• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Medline

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medlinesee related information
Date Initiated by FirmApril 08, 2024
Date PostedSeptember 06, 2024
Recall Status1 Open3, Classified
Recall NumberZ-2993-2024
Recall Event ID 94583
510(K)NumberK803086 
Product Classification Electrode, pacemaker, temporary - Product Code LDF
ProductMedline Convenience kits labeled as: 1) CATH LAB SPECIAL PROCEDURE, Pack Number 00-RDF036O; 2) CATH LAB SPECIAL PROCEDURE, Pack Number 00-RDF036OH; 3) CATH ACCESSORY PACK, Pack Number 0M-CP31433J; 4) 3" RADIAL ARTERIAL LINE KIT- OR, Pack Number ART1210; 5) PERF. ASSOC. CUT DOWN TRAY, Pack Number BM17BAD; 6) CATHETER ADD A CATH TRAY, Pack Number CVI4970; 7) BIOPSY TRAY, Pack Number DYNDH1129; 8) ARTHROGRAM TRAY, Pack Number DYNDH1134; 9) BRISTOL MYELOGRAM, Pack Number DYNDH1228; 10) LUMBAR TRAY, Pack Number DYNDH1540A; 11) RADIOLOGY BASIC PACK, Pack Number DYNDH1827; 12) BLOOD CULTURE KIT W/BOTTLES, Pack Number DYNDH1932; 13) BIOPSY PACK, Pack Number DYNDL1462B; 14) CATH LAB PACK-LF, Pack Number DYNJ0105963Q; 15) CARDIAC CATH HARPER PACK-LF, Pack Number DYNJ0161912G; 16) CATH LAB PACK-LF, Pack Number DYNJ0173400P; 17) CATH LAB PACK-LF, Pack Number DYNJ0194968AG; 18) PICC LINE PACK-LF, Pack Number DYNJ0275614F; 19) PEDI CATH PACK-LF, Pack Number DYNJ0429201L; 20) SPECIAL PROCEDURE ANGIO PACK, Pack Number DYNJ04386U; 21) RADIOLOGY PACK, Pack Number DYNJ04562S; 22) IR PACK, Pack Number DYNJ04674U; 23) IR PACK, Pack Number DYNJ04674UH; 24) LEFT HEART PACK-LF, Pack Number DYNJ0490280PH; 25) LT HEART/ANGIOGRAPHY PACK-LF, Pack Number DYNJ0537652AG; 26) SM CARDIAC CATH PACK-LF , Pack Number DYNJ0621195TH; 27) CATH LAB PACK-LF, Pack Number DYNJ0755738R; 28) CARDIAC CATH PACK-LF, Pack Number DYNJ0781476AAH ; 29) CARDIAC CATH PACK-LF, Pack Number DYNJ0781918NH; 30) CATH LAB PACK , Pack Number DYNJ07919GH; 31) PACEMAKER PACK-LF , Pack Number DYNJ0830939ACH ; 32) CARDIAC CATH PACK-LF, Pack Number DYNJ0843627G ; 33) PICC INSERTION PACK-LF, Pack Number DYNJ0880659D; 34) PEDS HEART CATH PACK-LF, Pack Number DYNJ0897998D; 35) RADIAL ARTERY CATH PACK-LF, Pack Number DYNJ0945414BH; 36) RADIAL ARTERY CATH PACK-LF, Pack Number DYNJ0945414CH; 37) CATH LAB PACK , Pack Number DYNJ10117JH; 38) CARDIAC CATH PACK, Pack Number DYNJ17081S; 39) DEPAUL ANGIO PROCEDURE PACK, Pack Number DYNJ19937I; 40) PV/PTA TRAY-RF, Pack Number DYNJ21944VH; 41) ANGIO TRAY, Pack Number DYNJ23438M; 42) ANGIO TRAY, Pack Number DYNJ23438N; 43) ANGIO TRAY, Pack Number DYNJ23438O; 44) ADULT CARDIAC CATH PACK, Pack Number DYNJ23455Q; 45) VIR PACK, Pack Number DYNJ23499MH; 46) HEART CATH-SMH, Pack Number DYNJ24483FH; 47) CATH LAB PACK , Pack Number DYNJ24802PH; 48) ANGIO DRAPE PACK, Pack Number DYNJ24806KH; 49) RADIOLOGY STD TRAY, Pack Number DYNJ25029RH; 50) CARDIAC CATH PACK, Pack Number DYNJ26305N; 51) DIAGNOSTIC ADULT-LF , Pack Number DYNJ27239DH; 52) HEART CATH LT PACK, Pack Number DYNJ31282I; 53) ANGIO PACK, Pack Number DYNJ31339I; 54) CARDIAC CATH PACK, Pack Number DYNJ31847I; 55) NEURO ANGIO TRAY-LF, Pack Number DYNJ34882M; 56) HEART CATH II PACK CHS-LF, Pack Number DYNJ35605M; 57) ISSAQUA CATH/IR PACK, Pack Number DYNJ36738G; 58) PEDS CARDIAC PACK, Pack Number DYNJ37418MH; 59) SPECIAL PROCEDURE PACK-LF, Pack Number DYNJ38050F; 60) SPECIAL PROCEDURE PACK-LF, Pack Number DYNJ38050FH; 61) SPECIAL PROCEDURE PACK, Pack Number DYNJ38314G; 62) MINOR SPECIAL PROCEDURE PACK, Pack Number DYNJ38326G; 63) CARDIAC CATH PACK, Pack Number DYNJ38866F; 64) PERM CATH TRAY, Pack Number DYNJ40423AH; 65) O R EMERGENCY PACK, Pack Number DYNJ40460CH; 66) CARDIAC CATH PACK, Pack Number DYNJ40511M; 67) CATH LAB PACK, Pack Number DYNJ40961A; 68) CARDIAC CATH TRAY, Pack Number DYNJ41050C; 69) CATH LAB PACK, Pack Number DYNJ43469D; 70) PK, RADIOLOGY, Pack Number DYNJ43816B; 71) CARDIAC CATH TRAY, Pack Number DYNJ45323F; 72) LEFT HEART, Pack Number DYNJ45984I; 73) CATH LAB PROCEDURE, Pack Number DYNJ46353F; 74) ANGIO PACK, Pack Number DYNJ46401B; 75) CARDIAC CATH TRAY, Pack Number DYNJ46402A; 76) CARDIAC CATH TRAY, Pack Number DYNJ46419C; 77) NEURO CATH PACK (NCLUA)227-LF, Pack Number DYNJ47646D; 78) NEURO CATH PACK (NCLUA)227-LF, Pack Number DYNJ47646F; 79) NEURO CATH PACK (NCLUA)227-LF, Pack Number DYNJ47646G; 80) NEURO
Code Information 00-RDF036O, Lot Number 24BMG670; 00-RDF036OH, Lot Number 24BMG670; 00-RDF036OH, Lot Number 22FMH197 ; 00-RDF036OH, Lot Number 22IMC476 ; 00-RDF036OH, Lot Number 22JMD690 ; 00-RDF036OH, Lot Number 22KMB816 ; 00-RDF036OH, Lot Number 22LMB962 ; 00-RDF036OH, Lot Number 23BMA905 ; 00-RDF036OH, Lot Number 23BMH681 ; 00-RDF036OH, Lot Number 23CMG412 ; 00-RDF036OH, Lot Number 23DMI750 ; 00-RDF036OH, Lot Number 23IMD552 ; 00-RDF036OH, Lot Number 23KMD849 ; 0M-CP31433J, Lot Number 24DBD373; ART1210, Lot Number 2023121890; BM17BAD, Lot Number 22FBA406; BM17BAD, Lot Number 22EBJ428; CVI4970, Lot Number 2023121880; DYNDH1129, Lot Number 2019012450; DYNDH1134, Lot Number 2019012450; DYNDH1228, Lot Number 23JBW148; DYNDH1228, Lot Number 23HBW751; DYNDH1228, Lot Number 23FBU652; DYNDH1228, Lot Number 23DBK822; DYNDH1228, Lot Number 23JBW148; DYNDH1228, Lot Number 23HBW751; DYNDH1228, Lot Number 23FBU652; DYNDH1228, Lot Number 23DBK822; DYNDH1540A, Lot Number 2023091990; DYNDH1827, Lot Number 24BMB809; DYNDH1932, Lot Number 2023121180; DYNDL1462B, Lot Number 2019012450; DYNJ0105963Q, Lot Number 22GMH630; DYNJ0105963Q, Lot Number 22EMG581; DYNJ0105963Q, Lot Number 22CMI727; DYNJ0105963Q, Lot Number 22CMA702; DYNJ0161912G, Lot Number 23KDB964; DYNJ0173400P, Lot Number 21FMC951; DYNJ0173400P, Lot Number 21DME907; DYNJ0173400P, Lot Number 21CMB877; DYNJ0173400P, Lot Number 21AMB468; DYNJ0173400P, Lot Number 20LME686; DYNJ0173400P, Lot Number 20LMC387; DYNJ0173400P, Lot Number 20JMC557; DYNJ0173400P, Lot Number 20IMB398; DYNJ0173400P, Lot Number 20GMC663; DYNJ0173400P, Lot Number 20EMA806; DYNJ0173400P, Lot Number 20EMA459; DYNJ0173400P, Lot Number 20BBR274; DYNJ0194968AG, Lot Number 24CMC638; DYNJ0194968AG, Lot Number 24BME100; DYNJ0275614F, Lot Number 24CMD821; DYNJ0275614F, Lot Number 24AMJ414; DYNJ0429201L, Lot Number 24BMI261; DYNJ04386U, Lot Number 24CMF538; DYNJ04562S, Lot Number 23BLA247; DYNJ04562S, Lot Number 23ALA434; DYNJ04562S, Lot Number 23ALA227; DYNJ04562S, Lot Number 22KLA758; DYNJ04562S, Lot Number 22KLA141; DYNJ04562S, Lot Number 22JLA659; DYNJ04562S, Lot Number 22ILA617; DYNJ04562S, Lot Number 22GLA997; DYNJ04674U, Lot Number 24BMJ236; DYNJ04674U, Lot Number 24BMB067; DYNJ04674UH, Lot Number 24BMJ236; DYNJ04674UH, Lot Number 24BMB067; DYNJ04674UH, Lot Number 23DMF209 ; DYNJ04674UH, Lot Number 23EMJ411 ; DYNJ04674UH, Lot Number 23FMD591 ; DYNJ04674UH, Lot Number 24AMB874 ; DYNJ0490280PH, Lot Number 21HMD962 ; DYNJ0490280PH, Lot Number 21JMH929 ; DYNJ0490280PH, Lot Number 19AKC431 ; DYNJ0490280PH, Lot Number 21KME862 ; DYNJ0490280PH, Lot Number 19BKB887 ; DYNJ0490280PH, Lot Number 19DKA387 ; DYNJ0490280PH, Lot Number 22AMB750 ; DYNJ0490280PH, Lot Number 22BMB089 ; DYNJ0490280PH, Lot Number 22CMF731 ; DYNJ0490280PH, Lot Number 22DME928 ; DYNJ0490280PH, Lot Number 22FMG329 ; DYNJ0490280PH, Lot Number 22HMG030 ; DYNJ0490280PH, Lot Number 22JMC980 ; DYNJ0490280PH, Lot Number 22LME104 ; DYNJ0490280PH, Lot Number 23AMB149 ; DYNJ0490280PH, Lot Number 23BMH102 ; DYNJ0490280PH, Lot Number 23EMG454 ; DYNJ0490280PH, Lot Number 23FMF799 ; DYNJ0490280PH, Lot Number 23GMD123 ; DYNJ0490280PH, Lot Number 23HMI667 ; DYNJ0490280PH, Lot Number 23JMA312 ; DYNJ0490280PH, Lot Number 23JMJ462 ; DYNJ0490280PH, Lot Number 24AMC475 ; DYNJ0537652AG, Lot Number 24BMB728; DYNJ0621195TH, Lot Number 21CMC148 ; DYNJ0621195TH, Lot Number 21DMA294 ; DYNJ0621195TH, Lot Number 21FMA079 ; DYNJ0621195TH, Lot Number 21GMD642 ; DYNJ0621195TH, Lot Number 19AKA773 ; DYNJ0621195TH, Lot Number 19CKA323 ; DYNJ0621195TH, Lot Number 21LMC489 ; DYNJ0621195TH, Lot Number 19DKC783 ; DYNJ0621195TH, Lot Number 21LMG876 ; DYNJ0621195TH, Lot Number 19GKB265 ; DYNJ0621195TH, Lot Number 19HKB762 ; DYNJ0621195TH, Lot Number 19IKC226 ; DYNJ0621195TH, Lot Number 19KKC114 ; DYNJ0621195TH, Lot Number 22CMB662 ; DYNJ0621195TH, Lot Number 22DME047 ; DYNJ0621195TH, Lot Number 22EMG769 ; DYNJ0621195TH, Lot Number 22EMG770 ; DYNJ0621195TH, Lot Number 22GMD234 ; DYNJ0621195TH, Lot Number 22HMF623 ; DYNJ0621195TH, Lot Number 22HMF626 ; DYNJ0621195TH, Lot Number 22HMF805 ; DYNJ0621195TH, Lot Number 22KMH017 ; DYNJ0621195TH, Lot Number 22LMH919 ; DYNJ0621195TH, Lot Number 23BMH224 ; DYNJ0621195TH, Lot Number 23DMC659 ; DYNJ0621195TH, Lot Number 23EMH238 ; DYNJ0621195TH, Lot Number 23FMF954 ; DYNJ0621195TH, Lot Number 23GME622 ; DYNJ0621195TH, Lot Number 23HMD011 ; DYNJ0621195TH, Lot Number 23IMH238 ; DYNJ0621195TH, Lot Number 23KME312 ; DYNJ0621195TH, Lot Number 23KMF280 ; DYNJ0621195TH, Lot Number 23KMF847 ; DYNJ0621195TH, Lot Number 23KMG596 ; DYNJ0621195TH, Lot Number 23LMH452 ; DYNJ0755738R, Lot Number 24BMC278; DYNJ0781476AAH , Lot Number 22EME967 ; DYNJ0781476AAH , Lot Number 22FMG782 ; DYNJ0781476AAH , Lot Number 22FMG783 ; DYNJ0781476AAH , Lot Number 22HMH711 ; DYNJ0781476AAH , Lot Number 23GMC354 ; DYNJ0781476AAH , Lot Number 23JMF525 ; DYNJ0781476AAH , Lot Number 23KMA657 ; DYNJ0781476AAH , Lot Number 23KME613 ; DYNJ0781476AAH , Lot Number 23KMJ481 ; DYNJ0781476AAH , Lot Number 24AMA131 ; DYNJ0781476AAH , Lot Number 24BMF082 ; DYNJ0781476AAH , Lot Number 24CMA858 ; DYNJ0781918NH, Lot Number 19LMA158 ; DYNJ07919GH, Lot Number 21GBB755 ; DYNJ07919GH, Lot Number 21HBU581 ; DYNJ07919GH, Lot Number 21JBZ187 ; DYNJ07919GH, Lot Number 21WBA117 ; DYNJ07919GH, Lot Number 21WBA928 ; DYNJ07919GH, Lot Number 21LBW658 ; DYNJ07919GH, Lot Number 22ABG228 ; DYNJ07919GH, Lot Number 22ABW646 ; DYNJ07919GH, Lot Number 19ABD174 ; DYNJ07919GH, Lot Number 19BBI258 ; DYNJ07919GH, Lot Number 19CBD215 ; DYNJ07919GH, Lot Number 22OBJ080 ; DYNJ07919GH, Lot Number 22DBN531 ; DYNJ07919GH, Lot Number 19DBK522 ; DYNJ07919GH, Lot Number 22EBN350 ; DYNJ07919GH, Lot Number 19EBA372 ; DYNJ07919GH, Lot Number 22GBJ853 ; DYNJ07919GH, Lot Number 19FBK929 ; DYNJ07919GH, Lot Number 19GBW574 ; DYNJ07919GH, Lot Number 22IBK744 ; DYNJ07919GH, Lot Number 19HBE646 ; DYNJ07919GH, Lot Number 19HBM345 ; DYNJ07919GH, Lot Number 22JBD628 ; DYNJ07919GH, Lot Number 22KBF159 ; DYNJ07919GH, Lot Number 19JBI461 ; DYNJ07919GH, Lot Number 23ABK220 ; DYNJ07919GH, Lot Number 19KBU453 ; DYNJ07919GH, Lot Number 23CBM139 ; DYNJ07919GH, Lot Number 23CBV596 ; DYNJ07919GH, Lot Number 23DBO846 ; DYNJ07919GH, Lot Number 23EBS913 ; DYNJ07919GH, Lot Number 23FBJ315 ; DYNJ07919GH, Lot Number 23HBJ763 ; DYNJ07919GH, Lot Number 23IBG534 ; DYNJ07919GH, Lot Number 23JBP625 ; DYNJ07919GH, Lot Number 23KBV909 ; DYNJ07919GH, Lot Number 24ABA869 ; DYNJ07919GH, Lot Number 21ABT102 ; DYNJ07919GH, Lot Number 21BBF501 ; DYNJ0830939ACH , Lot Number 23KME948 ; DYNJ0830939ACH , Lot Number 23LMH653 ; DYNJ0843627G , Lot Number 19JDA113 ; DYNJ0843627G , Lot Number 19LDA741 ; DYNJ0843627G , Lot Number 20ADD013 ; DYNJ0843627G , Lot Number 20CDB405 ; DYNJ0843627G , Lot Number 20EDA304 ; DYNJ0843627G , Lot Number 20FDB013 ; DYNJ0843627G , Lot Number 20HDA724 ; DYNJ0843627G , Lot Number 20KDC483 ; DYNJ0843627G , Lot Number 21ADB521 ; DYNJ0843627G , Lot Number 21BDB149 ; DYNJ0843627G , Lot Number 21DDA579 ; DYNJ0843627G , Lot Number 21EDB315 ; DYNJ0880659D, Lot Number 21JBZ445; DYNJ0880659D, Lot Number 21JBZ448; DYNJ0880659D, Lot Number 21IBG203; DYNJ0880659D, Lot Number 21GBA185; DYNJ0880659D, Lot Number 21CKA585; DYNJ0880659D, Lot Number 20KKA625; DYNJ0880659D, Lot Number 20FKA073; DYNJ0880659D, Lot Number 20EKA440; DYNJ0880659D, Lot Number 20BKB899; DYNJ0880659D, Lot Number 20BKA142; DYNJ0880659D, Lot Number 20AKA241; DYNJ0897998D, Lot Number 24BMD152; DYNJ0945414BH, Lot Number 21GLA816 ; DYNJ0945414BH, Lot Number 21ILA109 ; DYNJ0945414BH, Lot Number 21JLA361 ; DYNJ0945414BH, Lot Number 21KLB208 ; DYNJ0945414BH, Lot Number 21LLA833 ; DYNJ0945414BH, Lot Number 22BLA140 ; DYNJ0945414BH, Lot Number 22CLA144 ; DYNJ0945414BH, Lot Number 22DLA977 ; DYNJ0945414BH, Lot Number 22FLB042 ; DYNJ0945414BH, Lot Number 22HLA417 ; DYNJ0945414BH, Lot Number 22ILA909 ; DYNJ0945414BH, Lot Number 22JLA505 ; DYNJ0945414BH, Lot Number 23ALA510 ; DYNJ0945414BH, Lot Number 23BLA048 ; DYNJ0945414BH, Lot Number 23CLA942 ; DYNJ0945414BH, Lot Number 23ELA395 ; DYNJ0945414BH, Lot Number 23FLA777 ; DYNJ0945414BH, Lot Number 23GLA541 ; DYNJ0945414BH, Lot Number 23GLA853 ; DYNJ0945414BH, Lot Number 23HLA190 ; DYNJ0945414BH, Lot Number 23JLA036 ; DYNJ0945414BH, Lot Number 19BKA780 ; DYNJ0945414BH, Lot Number 19CKC609 ; DYNJ0945414BH, Lot Number 19EKB333 ; DYNJ0945414BH, Lot Number 19FKC401 ; DYNJ0945414BH, Lot Number 19GKC996 ; DYNJ0945414BH, Lot Number 19JKA507 ; DYNJ0945414BH, Lot Number 19LKA717 ; DYNJ0945414BH, Lot Number 21CKA003 ; DYNJ0945414BH, Lot Number 21CKA573 ; DYNJ0945414CH, Lot Number 24ALB040 ; DYNJ10117JH, Lot Number 22IBK960 ; DYNJ10117JH, Lot Number 22KBD253 ; DYNJ10117JH, Lot Number 22KBQ971 ; DYNJ10117JH, Lot Number 23CBI713 ; DYNJ10117JH, Lot Number 23CBJ471 ; DYNJ10117JH, Lot Number 23EBM713 ; DYNJ10117JH, Lot Number 23GBM013 ; DYNJ10117JH, Lot Number 23JBB557 ; DYNJ10117JH, Lot Number 23JBV608 ; DYNJ10117JH, Lot Number 23KBL195 ; DYNJ10117JH, Lot Number 23LBF433 ; DYNJ17081S, Lot Number 24BMI797; DYNJ19937I, Lot Number 24BMJ856; DYNJ19937I, Lot Number 24BMJ856; DYNJ21944VH, Lot Number 22DBH189 ; DYNJ21944VH, Lot Number 22DBL929 ; DYNJ21944VH, Lot Number 22FBK986 ; DYNJ21944VH, Lot Number 22IBB299 ; DYNJ21944VH, Lot Number 22IBU953 ; DYNJ21944VH, Lot Number 22LBP690 ; DYNJ21944VH, Lot Number 23BBS804 ; DYNJ21944VH, Lot Number 23CBA356 ; DYNJ21944VH, Lot Number 23EBD119 ; DYNJ21944VH, Lot Number 23FBN290 ; DYNJ21944VH, Lot Number 23GBO884 ; DYNJ21944VH, Lot Number 23HBW324 ; DYNJ21944VH, Lot Number 23IBU028 ; DYNJ21944VH, Lot Number 23JBV665 ; DYNJ21944VH, Lot Number 23LBD751 ; DYNJ23438M, Lot Number 23EME932; DYNJ23438M, Lot Number 23CMA632; DYNJ23438M, Lot Number 23AME649; DYNJ23438M, Lot Number 22KME553; DYNJ23438M, Lot Number 22JME730; DYNJ23438M, Lot Number 22GMC968; DYNJ23438M, Lot Number 22FMG446; DYNJ23438M, Lot Number 22EMH355; DYNJ23438M, Lot Number 22DMI674; DYNJ23438M, Lot Number 22CMC423; DYNJ23438N, Lot Number 23KMG282; DYNJ23438N, Lot Number 23JMF036; DYNJ23438N, Lot Number 23HMG428; DYNJ23438N, Lot Number 23GME450; DYNJ23438O, Lot Number 24BMG597; DYNJ23438O, Lot Number 23LME107; DYNJ23455Q, Lot Number 24BMH318; DYNJ23455Q, Lot Number 24BMG216; DYNJ23499MH, Lot Number 22CLA800 ; DYNJ23499MH, Lot Number 22CLB201 ; DYNJ23499MH, Lot Number 22DLA291 ; DYNJ23499MH, Lot Number 22FLA104 ; DYNJ23499MH, Lot Number 22FLA990 ; DYNJ23499MH, Lot Number 23ABJ995 ; DYNJ23499MH, Lot Number 23CBR249 ; DYNJ23499MH, Lot Number 23DBR035 ; DYNJ23499MH, Lot Number 23EBO509 ; DYNJ23499MH, Lot Number 23FBA509 ; DYNJ23499MH, Lot Number 23GBT721 ; DYNJ23499MH, Lot Number 23IBB622 ; DYNJ23499MH, Lot Number 23IBC646 ; DYNJ23499MH, Lot Number 23LBC976 ; DYNJ24483FH, Lot Number 21GBK154 ; DYNJ24483FH, Lot Number 21JBA911 ; DYNJ24483FH, Lot Number 21KBK768 ; DYNJ24483FH, Lot Number 21KBX225 ; DYNJ24483FH, Lot Number 21LBQ325 ; DYNJ24483FH, Lot Number 22BBG129 ; DYNJ24483FH, Lot Number 22EBE392 ; DYNJ24483FH, Lot Number 22FBC111 ; DYNJ24483FH, Lot Number 22HBM207 ; DYNJ24483FH, Lot Number 22HBN243 ; DYNJ24483FH, Lot Number 22IBO703 ; DYNJ24483FH, Lot Number 22JBU403 ; DYNJ24483FH, Lot Number 23ABJ189 ; DYNJ24483FH, Lot Number 23BBK035 ; DYNJ24483FH, Lot Number 23CBF253 ; DYNJ24483FH, Lot Number 23EBD541 ; DYNJ24483FH, Lot Number 23EBR279 ; DYNJ24483FH, Lot Number 23GBT069 ; DYNJ24483FH, Lot Number 23HBR771 ; DYNJ24483FH, Lot Number 23JBH882 ; DYNJ24483FH, Lot Number 23KBQ076 ; DYNJ24483FH, Lot Number 24ABT625 ; DYNJ24483FH, Lot Number 21EBC084 ; DYNJ24802PH, Lot Number 22JMG933 ; DYNJ24802PH, Lot Number 23AMJ220 ; DYNJ24802PH, Lot Number 23BME183 ; DYNJ24802PH, Lot Number 23FMA595 ; DYNJ24802PH, Lot Number 23IBC272 ; DYNJ24802PH, Lot Number 23JBM891 ; DYNJ24806KH, Lot Number 23EMH255 ; DYNJ24806KH, Lot Number 23GMF832 ; DYNJ24806KH, Lot Number 23HMB542 ; DYNJ24806KH, Lot Number 23IBL985 ; DYNJ24806KH, Lot Number 23JBC114 ; DYNJ24806KH, Lot Number 23KBI301 ; DYNJ24806KH, Lot Number 23LBB175 ; DYNJ25029RH, Lot Number 22CLA656 ; DYNJ25029RH, Lot Number 22DLA394 ; DYNJ25029RH, Lot Number 22FLA086 ; DYNJ25029RH, Lot Number 22GLA148 ; DYNJ25029RH, Lot Number 22GLA925 ; DYNJ25029RH, Lot Number 22JLA545 ; DYNJ25029RH, Lot Number 22KLA120 ; DYNJ25029RH, Lot Number 22KLA762 ; DYNJ25029RH, Lot Number 23ALA105 ; DYNJ25029RH, Lot Number 23BLA677 ; DYNJ25029RH, Lot Number 23CLA727 ; DYNJ25029RH, Lot Number 23DLA122 ; DYNJ25029RH, Lot Number 23DLA547 ; DYNJ25029RH, Lot Number 23ELB084 ; DYNJ25029RH, Lot Number 23FLA410 ; DYNJ25029RH, Lot Number 23HLA004 ; DYNJ25029RH, Lot Number 23ILA065 ; DYNJ25029RH, Lot Number 23JLA907 ; DYNJ25029RH, Lot Number 23LLA204 ; DYNJ25029RH, Lot Number 23LLA811 ; DYNJ26305N, Lot Number 22FMH026; DYNJ26305N, Lot Number 22DMF922; DYNJ26305N, Lot Number 22CMC069; DYNJ27239DH, Lot Number 21FBR668 ; DYNJ27239DH, Lot Number 21HBE872 ; DYNJ27239DH, Lot Number 21IBJ808 ; DYNJ27239DH, Lot Number 21KBK878 ; DYNJ27239DH, Lot Number 21LBW583 ; DYNJ27239DH, Lot Number 22ABS567 ; DYNJ27239DH, Lot Number 19ABD515 ; DYNJ27239DH, Lot Number 22NBA011 ; DYNJ27239DH, Lot Number 19ABW321 ; DYNJ27239DH, Lot Number 19CBB958 ; DYNJ27239DH, Lot Number 22OBH550 ; DYNJ27239DH, Lot Number 19CBK342 ; DYNJ27239DH, Lot Number 22EBT237 ; DYNJ27239DH, Lot Number 19EBD627 ; DYNJ27239DH, Lot Number 22GBG608 ; DYNJ27239DH, Lot Number 19FBS991 ; DYNJ27239DH, Lot Number 22GBS671 ; DYNJ27239DH, Lot Number 22IBJ991 ; DYNJ27239DH, Lot Number 19HBG466 ; DYNJ27239DH, Lot Number 19TBE261 ; DYNJ27239DH, Lot Number 22JBW876 ; DYNJ27239DH, Lot Number 19JBM769 ; DYNJ27239DH, Lot Number 23BBE043 ; DYNJ27239DH, Lot Number 23BBE354 ; DYNJ27239DH, Lot Number 19WBA110 ; DYNJ27239DH, Lot Number 23DBL781 ; DYNJ27239DH, Lot Number 23FBD165 ; DYNJ27239DH, Lot Number 23GBV377 ; DYNJ27239DH, Lot Number 23IBJ246 ; DYNJ27239DH, Lot Number 23JBT634 ; DYNJ27239DH, Lot Number 23LBD753 ; DYNJ27239DH, Lot Number 23LBS106 ; DYNJ27239DH, Lot Number 21DBR890 ; DYNJ31282I, Lot Number 22IMA443; DYNJ31282I, Lot Number 22FMH651; DYNJ31282I, Lot Number 22BMH568; DYNJ31339I, Lot Number 22HMF102; DYNJ31339I, Lot Number 22HMF103; DYNJ31339I, Lot Number 22FMH653; DYNJ31339I, Lot Number 22CMA673; DYNJ31847I, Lot Number 24BLA978; DYNJ34882M, Lot Number 23LBH460; DYNJ34882M, Lot Number 23JBI855; DYNJ34882M, Lot Number 23IBB900; DYNJ34882M, Lot Number 23GBS729; DYNJ34882M, Lot Number 23EBR807; DYNJ34882M, Lot Number 23EBB249; DYNJ34882M, Lot Number 23CBS242; DYNJ34882M, Lot Number 23ABS130; DYNJ34882M, Lot Number 22KBK755; DYNJ34882M, Lot Number 22DBT284; DYNJ34882M, Lot Number 22DBU024; DYNJ35605M, Lot Number 22IBE481; DYNJ36738G, Lot Number 22IMB298; DYNJ36738G, Lot Number 22FMH659; DYNJ36738G, Lot Number 22FMF231; DYNJ36738G, Lot Number 22EME957; DYNJ36738G, Lot Number 22CMD714; DYNJ37418MH, Lot Number 23JLB292 ; DYNJ37418MH, Lot Number 23LLA270 ; DYNJ37418MH, Lot Number 24ALA699 ; DYNJ38050F, Lot Number 24BMI233; DYNJ38050FH, Lot Number 24BMI233; DYNJ38050FH, Lot Number 21KMD092 ; DYNJ38050FH, Lot Number 21KMG565 ; DYNJ38050FH, Lot Number 22AMB990 ; DYNJ38050FH, Lot Number 22BMB319 ; DYNJ38050FH, Lot Number 22GME990 ; DYNJ38050FH, Lot Number 22GMG098 ; DYNJ38050FH, Lot Number 22KMC784 ; DYNJ38050FH, Lot Number 22LMD184 ; DYNJ38050FH, Lot Number 22LMD185 ; DYNJ38050FH, Lot Number 23AMF469 ; DYNJ38050FH, Lot Number 23AMJ204 ; DYNJ38050FH, Lot Number 23DMD884 ; DYNJ38050FH, Lot Number 23DMI461 ; DYNJ38050FH, Lot Number 23GMA758 ; DYNJ38050FH, Lot Number 23GMG025 ; DYNJ38050FH, Lot Number 23HMD508 ; DYNJ38050FH, Lot Number 23IMD354 ; DYNJ38050FH, Lot Number 23IME633 ; DYNJ38050FH, Lot Number 23JMF033 ; DYNJ38050FH, Lot Number 23JMF216 ; DYNJ38050FH, Lot Number 23LMB748 ; DYNJ38314G, Lot Number 24BMI892; DYNJ38314G, Lot Number 24BMD073; DYNJ38326G, Lot Number 24CME051; DYNJ38866F, Lot Number 24BMB893; DYNJ40423AH, Lot Number 21IBE188 ; DYNJ40423AH, Lot Number 21JBO288 ; DYNJ40423AH, Lot Number 21KBJ295 ; DYNJ40423AH, Lot Number 21LBG651 ; DYNJ40423AH, Lot Number 21LBX784 ; DYNJ40423AH, Lot Number 19BBC889 ; DYNJ40423AH, Lot Number 22OBI319 ; DYNJ40423AH, Lot Number 19CBM373 ; DYNJ40423AH, Lot Number 22EBK738 ; DYNJ40423AH, Lot Number 19PBA754 ; DYNJ40423AH, Lot Number 22FBJ109 ; DYNJ40423AH, Lot Number 19QBB003 ; DYNJ40423AH, Lot Number 22GBE426 ; DYNJ40423AH, Lot Number 19FBS884 ; DYNJ40423AH, Lot Number 22HBG322 ; DYNJ40423AH, Lot Number 19GBX387 ; DYNJ40423AH, Lot Number 22JBA875 ; DYNJ40423AH, Lot Number 19TBF451 ; DYNJ40423AH, Lot Number 19IBF089 ; DYNJ40423AH, Lot Number 22KBJ554 ; DYNJ40423AH, Lot Number 22KBV649 ; DYNJ40423AH, Lot Number 23ABI890 ; DYNJ40423AH, Lot Number 19KBE266 ; DYNJ40423AH, Lot Number 19KBJ482 ; DYNJ40423AH, Lot Number 19LBG354 ; DYNJ40423AH, Lot Number 23CBW953 ; DYNJ40423AH, Lot Number 23DBQ719 ; DYNJ40423AH, Lot Number 23EBK984 ; DYNJ40423AH, Lot Number 23GBJ696 ; DYNJ40423AH, Lot Number 23HBJ497 ; DYNJ40423AH, Lot Number 23JBB911 ; DYNJ40423AH, Lot Number 23JBC229 ; DYNJ40423AH, Lot Number 23LBN666 ; DYNJ40423AH, Lot Number 21ABP453 ; DYNJ40423AH, Lot Number 21ABR881 ; DYNJ40423AH, Lot Number 21DBA932 ; DYNJ40423AH, Lot Number 21DBQ047 ; DYNJ40460CH, Lot Number 21IBP649 ; DYNJ40460CH, Lot Number 21KBC190 ; DYNJ40460CH, Lot Number 21LBU526 ; DYNJ40460CH, Lot Number 19ABC249 ; DYNJ40460CH, Lot Number 22BBY181 ; DYNJ40460CH, Lot Number 19ABO821 ; DYNJ40460CH, Lot Number 19CBD212 ; DYNJ40460CH, Lot Number 22DBL145 ; DYNJ40460CH, Lot Number 19DBG997 ; DYNJ40460CH, Lot Number 22EBB393 ; DYNJ40460CH, Lot Number 19EBG847 ; DYNJ40460CH, Lot Number 22GBN835 ; DYNJ40460CH, Lot Number 19GBC026 ; DYNJ40460CH, Lot Number 19HBC448 ; DYNJ40460CH, Lot Number 22JBJ134 ; DYNJ40460CH, Lot Number 19IBP623 ; DYNJ40460CH, Lot Number 22KBR788 ; DYNJ40460CH, Lot Number 23ABJ762 ; DYNJ40460CH, Lot Number 19LBG348 ; DYNJ40460CH, Lot Number 23DBP773 ; DYNJ40460CH, Lot Number 23FBC379 ; DYNJ40460CH, Lot Number 23GBC555 ; DYNJ40460CH, Lot Number 23IBE135 ; DYNJ40460CH, Lot Number 23JBC231 ; DYNJ40460CH, Lot Number 23KBJ950 ; DYNJ40460CH, Lot Number 23LBD302 ; DYNJ40511M, Lot Number 24CMB343; DYNJ40511M, Lot Number 24BMB505; DYNJ40961A, Lot Number 24BMC065; DYNJ40961A, Lot Number 24BMC065; DYNJ41050C, Lot Number 24BMH762; DYNJ43469D, Lot Number 24BMJ703; DYNJ43469D, Lot Number 22GME883; DYNJ43816B, Lot Number 24BMC291; DYNJ45323F, Lot Number 24BMH826; DYNJ45323F, Lot Number 24BME359; DYNJ45323F, Lot Number 24BMH826; DYNJ45323F, Lot Number 24BME359; DYNJ45984I, Lot Number 24BMD224; DYNJ46353F, Lot Number 24BMJ357; DYNJ46353F, Lot Number 24BMJ357; DYNJ46401B, Lot Number 24BMJ852; DYNJ46402A, Lot Number 24BMD163; DYNJ46419C, Lot Number 24BMA912; DYNJ47646D, Lot Number 22BBU885; DYNJ47646D, Lot Number 21LBD129; DYNJ47646D, Lot Number 21KBQ988; DYNJ47646D, Lot Number 21KBH535; DYNJ47646D, Lot Number 21GBG146; DYNJ47646F, Lot Number 22EMF795; DYNJ47646F, Lot Number 22DMI820; DYNJ47646G, Lot Number 22LMF310; DYNJ47646G, Lot Number 22KMB755; DYNJ47646G, Lot Number 22GMC576; DYNJ47646I, Lot Number 24DMD242; DYNJ47804F, Lot Number 24BMC323; DYNJ47975C, Lot Number 24BMH340; DYNJ47975C, Lot Number 24BMA889; DYNJ49062DH, Lot Number 22OBK858 ; DYNJ49062DH, Lot Number 22DBM918 ; DYNJ49062DH, Lot Number 22GBM384 ; DYNJ49062DH, Lot Number 22GBV475 ; DYNJ49062DH, Lot Number 22HBG368 ; DYNJ49062DH, Lot Number 22KBO947 ; DYNJ49062DH, Lot Number 23CBK044 ; DYNJ49062DH, Lot Number 23JBI076 ; DYNJ49062DH, Lot Number 23JBV484 ; DYNJ49062DH, Lot Number 21ABL438 ; DYNJ49062DH, Lot Number 21DBP350 ; DYNJ49217C, Lot Number 24BMA571; DYNJ49217C, Lot Number 23KMH769; DYNJ49217C, Lot Number 23HMA560; DYNJ49217C, Lot Number 23FMF513; DYNJ49217C, Lot Number 23DME875; DYNJ49217C, Lot Number 23AMI574; DYNJ50212B, Lot Number 24CME981; DYNJ50212B, Lot Number 24BMD225; DYNJ50236B, Lot Number 23CDB845; DYNJ50236B, Lot Number 23BDA939; DYNJ50236B, Lot Number 22LDA599; DYNJ50236B, Lot Number 22KDA255; DYNJ50236B, Lot Number 22IDB847; DYNJ50236B, Lot Number 22GDA060; DYNJ50236B, Lot Number 22EDB782; DYNJ50779F, Lot Number 24BMF696; DYNJ51637CH, Lot Number 21HMA830 ; DYNJ51637CH, Lot Number 22CME390 ; DYNJ51637CH, Lot Number 22DMA354 ; DYNJ51637CH, Lot Number 22FMF840 ; DYNJ51637CH, Lot Number 22GMA602 ; DYNJ51637CH, Lot Number 22GMC773 ; DYNJ51637CH, Lot Number 22HMG082 ; DYNJ51637CH, Lot Number 22HMI131 ; DYNJ51637CH, Lot Number 23BME051 ; DYNJ51637CH, Lot Number 23BME054 ; DYNJ51637CH, Lot Number 23CMC268 ; DYNJ51637CH, Lot Number 23FME905 ; DYNJ51637CH, Lot Number 23HMB810 ; DYNJ51637CH, Lot Number 23JMF462 ; DYNJ51637CH, Lot Number 23KMA248 ; DYNJ51637CH, Lot Number 23LMA209 ; DYNJ51637CH, Lot Number 24AMC132 ; DYNJ51765BH, Lot Number 21BMA848 ; DYNJ51765BH, Lot Number 21CMA714 ; DYNJ51765BH, Lot Number 21DMC139 ; DYNJ51765BH, Lot Number 21EME407 ; DYNJ51765BH, Lot Number 21GMD394 ; DYNJ51765BH, Lot Number 21HMC715 ; DYNJ51765BH, Lot Number 21JMH560 ; DYNJ51765BH, Lot Number 19BKB894 ; DYNJ51765BH, Lot Number 21KMH433 ; DYNJ51765BH, Lot Number 19DKB022 ; DYNJ51765BH, Lot Number 21LMF423 ; DYNJ51765BH, Lot Number 22BMG600 ; DYNJ51765BH, Lot Number 22DMD612 ; DYNJ51765BH, Lot Number 22EME806 ; DYNJ51765BH, Lot Number 22FMH681 ; DYNJ51765BH, Lot Number 22HME321 ; DYNJ51765BH, Lot Number 22IMA668 ; DYNJ51765BH, Lot Number 23AMB196 ; DYNJ51765BH, Lot Number 19EBQ420 ; DYNJ51765BH, Lot Number 23EMB331 ; DYNJ51765BH, Lot Number 19RBA901 ; DYNJ51765BH, Lot Number 23GMD555 ; DYNJ51765BH, Lot Number 23HMF440 ; DYNJ51765BH, Lot Number 23IMG168 ; DYNJ51765BH, Lot Number 23JMH006 ; DYNJ51765BH, Lot Number 19HBV113 ; DYNJ51765BH, Lot Number 19JBQ069 ; DYNJ51765BH, Lot Number 19KBP770 ; DYNJ51765BH, Lot Number 19LBE738 ; DYNJ53225A, Lot Number 23CBX606; DYNJ53225A, Lot Number 23BBI160; DYNJ53225A, Lot Number 23ABE934; DYNJ53225A, Lot Number 22KBP555; DYNJ53225A, Lot Number 22JBB108; DYNJ53225A, Lot Number 22GBP836; DYNJ53225A, Lot Number 22EBJ911; DYNJ53225A, Lot Number 23CBX606; DYNJ53225A, Lot Number 23BBI160; DYNJ53225A, Lot Number 23ABE934; DYNJ53225A, Lot Number 22KBP555; DYNJ53225A, Lot Number 22JBB108; DYNJ53225A, Lot Number 22GBP836; DYNJ53225A, Lot Number 22EBJ911; DYNJ53241A, Lot Number 24ALB107; DYNJ53397C, Lot Number 24BMI147; DYNJ54704FH, Lot Number 21KBF606 ; DYNJ54704FH, Lot Number 21LBJ955 ; DYNJ54704FH, Lot Number 21LBW035 ; DYNJ54704FH, Lot Number 22ABN217 ; DYNJ54704FH, Lot Number 22DBC632 ; DYNJ54704FH, Lot Number 22EBS077 ; DYNJ54704FH, Lot Number 22GBL864 ; DYNJ54704FH, Lot Number 22HBS918 ; DYNJ54704FH, Lot Number 22JBM435 ; DYNJ54704FH, Lot Number 22KBD295 ; DYNJ54704FH, Lot Number 23ABQ511 ; DYNJ54704FH, Lot Number 23BBD735 ; DYNJ54704FH, Lot Number 23DBI843 ; DYNJ54704FH, Lot Number 23EBL306 ; DYNJ54704FH, Lot Number 23FBB640 ; DYNJ54704FH, Lot Number 23GBL713 ; DYNJ54704FH, Lot Number 23IBU922 ; DYNJ54704FH, Lot Number 23JBU316 ; DYNJ54759H , Lot Number 21KBQ998 ; DYNJ54759H , Lot Number 22BBT936 ; DYNJ54759H , Lot Number 22OBA976 ; DYNJ54759H , Lot Number 19CBS587 ; DYNJ54759H , Lot Number 19TBD937 ; DYNJ54759H , Lot Number 22LBR919 ; DYNJ54759H , Lot Number 23ABJ140 ; DYNJ54759H , Lot Number 19KBG955 ; DYNJ54759H , Lot Number 19LBU424 ; DYNJ54759H , Lot Number 23DBB536 ; DYNJ54759H , Lot Number 23FBN119 ; DYNJ54759H , Lot Number 23HBC296 ; DYNJ54759H , Lot Number 20DBP692 ; DYNJ54759H , Lot Number 20EBQ535 ; DYNJ54759H , Lot Number 20GBS306 ; DYNJ54759H , Lot Number 20HBL697 ; DYNJ54759H , Lot Number 20IBS035 ; DYNJ54759H , Lot Number 20LBM637 ; DYNJ55352A, Lot Number 23KBN830; DYNJ55352A, Lot Number 23JBO433; DYNJ55352A, Lot Number 23HBK208; DYNJ55352A, Lot Number 23GBI660; DYNJ55352A, Lot Number 23FBK324; DYNJ55352A, Lot Number 23EBF584; DYNJ55352A, Lot Number 23CBN279; DYNJ55352A, Lot Number 22LBE925; DYNJ55352A, Lot Number 22LBB760; DYNJ55352A, Lot Number 22KBH188; DYNJ55352A, Lot Number 22JBL274; DYNJ55352A, Lot Number 22IBC941; DYNJ55352A, Lot Number 22HBL854; DYNJ55352A, Lot Number 22GBH824; DYNJ55352A, Lot Number 22FBL491; DYNJ55352A, Lot Number 22EBH563; DYNJ55352A, Lot Number 22BBW822; DYNJ55352A, Lot Number 22MBA245; DYNJ55352A, Lot Number 21LBT347; DYNJ55352A, Lot Number 21VBB834; DYNJ55352A, Lot Number 21HBQ697; DYNJ55352A, Lot Number 21DBS097; DYNJ55352A, Lot Number 21BBM970; DYNJ55352A, Lot Number 20LBS769; DYNJ55352A, Lot Number 20JBL438; DYNJ55352A, Lot Number 20GBA541; DYNJ55352A, Lot Number 20BBK164; DYNJ55352A, Lot Number 20ABZ125; DYNJ56255B, Lot Number 24BLA925; DYNJ56378G, Lot Number 24CMC598; DYNJ56378G, Lot Number 24BMA499; DYNJ57685BH, Lot Number 21HMG832 ; DYNJ57685BH, Lot Number 22AMC145 ; DYNJ57685BH, Lot Number 22CMJ126 ; DYNJ57685BH, Lot Number 22EMA550 ; DYNJ57685BH, Lot Number 22FMA863 ; DYNJ57685BH, Lot Number 22GME902 ; DYNJ57685BH, Lot Number 22HMC011 ; DYNJ57685BH, Lot Number 22HMI037 ; DYNJ57685BH, Lot Number 22KMH699 ; DYNJ57685BH, Lot Number 23BMD485 ; DYNJ57685BH, Lot Number 23EMB679 ; DYNJ57685BH, Lot Number 23EMK046 ; DYNJ57685BH, Lot Number 23FMI778 ; DYNJ57685BH, Lot Number 23GME136 ; DYNJ57685BH, Lot Number 23IMD282 ; DYNJ57685BH, Lot Number 23JMA052 ; DYNJ57685BH, Lot Number 23JMC428 ; DYNJ57685BH, Lot Number 23LME792 ; DYNJ57685BH, Lot Number 24BMH831 ; DYNJ57723C, Lot Number 24BMD176; DYNJ59893BH, Lot Number 21KBW276 ; DYNJ59893BH, Lot Number 22BBJ057 ; DYNJ59893BH, Lot Number 22NBB317 ; DYNJ59893BH, Lot Number 22NBB319 ; DYNJ59893BH, Lot Number 22EBH526 ; DYNJ59893BH, Lot Number 22HBH440 ; DYNJ59893BH, Lot Number 22HBM205 ; DYNJ59893BH, Lot Number 22JBS681 ; DYNJ59893BH, Lot Number 22KBH312 ; DYNJ59893BH, Lot Number 22LBB108 ; DYNJ59893BH, Lot Number 23GBD440 ; DYNJ59893BH, Lot Number 23GBS355 ; DYNJ59893BH, Lot Number 23KBA502 ; DYNJ59893BH, Lot Number 23KBW947 ; DYNJ59893BH, Lot Number 23LBS294 ; DYNJ59898BH, Lot Number 21KBN438 ; DYNJ59898BH, Lot Number 21LBF430 ; DYNJ59898BH, Lot Number 22BBB648 ; DYNJ59898BH, Lot Number 22CBC054 ; DYNJ59898BH, Lot Number 22DBM996 ; DYNJ59898BH, Lot Number 22EBR846 ; DYNJ59898BH, Lot Number 22GBM877 ; DYNJ59898BH, Lot Number 22HBR328 ; DYNJ59898BH, Lot Number 22JBN705 ; DYNJ59898BH, Lot Number 22JBQ762 ; DYNJ59898BH, Lot Number 22LBE131 ; DYNJ59898BH, Lot Number 23ABR592 ; DYNJ59898BH, Lot Number 23BBO064 ; DYNJ59898BH, Lot Number 23DBM515 ; DYNJ59898BH, Lot Number 23EBQ916 ; DYNJ59898BH, Lot Number 23FBS147 ; DYNJ59898BH, Lot Number 23GBL692 ; DYNJ59898BH, Lot Number 23HBP704 ; DYNJ59898BH, Lot Number 23JBN221 ; DYNJ59898BH, Lot Number 23KBS374 ; DYNJ59931BH, Lot Number 22BBX590 ; DYNJ59931BH, Lot Number 22DBM388 ; DYNJ59931BH, Lot Number 22DBS611 ; DYNJ59931BH, Lot Number 22FBU003 ; DYNJ59931BH, Lot Number 22HBI355 ; DYNJ59931BH, Lot Number 22KBH299 ; DYNJ59931BH, Lot Number 22LBA917 ; DYNJ59931BH, Lot Number 23ABL095 ; DYNJ59931BH, Lot Number 23CBU118 ; DYNJ59931BH, Lot Number 23EBN753 ; DYNJ59931BH, Lot Number 23GBC772 ; DYNJ59931BH, Lot Number 23JBO886 ; DYNJ59931BH, Lot Number 23KBP469 ; DYNJ59931BH, Lot Number 23KBT133 ; DYNJ59931BH, Lot Number 23LBP220 ; DYNJ60112B, Lot Number 24BMG407; DYNJ60405C, Lot Number 24CMB841; DYNJ60405C, Lot Number 24BMC988; DYNJ60424A, Lot Number 24BMC104; DYNJ60533C, Lot Number 24CMB458; DYNJ60533C, Lot Number 24BMC106; DYNJ60533C, Lot Number 24CMB458; DYNJ60533C, Lot Number 24BMC106; DYNJ60655D, Lot Number 24BMG238; DYNJ60655D, Lot Number 24BMA787; DYNJ62086F, Lot Number 24BMI705; DYNJ62086F, Lot Number 24BMI705; DYNJ62609DH, Lot Number 22JLA172 ; DYNJ62609DH, Lot Number 22KLA005 ; DYNJ62609DH, Lot Number 23ALA693 ; DYNJ62609DH, Lot Number 23LLA772 ; DYNJ62609DH, Lot Number 24ALA003 ; DYNJ62609DH, Lot Number 24ALB059 ; DYNJ62980A, Lot Number 24AMB177; DYNJ62980A, Lot Number 23IMH731; DYNJ62980A, Lot Number 23IMC714; DYNJ62980A, Lot Number 23IMA210; DYNJ62980A, Lot Number 23HMH070; DYNJ62980A, Lot Number 23GMI101; DYNJ62980A, Lot Number 23GMB319; DYNJ62980A, Lot Number 23FMI809; DYNJ62980A, Lot Number 23EMC205; DYNJ62980A, Lot Number 23DME222; DYNJ62980A, Lot Number 23BMD380; DYNJ62980A, Lot Number 22LMF630; DYNJ62980A, Lot Number 22JMB790; DYNJ62980A, Lot Number 22IMC933; DYNJ62980A, Lot Number 22HMB197; DYNJ62980A, Lot Number 22EMH686; DYNJ62980A, Lot Number 22EMA467; DYNJ62980A, Lot Number 22CMJ137; DYNJ63686C, Lot Number 23JBP460; DYNJ64525BH, Lot Number 21JBR200 ; DYNJ64525BH, Lot Number 21KBM325 ; DYNJ64525BH, Lot Number 21LBJ452 ; DYNJ64525BH, Lot Number 22BBS908 ; DYNJ64525BH, Lot Number 22NBE425 ; DYNJ64525BH, Lot Number 22CBZ142 ; DYNJ64525BH, Lot Number 22EBH209 ; DYNJ64525BH, Lot Number 22GBT896 ; DYNJ64525BH, Lot Number 23BBD176 ; DYNJ64525BH, Lot Number 23BBO100 ; DYNJ64525BH, Lot Number 23EBJ205 ; DYNJ64525BH, Lot Number 23EBW847 ; DYNJ64525BH, Lot Number 23FBR366 ; DYNJ64525BH, Lot Number 23GBK785 ; DYNJ64525BH, Lot Number 23GBT353 ; DYNJ64525BH, Lot Number 23HBK725 ; DYNJ64525BH, Lot Number 23IBE961 ; DYNJ64525BH, Lot Number 23JBG437 ; DYNJ64525BH, Lot Number 23LBA330 ; DYNJ64546BH, Lot Number 21IBL728 ; DYNJ64546BH, Lot Number 21KBP107 ; DYNJ64546BH, Lot Number 23EBC124 ; DYNJ64546BH, Lot Number 23JBD109 ; DYNJ64548DH, Lot Number 23HBC766 ; DYNJ64548DH, Lot Number 23HBI072 ; DYNJ64587AH, Lot Number 22BBY939 ; DYNJ64587AH, Lot Number 22CBU206 ; DYNJ64587AH, Lot Number 22OBK428 ; DYNJ64587AH, Lot Number 22EBL781 ; DYNJ64587AH, Lot Number 22EBL782 ; DYNJ64587AH, Lot Number 22KBV934 ; DYNJ64587AH, Lot Number 23ABN267 ; DYNJ64587AH, Lot Number 23ABS667 ; DYNJ64587AH, Lot Number 23CBK426 ; DYNJ64587AH, Lot Number 23DBJ436 ; DYNJ64587AH, Lot Number 23EBI108 ; DYNJ64587AH, Lot Number 23FBH225 ; DYNJ64587AH, Lot Number 23GBB016 ; DYNJ64587AH, Lot Number 23HBB670 ; DYNJ64587AH, Lot Number 23JBA945 ; DYNJ64587AH, Lot Number 23JBM612 ; DYNJ64587AH, Lot Number 23KBL706 ; DYNJ64587AH, Lot Number 23LBT821 ; DYNJ64620H , Lot Number 21IBD120 ; DYNJ64620H , Lot Number 21JBC520 ; DYNJ64620H , Lot Number 21KBC208 ; DYNJ64620H , Lot Number 21LBG582 ; DYNJ64620H , Lot Number 22ABA558 ; DYNJ64620H , Lot Number 22CBA435 ; DYNJ64620H , Lot Number 22OBE449 ; DYNJ64620H , Lot Number 22DBU504 ; DYNJ64620H , Lot Number 22FBJ147 ; DYNJ64620H , Lot Number 22GBE449 ; DYNJ64620H , Lot Number 22IBK125 ; DYNJ64620H , Lot Number 22IBP210 ; DYNJ64620H , Lot Number 22KBE166 ; DYNJ64620H , Lot Number 22LBK893 ; DYNJ64620H , Lot Number 23ABJ191 ; DYNJ64620H , Lot Number 23BBQ870 ; DYNJ64620H , Lot Number 23EBA313 ; DYNJ64620H , Lot Number 23FBO928 ; DYNJ64620H , Lot Number 23GBL683 ; DYNJ64620H , Lot Number 23GBR414 ; DYNJ64620H , Lot Number 23JBM176 ; DYNJ64620H , Lot Number 23KBC219 ; DYNJ64620H , Lot Number 23LBB152 ; DYNJ64626BH, Lot Number 22HBF983 ; DYNJ64626BH, Lot Number 22HBO590 ; DYNJ64626BH, Lot Number 22IBJ761 ; DYNJ64626BH, Lot Number 22JBG356 ; DYNJ64626BH, Lot Number 22KBS705 ; DYNJ64626BH, Lot Number 22LBO971 ; DYNJ64626BH, Lot Number 23ABN271 ; DYNJ64626BH, Lot Number 23EBF819 ; DYNJ64626BH, Lot Number 23FBU846 ; DYNJ64626BH, Lot Number 23GBO963 ; DYNJ64626BH, Lot Number 23IBR519 ; DYNJ64626BH, Lot Number 23JBL760 ; DYNJ64626BH, Lot Number 23LBG554 ; DYNJ64936BH, Lot Number 22HBY850 ; DYNJ64936BH, Lot Number 22LBQ523 ; DYNJ64936BH, Lot Number 23ABK408 ; DYNJ64936BH, Lot Number 23BBP118 ; DYNJ64936BH, Lot Number 23DBP201 ; DYNJ64936BH, Lot Number 23EBI424 ; DYNJ64936BH, Lot Number 23FBF158 ; DYNJ64936BH, Lot Number 23GBM124 ; DYNJ64936BH, Lot Number 23HBW852 ; DYNJ64936BH, Lot Number 23JBA954 ; DYNJ64936BH, Lot Number 23KBI356 ; DYNJ64990B, Lot Number 24BMD874; DYNJ65162BH, Lot Number 23HBD541 ; DYNJ65162BH, Lot Number 23IBC495 ; DYNJ65162BH, Lot Number 23JBU189 ; DYNJ65162BH, Lot Number 23KBA866 ; DYNJ65162BH, Lot Number 23LBL832 ; DYNJ65162BH, Lot Number 23LBT273 ; DYNJ65238B, Lot Number 23GMB562; DYNJ65238B, Lot Number 22HMH793; DYNJ65238B, Lot Number 22FMI017; DYNJ65238B, Lot Number 22EMF834; DYNJ65238B, Lot Number 22BMG103; DYNJ65481, Lot Number 24BMJ722; DYNJ65481, Lot Number 24BMF056; DYNJ65578A, Lot Number 24BMJ184; DYNJ65959B, Lot Number 24DMD333; DYNJ66007, Lot Number 24BMK046; DYNJ66007, Lot Number 24BME714; DYNJ66131A, Lot Number 24AMH848; DYNJ66259A, Lot Number 23KMG710; DYNJ66259A, Lot Number 23JMD657; DYNJ66259A, Lot Number 23IMF851; DYNJ66259A, Lot Number 23HME574; DYNJ66259A, Lot Number 23HMB253; DYNJ66259A, Lot Number 23HMA625; DYNJ66259A, Lot Number 23GMC209; DYNJ66259A, Lot Number 23FMF796; DYNJ66259A, Lot Number 23DMA168; DYNJ66259A, Lot Number 23AMI207; DYNJ66259A, Lot Number 22LMF639; DYNJ66259A, Lot Number 22KMC724; DYNJ66259A, Lot Number 22JMD922; DYNJ66259A, Lot Number 22IMH587; DYNJ66259A, Lot Number 22HMG530; DYNJ66259A, Lot Number 22HMB016; DYNJ66260A, Lot Number 23IME568; DYNJ66260A, Lot Number 23HME573; DYNJ66260A, Lot Number 23FMC879; DYNJ66260A, Lot Number 23BME124; DYNJ66260A, Lot Number 22LMG896; DYNJ66260A, Lot Number 22JMD516; DYNJ66260A, Lot Number 22HMG895; DYNJ66260A, Lot Number 22GMH445; DYNJ66366D, Lot Number 24BMD428; DYNJ66724BH, Lot Number 21JBV119 ; DYNJ66724BH, Lot Number 21WBA185 ; DYNJ66724BH, Lot Number 21LBW926 ; DYNJ66724BH, Lot Number 22BBY511 ; DYNJ66724BH, Lot Number 22OBC003 ; DYNJ66724BH, Lot Number 22EBM894 ; DYNJ66724BH, Lot Number 22FBS448 ; DYNJ66724BH, Lot Number 22HBB047 ; DYNJ66724BH, Lot Number 22IBL266 ; DYNJ66724BH, Lot Number 22JBK074 ; DYNJ66724BH, Lot Number 22KBS734 ; DYNJ66724BH, Lot Number 23ABR605 ; DYNJ66724BH, Lot Number 23BBO068 ; DYNJ66724BH, Lot Number 23EBB043 ; DYNJ66724BH, Lot Number 23EBP768 ; DYNJ66724BH, Lot Number 23GBC395 ; DYNJ66724BH, Lot Number 23HBJ373 ; DYNJ66724BH, Lot Number 23IBU041 ; DYNJ66724BH, Lot Number 23JBW840 ; DYNJ66724BH, Lot Number 23KBS964 ; DYNJ66724BH, Lot Number 23LBU960 ; DYNJ67378B, Lot Number 24BMG688; DYNJ67567A, Lot Number 24BMC110; DYNJ67599, Lot Number 21IBN436; DYNJ67599, Lot Number 21HBC916; DYNJ67599, Lot Number 21GBL479; DYNJ67599, Lot Number 21IBN436; DYNJ67599, Lot Number 21HBC916; DYNJ67599, Lot Number 21GBL479; DYNJ67685, Lot Number 21HBE807; DYNJ67685, Lot Number 21HBB723; DYNJ67685B, Lot Number 24CMC608; DYNJ68548, Lot Number 21HBC818; DYNJ68548, Lot Number 21GBJ891; DYNJ68548A, Lot Number 22BBP809; DYNJ68548A, Lot Number 21LBU339; DYNJ68548A, Lot Number 21KBX361; DYNJ68548A, Lot Number 21KBJ138; DYNJ68548A, Lot Number 21IBS100; DYNJ68613, Lot Number 24CMA113; DYNJ69054, Lot Number 24BMF146; DYNJ69283A, Lot Number 24BMG983; DYNJ80163DH, Lot Number 23GBT488 ; DYNJ80163DH, Lot Number 23HBB653 ; DYNJ80163DH, Lot Number 23LBU923 ; DYNJ80163DH, Lot Number 23LBU924 ; DYNJ80163DH, Lot Number 24ABC425 ; DYNJ80163DH, Lot Number 24ABD343 ; DYNJ80163DH, Lot Number 24ABU088 ; DYNJ80367, Lot Number 22BBK528; DYNJ80461, Lot Number 22ABB295; DYNJ80758H , Lot Number 22FBJ131 ; DYNJ80758H , Lot Number 22GBR587 ; DYNJ80758H , Lot Number 23IBH541 ; DYNJ80758H , Lot Number 23KBG834 ; DYNJ80758H , Lot Number 23LBU095 ; DYNJ82079, Lot Number 24DMD246; DYNJ82079, Lot Number 24BMH760; DYNJ82079, Lot Number 24BMC151; DYNJ82176H , Lot Number 23FDA956 ; DYNJ82176H , Lot Number 23FDC367 ; DYNJ82176H , Lot Number 23HDA640 ; DYNJ82176H , Lot Number 23HDB105 ; DYNJ82176H , Lot Number 23HMI738 ; DYNJ82176H , Lot Number 23JMF362 ; DYNJ82186AH, Lot Number 23FDC203 ; DYNJ82186AH, Lot Number 23HDA807 ; DYNJ82186AH, Lot Number 23IDB076 ; DYNJ82186AH, Lot Number 23IDB979 ; DYNJ82186AH, Lot Number 23JDB622 ; DYNJ82186AH, Lot Number 23IMF653 ; DYNJ82186AH, Lot Number 23JMF016 ; DYNJ82186AH, Lot Number 23LMD570 ; DYNJ82189BH, Lot Number 23JDA072 ; DYNJ82189BH, Lot Number 23JMC255 ; DYNJ82191CH, Lot Number 23LMG771 ; DYNJ82207BH, Lot Number 23JDA803 ; DYNJ82207BH, Lot Number 23JMF293 ; DYNJ82207BH, Lot Number 23LMG278 ; DYNJ82269H , Lot Number 23GDA711 ; DYNJ82269H , Lot Number 23IMD001 ; DYNJ82269H , Lot Number 23JMD240 ; DYNJ82280BH, Lot Number 23IDB345 ; DYNJ82280BH, Lot Number 23KMB459 ; DYNJ82321H , Lot Number 23GDB512 ; DYNJ82321H , Lot Number 23IDC133 ; DYNJ82321H , Lot Number 23JMF075 ; DYNJ82321H , Lot Number 23KMC469 ; DYNJ82321H , Lot Number 23KMG811 ; DYNJ82323AH, Lot Number 23LMC240 ; DYNJ82323H , Lot Number 23HDA104 ; DYNJ82323H , Lot Number 23HDB457 ; DYNJ82323H , Lot Number 23IDB943 ; DYNJ82323H , Lot Number 23IDB944 ; DYNJ82323H , Lot Number 23IDC304 ; DYNJ82323H , Lot Number 23JDA664 ; DYNJ82323H , Lot Number 23JMH146 ; DYNJ82323H , Lot Number 23JMJ622 ; DYNJ82418, Lot Number 24DMC582; DYNJ82418, Lot Number 24CME395; DYNJ82418, Lot Number 23KMI875; DYNJ82418, Lot Number 23JMD311; DYNJ82418, Lot Number 23GME377; DYNJ82418, Lot Number 23AMD137; DYNJ82418, Lot Number 22HMI054; DYNJ83539H , Lot Number 23ABV291 ; DYNJ83539H , Lot Number 23DBA067 ; DYNJ83539H , Lot Number 23EBT383 ; DYNJ83539H , Lot Number 23GBC418 ; DYNJ83539H , Lot Number 23HBK248 ; DYNJ83539H , Lot Number 23IBL665 ; DYNJ83539H , Lot Number 23JBP839 ; DYNJ83539H , Lot Number 23LBG640 ; DYNJ83554, Lot Number 24DMC657; DYNJ83554, Lot Number 24BMH275; DYNJ83626, Lot Number 24BMH313; DYNJ83829H , Lot Number 23CBP984 ; DYNJ83829H , Lot Number 23FBH266 ; DYNJ83924AH, Lot Number 23KBN818 ; DYNJ83924AH, Lot Number 23KBQ984 ; DYNJ84632A, Lot Number 24BMJ660; DYNJ84908, Lot Number 24BMG718; DYNJ84908, Lot Number 24BMA886; DYNJ84908, Lot Number 24BMG718; DYNJ84908, Lot Number 24BMA886; DYNJ907263A, Lot Number 21IBM798; DYNJ907263A, Lot Number 21HBQ039; DYNJ908249AH , Lot Number 22JBO556 ; DYNJ908249AH , Lot Number 22KBI072 ; DYNJ908249AH , Lot Number 23ABV137 ; DYNJ908249AH , Lot Number 23BBK741 ; DYNJ908249AH , Lot Number 23GBC526 ; DYNJ908249AH , Lot Number 23GBN378 ; DYNJ908249AH , Lot Number 23IBD120 ; DYNJCEN03A, Lot Number 24CMC586; DYNJHTCLAB8, Lot Number 23EBI285; DYNJHTCLAB8, Lot Number 23DBF052; DYNJHTCLAB8, Lot Number 23CBH448; DYNJHTCLAB8, Lot Number 23BBA180; DYNJHTCLAB8, Lot Number 22LBO751; DYNJHTCLAB9, Lot Number 23HDB484; DYNJHTCLAB9, Lot Number 23GDC184; DYNJHTCLAB9, Lot Number 23FDB500; DYNJHTCLAB9, Lot Number 23EDC033; DYNJHTCLAB9, Lot Number 23EDB553; DYNJHTCLAB9, Lot Number 23EDA186; DYNJHTCLAB9, Lot Number 23DDB242; DYNJHTCLAB9, Lot Number 23BDC246; DYNJHTCLAB9, Lot Number 23ADB707; DYNJHTCLAB9, Lot Number 23ADB150; DYNJHTCLAB9, Lot Number 22LDA748; DYNJHTCLAB9, Lot Number 22KDB190; DYNJHTCLAB9, Lot Number 22JDB159; DYNJHTCLAB9, Lot Number 22IDA173; DYNJHTCLAB9, Lot Number 22HDA720; DYNJHTCLAB9, Lot Number 22GDA259; DYNJHTCLAB9, Lot Number 22FDB908; DYNJHTCLAB9, Lot Number 22EDC029; DYNJHTCOMNH, Lot Number 23IBI436 ; DYNJHTCOMNH, Lot Number 23IBN783 ; DYNJHTCOMNH, Lot Number 23LBF855 ; DYNJHTCOMNH, Lot Number 23LBV211 ; DYNJT2112F, Lot Number 24CME323; DYNJT2112F, Lot Number 24BMB398; DYNJV0013Y, Lot Number 22IMC754; DYNJV0013Y, Lot Number 22FMH633; DYNJV0282K, Lot Number 24BMG685; ECVC2740C, Lot Number 2024030590; ECVC3285A, Lot Number 2023013190; ECVC3285A, Lot Number 2023060990; ECVC3285A, Lot Number 2023071490; ECVC3285A, Lot Number 2023092890; ECVC3285A, Lot Number 2023110790; ECVC3285A, Lot Number 2024020990; ECVC3285A, Lot Number 2024030890; ECVC3286, Lot Number 2023030790; ECVC3286, Lot Number 2023040490; ECVC3286, Lot Number 2023051890; ECVC3286, Lot Number 2024020690; ECVC7625, Lot Number 2022121690; ECVC7625, Lot Number 2023022290; ECVC7625A, Lot Number 2023072790; ECVC7835B, Lot Number 2023061280; ECVC7835B, Lot Number 2023111380; ECVC8040, Lot Number 2022121690; ECVC8040A, Lot Number 2023072790; MNS3560, Lot Number 2023020690; PHS204857009H, Lot Number 21AMA021 ; PHS204857009H, Lot Number 21DMD360 ; PHS204857009H, Lot Number 21FMD611 ; PHS204857009H, Lot Number 21JME830 ; PHS204857009H, Lot Number 19AKC471 ; PHS204857009H, Lot Number 21LMD336 ; PHS204857009H, Lot Number 19EKA240 ; PHS204857009H, Lot Number 19HKA321 ; PHS204857009H, Lot Number 19JKA374 ; PHS204857009H, Lot Number 22BMF591 ; PHS204857009H, Lot Number 22EMI508 ; PHS204857009H, Lot Number 22GME955 ; PHS204857009H, Lot Number 22LMI908 ; PHS204857009H, Lot Number 23BMA870 ; PHS204857009H, Lot Number 23DMB267 ; PHS204857009H, Lot Number 23GMF664 ; PHS204857009H, Lot Number 23IMC416 ; PHS204857009H, Lot Number 23LMA315 ; S5078BF09, Lot Number 2023091990; SAMPC0660, Lot Number 21DDA622 ; SUT20955, Lot Number 2023091990; VASC1409, Lot Number 23IBN852; VASC1409, Lot Number 23FBI203; VASC1410, Lot Number 23FBI202; WCH02BGOD, Lot Number 24DBE047
FEI Number 1417592
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactKarin Johnson
886-359-1704
Manufacturer Reason
for Recall
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.
DistributionWorldwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = LDF
-
-