Class 2 Device Recall Siemens Atellica CH Reaction Ring Cuvette Segment

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Siemens Atellica CH Reaction Ring Cuvette Segmentsee related information
Date Initiated by FirmOctober 21, 2024
Date PostedNovember 20, 2024
Recall Status1 Open3, Classified
Recall NumberZ-0496-2025
Recall Event ID 95626
510(K)NumberK161954 
Product Classification Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
ProductSiemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain and incubate the sample reaction volumes and present them to each processing step. Siemens Material Number(SMN): 11099326 All lots starting with N15 or 15 are expected to have the defect.
Code Information Lot # UDI N1527024 (01)00630414596143(10)N1527024(17)20291002; N1526424 (01)00630414596143(10)N1526424(17)20291001; N1525924 (01)00630414596143(10)N1525924(17)20290924 ; N1525324 (01)00630414596143(10)N1525324(17)20290918 ; N1524224 (01)00630414596143(10)N1524224(17)20290911; N1523724 (01)00630414596143(10)N1523724(17)20290904; N1523024 (01)00630414596143(10)N1523024(17)20290827; N1521824 (01)00630414596143(10)N1521824(17)20290814; N1521424 (01)00630414596143(10)N1521424(17)20290808; N1519324 (01)00630414596143(10)N1519324(17)20290719; N1518924 (01)00630414596143(10)N1518924(17)20290716; N1518224 (01)00630414596143(10)N1518224(17)20290710; N1518324 (01)00630414596143(10)N1518324(17)20290710; N1516224 (01)00630414596143(10)N1516224(17)20290621; N1515624 (01)00630414596143(10)N1515624(17)20290612; N1514924 (01)00630414596143(10)N1514924(17)20290605; N1514124 (01)00630414596143(10)N1514124(17)20290530; N1513324 (01)00630414596143(10)N1513324(17)20290522; N1512824 (01)00630414596143(10)N1512824(17)20290515; N1512424 (01)00630414596143(10)N1512424(17)20290510; N1511224 (01)00630414596143(10)N1511224(17)20290506; N1510724 (01)00630414596143(10)N1510724(17)20290423; N1510124 (01)00630414596143(10)N1510124(17)20290419.
FEI Number 2432235
Recalling Firm/
Manufacturer
Siemens Healthcare Diagnostics, Inc.
511 Benedict Ave
Tarrytown NY 10591-5005
For Additional Information ContactSAME
914-631-8000
Manufacturer Reason
for Recall
A manufacturing defect causes Atellica CH Reaction Ring Cuvette Segments with lot numbers N15 on the packaging or 15 on the underside of the cuvette segment have a potential for imprecision and erroneously depressed or elevated QC and patient results for seven (7) analytes: Ammonia, Acetaminophen, Alanine Aminotransferase (ALT), Hemoglobin A1c, Creatinine Kinase (CK), Cystatin C, and Salicylate.
FDA Determined
Cause 2
Under Investigation by firm
ActionSiemens issued Urgent Medical Device Correction (UMDC) to l US affected customers via FedEx beginning on 10/21/2024. An Urgent Field Safety Notice (UFSN) was issued to all OUS Siemens reporting sites with affected customers beginning on 10/21/2024. In accordance with outside US country regulations, Siemens in-country personnel for the EU and non-EU regions follow local requirements for contacting their respective National Competent Authorities. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Review your current Atellica CH Reaction Ring Cuvette Segment stock. All Atellica CH Reaction Ring Cuvette Segments beginning with 11 are not part of this communication and can be used without restriction. Any other lot in your stock must be immediately discarded. Locate your analyzer s serial number within the Appendix B Serial Number List. If your analyzer s serial number is in the provided list, perform the instructions in Appendix A to assess, inspect, and/or replace the Atellica CH Reaction Ring Cuvettes Segments. Determine your laboratory s replacement needs and provide detailed information to Siemens Healthineers for reporting to authorities. Complete and return the Field Correction Effectiveness Check Form attached to this letter within thirty (30) days. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Siemens Healthineers has confirmed Atellica CH reaction ring segments beginning with lot number 11 are not impacted, and going forward all future lots will work as expected. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Quantity in Commerce9729 units
DistributionUS Nationwide and Worldwide distribution: Foreign: Algeria Algeria Argentina Australia Austria Bahrain Bangladesh Belgium Brazil Bulgaria Burkina Faso Canada Chile China Colombia Congo, Republic C¿te dIvoire Croatia Cyprus Czech Republic Denmark Ecuador Egypt Estonia Finland France Germany Greece Hong Kong Hungary India Iran, Islamic Republic of Iraq Ireland Israel Italy Japan Jordan Kenya Kuwait Latvia Libya Malaysia Mexico Moldova, Rep. of Morocco Nepal Netherlands Norway Oman Pakistan Paraguay Peru Philippines Poland Portugal Qatar Republic Korea Romania Russian Federation Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan, Province of China Thailand Tunisia Turkey Uganda Ukraine United Arab Emirates United Kingdom Great Britain Uruguay Vatikancity Vietnam
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = JJE
-
-