 Class 2 Device Recall Medlinesee related information
Date Initiated by FirmApril 08, 2024
Date PostedSeptember 06, 2024
Recall Status1 Open3, Classified
Recall NumberZ-3053-2024
Recall Event ID 94583
510(K)NumberK203526 
Product Classification Arthroscope - Product Code HRX
ProductMedline Convenience kits labeled as: 1) C SCOPE DISPOSABLE KIT, Pack Code 613034700; 2) C SCOPE DISPOSABLE KIT, Pack Code 613034701; 3) C SCOPE DISPOSABLE KIT, Pack Code 613034702
Code Information 613034700, Lot Number 20ADD340 ; 613034700, Lot Number 20BDC393 ; 613034700, Lot Number 20BDC472 ; 613034700, Lot Number 20BDC473 ; 613034700, Lot Number 20FDA390 ; 613034700, Lot Number 20FDA896 ; 613034700, Lot Number 20FDA897 ; 613034700, Lot Number 20FDA898 ; 613034700, Lot Number 20FDA896A; 613034700, Lot Number 20JDB120 ; 613034700, Lot Number 21CDA660 ; 613034700, Lot Number 21DDB193 ; 613034700, Lot Number 21IDA396 ; 613034700, Lot Number 21LDB900 ; 613034700, Lot Number 22ADB549 ; 613034701, Lot Number 21CDA617 ; 613034701, Lot Number 21DDB192 ; 613034701, Lot Number 21IDB791 ; 613034701, Lot Number 21JDB094 ; 613034701, Lot Number 22BDB700 ; 613034702, Lot Number 21CDA618 ; 613034702, Lot Number 21DDB222 ; 613034702, Lot Number 21IDB804 ; 613034702, Lot Number 21LDB713
FEI Number 1417592
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactKarin Johnson
886-359-1704
Manufacturer Reason
for Recall
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.
DistributionWorldwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = HRX
-
-