| | Class 2 Device Recall Allura Xper FD series and Allura Xper OR Table series |  |
| Date Initiated by Firm | November 11, 2024 |
| Date Posted | December 12, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0676-2025 |
| Recall Event ID |
95713 |
| 510(K)Number | K161563 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following:
1. Allura Xper FD10, Model Number: 722010.
2. Allura Xper FD10/10, Model Number: 722011.
3. Allura Xper FD20, Model Number: 722012.
4. Allura Xper FD20 Biplane, Model Number: 722013.
5. Allura Xper FD10 OR Table, Model Number: 722022.
6. Allura Xper FD20 OR Table, Model Number: 722023.
7. Allura Xper FD20 Biplane OR Table, Model Number: 722025. |
| Code Information |
1. Model Number: 722010. UDI (Serial Numbers): (01)00884838054189(21)164 (164); (01)00884838059030(21)272 (272); (01)00884838059030(21)819 (819); (01)00884838059030(21)828 (828); No UDI (33, 39, 42, 43, 65, 67, 69, 79, 107, 119, 127, 158, 160, 169, 177, 190, 191, 209, 213, 214, 215, 218, 227, 229, 239, 240, 241, 252, 274, 280, 285, 286, 290, 296, 298, 306, 308, 316, 342, 343, 344, 360, 364, 365, 395, 411, 414, 422, 423, 453, 468, 477, 478, 480, 485, 488, 499, 500, 505, 521, 527, 531, 567, 572, 585, 592, 604, 605, 611, 624, 636, 637, 638, 641, 646, 648, 651, 656, 662, 668, 670, 671, 673, 677, 679, 683, 690, 694, 698, 714, 728, 730, 755, 763, 764, 780, 785, 787, 788, 791, 792, 794, 809, 818, 830, 833, 838, 839, 873, 878, 882, 890, 896, 902, 905, 914, 917, 920, 921, 931, 932, 937, 939, 946, 968, 970, 979, 986, 987, 992, 999, 1005, 1009, 1011, 1025, 1028, 1040, 1041, 1042, 1043, 1048, 1064, 1069, 1078, 1082, 1084, 1109, 1112, 1118, 1123, 1126, 1128, 1132, 1144, 1170, 1173, 1176, 1181, 1183, 1209, 1210, 1214, 1230, 1233, 1234, 1235, 1236, 1261, 1263, 1275, 1279, 1283, 1292, 1297, 1442, 601, 654, 1122, 1198, 332, 777, 666, 301, 608, 1081, 426, 924, 577, 265, 1047, 1187, 219, 959, 901, 552, 888, 1140, 64, 935, 1008, 1213, 1015, 1016, 1017, 1018, 374, 524, 1153, 1192, 1194, 841, 964, 996, 233, 1211, 721, 1190, 1188, 806, 907, 50, 418, 71, 400, 1168, 1116, 1093, 53, 370, 596, 45, 712, 1148, 323, 825, 402, 713, 195, 176, 508, 123, 1197, 91, 602, 245, 535, 230, 231, 232, 991, 248, 92, 187, 502, 1124, 537, 1199, 941, 553, 302, 289, 173, 134, 1137, 948, 509, 619, 1034, 447, 707, 708, 1285, 928, 1113, 802, 803, 192, 613, 614, 393, 506, 369, 1206, 264, 699, 759, 722, 1100, 350, 748, 1220, 657, 658, 659, 854, 565, 734, 735, 736, 737, 228, 642, 1054, 490, 391, 425, 513, 898, 424, 105, 993, 732, 580, 1256, 704, 695, 696, 238, 706, 1253, 1327, 1032, 319, 1252, 469, 541, 731, 1019, 1023, 1215, 247, 1191, 856, 51, 971, 603, 1003, 270, 267, 268, 1242, 47, 1089, 831, 530, 1091, 193, 578, 322, 1161, 126, 197, 884, 146, 62, 1095, 273, 29, 63, 412, 346, 622, 371, 740, 715, 635, 1134, 1135, 520, 936, 990, 1002, 911, 49, 978, 586, 1004, 1036, 653, 1086, 1180, 776, 288, 1245, 72, 625, 1221, 1087, 984, 253, 398, 1125, 1269, 773, 80, 380, 224, 222, 1010, 957, 282, 847, 1151, 1083, 1268, 711, 685, 686, 1104, 663, 664, 863, 799, 1224, 1225, 771, 631, 494, 1250, 652, 1106, 383, 52, 1278, 861, 906, 872, 382, 409, 969, 139, 1202, 28, 410, 1050, 915, 544, 299, 769, 540, 1007, 1305, 1306, 612, 1203, 1216, 620, 300, 1058, 600, 384, 90, 811, 1138, 967, 795, 1154, 655, 610, 895, 136, 944, 255, 415, 866, 954, 720, 83, 1141, 86, 88, 246, 801, 498, 1039, 574, 495, 1265, 1177, 1178, 128, 1105, 1266, 97, 703, 526, 1237, 650, 98, 427, 496, 1286, 897, 649, 401, 963, 876, 275, 800, 1012, 102, 1238, 475, 1090, 261, 739, 356, 877, 34, 35, 249, 1196, 1276, 1294, 927, 375, 162, 333, 943, 293, 903, 1111, 797, 416, 955, 1145, 514, 100, 1273, 1157, 860, 942, 961, 287, 1021, 1033, 998, 733, 154, 353, 1150, 832, 474, 471, 962, 1037, 1053, 579, 211, 404, 428, 503, 493, 1120, 837, 406, 351, 528, 559, 217, 529, 58, 566, 588, 589, 590, 591, 593, 594, 891, 949, 1146, 1267, 518, 568, 842, 744, 106, 103, 104, 385, 956, 70, 850, 250, 1097, 904, 75, 199, 328, 182, 294, 77, 751, 810, 1280, 1300, 1077, 858, 362, 1207, 874, 131, 1218, 1130, 972, 710, 934, 587, 645, 855, 178, 790, 466, 94, 276, 952, 548, 115, 997, 116, 753, 606, 320, 114, 181, 1182, 1065, 562, 180, 789, 1067, 1129, 975, 626, 89, 523, 243, 223, 1617, 186, 555, 598, 1119, 702, 321, 338, 137, 1051, 660, 1088, 741, 742, 1029, 1136, 95, 844, 973, 1162, 1000, 394, 142, 561, 277, 269, 304, 558, 132, 804, 1044, 822, 852, 476, 862, 96, 396, 254, 373, 875, 312, 313, 900, 220, 563, 330, 367, 473, 515, 909, 112, 121, 762, 564, 1074, 1075, 20, 124, 357, 807, 834, 1066, 184, 697, 857, 634, 980, 782, 783, 125, 1096, 225, 812, 813, 1055, 913, 1185, 368, 349, 570, 781, 354, 386, 387, 525, 66, 388, 235, 236, 549, 251, 278, 868, 1006, 1060, 516, 517, 774, 60, 166, 501, 845, 846, 597, 381, 492, 908, 607, 507, 923, 775, 403, 719, 687, 78, 1030, 1031, 185, 1147, 689).
2. Model Number: 722011. No UDI. Serial Numbers: 157, 83, 308, 95, 107, 34, 59, 23, 319, 49, 170, 137, 173, 101, 84, 155, 133, 234, 345, 124, 332, 102, 270, 177, 132, 325, 15, 76, 79, 17, 87, 237, 255, 50, 171, 168, 22, 275, 344, 324, 140, 163, 172, 103, 82, 154, 215, 40, 69, 272, 286, 81, 224, 184, 266, 175, 146, 164, 334, 338, 165, 80, 231, 323, 327, 179, 145, 117, 6, 355, 151, 202, 21, 203, 214, 340, 18, 301, 302, 303, 304, 305, 306, 307, 257, 265, 127, 180, 259, 181, 211, 206, 217, 314, 98, 201, 341, 318, 347, 261, 178, 183, 271, 253, 230, 317, 198, 321, 204, 94, 104, 52, 290, 291, 244, 333, 287, 39, 282, 78, 216, 158, 329, 246, 295, 56, 337, 292, 125, 225, 300, 88, 309, 86, 222, 71, 156, 233, 31, 188, 77, 99, 238, 223, 51, 153, 68, 129, 186, 118, 119, 41, 55, 263, 278, 258, 32, 89, 108, 310, 311, 139, 141, 66, 274, 296, 243, 191, 227, 273, 161, 220, 242, 166, 189, 185, 100, 190, 299, 280, 199, 194, 74, 75, 228, 250, 251, 91, 213, 12, 212, 113, 152, 112, 111, 115, 114, 116, 43, 277, 313, 26, 27, 29.
3. Model Number: 722012. UDI (Serial Numbers): (01)00884838059054(21)1356 (1356); (01)00884838059054(21)191 (191); (01)00884838059054(21)351 (351); (01)00884838059054(21)583 (583); (01)00884838059054(21)655 (655); (01)00884838059054(21)796 (796); (01)00884838059054(21)950 (950); No UDI (1117, 1001, 1673, 1712, 585, 517, 1545, 70, 1715, 504, 757, 935, 1250, 1613, 2006, 1733, 1734, 2201, 1922, 741, 92, 1034, 1359, 1848, 865, 1908, 1166, 1113, 1659, 494, 335, 1804, 1473, 1408, 1401, 1588, 576, 595, 590, 2302, 1279, 2151, 481, 348, 543, 893, 1187, 1455, 2085, 2202, 423, 525, 1798, 1656, 399, 990, 1871, 951, 1652, 894, 1590, 1597, 596, 1054, 1189, 1099, 823, 1037, 566, 72, 448, 1606, 924, 1610, 686, 1108, 1294, 568, 500, 459, 469, 1422, 1077, 1537, 1035, 285, 243, 84, 1317, 1649, 1895, 1900, 1723, 878, 919, 1581, 2221, 2222, 1515, 1803, 1731, 1017, 1020, 380, 1811, 2171, 1497, 782, 418, 1053, 1558, 1336, 601, 1014, 684, 1265, 331, 389, 732, 751, 1324, 2276, 2086, 1285, 2219, 298, 814, 815, 1120, 278, 297, 1713, 1506, 580, 1098, 1852, 1307, 1858, 377, 758, 1598, 172, 2054, 856, 1690, 1407, 480, 690, 579, 1327, 733, 1091, 1254, 797, 213, 735, 1280, 2138, 828, 458, 1118, 740, 1249, 542, 1033, 2265, 490, 1075, 2203, 1321, 1396, 605, 606, 444, 66, 2092, 1269, 67, 244, 1002, 1022, 1023, 2264, 2234, 1805, 1192, 1823, 483, 484, 615, 2150, 892, 1478, 1262, 1688, 1961, 1732, 1062, 649, 650, 767, 1104, 1641, 1277, 1975, 276, 316, 1335, 1461, 159, 489, 766, 639, 669, 671, 1583, 451, 625, 651, 1047, 107, 1447, 99, 670, 124, 1616, 1660, 1083, 1807, 1924, 511, 1714, 122, 101, 75, 1406, 352, 492, 472, 1105, 642, 643, 1270, 577, 578, 1306, 855, 1456, 2269, 71, 1078, 2285, 1793, 1797, 1216, 948, 387, 498, 1662, 104, 1868, 1743, 369, 1239, 1561, 464, 2293, 1960, 616, 1508, 1084, 1038, 1058, 318, 1990, 1796, 1026, 1337, 685, 656, 1560, 1441, 2926, 1854, 2288, 473, 2068, 624, 1586, 1617, 1904, 1354, 1918, 1906, 160, 383, 560, 68, 114, 1948, 1195, 518, 888, 1329, 1992, 323, 379, 792, 1003, 1219, 89, 115, 2003, 2004, 135, 1832, 2133, 2166, 562, 1941, 760, 859, 1593, 1608, 1947, 1950, 731, 520, 1328, 848, 2292, 1946, 274, 1063, 611, 1587, 1226, 613, 1271, 1842, 205, 2147, 1799, 1665, 187, 1513, 2015, 2207, 319, 2161, 1116, 1993, 208, 943, 150, 1281, 1945, 1919, 110, 337, 1735, 1326, 788, 488, 2002, 132, 1218, 1519, 1533, 336, 694, 1334, 2013, 1405, 1141, 923, 925, 533, 406, 1293, 513, 1446, 1625, 1675, 1683, 1148, 1181, 207, 1305, 790, 499, 1716, 934, 1726, 85, 476, 1657, 1045, 1531, 1630, 851, 1827, 277, 1923, 1516, 404, 453, 321, 334, 197, 1869, 2215, 574, 1611, 1559, 2924, 1686, 1635, 858, 1224, 1345, 86, 2057, 1430, 1255, 1599, 302, 610, 1209, 920, 666, 1989, 2014, 2022, 1126, 1943, 1188, 1073, 1404, 1664, 103, 1272, 2063, 1467, 1596, 326, 926, 675, 756, 1027, 424, 1894, 2145, 2146, 2148, 2065, 1527, 1477, 2227, 1177, 1645, 93, 95.1167, 1595, 88, 1672, 909, 1210, 1512, 1101, 787, 1399, 123, 126, 136, 952, 953, 954, 1005, 1661, 2156, 320, 186, 455, 570, 1622, 1115, 1316, 1212, 975, 976, 1032, 1746, 986, 971, 1594, 233, 239, 640, 446, 69, 430, 1542, 76, 1607, 1211, 1004, 2046, 974, 1044, 505, 310, 192, 314, 852, 224, 1016, 1097, 315, 1532, 1964, 2135, 1357, 589, 1092, 248, 846, 963, 964, 1658, 1344, 1902, 328, 781, 1909, 1132, 1361, 1504, 1433, 229, 142, 962, 1260, 1094, 837, 333, 1681, 1815, 922, 1333, 774, 1492, 2047, 1655, 1855, 2098, 930, 1052, 1421, 627, 1428, 1994, 1802, 1956, 931, 1817, 309, 2168, 1168, 1562, 967, 167, 910, 632, 863, 957, 289, 137, 1040, 1444, 409, 993, 995, 57, 1072, 516, 2042, 1332, 809, 795, 1460, 170, 914, 1156, 347, 1291, 1971, 664, 180, 1376, 2107, 139, 350, 628, 1865, 550, 1872, 759, 1464, 1363, 1457, 2020, 553, 288, 1980, 1580, 1708, 1495, 1244, 556, 546, 806, 1289, 1284, 1144, 1145, 223, 1081, 1436, 726, 755, 648, 1391, 1972, 215, 871, 629, 1631, 125, 432, 966, 1157, 1623, 1702, 916, 2112, 2123, 2118, 98, 1297, 2103, 211, 1147, 1303, 2300, 2120, 199, 1833, 1874, 631, 1887, 891, 1382, 1471, 1818, 2270, 1878, 1462, 1384, 798, 165, 210, 1203, 1966, 1178, 1999, 793, 794, 667, 2196, 1669, 502, 1021, 1142, 1879, 1703, 2031, 659, 1572, 1076, 2291, 1093, 479, 1043, 1626, 2056, 608, 1624, 216, 2204, 1138, 839, 805, 381, 752, 555, 1548, 1814, 1808, 979, 390, 261, 269, 2188, 1056, 866, 1699, 619, 1267, 311, 477, 1174, 1485, 59, 541, 1415, 764, 388, 876, 1150, 1701, 1693, 2080, 1986, 772, 1241, 1819, 1898, 911, 978, 2093, 116, 56, 2181, 599, 2034, 1071, 622, 941, 1088, 1927, 1412, 181, 548, 582, 588, 1567, 1367, 1143, 1423, 1268, 279, 1592, 1122, 2028, 1375, 1059, 400, 166, 746, 771, 770, 2177, 280, 1202, 753, 2274, 2198, 1028, 581, 1309, 810, 1086, 524, 1390, 1721, 374, 738, 360, 992, 219, 1825, 1839, 652, 2102, 2127, 1834, 829, 1434, 665, 742, 737, 1360, 1349, 391, 415, 2095, 730, 1967, 1996, 763, 332, 867, 902, 917, 1366, 2096, 1201, 587, 1061, 2072, 1691, 1861, 407, 1741, 833, 554, 214, 779, 799, 2101, 1228, 2117, 534, 1348, 813, 1740, 1530, 998, 1266, 680, 1997, 441, 60, 803, 2195, 895, 509, 1736, 355, 1381, 112, 478, 227, 785, 291, 185, 1450, 540, 1976, 2016, 1591, 443, 106, 127, 382, 661, 727, 1137, 1795, 1128, 1377, 392, 1134, 1130, 296, 266, 1074, 376, 2067, 152, 1978, 663, 1678, 569, 1481, 2349, 436, 1929, 1654, 287, 944, 2176, 1413, 188, 2140, 1998, 635, 1089, 1668, 1256, 2272, 660, 1671, 743, 58, 128, 1287, 1079, 61, 982, 693, 1388, 184, 1479, 1801, 2074, 1342, 405, 1535, 1882, 1127, 980, 2058, 1705, 260, 539, 2278, 748, 1435, 528, 506, 1885, 1926, 2073, 572, 688, 903, 410, 901, 869, 728, 91, 218, 419, 275, 830, 2094, 1411, 1222, 1397, 272, 2153, 1711, 1965, 1480, 1937, 1298, 414, 338, 1237, 1151, 1152, 1347, 1977, 1836, 1131, 1242, 573, 1484, 117, 2082, 1695, 817, 1536, 630, 593, 773, 1214, 936, 679, 1816, 544, 2392, 1862, 1364, 2033, 1296, 2200, 2017, 2208, 1175, 1233, 1653, 1463, 955, 2087, 1698, 1064, 1958, 271, 1135, 1385, 870, 507, 2182, 1223, 1288, 2008, 398, 1234, 965, 662, 868, 1483, 1489, 647, 1232, 2175, 1984, 222, 840, 175, 176, 1261, 437, 557, 1082, 1633, 426, 672, 1379, 872, 1973, 1822, 144, 692, 1820, 1821, 1312, 2077, 161, 140, 1696, 678, 780, 73, 74, 1292, 1466, 34, 119, 2043, 155, 1837, 118, 1451, 598, 1158, 1153, 1663, 1546, 835, 177, 1720, 749, 1400, 1897, 1246, 1245, 1939, 897, 2137, 111, 2206, 1585, 880, 1845, 777, 403, 26, 523, 1951, 2144, 1494, 2921, 385, 1930, 609, 1748, 1425, 1988, 2000, 148, 592, 1870, 674, 1049, 510, 2294, 2316, 1829, 1838, 1915, 1172, 687, 1710, 80, 1987, 428, 602, 623, 2220, 775, 988, 1582, 1979, 928, 807, 1322, 529, 1891, 465, 1719, 808, 1048, 1024, 824, 745, 617, 985, 1259, 1136, 2927, 1534, 1350, 1452, 1620, 802, 913, 1180, 354, 1813, 522, 983, 1121, 421, 1867, 1198, 416, 1840, 1812, 2066, 212, 933, 1676, 1634, 2059, 1541, 1338, 1520, 1955, 178, 249, 1418, 1251, 1176, 1417, 1899, 1454, 1055, 1962, 1015, 1124, 1547, 2266, 1230, 1543, 1609, 1744, 2110, 1553, 2922, 940, 1639.657, 2141, 1619, 143, 1368, 1149, 292, 887, 113, 105, 1742, 2280, 1949, 475, 1850, 1901, 845, 2100, 273, 1319, 2036, 468, 1341, 545, 526, 769, 1554, 1651, 932, 1060, 1647, 1220, 1200, 1932, 1931, 1935, 491, 514, 294, 1227, 236, 408, 1847, 2089, 873, 874, 1046, 994, 1745, 173, 1700, 2097, 341, 342, 1648, 853, 744, 786, 1355, 1925, 1313, 1991, 373, 1171, 1551, 1794, 2091, 1615, 1439, 1911, 149, 1080, 1, 618, 471, 1844, 1697, 972, 353, 2114, 1283, 1448, 267, 1123, 1206, 497, 425, 1330, 1888, 1050, 2253, 1273, 1974, 1907, 818, 621, 890, 1374, 1747, 162, 1371, 1125, 1031, 1085, 1864, 1565, 179, 946, 1372, 945, 1449, 1517, 1186, 2078, 1318, 1264, 1351, 1640, 1667, 1248, 1637, 1576, 1575, 1493, 646, 1577, 1579, 1574, 329, 2069, 1914, 1173, 343, 304, 306, 307, 308, 1684, 433, 413, 1159, 429, 2010, 1490, 1670, 921, 882, 1051, 2347, 567, 422, 1304, 849, 1853, 1057, 1295, 2055, 603, 1090, 463, 1841, 1129, 1959, 604, 850, 454, 460, 2053, 812, 1468, 1921, 1896, 1982, 1503, 295, 1969, 2130, 411, 183, 375, 969, 1628, 634, 1555, 847, 1666, 240, 1320, 1706, 1528, 241, 242, 157, 156, 804, 1432, 1008, 1068, 658).
4. Model Number: 722013. No UDI. Serial Numbers: 505, 217, 313, 492, 34, 153, 410, 360, 290, 74, 83, 580, 160, 404, 126, 244, 106, 346, 461, 307, 1, 442, 435, 293, 297, 46, 289, 638, 342, 69, 267, 214, 402, 113, 86, 603, 82, 149, 424, 140, 254, 255, 266, 475, 518, 519, 303, 194, 115, 150, 496, 599, 441, 491, 124, 151, 179, 550, 377, 575, 562, 178, 309, 601, 327, 73, 77, 264, 78, 308, 141, 357, 85, 40, 216, 637, 422, 174, 392, 109, 218, 228, 511, 389, 152, 479, 262, 61, 401, 497, 84, 125, 241, 564, 573, 426, 556, 42, 100, 495, 500, 96, 515, 105, 428, 427, 531, 375, 316, 278, 347, 205, 246, 43, 639, 35, 480, 304, 147, 15, 52, 283, 168, 171, 265, 272, 332, 509, 444, 400, 292, 117, 598, 415, 454, 222, 215, 335, 416, 47, 384, 559, 361, 358, 549, 476, 133, 36, 315, 31, 490, 417, 48, 276, 247, 618, 390, 157, 53, 238, 525, 17, 528, 224, 76, 585, 271, 172, 50, 544, 570, 359, 296, 530, 232, 478, 457, 547, 183, 184, 185, 186, 187, 188, 189, 190, 413, 437, 107, 138, 227, 318, 341, 225, 29, 508, 103, 434, 473, 135, 551, 557, 541, 294, 128, 388, 207, 258, 429, 95, 411, 353, 542, 533, 523, 182, 574, 139, 137, 250, 563, 502, 450, 436, 275, 524, 354, 510, 395, 458, 261, 595, 586, 535, 569, 425, 334, 344, 355, 356, 465, 345, 501, 522, 64, 396, 208, 408, 555, 394, 57, 520, 493, 201, 366, 527, 65, 449, 167, 70, 312, 321, 161, 324, 311, 176, 181, 554, 104, 459, 130, 325, 470, 440, 144, 28, 385, 403, 81, 134, 487, 256, 374, 405, 380, 367, 108, 352, 71, 455, 468, 60, 301, 299, 191, 300, 302, 239, 243, 110, 111, 158, 378, 223, 343, 165, 453, 291, 532, 364, 370, 22, 193, 336, 132, 236, 163, 2266, 368, 177, 98, 156, 443, 259, 540.
5. Model Number: 722022. No UDI. Serial Number: 6.
6. Model Number: 722023. No UDI. Serial Numbers: 27, 116, 12, 175, 98, 35, 72, 142, 46, 64, 110, 49, 124, 66, 117, 109, 42, 138, 82, 83, 127, 75, 118, 41, 33, 146, 121, 119, 70, 112, 47, 137, 25, 114, 105, 90, 107, 74, 5, 9, 4, 14, 13, 16, 132, 88, 73, 99, 38, 141, 32, 84, 68, 134, 79, 52, 56, 69, 37, 21, 53, 123, 157, 102, 126, 145, 144, 15, 135, 152, 149, 60, 150, 108, 100, 95, 34, 54, 85, 87, 94, 57, 30, 130, 58, 86, 125, 96, 3, 103, 10, 31, 106, 67.
7. Model Number: 722025. No UDI. Serial Numbers: 3, 2, 6.
|
| FEI Number |
3002807880
|
Recalling Firm/ Manufacturer |
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands
|
Manufacturer Reason for Recall | Potential safety issues in Allura Xper R7.6- R8.1 systems where the Lifetime Extension (LTE) kit is not installed in a timely manner. Root cause is under investigation, but the recall reason is listed as for systems over 10 years of life with LTE kits are experiencing propeller motor clamping bolt and x-ray tube locking bolt breaks. If this happens, it may cause erratic C-arm movements and potential collision with patients or bystanders, or detection of a false collisions and subsequent blocked movement of the c-arm. Movement may also cause image noise and black area on the x-ray image, and possible need to expose the patient to additional unnecessary radiation during re-examination. |
FDA Determined Cause 2 | Equipment maintenance |
| Action | Philips Medical Systems notified consignees on 11/11/2024 via letter. Consignees were instructed that affected systems may continue to be used in accordance with their Indications for Use and Instructions for Use (IFU), to keep the Urgent Medical Device Correction letter with the documentation of the system until Philips corrects the system, circulate this notice to all users of the system so that they are aware of the issue, and if you experience erratic movement of the C-arm and/or if you see a portion of black area on the X-ray image, please call your local Philips representative to report the event. Philips will contact all affected customers to arrange for a Field Service Engineer visit to install an updated Lifetime Extension Kit. Consignees were also instructed to notify customers if further distributed and inform Philips about the transfer through the email account: IGT_Recalls@philips.com. Philips requested consignees complete and return the provided response form.
On 11/24/2025, Philips sent a follow up letter to consignees to provide an update to the Urgent Medical Device Correction (MDC) issued on 11/11/2024. The updated letter informed consignees that after further technical investigation on both the hydrogen embrittlement and fatigue conditions, Philips has confirmed that these failure mechanisms do not apply to the bolts in question. Therefore, replacement is not required, even after 10 years of use. As a result, the risks and actions outlined in the original MDC no longer apply to affected systems. |
| Quantity in Commerce | 2,782 units |
| Distribution | Nationwide distribution. International distribution to Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Oman, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = OWB
|
|
|
|