• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall A810 CP app

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall A810 CP appsee related information
Date Initiated by FirmNovember 21, 2024
Date PostedDecember 27, 2024
Recall Status1 Open3, Classified
Recall NumberZ-0774-2025
Recall Event ID 95845
Product Classification Pump, infusion, implanted, programmable - Product Code LKK
ProductSynchroMed II Model 8637 and SynchroMed III Model 8667 implantable pumps used with PROG CT900E CLINICIAN TABLET GLBL SMSNG and PROG CT900E CLINICIAN TABLET PAPER EN with the A810 CP app version 2.x, CFN CT900E
Code Information GTIN: 00763000447090 00763000729141 00763000737290 00763000447090; Serial Numbers: NPL3020725 NPL3021958 NPL3021963 NPL3021966 NPL3021969 NPL3021970 NPL3021971 NPL3021974 NPL3021983 NPL3021984 NPL3021990 NPL3021997 NPL3022002 NPL3022005 NPL3022014 NPL3022022 NPL3022031 NPL3022032 NPL3022038 NPL3022044 NPL3022052 NPL3022062 NPL3022069 NPL3022070 NPL3022071 NPL3022072 NPL3022073 NPL3022074 NPL3022109 NPL3022114 NPL3022118 NPL3022124 NPL3022125 NPL3022131 NPL3022139 NPL3022187 NPL3022190 NPL3022193 NPL3022194 NPL3022195 NPL3022212 NPL3022213 NPL3022216 NPL3022218 NPL3022245 NPL3022257 NPL3022278 NPL3022280 NPL3022289 NPL3022291 NPL3022304 NPL3022311 NPL3022313 NPL3022314 NPL3022328 NPL3022352 NPL3022357 NPL3022364 NPL3022365 NPL3022711 NPL3022713 NPL3022714 NPL3022715 NPL3022717 NPL3022735 NPL3022761 NPL3022762 NPL3022765 NPL3022767 NPL3022768 NPL3022769 NPL3022776 NPL3022791 NPL3022793 NPL3022795 NPL3022810 NPL3022811 NPL3022815 NPL3022817 NPL3022818 NPL3022831 NPL3022836 NPL3022838 NPL3022841 NPL3022844 NPL3022856 NPL3022858 NPL3022869 NPL3022870 NPL3022873 NPL3022874 NPL3022876 NPL3022877 NPL3022879 NPL3022893 NPL3022899 NPL3022900 NPL3022912 NPL3022918 NPL3022922 NPL3022926 NPL3022939 NPL3022940 NPL3022942 NPL3022954 NPL3022957 NPL3022958 NPL3022959 NPL3022968 NPL3022969 NPL3022973 NPL3022974 NPL3022981 NPL3022999 NPL3023030 NPL3023032 NPL3023035 NPL3023038 NPL3023042 NPL3023043 NPL3023048 NPL3023050 NPL3023054 NPL3023059 NPL3023061 NPL3023065 NPL3023067 NPL3023076 NPL3023078 NPL3023086 NPL3023093 NPL3023097 NPL3023123 NPL3023131 NPL3023132 NPL3023133 NPL3023136 NPL3023146 NPL3023148 NPL3023149 NPL3023150 NPL3023153 NPL3023156 NPL3023167 NPL3023175 NPL3023180 NPL3023192 NPL3023194 NPL3023197 NPL3023205 NPL3023218 NPL3023229 NPL3023237 NPL3023239 NPL3023241 NPL3023242 NPL3023245 NPL3023254 NPL3023256 NPL3023257 NPL3023260 NPL3023262 NPL3023268 NPL3023272 NPL3023278 NPL3023288 NPL3023292 NPL3023315 NPL3023317 NPL3023319 NPL3023322 NPL3023323 NPL3023330 NPL3023331 NPL3023340 NPL3023347 NPL3023351 NPL3023352 NPL3023364 NPL3023367 NPL3023368 NPL3023373 NPL3023380 NPL3023388 NPL3023389 NPL3023394 NPL3023409 NPL3023411 NPL3023415 NPL3023422 NPL3023423 NPL3023424 NPL3023425 NPL3023429 NPL3023432 NPL3023435 NPL3023437 NPL3023439 NPL3023440 NPL3023443 NPL3023446 NPL3023447 NPL3023448 NPL3023449 NPL3023450 NPL3023459 NPL3023460 NPL3023463 NPL3023464 NPL3023466 NPL3023467 NPL3023474 NPL3023491 NPL3023494 NPL3023498 NPL3023504 NPL3023510 NPL3023511 NPL3023512 NPL3023520 NPL3023526 NPL3023541 NPL3023548 NPL3023550 NPL3023554 NPL3023555 NPL3023556 NPL3023559 NPL3023563 NPL3023573 NPL3023583 NPL3023593 NPL3023597 NPL3023598 NPL3023599 NPL3023600 NPL3023602 NPL3023607 NPL3023610 NPL3023615 NPL3023620 NPL3023635 NPL3023638 NPL3023640 NPL3023642 NPL3023646 NPL3023647 NPL3023649 NPL3023651 NPL3023659 NPL3023660 NPL3023661 NPL3023664 NPL3023665 NPL3023666 NPL3023689 NPL3023690 NPL3023692 NPL3023695 NPL3023704 NPL3023708 NPL3023711 NPL3023716 NPL3023718 NPL3023722 NPL3023723 NPL3023728 NPL3023729 NPL3023734 NPL3023744 NPL3023751 NPL3023752 NPL3023760 NPL3023762 NPL3023776 NPL3023778 NPL3023781 NPL3023782 NPL3023784 NPL3023785 NPL3023792 NPL3023803 NPL3023809 NPL3023829 NPL3023830 NPL3023831 NPL3023837 NPL3023841 NPL3023852 NPL3023854 NPL3023855 NPL3023857 NPL3023858 NPL3023860 NPL3023868 NPL3023873 NPL3023885 NPL3023902 NPL3023905 NPL3023915 NPL3023922 NPL3023930 NPL3023933 NPL3023936 NPL3023938 NPL3023948 NPL3023949 NPL3023955 NPL3023957 NPL3023959 NPL3023974 NPL3023977 NPL3023980 NPL3023983 NPL3023984 NPL3023990 NPL3023991 NPL3023998 NPL3024013 NPL3024015 NPL3024022 NPL3024023 NPL3024024 NPL3024025 NPL3024026 NPL3024028 NPL3024031 NPL3024033 NPL3024037 NPL3024040 NPL3024043 NPL3024044 NPL3024046 NPL3024054 NPL3024055 NPL3024056 NPL3024059 NPL3024072 NPL3024079 NPL3024082 NPL3024084 NPL3024087 NPL3024088 NPL3024092 NPL3024099 NPL3024100 NPL3024102 NPL3024103 NPL3024108 NPL3024111 NPL3024113 NPL3024117 NPL3024122 NPL3024123 NPL3024124 NPL3024127 NPL3024129 NPL3024139 NPL3024145 NPL3024148 NPL3024149 NPL3024150 NPL3024151 NPL3024152 NPL3024153 NPL3024155 NPL3024158 NPL3024161 NPL3024162 NPL3024166 NPL3024167 NPL3024170 NPL3024176 NPL3024179 NPL3024185 NPL3024188 NPL3024193 NPL3024196 NPL3024197 NPL3024198 NPL3024199 NPL3024200 NPL3024206 NPL3024207 NPL3024208 NPL3024209 NPL3024210 NPL3024212 NPL3024218 NPL3024220 NPL3024221 NPL3024225 NPL3024230 NPL3024233 NPL3024234 NPL3024235 NPL3024239 NPL3024240 NPL3024245 NPL3024246 NPL3024247 NPL3024256 NPL3024257 NPL3024258 NPL3024261 NPL3024266 NPL3024274 NPL3024276 NPL3024277 NPL3024278 NPL3024280 NPL3024284 NPL3024294 NPL3024296 NPL3024298 NPL3024299 NPL3024300 NPL3024303 NPL3024305 NPL3024307 NPL3024308 NPL3024309 NPL3024310 NPL3024311 NPL3024312 NPL3024313 NPL3024316 NPL3024321 NPL3024324 NPL3024327 NPL3024328 NPL3024329 NPL3024337 NPL3024341 NPL3024348 NPL3024350 NPL3024355 NPL3024356 NPL3024358 NPL3024360 NPL3024367 NPL3024370 NPL3024379 NPL3024381 NPL3024384 NPL3024393 NPL3024395 NPL3024399 NPL3024404 NPL3024405 NPL3024407 NPL3024408 NPL3024416 NPL3024417 NPL3024418 NPL3024419 NPL3024420 NPL3024424 NPL3024428 NPL3024430 NPL3024437 NPL3024439 NPL3024440 NPL3024442 NPL3024452 NPL3024455 NPL3024461 NPL3024462 NPL3024464 NPL3024472 NPL3024478 NPL3024479 NPL3024480 NPL3024482 NPL3024484 NPL3024485 NPL3024486 NPL3024487 NPL3024490 NPL3024491 NPL3024492 NPL3024493 NPL3024498 NPL3024499 NPL3024501 NPL3024502 NPL3024507 NPL3024509 NPL3024518 NPL3024519 NPL3024523 NPL3024531 NPL3024533 NPL3024534 NPL3024540 NPL3024541 NPL3024542 NPL3024544 NPL3024545 NPL3024549 NPL3024550 NPL3024551 NPL3024552 NPL3024554 NPL3024555 NPL3024556 NPL3024557 NPL3024558 NPL3024559 NPL3024563 NPL3024565 NPL3024567 NPL3024569 NPL3024572 NPL3024574 NPL3024576 NPL3024578 NPL3024582 NPL3024583 NPL3024585 NPL3024587 NPL3024589 NPL3024592 NPL3024594 NPL3024596 NPL3024598 NPL3024600 NPL3024602 NPL3024603 NPL3024604 NPL3024606 NPL3024608 NPL3024610 NPL3024615 NPL3024618 NPL3024619 NPL3024621 NPL3024625 NPL3024626 NPL3024631 NPL3024632 NPL3024635 NPL3024637 NPL3024639 NPL3024642 NPL3024643 NPL3024647 NPL3024649 NPL3024654 NPL3024656 NPL3024657 NPL3024658 NPL3024659 NPL3024660 NPL3024661 NPL3024662 NPL3024664 NPL3024665 NPL3024669 NPL3024671 NPL3024673 NPL3024676 NPL3024678 NPL3024680 NPL3024688 NPL3024694 NPL3024702 NPL3024709 NPL3024712 NPL3024717 NPL3024721 NPL3024722 NPL3024726 NPL3024728 NPL3024729 NPL3024730 NPL3024731 NPL3024734 NPL3024736 NPL3024737 NPL3024739 NPL3024740 NPL3024748 NPL3024750 NPL3024751 NPL3024755 NPL3024758 NPL3024759 NPL3024764 NPL3024765 NPL3024769 NPL3024771 NPL3024772 NPL3024774 NPL3024775 NPL3024777 NPL3024778 NPL3024780 NPL3024782 NPL3024784 NPL3024785 NPL3024786 NPL3024787 NPL3024794 NPL3024797 NPL3024799 NPL3024800 NPL3024803 NPL3024807 NPL3024808 NPL3024817 NPL3024819 NPL3024821 NPL3024823 NPL3024827 NPL3024834 NPL3024838 NPL3024843 NPL3024851 NPL3024857 NPL3024858 NPL3024859 NPL3024861 NPL3024867 NPL3024869 NPL3024870 NPL3024872 NPL3024875 NPL3024884 NPL3024890 NPL3024896 NPL3024897 NPL3024899 NPL3024903 NPL3024905 NPL3024906 NPL3024909 NPL3024919 NPL3024921 NPL3024929 NPL3024933 NPL3024934 NPL3024935 NPL3024936 NPL3024939 NPL3024940 NPL3024943 NPL3024945 NPL3024950 NPL3024951 NPL3024953 NPL3024954 NPL3024956 NPL3024964 NPL3024970 NPL3024971 NPL3024972 NPL3024975 NPL3024980 NPL3024982 NPL3024983 NPL3024985 NPL3024988 NPL3024993 NPL3024995 NPL3024997 NPL3024998 NPL3024999 NPL3025002 NPL3025011 NPL3025021 NPL3025024 NPL3025025 NPL3025026 NPL3025047 NPL3025048 NPL3025057 NPL3025058 NPL3025060 NPL3025062 NPL3025068 NPL3025070 NPL3025074 NPL3025084 NPL3025090 NPL3025097 NPL3025103 NPL3025105 NPL3025117 NPL3025119 NPL3025132 NPL3025133 NPL3025146 NPL3025153 NPL3025158 NPL3025160 NPL3025164 NPL3025167 NPL3025169 NPL3025179 NPL3025180 NPL3025185 NPL3025191 NPL3025197 NPL3025207 NPL3025217 NPL3025224 NPL3025227 NPL3025228 NPL3025231 NPL3025234 NPL3025237 NPL3025240 NPL3025241 NPL3025244 NPL3025249 NPL3025252 NPL3025253 NPL3025257 NPL3025263 NPL3025265 NPL3025270 NPL3025271 NPL3025275 NPL3025281 NPL3025289 NPL3025300 NPL3025302 NPL3025304 NPL3025305 NPL3025309 NPL3025312 NPL3025313 NPL3025320 NPL3025322 NPL3025327 NPL3025329 NPL3025331 NPL3025332 NPL3025335 NPL3025338 NPL3025339 NPL3025344 NPL3025354 NPL3025356 NPL3025370 NPL3025382 NPL3025385 NPL3025386 NPL3025388 NPL3025392 NPL3025394 NPL3025397 NPL3025409 NPL3025410 NPL3025419 NPL3025424 NPL3025433 NPL3025436 NPL3025439 NPL3025441 NPL3025446 NPL3025448 NPL3025449 NPL3025451 NPL3025452 NPL3025454 NPL3025458 NPL3025461 NPL3025462 NPL3025465 NPL3025466 NPL3025468 NPL3025470 NPL3025477 NPL3025479 NPL3025480 NPL3025482 NPL3025483 NPL3025490 NPL3025494 NPL3025502 NPL3025506 NPL3025508 NPL3025509 NPL3025513 NPL3025514 NPL3025515 NPL3025518 NPL3025526 NPL3025527 NPL3025530 NPL3025532 NPL3025533 NPL3025534 NPL3025535 NPL3025538 NPL3025542 NPL3025547 NPL3025549 NPL3025558 NPL3025562 NPL3025563 NPL3025572 NPL3025576 NPL3025577 NPL3025580 NPL3025584 NPL3025588 NPL3025589 NPL3025591 NPL3025595 NPL3025596 NPL3025598 NPL3025600 NPL3025601 NPL3025602 NPL3025606 NPL3025613 NPL3025617 NPL3025620 NPL3025621 NPL3025622 NPL3025624 NPL3025625 NPL3025629 NPL3025633 NPL3025638 NPL3025642 NPL3025644 NPL3025646 NPL3025649 NPL3025650 NPL3025652 NPL3025653 NPL3025661 NPL3025667 NPL3025669 NPL3025673 NPL3025674 NPL3025676 NPL3025677 NPL3025686 NPL3025689 NPL3025690 NPL3025691 NPL3025697 NPL3025703 NPL3025704 NPL3025707 NPL3025712 NPL3025713 NPL3025723 NPL3025725 NPL3025727 NPL3025739 NPL3025740 NPL3025741 NPL3025742 NPL3025745 NPL3025751 NPL3025754 NPL3025755 NPL3025757 NPL3025758 NPL3025775 NPL3025776 NPL3025777 NPL3025787 NPL3025795 NPL3025797 NPL3025799 NPL3025800 NPL3025807 NPL3025811 NPL3025813 NPL3025815 NPL3025816 NPL3025819 NPL3025820 NPL3025823 NPL3025825 NPL3025826 NPL3025827 NPL3025828 NPL3025832 NPL3025834 NPL3025837 NPL3025838 NPL3025839 NPL3025842 NPL3025845 NPL3025849 NPL3025850 NPL3025853 NPL3025855 NPL3025873 NPL3025879 NPL3025881 NPL3025882 NPL3025892 NPL3025893 NPL3025895 NPL3025900 NPL3025901 NPL3025908 NPL3025909 NPL3025916 NPL3025918 NPL3025921 NPL3025929 NPL3025932 NPL3025937 NPL3025938 NPL3025939 NPL3025946 NPL3025947 NPL3025949 NPL3025951 NPL3025953 NPL3025955 NPL3025963 NPL3025964 NPL3025965 NPL3025967 NPL3025975 NPL3025981 NPL3025998 NPL3025999 NPL3026005 NPL3026008 NPL3026009 NPL3026010 NPL3026012 NPL3026017 NPL3026019 NPL3026020 NPL3026021 NPL3026035 NPL3026038 NPL3026044 NPL3026050 NPL3026052 NPL3026053 NPL3026054 NPL3026057 NPL3026060 NPL3026061 NPL3026066 NPL3026068 NPL3026071 NPL3026076 NPL3026077 NPL3026081 NPL3026085 NPL3026088 NPL3026092 NPL3026094 NPL3026097 NPL3026099 NPL3026101 NPL3026102 NPL3026104 NPL3026108 NPL3026109 NPL3026111 NPL3026118 NPL3026120 NPL3026121 NPL3026123 NPL3026127 NPL3026134 NPL3026141 NPL3026142 NPL3026143 NPL3026144 NPL3026147 NPL3026148 NPL3026150 NPL3026151 NPL3026152 NPL3026153 NPL3026154 NPL3026155 NPL3026159 NPL3026161 NPL3026163 NPL3026165 NPL3026168 NPL3026169 NPL3026172 NPL3026173 NPL3026177 NPL3026178 NPL3026183 NPL3026184 NPL3026185 NPL3026187 NPL3026192 NPL3026193 NPL3026195 NPL3026196 NPL3026197 NPL3026198 NPL3026199 NPL3026201 NPL3026202 NPL3026203 NPL3026204 NPL3026206 NPL3026207 NPL3026208 NPL3026209 NPL3026210 NPL3026212 NPL3026215 NPL3026216 NPL3026217 NPL3026218 NPL3026221 NPL3026223 NPL3026224 NPL3026229 NPL3026233 NPL3026238 NPL3026241 NPL3026244 NPL3026245 NPL3026250 NPL3026258 NPL3026259 NPL3026261 NPL3026262 NPL3026263 NPL3026264 NPL3026267 NPL3026270 NPL3026271 NPL3026272 NPL3026273 NPL3026274 NPL3026275 NPL3026286 NPL3026289 NPL3026291 NPL3026292 NPL3026296 NPL3026300 NPL3026301 NPL3026303 NPL3026307 NPL3026313 NPL3026315 NPL3026317 NPL3026319 NPL3026321 NPL3026324 NPL3026325 NPL3026333 NPL3026334 NPL3026335 NPL3026337 NPL3026338 NPL3026343 NPL3026350 NPL3026351 NPL3026352 NPL3026356 NPL3026357 NPL3026358 NPL3026360 NPL3026362 NPL3026367 NPL3026374 NPL3026381 NPL3026382 NPL3026383 NPL3026384 NPL3026390 NPL3026391 NPL3026393 NPL3026396 NPL3026397 NPL3026398 NPL3026402 NPL3026405 NPL3026406 NPL3026407 NPL3026408 NPL3026414 NPL3026421 NPL3026422 NPL3026423 NPL3026425 NPL3026427 NPL3026428 NPL3026430 NPL3026432 NPL3026435 NPL3026438 NPL3026440 NPL3026441 NPL3026442 NPL3026444 NPL3026445 NPL3026448 NPL3026451 NPL3026452 NPL3027458 NPL3027462 NPL3027463 NPL3027464 NPL3027465 NPL3027467 NPL3027477 NPL3027480 NPL3027482 NPL3027487 NPL3027496 NPL3027516 NPL3027518 NPL3027520 NPL3027522 NPL3027526 NPL3027535 NPL3027538 NPL3027545 NPL3027549 NPL3027571 NPL3027592 NPL3027595 NPL3027601 NPL3027604 NPL3027606 NPL3027607 NPL3027610 NPL3027611 NPL3027614 NPL3027620 NPL3027627 NPL3027631 NPL3027634 NPL3027635 NPL3027636 NPL3027651 NPL3027652 NPL3027656 NPL3027658 NPL3027661 NPL3027664 NPL3027674 NPL3027676 NPL3027678 NPL3027680 NPL3027681 NPL3027682 NPL3027684 NPL3027685 NPL3027686 NPL3027693 NPL3027695 NPL3027710 NPL3027712 NPL3027714 NPL3027716 NPL3027726 NPL3027727 NPL3027731 NPL3027732 NPL3027733 NPL3027734 NPL3027736 NPL3027737 NPL3027745 NPL3027753 NPL3027760 NPL3027767 NPL3027770 NPL3027773 NPL3027774 NPL3027777 NPL3027779 NPL3027780 NPL3027787 NPL3027790 NPL3027791 NPL3027796 NPL3027801 NPL3027803 NPL3027810 NPL3027816 NPL3027820 NPL3027823 NPL3027829 NPL3027830 NPL3027832 NPL3027834 NPL3027836 NPL3027839 NPL3027840 NPL3027841 NPL3027843 NPL3027844 NPL3027846 NPL3027848 NPL3027850 NPL3027851 NPL3027852 NPL3027853 NPL3027856 NPL3027857 NPL3027863 NPL3027864 NPL3027865 NPL3030786 NPL3030794 NPL3030796 NPL3030799 NPL3030806 NPL3030807 NPL3030809 NPL3030811 NPL3030814 NPL3030817 NPL3030820 NPL3030824 NPL3030825 NPL3030836 NPL3030846 NPL3030848 NPL3030860 NPL3030865 NPL3030867 NPL3030869 NPL3030870 NPL3030873 NPL3030875 NPL3030879 NPL3030880 NPL3030882 NPL3030884 NPL3030889 NPL3030893 NPL3030900 NPL3030901 NPL3030902 NPL3030904 NPL3030905 NPL3030908 NPL3030913 NPL3030916 NPL3030917 NPL3030926 NPL3030929 NPL3030942 NPL3030943 NPL3030950 NPL3030951 NPL3030954 NPL3030955 NPL3030962 NPL3030964 NPL3030971 NPL3030978 NPL3030979 NPL3030983 NPL3030987 NPL3030989 NPL3030990 NPL3030992 NPL3030996 NPL3031005 NPL3031010 NPL3031011 NPL3031012 NPL3031013 NPL3031014 NPL3031016 NPL3031018 NPL3031019 NPL3031022 NPL3031024 NPL3031026 NPL3031027 NPL3031030 NPL3031031 NPL3031032 NPL3031035 NPL3031039 NPL3031041 NPL3031042 NPL3031043 NPL3031047 NPL3031053 NPL3031056 NPL3031058 NPL3031060 NPL3031061 NPL3031062 NPL3031063 NPL3031074 NPL3031075 NPL3031079 NPL3031080 NPL3031081 NPL3031082 NPL3031083 NPL3031084 NPL3031085 NPL3031087 NPL3031088 NPL3031089 NPL3031090 NPL3031091 NPL3031092 NPL3031095 NPL3031096 NPL3031097 NPL3031099 NPL3031100 NPL3031101 NPL3031102 NPL3031103 NPL3031109 NPL3031113 NPL3031117 NPL3031118 NPL3031122 NPL3031123 NPL3031124 NPL3031127 NPL3031128 NPL3031134 NPL3031135 NPL3031136 NPL3031137 NPL3031141 NPL3031144 NPL3031148 NPL3031150 NPL3031151 NPL3031152 NPL3031153 NPL3031154 NPL3031156 NPL3031158 NPL3031159 NPL3031160 NPL3031161 NPL3031165 NPL3031166 NPL3031171 NPL3031172 NPL3031177 NPL3031178 NPL3031180 NPL3031183 NPL3031187 NPL3031188 NPL3031190 NPL3031191 NPL3031193 NPL3031194 NPL3031197 NPL3031198 NPL3031200 NPL3031201 NPL3031209 NPL3031212 NPL3031215 NPL3031220 NPL3031222 NPL3031225 NPL3031227 NPL3031228 NPL3031231 NPL3031232 NPL3031233 NPL3031239 NPL3031240 NPL3031241 NPL3031242 NPL3031244 NPL3031245 NPL3031247 NPL3031248 NPL3031250 NPL3031251 NPL3031252 NPL3031257 NPL3031258 NPL3031259 NPL3031261 NPL3031265 NPL3031266 NPL3031267 NPL3031273 NPL3031277 NPL3031278 NPL3031279 NPL3031281 NPL3031282 NPL3031283 NPL3031284 NPL3031290 NPL3031292 NPL3031294 NPL3031296 NPL3031300 NPL3031305 NPL3031306 NPL3031307 NPL3031308 NPL3031309 NPL3031310 NPL3031311 NPL3031312 NPL3031313 NPL3031314 NPL3031315 NPL3031317 NPL3031320 NPL3031321 NPL3031323 NPL3031331 NPL3031334 NPL3031336 NPL3031339 NPL3031349 NPL3031352 NPL3031354 NPL3031357 NPL3031358 NPL3031359 NPL3031361 NPL3031362 NPL3031363 NPL3031366 NPL3031367 NPL3031368 NPL3031369 NPL3031371 NPL3031373 NPL3031374 NPL3031375 NPL3031377 NPL3031378 NPL3031379 NPL3031380 NPL3031383 NPL3031387 NPL3031388 NPL3031393 NPL3031395 NPL3031396 NPL3031397 NPL3031402 NPL3031403 NPL3031404 NPL3031405 NPL3031407 NPL3031408 NPL3031413 NPL3031416 NPL3031419 NPL3031420 NPL3031421 NPL3031423 NPL3031428 NPL3031429 NPL3031430 NPL3031431 NPL3031432 NPL3031433 NPL3031437 NPL3031444 NPL3031447 NPL3031448 NPL3031451 NPL3031452 NPL3031453 NPL3031456 NPL3031459 NPL3031461 NPL3031463 NPL3031466 NPL3031467 NPL3031471 NPL3031472 NPL3031473 NPL3031475 NPL3031477 NPL3031478 NPL3031482 NPL3031483 NPL3031487 NPL3031488 NPL3031491 NPL3031495 NPL3031496 NPL3031500 NPL3031501 NPL3031504 NPL3031508 NPL3031509 NPL3031511 NPL3031514 NPL3031515 NPL3031518 NPL3031521 NPL3031522 NPL3031525 NPL3031529 NPL3031531 NPL3031534 NPL3031537 NPL3031538 NPL3031540 NPL3031544 NPL3031546 NPL3031547 NPL3031548 NPL3031550 NPL3031551 NPL3031555 NPL3031558 NPL3031562 NPL3031563 NPL3031566 NPL3031569 NPL3031570 NPL3031575 NPL3031577 NPL3031582 NPL3031585 NPL3031587 NPL3031591 NPL3031592 NPL3031593 NPL3031594 NPL3031595 NPL3031596 NPL3031597 NPL3031600 NPL3031602 NPL3031603 NPL3031604 NPL3031606 NPL3031608 NPL3031609 NPL3031612 NPL3031614 NPL3031620 NPL3031624 NPL3031627 NPL3031632 NPL3031635 NPL3031638 NPL3031640 NPL3031641 NPL3031644 NPL3031646 NPL3031648 NPL3031649 NPL3031650 NPL3031651 NPL3031653 NPL3031654 NPL3031657 NPL3031658 NPL3031659 NPL3031661 NPL3031665 NPL3031670 NPL3031671 NPL3031672 NPL3031674 NPL3031676 NPL3031679 NPL3031680 NPL3031687 NPL3031690 NPL3031693 NPL3031694 NPL3031695 NPL3031696 NPL3031698 NPL3031700 NPL3031701 NPL3031702 NPL3031705 NPL3031708 NPL3031709 NPL3031710 NPL3031713 NPL3031718 NPL3031725 NPL3031727 NPL3031735 NPL3031737 NPL3031740 NPL3031742 NPL3031743 NPL3031744 NPL3031746 NPL3031750 NPL3031758 NPL3031759 NPL3031760 NPL3031765 NPL3031767 NPL3031773 NPL3031775 NPL3031777 NPL3031782 NPL3031786 NPL3031790 NPL3031792 NPL3031795 NPL3031796 NPL3031805 NPL3031807 NPL3031809 NPL3031815 NPL3031816 NPL3031817 NPL3031819 NPL3031823 NPL3031824 NPL3031826 NPL3031830 NPL3031832 NPL3031840 NPL3031848 NPL3031849 NPL3031852 NPL3031859 NPL3031863 NPL3031864 NPL3031865 NPL3031868 NPL3031870 NPL3031872 NPL3031873 NPL3031877 NPL3031880 NPL3031881 NPL3031883 NPL3031884 NPL3031895 NPL3031900 NPL3031902 NPL3031905 NPL3031907 NPL3031911 NPL3031912 NPL3031913 NPL3031914 NPL3031920 NPL3031923 NPL3031927 NPL3031930 NPL3031931 NPL3031932 NPL3031938 NPL3031947 NPL3031950 NPL3031951 NPL3031952 NPL3031953 NPL3031954 NPL3031957 NPL3031964 NPL3031968 NPL3031971 NPL3031975 NPL3031979 NPL3031986 NPL3031987 NPL3031988 NPL3031993 NPL3031994 NPL3031995 NPL3032001 NPL3032005 NPL3032007 NPL3032009 NPL3032015 NPL3032017 NPL3032018 NPL3032019 NPL3032021 NPL3032023 NPL3032024 NPL3032025 NPL3032028 NPL3032031 NPL3032033 NPL3032035 NPL3032036 NPL3032037 NPL3032038 NPL3032041 NPL3032046 NPL3032049 NPL3032051 NPL3032054 NPL3032056 NPL3032058 NPL3032061 NPL3032062 NPL3032067 NPL3032068 NPL3032069 NPL3032072 NPL3032074 NPL3032076 NPL3032077 NPL3032078 NPL3032079 NPL3032081 NPL3032085 NPL3032087 NPL3032088 NPL3032089 NPL3032094 NPL3032095 NPL3032104 NPL3032106 NPL3032107 NPL3032109 NPL3032110 NPL3032111 NPL3032112 NPL3032117 NPL3032118 NPL3032120 NPL3032122 NPL3032123 NPL3032124 NPL3032135 NPL3032136 NPL3032143 NPL3032144 NPL3032145 NPL3032146 NPL3032150 NPL3032151 NPL3032153 NPL3032154 NPL3032163 NPL3032165 NPL3032167 NPL3032168 NPL3032173 NPL3032180 NPL3032183 NPL3032187 NPL3032190 NPL3032192 NPL3032193 NPL3032195 NPL3032200 NPL3032201 NPL3032202 NPL3032206 NPL3032207 NPL3032210 NPL3032212 NPL3032214 NPL3032215 NPL3032217 NPL3032218 NPL3032219 NPL3032223 NPL3032227 NPL3032229 NPL3032231 NPL3032232 NPL3032236 NPL3032237 NPL3032238 NPL3032239 NPL3032241 NPL3032243 NPL3032244 NPL3032246 NPL3032250 NPL3032251 NPL3032256 NPL3032261 NPL3032262 NPL3032263 NPL3032266 NPL3032268 NPL3032273 NPL3032277 NPL3032281 NPL3032283 NPL3032286 NPL3032291 NPL3032296 NPL3032308 NPL3032310 NPL3032311 NPL3032320 NPL3032328 NPL3032330 NPL3032344 NPL3032346 NPL3032347 NPL3032348 NPL3032356 NPL3032360 NPL3032361 NPL3032366 NPL3032373 NPL3032375 NPL3032377 NPL3032378 NPL3032380 NPL3032382 NPL3032383 NPL3032384 NPL3032385 NPL3032387 NPL3032390 NPL3032399 NPL3032401 NPL3032403 NPL3032407 NPL3032408 NPL3032410 NPL3032416 NPL3032418 NPL3032420 NPL3032421 NPL3032424 NPL3032426 NPL3032428 NPL3032435 NPL3032436 NPL3032437 NPL3032440 NPL3032441 NPL3032442 NPL3032450 NPL3032451 NPL3032453 NPL3032454 NPL3032455 NPL3032456 NPL3032457 NPL3032458 NPL3032462 NPL3032463 NPL3032465 NPL3032467 NPL3032470 NPL3032471 NPL3032474 NPL3032482 NPL3032483 NPL3032485 NPL3032486 NPL3032488 NPL3032489 NPL3032491 NPL3032492 NPL3032493 NPL3032495 NPL3032496 NPL3032497 NPL3032500 NPL3032501 NPL3032505 NPL3032507 NPL3032510 NPL3032512 NPL3032516 NPL3032522 NPL3032523 NPL3032530 NPL3032531 NPL3032532 NPL3032536 NPL3032547 NPL3032548 NPL3032549 NPL3032550 NPL3032551 NPL3032556 NPL3032557 NPL3032560 NPL3032562 NPL3032566 NPL3032568 NPL3032572 NPL3032573 NPL3032574 NPL3032577 NPL3032579 NPL3032583 NPL3032587 NPL3032588 NPL3032591 NPL3032592 NPL3032594 NPL3032595 NPL3032597 NPL3032598 NPL3032599 NPL3032601 NPL3032602 NPL3032603 NPL3032609 NPL3032610 NPL3032611 NPL3032612 NPL3032613 NPL3032614 NPL3032616 NPL3032620 NPL3032621 NPL3032622 NPL3032623 NPL3032624 NPL3032626 NPL3032627 NPL3032628 NPL3032629 NPL3032630 NPL3032632 NPL3032633 NPL3032640 NPL3032641 NPL3032642 NPL3032648 NPL3032650 NPL3032651 NPL3032652 NPL3032656 NPL3032660 NPL3032664 NPL3032665 NPL3032666 NPL3032668 NPL3032672 NPL3032681 NPL3032683 NPL3032687 NPL3032688 NPL3032689 NPL3032690 NPL3032692 NPL3032693 NPL3032696 NPL3032697 NPL3032701 NPL3032702 NPL3032703 NPL3032706 NPL3032709 NPL3032714 NPL3032715 NPL3032717 NPL3032718 NPL3032719 NPL3032722 NPL3032724 NPL3032727 NPL3032728 NPL3032729 NPL3032735 NPL3032736 NPL3032739 NPL3032740 NPL3032741 NPL3032742 NPL3032745 NPL3032746 NPL3032747 NPL3032748 NPL3032750 NPL3032756 NPL3032759 NPL3032761 NPL3032762 NPL3032763 NPL3032764 NPL3032770 NPL3032772 NPL3032773 NPL3032774 NPL3032776 NPL3032777 NPL3032778 NPL3032779 NPL3032781 NPL3032782 NPL3032783 NPL3032784 NPL3032790 NPL3032791 NPL3032797 NPL3032798 NPL3032799 NPL3032803 NPL3032806 NPL3032807 NPL3032810 NPL3032814 NPL3032816 NPL3032818 NPL3032819 NPL3032821 NPL3032823 NPL3032824 NPL3032826 NPL3032828 NPL3032833 NPL3032835 NPL3032837 NPL3032838 NPL3032839 NPL3032848 NPL3032850 NPL3032856 NPL3032858 NPL3032860 NPL3032862 NPL3032863 NPL3032867 NPL3032869 NPL3032871 NPL3032872 NPL3032876 NPL3032877 NPL3032878 NPL3032879 NPL3032880 NPL3032882 NPL3032883 NPL3032885 NPL3032888 NPL3032891 NPL3032892 NPL3032893 NPL3032895 NPL3032896 NPL3032900 NPL3032904 NPL3032906 NPL3032907 NPL3032909 NPL3032911 NPL3032915 NPL3032916 NPL3032918 NPL3032920 NPL3032921 NPL3032922 NPL3032926 NPL3032927 NPL3032929 NPL3032931 NPL3032932 NPL3032934 NPL3032938 NPL3032939 NPL3032942 NPL3032944 NPL3032945 NPL3032949 NPL3032952 NPL3032953 NPL3032954 NPL3032959 NPL3032965 NPL3032968 NPL3032970 NPL3032973 NPL3032975 NPL3032978 NPL3032980 NPL3032982 NPL3032983 NPL3032985 NPL3032986 NPL3032989 NPL3032991 NPL3032995 NPL3032996 NPL3032997 NPL3032998 NPL3033001 NPL3033003 NPL3033007 NPL3033008 NPL3033011 NPL3033015 NPL3033017 NPL3033019 NPL3033020 NPL3033022 NPL3033023 NPL3033025 NPL3033026 NPL3033027 NPL3033029 NPL3033030 NPL3033032 NPL3033035 NPL3033036 NPL3033037 NPL3033038 NPL3033039 NPL3033040 NPL3033042 NPL3033043 NPL3033046 NPL3033047 NPL3033048 NPL3033049 NPL3033051 NPL3033052 NPL3033054 NPL3033057 NPL3033059 NPL3033062 NPL3033064 NPL3033066 NPL3033067 NPL3033068 NPL3033069 NPL3033070 NPL3033071 NPL3033072 NPL3033073 NPL3033077 NPL3033079 NPL3033080 NPL3033081 NPL3033087 NPL3033088 NPL3033089 NPL3033094 NPL3033095 NPL3033096 NPL3033100 NPL3033102 NPL3033105 NPL3033106 NPL3033107 NPL3033109 NPL3033111 NPL3033113 NPL3033114 NPL3033115 NPL3033116 NPL3033122 NPL3033123 NPL3033124 NPL3033127 NPL3033130 NPL3033132 NPL3033133 NPL3033136 NPL3033137 NPL3033138 NPL3033139 NPL3033140 NPL3033142 NPL3033144 NPL3033145 NPL3033148 NPL3033150 NPL3033151 NPL3033155 NPL3033157 NPL3033158 NPL3033164 NPL3033165 NPL3033166 NPL3033167 NPL3033169 NPL3033172 NPL3033173 NPL3033174 NPL3033182 NPL3033185 NPL3033186 NPL3033187 NPL3033190 NPL3033191 NPL3033192 NPL3033194 NPL3033196 NPL3033199 NPL3033200 NPL3033201 NPL3033203 NPL3033205 NPL3033206 NPL3033207 NPL3033208 NPL3033210 NPL3033211 NPL3033216 NPL3033217 NPL3033219 NPL3033220 NPL3033221 NPL3033222 NPL3033225 NPL3033227 NPL3033228 NPL3033230 NPL3033233 NPL3033235 NPL3033237 NPL3033238 NPL3033239 NPL3033241 NPL3033242 NPL3033243 NPL3033244 NPL3033245 NPL3033247 NPL3033248 NPL3033249 NPL3033250 NPL3033253 NPL3033254 NPL3033256 NPL3033258 NPL3033259 NPL3033260 NPL3033262 NPL3033265 NPL3033266 NPL3033268 NPL3033270 NPL3033275 NPL3033278 NPL3033279 NPL3033282 NPL3033283 NPL3033284 NPL3033285 NPL3033286 NPL3033289 NPL3033290 NPL3033292 NPL3033293 NPL3033295 NPL3033298 NPL3033300 NPL3033302 NPL3033305 NPL3033306 NPL3033307 NPL3033308 NPL3033311 NPL3033318 NPL3033322 NPL3033323 NPL3033327 NPL3033330 NPL3033334 NPL3033338 NPL3033342 NPL3033345 NPL3033347 NPL3033348 NPL3033349 NPL3033351 NPL3033354 NPL3033357 NPL3033358 NPL3033362 NPL3033363 NPL3033364 NPL3033365 NPL3033368 NPL3033370 NPL3033371 NPL3033372 NPL3033373 NPL3033375 NPL3033376 NPL3033382 NPL3033383 NPL3033384 NPL3033386 NPL3033388 NPL3033390 NPL3033391 NPL3033396 NPL3033401 NPL3033407 NPL3033411 NPL3033415 NPL3033418 NPL3033426 NPL3033428 NPL3033429 NPL3033430 NPL3033432 NPL3033434 NPL3033439 NPL3033440 NPL3033442 NPL3033451 NPL3033463 NPL3033465 NPL3033467 NPL3033469 NPL3033471 NPL3033484 NPL3033488 NPL3033490 NPL3033492 NPL3033493 NPL3033498 NPL3033499 NPL3033500 NPL3033502 NPL3033504 NPL3033506 NPL3033508 NPL3033512 NPL3033513 NPL3033514 NPL3033519 NPL3033521 NPL3033523 NPL3033526 NPL3033528 NPL3033529 NPL3033530 NPL3033531 NPL3033532 NPL3033533 NPL3033535 NPL3033536 NPL3033537 NPL3033542 NPL3033547 NPL3033554 NPL3033557 NPL3033559 NPL3033560 NPL3033566 NPL3033567 NPL3033568 NPL3033573 NPL3033581 NPL3033583 NPL3033585 NPL3033588 NPL3033594 NPL3033598 NPL3033602 NPL3033604 NPL3033606 NPL3033609 NPL3033610 NPL3033611 NPL3033614 NPL3033616 NPL3033619 NPL3033620 NPL3033622 NPL3033623 NPL3033624 NPL3033625 NPL3033628 NPL3033629 NPL3033635 NPL3033638 NPL3033640 NPL3033641 NPL3033644 NPL3033645 NPL3033651 NPL3033653 NPL3033656 NPL3033659 NPL3033661 NPL3033663 NPL3033667 NPL3033673 NPL3033678 NPL3033680 NPL3033681 NPL3033685 NPL3033687 NPL3033696 NPL3033698 NPL3033699 NPL3033700 NPL3033706 NPL3033708 NPL3033716 NPL3033717 NPL3033718 NPL3033719 NPL3033721 NPL3033725 NPL3033726 NPL3033729 NPL3033731 NPL3033732 NPL3033733 NPL3033734 NPL3033735 NPL3033736 NPL3033739 NPL3033741 NPL3033742 NPL3033743 NPL3033746 NPL3033749 NPL3033750 NPL3033751 NPL3033753 NPL3033755 NPL3033756 NPL3033758 NPL3033760 NPL3033761 NPL3033762 NPL3033764 NPL3033765 NPL3033767 NPL3033768 NPL3033772 NPL3033775 NPL3033776 NPL3033777 NPL3033778 NPL3033779 NPL3033785 NPL3033786 NPL3033787 NPL3033790 NPL3033791 NPL3033796 NPL3033800 NPL3033802 NPL3033803 NPL3033805 NPL3033809 NPL3033811 NPL3033812 NPL3033813 NPL3033815 NPL3033816 NPL3033818 NPL3033822 NPL3033823 NPL3033824 NPL3033825 NPL3033828 NPL3033829 NPL3033830 NPL3033834 NPL3033836 NPL3033838 NPL3033839 NPL3033840 NPL3033841 NPL3033843 NPL3033844 NPL3033846 NPL3033849 NPL3033854 NPL3033857 NPL3033858 NPL3033859 NPL3033861 NPL3033862 NPL3033865 NPL3033868 NPL3033869 NPL3033872 NPL3033873 NPL3033874 NPL3033878 NPL3033879 NPL3033882 NPL3033883 NPL3033887 NPL3033888 NPL3033891 NPL3033893 NPL3033899 NPL3033900 NPL3033901 NPL3033902 NPL3033903 NPL3033908 NPL3033909 NPL3033911 NPL3033912 NPL3033913 NPL3033914 NPL3033916 NPL3033917 NPL3033918 NPL3033919 NPL3033922 NPL3033923 NPL3033925 NPL3033926 NPL3033927 NPL3033928 NPL3033930 NPL3033934 NPL3033935 NPL3033937 NPL3033940 NPL3033943 NPL3033944 NPL3033945 NPL3033948 NPL3033949 NPL3033954 NPL3033955 NPL3033956 NPL3033958 NPL3033959 NPL3033960 NPL3033962 NPL3033964 NPL3033966 NPL3033968 NPL3033975 NPL3033977 NPL3033978 NPL3033979 NPL3033980 NPL3033981 NPL3033983 NPL3033984 NPL3033987 NPL3033990 NPL3033991 NPL3033992 NPL3033993 NPL3033994 NPL3033995 NPL3033996 NPL3034000 NPL3034002 NPL3034004 NPL3034005 NPL3034007 NPL3034008 NPL3034009 NPL3034011 NPL3034013 NPL3034017 NPL3034018 NPL3034021 NPL3034022 NPL3034024 NPL3034026 NPL3034030 NPL3034032 NPL3034034 NPL3034035 NPL3034036 NPL3034039 NPL3034040 NPL3034042 NPL3034043 NPL3034044 NPL3034045 NPL3034046 NPL3034047 NPL3034048 NPL3034049 NPL3034050 NPL3034052 NPL3034054 NPL3034055 NPL3034056 NPL3034057 NPL3034058 NPL3034060 NPL3034062 NPL3034064 NPL3034065 NPL3034066 NPL3034067 NPL3034068 NPL3034072 NPL3034074 NPL3034075 NPL3034076 NPL3034078 NPL3034079 NPL3034080 NPL3034081 NPL3034083 NPL3034084 NPL3034085 NPL3034086 NPL3034088 NPL3034089 NPL3034090 NPL3034091 NPL3034092 NPL3034093 NPL3034095 NPL3034096 NPL3034097 NPL3034098 NPL3034100 NPL3034102 NPL3034103 NPL3034104 NPL3034105 NPL3034107 NPL3034108 NPL3034109 NPL3034111 NPL3034112 NPL3034115 NPL3034116 NPL3034119 NPL3034120 NPL3034121 NPL3034123 NPL3034124 NPL3034125 NPL3034126 NPL3034127 NPL3034128 NPL3034130 NPL3034131 NPL3034133 NPL3034134 NPL3034135 NPL3034137 NPL3034138 NPL3034141 NPL3034142 NPL3034143 NPL3034145 NPL3034148 NPL3034151 NPL3034152 NPL3034157 NPL3034158 NPL3034159 NPL3034161 NPL3034163 NPL3034165 NPL3034168 NPL3034169 NPL3034171 NPL3034172 NPL3034173 NPL3034178 NPL3034179 NPL3034181 NPL3034182 NPL3034183 NPL3034185 NPL3034186 NPL3034187 NPL3034188 NPL3034189 NPL3034192 NPL3034194 NPL3034198 NPL3034199 NPL3034200 NPL3034201 NPL3034203 NPL3034205 NPL3034206 NPL3034209 NPL3034210 NPL3034211 NPL3034215 NPL3034216 NPL3034226 NPL3034227 NPL3034229 NPL3034231 NPL3034232 NPL3034233 NPL3034234 NPL3034236 NPL3034237 NPL3034238 NPL3034239 NPL3034240 NPL3034243 NPL3034244 NPL3034247 NPL3034250 NPL3034252 NPL3034253 NPL3034256 NPL3034259 NPL3034260 NPL3034262 NPL3034264 NPL3034266 NPL3034267 NPL3034269 NPL3034270 NPL3034271 NPL3034275 NPL3034276 NPL3034277 NPL3034278 NPL3034279 NPL3034280 NPL3034285 NPL3034286 NPL3034290 NPL3034291 NPL3034292 NPL3034294 NPL3034295 NPL3034296 NPL3034299 NPL3034301 NPL3034303 NPL3034306 NPL3034307 NPL3034308 NPL3034310 NPL3034316 NPL3034317 NPL3034319 NPL3034323 NPL3034324 NPL3034325 NPL3034326 NPL3034330 NPL3034331 NPL3034332 NPL3034334 NPL3034335 NPL3034339 NPL3034341 NPL3034342 NPL3034343 NPL3034345 NPL3034347 NPL3034348 NPL3034351 NPL3034352 NPL3034353 NPL3034354 NPL3034355 NPL3034357 NPL3034359 NPL3034360 NPL3034361 NPL3034363 NPL3034366 NPL3034367 NPL3034368 NPL3034370 NPL3034371 NPL3034372 NPL3034373 NPL3034374 NPL3034378 NPL3034380 NPL3034381 NPL3034382 NPL3034383 NPL3034384 NPL3034385 NPL3034388 NPL3034390 NPL3034391 NPL3034392 NPL3034397 NPL3034400 NPL3034402 NPL3034404 NPL3034405 NPL3034406 NPL3034407 NPL3034418 NPL3034421 NPL3034425 NPL3034427 NPL3034431 NPL3034437 NPL3034440 NPL3034441 NPL3034444 NPL3034449 NPL3034450 NPL3034459 NPL3034460 NPL3034461 NPL3034464 NPL3034465 NPL3034466 NPL3034467 NPL3034468 NPL3034470 NPL3034472 NPL3034474 NPL3034475 NPL3034477 NPL3034478 NPL3034479 NPL3034480 NPL3034483 NPL3034484 NPL3034490 NPL3034492 NPL3034493 NPL3034494 NPL3034495 NPL3034496 NPL3034498 NPL3034500 NPL3034502 NPL3034505 NPL3034506 NPL3034507 NPL3034513 NPL3034519 NPL3034520 NPL3034529 NPL3034534 NPL3034536 NPL3034539 NPL3034543 NPL3034544 NPL3034547 NPL3034548 NPL3034549 NPL3034550 NPL3034552 NPL3034553 NPL3034555 NPL3034556 NPL3034557 NPL3034560 NPL3034563 NPL3034571 NPL3034572 NPL3034574 NPL3034575 NPL3034592 NPL3034595 NPL3034597 NPL3034605 NPL3034608 NPL3034610 NPL3034615 NPL3034616 NPL3034618 NPL3034619 NPL3034634 NPL3034635 NPL3034636 NPL3034638 NPL3034641 NPL3034644 NPL3034648 NPL3034649 NPL3034651 NPL3034652 NPL3034658 NPL3034659 NPL3034661 NPL3034662 NPL3034665 NPL3034669 NPL3034670 NPL3034671 NPL3034673 NPL3034674 NPL3034676 NPL3034678 NPL3034679 NPL3034680 NPL3034681 NPL3034683 NPL3034685 NPL3034686 NPL3034688 NPL3034692 NPL3034693 NPL3034697 NPL3034702 NPL3034703 NPL3034706 NPL3034711 NPL3034714 NPL3034718 NPL3034720 NPL3034722 NPL3034723 NPL3034724 NPL3034728 NPL3034730 NPL3034732 NPL3034734 NPL3034737 NPL3034738 NPL3034740 NPL3034744 NPL3034747 NPL3034748 NPL3034752 NPL3034753 NPL3034757 NPL3034758 NPL3034759 NPL3034761 NPL3034762 NPL3034769 NPL3034772 NPL3034774 NPL3034776 NPL3034777 NPL3034778 NPL3034781 NPL3034785 NPL3036791 NPL3036795 NPL3036798 NPL3036802 NPL3036805 NPL3036806 NPL3036808 NPL3036810 NPL3036811 NPL3036817 NPL3036818 NPL3036819 NPL3036820 NPL3036821 NPL3036822 NPL3036824 NPL3036825 NPL3036827 NPL3036835 NPL3036842 NPL3036843 NPL3036859 NPL3036860 NPL3036863 NPL3036865 NPL3036866 NPL3036867 NPL3036871 NPL3036872 NPL3036875 NPL3036876 NPL3036877 NPL3036883 NPL3036885 NPL3036886 NPL3036889 NPL3036890 NPL3036902 NPL3036905 NPL3036908 NPL3036910 NPL3036916 NPL3036926 NPL3036927 NPL3036932 NPL3036934 NPL3036936 NPL3036940 NPL3036942 NPL3036944 NPL3036945 NPL3036946 NPL3036947 NPL3036948 NPL3036954 NPL3036956 NPL3036957 NPL3036959 NPL3036960 NPL3036962 NPL3036963 NPL3036967 NPL3036968 NPL3036969 NPL3036985 NPL3036991 NPL3036997 NPL3036999 NPL3037014 NPL3037022 NPL3037030 NPL3037038 NPL3037048 NPL3037052 NPL3037061 NPL3037062 NPL3037064 NPL3037066 NPL3037070 NPL3037072 NPL3037083 NPL3037085 NPL3037087 NPL3037088 NPL3037089 NPL3037094 NPL3037107 NPL3037108 NPL3037111 NPL3037112 NPL3037113 NPL3037115 NPL3037118 NPL3037119 NPL3037121 NPL3037123 NPL3037124 NPL3037126 NPL3037127 NPL3037131 NPL3037132 NPL3037136 NPL3037137 NPL3037139 NPL3037141 NPL3037143 NPL3037145 NPL3037146 NPL3037147 NPL3037148 NPL3037149 NPL3037150 NPL3037151 NPL3037152 NPL3037160 NPL3037163 NPL3037165 NPL3037167 NPL3037168 NPL3037171 NPL3037172 NPL3037173 NPL3037174 NPL3037176 NPL3037177 NPL3037178 NPL3037179 NPL3037180 NPL3037181 NPL3037183 NPL3037186 NPL3037188 NPL3037189 NPL3037191 NPL3037194 NPL3037198 NPL3037201 NPL3037203 NPL3037205 NPL3037208 NPL3037211 NPL3037212 NPL3037213 NPL3037215 NPL3037216 NPL3037217 NPL3037219 NPL3037221 NPL3037224 NPL3037225 NPL3037228 NPL3037229 NPL3037230 NPL3037231 NPL3037233 NPL3037237 NPL3037239 NPL3037240 NPL3037241 NPL3037243 NPL3037244 NPL3037246 NPL3037248 NPL3037250 NPL3037251 NPL3037252 NPL3037253 NPL3037254 NPL3037256 NPL3037260 NPL3037262 NPL3037264 NPL3037265 NPL3037266 NPL3037267 NPL3037269 NPL3037270 NPL3037271 NPL3037272 NPL3037277 NPL3037278 NPL3037280 NPL3037281 NPL3037283 NPL3037284 NPL3037285 NPL3037289 NPL3037292 NPL3037293 NPL3037294 NPL3037295 NPL3037297 NPL3037303 NPL3037305 NPL3037306 NPL3037308 NPL3037311 NPL3037319 NPL3037323 NPL3037324 NPL3037326 NPL3037327 NPL3037328 NPL3037329 NPL3037330 NPL3037331 NPL3037334 NPL3037336 NPL3037337 NPL3037340 NPL3037344 NPL3037346 NPL3037349 NPL3037352 NPL3037353 NPL3037355 NPL3037361 NPL3037362 NPL3037363 NPL3037364 NPL3037365 NPL3037366 NPL3037368 NPL3037369 NPL3037370 NPL3037371 NPL3037380 NPL3037381 NPL3037382 NPL3037383 NPL3037384 NPL3037387 NPL3037388 NPL3037389 NPL3037391 NPL3037396 NPL3037397 NPL3037399 NPL3037400 NPL3037402 NPL3037403 NPL3037404 NPL3037405 NPL3037406 NPL3037408 NPL3037411 NPL3037412 NPL3037414 NPL3037418 NPL3037420 NPL3037421 NPL3037425 NPL3037426 NPL3037429 NPL3037430 NPL3037431 NPL3037432 NPL3037433 NPL3037436 NPL3037440 NPL3037441 NPL3037444 NPL3037445 NPL3037446 NPL3037451 NPL3037456 NPL3037457 NPL3037459 NPL3037460 NPL3037462 NPL3037465 NPL3037466 NPL3037469 NPL3037476 NPL3037491 NPL3037493 NPL3037495 NPL3037499 NPL3037501 NPL3037503 NPL3037520 NPL3037526 NPL3037528 NPL3037530 NPL3037531 NPL3037542 NPL3037559 NPL3037563 NPL3037567 NPL3037568 NPL3037569 NPL3037571 NPL3037581 NPL3037586 NPL3037594 NPL3037596 NPL3037598 NPL3037601 NPL3037602 NPL3037603 NPL3037604 NPL3037607 NPL3037611 NPL3037614 NPL3037616 NPL3037617 NPL3037618 NPL3037620 NPL3037621 NPL3037624 NPL3037628 NPL3037629 NPL3037630 NPL3037631 NPL3037634 NPL3037635 NPL3037636 NPL3037637 NPL3037640 NPL3037641 NPL3037642 NPL3037645 NPL3037646 NPL3037647 NPL3037648 NPL3037650 NPL3037651 NPL3037652 NPL3037653 NPL3037654 NPL3037655 NPL3037657 NPL3037658 NPL3037659 NPL3037660 NPL3037661 NPL3037662 NPL3037664 NPL3037668 NPL3037669 NPL3037670 NPL3037672 NPL3037673 NPL3037674 NPL3037677 NPL3037678 NPL3037679 NPL3037681 NPL3037684 NPL3037688 NPL3037689 NPL3037690 NPL3037691 NPL3037695 NPL3037699 NPL3037700 NPL3037703 NPL3037704 NPL3037705 NPL3037706 NPL3037707 NPL3037708 NPL3037711 NPL3037713 NPL3037714 NPL3037715 NPL3037719 NPL3037720 NPL3037725 NPL3037728 NPL3037729 NPL3037732 NPL3037736 NPL3037737 NPL3037740 NPL3037741 NPL3037742 NPL3037745 NPL3037746 NPL3037747 NPL3037748 NPL3037750 NPL3037757 NPL3037759 NPL3037761 NPL3037762 NPL3037764 NPL3037765 NPL3037766 NPL3037767 NPL3037768 NPL3037771 NPL3037774 NPL3037776 NPL3037779 NPL3037780 NPL3037784 NPL3037785 NPL3037786 NPL3037789 NPL3037793 NPL3037796 NPL3037800 NPL3037802 NPL3037805 NPL3037811 NPL3037813 NPL3037814 NPL3037817 NPL3037818 NPL3037823 NPL3037826 NPL3037828 NPL3037830 NPL3037832 NPL3037833 NPL3037835 NPL3037837 NPL3037839 NPL3037842 NPL3037843 NPL3037844 NPL3037845 NPL3037847 NPL3037849 NPL3037850 NPL3037852 NPL3037855 NPL3037858 NPL3037860 NPL3037865 NPL3037870 NPL3037872 NPL3037879 NPL3037883 NPL3037884 NPL3037889 NPL3037899 NPL3037902 NPL3037907 NPL3037909 NPL3037912 NPL3037914 NPL3037915 NPL3037917 NPL3037922 NPL3037925 NPL3037928 NPL3037929 NPL3037930 NPL3037931 NPL3037932 NPL3037934 NPL3037937 NPL3037939 NPL3037944 NPL3037946 NPL3037948 NPL3037954 NPL3037956 NPL3037958 NPL3037960 NPL3037963 NPL3037965 NPL3037966 NPL3037973 NPL3037974 NPL3037975 NPL3037976 NPL3037977 NPL3037978 NPL3037979 NPL3037986 NPL3037990 NPL3037992 NPL3037996 NPL3037999 NPL3038000 NPL3038005 NPL3038012 NPL3038014 NPL3038015 NPL3038019 NPL3038020 NPL3038041 NPL3038042 NPL3038044 NPL3038045 NPL3038046 NPL3038048 NPL3038049 NPL3038051 NPL3038060 NPL3038062 NPL3038063 NPL3038064 NPL3038067 NPL3038069 NPL3038071 NPL3038072 NPL3038074 NPL3038075 NPL3038077 NPL3038078 NPL3038079 NPL3038080 NPL3038081 NPL3038091 NPL3038093 NPL3038097 NPL3038100 NPL3038107 NPL3038108 NPL3038109 NPL3038110 NPL3038113 NPL3038115 NPL3038119 NPL3038125 NPL3038126 NPL3038128 NPL3038129 NPL3038130 NPL3038133 NPL3038134 NPL3038136 NPL3038137 NPL3038138 NPL3038140 NPL3038142 NPL3038143 NPL3038151 NPL3038153 NPL3038154 NPL3038156 NPL3038160 NPL3038161 NPL3038162 NPL3038164 NPL3038167 NPL3038170 NPL3038173 NPL3038174 NPL3038175 NPL3038176 NPL3038177 NPL3038179 NPL3038180 NPL3038181 NPL3038184 NPL3038186 NPL3038187 NPL3038188 NPL3038189 NPL3038191 NPL3038193 NPL3038199 NPL3038205 NPL3038206 NPL3038214 NPL3038215 NPL3038217 NPL3038221 NPL3038222 NPL3038223 NPL3038224 NPL3038227 NPL3038229 NPL3038232 NPL3038233 NPL3038234 NPL3038239 NPL3038240 NPL3038243 NPL3038245 NPL3038246 NPL3038249 NPL3038251 NPL3038252 NPL3038253 NPL3038255 NPL3038256 NPL3038257 NPL3038260 NPL3038261 NPL3038263 NPL3038264 NPL3038265 NPL3038266 NPL3038269 NPL3038274 NPL3038276 NPL3038277 NPL3038280 NPL3038282 NPL3038283 NPL3038286 NPL3038287 NPL3038288 NPL3038291 NPL3038292 NPL3038293 NPL3038296 NPL3038297 NPL3038298 NPL3038300 NPL3038310 NPL3038316 NPL3038319 NPL3038320 NPL3038321 NPL3038323 NPL3038324 NPL3038325 NPL3038326 NPL3038327 NPL3038331 NPL3038334 NPL3038337 NPL3038343 NPL3038344 NPL3038346 NPL3038347 NPL3038348 NPL3038349 NPL3038350 NPL3038352 NPL3038355 NPL3038356 NPL3038361 NPL3038363 NPL3038365 NPL3038366 NPL3038370 NPL3038371 NPL3038372 NPL3038373 NPL3038374 NPL3038375 NPL3038376 NPL3038379 NPL3038380 NPL3038381 NPL3038385 NPL3038386 NPL3038388 NPL3038390 NPL3038395 NPL3038397 NPL3038399 NPL3038400 NPL3038401 NPL3038402 NPL3038403 NPL3038404 NPL3038405 NPL3038409 NPL3038412 NPL3038413 NPL3038414 NPL3038416 NPL3038418 NPL3038419 NPL3038421 NPL3038423 NPL3038427 NPL3038429 NPL3038430 NPL3038432 NPL3038434 NPL3038440 NPL3038443 NPL3038445 NPL3038446 NPL3038461 NPL3038462 NPL3038475 NPL3038494 NPL3038495 NPL3038496 NPL3038497 NPL3038509 NPL3038510 NPL3038512 NPL3038513 NPL3038514 NPL3038515 NPL3038516 NPL3038517 NPL3038518 NPL3038519 NPL3038520 NPL3038522 NPL3038524 NPL3038526 NPL3038527 NPL3038528 NPL3038530 NPL3038532 NPL3038534 NPL3038539 NPL3038545 NPL3038550 NPL3038553 NPL3038580 NPL3038581 NPL3038585 NPL3038588 NPL3038599 NPL3038603 NPL3038604 NPL3038614 NPL3038617 NPL3038630 NPL3038714 NPL3038718 NPL3038719 NPL3038736 NPL3038737 NPL3038738 NPL3038741 NPL3038742 NPL3038779 NPL3038780 NPL3038781 NPL3038802 NPL3038804 NPL3038827 NPL3038828 NPL3038829 NPL3038830 NPL3038831 NPL3038832 NPL3038833 NPL3038847 NPL3038849 NPL3038851 NPL3038852 NPL3038854 NPL3038856 NPL3038865 NPL3038873 NPL3038875 NPL3038876 NPL3038877 NPL3038885 NPL3038888 NPL3038891 NPL3038892 NPL3038895 NPL3038898 NPL3038899 NPL3038900 NPL3038905 NPL3038908 NPL3038909 NPL3038911 NPL3038916 NPL3038917 NPL3038918 NPL3038920 NPL3038921 NPL3038928 NPL3038946 NPL3038947 NPL3038951 NPL3038953 NPL3038954 NPL3038955 NPL3038956 NPL3038958 NPL3038959 NPL3038966 NPL3038972 NPL3038973 NPL3038974 NPL3038975 NPL3038976 NPL3038982 NPL3038984 NPL3038985 NPL3038987 NPL3038989 NPL3038990 NPL3038992 NPL3038996 NPL3039002 NPL3039003 NPL3039012 NPL3039013 NPL3039014 NPL3039015 NPL3039016 NPL3039019 NPL3039020 NPL3039022 NPL3039023 NPL3039024 NPL3039025 NPL3039028 NPL3039033 NPL3039034 NPL3039035 NPL3039036 NPL3039040 NPL3039043 NPL3039044 NPL3039046 NPL3039047 NPL3039055 NPL3039058 NPL3039059 NPL3039060 NPL3039061 NPL3039062 NPL3039063 NPL3039067 NPL3039071 NPL3039072 NPL3039073 NPL3039074 NPL3039076 NPL3039084 NPL3039085 NPL3039087 NPL3039088 NPL3039097 NPL3039119 NPL3039121 NPL3039123 NPL3039127 NPL3039131 NPL3039132 NPL3039134 NPL3039137 NPL3039138 NPL3039152 NPL3039154 NPL3039168 NPL3039179 NPL3039182 NPL3039183 NPL3039187 NPL3039188 NPL3039189 NPL3039190 NPL3039194 NPL3039195 NPL3039201 NPL3039209 NPL3039211 NPL3039212 NPL3039231 NPL3039232 NPL3039233 NPL3039234 NPL3039237 NPL3039239 NPL3039240 NPL3039241 NPL3039242 NPL3039244 NPL3039245 NPL3039246 NPL3039252 NPL3039254 NPL3039255 NPL3039257 NPL3039260 NPL3039262 NPL3039263 NPL3039265 NPL3039266 NPL3039267 NPL3039268 NPL3039269 NPL3039270 NPL3039271 NPL3039272 NPL3039273 NPL3039274 NPL3039278 NPL3039279 NPL3039280 NPL3039291 NPL3039292 NPL3039293 NPL3039296 NPL3039305 NPL3039308 NPL3039309 NPL3039310 NPL3039326 NPL3039327 NPL3039329 NPL3039330 NPL3039331 NPL3039332 NPL3039333 NPL3039334 NPL3039335 NPL3039340 NPL3039345 NPL3039346 NPL3039347 NPL3039348 NPL3039349 NPL3039353 NPL3039354 NPL3039356 NPL3039357 NPL3039359 NPL3039360 NPL3039364 NPL3039365 NPL3039374 NPL3039375 NPL3039376 NPL3039378 NPL3039382 NPL3039383 NPL3039384 NPL3039385 NPL3039386 NPL3039387 NPL3039388 NPL3039389 NPL3039390 NPL3039391 NPL3039392 NPL3039393 NPL3039394 NPL3039395 NPL3039396 NPL3039397 NPL3039398 NPL3039399 NPL3039400 NPL3039401 NPL3039402 NPL3039403 NPL3039404 NPL3039405 NPL3039406 NPL3039407 NPL3039408 NPL3039419 NPL3039420 NPL3039429 NPL3039430 NPL3039435 NPL3039436 NPL3039440 NPL3039456 NPL3039457 NPL3039460 NPL3039479 NPL3039484 NPL3039485 NPL3039486 NPL3039488 NPL3039489 NPL3039490 NPL3039495 NPL3039496 NPL3039497 NPL3039498 NPL3039501 NPL3039504 NPL3039509 NPL3039510 NPL3039511 NPL3039512 NPL3039513 NPL3039514 NPL3039515 NPL3039516 NPL3039517 NPL3039518 NPL3039519 NPL3039523 NPL3039525 NPL3039526 NPL3039528 NPL3039530 NPL3039537 NPL3039542 NPL3039543 NPL3039544 NPL3039545 NPL3039548 NPL3039555 NPL3039556 NPL3039557 NPL3039558 NPL3039560 NPL3039562 NPL3039611 NPL3039612 NPL3039613 NPL3039614 NPL3039615 NPL3039619 NPL3039622 NPL3039625 NPL3039626 NPL3039627 NPL3039640 NPL3039642 NPL3039647 NPL3039656 NPL3039657 NPL3039659 NPL3039660 NPL3039664 NPL3039672 NPL3039678 NPL3039679 NPL3039680 NPL3039681 NPL3039686 NPL3039688 NPL3039701 NPL3039703 NPL3039704 NPL3039705 NPL3039709 NPL3039713 NPL3039732 NPL3039735 NPL3039736 NPL3039776 NPL3039801 NPL3039806 NPL3039844 NPL3039850 NPL3039852 NPL3039862 NPL3039865 NPL3039866 NPL3026119 NPL3026156 NPL3038531 NPL3025012 NPL3026130
Recalling Firm/
Manufacturer
Medtronic Neuromodulation
7000 Central Ave Ne
Minneapolis MN 55432-3568
For Additional Information ContactJeffrey Trauring
763-514-2000
Manufacturer Reason
for Recall
Software issues were identified in application version 2.x.
FDA Determined
Cause 2
Under Investigation by firm
ActionAn URGENT: MEDICAL DEVICE CORRECTION notification letter dated November 2024 was sent to customers. Actions Share this notice with all those who need to be aware of this issue within your organization and maintain a copy of this notice in your records. Please complete and return the Customer Confirmation Form enclosed with this letter, acknowledging that you have received this information. The form may be emailed to neuro.quality@medtronic.com. Additional Information Medtronic is actively working on a permanent solution and will inform you as soon as it becomes available. Medtronic is communicating this information to the appropriate regulatory agencies. Adverse events or quality problems experienced with this product should be reported to Medtronic and the FDA MedWatch Adverse Event Reporting program via: Medtronic email at RS.NeuAdverseEventReporting@Medtronic.com or Medtronic Technical Services at 1-800-707-0933 weekdays 7:00 AM 6:00 PM CT.
Quantity in Commerce3981 systems
DistributionWorldwide - US Nationwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-