| | Class 2 Device Recall Smith & Nephew REAL INTELLIGENCE" CORI" (CORI) |  |
| Date Initiated by Firm | May 07, 2024 |
| Date Posted | June 11, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2135-2024 |
| Recall Event ID |
94635 |
| 510(K)Number | K240139 |
| Product Classification |
Orthopedic stereotaxic instrument - Product Code OLO
|
| Product | CORI Starter Kit, Real Intelligence CORI, CORI STARTER KIT - UNIVERSAL.Intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures
Product Numbers:
(1) ROB10000 CORI Starter Kit
(2) ROB10024 Real Intelligence CORI
(3) ROB10197 CORI STARTER KIT - UNIVERSAL
Note: the three product numbers cover the three different configurations in which the CORI Console is sold |
| Code Information |
UDI-DI:
(1) ROB10000t 00885556755068;
(2) ROB10024 00885556757420;
(3) ROB10197 00885556796849
Serial Numbers:
SN000001 SN000048 SN000095 SN000142 SN000189 SN000236
SN000002 SN000049 SN000100 SN000143 SN000190 SN000237
SN000003 SN000050 SN000101 SN000144 SN000191 SN000238
SN000004 SN000051 SN000102 SN000145 SN000192 SN000239
SN000005 SN000052 SN000103 SN000146 SN000193 SN000240
SN000006 SN000053 SN000096 SN000147 SN000194 SN000241
SN000007 SN000054 SN000097 SN000148 SN000195 SN000242
SN000008 SN000055 SN000098 SN000149 SN000196 SN000243
SN000009 SN000056 SN000099 SN000150 SN000197 SN000244
SN000010 SN000057 SN000104 SN000151 SN000198 SN000245
SN000011 SN000058 SN000105 SN000152 SN000199 SN000246
SN000012 SN000059 SN000106 SN000153 SN000200 SN000247
SN000013 SN000060 SN000107 SN000154 SN000201 SN000248
SN000014 SN000061 SN000108 SN000155 SN000202 SN000249
SN000015 SN000062 SN000109 SN000156 SN000203 SN000250
SN000016 SN000063 SN000110 SN000157 SN000204 SN000251
SN000017 SN000064 SN000111 SN000158 SN000205 SN000252
SN000018 SN000065 SN000112 SN000159 SN000206 SN000253
SN000019 SN000066 SN000113 SN000160 SN000207 SN000254
SN000020 SN000067 SN000114 SN000161 SN000208 SN000255
SN000021 SN000068 SN000115 SN000162 SN000209 SN000256
SN000022 SN000069 SN000116 SN000163 SN000210 SN000257
SN000023 SN000070 SN000117 SN000164 SN000211 SN000258
SN000024 SN000071 SN000118 SN000165 SN000212 SN000259
SN000025 SN000072 SN000119 SN000166 SN000213 SN000260
SN000026 SN000073 SN000120 SN000167 SN000214 SN000261
SN000027 SN000074 SN000121 SN000168 SN000215 SN000262
SN000028 SN000075 SN000122 SN000169 SN000216 SN000263
SN000029 SN000076 SN000123 SN000170 SN000217 SN000264
SN000030 SN000077 SN000124 SN000171 SN000218 SN000265
SN000031 SN000078 SN000125 SN000172 SN000219 SN000266
SN000032 SN000079 SN000126 SN000173 SN000220 SN000267
SN000033 SN000080 SN000127 SN000174 SN000221 SN000268
SN000034 SN000081 SN000128 SN000175 SN000222 SN000269
SN000035 SN000082 SN000129 SN000176 SN000223 SN000270
SN000036 SN000083 SN000130 SN000177 SN000224 SN000271
SN000037 SN000084 SN000131 SN000178 SN000225 SN000272
SN000038 SN000085 SN000132 SN000179 SN000226 SN000273
SN000039 SN000086 SN000133 SN000180 SN000227 SN000274
SN000040 SN000087 SN000134 SN000181 SN000228 SN000275
SN000041 SN000088 SN000135 SN000182 SN000229 SN000276
SN000042 SN000089 SN000136 SN000183 SN000230 SN000277
SN000043 SN000090 SN000137 SN000184 SN000231 SN000278
SN000044 SN000091 SN000138 SN000185 SN000232 SN000279
SN000045 SN000092 SN000139 SN000186 SN000233 SN000280
SN000046 SN000093 SN000140 SN000187 SN000234 SN000281
SN000047 SN000094 SN000141 SN000188 SN000235 SN000282
SN000283 SN000330 SN000387 SN000434 SN000481 SN000528
SN000284 SN000331 SN000388 SN000435 SN000482 SN000529
SN000285 SN000332 SN000389 SN000436 SN000483 SN000530
SN000286 SN000333 SN000390 SN000437 SN000484 SN000531
SN000287 SN000334 SN000391 SN000438 SN000485 SN000532
SN000288 SN000335 SN000392 SN000439 SN000486 SN000533
SN000289 SN000336 SN000393 SN000440 SN000487 SN000534
SN000290 SN000337 SN000394 SN000441 SN000488 SN000535
SN000291 SN000338 SN000395 SN000442 SN000489 SN000536
SN000292 SN000339 SN000396 SN000443 SN000490 SN000537
SN000293 SN000340 SN000397 SN000444 SN000491 SN000538
SN000294 SN000341 SN000398 SN000445 SN000492 SN000539
SN000295 SN000342 SN000399 SN000446 SN000493 SN000540
SN000296 SN000343 SN000400 SN000447 SN000494 SN000541
SN000297 SN000344 SN000401 SN000448 SN000495 SN000542
SN000298 SN000345 SN000402 SN000449 SN000496 SN000543
SN000299 SN000346 SN000403 SN000450 SN000497 SN000544
SN000300 SN000347 SN000404 SN000451 SN000498 SN000545
SN000301 SN000348 SN000405 SN000452 SN000499 SN000546
SN000302 SN000349 SN000406 SN000453 SN000500 SN000547
SN000303 SN000350 SN000407 SN000454 SN000501 SN000548
SN000304 SN000351 SN000408 SN000455 SN000502 SN000549
SN000305 SN000352 SN000409 SN000456 SN000503 SN000550
SN000306 SN000353 SN000410 SN000457 SN000504 SN000551
SN000307 SN000354 SN000411 SN000458 SN000505 SN000552
SN000308 SN000355 SN000412 SN000459 SN000506 SN000553
SN000309 SN000356 SN000413 SN000460 SN000507 SN000554
SN000310 SN000357 SN000414 SN000461 SN000508 SN000555
SN000311 SN000358 SN000415 SN000462 SN000509 SN000556
SN000312 SN000359 SN000416 SN000463 SN000510 SN000557
SN000313 SN000360 SN000417 SN000464 SN000511 SN000558
SN000314 SN000361 SN000418 SN000465 SN000512 SN000559
SN000315 SN000362 SN000419 SN000466 SN000513 SN000560
SN000316 SN000363 SN000420 SN000467 SN000514 SN000561
SN000317 SN000364 SN000421 SN000468 SN000515 SN000562
SN000318 SN000375 SN000422 SN000469 SN000516 SN000563
SN000319 SN000376 SN000423 SN000470 SN000517 SN000564
SN000320 SN000377 SN000424 SN000471 SN000518 SN000565
SN000321 SN000378 SN000425 SN000472 SN000519 SN000566
SN000322 SN000379 SN000426 SN000473 SN000520 SN000567
SN000323 SN000380 SN000427 SN000474 SN000521 SN000568
SN000324 SN000381 SN000428 SN000475 SN000522 SN000569
SN000325 SN000382 SN000429 SN000476 SN000523 SN000570
SN000326 SN000383 SN000430 SN000477 SN000524 SN000571
SN000327 SN000384 SN000431 SN000478 SN000525 SN000572
SN000328 SN000385 SN000432 SN000479 SN000526 SN000573
SN000329 SN000386 SN000433 SN000480 SN000527 SN000574
SN000575 SN000622 SN000669 SN000716 SN000763 SN000810
SN000576 SN000623 SN000670 SN000717 SN000764 SN000811
SN000577 SN000624 SN000671 SN000718 SN000765 SN000812
SN000578 SN000625 SN000672 SN000719 SN000766 SN000813
SN000579 SN000626 SN000673 SN000720 SN000767 SN000814
SN000580 SN000627 SN000674 SN000721 SN000768 SN000815
SN000581 SN000628 SN000675 SN000722 SN000769 SN000816
SN000582 SN000629 SN000676 SN000723 SN000770 SN000817
SN000583 SN000630 SN000677 SN000724 SN000771 SN000818
SN000584 SN000631 SN000678 SN000725 SN000772 SN000819
SN000585 SN000632 SN000679 SN000726 SN000773 SN000820
SN000586 SN000633 SN000680 SN000727 SN000774 SN000821
SN000587 SN000634 SN000681 SN000728 SN000775 SN000822
SN000588 SN000635 SN000682 SN000729 SN000776 SN000823
SN000589 SN000636 SN000683 SN000730 SN000777 SN000824
SN000590 SN000637 SN000684 SN000731 SN000778 SN000825
SN000591 SN000638 SN000685 SN000732 SN000779 SN000826
SN000592 SN000639 SN000686 SN000733 SN000780 SN000827
SN000593 SN000640 SN000687 SN000734 SN000781 SN000828
SN000594 SN000641 SN000688 SN000735 SN000782 SN000829
SN000595 SN000642 SN000689 SN000736 SN000783 SN000830
SN000596 SN000643 SN000690 SN000737 SN000784 SN000831
SN000597 SN000644 SN000691 SN000738 SN000785 SN000832
SN000598 SN000645 SN000692 SN000739 SN000786 SN000833
SN000599 SN000646 SN000693 SN000740 SN000787 SN000834
SN000600 SN000647 SN000694 SN000741 SN000788 SN000835
SN000601 SN000648 SN000695 SN000742 SN000789 SN000836
SN000602 SN000649 SN000696 SN000743 SN000790 SN000837
SN000603 SN000650 SN000697 SN000744 SN000791 SN000838
SN000604 SN000651 SN000698 SN000745 SN000792 SN000839
SN000605 SN000652 SN000699 SN000746 SN000793 SN000840
SN000606 SN000653 SN000700 SN000747 SN000794 SN000841
SN000607 SN000654 SN000701 SN000748 SN000795 SN000842
SN000608 SN000655 SN000702 SN000749 SN000796 SN000843
SN000609 SN000656 SN000703 SN000750 SN000797 SN000844
SN000610 SN000657 SN000704 SN000751 SN000798 SN000845
SN000611 SN000658 SN000705 SN000752 SN000799 SN000846
SN000612 SN000659 SN000706 SN000753 SN000800 SN000847
SN000613 SN000660 SN000707 SN000754 SN000801 SN000848
SN000614 SN000661 SN000708 SN000755 SN000802 SN000849
SN000615 SN000662 SN000709 SN000756 SN000803 SN000850
SN000616 SN000663 SN000710 SN000757 SN000804 SN000851
SN000617 SN000664 SN000711 SN000758 SN000805 SN000852
SN000618 SN000665 SN000712 SN000759 SN000806 SN000853
SN000619 SN000666 SN000713 SN000760 SN000807 SN000854
SN000620 SN000667 SN000714 SN000761 SN000808 SN000855
SN000621 SN000668 SN000715 SN000762 SN000809 SN000856
SN000857 SN000904 SN000951 SN000998 SN001045 SN001092
SN000858 SN000905 SN000952 SN000999 SN001046 SN001093
SN000859 SN000906 SN000953 SN001000 SN001047 SN001094
SN000860 SN000907 SN000954 SN001001 SN001048 SN001095
SN000861 SN000908 SN000955 SN001002 SN001049 SN001096
SN000862 SN000909 SN000956 SN001003 SN001050 SN001097
SN000863 SN000910 SN000957 SN001004 SN001051 SN001098
SN000864 SN000911 SN000958 SN001005 SN001052 SN001099
SN000865 SN000912 SN000959 SN001006 SN001053 SN001100
SN000866 SN000913 SN000960 SN001007 SN001054 SN001101
SN000867 SN000914 SN000961 SN001008 SN001055 SN001102
SN000868 SN000915 SN000962 SN001009 SN001056 SN001103
SN000869 SN000916 SN000963 SN001010 SN001057 SN001104
SN000870 SN000917 SN000964 SN001011 SN001058 SN001105
SN000871 SN000918 SN000965 SN001012 SN001059 SN001106
SN000872 SN000919 SN000966 SN001013 SN001060 SN001107
SN000873 SN000920 SN000967 SN001014 SN001061 SN001108
SN000874 SN000921 SN000968 SN001015 SN001062 SN001109
SN000875 SN000922 SN000969 SN001016 SN001063 SN001110
SN000876 SN000923 SN000970 SN001017 SN001064 SN001111
SN000877 SN000924 SN000971 SN001018 SN001065 SN001112
SN000878 SN000925 SN000972 SN001019 SN001066 SN001113
SN000879 SN000926 SN000973 SN001020 SN001067 SN001114
SN000880 SN000927 SN000974 SN001021 SN001068 SN001115
SN000881 SN000928 SN000975 SN001022 SN001069 SN001116
SN000882 SN000929 SN000976 SN001023 SN001070 SN001117
SN000883 SN000930 SN000977 SN001024 SN001071 SN001118
SN000884 SN000931 SN000978 SN001025 SN001072 SN001119
SN000885 SN000932 SN000979 SN001026 SN001073 SN001120
SN000886 SN000933 SN000980 SN001027 SN001074 SN001121
SN000887 SN000934 SN000981 SN001028 SN001075 SN001122
SN000888 SN000935 SN000982 SN001029 SN001076 SN001123
SN000889 SN000936 SN000983 SN001030 SN001077 SN001124
SN000890 SN000937 SN000984 SN001031 SN001078 SN001125
SN000891 SN000938 SN000985 SN001032 SN001079 SN001250
SN000892 SN000939 SN000986 SN001033 SN001080 SN001251
SN000893 SN000940 SN000987 SN001034 SN001081 SN001252
SN000894 SN000941 SN000988 SN001035 SN001082 SN001253
SN000895 SN000942 SN000989 SN001036 SN001083 SN001254
SN000896 SN000943 SN000990 SN001037 SN001084 SN001255
SN000897 SN000944 SN000991 SN001038 SN001085 SN001256
SN000898 SN000945 SN000992 SN001039 SN001086 SN001257
SN000899 SN000946 SN000993 SN001040 SN001087 SN001258
SN000900 SN000947 SN000994 SN001041 SN001088 SN001259
SN000901 SN000948 SN000995 SN001042 SN001089 SN001260
SN000902 SN000949 SN000996 SN001043 SN001090 SN001261
SN000903 SN000950 SN000997 SN001044 SN001091 SN001262
SN001263 SN001310 SN001357 SN001404
SN001264 SN001311 SN001358
SN001265 SN001312 SN001359
SN001266 SN001313 SN001360
SN001267 SN001314 SN001361
SN001268 SN001315 SN001362
SN001269 SN001316 SN001363
SN001270 SN001317 SN001364
SN001271 SN001318 SN001365
SN001272 SN001319 SN001366
SN001273 SN001320 SN001367
SN001274 SN001321 SN001368
SN001275 SN001322 SN001369
SN001276 SN001323 SN001370
SN001277 SN001324 SN001371
SN001278 SN001325 SN001372
SN001279 SN001326 SN001373
SN001280 SN001327 SN001374
SN001281 SN001328 SN001375
SN001282 SN001329 SN001376
SN001283 SN001330 SN001377
SN001284 SN001331 SN001378
SN001285 SN001332 SN001379
SN001286 SN001333 SN001380
SN001287 SN001334 SN001381
SN001288 SN001335 SN001382
SN001289 SN001336 SN001383
SN001290 SN001337 SN001384
SN001291 SN001338 SN001385
SN001292 SN001339 SN001386
SN001293 SN001340 SN001387
SN001294 SN001341 SN001388
SN001295 SN001342 SN001389
SN001296 SN001343 SN001390
SN001297 SN001344 SN001391
SN001298 SN001345 SN001392
SN001299 SN001346 SN001393
SN001300 SN001347 SN001394
SN001301 SN001348 SN001395
SN001302 SN001349 SN001396
SN001303 SN001350 SN001397
SN001304 SN001351 SN001398
SN001305 SN001352 SN001399
SN001306 SN001353 SN001400
SN001307 SN001354 SN001401
SN001308 SN001355 SN001402
SN001309 SN001356 SN001403
|
Recalling Firm/ Manufacturer |
Blue Belt Technologies, Inc 2875 Railroad St Pittsburgh PA 15201
|
| For Additional Information Contact | David Snyder 978-749-1440 |
Manufacturer Reason for Recall | CORI Surgical System with Real Intelligence Software exhibits an increased Marker Registration Error (MRE) which may cause tracker arrays to flicker on screen of the CORI Surgical System during use, and may cause surgical delay |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Smith & Nephew (Blue Belt wholly owned subsidiary of Smith & Nephew, Inc.) issued Urgent Medical Device Correction Notice C-2024-03 notification on May 7, 2024. Letter states reason for recall, health risk and action to take:
Required Actions for user/hospital:
1. Please see attached IFU Addendum for troubleshooting tips and notes to enhance the consistency of visibility of the markers and tracking arrays. This IFU Addendum can be found at https://ifu.smith-nephew.com/ by searching IFU number 1000264245.
2. Complete Correction Notice Acknowledgement Form and submit to FieldActions@smith-nephew.com.
3. A Smith & Nephew Robotics Representative will be deployed to your site to implement a new software version to correct the issue once it is available. |
| Quantity in Commerce | 916 units |
| Distribution | Nationwide
Foreign:
Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Korea,
Kuwait, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Puerto Rico, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Taiwan, Thailand, Turkey, UAE, UK |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = OLO
|
|
|
|