• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Zimmer Periarticular Locking Plate System Distal Lateral Fibula Locking Plate

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Zimmer Periarticular Locking Plate System Distal Lateral Fibula Locking Platesee related information
Date Initiated by FirmMarch 20, 2024
Date PostedApril 30, 2024
Recall Status1 Open3, Classified
Recall NumberZ-1719-2024
Recall Event ID 94276
510(K)NumberK070906 
Product Classification Screw, fixation, bone - Product Code HWC
ProductZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 12 Holes, 184 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 47-2357-018-12
Code Information UDI: (01)00889024088603(17)290512(10)64381071 (01)00889024088603(17)310324(10)65038410 (01)00889024088603(17)310324(10)65038411 (01)00889024088603(17)310324(10)65038412 (01)00889024088603(17)310821(10)65193244 (01)00889024088603(17)311018(10)65233124 (01)00889024088603(17)311017(10)65233125 Lot Numbers: 64381071 65038410 65038411 65038412 65193244 65233124 65233125
FEI Number 1000220733
Recalling Firm/
Manufacturer
Zimmer, Inc.
1800 W Center St
Warsaw IN 46580-2304
For Additional Information Contact411 Technical Services
574-371-3071
Manufacturer Reason
for Recall
Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
FDA Determined
Cause 2
Under Investigation by firm
ActionZimmer issued Urgent Medical Device Recall Letter to Distributors Risk Managers and Surgeons on March 20, 2024 via mail/email. Letter states reason for recall, health risk and action to take: Our records indicate that you may have received one or more of the affected products. The affected units were distributed between March 2019 and January 2024. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice for hospitals or surgeons and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. 4. Retain a copy of the acknowledgeme
DistributionWorldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = HWC
-
-