• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Philips SmartHopping 2.0 AP 1.4 GHz

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Philips SmartHopping 2.0 AP 1.4 GHzsee related information
Date Initiated by FirmNovember 20, 2024
Date PostedDecember 17, 2024
Recall Status1 Open3, Classified
Recall NumberZ-0686-2025
Recall Event ID 95847
Product Classification Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
ProductPhilips Smart-Hopping 2.0 AP 1.4 GHz. Model Number: 867216
Code Information GTIN: 00884838099876. Model Number: 867216. All software versions are affected. UDI Numbers: (01)00884838099876(21)DK34706344, (01)00884838099876(21)DK33603995, (01)00884838099876(21)DK34706512, (01)00884838099876(21)DK33402245, (01)00884838099876(21)DK34706012, (01)00884838099876(21)DK33402324, (01)00884838099876(21)DK33603538, (01)00884838099876(21)DK34405346, (01)00884838099876(21)DK32801033, (01)00884838099876(21)DK34705955, (01)00884838099876(21)DK40307155, (01)00884838099876(21)DK34405363, (01)00884838099876(21)DK33604046, (01)00884838099876(21)DK34706011, (01)00884838099876(21)DK34706010, (01)00884838099876(21)DK33402486, (01)00884838099876(21)DK34405820, (01)00884838099876(21)DK34405539, (01)00884838099876(21)DK34405788, (01)00884838099876(21)DK33403107, (01)00884838099876(21)DK33603456, (01)00884838099876(21)DK34706099, (01)00884838099876(21)DK32801250, (01)00884838099876(21)DK32801095, (01)00884838099876(21)DK33604249, (01)00884838099876(21)DK33403062, (01)00884838099876(21)DK34706432, (01)00884838099876(21)DK33403214, (01)00884838099876(21)DK33402554, (01)00884838099876(21)DK40307129, (01)00884838099876(21)DK33603678, (01)00884838099876(21)DK32801039, (01)00884838099876(21)DK33402660, (01)00884838099876(21)DK34706048, (01)00884838099876(21)DK33904683, (01)00884838099876(21)DK33402439, (01)00884838099876(21)DK34405702, (01)00884838099876(21)DK40306709, (01)00884838099876(21)DK32800996, (01)00884838099876(21)DK34405910, (01)00884838099876(21)DK32800867, (01)00884838099876(21)DK33603918, (01)00884838099876(21)DK33604048, (01)00884838099876(21)DK33904818, (01)00884838099876(21)DK34706539, (01)00884838099876(21)DK34405884, (01)00884838099876(21)DK33403219, (01)00884838099876(21)DK33603727, (01)00884838099876(21)DK34706371, (01)00884838099876(21)DK33402675, (01)00884838099876(21)DK33402648, (01)00884838099876(21)DK33604588, (01)00884838099876(21)DK33604264, (01)00884838099876(21)DK33402127, (01)00884838099876(21)DK34706378, (01)00884838099876(21)DK32801318, (01)00884838099876(21)DK33403095, (01)00884838099876(21)DK33603562, (01)00884838099876(21)DK34706181, (01)00884838099876(21)DK33604554, (01)00884838099876(21)DK33904888, (01)00884838099876(21)DK33402343, (01)00884838099876(21)DK40306696, (01)00884838099876(21)DK40306992, (01)00884838099876(21)DK33603488, (01)00884838099876(21)DK32800866, (01)00884838099876(21)DK34706255, (01)00884838099876(21)DK34706320, (01)00884838099876(21)DK34405767, (01)00884838099876(21)DK34706095, (01)00884838099876(21)DK33402405, (01)00884838099876(21)DK33603933, (01)00884838099876(21)DK33604442, (01)00884838099876(21)DK40306831, (01)00884838099876(21)DK33403181, (01)00884838099876(21)DK34405435, (01)00884838099876(21)DK34405878, (01)00884838099876(21)DK33905047, (01)00884838099876(21)DK33403263, (01)00884838099876(21)DK33402108, (01)00884838099876(21)DK33904987, (01)00884838099876(21)DK33402141, (01)00884838099876(21)DK33402149, (01)00884838099876(21)DK40306877, (01)00884838099876(21)DK33603419, (01)00884838099876(21)DK33402615, (01)00884838099876(21)DK33604419, (01)00884838099876(21)DK32801254, (01)00884838099876(21)DK32801299, (01)00884838099876(21)DK33604191, (01)00884838099876(21)DK33403177, (01)00884838099876(21)DK33904965, (01)00884838099876(21)DK34405697, (01)00884838099876(21)DK33402251, (01)00884838099876(21)DK33402899, (01)00884838099876(21)DK33604612, (01)00884838099876(21)DK33905057, (01)00884838099876(21)DK32801016, (01)00884838099876(21)DK33403251, (01)00884838099876(21)DK32801153, (01)00884838099876(21)DK40306893, (01)00884838099876(21)DK33403133, (01)00884838099876(21)DK33904707, (01)00884838099876(21)DK33402794, (01)00884838099876(21)DK34405477, (01)00884838099876(21)DK40306707 ,(01)00884838099876(21)DK32800977, (01)00884838099876(21)DK33603959, (01)00884838099876(21)DK33603543, (01)00884838099876(21)DK33402371, (01)00884838099876(21)DK34405486, (01)00884838099876(21)DK33403272, (01)00884838099876(21)DK33603742, (01)00884838099876(21)DK32801045, (01)00884838099876(21)DK33604374, (01)00884838099876(21)DK32801218, (01)00884838099876(21)DK33604349, (01)00884838099876(21)DK33905158, (01)00884838099876(21)DK33603443, (01)00884838099876(21)DK34706451, (01)00884838099876(21)DK33402451, (01)00884838099876(21)DK34405777, (01)00884838099876(21)DK34706063, (01)00884838099876(21)DK33402692, (01)00884838099876(21)DK34706448, (01)00884838099876(21)DK33604062, (01)00884838099876(21)DK32800961, (01)00884838099876(21)DK33604477, (01)00884838099876(21)DK33904802, (01)00884838099876(21)DK40307072, (01)00884838099876(21)DK32801156, (01)00884838099876(21)DK32801317, (01)00884838099876(21)DK32800916, (01)00884838099876(21)DK33603989, (01)00884838099876(21)DK33905142, (01)00884838099876(21)DK33904796, (01)00884838099876(21)DK33604177, (01)00884838099876(21)DK33603571, (01)00884838099876(21)DK33604137, (01)00884838099876(21)DK33604529, (01)00884838099876(21)DK33604398. .
FEI Number 3016618143
Recalling Firm/
Manufacturer
Philips North America Llc
222 Jacobs St
Cambridge MA 02141-2289
For Additional Information ContactPhilips Customer Services
800-722-9377
Manufacturer Reason
for Recall
The following four issues are included: 1. Certain Channel Settings may result in an Access Point preventing a connected client device from roaming to another Access Point potentially causing the client device to appear to drop out from the network. 2. if the Access Point experiences a disconnection from the Access Point Controller exceeding 4 minutes, the Access Point may not transition through the desired state to all for clean reconnection of the client radio link. 3. Potential to experience higher than expected data dropouts, device disassociations/reassociations for their wireless clients, and frequent "Loss of AP" alerts. 4. Potential for the Smart-hopping Access Point is affected by a software issue where the Access Point will resent after 82.85 days when the frame counter reaches its limit. This causes the latest synchronization check to fail and potential for loss of patient data.
FDA Determined
Cause 2
Under Investigation by firm
ActionConsignees were notified on about 11/20/2024 via certified mail. Consignees were instructed to identify affected units, share the communication with clinical staff, and place the Urgent Medical Device Correction with the documentation of the Philips Smart-hopping 2.0 AP 1.4 GHz and associated devices. A Philips representative will contact customers to arrange a software upgrade to E.19.02 (5.0.19.02) or greater to correct the issues. Distributors were instructed to contact customers and provide the Medical Device Correction to each.
Quantity in Commerce4,811 units
DistributionNationwide distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-