| | Class 2 Device Recall Stryker iBur hubs and cutting accessories |  |
| Date Initiated by Firm | July 24, 2024 |
| Date Posted | August 26, 2024 |
| Recall Status1 |
Completed |
| Recall Number | Z-2728-2024 |
| Recall Event ID |
95040 |
| 510(K)Number | K210377 |
| Product Classification |
Drills, burrs, trephines & accessories (simple, powered) - Product Code HBE
|
| Product | Stryker iBur 4.0mm Coarse Diamond Round, Distal Bend -
intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications
Ref: 8431-013-040DC |
| Code Information |
GTIN: 07613327501155
Lot Numbers:
22248017
22263017
22334017
22348017
22348027
23016027
23095027
23128017
23163017
23179017
23205017
23205027
23205037
23221017
23311017
24004017
24004027
24004037
|
| FEI Number |
3015967359
|
Recalling Firm/ Manufacturer |
Stryker Corporation 1941 Stryker Way Portage MI 49002-9711
|
| For Additional Information Contact | SAME 269-800-1941 |
Manufacturer Reason for Recall | Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Stryker issued Urgent Medical Device Recall letter on 7/24/424 to Materials Manager, Risk Manager, OR Director via Fed'x. Letter states reason for recall, health risk, and action to take:
1. Immediately review your inventory to locate and quarantine any affected products at your facility.
2. Return the enclosed Business Reply Form (BRF), even if the affected product is no longer in inventory, to the email Instruments.Recalls@stryker.com.
3. Upon receipt of the completed BRF, Stryker will provide a shipping label to return recalled product on-hand.
4. A credit will be provided upon receipt of the recalled product.
5. Maintain awareness of this communication internally and inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients accordingly.
If you have any questions about this matter, contact Stryker Instruments by phone (269) 568-1481 or email at Instruments.Recalls@stryker.com.
|
| Quantity in Commerce | 5031 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Australia, Canada, France, Hong Kong, India, Italy, Japan, Japan, Netherlands, South Korea, Spain, Sweden, Taiwan, United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
|
| 510(K) Database | 510(K)s with Product Code = HBE
|
|
|
|