• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall Medline

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Medlinesee related information
Date Initiated by FirmNovember 15, 2023
Date PostedFebruary 25, 2024
Recall Status1 Open3, Classified
Recall NumberZ-1126-2024
Recall Event ID 93800
Product Classification Eye tray - Product Code OJK
ProductMedline Kits, trays, and packs labeled as follows: a) CATARACT PACK, REF VAL011CTASD; b) CATARACT PACK, REF DYNJ68875F; c) CATARACT PACK, REF DYNJ54869F; d) CATARACT PREP PACK, REF DYNJ44917L; e) E-BLEPH CDS, REF CDS983585G; f) EYE CDS, REF CDS983916G; g) EYE KIT, REF DYKM2222; h) EYE PLASTIC PACK, REF DYNJ47859C; i) K PACK, REF LYN030KAGVD; j) OCULOPLASTY PACK, REF DYNJ64572A; k) OL OCULAR, REF DYNJ906183J; l) OL OCULAR, REF DYNJ906183K; m) P M CATARACT, REF DYNJ44917G; n) P M CATARACT, REF DYNJ44917I; o) P PACK, REF LYN030POGVD; p) R VITRECTOMY, REF EYP087W; q) VITRECTOMY PACK, REF DYNJ68884B
Code Information a) REF VAL011CTASD, UDI/DI 40195327152032 (case), 10195327152031 (unit), Lot Numbers: 22LBB812, 22LBH455; b) REF DYNJ68875F, UDI/DI 40195327382125 (case), 10195327382124 (unit), Lot Numbers: 23FBQ273, 23GBS122, 23GBS486, 23HBP752; c) REF DYNJ54869F, UDI/DI 40195327407286 (case), 10195327407285 (unit), Lot Numbers: 23FBQ273, 23GBS122, 23GBS486, 23HBP752; d) REF DYNJ44917L, UDI/DI 40195327472604 (case), 10195327472603 (unit), Lot Numbers: 23HBS811; e) REF CDS983585G, UDI/DI 40193489460781 (case), 10193489460780 (unit), Lot Numbers: 22DBI589, 22HBA111, 22JBQ351, 23EBG770; f) REF CDS983916G, UDI/DI 40193489803861 (case), 10193489803860 (unit), Lot Numbers: 22CBV428, 22GBP859, 22HBP383, 23EBP259, 23GBG620, 23IBS207, 23JBW354; g) REF DYKM2222, UDI/DI 40195327337255 (case), 10195327337254 (unit), Lot Numbers: 23DBI071; h) REF DYNJ47859C, UDI/DI 40193489916066 (case), 10193489916065 (unit), Lot Numbers: 23EBA347; i) REF LYN030KAGVD, UDI/DI 40193489956574 (case), 10193489956573 (unit), Lot Numbers: 22ABH632, 22ABW357, 22CBL660, 22IBS624; j) REF DYNJ64572A, UDI/DI 40195327112555 (case), 10195327112554 (unit), Lot Numbers: 22GBL794; k) REF DYNJ906183J, UDI/DI 40195327279449 (case), 10195327279448 (unit), Lot Numbers: 22LMC016, 22LMH698, 23AME088, 23AMF669, 23BMH365, 23CMB182, 23EMA555; l) REF DYNJ906183K, UDI/DI 40195327385522 (case), 10195327385521 (unit), Lot Numbers: 23FMC104, 23FME336, 23GMD288, 23HMH680; m) REF DYNJ44917G, UDI/DI 40195327031481 (case), 10195327031480 (unit), Lot Numbers: 22BBV644; n) REF DYNJ44917I, UDI/DI 40195327156375 (case), 10195327156374 (unit), Lot Numbers: 22FBN595, 22IBK002; o) REF LYN030POGVD, UDI/DI 22ABM649, 22ABW358, 22BBK347, 22CBT855, 22EBR210, 22HBR247, 22KBH816, 22LBA956, 22OBF733, 23ABB019, 23BBQ119; p) REF EYP087W, UDI/DI 50653160050221 (case), 00653160050226 (unit), Lot Numbers: 40195327037254 (case), 10195327037253 (unit), Lot Numbers: 22CBF187, 22EBE045, 22FBO783, 22GBZ358, 22IBF001, 22JBC220, 22KBN864, 22OBE478, 23ABG110, 23BBR173, 23DBH202; q) REF DYNJ68884B, UDI/DI 40195327059010 (case), 10195327059019 (unit), Lot Numbers: 22GBG399
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactKarin Johnson
886-359-1704
Manufacturer Reason
for Recall
Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionMEDLINE issued an IMMEDIATE ACTION REQUIRED NURSE ASSIST PRODUCT RECALL notice to its consignees on 11/16/2023 via USPS first class mail and email. The notice explained the problem, requested log on to their portal to identify the products and check their inventory so the product could be quarantined. Once Medline receives the completed response form, over-labels will be shipped to the consignee. Medline Industries, LP instructed customers to over-label the convenience kits with a warning label indicating to remove the sodium chloride irrigation or Sterile water and replace from a different sterile supply. Distributors or consignees who further distributed the affected products were directed to notify their customers. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity in Commerce5724 units
DistributionDistribution US nationwide including Puerto Rico, Canada, UAE, and Panama
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-