| | Class 2 Device Recall MiniCap Extended Life PD Transfer Set |  |
| Date Initiated by Firm | October 23, 2024 |
| Date Posted | November 27, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0562-2025 |
| Recall Event ID |
95524 |
| 510(K)Number | K192705 |
| Product Classification |
Set, administration, for peritoneal dialysis, disposable - Product Code KDJ
|
| Product | Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number 5C4483; use in Peritoneal Dialysis |
| Code Information |
UDI/DI 00085412007748, Lot/Serial Numbers: H23J10051, H23J23062, H23K20066, H24C08033, H24E16080, H24F12046, H24F17045, H24G02045, H24I02090 |
| FEI Number |
1417572
|
Recalling Firm/ Manufacturer |
Baxter Healthcare Corporation 1 Baxter Pkwy Deerfield IL 60015-4625
|
| For Additional Information Contact | Center for One Baxter 800-422-9837 |
Manufacturer Reason for Recall | There has been an increase in the number of complaints related to the separation of the female connector from the main body of the transfer set. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Baxter issued a SAFETY ALERT notice to its consignees on 10/23/2024 via USPS first class mail. The notice explains the issue and potential hazard involved. The notice addressed to dialysis providers indicated that if a patient contacts your clinic experiencing separation of the transfer set, to replace their transfer set and contact Baxter. Notices addressed to peritoneal patients provided the following instructions:
"If you experience a separation of the transfer set, please close the transfer set twist clamp and do not cut the patient line. Contact your doctor and/or nurse immediately to arrange for a transfer set replacement and further guidance."
Distributors were directed to notify their customers.
For general questions regarding this communication, please contact Baxter Renal Home Care Services which can be reached at 800-284-4060, option 1, between the hours of 7:00 am and 6:00 pm Central Time, Monday through Friday.
|
| Quantity in Commerce | 62,718 |
| Distribution | Worldwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = KDJ
|
|
|
|