• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall 3M Specialty Blade Assembly

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall 3M Specialty Blade Assemblysee related information
Date Initiated by FirmJanuary 12, 2024
Date PostedFebruary 21, 2024
Recall Status1 Open3, Classified
Recall NumberZ-1169-2024
Recall Event ID 93825
Product Classification Razor, surgical - Product Code LWK
Product3M Specialty Blade Assembly, REF 9660,70-2011-8462-2, 7100213114, used with used with 3M Surgical Clipper Catalog Number 9661L
Code Information UDI/DI 50707387789003 (shipper), 30707387789009 (tray), all lot codes
Recalling Firm/
Manufacturer
3M Company - Health Care Business
3m Center 2510 Conway Ave
Bldg 275-5w-06
Saint Paul MN 55144-0001
For Additional Information Contact3M Helath Care Customer Helpline
800-228-3957 Ext. 7
Manufacturer Reason
for Recall
3M has added the following cautionary statement regarding the use of the product: "Caution - To reduce the risk of burns, which if not avoided may result in minor injury: " Do not use on injured skin. " Do not leave clipper running unused in air without applying to skin for more than 1 minute as blade temperature may exceed 41C (105.8F). " Do not leave the clipper blade applied to the same location on the patient s skin for longer than 1 second as potential injury may occur; continuous motion of the clipper during use is required. Note: If clipper is left running unused in air for 10 minutes without applying to skin, the Model 9690 blade maximum temperature may reach 60C (140F) and the Model 9660 blade maximum temperature may reach 77C (170.6F)."
FDA Determined
Cause 2
Under Investigation by firm
Action3M issued a Medical Device Safety Notification - First Notice to its consignees on 01/12/2024 via USPS certified mail and email. The notice explained the issue and its decision to notify its consignees of an added caution statement regarding the use of the product: Caution - To reduce the risk of burns, which if not avoided may result in minor injury: " Do not use on injured skin. " Do not leave clipper running unused in air without applying to skin for more than 1 minute as blade temperature may exceed 41C (105.8F). " Do not leave the clipper blade applied to the same location on the patient s skin for longer than 1 second as potential injury may occur; continuous motion of the clipper during use is required. Note: If clipper is left running unused in air for 10 minutes without applying to skin, the Model 9690 blade maximum temperature may reach 60C (140F) and the Model 9660 blade maximum temperature may reach 77C (170.6F). Users were asked to take the following actions: "" Ensure this safety notification is passed to all those who need to be aware within your organization or to any department where the affected product has been transferred. " Continue to use the product as intended, following the additional caution statement above. " Complete the attached Customer Acknowledgement Form acknowledging that you have read and understood this letter and the additional caution statement. " Email the completed form to 3MFieldSafetyAction@montage72.com." For additional questions or need to report an adverse event, contact the 3M Health Care Customer Helpline at 1-800-228-3957, option 7.
Quantity in Commerce52,833,750 blades
DistributionWorldwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-