• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall StealthStation S8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall StealthStation S8see related information
Date Initiated by FirmAugust 22, 2024
Date PostedOctober 10, 2024
Recall Status1 Open3, Classified
Recall NumberZ-0050-2025
Recall Event ID 95283
510(K)NumberK201189 
Product Classification Orthopedic stereotaxic instrument - Product Code OLO
ProductStealthStation S8 App versions 1.3.0 and 1.3.2, (Model # 9735762) The StealthStation System, with StealthStation Spine software, is intended as an aid for precisely locating anatomical structures in either open or percutaneous neurosurgical and orthopedic procedures.
Code Information Product Number: 9735762 UDI-DI code: 00763000456986 Serial Numbers: 420002290 N06805776 N07323682 N07323683 N07323687 N07326235 N07334454 N07334475 N07338479 N07338480 N07338482 N07338929 N07354639 N07354642 N07354643 N07354644 N07354645 N07355155 N07357492 N07357493 N07360505 N07360506 N07362322 N07362323 N07364218 N07364220 N07364221 N07365158 N07365160 N07365163 N07365164 N07369302 N07369307 N07401810 N07401811 N07401812 N07401813 N07407569 N07418140 N07418141 N07426193 N07427592 N07427593 N07427597 N07428130 N07428133 N07428134 N07428190 N07428192 N07428193 N07428195 N07428197 N07429001 N07436071 N07436073 N07436074 N07438127 N07438128 N07441426 N07441431 N07448170 N07448371 N07448372 N07449575 N07460062 N07460064 N07461619 N07465933 N07469095 N07469096 N07469097 N07469100 N07469789 N07470357 N07470358 N07471822 N07471823 N07475369 N07475370 N07482977 N07505369 N07508757 N07511354 N07511355 N07511359 N07516935 N07525093 N07525094 N07525099 N07529455 N07533927 N07533928 N07533929 N07533930 N07533931 N07533932 N07533933 N07534178 N07534179 N07534180 N07534655 N07534660 N07536564 N07540834 N07540836 N07540840 N07546674 N07546675 N07546676 N07546679 N07546680 N07547570 N07549763 N07549764 N07550571 N07550577 N07552479 N07565206 N07565587 N07579210 N07579211 N07579475 N07581415 N07585678 N07587632 N07587633 N07587635 N07587636 N07587637 N07587859 N07587861 N07587862 N07587863 N07590349 N07590350 N07590351 N07594577 N07594578 N07594581 N07595330 N07595342 N07596952 N07598116 N07605593 N07612697 N07618329 N07620549 N07803571 N07803573 N07803575 N07806782 N07807286 N07807287 N07807288 N07807289 N07807290 N07807291 N07807292 N07807293 N07807294 N07818055 N07818056 N07819346 N07819352 N07819353 N07822371 N07822379 N07826933 N07828615 N07844635 N07845660 N07845663 N07845671 N07858648 N07858649 N07858650 N07858651 N07858652 N07861405 N07864096 N07864097 N07864098 N07874891 N07874892 N07874895 N07874896 N07874899 N07882341 N08074937 N08074938 N08078498 N08081904 N08085721 N08085722 N08087068 N08089998 N08089999 N08091630 N08098821 N08098826 N08108825 N08108827 N08108828 N08112922 N08117758 N08117759 N08118873 N08122313 N08123244 N08127267 N08127268 N08128176 N08128180 N08128181 N08135030 N08135036 N08135049 N08135455 N08135456 N08146120 N08147667 N08147668 N08147669 N08149571 N08149572 N08149573 N08149574 N08149575 N08149671 N08153886 N08158320 N08159975 N08159977 N08159978 N08161541 N08161542 N08166269 N08167831 N08169640 N08169641 N08169642 N08176344 N08176345 N08177679 N08177680 N08184135 N08184138 N08184139 N08184856 N08184857 N08184858 N08184861 N08184862 N08184863 N08188260 N08188261 N08191893 N08198975 N08200845 N08200846 N08201073 N08203529 N08203530 N08206836 N08208307 N08208308 N08208309 N08212626 N08212627 N08213736 N08213737 N08213739 N08214754 N08218322 N08218323 N08218324 N08223527 N08223864 N08223865 N08223867 N08223868 N08223869 N08224684 N08224688 N08226285 N08231672 N08231673 N08231674 N08231675 N08231677 N08231678 N08231679 N08232081 N08232083 N08232084 N08232597 N08238074 N08238077 N08238078 N08238079 N08239471 N08239472 N08239473 N08239474 N08239475 N08245854 N08245855 N08245856 N08245863 N08246867 N08246870 N08248354 N08248355 N08248357 N08248358 N08248359 N08250623 N08250625 N08257411 N08257412 N08257413 N08259495 N09734896 N09736342 N09736343 N09736344 N09736346 N09736588 N09736594 N09736595 N09736596 N09736597 N09739964 N09739965 N09739966 N09739967 N09740159 N09740160 N11707446 N11707447 N11707774 N11707775 N11707776 N11707777 N11715900 N11715901 N11717901 N11724973 N29251669 N29251670 N29255589 N29255590 N29259317 N29259318 N29259320 N29259321 N29259324 N29457787 N29457788 N29457789 N29457790 N29457791 N29457792 N29475311 N29475312 N29475313 N29475314 N29475318 N29475319 N29475325 N29475326 N29475327 N29475328 N29476198 N29476199 N29476770 N29476771 N29476772 N29483351 N29487528 N29487529 N29487530 N29487531 N29500856 N29500951 N29501829 N29501831 N29507962 N29507963 N29509090 N29509091 N29510153 N29510154 N29510155 N29510156 N29510158 N29511200 N29512714 N29520356 N29520359 N29525793 N29534150 N29534379 N29534572 N29534602 N29536086 N29536088 N29536625 N29536626 N29536627 N29536682 N29536683 N29538921 N29538922 N29540603 N29540605 N29540606 N29540607 N29540609 N29540610 N29549401 N29551602 N29551608 N29553205 N29553206 N29553207 N29553274 N29553275 N29553277 N295533206 N29553585 N29553586 N29553587 N29553588 N29553590 N29553591 N29553592 N29553593 N29553594 N29556818 N29556820 N29556821 N29556822 N29557731 N29557740 N29558669 N29561613 N29561614 N29561615 N29561616 N29773604 N29773618 N29773620 N29773621 N29773622 N29773623 N29773624 N29773626 N29773661 N29773662 N29773663 N29773664 N29773668 N29773669 N29776051 N29776053 N29776697 N29776700 N29776876 N29776877 N29776880 N29776881 N29780152 N29780153 N29780154 N29788535 N29788536 N29788537 N29788542 N29788543 N29793116 N29794792 N29794901 N29796751 N29796752 N29801228 N29809720 N29809721 N29810224 N29811781 N29811783 N29812293 N29814820 N29822158 N29822160 N29822161 N29824309 N29824310 N29824311 N29824500 N29825187 N29825237 N29825261 N29827986 N29830913 N29830919 N29830920 N29830924 N29830925 N29836062 N29836064 N29836065 N29836430 N29836431 N29836433 N29836435 N29836437 N29836438 N29838839 N29838840 N29838841 N29838844 N29838845 N29843734 N29843735 N29843736 N29843737 N29843738 N29843739 N29843740 N29843741 N29843743 N29843744 N29843745 N29845224 N29845225 N29848894 N29849572 N29849573 N29849574 N29849575 N29849652 N29849655 N29849657 N29851124 N29851443 N29851445 N29853878 N29853879 N29853880 N29853881 N29856808 N29856809 N29856812 N29858116 N29858117 N29858122 N29858126 N29858230 N29858231 N29859396 N29859398 N29859400 N29860249 N29861320 N29861321 N29862026 N29862029 N29862030 N29864623 N29867546 N29867547 N29867548 N29867549 N29867586 N29867587 N29867889 N29868014 N29868807 N29868851 N29868852 N29868853 N29868855 N29868857 N29868858 N29869161 N29869162 N29869163 N29871247 N29871248 N29872557 N29872730 N29872737 N29872738 N29873364 N29875639 N29879267 N29881681 N29881682 N29881683 N29882616 N29882804 N29882806 N29882809 N29882815 N29885064 N29885088 N29885089 N29885090 N29885091 N29885092 N29885094 N29885095 N29885096 N29885153 N29885730 N29885732 N29886427 N29886428 N29886429 N29886430 N29886522 N29886523 N29886528 N29890096 N29890097 N29890098 N29890099 N29890100 N29891403 N29891404 N29891405 N29892098 N29896152 N29896153 N29898305 N29898306 N29898307 N29898308 N29898312 N29898372 N29898374 N29898375 N29898376 N29898377 N29898378 N29899351 N29900706 N29900707 N29900708 N29900709 N29900710 N29900711 N29900712 N29900713 N29900970 N29900971 N29900972 N29900973 N29900974 N29900995 N29901101 N29901102 N29901103 N29901971 N29901972 N29901973 N29901974 N29901976 N29901977 N29901978 N29901979 N29903582 N29903715 N29903791 N29903795 N29909843 N29913941 N29913943 N29916480 N29919195 N29921183 N29925864 N29925865 N29925867 N29925868 N29925869 N29925871 N29930372 N29930373 N29930375 N29930376 N29930377 N29930378 N29930379 N29930385 N29930386 N29932139 N29934766 N29934767 N29938110 N29939493 N29943214 N29943215 N29946821 N29946822 N29948628 N29956579 N29956580 N29956581 N29956584 N29956586 N29956587 N29957251 N29957252 N29957253 N29957254 N29960072 N29960078 N29960894 N29960895 N29960896 N29960897 N29960899 N29967249 N29967298 N29967299 N29969352 N29969353 N29975413 N29977065 N29977067 N29977068 N29977070 N29977071 N29980907 N29980908 N29980912 N29980913 N29980914 N29987099 N29987100 N29988801 N29988802 N29988803 N29994603 N29994604 N29994605 N29994607 N29994608 N29994609 N29994610 N29994612 N29998256 N29998258 N30001585 N30001587 N30001588 N30001589 N30001590 N30001591 N30001597 N30003181 N30005600 N30007080 N30007507 N30007508 N30007509 N30007511 N30007522 N30008601 N30008602 N30008604 N30009129 N30009130 N30009134 N30009136 N30018137 N30019272 N30019275 N30019277 N30019278 N30019279 N30019280 N30019281 N30020326 N30020328 N30020330 N30025650 N30026807 N30027807 N30027907 N30027908 N30027909 N30038958 N30038959 N30038960 N30038961 N30038962 N30038963 N30038966 N30043563 N30043564 N30043565 N30043566 N30043567 N30043568 N30046043 N30046048 N30046049 N30046050 N30050854 N30053253 N30053254 N30053255 N30053256 N30054924 N30054925 N30054926 N30055150 N30058285 N30067119 N30067124 N30067131 N30073492 N30073495 N30073498 N30073499 N30384001 N30384007 N30389701 N30389702 N30389703 N30389704 N30389705 N30389706 N30389708 N30389709 N30389710 N30391935 N30391936 N30391937 N30391938 N30396657 N30396658 N30396660 N30396662 N30399399 N30399400 N30402628 N30402629 N30414351 N30414352 N30416387 N30416389 N30416390 N30416392 N30416393 N30416394 N30416395 N30416396 N30416397 N30416398 N30416400 N30420851 N30422271 N30422272 N30422273 N30422275 N30422276 N30422278 N30422279 N30422280 N30422909 N30429440 N30431709 N30431713 N30431719 N30431723 N30431724 N30443160 N30448807 N30448820 N30449284 N30449286 N30449287 N30450675 N30450676 N30450677 N30451163 N30451175 N30451176 N30462308 N30462309 N30462772 N30466851 N30466857 N30466858 N30466861 N30466865 N30466866 N30467500 N30469301 N30469302 N30469304 N30469307 N30469310 N30469313 N30469316 N30469317 N30469320 N30469321 N30469322 N30471330 N30471331 N30471332 N30471333 N30471334 N30471335 N30471336 N30471337 N30471338 N30471339 N30471340 N30471341 N30471342 N30471343 N30471454 N30475008 N30475010 N30480378 N30480380 N30480382 N30480383 N30480384 N30480385 N30480386 N30480387 N30480388 N30480391 N30480392 N30480393 N30480394 N30483897 N30490874 N30490875 N30490877 N30490882 N30490883 N30490884 N30490885 N30490886 N30490887 N30490889 N30492389 N30492391 N30492393 N30492394 N30492396 N30497753 N30497764 N30504364 N30504366 N30504367 N30504377 N30504378 N30508880 N30508881 N30508884 N30508885 N30508890 N30508894 N30510249 N30513601 N30513602 N30513604 N30513606 N30513614 N30513618 N30513620 N30513622 N30513624 N30513625 N30513626 N30513627 N30519272 N30519274 N30519277 N30526707 N30526708 N30530249 N30534707 N30536928-1 N30536930-1 N30537285 N30537286 N30537287 N30537288 N30537289 N30537290 N30537291 N30537297-1 N30537298 N30537299 N30537301 N30537302 N30541201 N30541202 N30541203 N30541204 N30541208 N30544661 N30544671 N30545717 N30545718 N30549518 N30549519 N30549520 N30549521 N30549522 N30549528 N30549536 N30549537 N30549538 N30549540 N30558744 N30569641 N30572111 N30572117 N30573154 N30574297 N30577598 N30577600 N30584145 N30584146 N30584149 N30584150 N30586194 N30586196 S8 N30586197 N30593951 N30593961 N30598453 N30598455 N30598460 N30598463 N30598464 N30598465 N30598469 N30609087 N30609088 N30609089 N30609090 N30609091 N30609092 N30609094 N30610354 N30610355 N30610356 N30610357 N30610822 N30610823 N30613055 N30613056 N30613057 N30613058 N30613629 N30614065 N30614066 N30614067 N30614068 N30614069 N30614071 N30614072 N30614952 N30614955 N30614956 N30614957 N30614958 N30617616 N30617624 N30617629 N30619777 N30619779 N30619780 N30619781 N30619791 N30619792 N30629829 N30629830 N30629831 N30633466 N30633467 N30636361 N30645242 N30645243 N30645248 N30645250 N30650691 N30650692 N30650694 N30650696 N30655264 N30655265 N30655267 N30655268 N30655269 N30655270 N30655275 N30655276 N30655277 N30660153 N30660155 N30660156 N30660157 N30660158 N30660159 N30660160 N30660161 N30660164 N30662349 N30663086 N30663087 N30663088 N30663089 N30663091 N30663092 N30663093 N30663094 N30666234 N30666235 N30666236 N30666238 N30666239 N30666240 N30666241 N30676054 N30681929 N30681937 N30681943 N30681944 N30681945 N30681946 N30681947 N30681949 N30681950 N30682922 N30682951 N30682952 N30682953 N30682954 N30682955 N30682956 N30682957 N30682958 N30685584 N30685585 N30685586 N30685587 N30685588 N30685758 N30685759 N30685760 N30685761 N30685762 N30685763 N30685765 N30685766 N30685771 N30693296 N30693297 N30693298 N30695288 N30695289 N30695401 N30695402 N30695403 N30695404 N30695405 N30695406 N30695407 N30695408 N30695410 N30695411 N30695412 N30695413 N30695414 N30695417-1 N30695419 N30695420 N30701887 N30701888 N30701889 N30701890 N30701891 N30701892 N30703010 N30703012 N30703013 N30703014 N30703015 N30703016 N30703030 N30703033 N30709428 N30709429 N30709430 N30709431 N30709432 N30709433 N30717670 N30717672 N30717673 N30717674 N30717675 N30717677 N30719408 N30720663 N30720664 N30720665 N30720677 N30720678 N30720689 N30727243 N30727245 N30727247 N30727404 N30727405 N30727406 N30727407 N30729792 N30729796 N30733003 N30733004 N30733005 N30733007 N30733008 N30735176 N30735185 N30735187 N30735189 N30735191 N30740378 N30740380 N30740381 N30740383 N30740384 N30745968 N30745969 N30745970 N30745971 N30745972 N30745973 N30745975 N30757986 N30757987 N30757988 N30757989 N30757990 N30757993 N30757996 N30758699 N30758700 N30760980 N30760981 N30760982 N30760998 N30760999 N30761000 N30762955 N30762956 N30762957 N30765643 N30765644 N30765648 N30765649 N30765650 N30767501 N30767502 N30767504 N30768301 N30768304 N30768305 N30778123 N30778126 N30778127 N30778128 N30778131 N30778132 N30778134 N30778611 N30778614 N30778621 N30779301 N30779302 N30779303 N30779304 N30779306 N30779312 N30783989 N30783990 N30791195 N30791196 N30791698 N30792236 N30792237 N30792238 N30797083 N30797461 N30797462 N30797463 N30797465 N30797466 N30797468 N30801654 N30801655 N30801661 N30803334 N30803341 N30803344 N30805003 N30808863 N30808864 N30808865 N30813971 N30813972 N30813973 N30813975 N30813976 N30813977 N30813978 N30813981 N30816154 N30816155 N30816156 N30816157 N30816803 N30816805 N30816806 N30816807 N30816810 N30816812 N30840991 N30847500 N30848059 N30848065 N30848066 N30848068 N30873776 N30876676 N30880282 N30880952 N30880953 N30881461 N30881462 N30881464 N30881465 N30881466 N30881468 N30881469 N30881470 N30884083 N30884086 N30884092 N30884093 N30884096 N30884097 N30886547 N30886548 N30887960 N30890089 N30890090 N30890092 N30890093 N30891093 N30891094 N30891095 N30891096 N30891098 N30891100 N30894273 N30894274 N30894959 N30894961 N30901483 N30901484 N30901485 N30901486 N30901487 N30901488 N30901489 N30901490 N30901492 N30902424 N30902425 N30902426 N30902427 N30902428 N30902429 N30902430 N30902431 N30902432 N30902433 N30913434 N30913435 N30913436 N30913438 N30913439 N30916526 N30917086 N30917088 N30917089 N30927582 N30927595 N30933436 N30933439 N30933440 N30933445 N30949202 N30949203 N30949204 N30949205 N30955555 N30961315 N30961325 N30962994 N30962998 N30963485 N30963486 N30963500 N30964151 N30964152 N30964156 N30964158 N30964166 N30971117 N30971896 N30972464 N30972470 N30972476 N30972477 N30972478 N30972479 N30972481 N30972485 N30972486 N30972487 N30972971 N30972981 N30979946 N30986577 N30986578 N30986583 N30986584 N30986586 N30986589 N30986591 N30995390 N30995391 N30995392 N30995393 N30995395 N30995396 N30995940 N31004174 N31004175 N31004176 N31004177 N31004178 N31004179 N31004191 N31004192 N31004193 N31004194 N31011631 N31011632 N31011633 N31011634 N31011840 N31016934 N31016935 N31016936 N31016937 N31016938 N31016939 N31016940 N31020752 N31020753 N31022440 N31022441 N31022443 N31022444 N31023738 N31023739 N31023740 N31023742 N31023745 N31023746 N31031026 N31031032 N31031033 N31035032 N31035035 N31035039 N31035046 N31041494 N31041496 N31046712 N31046713 N31046714 N31046715 N31049473 N31049481 N31049482 N31049488 N31049489 N31049490 N31053593 N31053600 N31058498 N31066805 N31067153 N31067156 N31067157 N31067159 N31067160 N31067163 N31077413 N31077414 N31077416 N31077417 N31077418 N31077419 N31077420 N31077421 N31077422 N31077424 N31077425 N31088563 N31088564 N31088567 N31093271 N31093276 N31093277 N31093278 N31093279 N31093280 N31093284 N31093288 N31102006 N31102007 N31102008 N31102009 N31102010 N31102011 N31102012 N31102013 N31102014 N31102015 N31102016 N31102017 N31102018 N31102020 N31102021 N31102023 N31102024 N31102025 N31102026 N31104164 N31104167 N31104168 N31104169 N31104170 N31104171 N31104172 N31104173 N31104174 N31104175 N31104176 N31104180 N31104184 N31104185 N31104186 N31104187 N31104188 N31104189 N31104190 N31104193 N31104194 N31104195 N31110727 N31110728 N31110731 N31110732 N31110742 N31110744 N31110745 N31118470 N31118474 N31118840 N31118841 N31118843 N31120564 N31120579 N31123968 N31123971 N31126501 N31126502 N31131378 N31131379 N31131380 N31131388 N31131389 N31131390 N31131391 N31144935 N31144936 N31144941 N31144942 N31149393 N31157834 N31157837 N31157845 N31157846 N31157848 N31162827 N31162830 N31162831 N31162841 N31169936 N31175283 N31175291 N31175294 N31175296 N31180547 N31180550 N31181858 N31181859 N31188058 N31189938 N31189939 N31189940 N31208836 N31210271 N31210277 N31212587 N31213136 N31213138 N31217618 N31230148 N31234491 N31242231 N31242232 N31249930 N31249932 N31249938 N31249939 N31249943 N31258584 N31258592 N31265184 N31265826 N31265829 N31271212 N31272133 N31272134 N31272135 N31272144 N31272145 N31272146 N31272147 N31272148 N31272149 N31273273 N31273274 N31273276 N31280253 N31280258 N31280262 N31286158 N31286159 N31286161 N31286163 N31286166 N31286167 N31286172 N31293530 N31294358 N31300767 N31301416 N31301419 N31301420 N31301421 N31301422 N31306925 N31313634 N313184342 N31318446 N31322964 N31322965 N31328643 N31328903 N31328905 N31332723 N31332729 N31336927 N31336928 N31338251 N31338256 N31338260 N31338263 N31341365 N31341367 N31341375 N31341376 N31347987 N31347998 N31352351 N31352352 N31352353 N31352358 N31367836 N31367840 N31376271 N31384772 N31384775 N31384776 N31384777 N31384778 N31384779 N31384784 N31384785 N31384791 N31385165 N31385166 N31385169 N31389386 N31389389 N31392891 N31398039 N31398040 N31400471 N31400475 N31400476 N31400477 N31406371 N31406372 N31406376 N31406388 N31406863 N31406864 N31406865 N31406869 N31406877 N31406880 N31407925 N31407926 N31413729 N31422273 N31431042 N31431043 N31438953 N31444258 N31445660 N31445689 N31464395 N31464398 N31464399 N31464903 UDI-DI code: 00763000368494 Serial Numbers: N07370801 N07441433 N07521588 N08159976 N08223525 N08223526 N08245859 N11710650 N29255594 N29529070 N29531246 N29810113 N29855918 N29934774 N30001599 N30009142 N30043574 N30046047 N30067122 N30067123 N30389711 N30416388 N30449098 N30462311 N30610820 N30803335 N30927583 N30927584 N31022442 N31049476 N31088569 N31088570
FEI Number 1000517638
Recalling Firm/
Manufacturer
Medtronic Navigation, Inc.
200 Medtronic Drive
Lafayette CO 80026
For Additional Information ContactRita Greenberg
612-385-3537
Manufacturer Reason
for Recall
Due to two software anomalies that can occur within a spine procedure. Two issues are: 1. Navigation Orientation - when lateral orientation radiographic images are taken, information may appear flipped when utilizing the system (180 degree flipped). 2. Pop-up Message - when using CT+Fluouroscopy images with a slice spacing thickness greater than 2mm, the system does not display a pop-up message indicating the slice spacing is not optimal.
FDA Determined
Cause 2
Software design
ActionOn 08/22/2024, an "URGENT: MEDICAL DEVICE CORRECTION" Letter was mailed to customers informing them of 2 software anomalies impacting spine procedures on the StealthStation S8 system (StealthStation S8 Basic, StealthStation S8 Plus and StealthStation S8 Premium) with Application software version 1.3.0 or 1.3.2. Issues are: 1.Navigation Orientation: When lateral orientation radiographic images are taken using a navigation-enabled Ziehm Vision FD Vario 3D C-arm, the orientation information appears flipped when displayed by the StealthStation S8 system. The navigation information is displayed with a 180-degree rotation around the superior/inferior(S/I) axis with the left/right (L/R) axis and anterior/posterior (A/P) axis appearing as mirror images. The exam image orientation and related indication labels on the exams remain correct. Note -this anomaly only impacts StealthStation" S8 systems where the Ziehm Vision FDVario 3D C-arm Interface Kit, 9734486, installed and calibrated on a Ziehm FD system in the lateral orientation; other orientations settings are not impacted. Medtronic Recommends Do not use the lateral orientation setting on the Ziehm Vision FD Vario 3D C-arm, if using the StealthStation" S8 with software version 1.3 or 1.3.2. 2.Pop-up Message: When using CT + Fluoroscopy images with a slice spacing thickness greater than 2 mm, the system does not display a pop-up message indicating the slice spacing is not optimal. Navigation remains accurate in this situation. Customers are instructed to: -Review Appendix A which provides detailed examples of the issues and how to identify if your system is affected. -Medtronic will work expeditiously with facilities to coordinate a software update to version 2.1 at no charge to resolve anomalies. Questions, contact Medtronic Technical Services at 1-888-826-5603 or email at rs.navtechsupport@medtronic.com.
Quantity in Commerce1,801 systems
DistributionWorldwide - U.S Nationwide distribution including in the states of AK, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NM, NV, NY, OH, OK, PA, Puerto Rico, SC, TN, TX, UT, VA, WA, WI and WV. The countries of Argentina, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, Brazil, BULGARIA, CANADA, CHILE, China, Colombia, Costa Rica, CROATIA, CYPRUS, DENMARK, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE, HUNGARY, ICELAND, India, ISRAEL, ITALY, Japan, Jordan, KAZAKHSTAN, KENYA, LEBANON, LIBYAN ARAB JAMAHIRIYA, LITHUANIA, Malaysia, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NICARAGUA, OMAN, PAKISTAN, PANAMA, Philippines, POLAND, PUERTO RICO, QATAR, Republic of Korea, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SERBIA, Singapore, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, Turkey, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = OLO
-
-