• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall OEC Elite

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall OEC Elitesee related information
Date Initiated by FirmNovember 21, 2024
Date PostedDecember 18, 2024
Recall Status1 Open3, Classified
Recall NumberZ-0704-2025
Recall Event ID 95846
510(K)NumberK170752 K171565 K172550 K192819 
Product Classification Image-intensified fluoroscopic x-ray system, mobile - Product Code OXO
ProductOEC Elite REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC Elite mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventional, and surgical procedures.
Code Information Product Name: OEC Elite GTIN code: 00001234567895 Lot/Serial Numbers / UDI codes: FAXXXE00254 (01)00840682121729(11)190400(21)FAXXXE00254 FBHXTX00837 (01)00840682121729(11)230300(21)FBHXTX00837 FBHXTX00912 (01)00840682121729(11)230500(21)FBHXTX00912 FBHXTX01278 (01)00840682121729(11)240800(21)FBHXTX01278 FBXXXE00731 (01)00840682121729(11)190500(21)FBXXXE00731 FBMHTX00470 (01)00840682121729(11)240900(21)FBMHTX00470 F2XXTX00801 (01)00840682121729(11)240800(21)F2XXTX00801 FBHXXX00171 (01)00840682121729(11)180600(21)FBHXXX00171 FBXXXE00715 (01)00840682121729(11)190500(21)FBXXXE00715 FBMHTX00444 (01)00840682121729(11)240600(21)FBMHTX00444 FBXXTE02974 (01)00840682121729(11)240800(21)FBXXTE02974 FBMHTX00466 (01)00840682121729(11)240900(21)FBMHTX00466 FAXXTE00672 (01)00840682121729(11)221200(21)FAXXTE00672 F9XXTX00276 (01)00840682121729(11)230900(21)F9XXTX00276 FAXXTE01160 (01)00840682121729(11)240800(21)FAXXTE01160 FBMHTX00454 (01)00840682121729(11)240800(21)FBMHTX00454 FBMHTX00455 (01)00840682121729(11)240800(21)FBMHTX00455 FBMHTX00465 (01)00840682121729(11)240900(21)FBMHTX00465 FBXXTE02975 (01)00840682121729(11)240800(21)FBXXTE02975 FBXXTE02976 (01)00840682121729(11)240800(21)FBXXTE02976 FBXXXE00312 (01)00840682121729(11)180900(21)FBXXXE00312 FAXXTE00664 (01)00840682121729(11)221100(21)FAXXTE00664 FBMHTX00450 (01)00840682121729(11)240700(21)FBMHTX00450 FAXXTE01154 (01)00840682121729(11)240700(21)FAXXTE01154 FBXXTE02994 (01)00840682121729(11)240800(21)FBXXTE02994 FAXXTE00544 (01)00840682121729(11)220600(21)FAXXTE00544 FBXXTE01055 (01)00840682121729(11)211200(21)FBXXTE01055 FBXXTE02966 (01)00840682121729(11)240800(21)FBXXTE02966 FBMHTX00467 (01)00840682121729(11)240900(21)FBMHTX00467 FAXXTE00339 (01)00840682121729(11)210600(21)FAXXTE00339 FSXXTX00282 (01)00840682121729(11)220500(21)FSXXTX00282 FAXXXE00231 (01)00840682121729(11)190300(21)FAXXXE00231 FBXXXE00653 (01)00840682121729(11)190400(21)FBXXXE00653 F2XXTX00381 (01)00840682121729(11)220100(21)F2XXTX00381 FBXXXE00008 (01)00840682121729(11)171000(21)FBXXXE00008 FBMHTX00468 (01)00840682121729(11)240900(21)FBMHTX00468 FBXXTP00008 (01)00840682121729(11)240700(21)FBXXTP00008 FSXXTX00501 (01)00840682121729(11)240400(21)FSXXTX00501 F2XXTX00802 (01)00840682121729(11)240800(21)F2XXTX00802 FAXXTE01157 (01)00840682121729(11)240700(21)FAXXTE01157 FBMHTX00447 (01)00840682121729(11)240600(21)FBMHTX00447 FBMHDX00049 (01)00195278096944(11)220100(21)FBMHDX00049 FBXXTE02312 (01)00840682121729(11)230900(21)FBXXTE02312 F9XXTX00301 (01)00840682121729(11)240100(21)F9XXTX00301 FBXXTE02965 (01)00840682121729(11)240800(21)FBXXTE02965 FBXXXE00654 (01)00840682121729(11)190400(21)FBXXXE00654 FBMHTX00452 (01)00840682121729(11)240800(21)FBMHTX00452 FBXXTE00561 (01)00840682121729(11)201100(21)FBXXTE00561 FBXXTE02963 (01)00840682121729(11)240800(21)FBXXTE02963 FBXXTE02109 (01)00840682121729(11)230600(21)FBXXTE02109 FBXXTP00006 (01)00840682121729(11)240700(21)FBXXTP00006 FBXXTP00007 (01)00840682121729(11)240700(21)FBXXTP00007 F2XXTX00021 (01)00840682121729(11)191100(21)F2XXTX00021 FBXXTE02991 (01)00840682121729(11)240800(21)FBXXTE02991 FBXXTE02992 (01)00840682121729(11)240800(21)FBXXTE02992 F2XXTX00799 (01)00840682121729(11)240800(21)F2XXTX00799 FSXXTX00514 (01)00840682121729(11)240600(21)FSXXTX00514 FBXXTE01754 (01)00840682121729(11)230100(21)FBXXTE01754 FBXXTE02914 (01)00840682121729(11)240600(21)FBXXTE02914 FBXXTE00515 (01)00840682121729(11)201000(21)FBXXTE00515 FAXXXE00049 (01)00840682121729(11)180200(21)FAXXXE00049 FBXXXE00272 (01)00840682121729(11)180900(21)FBXXXE00272 FSXXXX00004 (01)00840682121729(11)190400(21)FSXXXX00004 FBMHTX00433 (01)00840682121729(11)240400(21)FBMHTX00433 FBMHTX00446 (01)00840682121729(11)240600(21)FBMHTX00446 FBXXTE01644 (01)00840682121729(11)221200(21)FBXXTE01644 FBMHTX00464 (01)00840682121729(11)240900(21)FBMHTX00464 FBXXTE02978 (01)00840682121729(11)240800(21)FBXXTE02978 F2XXTX00798 (01)00840682121729(11)240800(21)F2XXTX00798 FBXXTE02979 (01)00840682121729(11)240800(21)FBXXTE02979 FBXXTE01932 (01)00840682121729(11)230300(21)FBXXTE01932 FBXXTE02977 (01)00840682121729(11)240800(21)FBXXTE02977 FBXXTE02993 (01)00840682121729(11)240800(21)FBXXTE02993 FBHXTX00821 (01)00840682121729(11)230200(21)FBHXTX00821 F2XXTX00800 (01)00840682121729(11)240800(21)F2XXTX00800 FSXXTX00284 (01)00840682121729(11)220500(21)FSXXTX00284 FBXXTE00575 (01)00840682121729(11)201200(21)FBXXTE00575 FBMHTX00469 (01)00840682121729(11)240900(21)FBMHTX00469 FBXXTE02323 (01)00840682121729(11)230900(21)FBXXTE02323 FSXXTX00422 (01)00840682121729(11)230300(21)FSXXTX00422 FAHXTX00538 (01)00840682121729(11)240600(21)FAHXTX00538 FBXXTE02996 (01)00840682121729(11)240800(21)FBXXTE02996 FSXXTX00418 (01)00840682121729(11)230300(21)FSXXTX00418 FBXXTE01863 (01)00840682121729(11)230300(21)FBXXTE01863 FBMHTX00453 (01)00840682121729(11)240800(21)FBMHTX00453 FBMHTX00329 (01)00840682121729(11)230200(21)FBMHTX00329 FAXXTE01158 (01)00840682121729(11)240800(21)FAXXTE01158 FBXXTE02995 (01)00840682121729(11)240800(21)FBXXTE02995 FAXXTE01159 (01)00840682121729(11)240800(21)FAXXTE01159 F9XXXX00064 (01)00840682121729(11)191200(21)F9XXXX00064 FBXXTE02980 (01)00840682121729(11)240800(21)FBXXTE02980 FBXXTE02981 (01)00840682121729(11)240800(21)FBXXTE02981 FBXXTE02982 (01)00840682121729(11)240800(21)FBXXTE02982 FBXXTE02983 (01)00840682121729(11)240800(21)FBXXTE02983 FBXXTE02984 (01)00840682121729(11)240800(21)FBXXTE02984 FBXXTE02986 (01)00840682121729(11)240800(21)FBXXTE02986 FBXXTE02987 (01)00840682121729(11)240800(21)FBXXTE02987 FBMHTX00456 (01)00840682121729(11)240800(21)FBMHTX00456 FBMHTX00449 (01)00840682121729(11)240700(21)FBMHTX00449 F9XXTX00329 (01)00840682121729(11)240800(21)F9XXTX00329
FEI Number 1720753
Recalling Firm/
Manufacturer
GE OEC Medical Systems, Inc
384 N Wright Brothers Dr
Salt Lake City UT 84116-2862
For Additional Information ContactGE HealthCare Service
800-437-1171
Manufacturer Reason
for Recall
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
FDA Determined
Cause 2
Process control
ActionOn 11/21/2024, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter informing customers that GE HealthCare has become aware that recently manufactured OEC Elite and OEC 3D mobile C-arms, as well as recently replaced X-ray tubes on certain OEC Elite, OEC 3D and OEC 9900/OEC 9800 mobile C-arms potentially have insufficient sealing that can result in an oil leak. This issue could result in loss of fluoroscopic imaging capability. Customers are instructed to: 1.Continue to use their OEC Elite, OEC 3D, OEC 9900 and/or OEC9800 mobile C-arm. 2.If oil is found leaking from the X-ray tube on their OEC Elite, OEC 3D, OEC9900 and/or OEC 9800 mobile C-arm, contact the local GE HealthCareService representative. 3.Ensure all potential users in their facility are made aware of this safety notification and the recommended actions. 4.Please complete and return the attached acknowledgement form torecall.15150@gehealthcare.com. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to plan and schedule for the inspection and correction if needed. for any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Quantity in Commerce100 systems
DistributionU.S.: AK, AZ, CA, CO, CT, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Australia, Brazil, Canada, Chile, Costa Rica, India, Italy, Philippines, Poland, Portugal, Sweden, Switzerland, and United Kingdom
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = OXO
510(K)s with Product Code = OXO
-
-