201811457
2124300076
2201811454
2201811453
2201811451
2201811449
2201811445
2201811444
2201811443
2124300084
2201811439
2201811434
2201811433
2201811432
2124300089
2201811431
2201811430
2201811429
2201811428
2124300094
2201811419
2124300096
2201811416
2201811415
2201811414
2124300100
2201811412
2201811411
2201811409
2201811405
2201811404
2201811400
2201811399
2201811398
2124300109
2124300110
2201811397
2201811395
2201811391
2201811390
2201811388
2201811385
2201811384
2201811383
2201811372
2201811367
2201811365
2201811364
2201811363
2124300124
2201811362
2201811361
2201811360
2201811359
2124300129
2201811355
2201811352
2201811349
2201811348
2124300134
2201811347
2124300136
2124300137
2201811339
2201811335
2201811334
2201811327
2201811325
2201811322
2124300144
2201811318
2201811309
2136400804
2136400803
2136400800
2136400799
2124300151
2136400798
2136400796
2136400795
2136400794
2136400792
2136400791
2136400789
2136400787
2136400786
2136400784
2136400781
2136400780
2136400779
2136400776
2136400775
2136400774
2136400772
2136400767
2136400765
2136400758
2136400757
2136400753
2136400752
2136400751
2136400750
2136400749
2136400747
2136400746
2136400744
2136400743
2136400742
2124300183
2136400736
2136400735
2136400733
2136400732
2136400731
2136400730
2136400728
2136400727
2136400725
2136400723
2136400722
2136400720
2136400718
2136400717
2136400715
2136400712
2136400711
2136400708
2136400707
2136400702
2136400700
2136400699
2136400690
2136400688
2136400683
2136400680
2136400679
2136400676
2136400674
2136400671
2136400670
2136400666
2125900452
2136400660
2136400659
2136400658
2136400657
2136400656
2136400655
2136400654
2136400651
2136400650
2136400649
2136400648
2136400647
2136400643
2136400641
2136400640
2136400636
2136400634
2136400633
2136400630
2136400629
2136400628
2136400626
2136400624
2136400617
2136400615
2136400612
2136400611
2136400610
2136400604
2136400600
2136400598
2136400597
2136400596
2125900486
2136400595
2136400594
2136400592
2136400591
2136400590
2136400588
2136400587
2125900494
2136400586
2136400583
2136400582
2136400581
2136400579
2136400578
2136400574
2136400573
2136400571
2136400570
2136400567
2136400566
2136400562
2136400556
2135000750
2135000749
2135000748
2135000746
2135000742
2135000741
2135000739
2125900516
2135000737
2135000734
2135000730
2135000729
2135000728
2135000727
2135000725
2135000722
2135000721
2135000718
2135000717
2135000715
2135000714
2135000710
2135000707
2135000702
2135000701
2135000699
2135000697
2135000696
2135000695
2135000690
2135000689
2135000688
2135000687
2135000683
2135000682
2135000680
2135000673
2135000670
2135000664
2135000663
2135000661
2135000660
2135000659
2135000658
2135000656
2135000655
2135000654
2135000645
2135000639
2135000636
2135000633
2135000632
2135000628
2135000625
2135000624
2135000623
2135000621
2135000615
2135000613
2135000612
2135000610
2135000604
2135000603
2135000602
2135000600
2135000599
2135000597
2135000589
2135000588
2135000586
2135000585
2135000583
2135000581
2135000575
2135000573
2135000572
2135000571
2135000570
2135000568
2135000564
2135000563
2135000562
2135000561
2135000558
2135000557
2135000555
2135000554
2135000546
2135000542
2135000540
2135000533
2135000529
2135000526
2135000523
2135000520
2135000512
2135000509
2130700984
2130700975
2130700961
2213800316
2213800288
2213800282
2213800250
2213800239
2213800184
2212201117
2212201114
2212201108
2212201105
2212201104
2212201098
2212201097
2212201086
2212201082
2212201081
2212201073
2212201069
2212201068
2212201064
2212201058
2212201057
2212201056
2212201050
2212201042
2212201034
2212201027
2212201019
2212201016
2212201013
2212201012
2212200998
2212200996
2212200995
2212200991
2212200989
2212200988
2212200985
2212200976
2212200973
2212200969
2212200966
2212200962
2212200961
2212200947
2212200946
2212200944
2212200939
2212200937
2212200935
2212200931
2212200930
2212200928
2212200926
2212200924
2212200923
2212200922
2212200921
2212200916
2212200910
2212200909
2212200905
2205900173
2205900164
2205900162
2205900159
2205900151
2205900148
2118200008
2205900138
2205900130
2205900122
2205900121
2205900120
2205900117
2205900102
2205900100
2205900099
2205900087
2205900086
2119300414
2119300415
2205900085
2119300417
2205900083
2205900082
2205900081
2119300421
2205900079
2205900075
2205900068
2205900065
2205900064
2201811618
2201811616
2119300429
2201811609
2201811604
2201811595
2201811593
2201811592
2201811589
2201811583
2201811581
2201811578
2119300439
2201811573
2201811572
2201811566
2201811559
2201811558
2201811557
2201811553
2201811547
2201811543
2201811536
2201811535
2201811528
2124300058
2201811525
2201811522
2201811513
2201811511
2124300063
2201811506
2201811505
2124300066
2201811502
2201811495
2201811494
2201811491
2201811483
2201811480
2124300073
2124300074
2201811478
2201811475
2201811473
2124300078
2124300079
2201811469
2201811468
2201811466
2201811462
2201811456
2124300085
2201811455
2201811452
2201811450
2201811441
2124300090
2201811440
2201811438
2201811436
2201811426
2201811424
2201811422
2201811407
2201811402
2124300099
2201811401
2124300101
2124300102
2124300103
2201811396
2201811393
2201811386
2124300107
2201811382
2201811381
2201811377
2201811375
2124300112
2201811370
2201811366
2201811356
2201811354
2201811351
2201811350
2201811346
2201811342
2201811338
2201811333
2124300123
2201811330
2124300125
2201811329
2124300127
2201811316
2201811314
2201811310
2201811308
2124300132
2124300133
2136400797
2136400790
2136400785
2136400782
2124300138
2136400778
2124300140
2124300141
2136400773
2124300143
2136400766
2136400763
2136400761
2136400760
2136400755
2136400741
2136400734
2136400721
2136400716
2136400701
2136400698
2136400697
2124300156
2136400696
2136400691
2124300159
2136400689
2136400685
2136400677
2136400668
2136400645
2136400642
2136400638
2124300167
2136400621
2136400619
2136400618
2136400614
2136400613
2136400609
2136400608
2136400607
2136400601
2136400589
2124300178
2136400585
2136400577
2136400572
2124300182
2136400569
2124300184
2136400568
2136400559
2135000745
2135000744
2135000740
2125900426
2135000738
2135000735
2135000732
2135000724
2125900431
2125900432
2135000723
2135000713
2125900435
2125900436
2135000711
2135000708
2135000703
2135000700
2135000694
2125900442
2135000693
2135000691
2135000685
2135000684
2135000679
2125900448
2135000677
2135000669
2135000666
2135000665
2135000657
2135000653
2135000651
2135000649
2135000644
2135000638
2125900459
2135000634
2135000631
2125900462
2135000620
2135000617
2125900465
2135000616
2135000608
2135000605
2135000601
2135000598
2135000595
2135000593
2135000592
2135000590
2135000587
2135000582
2125900477
2135000553
2125900479
2135000547
2135000539
2135000536
2125900483
2135000535
2135000531
2135000528
2135000527
2135000516
2135000514
2125900490
2135000513
2125900492
2135000502
2135000501
2130700983
2130700962
2130700958
2130700955
2130700954
2130700952
2130700950
2125900502
2130700948
2130700945
2130700943
2125900506
2130700942
2130700939
2130700936
2125900510
2130700933
2130700930
2130700927
2130700926
2130700924
2130700916
2125900517
2130700915
2130700911
2130700910
2130700908
2130700902
2125900523
2130700898
2130700897
2130700892
2125900527
2130700891
2130700886
2130700878
2130700876
2125900532
2130700870
2130700868
2130700867
2125900536
2130700865
2130700863
2130700853
2125900540
2130700851
2130700848
2130700847
2130700841
2130700839
2130700838
2130700835
2125900620
2125900549
2125900619
2125900610
2125900609
2125900553
2125900554
2125900555
2125900608
2125900606
2125900603
2125900600
2125900599
2125900596
2125900590
2125900587
2125900584
2125900581
2125900575
2125900569
2213800358
2213800355
2213800347
2213800345
2213800339
2213800336
2213800329
2213800324
2213800319
2213800314
2213800309
2213800279
2213800277
2213800274
2213800271
2213800261
2213800258
2213800255
2213800229
2213800228
2213800220
2213800208
2213800204
2213800190
2213800177
2213800172
2213800166
2212201112
2212201110
2212201107
2212201106
2212201099
2212201094
2212201077
2212201075
2212201070
2212201065
2212201049
2212201047
2212201045
2212201041
2212201039
2212201032
2212201026
2212201018
2212201014
2212201007
2212200992
2212200984
2212200977
2212200974
2212200918
2213800337
2213800334
2213800323
2213800206
2212201079
2212201048
2212201033
2205900178
2205900136
2205900093
2201811575
2201811496
2201811489
2136400762
2136400605
2135000594
2125900566
2124300177
2124300174
2124300150
2124300114
2124300111
2124300087
2124300086
2124300082
2124300070
2124300068
2119300432
2119300430
2119300428
2119300426
2119300411
2213800352
2213800338
2213800307
2213800286
2213800264
2213800238
2213800213
2213800197
2213800188
2212201060
2212200979
2212200972
2212200933
2136400603
2124300157
2213800298
2213800297
2213800193
2212201116
2212200970
2205900161
2135000607
2130700826
2125900501
2125900497
2213800342
2213800317
2213800212
2213800231
|
Recalling Firm/ Manufacturer |
Fresenius Kabi USA, LLC 50 High St Ste 50 North Andover MA 01845-2620
|
| For Additional Information Contact | Customer Support 978-775-8100 |
Manufacturer Reason for Recall | The device may experience mechanical interference on the Fluid Valve pins, which will trigger a Pump Problem alarm. The issue may lead to delay or interruption of therapy. Depending on the therapy and duration of delay/interruption, the issue may lead to serious harm or death. |
FDA Determined Cause 2 | Process change control |
| Action | On November 29, 2023, the firm began notifying affected customers via" URGENT-Medical Device Recall" letter dated November 28, 2023.
Customers were provided with precautions to take until the affected LVPs are able to be returned for repair or replacement of internal hardware.
Customers will be contacted by a Fresenius Kabi service representative to create a plan for addressing any affected LVPs at their facility. This effort is expected to begin in January 2024. Fresenius Kabi will repair LVPs that may encounter mechanical interference causing the Fluid Valve Actuation System not to move properly. If the LVP cannot be repaired, it will be replaced at no cost to the customer.
If you have any questions contact Customer support at, email: Ivenix_support@fresenius-kabi.com or phone at 1-978-775-8100. |
| Quantity in Commerce | 938 US |
| Distribution | US distribution to states of: CA, CO, NJ, UT, and WI.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = FRN
|