• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall BD Alaris

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall BD Alarissee related information
Date Initiated by FirmAugust 15, 2024
Date PostedSeptember 13, 2024
Recall Status1 Open3, Classified
Recall NumberZ-3139-2024
Recall Event ID 94937
510(K)NumberK211218 
Product Classification Pump, infusion - Product Code FRN
ProductAlaris PCA Module Model 8120, a modular infusion pump and monitoring system
Code Information Model Number: 8120 UDI-DI codes: 10885403516030, 10885403515316, 10885403812033, 10885403812026, 10885403494291, 10885403812002 Serial Numbers: 16473112 16474187 16458453 16458381 16458632 16472825 16456729 16457626 16453240 16458585 16450364 16459191 16458699 16473046 16458427 16473677 16472463 16459701 16472829 16458736 16458700 16457828 16460132 16455368 16472831 16473116 16457873 16460396 16473618 16473688 16458544 16458121 16473391 16527718 16527910 16441259 16441394 16528387 16441214 16477919 16441048 16525578 16440759 16441453 16525916 16441034 16441160 16441164 16450345 16441601 16454386 16442620 16479558 16452454 16358256 16452960 16453953 16452089 16472866 16452442 16449483 16282569 16452076 16473123 16453256 16448217 16452996 16438128 16453852 16282541 16453263 16453423 16454448 16452576 16450489 16453424 16453702 16453853 16452088 16452445 16453527 16298268 16454345 16282360 16452394 16280503 16452485 16298349 16440752 16441277 16453848 16451983 16453700 16454300 16453421 16438875 16452129 16453535 16452508 16472830 16472480 16442014 16282559 16441295 16450415 16450090 16452417 16471744 16452387 16453522 16453579 16452626 16452536 16282579 16441545 16472995 16282616 16438173 16449362 16449280 16449336 16527580 16441101 16525189 16526900 16526177 16511913 16441474 16515632 16441290 16527326 16440746 16440745 16283136 16284350 16284295 16278278 16282406 16282701 16284376 16284539 16285656 16283024 16285385 16280047 16282941 16282871 16284654 16285713 16280325 16284548 16282671 16284131 16282990 16282603 16285595 16280391 16285788 16285305 16284591 16285009 16280279 16283279 16284313 16285063 16282546 16283207 16282639 16282557 16283012 16284466 16279819 16283281 16282407 16282805 16282548 16284794 16285661 16283272 16282566 16285657 16284238 16285665 16285617 16279174 16283177 16283230 16285241 16282586 16280036 16280501 16284709 16284608 16285706 16285658 16285662 16282640 16282259 16282680 16284959 16284777 16285378 16285693 16279615 16284070 16282678 16279855 16278366 16285509 16282857 16282700 16283876 16279578 16282608 16285592 16285664 16282596 16284183 16282583 16359215 16359421 16399909 16359158 16359537 16397508 16358882 16359302 16281960 16399657 16282573 16359456 16280308 16358349 16359495 16282611 16282367 16282580 16360101 16359586 16360086 16285672 16282543 16282001 16359475 16282605 16359857 16282571 16358861 16359086 16282595 16359390 16359524 16283204 16282197 16359695 16359703 16359858 16358822 16285395 16358710 16285702 16283055 16285366 16360119 16282235 16359217 16279603 16314048 16301945 16279654 16282362 16282842 16282663 16279860 16278456 16282590 16283210 16284330 16284770 16285653 16282636 16284375 16283228 16282621 16282648 16284005 16285659 16282553 16282830 16285675 16284334 16282874 16283892 16279617 16285227 16279562 16280023 16279550 16280182 16282646 16278386 16284135 16282593 16279990 16285593 16280141 16282563 16285363 16280307 16282770 16279992 16280298 16284382 16282612 16282937 16282552 16282674 16280235 16282361 16282649 16280435 16282683 16285650 16284115 16285505 16284237 16282404 16282572 16283068 16280496 16282578 16280175 16282561 16284296 16280253 16281983 16282973 16285498 16285618 16282664 16279988 16282619 16399569 16359432 16359446 16359836 16359584 16358849 16282554 16280762 16359793 16359310 16285590 16358921 16282581 16359106 16359511 16282576 16358703 16359304 16284936 16358693 16359612 16358301 16283277 16282589 16282670 16359092 16359216 16279606 16282642 16285701 16359108 16283900 16280559 16358919 16283098 16358975 16397963 16359332 16398741 16284393 16283001 16359481 16284180 16282200 16282604 16280505 16282026 16282196 16282531 16359219 16544301 16543199 16543502 16540811 16544307 16544273 16536911 16539663 16544195 16540987 16539173 16540578 16540229 16537675 16539699 16543268 16540652 16536923 16542593 16540383 16537836 16544131 16542622 16543170 16540954 16539018 16540552 16544042 16540045 16544718 16540942 16540381 16539976 16540988 16543496 16536496 16540366 16539336 16543505 16540187 16540809 16453951 16450043 16454385 16572240 16570911 16569600 16570273 16458404 16450170 16459870 16471882 16569153 16572075 16458536 16458021 16459508 16570948 16572402 16458681 16570271 16569613 16571538 16570950 16458730 16453259 16458695 16460172 16450044 16459761 16460236 16473542 16469900 16449731 16472536 16458437 16460153 16458734 16459124 16573320 16570159 16473043 16458339 16573214 16572313 16572388 16572873 16458892 16563985 16458610 16458620 16474196 16471924 16458110 16569208 16459610 16568846 16458644 16569227 16573351 16571715 16572920 16472991 16456905 16458801 16573195 16456912 16459069 16573370 16436691 16572012 16458633 16459752 16571425 16572919 16569035 16458698 16458660 16573280 16458732 16459811 16572136 16301511 16571427 16568836 16569431 16573349 16570747 16571440 16567413 16569898 16570106 16621932 16568873 16568783 16568793 16569908 16566950 16568859 16570742 16569899 16569902 16572085 16571988 16571398 16569304 16569318 16568540 16570105 16567676 16567867 16569188 16566626 16571426 16567827 16570101 16571701 16571797 16567567 16571807 16569616 16567821 16563504 16567729 16569599 16566377 16570093 16566381 16571709 16572492 16570155 16570797 16573211 16566734 16569384 16569623 16567581 16570149 16569444 16570150 16569907 16569312 16570693 16570750 16563865 16571362 16570749 16568694 16568863 16569102 16570695 16569909 16571678 16568515 16571554 16573362 16567185 16567866 16568432 16572039 16569369 16571940 16572114 16567176 16570095 16569994 16570098 16569608 16571428 16570743 16569012 16569042 16563410 16473041 16472651 16472467 16472827 16472275 16472741 16211076 16211106 16211075 16210448 16211090 16211087 16211149 16211166 16210628 16211132 16211079 16211111 16211103 16211091 16211104 16211067 16211085 16211061 16211063 16570094 16570913 16569222 16569476 16567739 16568431 16568533 16568893 16569996 16569382 16569416 16566240 16569579 16569440 16567403 16573365 16567399 16568691 16571443 16562026 16566600 16564928 16567172 16572188 16567576 16566061 16572103 16567179 16567573 16572093 16571515 16573256 16572051 16571652 16569632 16570677 16571635 16569905 16566940 16569584 16572200 16568510 16572298 16571848 16572024 16572333 16569224 16571705 16568507 16570280 16569351 16567728 16572176 16566986 16569604 16568479 16571736 16570686 16566952 16568534 16571969 16564116 16567579 16571621 16570744 16570905 16566260 16567578 16568909 16570281 16572285 16570274 16562517 16569900 16569278 16569374 16569606 16570354 16569541 16568499 16571581 16564115 16569903 16569171 16571431 16569602 16471126 16473336 16472865 16471583 16210612 16211119 16210601 16211074 16211078 16211093 16211089 16211065 16211088 16210586 16211081 16211084 16211077 16211069 16211100 16211073 16211096 16210595 16211056 16211059 16211126 16211092 16318474 16512048 16316366 16487159 16514655 16414626 16318551 16278075 16402616 16211108 16210515 16210675 16278284 16278468 16278276 16277388 16278230 16278182 16488733 16515267 16557702 16517550 16513343 16518487 16517164 16486369 16489069 16621230 16518678 16502333 16502759 16622764 16557924 16517858 16622722 16517050 16515628 16490087 16514941 16549560 16517855 16557748 16517083 16518314 16514798 16517249 16617311 16513576 16621984 16482742 16490068 16517218 16557859 16518014 16211068 16277323 16278517 16278309 16556057 16490310 16508114 16487151 16518325 16510680 16515334 16514942 16557944 16557849 16457971 16513019 16517562 16623695 16513379 16518011 16502164 16518262 16517202 16555829 16515263 16515634 16516954 16620969 16558236 16622976 16518320 16519036 16558409 16622686 16557576 16558566 16556112 16503603 16516795 16486631 16516531 16489483 16557995 16557899 16486346 16518010 16557539 16519010 16517849 16622667 16517850 16486308 16556063 16510458 16211055 16211058 16210435 16211110 16211086 16211083 16515556 16516876 16516998 16401732 16518681 16317806 16516746 16490510 16516543 16330859 16301611 16315505 16277633 16278094 16278996 16278431 16278831 16278770 16277775 16279089 16278952 16278294 16279024 16278251 16279001 16277762 16278833 16278478 16277921 16277609 16278999 16278357 16278238 16278718 16278538 16278861 16278794 16278124 16278892 16278974 16279000 16278128 16278316 16278954 16277761 16277253 16278951 16278125 16278686 16278582 16278550 16278511 16278271 16278891 16278893 16278489 16277768 16278069 16277926 16278650 16278753 16277989 16278712 16277648 16335111 16278998 16278204 16278498 16279090 16278930 16278446 16278022 16279067 16278888 16278916 16278846 16278329 16278922 16278397 16278852 16278546 16278057 16278860 16278272 16279105 16279053 16279016 16277709 16278425 16278179 16278220 16278105 16277811 16278148 16278742 16278155 16278944 16278859 16277984 16278787 16278890 16278731 16278252 16279015 16278177 16278647 16278675 16278980 16278773 16278730 16398479 16402620 16398792 16238919 16238624 16238924 16620807 16622166 16609129 16620901 16609513 16620365 16620079 16621995 16620358 16620856 16609580 16620602 16619485 16620575 16609703 16609191 16609709 16620805 16620484 16621290 16620812 16397446 16620962 16620334 16621816 16619791 16620580 16609579 16620574 16609131 16609197 16621954 16621576 16621624 16620496 16619957 16609771 16621627 16609646 16617048 16617716 16619448 16620761 16620149 16609130 16620760 16608947 16620582 16619437 16609705 16620759 16621292 16609765 16620738 16620440 16619746 16620518 16619714 16609004 16619712 16620276 16620067 16609704 16619341 16620581 16621735 16620497 16620586 16620898 16609758 16609768 16620968 16621978 16620425 16619895 16620066 16606394 16619998 16189373 16189785 16189387 16189384 16190370 16189867 16189451 16189472 16189354 16190501 16189394 16189479 16189570 16190644 16189435 16189825 16191041 16190550 16191016 16190629 16191037 16191071 16189430 16189620 16190614 16189413 16189837 16189860 16191031 16190330 16189518 16189580 16189371 16189400 16189619 16189736 16189641 16189447 16193530 16194851 16191549 16193360 16191799 16191582 16191640 16191642 16193439 16191749 16191626 16191555 16191795 16193618 16191608 16191687 16191659 16191854 16194831 16191013 16193279 16191536 16191570 16194856 16191815 16193524 16191614 16193278 16189424 16189757 16189875 16189420 16190285 16190298 16190273 16190420 16189902 16189380 16189851 16189934 16189408 16189943 16190222 16189366 16189440 16190284 16190591 16189781 16189405 16191083 16189443 16190575 16189391 16189618 16190269 16189476 16190494 16189457 16191026 16189842 16189469 16190218 16190415 16190230 16189652 16190212 16189461 16190257 16189770 16190234 16191672 16191627 16191563 16191754 16193482 16191628 16190632 16191737 16191602 16194840 16191577 16193294 16193547 16193304 16191590 16191743 16191597 16193483 16191827 16191874 16238911 16397290 16621964 16621023 16620588 16608929 16620811 16617876 16618914 16619607 16621688 16619296 16620584 16621906 16620970 16621251 16620860 16609001 16609828 16620737 16622157 16621225 16620971 16622116 16620857 16606202 16620739 16619965 16609511 16621799 16609515 16620809 16621024 16620900 16620453 16620590 16618724 16620896 16609832 16620313 16621644 16609655 16620300 16606199 16609582 16608846 16608976 16620589 16609650 16609700 16621114 16620576 16621641 16609510 16609774 16609584 16620578 16608848 16609779 16609652 16621249 16620367 16619900 16621055 16619076 16620347 16617369 16620859 16620585 16609194 16609187 16620350 16621275 16608472 16617127 16621886 16609756 16620577 16619444 16620587 16621632 16619667 16621614 16621778 16401343 16313857 16226234 16550496 16551102 16543241 16543510 16225700 16211107 16225344 16211159 16225558 16210440 16549740 16542686 16544129 16550118 16543026 16301652 16465486 16314952 16301964 16315686 16315508 16301608 16315585 16465458 16301542 16550539 16544609 16211071 16225720 16210100 16211098 16542612 16210568 16544234 16548118 16210420 16211060 16536631 16542746 16550678 16314519 16314308 16315480 16315601 16313787 16314432 16542282 16544673 16544726 16540480 16542166 16549460 16536737 16541815 16542142 16542034 16541947 16539363 16544362 16550222 16542015 16541800 16550620 16544720 16544434 16540986 16542091 16544714 16541824 16543283 16541801 16550202 16550078 16266464 16266450 16266665 16269006 16452322 16452010 16266545 16266326 16268946 16266418 16266599 16266698 16266204 16268930 16268947 16266593 16266619 16269033 16266680 16269060 16266414 16267020 16452248 16266456 16266679 16266351 16267039 16266397 16266586 16266950 16268932 16266572 16450298 16541823 16543274 16540126 16541974 16542114 16550177 16543506 16542563 16550556 16542163 16540718 16549916 16544632 16542340 16542590 16540820 16545931 16542969 16544806 16542258 16540751 16544390 16537889 16266533 16266743 16266476 16266705 16269003 16269023 16266610 16267089 16266502 16266620 16268983 16266515 16266277 16266501 16266623 16266253 16452175 16361507 16266576 16266301 16266575 16266291 16452256 16452026 16266489 16266809 16266383 16266276 16266800 16266463 16268980 16267006 16267023 16269083 16266443 16360258 16608687 16606268 16609007 16608671 16608518 16605955 16606388 16608999 16606265 16605883 16608561 16606389 16606395 16609005 16609134 16606390 16605797 16608479 16606078 16609003 16608844 16608531 16608670 16608850 16608537 16606269 16608614 16608695 16535312 16536534 16536858 16535351 16536725 16535401 16536530 16536890 16535506 16536934 16536941 16536671 16536740 16536499 16536776 16536477 16536516 16532035 16536601 16536470 16536619 16536563 16536559 16536813 16536853 16536650 16536764 16536798 16536546 16536745 16536884 16536947 16531246 16536730 16536752 16534577 16536517 16534895 16536894 16536921 16534890 16531132 16535759 16536668 16536475 16536865 16536845 16536490 16536942 16536605 16536565 16536924 16536561 16536887 16536708 16536503 16536889 16536132 16536883 16536673 16536542 16534653 16535437 16535778 16536212 16536896 16506716 16505755 16317854 16332104 16536864 16238719 16317732 16505883 16483085 16606267 16608934 16608840 16608540 16606393 16606601 16608471 16608552 16608403 16609087 16606261 16608948 16608534 16608841 16609132 16608373 16226135 16536550 16536932 16536317 16536683 16530181 16524969 16536952 16536557 16557807 16536927 16535798 16536523 16536843 16558221 16536547 16536795 16536035 16536812 16536165 16536562 16536891 16535947 16535446 16534922 16526873 16536759 16536907 16535701 16536914 16536893 16536716 16536880 16536472 16536791 16536920 16536484 16534586 16536903 16536567 16536316 16535571 16536570 16536885 16536602 16536722 16534718 16536905 16536939 16536699 16536042 16535811 16536950 16536569 16536665 16536940 16536467 16536486 16557871 16536790 16536929 16536954 16536734 16536938 16536621 16318453 16506690 16318570 16507535 16455927 16332053 16318436 16398833 16209944 16509537 16318490 16317872 16548328 16534373 16531811 16534944 16549852 16534742 16530974 16532593 16548042 16528390 16594895 16546862 16532869 16573293 16598924 16532568 16534238 16598985 16534721 16531896 16536963 16594834 16599118 16524754 16537865 16550362 16600794 16533043 16533917 16532855 16532867 16534049 16549522 16546119 16532894 16532620 16548113 16549566 16547900 16525914 16606076 16535141 16546390 16525583 16527157 16599071 16547895 16548395 16548638 16529074 16599193 16536951 16598961 16547905 16532556 16538054 16474806 16533057 16546216 16548233 16533966 16540799 16533076 16534641 16532532 16534215 16549618 16548627 16599053 16533064 16534018 16526879 16534936 16534334 16511306 16517684 16548150 16532564 16542744 16547558 16548514 16536226 16547701 16548626 16534196 16535524 16544801 16532957 16547707 16549505 16544788 16527630 16533033 16550301 16532893 16534902 16605994 16532477 16532849 16544963 16535152 16547703 16546277 16547087 16549804 16536447 16542205 16538627 16541889 16547896 16531817 16531094 16599035 16599117 16538951 16606068 16545905 16532918 16547453 16524746 16548112 16539051 16599012 16532305 16547904 16534168 16534912 16547022 16548486 16534933 16532860 16534700 16533004 16525995 16532558 16534060 16549554 16534939 16546143 16530624 16546995 16547434 16546084 16605993 16542946 16548148 16532345 16600941 16598963 16532622 16525386 16600993 16548412 16525182 16534753 16548107 16531292 16532108 16532450 16534919 16548633 16532604 16549511 16548330 16548518 16532392 16599202 16533842 16532880 16538908 16547164 16599201 16605907 16225920 16193571 16193350 16226127 16226120 16226124 16225374 16226121 16193331 16226117 16225821 16225401 16225905 16226190 16225396 16193540 16226051 16226247 16226122 16193319 16226130 16226275 16193856 16225382 16225906 16225815 16193282 16226192 16225696 16193268 16226108 16226125 16226126 16193690 16225822 16225378 16225349 16225368 16226191 16535263 16535243 16598967 16532885 16532592 16533036 16532625 16531099 16548516 16536959 16530596 16605953 16511312 16526985 16548063 16528156 16548483 16528393 16550405 16533850 16534351 16545849 16548109 16547893 16548630 16464998 16536965 16606194 16534546 16544954 16598913 16534046 16525179 16547903 16536912 16594869 16533780 16549874 16529191 16545042 16536904 16531504 16525383 16547551 16519023 16532675 16538718 16595171 16533162 16534998 16539601 16548515 16532877 16534410 16538990 16605888 16544106 16534453 16533054 16605996 16532998 16548398 16534188 16532015 16545882 16532865 16534931 16532834 16532291 16534762 16532847 16547899 16531545 16548629 16532337 16599069 16534112 16533138 16532327 16534758 16527879 16548324 16529189 16530045 16594914 16599008 16550847 16527071 16536910 16599121 16606067 16534464 16531820 16549791 16539321 16573286 16533862 16540698 16547446 16538953 16606065 16532598 16524750 16534785 16527711 16532980 16594837 16526863 16527181 16606079 16535163 16525391 16547686 16531634 16538399 16532756 16546315 16525690 16531124 16531193 16532577 16606077 16529328 16549716 16533792 16547709 16530608 16532277 16534053 16547897 16525872 16599070 16524752 16548402 16533820 16534021 16546797 16528994 16534165 16532702 16533060 16546033 16532788 16533908 16541802 16595223 16598929 16528371 16529302 16599261 16532839 16537766 16531648 16532602 16535515 16548331 16606197 16534565 16542084 16529453 16529536 16542728 16535420 16529076 16599300 16534248 16548116 16606385 16532408 16532770 16532621 16534184 16599002 16532361 16535537 16518485 16599001 16598957 16528234 16601023 16527864 16542293 16532590 16540961 16531000 16599184 16531807 16532897 16536955 16542319 16533055 16542082 16534889 16530442 16532851 16531203 16531283 16534014 16535049 16527924 16535106 16605989 16536958 16225737 16225734 16226089 16226101 16193323 16225819 16226248 16225965 16226193 16225471 16226104 16225535 16193307 16226091 16225845 16226019 16225820 16193342 16225547 16509321 16314092 16507537 16550706 16464807 16553293 16553174 16465139 16551579 16466272 16538704 16539312 16465291 16538843 16539583 16554065 16538900 16540167 16539932 16553624 16553285 16465297 16550513 16555014 16559689 16558842 16557833 16557796 16558355 16559102 16558651 16558303 16558820 16562004 16558445 16558510 16554445 16558206 16555371 16557962 16555841 16559606 16557584 16558456 16557949 16558874 16557566 16559433 16225386 16226276 16225348 16225538 16225392 16225586 16225347 16225817 16225532 16225594 16225196 16225481 16225501 16225132 16225599 16226119 16225242 16238916 16225340 16225343 16225650 16225516 16225352 16225339 16225576 16225363 16474334 16464941 16472552 16473563 16474384 16474003 16473592 16474636 16474740 16473623 16467314 16473696 16474645 16474946 16474023 16475993 16472532 16475019 16472252 16472652 16474708 16474512 16475475 16474568 16473659 16473330 16467030 16474613 16466443 16475042 16474901 16474579 16475015 16474192 16475092 16475790 16474345 16474913 16474760 16475316 16475013 16474202 16473649 16474629 16474594 16238758 16461099 16463453 16334581 16477232 16467527 16464589 16539307 16464622 16465568 16336935 16481759 16399498 16482163 16401252 16461074 16360974 16468365 16466433 16360780 16468022 16468509 16458103 16399526 16465709 16479736 16468026 16468376 16398911 16464512 16468529 16481960 16463936 16399512 16468017 16238755 16554425 16565161 16468036 16565985 16502207 16468011 16569329 16464939 16569901 16461243 16465625 16532819 16477936 16468019 16505357 16467403 16468001 16481746 16466619 16467319 16479316 16565117 16464616 16458682 16566266 16573736 16465842 16463859 16473048 16538433 16461130 16508020 16466494 16532762 16504271 16464040 16466790 16554778 16467762 16468018 16463661 16463858 16466533 16464431 16464008 16460259 16467346 16502116 16465143 16571861 16535953 16464515 16464628 16464994 16464868 16573090 16478712 16479923 16544665 16464081 16479928 16466514 16463563 16503351 16464162 16467316 16476687 16481894 16510046 16466174 16476214 16467353 16468012 16555015 16468317 16463409 16459500 16482158 16572923 16503124 16502946 16567180 16467065 16503644 16505586 16555626 16508105 16464687 16555763 16476808 16463957 16481828 16468016 16566102 16463230 16464426 16569083 16481752 16468014 16573084 16475378 16458630 16468028 16465001 16565399 16464686 16468034 16508005 16464079 16463646 16555861 16464591 16468876 16475464 16464626 16477505 16468303 16570745 16466796 16465147 16464577 16506969 16466209 16238925 16565414 16458523 16468010 16479734 16555007 16460501 16475290 16477276 16467267 16467458 16573243 16464509 16466030 16554208 16464325 16465939 16476621 16479929 16555616 16450091 16566363 16502592 16555260 16466621 16466965 16555239 16464334 16464931 16465807 16467865 16553439 16466426 16468023 16554436 16466503 16573678 16465596 16468004 16505452 16569904 16469366 16468968 16464627 16476216 16468037 16476740 16460384 16464575 16464100 16503608 16532591 16467003 16465150 16555257 16468449 16573668 16506336 16482557 16565610 16554218 16442242 16534822 16481749 16565336 16464618 16466971 16468382 16452653 16477628 16464804 16465673 16475195 16463328 16465292 16465695 16573410 16467329 16566981 16465163 16481745 16465806 16465135 16466151 16477878 16555019 16566593 16468521 16464071 16476497 16468517 16556062 16534428 16573522 16554212 16530880 16475906 16466340 16556109 16458680 16460521 16465746 16479729 16566967 16566954 16467040 16467896 16481751 16555242 16460573 16465804 16573720 16468345 16570690 16481756 16565143 16551379 16465541 16513767 16468557 16466388 16573456 16468372 16566955 16468614 16477192 16477661 16471825 16477791 16477940 16479926 16464596 16457860 16477654 16466091 16476475 16464756 16540005 16460801 16479731 16466961 16464075 16465885 16468013 16467997 16466121 16479560 16465438 16466578 16467359 16401610 16468595 16468031 16461214 16468386 16361393 16399102 16465382 16461137 16482448 16461120 16468430 16482409 16461187 16468909 16399279 16481939 16468438 16505694 16504438 16506473 16505593 16507198 16460366 16508472 16504143 16503890 16452998 16503889 16502460 16508008 16491071 16503600 16510012 16503132 16509601 16463245 16507558 16502409 16503438 16508101 16503437 16503888 16502583 16506878 16503623 16468020 16503606 16509310 16503116 16502057 16502322 16504272 16508002 16459355 16507543 16501975 16506263 16506980 16503348 16509508 16504639 16502880 16503891 16484594 16506313 16506664 16509244 16502675 16503605 16505747 16506124 16506213 16501852 16503894 16488382 16484320 16502255 16510089 16508465 16505699 16506875 16505442 16502896 16505467 16505660 16506651 16503353 16502400 16505675 16509485 16506455 16508468 16509426 16484588 16505596 16507550 16503611 16505446 16505806 16467599 16504640 16459105 16506971 16504442 16502808 16460999 16488126 16507329 16502832 16505380 16506027 16502259 16508109 16503428 16503892 16506023 16504443 16506151 16505639 16502047 16508318 16509260 16502479 16503352 16504646 16507207 16509278 16506087 16504444 16314809 16314539 16301536 16457744 16531617 16531791 16530806 16531935 16532600 16460683 16532314 16532070 16460778 16531639 16460287 16532364 16531590 16531510 16532601 16531630 16530956 16458696 16531232 16531825 16460004 16532528 16531041 16528356 16460451 16458998 16532417 16531502 16531795 16314193 16531919 16530077 16460563 16530660 16454676 16531860 16460455 16528790 16457728 16531907 16530425 16530972 16532599 16482454 16466052 16483071 16460203 16482748 16482179 16482743 16468021 16483383 16481765 16467295 16481754 16483075 16482141 16481931 16467998 16482561 16479296 16477547 16483403 16479725 16463907 16475412 16482621 16460094 16483389 16478185 16475705 16482570 16482414 16458629 16482769 16479295 16459712 16483385 16482411 16482741 16482558 16479049 16479053 16479299 16478422 16478875 16571411 16475473 16479550 16479717 16538343 16479293 16478710 16475872 16479552 16474374 16477536 16475905 16573739 16481888 16475793 16478879 16479056 16476664 16478705 16478304 16475984 16473117 16570277 16478543 16569379 16477584 16467676 16478310 16568512 16569485 16476365 16479730 16479551 16478873 16464802 16458655 16573539 16476206 16478182 16476277 16478877 16475702 16476276 16479088 16464685 16478713 16475601 16479548 16475913 16475371 16479547 16573724 16478659 16573735 16478872 16479054 16573665 16573677 16473431 16573692 16477763 16476092 16478874 16479297 16473543 16573652 16464044 16476484 16479083 16469637 16479307 16573472 16573498 16475153 16573478 16478183 16476430 16478711 16479045 16478101 16572989 16481906 16572272 16475495 16478061 16573722 16464579 16573486 16478093 16465928 16479723 16474083 16538836 16474250 16479051 16462193 16238761 16358619 16469780 16441946 16399967 16359894 16461935 16402509 16285683 16438031 16285676 16398740 16463031 16361805 16436924 16284484 16285809 16285674 16360233 16285682 16463460 16361891 16440714 16463877 16359741 16285093 16437439 16285161 16441608 16360651 16436677 16360177 16437589 16318927 16361750 16285695 16400109 16401389 16399464 16334589 16361985 16397218 16361313 16398272 16285704 16361789 16436771 16361773 16436956 16361620 16401250 16437372 16360737 16361711 16399630 16402361 16285036 16285663 16436674 16360250 16463856 16285268 16333816 16461528 16360261 16318663 16398505 16462204 16399712 16436890 16283961 16359585 16361740 16360272 16285939 16301939 16361769 16436676 16337081 16285648 16436770 16282585 16436741 16401492 16301684 16399224 16437236 16359081 16462898 16336106 16463133 16437662 16462086 16436747 16436996 16401761 16438104 16334113 16441546 16400023 16402443 16450586 16361733 16285649 16401158 16360072 16189506 16361845 16361373 16361912 16463445 16285794 16441156 16463244 16399912 16361311 16285616 16470229 16285843 16361525 16506078 16400137 16470181 16414469 16463236 16391356 16474016 16399150 16361881 16441483 16505587 16473224 16436740 16399662 16471606 16470417 16468009 16462088 16285252 16401281 16285667 16226134 16361869 16461929 16469611 16401306 16463110 16461943 16401396 16462776 16619808 16401107 16285789 16462902 16468000 16437135 16360360 16359888 16301999 16436861 16318789 16437176 16285619 16285810 16461699 16461809 16437089 16285091 16361523 16402302 16285812 16360839 16238901 16190413 16285793 16360116 16360912 16285666 16361228 16463162 16285681 16463086 16360132 16360575 16398025 16401228 16463150 16361364 16360164 16285647 16462876 16360883 16441480 16462833 16400600 16361024 16397356 16463343 16285668 16335997 16359447 16359791 16361009 16436737 16441599 16472456 16285703 16360207 16462680 16285767 16361240 16399715 16361312 16360152 16463045 16285811 16441597 16436721 16463190 16461293 16336266 16398379 16401307 16361617 16330986 16399668 16361988 16285679 16401245 16461284 16285671 16463228 16333007 16469734 16285673 16441581 16285670 16285842 16361703 16360183 16361518 16361720 16361527 16361775 16398012 16359779 16301822 16468577 16361033 16462093 16440586 16302000 16285792 16441303 16333135 16336319 16436604 16467757 16441544 16461600 16359509 16361376 16457630 16437128 16285791 16461287 16301826 16361367 16399871 16449381 16462780 16336644 16361754 16400383 16462191 16462197 16462206 16436753 16285684 16361781 16335502 16359626 16336169 16335503 16335534 16360334 16360215 16360242 16336573 16336440 16360514 16334853 16360190 16359868 16336320 16400037 16331334 16336330 16401355 16336545 16334238 16336726 16335528 16396945 16335532 16334436 16334745 16334808 16335785 16336743 16360462 16361382 16334566 16360142 16334159 16334234 16359865 16336007 16335265 16400341 16399940 16400519 16335579 16335058 16336565 16335452 16335076 16336623 16335287 16397931 16397145 16400155 16335628 16335338 16335554 16336318 16360249 16334112 16397032 16360815 16335381 16335371 16396999 16336390 16359642 16335122 16335382 16334847 16390337 16335533 16401368 16334587 16335742 16336016 16359736 16335379 16335555 16335773 16336575 16359644 16335270 16359696 16336147 16335819 16334325 16336517 16398073 16334562 16397106 16401249 16360451 16360353 16335182 16360108 16335313 16401095 16335067 16335498 16336101 16332991 16332539 16332701 16333550 16332580 16330191 16331635 16333498 16332361 16333232 16333065 16332631 16333103 16333499 16333521 16333590 16238680 16238904 16568670 16558439 16556058 16558339 16555258 16557821 16566997 16553768 16558542 16572986 16558266 16562244 16571724 16555568 16553222 16548408 16548517 16557778 16558625 16554389 16558488 16558763 16331392 16333449 16333188 16398385 16333535 16332797 16333182 16333157 16332828 16331024 16333497 16333126 16333565 16332126 16333536 16333013 16332327 16226131 16559057 16558276 16558255 16570160 16558289 16572291 16558401 16189549 16189530 16298138 16298124 16298187 16298284 16297882 16298569 16297908 16298497 16297875 16298370 16297663 16298341 16297830 16298355 16297429 16400902 16297783 16400838 16297522 16300704 16334732 16297363 16300645 16300752 16333483 16297855 16297685 16296823 16190400 16190402 16193512 16336774 16302161 16300660 16297476 16301427 16301002 16332033 16302065 16301682 16300741 16297828 16300654 16302060 16336742 16319638 16301399 16300928 16301039 16296846 16302160 16332848 16302120 16296707 16302139 16335559 16302193 16319671 16300694 16301238 16300653 16302212 16296844 16297291 16300682 16336546 16319029 16302300 16335113 16300617 16301000 16301318 16301169 16300790 16297539 16301352 16301240 16302112 16302164 16336758 16297138 16302140 16302156 16297337 16300625 16301672 16301110 16319617 16297004 16300637 16336446 16296678 16300709 16296865 16335928 16296640 16319599 16301393 16302105 16332310 16318620 16300624 16319020 16296819 16297028 16318984 16300977 16319570 16301161 16297404 16297288 16297619 16301143 16302172 16301004 16319041 16300923 16300962 16300650 16301309 16301163 16301646 16301187 16301109 16301185 16301871 16318951 16297747 16301173 16302296 16302138 16302274 16302159 16301375 16300687 16193517 16193351 16190247 16193507 16193653 16193275 16190327 16302092 16301294 16300656 16301314 16300670 16300582 16300690 16300866 16297688 16300933 16300862 16300763 16301009 16301319 16319526 16301108 16301317 16301561 16297046 16300672 16302158 16300658 16300667 16302185 16302119 16300702 16302113 16301311 16331457 16300675 16300974 16301003 16300666 16320010 16319933 16297137 16300681 16300697 16301332 16302162 16336897 16301307 16302137 16300577 16296454 16296859 16300657 16301233 16300705 16335756 16297752 16300689 16296658 16302306 16302199 16300874 16300628 16301829 16301308 16302179 16301007 16302165 16297023 16297265 16301995 16302080 16300999 16319884 16301186 16301270 16300683 16297790 16300692 16301031 16319626 16300971 16301008 16297054 16300680 16302269 16319039 16318974 16296830 16319584 16301377 16302103 16302262 16300686 16302132 16297118 16300972 16319994 16300647 16301683 16300856 16301316 16302163 16297058 16300688 16296805 16333534 16191809 16189556 16190500 16193290 16193311 16297151 16300469 16335914 16297416 16336070 16336142 16297420 16300608 16300651 16332364 16297756 16319971 16334616 16297716 16300662 16296432 16300685 16469431 16469432 16527954 16527792 16468425 16225345 16225350 16225351 16225483 16225489 16225517 16225337 16225341 16528361 16525738 16524935 16529402 16530490 16468396 16469740 16469279 16469358 16469705 16470125 16225346 16225338 16529186 16469225 16469361 16469676 16527363 16528152 16527837 16453422 16440482 16453253 16438249 16448219 16452989 16454332 16442646 16440467 16452176 16437552 16454313 16441769 16454447 16453711 16440569 16454298 16438091 16454469 16454392 16452958 16452988 16442834 16563418 16400724 16401450 16398274 16401369 16400131 16449382 16440523 16454105 16438254 16439251 16454330 16401482 16453705 16454320 16452131 16440634 16401518 16449358 16452654 16453849 16564003 16440478 16453959 16454295 16439247 16454381 16452762 16454389 16454418 16453707 16449376 16438222 16454104 16438241 16401625 16397447 16439395 16397940 16454289 16440447 16439108 16442228 16440643 16440516 16439766 16440564 16315633 16301616 16314970 16313693 16314963 16467043 16511301 16512032 16510684 16528057 16514134 16512999 16515555 16527980 16528229 16525573 16526147 16526978 16515626 16528384 16526903 16527920 16513108 16528810 16524961 16527144 16526929 16501960 16525986 16515554 16525846 16529071 16529083 16528239 16525192 16515629 16528169 16300970 16316112 16315548 16302157 16466859 16525576 16527331 16525870 16527289 16514807 16513190 16525839 16527685 16525390 16528806 16527038 16529193 16527657 16525387 16529386 16514934 16528148 16516473 16525188 16401779 16208143 16211072 16211062 16535260 16540452 16536518 16540569 16401292 16543517 16277352 16542639 16544310 16538474 16544143 16536489 16536620 16536446 16536792 16541964 16544790 16532624 16536799 16538425 16277784 16277249 16316348 16557581 16555012 16301645 16301713 16316021 16550834 16557972 16557756 16557890 16557601 16551059 16286004 16283082 16266749 16266815 16300655 16315498 16542583 16279121 16397698 16280221 16335007 16335988 16280339 16278012 16332351 16334162 16335130 16542875 16334584 16402381 16335795 16336868 16278439 16280380 16278171 16279847 16278262 16279851 16278405 16455103 16454882 16570681 16335582 16454833 16572396 16573690 16336850 16562738 16336866 16573323 16566677 16566785 16565530 16549838 16454849 16564861 16566730 16553201 16573653 16565424 16566372 16455057 16566169 16568798 16454499 16455428 16566724 16572306 16566211 16565016 16550386 16572494 16455033 16561592 16567167 16571731 16455093 16571733 16565568 16454705 16336865 16561488 16572350 16544659 16564320 16550890 16566360 16454832 16336853 16566181 16567184 16454865 16562290 16335877 16572318 16545947 16549887 16565645 16563417 16336223 16566182 16544797 16566367 16566258 16550045 16572343 16566732 16566971 16572356 16558769 16567170 16279581 16280438 16279106 16278341 16279612 16280214 16279621 16455530 16336014 16454708 16573083 16550656 16335990 16550526 16563583 16561674 16335945 16455092 16565793 16563102 16336834 16549955 16563581 16541932 16566613 16572825 16572985 16454769 16573074 16562656 16564309 16564987 16336752 16565895 16573654 16336888 16562473 16562496 16454737 16563055 16573433 16337092 16564315 16539963 16565958 16560194 16562671 16540543 16567181 16549980 16565933 16565682 16565775 16454814 16550246 16454889 16454939 16566953 16571726 16562330 16566938 16573196 16571832 16543122 16455214 16566974 16336782 16572397 16565672 16567731 16454744 16571746 16547353 16563871 16569234 16573037 16565910 16564981 16550602 16566379 16562258 16567400 16566028 16569532 16573040 16454498 16474352 16572067 16563290 16454845 16566620 16567182 16562100 16401710 16336312 16278972 16335824 16279841 16336718 16279171 16279133 16336290 16334586 16333570 16279114 16333861 16282567 16398640 16454426 16268970 16269066 16269069 16442697 16391102 16390500 16336674 16268984 16269034 16336332 16268973 16269035 16449245 16450117 16269049 16336931 16269162 16269042 16336923 16269101 16269108 16268975 16401171 16269326 16454419 16336317 16282591 16268989 16269041 16336490 16283235 16269070 16269104 16269119 16282575 16282901 16282597 16283270 16283062 16282638 16282661 16282644 16283266 16282817 16282702 16282925 16284196 16282645 16283167 16283182 16450670 16454500 16465515 16490483 16485514 16458428 16458808 16449394 16359006 16388552 16449354 16486588 16490519 16360277 16487166 16458312 16458393 16458806 16485500 16458332 16457650 16458347 16483890 16450093 16485660 16454315 16486837 16487150 16452443 16456713 16457802 16457832 16490405 16455627 16457866 16388743 16486622 16458118 16490113 16455444 16458367 16361404 16361783 16453957 16488871 16455216 16455430 16454481 16487064 16457782 16361001 16458336 16485653 16454771 16359664 16458513 16486848 16449239 16454485 16388531 16486942 16456269 16457788 16390179 16456917 16486634 16486386 16388715 16456926 16488388 16455210 16457899 16458487 16390322 16458308 16488392 16455065 16458157 16450319 16486939 16457592 16457653 16455871 16454872 16455500 16484757 16485517 16457862 16487305 16455569 16458290 16479044 16457817 16488390 16390291 16455426 16455466 16482739 16486000 16458573 16455425 16457587 16486620 16455498 16335777 16452537 16455469 16455098 16485971 16358716 16490894 16486851 16455062 16452765 16455593 16478306 16449625 16458576 16456611 16456111 16282622 16283257 16283248 16282870 16282856 16282684 16283212 16283161 16282601 16283268 16284772 16282726 16283251 16283149 16388475 16361145 16450088 16454932 16455366 16457658 16388959 16455091 16483895 16388608 16456404 16390583 16457618 16454810 16452260 16485676 16454326 16454480 16475819 16457806 16388583 16442999 16457718 16388768 16452992 16487051 16458471 16449726 16455048 16455193 16458572 16455626 16486941 16490756 16455304 16450089 16458087 16457749 16450337 16458563 16455050 16361288 16457643 16458012 16486937 16388511 16399614 16461811 16457735 16458621 16455212 16390604 16454805 16389058 16466087 16488160 16453524 16388725 16457964 16487167 16486842 16487077 16397518 16457794 16450035 16457667 16487198 16456569 16454479 16483892 16487049 16458069 16389324 16488157 16454483 16458296 16359101 16486938 16441494 16452481 16455496 16455815 16455427 16457773 16487076 16485655 16450365 16487165 16490395 16458556 16485495 16455209 16486322 16454928 16458254 16458323 16454672 16457776 16452303 16456212 16336739 16333191 16269024 16268995 16452262 16335883 16332630 16268993 16449617 16450388 16450145 16269028 16269088 16268933 16268977 16399636 16269012 16269008 16336980 16269019 16269048 16269106 16268949 16336844 16450379 16316291 16313923 16279906 16280518 16279769 16357960 16358041 16358356 16357973 16280830 16280537 16358440 16358202 16280569 16358441 16280423 16357992 16358123 16281026 16281057 16358341 16280748 16280509 16358442 16358282 16358430 16280513 16280951 16280561 16280112 16280502 16280746 16279170 16280530 16280444 16281070 16280412 16280550 16280926 16279626 16280523 16280275 16280583 16358143 16358689 16358147 16358580 16357995 16280209 16280436 16280507 16280723 16279924 16280546 16279634 16280566 16280695 16280485 16280517 16280739 16280539 16358187 16358543 16358690 16358005 16358044 16358092 16358055 16358222 16357983 16358140 16358194 16358184 16358665 16358265 16358116 16280883 16358312 16357979 16280775 16280515 16358253 16358001 16279767 16358193 16358625 16280241 16358195 16358293 16358266 16280521 16279169 16301642 16302044 16296576 16297376 16296795 16296535 16296822 16466774 16467034 16400206 16467299 16398302 16466596 16466648 16400851 16401126 16467432 16466751 16467185 16397304 16467063 16466646 16397951 16465730 16467999 16399706 16466475 16459400 16391641 16466843 16467594 16467186 16466720 16467317 16467219 16398034 16466667 16466783 16397979 16399481 16401502 16466822 16437157 16467268 16467453 16466960 16391332 16438201 16467256 16399993 16402538 16563118 16562266 16562146 16391497 16466887 16561388 16562373 16565861 16562385 16562464 16562247 16562349 16562766 16398977 16399247 16565550 16562479 16562393 16563505 16562425 16399025 16562305 16560137 16562739 16437982 16564334 16562467 16563122 16562342 16562232 16562046 16397104 16399520 16397679 16562275 16562567 16562644 16564278 16562772 16399603 16465985 16398131 16466985 16464428 16466439 16466482 16398235 16398260 16401473 16467355 16401393 16401157 16398976 16467906 16467184 16401115 16401497 16467257 16400091 16466756 16399804 16466553 16466653 16466660 16467218 16397801 16467266 16391386 16464689 16464871 16466698 16438024 16398438 16399277 16467052 16391372 16398276 16398935 16399496 16400192 16467433 16269022 16269115 16269044 16269025 16269054 16269105 16269084 16269103 16269045 16268959 16269050 16268986 16269014 16269071 16269011 16269053 16268927 16269116 16269062 16268960 16269091 16269059 16268938 16268963 16268997 16269037 16269074 16269055 16269068 16268950 16269056 16268937 16269065 16266910 16268948 16269016 16269043 16268972 16269031 16268976 16269007 16268951 16269040 16269078 16269075 16269013 16269118 16282900 16268969 16268956 16269009 16269117 16269010 16268965 16269079 16269038 16268935 16269080 16268936 16269021 16269052 16268929 16269030 16269067 16269004 16268966 16268994 16269001 16268990 16269100 16269076 16268974 16268961 16269077 16269102 16266362 16268982 16269020 16268962 16269032 16269046 16269081 16268985 16268978 16269029 16269057 16269072 16401693 16361641 16358432 16361838 16401137 16361270 16401769 16361878 16398227 16361859 16402104 16361314 16361256 16361391 16335053 16525325 16527702 16335864 16334100 16336927 16337085 16336283 16336783 16336099 16336510 16336781 16336260 16336909 16336675 16336587 16336001 16336306 16336856 16336667 16337170 16336845 16336010 16336311 16336901 16336886 16335776 16336837 16528176 16525580 16315706 16331969 16389666 16331532 16331230 16390435 16389345 16517682 16336300 16335919 16389758 16335179 16336310 16332159 16336600 16361375 16359665 16336282 16335911 16336315 16332098 16388984 16389134 16359069 16360155 16359521 16360510 16389342 16389704 16361601 16335634 16335180 16336066 16335755 16336769 16336302 16359781 16335745 16336269 16519150 16331426 16506975 16336071 16334871 16336301 16361508 16389885 16391202 16389333 16335820 16360400 16336233 16331488 16388703 16336051 16336456 16334331 16390600 16332354 16336924 16331625 16334319 16336072 16391061 16503604 16359859 16335553 16390403 16336444 16390789 16334160 16358422 16437110 16334594 16335829 16389322 16388605 16389990 16335778 16389080 16336097 16335365 16332433 16336115 16335333 16359792 16331950 16391097 16391116 16335526 16333912 16436974 16336201 16334615 16390317 16336048 16331114 16360191 16336983 16336580 16336267 16331183 16388675 16389008 16390275 16336428 16336646 16360470 16336321 16389344 16389820 16360966 16360256 16333238 16336487 16361479 16331552 16336141 16335051 16517377 16335041 16391146 16388858 16359861 16336309 16359871 16389343 16390362 16331925 16331590 16390106 16389876 16450236 16482993 16504142 16474895 16504274 16302071 16301650 16314917 16437304 16437305 16437668 16436768 16437524 16437462 16436940 16390248 16333592 16333511 16333485 16388795 16388938 16333190 16333133 16388971 16389317 16389686 16333608 16388830 16388820 16400153 16388805 16333029 16332442 16390346 16332386 16332421 16389519 16333463 16333038 16333834 16332878 16389472 16388731 16333081 16333230 16333561 16301953 16283116 16331327 16333537 16331957 16333662 16283260 16282587 16282577 16282691 16388783 16565844 16333134 16333597 16333183 16389548 16361622 16333508 16332623 16332340 16332196 16389639 16333480 16333158 16332446 16388742 16333457 16332326 16388850 16388973 16361289 16332274 16333239 16333486 16361514 16333693 16333439 16401156 16389606 16389502 16333567 16333119 16333622 16332621 16332288 16389762 16333717 16389417 16333179 16333032 16315302 16300415 16301715 16299986 16300592 16299755 16297865 16299432 16298165 16299399 16299446 16298239 16299917 16299426 16301544 16299678 16300857 16300359 16299753 16300623 16299936 16299435 16301677 16296854 16300568 16299927 16300626 16301280 16301346 16299385 16300593 16300618 16299722 16298199 16299721 16299853 16300023 16301541 16299402 16299995 16299934 16297906 16301301 16299689 16301609 16298269 16300008 16299989 16300468 16300545 16300389 16300364 16300589 16296754 16300015 16300564 16300774 16300467 16299868 16299798 16297554 16298575 16298100 16301521 16300404 16300587 16300616 16299439 16299997 16300584 16301414 16301513 16298246 16299314 16300068 16299489 16300599 16296688 16300057 16300367 16301647 16301502 16300609 16299443 16299937 16299475 16299915 16299965 16300387 16297916 16299775 16296333 16299339 16299920 16301348 16300572 16300551 16300930 16298210 16301306 16300001 16331944 16330964 16361119 16301535 16299501 16300031 16300362 16300621 16299488 16300627 16300397 16300524 16301546 16296597 16301469 16300361 16299512 16299365 16296298 16300876 16296869 16300597 16301305 16300612 16301497 16299397 16299515 16301607 16296488 16301155 16299507 16300713 16299852 16301475 16301451 16300373 16300955 16299975 16299479 16300610 16296618 16300570 16299690 16301315 16300388 16300059 16299371 16299403 16301292 16300084 16299929 16301371 16301444 16300058 16299756 16299979 16298458 16301403 16301472 16301674 16300056 16297544 16300549 16299400 16299418 16300603 16296732 16299497 16300371 16301112 16301553 16300553 16296458 16300652 16300684 16299424 16300691 16297060 16299401 16300465 16300715 16299801 16299827 16296811 16301310 16299994 16300482 16300466 16301471 16301313 16301518 16299774 16300594 16300406 16300555 16301111 16301508 16300043 16296489 16299445 16299688 16299448 16300622 16300994 16300869 16300983 16299719 16299871 16300590 16297930 16301312 16301659 16299999 16300548 16299051 16300607 16299796 16300006 16301411 16299873 16299928 16299830 16300620 16360425 16331179 16361483 16361467 16331136 16361348 16401411 16438141 16440522 16440496 16238902 16401432 16399638 16399544 16401639 16315687 16313751 16551582 16551033 16547688 16550844 16545861 16547148 16545800 16546617 16542544 16542664 16546567 16547562 16544598 16547170 16546632 16547121 16538273 16546173 16546589 16546103 16545833 16546099 16546681 16551533 16540191 16546871 16546095 16547554 16547564 16599307 16550836 16540574 16544795 16536946 16550187 16536972 16542137 16551099 16551105 16545040 16551531 16540773 16546880 16546601 16546890 16551047 16544786 16542051 16546228 16548628 16550883 16546299 16546052 16547439 16546407 16544965 16538567 16547693 16544969 16546341 16545039 16552043 16546133 16547894 16551969 16542107 16552121 16551090 16551369 16546907 16551089 16546155 16283105 16282545 16559991 16282758 16559776 16559380 16283130 16282804 16539301 16453420 16459010 16458679 16459910 16459236 16454679 16455063 16455720 16452591 16482563 16550316 16442789 16540487 16450156 16459617 16559345 16559794 16563492 16538084 16455493 16539344 16455413 16458678 16454809 16452552 16455450 16452242 16564273 16455411 16454379 16559122 16559572 16559667 16460040 16464331 16540993 16540821 16564958 16456025 16550375 16560443 16563412 16564325 16564188 16539290 16450408 16559774 16562578 16536311 16562756 16460786 16565231 16458598 16460696 16449283 16562672 16456804 16458001 16454927 16459747 16541818 16540247 16460941 16564006 16459041 16537947 16559534 16455813 16563586 16536974 16563589 16564319 16561472 16438195 16454719 16455204 16563120 16458882 16442742 16454753 16454663 16459495 16455719 16455512 16459765 16458884 16550876 16560404 16454497 16437567 16459834 16455224 16560438 16458812 16454478 16460114 16459803 16540515 16459675 16534199 16458810 16559516 16442615 16557738 16549996 16558790 16459733 16540815 16563058 16455506 16459612 16558683 16559775 16468003 16461170 16540583 16458643 16460553 16561474 16532561 16457607 16564856 16458642 16564910 16460538 16458638 16563999 16455068 16459831 16536735 16540093 16538901 16563408 16563489 16459750 16441217 16456065 16283255 16282855 16565192 16538559 16565133 16459928 16540351 16555016 16455539 16562741 16460405 16551529 16454828 16539059 16467734 16560436 16539063 16458051 16547892 16454690 16450040 16538362 16559161 16565419 16458294 16562462 16539980 16560448 16460741 16460387 16449640 16454496 16540581 16562783 16557980 16538588 16460489 16555259 16564951 16540014 16457957 16559207 16541789 16558600 16449379 16561216 16454693 16458592 16460183 16539827 16542726 16540063 16559771 16458661 16440333 16455969 16549867 16454748 16559792 16562670 16450224 16454484 16454688 16540073 16559998 16455255 16554345 16536762 16558524 16563407 16459957 16460379 16541799 16454721 16559778 16540150 16460466 16565432 16539014 16563103 16460072 16540060 16539842 16454707 16559711 16539892 16558782 16558798 16538617 16564955 16559993 16559790 16459502 16542716 16558573 16460372 16559987 16454816 16459505 16564311 16455418 16458658 16563503 16458520 16460721 16560450 16283240 16283114 16282975 16559772 16561426 16560001 16502942 16475546 16488737 16505615 16461118 16473588 16475810 16390547 16474538 16482428 16479310 16488582 16505888 16486273 16484603 16504445 16483451 16486007 16486028 16505885 16502098 16485983 16502833 16475909 16476644 16501780 16475709 16481747 16477428 16488757 16476211 16502834 16482746 16502163 16482562 16473044 16476257 16485652 16459780 16502008 16503135 16401784 16475417 16475375 16478882 16398105 16477289 16501827 16483818 16484317 16477133 16475599 16398326 16475249 16475553 16485671 16458433 16484297 16396998 16482458 16460601 16483897 16474711 16475706 16479064 16475136 16506142 16477201 16485499 16474507 16501754 16458450 16485502 16486254 16475073 16501756 16503114 16502253 16488870 16488556 16503141 16485666 16490645 16503129 16398469 16503607 16461806 16503345 16458640 16490947 16485657 16397046 16503113 16477448 16476688 16484759 16477561 16479927 16484761 16460767 16502955 16473993 16484312 16503602 16475056 16476043 16483798 16486589 16503601 16501836 16501846 16483885 16488735 16460507 16475288 16485492 16486014 16501807 16475703 16477292 16473653 16505355 16476645 16504141 16491058 16505791 16485663 16482730 16504436 16474932 16484323 16488727 16475318 16504447 16485673 16485665 16503435 16476279 16479059 16485490 16486009 16472548 16483089 16503436 16461530 16502950 16473567 16473717 16476280 16503433 16502868 16486018 16483447 16475115 16488730 16501912 16479089 16475596 16503121 16472404 16475126 16501842 16482433 16460358 16476057 16502853 16503138 16459615 16506292 16358630 16359303 16279140 16358724 16572302 16572370 16570819 16279103 16611608 16616459 16616974 16617125 16616512 16616523 16611216 16616409 16617283 16609577 16617255 16601108 16568465 16617289 16617287 16617182 16616403 16616535 16616555 16226250 16279586 16616907 16617191 16616490 16616934 16616598 16568817 16617190 16617291 16225274 16225630 16226123 16225574 16280252 16359301 16358684 16279623 16279597 16331026 16554430 16551378 16555018 16550856 16554269 16551530 16553779 16554280 16552109 16553545 16554772 16551523 16547898 16551092 16553008 16553079 16553261 16551374 16553212 16549764 16551111 16553941 16554047 16554777 16553778 16551367 16551583 16551522 16543514 16550508 16547683 16554771 16554226 16553915 16554439 16563049 16542894 16552036 16552039 16553240 16553177 16553857 16550633 16554250 16552994 16551521 16547084 16553385 16544781 16553454 16549453 16553050 16551965 16551382 16554413 16549605 16553928 16553232 16553058 16553076 16551580 16554253 16553030 16554099 16553877 16552116 16553703 16550277 16553060 16554347 16551526 16551386 16551528 16553136 16552977 16554240 16552976 16554779 16553423 16553593 16553608 16554220 16536968 16551376 16554433 16553056 16554024 16554157 16551371 16390493 16390787 16390770 16390518 16401633 16390971 16401711 16402598 16390725 16390706 16402628 16402412 16390690 16389772 16390128 16390441 16390467 16402532 16390511 16402329 16402681 16390263 16390499 16313976 16301382 16302043 16314086 16280324 16298327 16297932 16280406 16281108 16280326 16282277 16277257 16298439 16298388 16280857 16280700 16298540 16297725 16298425 16296600 16282319 16297533 16280014 16298366 16297418 16297092 16514082 16513076 16314769 16285997 16284396 16284465 16284531 16284328 16284786 16285690 16284739 16285696 16285698 16282558 16284714 16285660 16285677 16285691 16285998 16286002 16285870 16285651 16285931 16284744 16284700 16284502 16284696 16282556 16285347 16285935 16285922 16285171 16285800 16285798 16284737 16284803 16284032 16284473 16284546 16285895 16285524 16285694 16282339 16284094 16284790 16285594 16284758 16285796 16280465 16282564 16285692 16284297 16284508 16284222 16284716 16284403 16285423 16285926 16285996 16282364 16285797 16285841 16284681 16282198 16285697 16285081 16285824 16285209 16282199 16284433 16284761 16285611 16334846 16279995 16284517 16284482 16285999 16285938 16285937 16286003 16285012 16284726 16285699 16282570 16285790 16438208 16285921 16285934 16284302 16285799 16285391 16284307 16284464 16285920 16284024 16286001 16284678 16286000 16284703 16285504 16285680 16606391 16601082 16606378 16608494 16608375 16608673 16605908 16599036 16605909 16454476 16450677 16450429 16454420 16454382 16450086 16454291 16398709 16449393 16454428 16452168 16449319 16442139 16476759 16450071 16454421 16454393 16452012 16452177 16449318 16452173 16450038 16454322 16452767 16452138 16452652 16397784 16488568 16454383 16448221 16438121 16441404 16454449 16452471 16452142 16398076 16454446 16453955 16449360 16452368 16449242 16452458 16449602 16277264 16277456 16277552 16277240 16334628 16277627 16277568 16277615 16277525 16282408 16335106 16336762 16437073 16437692 16238922 16238718 16438156 16437016 16437941 16437266 16238912 16225342 16225311 16461702 16450330 16518479 16461597 16461190 16462187 16461247 16461188 16461281 16518899 16552125 16476697 16476800 16554784 16555235 16476417 16554206 16478196 16554162 16477644 16550502 16554400 16478488 16478315 16476780 16477854 16553766 16554775 16553407 16478548 16553688 16553040 16476496 16554783 16558861 16553351 16553358 16555008 16554420 16553759 16551581 16548519 16553649 16553889 16553340 16554785 16478190 16551384 16478188 16476502 16553722 16478180 16551525 16478421 16226133 16238720 16238906 16238673 16226132 16238923 16238908 16238920 16226129 16238905 16225934 16238651 16238910 16238903 16238883 16238737 16238813 16238918 16238675 16238913 16238921 16257034 16254409 16254399 16257020 16197332 16197987 16197415 16197427 16197376 16197985 16197286 16197938 16197347 16197999 16197483 16197891 16197854 16197380 16197405 16257016 16197491 16197512 16197342 16197992 16197283 16197874 16197475 16197861 16197900 16197969 16197886 16197349 16197506 16197885 16197281 16197443 16197485 16196976 16197871 16197367 16197943 16197921 16197528 16197995 16197318 16257027 16197988 16197282 16197409 16257031 16197321 16197882 16254413 16197459 16197474 16197905 16197918 16197372 16197912 16197500 16197364 16197944 16197986 16197486 16290537 16290270 16290585 16290445 16290198 16290144 16290196 16290516 16290205 16290564 16290580 16290500 16290055 16290591 16290047 16290582 16290450 16290014 16290030 16290368 16433397 16433379 16290467 16290511 16433737 16257021 16257024 16290382 16290022 16433453 16290509 16290024 16290508 16290468 16433630 16290570 16290027 16433692 16290430 16433295 16290016 16433802 16290592 16290481 16290263 16290215 16290446 16433719 16290013 16290319 16433213 16290422 16257030 16197285 16290033 16257045 16197323 16290019 16290034 16433694 16290011 16433215 16290026 16433827 16290240 16290466 16197329 16290195 16197896 16290313 16196952 16197287 16197284 16433843 16290079 16290317 16433364 16290015 16290370 16290192 16290020 16290449 16290490 16290139 16433723 16433767 16290528 16290573 16290031 16433681 16290393 16290078 16290249 16290327 16433335 16290400 16290342 16257028 16254406 16290023 16257037 16257025 16197929 16290150 16433770 16433429 16290012 16290527 16290476 16433422 16290510 16290581 16257035 16254395 16227405 16227418 16227415 16227414 16227863 16211296 16211294 16211286 16211293 16211287 16211292 16211288 16211290 16211289 16211291 16211285 16211297 16195948 16195911 16430736 16431083 16429624 16429460 16429393 16429639 16430708 16429205 16430636 16431245 16431589 16429213 16430566 16430545 16430165 16431274 16430245 16429144 16430467 16429613 16431342 16429016 16430301 16431036 16429293 16430277 16430229 16431515 16430439 16430623 16429629 16430683 16429645 16431558 16430507 16430742 16429127 16429294 16430681 16430396 16429208 16430587 16429612 16430982 16429137 16430632 16429042 16430930 16430965 16430963 16429539 16429110 16431588 16429063 16429593 16431555 16429479 16431129 16429233 16429456 16430257 16431590 16429106 16429047 16429399 16429287 16429604 16430495 16429298 16429617 16431325 16429132 16429396 16429499 16431557 16430387 16430425 16429114 16429616 16430962 16430969 16429401 16430615 16429200 16430314 16431294 16429066 16429202 16429572 16429570 16429237 16429490 16429236 16429240 16429058 16431486 16431424 16429569 16429588 16429087 16430595 16429071 16430136 16429578 16429557 16430904 16431613 16429291 16257593 16260298 16257497 16259012 16259549 16227680 16227895 16227955 16259104 16259131 16259071 16259159 16259454 16259112 16259447 16259155 16258985 16227957 16257531 16257478 16260097 16257561 16259338 16259175 16260057 16259116 16257506 16259233 16257489 16259313 16227782 16259193 16257584 16259102 16259363 16257484 16259034 16259054 16257705 16259580 16259198 16259063 16260141 16259085 16259167 16211295 16211284 16429637 16227956 16260073 16429383 16259158 16227795 16257471 16259444 16259424 16259017 16259453 16258990 16259126 16429636 16259180 16259050 16430212 16259206 16258999 16259076 16429568 16429515 16430166 16429582 16429623 16429590 16429605 16430190 16429508 16400309 16399079 16474012 16399103 16277887 16398301 16398268 16484611 16484356 16483887 16482834 16278416 16482788 16485669 16484292 16461063 16482736 16482751 16279618 16278106 16277558 16278335 16278254 16279605 16278127 16277924 16279585 16463680 16482737 16484760 16482998 16485498 16483013 16483091 16468015 16473982 16483820 16473573 16476624 16458893 16484319 16478267 16483111 16483796 16483096 16478426 16606000 16546108 16482744 16609829 16483394 16279764 16278221 16277486 16279631 16278904 16279571 16278782 16279793 16278183 16279609 16278889 16483080 16484764 16475313 16482755 16481949 16476389 16599304 16482560 16483782 16472739 16609653 16486005 16599302 16476376 16485512 16484595 16484762 16479559 16476495 16485507 16476859 16479555 16542064 16483884 16600700 16606373 16479721 16472661 16481923 16600665 16477556 16476278 16476492 16401397 16483349 16399641 16468496 16401074 16398680 16279087 16482605 16479557 16484608 16399289 16399533 16474212 16399245 16400077 16277715 16483353 16398483 16483122 16399879 16481954 16484596 16484758 16482421 16401152 16399590 16315264 16314031 16619052 16618920 16619398 16617367 16618773 16617600 16534701 16619656 16619483 16619910 16619726 16617475 16619489 16617785 16616402 16617629 16620040 16617536 16617719 16619069 16534340 16617486 16619821 16619864 16619230 16618649 16617597 16619075 16617313 16617472 16620241 16619686 16619906 16617601 16618538 16617626 16617786 16619068 16619333 16618540 16618746 16524984 16617628 16618789 16617370 16619681 16619306 16618718 16618664 16618537 16619047 16619036 16618849 16618655 16618982 16618935 16619812 16620207 16618683 16619663 16619580 16619478 16617711 16534622 16619893 16619987 16619464 16617478 16617578 16617532 16619418 16619639 16619938 16620047 16619949 16618536 16619962 16618996 16617186 16618961 16618546 16618643 16620031 16616477 16616930 16617887 16617371 16619472 16618541 16534779 16401731 16401754 16398015 16403968 16399969 16397355 16397894 16397519 16402641 16397777 16397489 16399604 16401114 16400536 16401777 16399838 16398102 16402665 16401915 16399524 16397011 16402506 16401372 16401782 16401323 16400207 16397602 16399963 16401741 16402614 16400356 16401256 16399225 16398913 16402413 16397732 16399891 16399680 16398088 16401712 16397806 16399707 16401664 16398011 16399935 16397376 16399601 16400142 16403999 16399936 16399527 16452304 16400119 16401341 16399710 16402192 16397435 16397378 16399905 16399029 16397219 16398832 16400272 16400491 16400372 16402531 16401966 16403966 16399149 16400122 16452352 16401775 16399585 16400270 16398440 16399582 16285795 16298422 16297453 16298274 16297801 16297847 16297740 16297384 16414675 16331947 16331576 16297239 16298163 16297593 16285669 16285808 16297644 16285700 16298404 16314938 16315047 16315605 16280015 16280498 16280524 16280434 16280500 16398026 16280042 16280019 16278506 16298244 16298478 16298222 16297925 16298111 16298305 16402626 16401713 16401120 16400442 16400079 16401883 16400611 16400771 16298306 16298209 16298292 16297902 16298399 16401342 16400812 16402350 16399814 16397855 16297327 16297775 16298171 16398424 16398516 16400012 16402559 16290424 16290374 16290032 16290489 16290021 16290077 16290448 16290469 16290018 16290029 16290194 16290168 16290369 16290405 16290367 16290447 16290507 16290495 16290421 16290222 16290389 16290256 16290337 16290229 16290028 16290017 16400896 16398597 16402433 16401143 16361262 16286581 16430177 15651629 15545133 15545224 15776088 15776052 15545145 15776002 15545258 15775999 15776053 15776056 15775996 15545159 15775990 15776070 15546837 15776104 15545275 15545215 15775983 15775997 15584267 15776096 15897960 15545270 15545278 15545201 15775969 15775994 15776118 15545011 15776023 15776029 15776031 15544903 15776067 15545362 15898509 15776054 15545337 15546840 15544918 15545188 15545326 15776039 16101862 15775973 15776022 15886679 15544957 15545067 16606236 16617838 16617101 16617929 16617373 16617594 16617078 16617024 16617525 16617883 16616942 16617292 16606313 16606455 16617279 16617329 16606414 16617402 16617414 16617651 16617884 16617760 16617748 16617931 16617140 16617251 16606548 16617678 16617200 16606583 16606567 16618484 16606249 16617317 16606305 16617580 16606401 16617381 16617481 16622547 16621130 16619073 16622007 16621039 16619748 16622713 16621810 16621842 16622328 16622646 16620078 16622539 16622410 16622144 16623523 16620740 16623516 16623603 16621019 16622209 16623037 16621742 16620573 16621131 16622380 16621859 16622418 16620323 16622415 16622968 16621649 16622098 16619190 16621755 16622360 16622213 16616445 16622364 16621233 16621971 16622655 16622443 16622302 16622477 16609000 16622670 16623797 16622355 16616406 16624000 16622762 16624020 16623771 16611204 16623776 16619439 16622709 16620579 16622961 16623757 16623693 16624179 16624181 16621643 16623690 16620858 16624180 16624175 16623469 16619623 16622694 16624167 16624651 16620289 16624011 16623609 16619067 16623527 16623602 16625314 16622745 16622172 16619695 16622690 16624442 16618938 16621717 16620043 16620905 16611550 16623519 16617872 16624164 16622704 16625318 16625317 16620021 16623798 16617778 16623765 16620235 16617454 16624484 16624022 16622005 16622960 16624591 16623475 16624668 16622739 16624135 16623953 16623697 16619943 16606272 16624134 16624100 16606358 16621252 16622959 16623605 16622719 16622974 16618751 16619747 16616951 16617498 16606351 16607969 16331434 16297347 16335932 16318855 16334601 16334143 16334546 16333734 16335010 16333258 16334963 16333838 16334344 16335110 16333058 16334342 16334800 16334731 16335139 16334859 16333833 16334270 16334036 16334429 16335282 16334844 16333965 16332057 16334733 16334744 16334012 16333648 16335288 16334569 16333946 16335524 16333096 16335378 16331146 16334163 16333866 16334271 16333851 16333998 16333954 16334183 16331891 16331964 16335014 16334472 16333725 16334870 16334339 16334845 16333478 16334799 16334837 16334956 16334727 16335035 16331028 16334291 16334156 16334115 16334605 16334714 16315437 16334024 16316950 16317452 16317588 16334722 16334787 16334031 16317447 16333669 16333730 16333804 16317047 16333661 16317127 16317179 16334061 16317515 16334403 16317186 16316126 16332138 16334730 16333229 16334187 16317839 16334034 16333660 16317094 16317161 16317466 16335163 16333817 16333863 16335046 16333782 16333062 16333911 16317701 16334907 16332569 16333703 16334302 16316095 16334062 16334769 16334158 16302401 16317060 16334729 16316937 16317689 16317656 16333913 16317157 16334466 16315576 16334107 16334064 16334148 16315619 16333476 16317816 16332306 16334992 16317475 16334488 16333586 16334834 16316334 16334674 16333741 16317191 16333836 16302677 16316012 16335021 16334175 16333231 16317198 16317145 16316997 16333942 16334037 16333656 16334565 16333961 16334231 16334892 16334842 16317181 16317644 16317151 16314836 16317674 16334437 16334823 16333962 16334583 16317534 16334588 16317493 16317615 16334728 16332516 16333259 16333944 16316974 16333787 16333160 16334493 16333632 16333739 16333159 16335311 16333617 16317433 16333793 16316986 16333627 16333472 16334161 16334872 16334288 16331010 16333107 16334032 16334873 16302678 16317600 16335074 16317553 16334109 16317136 16315471 16334365 16333176 16334683 16333566 16333173 16315983 16332012 16332752 16467282 16401263 16332548 16333084 16361420 16359780 16449574 16397920 16441159 16466833 16466688 16536698 16462205 16398393 16398139 16475991 16510301 16332804 16399659 16277907 16536445 16452178 16550773 16536285 16478657 16332184 16333189 16398117 16313713 16277333 16361520 16361000 16361129 16361634 16361647 16450097 16398555 16361637 16277936 16390135 16401951 16390708 16269026 16361457 16297796 16359299 16391476 16331116 16397509 16536888 16398678 16401395 16331121 16477844 16330865 16318896 16298413 16278548 16535826 16361120 16391617 16360211 16401614 16361128 16398711 16361214 16361378 16399746 16332754 16467156 16536527 16398571 16361025 16360055 16330907 16359785 16361729 16315468 16332622 16361207 16282844 16332748 16400817 16440704 16397664 16330823 16401666 16391466 16401776 16479712 16330840 16332368 16318483 16450350 16390689 16438144 16441611 16361359 16360982 16361461 16390137 16398344 16361300 16361397 16391627 16360106 16440521 16360956 16331330 16390644 16455412 16361168 16361862 16361031 16318529 16282582 16282592 16284113 16283259 16315003 16314003 16315967 16314500 16301816 16301740 16300649 16314244 16314528 16315287 16301717 16314852 16316966 16314078 16302084 16301680 16301934 16314263 16314800 16314016 16315670 16315453 16314665 16302054 16302104 16314866 16316330 16315749 16302026 16302129 16314526 16315442 16314449 16313963 16314780 16316036 16315675 16313991 16314792 16314822 16314065 16330995 16314755 16314534 16360757 16519395 16400527 16519396 16519398 16454299 16361143 16519397 16625916 16400078 16571389 16519394 16238623 16238681 16238652 16238757 16238736 16238655 16225610 16238653 16226118 16238756 16238622 16238914 16238669 16238909 16226128 16238915 16238619 16238618 16238899 16226261 16238620 16238621 16238733 16238907 16238625 16616968 16619597 16193801 16430414 16622060 16457995 16623796 16543480 16501009 16536849 16622676 16508007 16568713 16561438 16549574 16450154 16536852 16624649 16542873 16453851 16506876 16536850 16536848 16536851 16543460 16487083 16627365 16625452 16627660 16628498 16628296 16627880 16628049 16627963 16627884 16626276 16625519 16628501 16626580 16628048 16628426 16627231 16628302 16628045 16625351 16627255 16626774 16627959 16628311 16628042 16627672 16628305 16627668 16627367 16627246 16626602 16626708 16626605 16627962 16628711 16629010 16628629 16628528 16628639 16628926 16629056 16628920 16628717 16628712 16628531 16628527 16629011 16628924 16629012 16628997 16628699 16628998 16628523 16628698 16628701 16628633 16628635 16628974 16628643 16628718 16628700 16629052 16628919 16631252 16629626 16629641 16630142 16629640 16631963 16629483 16631629 16629440 16630354 16629796 16630129 16631641 16631786 16629814 16629793 16630340 16631507 16630391 16629633 16631779 16629630 16629629 16630369 16630317 16630345 16631777 16629635 16631632 16630117 16630322 16629758 16629739 16630500 16631787 16629475 16629486 16631258 16629789 16630067 16631132 16631784 16631245 16629459 16631628 16631505 16631080 16631300 16630522 16629472 16629489 16630285 16631979 16631292 16629638 16631774 16631778 16630104 16631635 16631098 16630143 16629632 16629753 16629826 16629478 16629836 16630374 16630438 16631516 16629642 16631783 16631512 16629631 16629634 16629627 16630313 16629474 16629733 16631269 16629481 16631630 16629473 16631265 16631782 16631073 16630412 16631978 16630076 16631297 16629460 16629461 16631517 16631634 16631105 16631123 16629628 16629809 16629477 16629805 16630144 16629616 16631074 16260488 16260482 16562750 16257576 16260122 16259329 16563993 16260498 16259016 16260377 16259003 16210787 16259375 16260072 16209994 16259434 16260177 16562548 16259582 16233345 16227912 16260366 16563288 16260369 16260063 16562743 16563851 16227989 16259586 16260306 16259520 16561302 16259479 16259581 16563857 16562746 16260227 16562753 16563862 16259186 16260082 16257501 16260076 16429579 16259478 16429651 16209931 16257522 16257511 16260241 16259358 16259271 16259487 16260103 16260372 16259507 16260090 16233354 16260507 16430590 16257713 16260506 16208121 16260517 16257517 16260483 16257706 16260098 16259640 16430479 16210452 16259596 16260045 16259595 16259437 16260257 16430383 16260508 16259613 16429239 16260230 16260070 16431415 16625609 16626575 16629457 16628526 16626355 16624547 16627889 16625724 16625371 16625779 16628631 16624697 16627227 16626726 16627353 16625471 16625919 16627957 16626733 16625391 16627251 16623760 16627886 16626666 16625225 16628421 16627232 16622411 16626664 16627245 16626383 16626391 16627665 16622734 16627240 16627844 16625648 16628529 16625755 16628173 16624170 16626655 16628500 16628310 16625372 16624122 16626268 16625878 16616401 16626650 16626675 16625543 16625284 16626653 16617130 16626671 16611557 16616451 16625494 16626764 16627211 16616949 16627221 16625581 16626777 16625381 16617049 16625899 16627000 16624550 16625369 16626600 16625790 16623896 16625676 16616950 16625451 16624117 16625368 16620591 16616904 16611112 16611456 16616915 16626233 16624572 16626306 16616428 16623767 16611417 16626527 16626252 16616975 16611756 16622081 16626684 16625370 16627237 16624670 16625397 16611183 16611111 16611325 16625281 16611246 16611228 16626701 16626718 16627357 16617054 16625224 16625508 16626132 16626820 16611429 16624505 16626661 16627360 16617632 16606844 16606344 16617878 16607920 16616946 16617047 16617880 16606291 16617782 16606319 16607891 16617399 16617664 16617605 16606611 16617614 16607948 16617877 16617662 16606599 16606820 16617017 16617849 16617300 16616918 16606283 16617132 16606438 16617787 16606806 16617625 16606594 16606816 16608025 16617066 16617164 16606616 16617767 16617519 16617119 16606621 16617853 16606300 16606696 16617053 16617231 16608010 16606398 16632029 16632030 16632031 16631999 16632012 16631136 16569577 16571994 16566246 16566222 16565149 16573035 16601175 16566617 16565808 16566939 16566624 16569393 16569376 16567409 16563117 16572394 16566391 16572181 16599303 16566384 16572871 16565627 16600619 16572872 16573462 16563413 16564323 16513333 16566598 16562607 16566009 16566382 16563977 16565384 16566257 16563588 16514338 16565175 16563495 16565252 16567418 16566678 16563273 16565756 16566389 16568986 16566605 16514344 16573086 16514432 16564185 16571605 16564113 16562594 16565883 16563499 16571447 16566608 16570275 16572167 16563853 16556116 16563584 16564333 16563297 16564963 16556293 16556299 16564118 16556060 16565953 16557558 16556111 16555717 16555374 16555553 16557594 16554774 16555013 16314631 16557726 16556290 16556294 16555010 16555714 16556114 16557580 16556103 16557590 16564102 16563051 16510008 16506293 16504257 16508313 16598815 16510151 16477888 16506914 16599308 16507540 16599014 16506371 16502223 16600505 16510070 16507554 16599119 16601367 16507208 16503614 16506917 16515396 16508017 16476875 16599166 16600681 16600649 16502449 16508116 16509607 16502657 16515631 16478303 16515547 16506620 16508319 16508123 16507221 16507206 16507220 16600633 16599056 16509675 16599306 16507531 16506466 16477417 16503878 16504435 16599068 16599013 16599188 16599266 16516505 16601154 16508112 16600498 16515264 16511752 16477570 16314666 16536926 16536579 16538023 16513157 16514143 16513976 16513175 16514439 16514440 16484599 16514936 16517072 16516464 16504643 16513902 16516434 16517482 16516767 16517189 16517469 16482172 16516663 16515633 16514017 16514793 16513045 16513320 16513751 16503432 16513914 16514149 16514427 16514127 16510118 16514940 16513990 16513764 16513481 16514007 16512047 16513489 16516488 16516893 16515391 16516597 16516409 16515636 16517467 16506977 16517279 16517559 16515635 16517497 16516786 16484592 16516904 16517466 16516931 16490060 16518684 16473225 16477957 16477257 16489551 16511574 16487153 16477208 16489907 16511888 16439061 16518016 16477770 16440368 16486286 16502794 16461045 16486267 16490369 16477179 16512019 16518013 16477443 16437406 16440710 16489528 16510375 16460783 16511322 16483382 16440351 16440306 16486300 16440327 16477502 16515630 16437961 16502717 16486603 16504259 16438980 16476498 16507330 16511314 16486278 16511757 16519024 16486264 16437286 16437591 16477524 16485662 16512018 16510686 16486614 16440314 16476838 16510500 16486846 16511914 16459309 16440489 16515323 16513059 16438131 16485996 16517522 16489740 16486012 16439192 16477579 16440561 16440453 16513272 16518323 16515335 16516696 16440343 16440419 16485981 16517487 16519141 16486293 16486825 16484763 16506170 16440394 16486259 16440364 16437634 16504146 16438112 16439086 16437451 16490493 16513267 16440385 16503434 16511750 16486835 16511557 16513000 16508316 16511862 16510678 16440683 16486639 16477410 16477638 16513611 16478164 16486325 16477782 16513909 16477610 16489554 16517678 16486599 16511572 16315640 16313956 16316169 16301824 16302270 16302402 16302556 16302304 16302605 16302494 16302268 16302278 16302671 16302673 16302335 16302308 16302031 16302336 16302310 16302608 16302167 16302107 16302302 16302325 16302334 16302668 16332690 16331496 16332458 16302256 16302667 16302555 16302289 16302066 16302192 16302675 16302672 16302376 16278754 16302646 16302173 16302101 16302403 16302188 16302606 16278447 16302077 16302559 16302373 16302324 16278348 16302130 16301943 16278536 16269063 16302370 16302314 16302642 16302496 16331903 16414649 16331869 16332850 16331898 16563292 16501881 16501815 16501855 16459155 16314904 16301818 16315582 16599160 16391170 16389002 16391163 16400004 16562762 16563843 16562635 16564967 16391081 16390506 16391128 16331489 16399934 16401440 16401714 16536545 16398490 16536575 16530677 16518812 16536626 16536454 16401139 16402270 16529079 16534908 16536423 16536315 16535247 16397834 16573723 16401138 16531632 16536572 16570278 16536711 16572835 16536689 16535810 16530355 16573651 16398600 16535849 16536532 16573270 16569308 16532321 16536687 16401394 16570161 16536641 16536627 16536512 16536284 16536509 16536642 16402494 16573085 16536702 16536720 16573213 16401261 16573510 16536312 16535761 16535568 16534709 16536782 16536696 16570916 16536646 16401242 16573650 16536514 16573332 16572216 16531109 16536279 16536564 16401668 16536577 16531508 16536636 16536452 16536886 16536622 16536511 16535564 16402115 16536487 16534929 16399081 16535806 16535807 16536524 16536469 16301467 16301610 16314301 16313825 16535251 16535245 16535329 16535277 16535288 16535246 16535682 16535976 16536617 16530083 16536115 16535755 16534725 16536040 16534590 16536494 16534699 16535552 16534584 16531412 16535818 16525384 16536013 16534606 16525993 16534624 16536522 16535851 16536654 16535813 16534409 16534570 16536457 16536277 16536239 16536538 16534571 16535455 16536635 16536183 16534732 16535175 16400056 16455240 16360820 16361910 16440733 16440750 16461192 16440650 16361234 16360387 16335648 16461100 16463691 16335950 16460592 16461013 16460928 16335862 16333760 16391173 16461145 16335849 16458657 16462804 16440732 16461213 16361762 16335372 16335787 16440755 16335531 16360682 16331413 16331328 16460751 16358924 16390302 16440751 16331503 16440756 16441131 16463499 16440753 16360307 16360549 16440741 16461594 16361381 16360276 16464333 16441029 16441006 16441109 16440743 16460955 16461147 16336166 16360253 16441060 16441114 16440740 16441018 16333910 16331172 16334114 16440964 16360932 16441051 16360269 16441287 16360378 16331638 16441194 16440623 16438041 16360223 16440748 16360903 16440687 16441213 16558365 16534616 16535311 16532355 16536466 16534807 16534599 16536525 16535319 16535249 16535237 16535872 16535171 16534073 16535929 16534608 16534646 16536844 16536488 16535554 16536018 16536299 16536502 16534530 16535559 16534887 16535463 16535254 16535297 16535252 16534621 16535612 16536501 16534252 16536783 16536746 16535577 16536271 16535808 16535957 16535726 16525171 16535130 16536478 16536485 16534107 16524722 16536638 16535476 16536200 16390419 16461034 16336581 16361374 16441061 16463524 16360221 16336065 16460987 16336835 16389466 16440754 16360498 16336721 16334438 16335639 16391489 16361271 16334755 16361146 16441062 16332016 16360275 16460821 16333835 16440692 16361369 16336150 16360940 16360812 16440749 16360393 16334236 16461117 16335775 16360087 16332543 16358431 16336178 16335045 16332695 16441171 16441040 16360203 16360440 16390589 16441128 16336567 16390707 16440742 16334439 16464052 16360366 16441466 16335075 16441216 16336299 16397731 16361383 16440724 16360148 16360529 16441215 16440744 16335905 16440747 16334576 16334435 16360098 16440757 16399649 16361926 16336392 16441245 16441054 16359321 16361272 16440734 16360761 16461121 16334573 16359540 16441185 16441122 16195237 16283108 16282677 16282675 16282803 16283262 16284178 16283127 16283050 16282781 16282676 16283106 16283159 16282672 16284176 16283264 16282565 16283253 16226116 16225818 16226156 16283242 16282899 16282757 16283044 16282759 16282816 16282602 16283243 16284338 16282685 16282690 16282985 16284137 16283926 16300065 16535585 16532270 16540618 16540033 16533015 16403988 16540707 16319795 16536898 16537902 16536473 16300005 16536981 16541817 16536507 16535745 16540779 16540989 16538920 16536221 16538016 16300408 16540211 16532378 16536915 16546649 16538005 16277364 16534540 16540893 16542305 16530133 16546831 16539630 16516653 16536964 16300090 16300075 16542712 16319652 16320124 16277242 16536329 16300429 16300409 16319807 16547007 16299983 16536460 16319645 16536294 16535904 16536937 16277282 16277502 16534609 16536033 16536041 16277963 16320134 16542232 16536879 16540356 16277292 16541804 16542661 16540786 16318724 16535827 16535238 16535305 16238926 16282897 16284672 16282873 16282682 16282598 16282829 16283090 16283271 16284208 16282681 16540390 16542191 16536458 16537799 16320118 16547096 16542213 16536966 16536039 16536618 16540813 16536260 16534649 16536794 16538375 16277271 16319810 16538424 16536500 16536973 16542669 16535486 16319531 16547902 16536319 16537810 16539992 16540549 16539800 16277262 16277335 16277212 16542977 16533058 16300039 16538329 16277671 16536001 16536928 16538594 16542682 16320141 16537966 16536038 16540087 16535562 16516412 16536034 16536223 16537950 16536225 16535812 16300011 16277311 16277329 16319789 16536953 16542023 16542904 16277579 16277419 16534803 16541814 16300029 16539347 16542331 16319717 16538899 16536228 16320114 16300390 16545916 16538692 16535986 16544959 16536900 16534650 16540377 16319823 16300358 16319745 16320049 16320149 16540829 16540519 16535586 16537914 16536320 16536463 16536321 16533059 16542184 16538470 16542005 16536146 16535557 16536455 16541797 16536037 16277980 16282788 16283174 16284765 16284219 16315046 16313742 16301606 16302002 16314659 16210042 16209917 16208132 16210430 16210417 16195515 16195490 16210431 16210526 16208138 16210053 16210898 16199925 16195424 16210547 16209927 16210886 16210418 16210007 16208205 16207650 16211097 16209936 16208059 16210380 16210443 16209947 16210017 16207632 16195349 16211080 16210437 16210442 16208208 16209933 16209914 16195402 16209919 16210056 16209938 16195260 16210502 16210140 16207636 16210446 16210441 16209943 16193666 16208048 16208206 16210091 16208070 16211052 16210455 16211165 16211082 16210081 16210422 16210541 16210487 16195225 16209948 16209922 16207652 16208155 16210434 16209912 16210147 16210064 16208160 16191873 16208209 16208204 16209942 16210120 16195403 16210416 16211101 16208199 16210566 16210438 16210439 16210004 16208170 16195386 16209925 16210578 16210144 16210620 16211057 16211053 16209916 16210415 16208092 16208097 16209928 16210427 16208103 16210414 16470833 16402575 16490938 16491065 16463981 16510481 16488958 16486627 16506031 16471459 16472473 16471225 16468886 16489631 16398429 16486607 16491227 16472657 16490861 16488963 16510228 16468868 16488966 16489499 16508474 16488563 16468764 16475500 16509527 16510191 16511570 16471715 16471952 16510873 16397146 16470295 16470934 16489670 16508315 16397945 16469584 16489561 16506577 16470911 16469901 16490430 16469299 16483099 16469459 16469677 16506733 16506970 16471456 16490322 16469198 16510672 16469640 16490602 16469581 16490942 16468356 16488144 16510874 16510869 16471816 16471524 16504262 16506584 16468005 16468542 16489406 16490536 16469474 16510696 16471727 16488159 16506140 16471323 16468486 16469552 16472169 16485976 16511154 16489448 16489068 16510870 16390786 16490209 16490390 16506827 16469180 16471729 16489600 16509435 16488718 16489602 16402364 16469511 16510338 16511150 16506320 16468443 16490164 16507546 16398868 16472542 16469437 16506953 16510424 16457759 16470938 16486047 16468464 16471533 16509218 16470973 16471877 16510668 16470286 16488872 16509693 16469016 16470283 16468329 16471934 16511561 16470940 16508011 16397598 16469094 16489691 16469164 16470080 16489509 16510004 16510876 16489772 16506300 16468033 16489850 16490260 16504434 16490750 16468599 16488553 16503636 16490053 16398082 16510385 16490724 16504137 16489788 16469033 16472395 16468029 16489525 16489655 16510515 16470476 16471162 16509564 16490763 16472824 16468032 16472659 16489474 16470955 16510025 16490906 16490733 16509268 16471243 16469680 16487155 16471078 16491176 16398869 16468856 16402241 16482724 16510463 16471721 16488395 16469004 16471228 16470474 16471590 16316230 16314510 16315484 16302088 16210021 16210510 16208126 16209915 16210451 16194878 16211054 16208198 16210995 16208178 16208210 16195406 16207966 16210392 16207642 16210425 16208203 16208116 16195245 16207637 16210444 16210445 16208165 16199879 16209932 16209939 16208215 16210453 16209978 16210436 16209989 16210603 16199944 16208213 16207653 16209960 16210449 16208113 16210887 16209940 16210074 16194880 16210428 16210421 16207646 16210233 16209929 16210264 16210426 16208214 16209937 16210572 16199829 16210256 16210533 16210742 16210450 16195502 16210521 16208082 16207907 16210447 16209913 16195397 16209945 16210454 16209926 16210839 16209941 16207648 16207673 16208150 16210423 16210550 16210419 16209935 16194879 16209930 16207672 16210497 16209946 16209911 16195218 16210066 16208200 16210429 16210424 16207886 16209934 16194860 16210432 16208212 16471873 16502321 16471870 16474726 16504652 16390788 16401359 16507562 16511915 16469506 16506166 16510308 16471098 16506342 16471753 16490048 16400158 16441427 16487196 16468027 16471991 16510163 16506638 16471596 16468025 16489892 16489537 16510302 16471261 16476210 16398790 16469063 16489104 16510867 16470495 16471249 16488866 16469613 16490313 16509639 16511148 16472400 16489637 16506555 16511156 16472163 16509572 16507541 16469360 16478318 16503885 16468602 16511398 16511760 16489105 16471477 16488551 16488579 16506728 16470922 16511566 16391307 16471485 16489664 16491215 16399726 16471738 16468578 16461146 16461171 16484581 16470905 16563587 16472008 16485491 16472654 16504269 16489837 16510872 16439442 16502589 16507565 16472531 16489678 16468901 16471511 16461173 16469408 16469704 16508462 16488873 16503053 16507204 16490469 16508124 16471235 16469429 16506673 16485692 16491208 16508119 16489879 16471066 16490137 16490631 16207629 16469583 16471139 16501955 16510868 16506610 16470252 16472250 16509205 16511143 16397601 16468779 16472001 16469514 16468337 16468418 16475668 16489070 16399201 16469374 16488952 16489531 16489547 16489757 16490501 16490656 16469638 16489682 16471118 16511310 16401825 16468035 16478704 16391193 16468590 16469397 16490232 16490912 16491210 16489463 16469046 16402181 16469423 16472053 16489816 16510679 16483378 16471979 16400459 16469359 16489452 16469106 16486582 16502603 16506880 16468030 16508317 16472989 16469585 16398933 16397303 16471220 16488869 16509547 16468745 16469235 16469898 16471453 16469489 16361528 16450139 16336477 16361545 16361371 16336368 16336720 16331097 16336218 16361465 16336421 16361151 16359666 16359784 16331133 16359797 16361925 16361411 16335804 16331940 16361484 16268971 16269327 16269017 16268939 16268957 16268928 16269107 16269047 16268944 16269085 16268955 16269272 16269036 16268931 16269002 16534459 16534604 16269064 16269288 16268945 16268926 16267062 16525382 16529509 16530560 16510669 16538277 16540154 16538057 16539549 16510403 16529452 16510133 16529556 16331085 16333016 16509622 16511311 16529642 16536909 16537844 16536980 16331118 16533957 16529394 16401413 16539174 16399583 16529390 16536931 16397085 16529531 16533030 16529686 16540818 16330842 16539668 16542784 16529490 16540184 16538265 16529299 16330195 16330833 16538300 16538385 16330836 16534219 16333106 16540722 16414468 16534345 16530892 16528616 16538775 16528346 16538018 16318632 16539105 16529666 16531038 16540264 16518685 16536899 16537925 16511146 16532681 16539294 16331110 16332795 16510877 16530601 16539914 16539340 16532758 16534569 16333068 16530520 16529353 16538365 16537742 16538690 16536960 16538958 16536945 16539118 16540359 16333044 16482616 16530148 16531044 16532684 16531604 16333050 16538769 16536882 16536922 16477820 16539107 16534602 16536948 16330975 16478740 16330951 16538572 16534589 16536767 16539298 16332872 16318747 16529308 16528606 16529399 16524967 16531840 16536969 16502084 16317830 16536573 16302313 16485493 16529535 16536977 16538857 16530375 16532679 16536983 16511553 16533056 16531032 16536917 16330183 16525186 16530497 16533861 16527770 16511559 16333000 16318758 16536291 16330849 16330175 16398245 16333069 16478314 16538027 16534009 16391379 16531342 16529451 16530192 16331102 16530091 16530150 16534160 16539646 16529436 16536971 16532682 16318940 16478307 16331213 16539315 16330162 16318650 16531514 16536943 16333010 16538006 16529520 16331089 16536918 16538612 16333187 16538935 16538846 16333051 16530365 16530446 16332858 16524732 16330864 16536906 16534177 16540353 16540372 16331468 16478549 16528392 16529196 16330913 16530405 16531519 16531947 16478707 16534187 16537917 16332753 16333066 16510875 16530141 16540502 16333072 16538584 16532768 16537980 16540550 16333121 16541012 16525175 16332875 16529201 16534381 16332868 16530476 16532829 16527159 16529597 16538033 16538553 16529510 16225998 16269126 16238917 16195368 16211094 16210893 16209983 16531636 16551578 16549680 16550258 16551094 16544971 16561255 16550294 16551054 16548040 16547112 16238900 16238674 16238682 16269005 16266392 16269061 16268979 16268934 16269058 16268981 16268925 16269086 16268958 16269090 16268987 16268967 16269089 16268943 16540468 16539277 16540537 16536901 16539331 16536892 16536956 16536897 16538335 16536881 16538353 16538975 16540179 16536982 16536933 16540990 16536919 16539691 16537815 16536895 16538970 16540917 16536925 16540343 16540582 16539287 16531981 16540183 16538777 16536978 16535846 16537808 16534789 16540290 16538340 16536976 16538671 16535318 16538652 16536975 16538454 16536913 16536962 16537961 16538732 16540691 16537995 16538999 16539569 16539882 16540342 16536930 16536556 16539677 16538416 16525743 16532078 16537894 16529444 16332624 16530401 16330881 16535116 16330967 16529335 16478186 16529712 16534502 16530838 16332987 16510017 16530145 16331399 16511151 16529527 16332279 16526959 16332995 16330177 16452116 16529305 16485496 16529396 16529456 16529086 16332330 16529533 16333122 16391633 16330953 16332059 16330981 16534613 16397289 16526931 16510468 16331016 16331995 16332994 16454429 16531141 16331115 16529199 16534446 16529487 16438226 16534814 16530156 16332952 16318769 16318882 16530329 16330920 16529317 16330184 16330940 16400089 16331156 16398638 16524962 16528973 16534770 16530373 16534568 16532844 16332869 16330905 16530587 16531214 16529455 16533053 16529313 16534552 16529518 16398457 16399931 16533854 16333104 16529641 16331918 16530866 16330879 16330868 16331370 16450401 16331938 16511575 16530984 16532696 16391625 16534179 16534539 16529521 16529330 16533052 16530385 16534760 16332796 16332691 16331119 16529457 16318505 16332870 16529253 16533840 16330884 16526866 16530138 16532863 16331955 16525702 16530271 16331166 16330924 16333156 16332637 16398477 16534479 16534391 16530540 16534429 16535026 16529432 16534951 16529450 16531096 16528811 16330936 16401103 16530139 16530690 16528977 16530616 16534123 16330875 16534348 16225816 16269073 16277345 16195299 16209918 16210433 16208037 16531166 16531536 16532779 16551087 16548114 16542990 16544473 16540590 16550865 16551524 16551097 16542552 16549495 16550870 16538955 16551091 16543243 16551532 16544958 16539103 16467494 16190208 16190261 16191054 16233351 16233352 16233341 16233346 16233356 16233355 16233343 16233347 16233358 16233353 16233340 16233348 16233342 16233357 16468007 16468002 16191018 16190297 16189561 16191079 16190251 16190243 16191064 16189544 16189525 16467868 16302015 16282659 16282650 16282643 16282641 16282625 16282665 16282647 16282560 16282599 16282624 16282613 16282673 16282660 16282600 16282588 16282614 16282594 16282948 16282615 16282620 16536526 16536492 16538429 16535241 16536422 16283032 16285927 16268968 16268988 16285932 16269120 16282708 16285936 16269082 16269318 16268964 16285596 16267086 16285933 16269018 16284105 16269000 16282607 16283221 16269051 16269039 16269027 16285678 16269281 16284747 16269015 16268952 16269087 16314014 16313730 16282617 16282201 16282532 16282568 16282547 16282606 16282542 16282549 16282379 16282618 16282550 16282584 16282551 16282409 16282405 16282555 16282533 16282637 16282370 16282623 16282662 16282195 16282544 16282574 16282562 16517834 16452449 16452101 16536902 16536908 16462816 16449356 16489714 16536753 16450134 16450155 16464038 16489863 16536944 16461697 16507997 16488376 16450455 16450318 16452264 16490294 16536957 16488950 16449735 16450453 16450073 16441698 16449489 16463604 16441334 16440758 16488397 16536949 16297154 16297306 16278253 16297530 16297472 16559791 16560403 16561322 16452364 16504439 16449391 16559984 16561224 16561640 16437297 16561518 16449668 16506978 16448218 16449787 16562485 16450052 16562113 16561753 16449725 16561295 16452058 16507560 16452318 16562140 16561766 16452106 16560453 16561276 16560206 16297704 16297037 16297385 16559995 16560143 16449361 16505374 16561573 16561553 16559299 16562077 16573530 16505716 16562089 16561625 16561528 16561727 16452139 16560222 16559265 16561250 16452149 16559777 16491068 16561931 16561739 16561706 16504446 16572251 16450101 16555825 16569324 16452300 16452133 16488722 16452343 16536970 16467592 16464166 16536935 16437484 16536967 16463779 16536936 16536916 16449744 16487161 16450412 16490183 16536750 16490459 16464433 16463387 16464588 16450387 16491222 16536961 16452231 16536979 16468006 16450084 16463727 16301605 16301555 16314906 16463420 16463538 16476056 16468008 16463746 16473435 16463771 16509459 16467351 16464054 16463183 16467566 16438058 16462805 16467402 16467406 16462685 16467502 16473978 16463631 16476762 16467434 16467929 16467303 16468024 16467597 16460975 16450294 16463738 16462694 16466437 16463170 16467913 16474340 16473387 16450795 16449230 16449255 16441574 16450491 16454310 16454450 16449313 16450456 16454380 16452441 16454317 16454387 16450435 16463338 16452063 16474090 16464104 16473114 16467456 16467448 16463901 16474218 16463383 16463583 16449589 16474788 16464074 16467450 16467759 16467404 16452266 16463198 16475785 16467866 16463674 16459760 16467217 16449286 16452136 16452292 16450245 16454384 16454297 16449265 16454372 16454371 16454368 16449812 16450039 16454328 16452090 16452371 16454307 16454451 16452038 16454390 16470292 16470391 16471044 16390482 16469904 16471055 16389783 16468570 16471803 16469520 16470416 16389987 16473113 16470419 16469902 16469550 16471046 16474081 16469785 16470393 16469906 16210159 16209920 16469706 16469735 16469788 16473714 16528623 16518481 16527252 16531311 16531996 16510667 16519045 16518265 16531658 16527892 16527727 16474358 16527095 16531786 16528622 16525740 16469737 16530465 16528624 16527369 16518264 16473047 16470251 16529343 16519047 16474188 16473001 16531595 16469743 16470077 16488867 16518473 16528802 16514643 16529454 16519143 16519105 16513151 16513354 16524965 16517008 16532279 16528809 16470082 16532333 16524979 16469643 16525988 16469702 16517543 16530857 16511748 16517853 16517179 16469615 16518683 16516444 16530830 16531962 16530822 16518756 16516623 16512718 16514160 16524755 16519146 16525388 16532278 16473386 16517267 16473226 16515390 16506188 16527218 16528970 16518988 16512720 16530154 16513778 16511157 16510509 16511159 16528777 16514659 16524973 16518995 16514802 16526928 16502606 16513417 16518475 16527801 16519015 16526923 16513288 16518675 16487212 16510368 16525741 16513634 16527575 16517212 16527328 16515394 16510318 16515550 16519016 16472533 16472748 16469612 16469781 16469678 16472743 16531799 16514651 16528795 16511883 16530899 16531602 16517634 16525990 16531549 16524976 16528162 16513905 16519054 16531789 16527187 16528380 16518789 16525911 16515312 16532778 16513900 16532582 16524741 16529067 16510676 16512723 16517472 16532551 16459981 16515331 16531578 16527378 16531191 16530646 16513300 16527814 16518468 16530971 16528620 16527604 16528974 16519009 16532115 16517577 16519067 16527356 16513098 16512716 16518992 16527169 16531238 16528172 16473437 16518471 16528376 16510689 16472746 16519012 16516579 16518267 16507561 16514436 16518811 16528626 16512717 16518015 16517539 16515392 16526191 16473317 16519081 16515265 16514791 16504256 16516725 16470127 16515266 16513572 16516985 16515393 16518322 16518483 16527344 16514937 16511298 16515317
FEI Number 2016493
Recalling Firm/
Manufacturer
CareFusion 303, Inc.
10020 Pacific Mesa Blvd
San Diego CA 92121-4386
For Additional Information ContactStephanie Chavez
858-336-8754
Manufacturer Reason
for Recall
Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.
FDA Determined
Cause 2
Process control
ActionOn 08/15/2024, the firm sent an "URGENT MEDICAL DEVICE PRODUCT CORRECTION" Letter via FedEx /2-day or email to customers informing them that, BD has identified an increase in bent contacts on the new Female/Left IUI connectors which may cause delays in initiating therapy. Customer are instructed to: 1. Per current labeling, customers should inspect all IUI connectors prior to use. You may also refer to Appendix A for further details. 2. If you identify a bent/lifted IUI pin prior to use, do not use the device and send the device to biomedical engineering or Qualified Service Personnel for repair. 3. Please distribute this notice to all facility personnel responsible for inspection of BD Alaris" devices. For questions/assistance - Technical Questions - Technical Support - Phone: 1-866-488-1408, Phone hours: 5:00am to 5:00pm PT, Monday Friday. Product Complaints - BD Designated Complaint Handling Unit - Phone: 1-888-812-3266 Phone hours: 7:00am to 5:00pm PT, Monday Friday, Email: productcomplaints@bd.com. Email for receipt of customer response form and recall related questions -
Quantity in Commerce10460
DistributionWorldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. The countries of United Arab Emirates, Australia, Belgium, Canada, Hungary, Kuwait, Qatar, Saudi Arabia, Singapore, Taiwan, and New Zealand.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = FRN
-
-