• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall SenTiva"

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall SenTiva"see related information
Date Initiated by FirmJanuary 03, 2025
Date PostedJanuary 30, 2025
Recall Status1 Open3, Classified
Recall NumberZ-1067-2025
Recall Event ID 96063
PMA NumberP970003 
Product Classification Stimulator, autonomic nerve, implanted for epilepsy - Product Code LYJ
ProductBrand Name: SenTiva" Product Name: VNS Therapy" SenTiva", Model 1000 Model/Catalog Number: 1000 Software Version: NA Product Description: The LivaNova VNS Therapy" System, used for Vagus Nerve Stimulation (VNS), consists of an implantable VNS Therapy generator, lead, and external programming system used to change stimulation settings. The generator is an implantable, multi-programmable pulse generator that delivers electrical signals to the vagus nerve through the lead. The generator is housed in a hermetically sealed titanium case and is powered by a single battery. Electrical signals are transmitted from the generator to the vagus nerve by the lead. The lead and the generator make up the implantable portion of the VNS Therapy System. The VNS Therapy Programming System includes a computer pre-installed with VNS Therapy programming software and a programming wand. The physician uses the programming system to read and change generator settings. The Model 1000 Generator was commercially approved for distribution in the U.S. on October 3, 2017 via PMA P970003/S210. Component: The LivaNova VNS Therapy" System, used for Vagus Nerve Stimulation (VNS).
Code Information Lot Code: Model No 1000 UDI-DI: 05425025750405 Lot Numbers/Serial Numbers: 21517 21518 215190 215200 215210 215220 21530 21531 21532 21533 21534 21535 21536 215380 215400 21541 21542 21543 21544 21545 21546 21547 21548 21550 21551 21553 21554 21555 21556 21558 21559 21560 21562 21563 215640 215670 215680 21569 21570 21571 21573 21575 215770 21580 215810 215820 21583 215840 21585 21586 21589 215900 21591 21593 215940 215950 21597 21598 215990 216000 216020 216030 216040 216050 21606 21608 21609 216110 21616 21617 216180 216200 21621 216220 216250 216270 216280 216290 216300 216310 21632 216330 216340 21635 21636 21637 21638 21639 21640 216410 21642 216430 216440 216450 216460 216470 21648 216500 21651 21653 21654 21655 21659 21660 216620 21663 21664 216650 216670 21669 21670 21672 21673 21676 21677 21678 216790 216800 21681 216830 21684 21685 21686 21687 21688 21689 21690 21691 21692 21693 21694 21695 21696 21697 21699 21700 21701 21702 21704 21705 21706 21708 21710 21711 21713 21714 21715 21718 21719 21720 21721 21722 21723 21724 21725 21726 21728 21729 21731 21733 217350 21736 217380 21749 21750 21752 21754 21757 21758 21760 21762 21763 21766 21768 21769 21771 21772 21774 21775 21776 21777 21778 21779 21782 21783 21784 21786 21787 21788 217890 21790 21791 21792 21794 21800 21801 21802 21803 21804 21825 21826 21827 21829 21830 21832 21833 21834 21835 21836 21837 21841 21844 218460 21849 21851 21852 21853 21854 21855 21856 21857 21860 21861 21862 21863 21864 21865 21866 21867 218700 21871 21872 21873 21875 21880 21881 21882 21883 218840 21887 21888 21889 21892 21894 21896 21898 21899 21900 21903 21905 21907 219080 21909 219100 21911 21912 21913 219140 21917 219180 219190 219250 21926 21927 21928 21929 21931 21932 21937 21940 21941 21943 21944 21946 21947 21949 219500 219510 219520 219540 21955 21956 21957 21958 219600 21961 21963 21964 21965 21966 21967 21968 21969 21970 21971 21973 21974 21975 219760 219770 21978 21979 21980 21981 21982 21983 21984 21986 21987 21988 21990 21991 21992 21993 21994 21995 21996 21997 21998 21999 22001 22002 220030 22004 220050 22006 22007 22008 22009 22010 22013 22015 22016 22020 22021 22022 22023 22026 22027 22028 22029 22030 22031 22032 22033 22034 22037 220390 220400 220410 22042 22043 22044 22045 22046 22047 22049 220500 220510 22052 22053 220540 22055 220570 220580 220590 220600 220610 220630 220640 220650 220680 220690 220700 220710 220720 220730 220750 220760 220780 220790 220800 220840 220850 220860 220870 220880 220890 220900 220920 220950 220960 220970 220980 221000 221030 221050 221060 221080 221090 221100 221110 221120 221140 221150 221160 221180 221200 221220 221230 221250 22127 221280 221290 221310 221320 221340 221350 221360 221390 221400 221410 221470 221480 221490 221500 221530 221550 221570 221590 221600 221620 221650 221660 221670 221680 221690 221700 221710 221760 22178 22179 22180 22181 22184 22186 22187 22188 22190 22191 22192 22194 22195 221960 22197 22198 222000 222010 222020 222030 222040 222060 222080 222090 222110 222120 222130 222140 222150 222160 222200 222220 22223 222250 222260 222270 222340 222350 222400 222410 222420 222440 222470 222490 222500 222520 222540 222590 222600 222610 222620 222630 222640 222670 222690 222720 222730 222740 222750 222760 222770 222780 222790 222800 222810 222820 222830 222840 222850 222860 22287 222880 222910 222920 222930 22294 222950 222970 223000 223020 223050 22308 22309 223100 223140 223160 223190 223200 223220 223230 223240 223250 223260 223290 22330 223320 223340 22335 22336 223380 223390 223400 223430 223440 223450 223480 223500 223580 223600 223610 223640 223650 223660 223670 223680 223690 223710 223720 223740 223750 223770 223780 223790 223820 223830 223870 223900 223920 223930 223940 223950 223960 22397 223980 224000 22401 224030 224040 22406 224080 224090 224100 224110 224120 224130 224140 22415 224160 22417 22418 224190 224200 224210 224220 22423 224240 224250 224270 22428 22429 22430 224320 224350 224390 22440 224410 22442 22443 224440 224490 224500 224520 22453 224540 224550 224560 22457 224590 22460 224610 224620 224630 224640 224650 22466 22467 22468 22470 22473 224750 22477 22480 224840 22485 224860 224870 224880 224890 224900 224920 224940 22495 224960 224970 224980 22499 225000 225010 225030 225050 225060 22507 225080 225090 225120 225130 225150 225190 225250 225270 225290 225420 225430 225490 225510 225550 225580 225600 225620 225690 225700 225730 225750 225790 225860 22587 225880 22589 225900 225970 226000 226010 226020 226040 226050 226070 226090 226100 226120 22614 22615 226170 226210 226230 226240 22625 22626 22629 226330 226340 226350 226360 22637 22639 226400 22642 226430 22644 22645 22646 22647 22648 226510 22652 22653 226540 226560 226590 22660 226610 22662 22663 226650 226670 226680 226690 22670 226710 226740 226750 226760 226770 226780 226790 226800 226810 226860 226870 226880 226890 226900 226920 226930 226940 226950 226960 226970 226990 227000 227010 227020 227040 227050 227060 227110 227130 227160 227170 227180 227190 22720 22721 22723 22724 22726 22728 22730 22731 227320 22734 227350 22736 22737 22738 22739 227410 227420 227440 22745 22747 227490 22751 22752 22754 22755 22756 22757 227580 227590 227610 227620 22763 227650 227670 227680 227690 227700 227710 227720 227730 22774 227750 227770 22779 227820 22783 227840 22785 227860 22787 227880 227910 227920 227940 227950 227960 227970 227980 228010 228020 228030 228040 228050 228080 228090 228110 228120 228130 228140 228160 228170 228180 228190 228200 228210 228230 228240 228260 228270 228280 228300 228310 228320 228330 228340 22835 228360 228380 228390 228400 228410 228430 228440 228450 228470 228480 22849 22850 22851 22852 22853 22854 22855 22856 22857 228590 22860 228610 228620 228630 228650 228660 22869 22870 228710 228730 22874 22875 228760 228770 22881 22883 22884 22885 22887 22892 22893 22895 22896 22899 22902 229040 229050 229060 229070 229080 229100 229130 229140 229170 229180 229200 229250 229260 229270 229280 229300 229310 229330 229350 229360 229370 229390 229400 229410 229420 229430 229440 229460 229470 229480 229490 229500 229510 229520 229530 229540 229550 229560 229570 22958 22959 229600 22961 22962 22965 229670 22968 22980 22981 22982 229830 229840 229860 229880 229890 229920 229930 229950 229960 229980 22999 230010 23003 230050 23006 230070 230080 23010 230120 230130 230140 230150 230160 230170 230180 230190 230200 230210 230220 230240 230260 230280 230290 230300 230320 230330 230340 230350 230360 230370 230380 23039 23040 23041 23042 23044 230450 230460 23047 23048 230490 230510 230530 230540 230550 230560 230570 230590 230600 230610 230620 230630 230660 230670 230680 230690 230720 23074 230750 230760 230780 23079 230800 230820 230830 230850 230880 230890 230900 230910 230930 230970 231000 231020 231030 231040 231050 231070 231090 231100 231120 231130 231140 23115 231160 231170 231180 231190 231200 23121 231220 23123 231240 23127 23129 231320 231340 231360 231370 231380 231390 231400 231410 231420 231430 231440 231460 231470 231490 231500 231520 231530 231540 231550 231560 231570 231580 231590 231600 231610 231640 231650 231660 231670 231680 231690 231700 231710 231730 231750 231770 231780 231810 231840 231860 231870 231890 231900 231910 231950 23198 231990 232000 232030 232060 23209 232120 232140 232160 232170 232180 232190 232200 232220 232230 232250 232260 232310 232320 232350 232380 232390 232400 232410 232460 232490 23252 232530 232560 232570 23258 232590 232610 232620 232640 232650 232670 232680 232690 306081 306082 306083 306087 306088 306089 306091 306092 306093 306094 306095 306096 306098 306099 306100 306102 306103 306104 306106 306107 306108 306109 306110 306111 306113 306114 306115 306116 306117 306118 306120 306122 306123 306124 306125 306126 306127 306128 306129 306130 306131 306132 306133 306134 306135 306136 306137 306138 306139 306140 306141 306142 306144 306145 306146 306147 306148 306149 306150 306151 306152 306153 306154 306155 306157 306158 306159 306160 306161 306162 306163 306164 306165 306166 306167 306168 306169 306170 306171 306172 306174 306175 306176 306177 306178 306179 306180 306182 306183 306184 306185 306186 306187 306188 306189 306190 306191 306192 306193 306194 306196 306197 306198 306199 306200 306201 306202 306203 306204 306205 306206 306207 306208 306209 306210 306211 306212 306213 306214 306215 306216 306217 306219 306220 306221 306222 306223 306224 306225 306226 306227 306228 306229 306230 306231 306232 306233 306234 306235 306236 306237 306238 306240 306241 306242 306243 306244 306245 306246 306247 306248 306249 306250 306251 306252 306253 306254 306255 306256 306257 306258 306261 306264 306265 306266 306267 306269 306270 306271 306272 306273 306275 306277 306278 306279 306280 306281 306282 306283 306284 306285 306286 306287 306289 306290 306291 306292 306293 306294 306295 306296 306297 306298 306300 306301 306302 306303 306304 306306 306307 306308 306310 306311 306312 306313 306314 306315 306316 306319 306320 306321 306322 306323 306325 306326 306327 306328 306329 306330 306331 306332 323456 323457 323458 323459 323460 323461 323462 323463 323465 323466 323467 323468 323469 323470 323472 323473 323474 323475 323476 323477 323478 323479 323480 323481 323483 323485 323486 323487 323488 323489 323490 323491 323493 323494 323495 323496 323497 323498 323499 323500 323501 323502 323503 323504 323505 323506 323507 323508 323510 323511 323512 323513 323514 323515 323516 323517 323518 323519 323520 323521 323522 323523 323524 323525 323526 323528 323529 323530 323531 323533 323535 323536 323537 323538 323539 323540 323542 323543 323544 323545 323548 323549 323550 323551 323552 323553 323554 323555 323556 323558 323559 323560 323561 323562 323563 323564 323565 323566 323567 323569 323570 323571 323572 323573 323574 323575 323576 323577 323578 323579 323580 323581 323582 323583 323584 323585 323586 323587 323588 323589 323590 323591 323593 323594 323595 323596 323597 323598 323599 323600 323601 323602 323603 323604 323605 323606 323607 323608 323609 323610 323611 323613 323614 323615 323616 323617 323618 323619 323620 323621 323622 323623 323624 323625 323626 323627 323628 323629 323631 323632 323633 323634 323635 323636 323637 323638 323639 323640 323641 323642 323643 323644 323645 323646 323647 323648 323649 323650 323651 323652 323653 323654 323655 323656 323657 323659 323661 323662 323663 323664 323666 323667 323668 323669 323670 323671 323672 323673 323674 323676 323677 323679 323681 323682 323683 323684 323685 323686 323687 323688 323689 323690 323691 323692 323693 323695 323696 323697 323700 323701 323702 323703 323704 323705 323706 323707 323708 323709 323710 323711 323712 323713 323714 323715 323716 323717 323718 323720 323721 323722 323723 323724 323725 323726 323727 323728 323729 323730 323731 323732 323733 323734 323735 323736 323737 323738 323739 323740 323741 323742 323743 323744 323745 323746 323747 323748 323749 323750 323751 323754 323755 323756 323757 323758 323759 323760 323762 323763 323764 323765 323766 323767 323768 323769 323770 323771 323772 323773 323774 323775 323776 323777 323778 323779 323780 323781 323782 323783 323784 323785 323786 323787 323789 323791 323792 323793 323794 323795 323796 323797 323798 323799 323800 323801 323802 323803 323804 323805 323806 323807 323808 323810 323811 323812 323813 323814 323815 323816 323817 323820 323821 323823 323824 323826 323827 323828 323829 323830 323831 323832 323833 323834 323835 323836 323837 323838 323839 323840 323841 323842 323843 323844 323845 323846 323847 323848 323849 323851 323852 323853 323854 323855 323856 323857 323858 323859 323860 323861 323862 323863 323864 323865 323866 323867 323868 323869 323870 323871 323872 323873 323874 323875 323876 323877 323878 323879 323881 323883 323884 323885 323886 323887 323888 323889 323890 323891 323892 323893 323894 323895 323896 323897 323898 323899 323900 323901 323902 323903 323904 323905 323906 323907 323908 323909 323910 323911 323912 323913 323914 323915 323916 323917 323918 323919 323920 323921 323924 323925 323926 323927 323928 323929 323930 323931 323933 323934 323935 323936 323937 323938 323939 323940 323942 323945 323946 323947 323948 323949 323950 323951 323952 323953 323954 323955 323956 323957 323958 323959 323960 323961 323962 323963 323964 323965 323966 323967 323968 323970 323971 323972 323973 323974 323976 323977 323978 323979 323980 323981 323982 323983 323985 323986 323987 323988 323989 323990 323991 323992 323993 323994 323995 323996 323997 323998 323999 324000 324001 324002 324003 324004 324006 324007 324008 324009 324010 324011 324012 324013 324014 324015 324016 324017 324018 324019 324020 324022 324023 324024 324025 324026 324027 324028 324029 324030 324031 324032 324033 324035 324037 324038 324039 324040 324041 324042 324044 324045 324046 324047 324048 324049 324050 324053 324055 324056 324057 324058 324062 324063 324064 324065 324066 324067 324069 324070 324071 324072 324073 324074 324075 324076 324077 324078 324079 324080 324081 324082 324083 324084 324085 324086 324087 324088 324089 324090 324091 324092 324093 324094 324095 324096 324097 324098 324099 324100 324101 324102 324103 324104 324105 324106 324107 324108 324109 324110 324111 324112 324113 324114 324115 324116 324117 324118 324119 324120 324121 324122 324123 324124 324125 324126 324127 324128 324129 324130 324131 324132 324133 324134 324135 324136 324137 324138 324139 324140 324141 324142 324143 324145 324146 324147 324148 324149 324150 324151 324152 324153 324154 324155 324156 324157 324158 324159 324160 324161 324162 324163 324165 324166 324167 324168 324169 324170 324171 324172 324173 324174 324176 324177 324178 324179 324181 324182 324183 324184 324185 324186 324187 324188 324189 324191 324192 324193 324194 324195 324196 324197 324198 324199 324200 324201 324202 324203 324204 324205 324206 324207 324208 324209 324210 324211 324212 324213 324214 324215 324216 324217 324218 324220 324221 324222 324223 324224 324225 324226 324227 324228 324229 324230 324231 324232 324233 324235 324236 324237 324238 324241 324242 324243 324244 324245 324246 324247 324248 324249 324250 324251 324252 324253 324255 324256 324257 324258 324259 324260 324261 324263 324264 324265 324267 324268 324269 324270 324272 324273 324274 324275 324276 324277 324279 324280 324281 324282 324283 324284 324285 324286 324287 324288 324289 324290 324291 324292 324293 324294 324296 324297 324298 324299 324300 324301 324302 324303 324304 324305 324306 324307 324308 324310 324311 324312 324313 324314 324315 324316 324317 324318 324320 324321 324322 324323 324324 324325 324326 324327 324328 324329 324330 324331 324332 324333 324334 324335 324336 324337 324338 324339 324340 324341 324344 324345 324346 324347 324348 324349 324350 324351 324353 324354 324355 324356 324357 324358 324359 324360 324361 324362 324363 324365 324366 324367 324368 324369 324370 324371 324372 324373 324375 324376 324377 324378 324379 324380 324381 324382 324383 324384 324385 324386 324387 324388 324389 324390 324391 324392 324393 324394 324395 324396 324397 324398 324399 324401 324402 324403 324404 324405 324406 324407 324408 324409 324410 324411 324412 324413 324414 324416 324417 324418 324419 324420 324421 324422 324423 324425 324427 324428 324429 324430 324432 324433 324434 324435 324436 324437 324438 324439 324440 324441 324442 324444 324445 324446 324447 324448 324449 324450 324451 324452 324453 324455 324456 324457 324458 324459 324460 324461 324462 324463 324464 324465 324466 324468 324469 324470 324471 324472 324473 324474 324475 324476 324477 324478 324479 324480 324481 324482 324484 324485 324486 324487 324488 324489 324490 324491 324492 324493 324494 324495 324496 324497 324498 324499 324500 324501 324502 324503 324504 324505 324506 324507 324508 324509 324510 324511 324512 324513 324514 324515 324516 324517 324518 324519 324520 324521 324522 324523 324524 324525 324527 324528 324529 324530 324531 324532 324533 324534 324535 324536 324537 324538 324539 324540 324541 324543 324544 324545 324546 324547 324548 324549 324551 324552 324553 324554 324555 324556 324557 324558 324559 324560 324561 324563 324564 324565 324566 324567 324570 324571 324572 324573 324574 324575 324576 324577 324578 324579 324580 324581 324582 324584 324585 324586 324587 324588 324652 324653 324654 324655 324657 324658 324659 324661 324662 324663 324664 324666 324667 324668 324669 324670 324671 324672 324674 324675 324676 324678 324679 324680 324681 324682 324683 324684 324685 324686 324687 324688 324689 324690 324691 324692 324693 324694 324695 324696 324697 324698 324699 324700 324701 324702 324703 324704 324705 324706 324707 324708 324709 324710 324711 324712 324714 324715 324716 324717 324718 324719 324720 324721 324722 324723 324724 324725 324726 324727 324728 324729 324730 324732 324733 324734 324735 324736 324737 324738 324740 324741 324743 324744 324745 324746 324747 324748 324749 324750 324751 324752 324754 324755 324756 324757 324759 324760 324761 324762 324763 324764 324765 324766 324767 324768 324770 324771 324772 324773 324774 324775 324776 324777 324778 324779 324780 324781 324782 324784 324785 324786 324787 324788 324789 324790 324792 324793 324794 324795 324796 324797 324798 324799 324800 324801 324802 324803 324804 324805 324806 324807 324808 324809 324810 324811 324812 324813 324814 324815 324816 324817 324818 324819 324820 324821 324822 324823 324824 324825 324826 324827 324828 324829 324830 324831 324832 324833 324834 324835 324836 324839 324840 324853 324854 324855 324856 324857 324858 324859 324860 324861 324862 324863 324864 324865 324866 324867 324868 324869 324870 324871 324872 324873 324874 324875 324876 324877 324878 324879 324880 324881 324882 324883 324884 324885 324886 324887 324888 324889 324890 324891 324892 324893 324894 324895 324896 324897 324898 324899 324900 324902 324903 324904 324905 324906 324907 324908 324909 324910 324911 324912 324913 324914 324915 324916 324917 324918 324919 324920 324921 324922 324923 324924 324925 324926 324927 324928 324929 324930 324931 324932 324933 324934 324935 324936 324937 324938 324939 324940 324941 324942 324943 324944 324945 324946 324947 324948 324949 324950 324951 324952 324953 324954 324955 324956 324958 324959 324960 324961 324962 324963 324964 324965 324966 324967 324968 324969 324971 324972 324973 324974 324975 324976 324977 324978 324979 324980 324981 324982 324983 324984 324985 324986 324987 324988 324989 324990 324991 324992 324993 324995 324996 324998 324999 325001 325002 325003 325004 325005 325006 325007 325008 325009 325010 325011 325012 325013 325015 325016 325017 325018 325019 325020 325022 325023 325024 325025 325026 325027 325028 325029 325030 325031 325032 325033 325034 325035 325036 325037 325038 325039 325040 325041 325042 325043 325044 325045 325047 325048 325049 325050 325051 325052 325053 325054 325055 325056 325058 325059 325060 325061 325062 325063 325064 325065 325066 325067 325068 325069 325070 325071 325072 325073 325074 325075 325076 325077 325078 325079 325080 325081 325082 325083 325084 325085 325086 325087 325088 325089 325090 325091 325092 325093 325094 325095 325096 325097 325099 325101 325102 325103 325104 325105 325106 325107 325108 325109 325110 325111 325112 325113 325114 325116 325117 325118 325119 325120 325121 325122 325123 325124 325125 325126 325127 325128 325129 325130 325131 325132 325134 325135 325137 325138 325139 325140 325141 325142 325143 325144 325145 325146 325147 325148 325149 325150 325151 325152 325153 325154 325155 325156 325157 325158 325159 325160 325161 325162 325163 325164 325165 325166 325167 325168 325170 325171 325172 325175 325176 325178 325179 325180 325181 325182 325183 325184 325185 325186 325187 325188 325189 325190 325191 325192 325193 325194 325195 325197 325199 325200 325201 325202 325203 325204 325205 325206 325207 325208 325209 325210 325211 325212 325213 325214 325217 325218 325219 325220 325221 325222 325224 325225 325226 325227 325229 325230 325231 325232 325233 325234 325235 325236 325237 325238 325239 325240 325241 325242 325243 325244 325245 325246 325247 325248 325249 325250 325251 325252 325253 325254 325256 325257 325258 325259 325260 325262 325263 325264 325265 325266 325267 325269 325271 325272 325273 325274 325275 325276 325277 325278 325279 325280 325281 325282 325283 325284 325285 325286 325287 325288 325289 325290 325291 325292 325293 325294 325295 325296 325297 325298 325299 325300 325302 325303 325304 325305 325306 325307 325308 325309 325310 325311 325312 325313 325314 325315 325316 325317 325318 325319 325320 325321 325322 325323 325324 325325 325326 325327 325328 325329 325330 325331 325332 325333 325334 325335 325336 325337 325338 325340 325341 325342 325343 325344 325345 325346 325347 325348 325349 325350 325351 325352 325353 325354 325355 325356 325357 325358 325359 325360 325361 325362 325363 325364 325365 325367 325368 325369 325370 325371 325372 325373 325374 325375 325376 325377 325378 325379 325380 325381 325382 325383 325384 325385 325386 325387 325388 325389 325390 325391 325392 325394 325396 325397 325398 325399 325401 325402 325403 325404 325405 325406 325407 325408 325409 325410 325411 325412 325413 325414 325415 325418 325419 325420 325421 325422 325424 325425 325427 325428 325429 325430 325431 325432 325433 325434 325435 325436 325437 325438 325440 325441 325442 325443 325445 325446 325447 325448 325449 325450 325451 325452 325453 325454 325455 325456 325457 325458 325459 325460 325461 325462 325463 325464 325465 325466 325467 325468 325469 325470 331543 331544 331545 331546 331548 331549 331550 331551 331552 331553 331554 331555 331556 331557 331558 331559 331560 331561 331562 331563 331564 331565 331566 331567 331568 331569 331570 331571 331572 331573 331574 331575 331576 331577 331578 331579 331581 331582 331583 331584 331585 331586 331587 331588 331589 331590 331591 331592 331593 331594 331595 331596 331597 331598 331599 331600 331601 331602 331603 331604 331605 331606 331607 331608 331609 331610 331611 331612 331613 331614 331615 331616 331617 331618 331619 331621 331622 331623 331624 331625 331626 331627 331628 331629 331630 331631 331632 331633 331635 331636 331637 331638 331639 331640 331641 331642 331643 331644 331645 331646 331648 331649 331650 331651 331654 331656 331657 331658 331659 331660 331662 331663 331664 331665 331666 331667 331668 331670 331671 331672 331674 331675 331676 331677 331678 331679 331680 331681 331682 331683 331684 331685 331686 331687 331688 331689 331691 331692 331693 331694 331695 331696 331697 331698 331699 331700 331701 331702 331703 331704 331705 331706 331707 331708 331709 331710 331711 331712 331713 331714 331715 331716 331717 331718 331719 331720 331721 331722 331723 331724 331725 331726 331727 331728 331729 331730 331731 331732 331733 331734 331736 331738 331741 331742 331743 331744 331745 331748 331749 331750 331752 331753 331754 331755 331756 331757 331758 331759 331760 331761 331762 331763 331764 331765 331766 331767 331768 331769 331771 331772 331773 331774 331775 331776 331777 331778 331779 331780 331781 331782 331783 331784 331785 331786 331787 331788 331789 331790 331791 331792 331793 331795 331796 331797 331798 331799 331801 331802 331803 331804 331805 331806 331807 331808 331809 331810 331811 331812 331813 331814 331815 331816 331817 331818 331819 331820 331821 331822 331824 331825 331826 331827 331828 331829 331830 331831 331832 331833 331834 331835 331836 331837 331838 331839 331840 331841 331842 331843 331844 331845 331846 331847 331848 331851 331852 331854 331855 331856 331857 331858 331859 331860 331861 331862 331863 331864 331865 331866 331867 331868 331869 331870 331871 331872 331873 331874 331875 331876 331877 331878 331879 331880 331881 331882 331883 331884 331885 331886 331887 331888 331890 331891 331892 331893 331895 331896 331897 331898 331899 331900 331901 331902 331903 331904 331905 331906 331907 331908 331909 331910 331911 331912 331913 331914 331915 331916 331917 331918 331919 331920 331922 331923 331924 331925 331926 331927 331928 331929 331930 331931 331933 331934 331935 331936 331937 331938 331940 331942 331943 331944 331945 331946 331947 331948 331949 331950 331951 331952 331953 331954 331955 331956 331957 331958 331959 331960 331961 331962 331963 331964 331965 331966 331967 331968 331969 331970 331971 331972 331973 331974 331976 331977 331978 331979 331980 331981 331982 331983 331984 331985 331986 331987 331989 331990 331991 331992 331993 331994 331995 331996 331997 331998 331999 332000 332001 332002 332003 332004 332005 332006 332007 332008 332009 332010 332011 332012 332013 332014 332015 332016 332017 332018 332019 332020 332021 332022 332023 332024 332025 332026 332027 332028 332029 332030 332031 332032 332033 332034 332035 332036 332037 332038 332039 332040 332041 332043 332044 332045 332046 332047 332048 332049 332050 332051 332052 332053 332054 332055 332056 332058 332059 332060 332061 332064 332065 332066 332067 332069 332070 332071 332073 332074 332075 332076 332078 332079 332080 332081 332082 332083 332084 332085 332086 332087 332088 332091 332092 332093 332094 332095 332096 332097 332098 332099 332100 332101 332103 332104 332105 332106 332107 332108 332110 332113 332114 332115 332116 332117 332120 332121 332122 332123 332124 332126 332127 332128 332129 332130 332131 332132 332133 332134 332135 332136 332137 332138 332139 332140 332141 332142 332143 332145 332146 332147 332148 332149 332150 332151 332152 332153 332154 332155 332156 332157 332158 332159 332160 332161 332162 332163 332164 332165 332166 332167 332168 332169 332170 332171 332172 332174 332176 332177 332178 332179 332180 332182 332183 332186 332187 332188 332189 332190 332192 332193 332194 332195 332196 332197 332198 332199 332200 332201 332202 332203 332204 332206 332207 332208 332209 332210 332211 332212 332213 332214 332215 332216 332217 332218 332219 332221 332222 332223 332224 332225 332227 332228 332229 332230 332231 332232 332233 332234 332235 332237 332238 332239 332240 332241 332242 332243 332244 332245 332246 332247 332248 332249 332250 332251 332252 332253 332254 332255 332256 332257 332259 332260 332261 332262 332263 332264 332265 332266 332267 332268 332269 332270 332271 332272 332273 332274 332275 332276 332277 332278 332279 332280 332281 332282 332283 332284 332285 332286 332287 332288 332289 332290 332291 332292 332293 332294 332295 332297 332298 332299 332300 332301 332302 332303 332304 332305 332306 332307 332308 332309 332310 332311 332312 332314 332315 332316 332317 332318 332319 332320 332321 332322 332323 332324 332325 332326 332327 332328 332329 332330 332331 332332 332333 332335 332336 332337 332338 332339 332340 332343 332345 332346 332347 332348 332349 332350 332351 332352 332353 332354 332355 332356 332357 332358 332359 332360 332361 332362 332363 332365 332366 332367 332368 332369 332370 332371 332372 332373 332375 332376 332377 332378 332379 332380 332381 332382 332383 332384 332385 332387 332388 332389 332390 332391 332392 332393 332394 332395 332396 332397 332400 332401 332402 332403 332404 332405 332406 332407 332410 332411 332412 332413 332414 332415 332416 332417 332418 332419 332420 332421 332422 332423 332424 332425 332426 332427 332428 332429 332430 332431 332432 332434 332435 332436 332440 332441 332442 332443 332444 332446 332447 332449 332450 332451 332452 332453 332454 332455 332456 332457 332458 332459 332460 332462 332463 332464 332465 332466 332467 332468 332469 332470 332471 332472 332473 332474 332475 332476 332477 332478 332479 332480 332481 332482 332483 332484 332485 332486 332487 332489 332490 332491 332492 332493 332494 332495 332497 332498 332499 332500 332501 332502 332503 332504 332505 332506 332507 332508 332509 332510 332511 332512 332513 332515 332516 332517 332518 332519 332520 332522 332523 332524 332525 332526 332527 332528 332529 332530 332531 332532 332533 332534 332535 332537 332538 332539 332540 332541 332542 332543 332544 332545 332546 332547 332548 332549 332550 332551 332552 332553 332555 332556 332557 332558 332559 332560 332561 332562 332563 332564 332565 332566 332567 332568 332569 332570 332571 332572 332573 332574 332577 332578 332579 332580 332581 332582 332583 332584 332585 332586 332587 332588 332589 332590 332591 332592 332593 332594 332595 332596 332597 332598 332599 332600 332601 332602 332603 332604 332605 332606 332607 332608 332609 332610 332611 332612 332613 332614 332615 332616 332617 332618 332620 332621 332622 332623 332624 332625 332626 332627 332628 332629 332630 332631 332632 332633 332635 332636 332637 332638 332639 332640 332641 332642 332643 332644 332646 332647 332648 332649 332650 332651 332652 332653 332656 332657 332658 332659 332661 332662 332663 332664 332665 332666 332667 332668 332670 332671 332672 332674 332675 332676 332680 332711 332712 332713 332714 332715 332716 332717 332719 332720 332722 332723 332724 332725 332727 332728 332730 332731 332732 332733 332734 332735 332736 332737 332738 332739 332740 332741 332742 332743 332744 332745 332746 332748 332749 332750 332751 332752 332753 332754 332755 332756 332757 332758 332759 332760 332761 332762 332763 332764 332765 332766 332767 332768 332769 332770 332771 332772 332773 332774 332775 332776 332777 332778 332779 332780 332781 332782 332783 332784 332785 332786 332787 332788 332789 332790 332791 332792 332793 332794 332795 332796 332797 332798 332799 332800 332803 332804 332805 332806 332807 332808 332809 332834 332835 332836 332837 332838 332839 332840 332841 332842 332843 332844 332845 332846 332847 332848 332849 332850 332852 332853 332854 332855 332856 332857 332858 332859 332861 332862 332863 332864 332865 332866 332867 332868 332869 332871 332872 332873 332874 332875 332876 332877 332878 332879 332880 332881 332882 332883 332884 332885 332886 332887 332888 332889 332890 332891 332894 332895 332896 332897 332898 332899 332901 332902 332903 332904 332905 332906 332907 332908 332909 332910 332911 332912 332913 332914 332915 332916 332917 332918 332919 332921 332923 332924 332925 332926 332927 332928 332929 332930 332931 332932 332933 332934 332935 332937 332938 332939 332940 332941 332942 332943 332944 332945 332946 332947 332948 332949 332950 332951 332952 332953 332954 332955 332956 332957 332958 332959 332960 332961 332963 332964 332965 332966 332968 332969 332970 332971 332972 332973 332974 332975 332976 332977 332978 332979 332980 332981 332982 332983 332985 332986 332987 332988 332989 332990 332991 332992 332993 332994 332995 332996 332998 332999 333000 333002 333003 333004 333005 333006 333007 333008 333009 333010 333011 333012 333013 333015 333016 333017 333018 333019 333020 333021 333022 333023 333024 333025 333027 333028 333029 333030 333031 333035 333036 333037 333038 333039 333040 333041 333042 333043 333044 333045 333047 333048 333049 333050 333051 333052 333054 333055 333056 333057 333060 333061 333062 333063 333064 333065 333066 333068 333069 333070 333071 333072 333073 333074 333075 333077 333078 333079 333080 333081 333082 333083 333084 333085 333086 333088 333089 333090 333091 333092 333093 333094 333095 333096 333097 333098 333099 333100 333101 333102 333103 333105 333106 333107 333108 333109 333110 333111 333112 333113 333114 333115 333116 333117 333118 333120 333121 333122 333123 333124 333125 333126 333127 333128 333129 333130 333132 333133 333134 333135 333136 333137 333138 333139 333141 333142 333143 333144 333145 333146 333147 333148 333149 333150 333151 333152 333153 333154 333155 333156 333157 333158 333160 333161 333163 333164 333165 333166 333167 333168 333169 333170 333171 333172 333173 333174 333175 333176 333177 333178 333179 333180 333181 333183 333184 333185 333186 333187 333188 333189 333190 333191 333192 333193 333194 333196 333197 333198 333199 333201 333202 333203 333204 333205 333206 333208 333209 333211 333212 333214 333215 333216 333217 333218 333219 333220 333221 333222 333223 333224 333225 333226 333228 333229 333230 333231 333232 333233 333234 333236 333237 333238 333239 333240 333241 333242 333243 333244 333246 333247 333248 333249 333250 333251 333252 333253 333254 333255 333256 333257 333258 333259 333260 333261 333262 333264 333265 333266 333267 333268 333269 333270 333271 333272 333273 333274 333275 333276 333277 333278 333279 333280 333281 333282 333283 333284 333286 333287 333288 333289 333290 333291 333292 333293 333294 333295 333296 333297 333298 333299 333300 333301 333302 333303 333304 333305 333306 333307 333308 333309 333310 333311 333312 333314 333315 333316 333317 333318 333319 333320 333321 333323 333324 333325 333326 333327 333328 333329 333330 333331 333332 333333 333334 333336 333337 333338 333339 333340 333341 333342 333343 333344 333346 333347 333348 333349 333350 333351 333352 333353 333354 333355 333356 333357 333358 333359 333360 333362 333363 333364 333365 333366 333367 333368 333370 333372 333373 333375 333377 333380 333381 333382 333383 333384 333385 333386 333387 333388 333389 333390 333391 333392 333393 333394 333397 333398 333399 333400 333401 333403 333404 333405 333406 333407 333408 333409 333411 333412 333413 333414 333416 333417 333418 333419 333420 333421 333422 333423 333424 333425 333426 333427 333428 333429 333430 333431 333432 333433 333434 333435 333436 333438 333439 333440 333441 333442 333443 333444 333446 333447 333448 333449 333450 333451 333452 333453 333454 333455 333456 333457 333458 333459 333460 333461 333462 333463 333464 333465 333466 333468 333469 333470 333471 333472 333473 333474 333475 333476 333478 333479 333480 333481 333482 333483 333484 333485 333486 333487 333488 333489 333490 333491 333492 333493 333494 333495 333497 333498 333499 333500 333501 333502 333503 333505 333506 333507 333508 333509 333510 333511 333512 333513 333514 333515 333516 333517 333518 333519 333521 333522 333523 333524 333525 333526 333528 333530 333531 333532 333533 333534 333535 333536 333537 333538 333539 333540 333541 333542 333544 333545 333546 333547 333548 333549 333550 333551 333552 333554 333555 333556 333557 333558 333559 333560 333561 333562 333563 333564 333565 333566 333567 333568 333569 333570 333571 333572 333573 333574 333575 333576 333577 333578 333579 333580 333581 333582 333583 333584 333585 333586 333587 333588 333589 333590 333591 333592 333593 333594 333595 333596 333597 333598 333599 333600 333601 333603 333604 333605 333606 333608 333609 333610 333611 333613 333614 333615 333616 333617 333618 333619 333620 333621 333623 333625 333626 333627 333628 333629 333630 333631 333632 333633 333634 333635 333637 333638 333639 333640 333641 333643 333644 333645 333646 333647 333648 333649 333650 333651 333652 333653 333654 333655 333656 333657 333658 333659 333660 333661 333662 333663 333664 333665 333666 333667 333668 333669 333670 333671 333672 333673 333674 333675 333676 333677 333678 333679 333680 333681 333682 333684 333685 333686 333687 333688 333689 333690 333692 333693 333694 333695 333696 333697 333698 333699 333700 333701 333702 333705 333706 333707 333709 333710 333711 333713 333714 333715 333716 333717 333718 333719 333724 333725 333726 333727 333728 333729 333730 333731 333732 333734 333735 333736 333737 333738 333739 333740 333741 333742 333743 333744 333745 333746 333747 333748 333750 333751 333752 333753 333754 333755 333756 333757 333759 333760 333761 333762 333763 333764 333765 333766 333767 333768 333769 333771 333772 333773 333777 333779 333780 333781 333782 333783 333784 333785 333786 333787 333788 333790 333791 333792 333793 333795 333796 333797 333798 333799 333803 333804 333805 333806 333807 333808 333809 333810 333811 333812 333813 333814 333824 333825 333826 333828 333829 333830 333831 333832 333833 333834 333835 333836 333837 333839 333840 333841 333842 333843 333844 333845 333846 333847 333848 333850 333851 333852 333853 333854 333855 333857 333858 333859 333860 333861 333862 333863 333864 333865 333866 333867 333868 333869 333870 333871 333872 333873 333874 333875 333876 333877 333878 333879 333880 333881 333882 333883 333884 333885 333886 333887 333888 333889 333890 333891 333894 333895 333896 333897 333898 333899 333900 333902 333903 333904 333905 333906 333907 333908 333909 333910 333911 333912 333913 333914 333916 333917 333918 333919 333920 333921 333922 333923 333925 333926 333927 333928 333929 333930 333931 333933 333934 333935 333936 333937 333938 333939 333940 333941 333942 333943 333944 333945 333946 333947 333948 333949 333950 333951 333952 333953 333954 333955 333956 333957 333959 333960 333961 333962 333963 333964 333965 333966 333967 333968 333969 333970 333971 333972 333973 333975 333978 333979 333980 333981 333982 333983 333984 333985 333986 333987 333988 333989 333991 333992 333993 333994 333995 333996 333997 333998 333999 334000 334002 334003 334004 334005 334006 334007 334008 334009 334011 334012 334013 334014 334015 334016 334017 334018 334019 334020 334021 334022 334023 334024 334025 334026 334027 334028 334029 334030 334031 334033 334034 334035 334036 334037 334039 334040 334042 334043 334044 334045 334046 334047 334048 334049 334050 334052 334053 334054 334055 334056 334057 334058 334059 334060 334062 334063 334064 334065 334066 334067 334068 334069 334070 334071 334072 334073 334074 334075 334076 334077 334079 334080 334081 334082 334083 334084 334085 334086 334087 334088 334089 334090 334091 334093 334094 334095 334096 334097 334098 334099 334100 334101 334102 334103 334104 334105 334106 334107 334108 334109 334110 334111 334112 334113 334114 334115 334116 334117 334118 334119 334120 334121 334123 334124 334125 334126 334127 334128 334129 334130 334132 334133 334134 334135 334136 334137 334138 334139 334140 334141 334142 334143 334145 334146 334147 334148 334149 334150 334151 334153 334154 334155 334156 334157 334158 334159 334160 334161 334162 334164 334165 334166 334167 334168 334169 334170 334171 334172 334173 334174 334175 334176 334177 334178 334179 334180 334181 334182 334183 334184 334185 334186 334187 334188 334189 334190 334191 334192 334193 334194 334196 334197 334198 334199 334200 334201 334202 334203 334205 334206 334207 334208 334209 334210 334211 334212 334213 334214 334215 334216 334217 334218 334221 334222 334223 334224 334225 334226 334227 334228 334230 334231 334232 334234 334235 334236 334237 334238 334240 334241 334242 334243 334244 334245 334247 334249 334251 334252 334253 334254 334255 334256 334257 334258 334259 334260 334261 334262 334263 334264 334265 334266 334267 334268 334269 334270 334272 334273 334274 334275 334276 334277 334278 334279 334280 334281 334282 334284 334285 334286 334287 334288 334289 334290 334291 334292 334294 334295 334296 334297 334298 334299 334300 334301 334302 334303 334304 334305 334306 334307 334308 334309 334310 334312 334313 334315 334316 334317 334318 334319 334320 334321 334322 334323 334324 334325 334326 334328 334329 334330 334331 334332 334333 334334 334335 334336 334337 334338 334340 334342 334343 334344 334345 334346 334347 334348 334349 334350 334351 334352 334354 334355 334356 334357 334358 334359 334360 334361 334362 334363 334364 334365 334366 334367 334368 334369 334370 334371 334372 334373 334374 334375 334376 334378 334379 334380 334381 334382 334383 334384 334385 334386 334387 334388 334389 334390 334391 334392 334393 334394 334395 334396 334397 334398 334399 334400 334401 334403 334404 334405 334406 334407 334408 334409 334410 334412 334413 334414 334415 334416 334417 334418 334419 334420 334421 334422 334423 334424 334425 334429 334430 334431 334432 334435 334436 334437 334438 334439 334440 334442 334443 334444 334445 334446 334447 334448 334449 334450 334451 334452 334453 334454 334455 334456 334458 334459 334460 334461 334462 334463 334464 334465 334466 334467 334468 334469 334470 334471 334472 334473 334474 334475 334476 334477 334478 334479 334480 334481 334482 334484 334485 334487 334488 334489 334490 334491 334492 334493 334494 334495 334496 334497 334498 334499 334500 334501 334502 334504 334505 334506 334507 334508 334510 334511 334512 334513 334514 334515 334516 334517 334518 334519 334520 334521 334522 334523 334525 334527 334528 334529 334530 334531 334533 334534 334535 334536 334537 334538 334539 334540 334541 334542 334543 334544 334545 334546 334547 334548 334549 334550 334551 334552 334553 334554 334555 334556 334558 334559 334560 334561 334562 334563 334564 334565 334566 334567 334568 334569 334570 334571 334572 334574 334575 334576 334577 334578 334579 334580 334581 334582 334583 334584 334585 334586 334587 334588 334589 334590 334591 334592 334593 334594 334595 334596 334597 334599 334600 334601 334602 334603 334604 334605 334606 334607 334608 334609 334610 334611 334612 334613 334614 334615 334616 334617 334618 334620 334621 334622 334623 334624 334625 334626 334628 334629 334630 334631 334633 334634 334635 334636 334637 334638 334639 334640 334641 334642 334643 334644 334646 334647 334648 334650 334651 334652 334653 334654 334655 334657 334659 334660 334661 334662 334663 334664 334665 334666 334667 334669 334670 334671 334673 334674 334676 334677 334679 334680 334681 334682 334683 334684 334685 334686 334687 334688 334689 334690 334691 334692 334693 334695 334696 334697 334698 334699 334700 334701 334702 334703 334704 334705 334706 334707 334709 334710 334711 334712 334713 334714 334715 334716 334717 334718 334719 334720 334722 334723 334724 334725 334726 334729 334730 334731 334732 334733 334734 334735 334736 334738 334739 334740 334741 334742 334743 334744 334745 334746 334748 334749 334750 334751 334752 334753 334754 334756 334757 334758 334759 334760 334761 334762 334763 334764 334766 334767 334768 334769 334770 334771 334772 334773 334774 334775 334776 334777 334779 334780 334781 334782 334783 334784 334785 334786 334787 334788 334789 334790 334791 334792 334793 334795 334796 334797 334798 334799 334800 334801 334802 334803 334806 334807 334808 334809 334810 334811 334812 334813 334814 334815 334816 334817 334818 334819 334820 334821 334822 334823 334825 334826 334828 334829 334830 334831 334832 334833 334834 334835 334836 334838 334839 334840 334841 334842 334844 334845 334846 334847 334848 334849 334850 334851 334853 334854 334856 334857 334858 334859 334861 334862 334863 334864 334865 334867 334868 334869 334870 334871 334872 334873 334874 334875 334876 334878 334879 334880 334881 334882 334884 334886 334887 334888 334889 334890 334891 334893 334894 334895 334896 334897 334899 334900 334901 334902 334903 334904 334905 334906 334907 334909 334911 334913 334914 334915 334916 334917 334918 334919 334922 334923 334924 334925 334927 334928 334930 334931 334932 334933 334935 334936 334937 334938 334939 334940 334941 334942 334943 334944 334945 334946 334947 334948 334949 334950 334951 334952 334953 334955 334956 334957 334958 334960 334961 334962 334963 334964 334965 334966 334967 334968 334969 334970 334971 334972 334973 334974 334975 334976 334977 334978 334979 334980 334981 334982 334983 334984 334985 334986 334987 334988 334989 334990 334991 334992 334993 334994 334995 334996 334997 334998 334999 335000 335001 335002 335003 335004 335005 335006 335008 335009 335010 335011 335013 335014 335015 335016 335017 335018 335019 335020 335021 335022 335023 335024 335026 335027 335028 335029 335030 335031 335032 335033 335034 335035 335036 335037 335038 335039 335040 335041 335042 335043 335044 335045 335046 335047 335048 335049 335050 335051 335052 335053 335054 335055 335057 335058 335059 335060 335061 335062 335063 335064 335065 335066 335067 335068 335070 335071 335072 335073 335075 335076 335077 335078 335079 335080 335081 335082 335084 335085 335086 335087 335088 335089 335090 335091 335092 335093 335094 335095 335096 335097 335098 335099 335100 335102 335103 335105 335106 335107 335108 335109 335110 335111 335112 335113 335114 335115 335116 335117 335118 335119 335120 335121 335122 335123 335124 335125 335126 335127 335128 335129 335130 335131 335132 335133 335134 335135 335136 335137 335138 335139 335140 335141 335142 335143 335144 335145 335146 335147 335148 335149 335150 335151 335152 335153 335154 335155 335156 335157 335158 335159 335160 335161 335162 335163 335164 335165 335166 335167 335169 335171 335173 335174 335175 335176 335177 335178 335179 335180 335182 335183 335184 335185 335186 335187 335188 335189 335190 335191 335192 335193 335194 335195 335196 335197 335198 335199 335200 335201 335202 335203 335204 335205 335206 335207 335208 335209 335210 335211 335213 335214 335215 335216 335217 335218 335219 335220 335221 335222 335223 335224 335225 335226 335227 335228 335229 335230 335231 335232 335233 335234 335235 335236 335237 335238 335239 335240 335241 335242 335243 335244 335245 335246 335248 335249 335250 335252 335253 335254 335255 335256 335257 335259 335260 335261 335262 335263 335264 335265 335266 335267 335268 335269 335270 335271 335272 335273 335274 335275 335276 335277 335278 335279 335280 335281 335282 335283 335284 335285 335286 335287 335288 335289 335290 335291 335292 335293 335294 335295 335296 335297 335298 338657 338659 338660 338661 338662 338663 338664 338665 338666 338667 338668 338669 338670 338671 338672 338673 338674 338675 338676 338678 338679 338680 338681 338682 338683 338684 338685 338686 338687 338688 338690 338691 338692 338693 338694 338695 338696 338698 338699 338700 338701 338702 338703 338705 338706 338707 338708 338709 338710 338711 338712 338713 338714 338715 338717 338718 338720 338721 338722 338723 338724 338725 338726 338727 338728 338729 338730 338731 338732 338733 338734 338735 338737 338738 338739 338740 338741 338742 338743 338744 338745 338746 338747 338748 338749 338750 338751 338752 338753 338755 338756 338757 338758 338759 338760 338762 338763 338764 338766 338767 338768 338769 338770 338771 338772 338773 338774 338775 338776 338777 338778 338780 338781 338783 338784 338785 338786 338787 338788 338789 338790 338791 338792 338793 338794 338795 338796 338798 338799 338800 338801 338802 338803 338804 338805 338806 338807 338808 338809 338810 338812 338813 338814 338816 338817 338818 338819 338820 338821 338822 338823 338825 338826 338827 338828 338829 338830 338831 338832 338834 338835 338836 338837 338838 338839 338840 338841 338844 338845 338846 338847 338848 338849 338850 338851 338852 338853 338854 338855 338856 338857 338858 338859 338862 338863 338864 338865 338866 338867 338868 338869 338870 338871 338872 338873 338874 338875 338876 338877 338878 338879 338880 338881 338882 338883 338884 338885 338886 338887 338888 338889 338890 338891 338892 338893 338894 338895 338896 338897 338898 338899 338900 338901 338902 338903 338904 338905 338906 338907 338908 338910 338911 338912 338913 338914 338915 338916 338917 338918 338919 338920 338921 338922 338924 338925 338926 338927 338928 338930 338931 338934 338935 338938 338939 338940 338941 338942 338943 338944 338945 338946 338947 338948 338949 338950 338951 338952 338954 338955 338956 338957 338958 338959 338961 338962 338963 338964 338965 338966 338967 338968 338969 338970 338971 338972 338973 338974 338976 338977 338978 338980 338981 338983 338984 338985 338986 338987 338988 338989 338990 338991 338992 338993 338994 338995 338996 338997 338998 339000 339004 339005 339006 339007 339008 339010 339011 339012 339013 339014 339015 339016 339017 339018 339019 339020 339021 339023 339024 339025 339026 339027 339028 339029 339031 339032 339033 339034 339036 339037 339038 339040 339041 339042 339043 339044 339045 339046 339047 339048 339049 339050 339051 339052 339054 339055 339056 339057 339058 339059 339060 339061 339062 339063 339064 339065 339066 339067 339068 339069 339070 339071 339072 339073 339074 339075 339076 339077 339078 339079 339080 339081 339082 339083 339084 339085 339087 339088 339089 339090 339091 339092 339093 339094 339095 339096 339097 339099 339100 339102 339103 339104 339105 339106 339107 339108 339109 339110 339111 339112 339113 339114 339115 339116 339117 339118 339119 339120 339121 339122 339123 339124 339125 339127 339128 339129 339130 339131 339132 339133 339134 339135 339136 339137 339138 339139 339140 339141 339142 339143 339144 339145 339146 339147 339148 339149 339150 339151 339152 339153 339154 339155 339156 339157 339158 339159 339164 339165 339166 339167 339168 339169 339170 339171 339173 339174 339176 339177 339178 339180 339182 339183 339185 339186 339188 339190 339195 339197 339198 339199 339201 339202 339203 339204 339205 339206 339207 339208 339209 339211 339212 339214 339215 339216 339217 339218 339219 339220 339221 339223 339224 339225 339226 339227 339228 339229 339230 339231 339232 339233 339234 339236 339238 339239 339240 339241 339242 339243 339244 339245 339247 339248 339250 339251 339252 339253 339254 339255 339256 339257 339258 339259 339260 339261 339262 339263 339264 339265 339266 339267 339269 339270 339271 339272 339273 339274 339275 339276 339277 339278 339279 339280 339281 339282 339283 339284 339285 339286 339287 339288 339289 339291 339293 339294 339295 339297 339298 339299 339300 339301 339302 339303 339304 339306 339307 339308 339310 339311 339312 339313 339314 339315 339316 339317 339318 339319 339320 339321 339322 339323 339324 339325 339326 339327 339328 339329 339330 339331 339332 339333 339334 339335 339336 339338 339339 339340 339342 339343 339344 339345 339346 339347 339348 339349 339350 339351 339352 339353 339355 339356 339359 339360 339361 339362 339363 339364 339365 339366 339367 339368 339433 339434 339435 339436 339437 339439 339440 339441 339442 339443 339444 339445 339446 339447 339448 339449 339450 339451 339452 339453 339454 339455 339456 339457 339458 339459 339460 339461 339462 339463 339464 339465 339466 339467 339468 339470 339471 339472 339473 339474 339475 339476 339477 339478 339479 339480 339481 339483 339484 339485 339486 339487 339489 339490 339491 339492 339493 339494 339495 339496 339497 339498 339499 339500 339501 339502 339503 339504 339505 339506 339507 339508 339510 339511 339512 339513 339515 339516 339517 339518 339519 339520 339521 339522 339523 339524 339525 339526 339527 339528 339529 339530 339531 339532 339533 339534 339536 339537 339538 339539 339540 339543 339544 339545 339546 339547 339548 339549 339550 339551 339553 339554 339555 339556 339557 339558 339560 339561 339562 339563 339565 339566 339567 339568 339569 339570 339571 339572 339573 339574 339575 339576 339577 339578 339579 339580 339581 339582 339584 339585 339586 339587 339588 339589 339590 339591 339592 339593 339594 339595 339596 339597 339598 339599 339600 339601 339602 339603 339604 339606 339607 339608 339609 339610 339611 339612 339613 339614 339615 339617 339618 339619 339620 339621 339622 339623 339624 339625 339626 339627 339628 339629 339630 339631 339632 339633 339634 339635 339636 339637 339638 339639 339640 339641 339642 339643 339644 339645 339646 339647 339648 339649 339650 339651 339652 339653 339654 339655 339657 339658 339659 339660 339661 339662 339663 339664 339665 339667 339668 339670 339671 339672 339673 339674 339675 339676 339677 339678 339679 339680 339681 339682 339683 339684 339685 339686 339687 339688 339689 339690 339691 339692 339693 339694 339695 339696 339697 339698 339699 339700 339701 339702 339703 339704 339705 339706 339707 339708 339709 339710 339711 339712 339713 339714 339715 339716 339717 339718 339719 339720 339721 339722 339723 339724 339725 339726 339727 339728 339729 339731 339732 339733 339734 339735 339736 339737 339738 339739 339740 339741 339742 339743 339744 339745 339746 339747 339748 339749 339750 339751 339752 339753 339754 339755 339756 339757 339758 339759 339762 339763 339764 339765 339766 339767 339768 339769 339770 339771 339773 339774 339775 339776 339777 339778 339779 339780 339781 339782 339783 339784 339785 339786 339787 339788 339789 339791 339792 339793 339794 339795 339797 339798 339800 339801 339803 339804 339805 339806 339807 339809 339810 339811 339812 339813 339814 339815 339816 339817 339818 339819 339820 339821 339822 339823 339824 339825 339826 339829 339831 339832 339833 339834 339835 339836 339837 339838 339839 339840 339841 339842 339843 339844 339845 339846 339847 339848 339849 339850 339851 339852 339853 339854 339855 339856 339857 339858 339859 339861 339862 339863 339864 339865 339866 339867 339868 339869 339870 339871 339872 339873 339874 339875 339876 339877 339878 339879 339881 339882 339883 339884 339885 339886 339887 339888 339889 339890 339891 339892 339893 339894 339895 339897 339898 339899 339900 339901 339902 339903 339905 339906 339907 339909 339910 339911 339912 339913 339914 339917 339918 339919 339920 339921 339922 339923 339924 339925 339926 339927 339928 339929 339930 339931 339932 339933 339934 339935 339936 339937 339938 339939 339940 339941 339942 339944 339945 339946 339947 339948 339949 339951 339952 339953 339954 339955 339956 339957 339958 339960 339961 339962 339963 339964 339965 339966 339968 339969 339971 339972 339973 339974 339975 339976 339977 339978 339979 339980 339981 339982 339983 339984 339985 339986 339987 339988 339990 339991 339992 339993 339994 339995 339996 339997 339998 340002 340003 340004 340006 340007 340008 340009 340010 340011 340012 340013 340014 340015 340016 340017 340018 340019 340020 340021 340022 340023 340024 340025 340026 340027 340028 340029 340030 340031 340032 340033 340034 340035 340036 340037 340038 340039 340040 340041 340042 340043 340044 340046 340047 340048 340049 340050 340051 340052 340053 340054 340057 340058 340059 340060 340061 340062 340063 340064 340065 340066 340067 340068 340069 340070 340071 340074 340077 340078 340079 340080 340081 340083 340084 340085 340087 340088 340090 340091 340092 340093 340094 340095 340096 340097 340099 340100 340101 340102 340103 340104 340105 340107 340123 340124 340125 340126 340127 340128 340130 340131 340132 340133 340134 340136 340137 340138 340139 340140 340141 340143 340144 340145 340146 340147 340148 340149 340150 340151 340152 340153 340154 340155 340156 340158 340159 340161 340162 340163 340164 340165 340166 340167 340168 340169 340170 340171 340172 340173 340174 340175 340176 340177 340178 340179 340180 340182 340184 340185 340186 340187 340188 340189 340190 340191 340192 340193 340194 340195 340196 340197 340198 340199 340200 340201 340202 340203 340205 340206 340209 340210 340211 340212 340213 340215 340216 340217 340218 340219 340220 340221 340222 340223 340224 340225 340226 340227 340228 340229 340230 340231 340232 340233 340234 340235 340236 340237 340238 340239 340240 340241 340242 340244 340245 340246 340247 340249 340250 340251 340252 340253 340254 340255 340257 340258 340259 340265 340266 340267 340268 340270 340272 340273 340274 340275 340276 340277 340278 340279 340281 340282 340283 340284 340285 340286 340287 340288 340289 340290 340291 340292 340296 340297 340298 340299 340300 340301 340302 340303 340304 340305 340307 340308 340309 340310 340311 340312 340313 340314 340315 340316 340317 340318 340319 340321 340322 340324 340325 340328 340330 340331 340332 340334 340335 340336 340337 340338 340339 340340 340341 340342 340343 340344 340345 340346 340347 340348 340349 340350 340351 340352 340353 340354 340355 340356 340357 340358 340359 340360 340361 340362 340363 340365 340366 340367 340368 340369 340370 340371 340372 340373 340375 340377 340378 340381 340382 340383 340384 340386 340387 340388 340389 340390 340391 340392 340393 340394 340395 340396 340397 340398 340399 340400 340401 340402 340403 340404 340405 340407 340408 340409 340410 340411 340413 340414 340415 340416 340417 340418 340420 340421 340422 340423 340424 340428 340429 340430 340431 340432 340433 340434 340435 340436 340437 340438 340439 340440 340441 340444 340445 340446 340449 340451 340452 340455 340456 340457 340458 340459 340460 340461 340462 340463 340464 340465 340467 340468 340469 340470 340472 340473 340474 340475 340476 340477 340478 340479 340480 340482 340483 340485 340486 340488 340489 340491 340492 340493 340494 340495 340496 340497 340499 340500 340501 340502 340503 340504 340506 340507 340508 340509 340510 340511 340512 340513 340514 340515 340516 340517 340518 340519 340520 340521 340522 340523 340524 340525 340526 340527 340528 340531 340532 340533 340534 340535 340536 340537 340538 340539 340540 340541 340542 340543 340544 340545 340546 340547 340548 340549 340550 340552 340553 340554 340555 340556 340557 340558 340559 340560 340561 340562 340563 340564 340565 340566 340568 340590 340591 340592 340593 340594 340595 340596 340597 340599 340600 340601 340602 340603 340604 340605 340606 340607 340608 340609 340610 340611 340612 340613 340614 340615 340616 340617 340618 340619 340620 340621 340622 340624 340625 340627 340628 340629 340630 340631 340632 340633 340634 340635 340636 340637 340638 340639 340640 340641 340642 340643 340644 340645 340646 340647 340648 340649 340650 340651 340652 340653 340654 340656 340657 340658 340659 340660 340661 340662 340664 340665 340666 340668 340669 340671 340672 340674 340675 340676 340677 340678 340679 340680 340681 340682 340683 340684 340685 340686 340687 340688 340689 340690 340691 340692 340693 340694 340695 340696 340697 340698 340699 340700 340701 340702 340704 340705 21517 21518 215190 215200 215210 215220 21530 21531 21532 21533 21534 21535 21536 215380 215400 21541 21542 21543 21544 21545 21546 21547 21548 21550 21551 21553 21554 21555 21556 21558 21559 21560 21562 21563 215640 215670 215680 21569 21570 21571 21573 21575 215770 21580 215810 215820 21583 215840 21585 21586 21589 215900 21591 21593 215940 215950 21597 21598 215990 216000 216020 216030 216040 216050 21606 21608 21609 216110 21616 21617 216180 216200 21621 216220 216250 216270 216280 216290 216300 216310 21632 216330 216340 21635 21636 21637 21638 21639 21640 216410 21642 216430 216440 216450 216460 216470 21648 216500 21651 21653 21654 21655 21659 21660 216620 21663 21664 216650 216670 21669 21670 21672 21673 21676 21677 21678 216790 216800 21681 216830 21684 21685 21686 21687 21688 21689 21690 21691 21692 21693 21694 21695 21696 21697 21699 21700 21701 21702 21704 21705 21706 21708 21710 21711 21713 21714 21715 21718 21719 21720 21721 21722 21723 21724 21725 21726 21728 21729 21731 21733 217350 21736 217380 21749 21750 21752 21754 21757 21758 21760 21762 21763 21766 21768 21769 21771 21772 21774 21775 21776 21777 21778 21779 21782 21783 21784 21786 21787 21788 217890 21790 21791 21792 21794 21800 21801 21802 21803 21804 21825 21826 21827 21829 21830 21832 21833 21834 21835 21836 21837 21841 21844 218460 21849 21851 21852 21853 21854 21855 21856 21857 21860 21861 21862 21863 21864 21865 21866 21867 218700 21871 21872 21873 21875 21880 21881 21882 21883 218840 21887 21888 21889 21892 21894 21896 21898 21899 21900 21903 21905 21907 219080 21909 219100 21911 21912 21913 219140 21917 219180 219190 219250 21926 21927 21928 21929 21931 21932 21937 21940 21941 21943 21944 21946 21947 21949 219500 219510 219520 219540 21955 21956 21957 21958 219600 21961 21963 21964 21965 21966 21967 21968 21969 21970 21971 21973 21974 21975 219760 219770 21978 21979 21980 21981 21982 21983 21984 21986 21987 21988 21990 21991 21992 21993 21994 21995 21996 21997 21998 21999 22001 22002 220030 22004 220050 22006 22007 22008 22009 22010 22013 22015 22016 22020 22021 22022 22023 22026 22027 22028 22029 22030 22031 22032 22033 22034 22037 220390 220400 220410 22042 22043 22044 22045 22046 22047 22049 220500 220510 22052 22053 220540 22055 220570 220580 220590 220600 220610 220630 220640 220650 220680 220690 220700 220710 220720 220730 220750 220760 220780 220790 220800 220840 220850 220860 220870 220880 220890 220900 220920 220950 220960 220970 220980 221000 221030 221050 221060 221080 221090 221100 221110 221120 221140 221150 221160 221180 221200 221220 221230 221250 22127 221280 221290 221310 221320 221340 221350 221360 221390 221400 221410 221470 221480 221490 221500 221530 221550 221570 221590 221600 221620 221650 221660 221670 221680 221690 221700 221710 221760 22178 22179 22180 22181 22184 22186 22187 22188 22190 22191 22192 22194 22195 221960 22197 22198 222000 222010 222020 222030 222040 222060 222080 222090 222110 222120 222130 222140 222150 222160 222200 222220 22223 222250 222260 222270 222340 222350 222400 222410 222420 222440 222470 222490 222500 222520 222540 222590 222600 222610 222620 222630 222640 222670 222690 222720 222730 222740 222750 222760 222770 222780 222790 222800 222810 222820 222830 222840 222850 222860 22287 222880 222910 222920 222930 22294 222950 222970 223000 223020 223050 22308 22309 223100 223140 223160 223190 223200 223220 223230 223240 223250 223260 223290 22330 223320 223340 22335 22336 223380 223390 223400 223430 223440 223450 223480 223500 223580 223600 223610 223640 223650 223660 223670 223680 223690 223710 223720 223740 223750 223770 223780 223790 223820 223830 223870 223900 223920 223930 223940 223950 223960 22397 223980 224000 22401 224030 224040 22406 224080 224090 224100 224110 224120 224130 224140 22415 224160 22417 22418 224190 224200 224210 224220 22423 224240 224250 224270 22428 22429 22430 224320 224350 224390 22440 224410 22442 22443 224440 224490 224500 224520 22453 224540 224550 224560 22457 224590 22460 224610 224620 224630 224640 224650 22466 22467 22468 22470 22473 224750 22477 22480 224840 22485 224860 224870 224880 224890 224900 224920 224940 22495 224960 224970 224980 22499 225000 225010 225030 225050 225060 22507 225080 225090 225120 225130 225150 225190 225250 225270 225290 225420 225430 225490 225510 225550 225580 225600 225620 225690 225700 225730 225750 225790 225860 22587 225880 22589 225900 225970 226000 226010 226020 226040 226050 226070 226090 226100 226120 22614 22615 226170 226210 226230 226240 22625 22626 22629 226330 226340 226350 226360 22637 22639 226400 22642 226430 22644 22645 22646 22647 22648 226510 22652 22653 226540 226560 226590 22660 226610 22662 22663 226650 226670 226680 226690 22670 226710 226740 226750 226760 226770 226780 226790 226800 226810 226860 226870 226880 226890 226900 226920 226930 226940 226950 226960 226970 226990 227000 227010 227020 227040 227050 227060 227110 227130 227160 227170 227180 227190 22720 22721 22723 22724 22726 22728 22730 22731 227320 22734 227350 22736 22737 22738 22739 227410 227420 227440 22745 22747 227490 22751 22752 22754 22755 22756 22757 227580 227590 227610 227620 22763 227650 227670 227680 227690 227700 227710 227720 227730 22774 227750 227770 22779 227820 22783 227840 22785 227860 22787 227880 227910 227920 227940 227950 227960 227970 227980 228010 228020 228030 228040 228050 228080 228090 228110 228120 228130 228140 228160 228170 228180 228190 228200 228210 228230 228240 228260 228270 228280 228300 228310 228320 228330 228340 22835 228360 228380 228390 228400 228410 228430 228440 228450 228470 228480 22849 22850 22851 22852 22853 22854 22855 22856 22857 228590 22860 228610 228620 228630 228650 228660 22869 22870 228710 228730 22874 22875 228760 228770 22881 22883 22884 22885 22887 22892 22893 22895 22896 22899 22902 229040 229050 229060 229070 229080 229100 229130 229140 229170 229180 229200 229250 229260 229270 229280 229300 229310 229330 229350 229360 229370 229390 229400 229410 229420 229430 229440 229460 229470 229480 229490 229500 229510 229520 229530 229540 229550 229560 229570 22958 22959 229600 22961 22962 22965 229670 22968 22980 22981 22982 229830 229840 229860 229880 229890 229920 229930 229950 229960 229980 22999 230010 23003 230050 23006 230070 230080 23010 230120 230130 230140 230150 230160 230170 230180 230190 230200 230210 230220 230240 230260 230280 230290 230300 230320 230330 230340 230350 230360 230370 230380 23039 23040 23041 23042 23044 230450 230460 23047 23048 230490 230510 230530 230540 230550 230560 230570 230590 230600 230610 230620 230630 230660 230670 230680 230690 230720 23074 230750 230760 230780 23079 230800 230820 230830 230850 230880 230890 230900 230910 230930 230970 231000 231020 231030 231040 231050 231070 231090 231100 231120 231130 231140 23115 231160 231170 231180 231190 231200 23121 231220 23123 231240 23127 23129 231320 231340 231360 231370 231380 231390 231400 231410 231420 231430 231440 231460 231470 231490 231500 231520 231530 231540 231550 231560 231570 231580 231590 231600 231610 231640 231650 231660 231670 231680 231690 231700 231710 231730 231750 231770 231780 231810 231840 231860 231870 231890 231900 231910 231950 23198 231990 232000 232030 232060 23209 232120 232140 232160 232170 232180 232190 232200 232220 232230 232250 232260 232310 232320 232350 232380 232390 232400 232410 232460 232490 23252 232530 232560 232570 23258 232590 232610 232620 232640 232650 232670 232680 232690 306081 306082 306083 306087 306088 306089 306091 306092 306093 306094 306095 306096 306098 306099 306100 306102 306103 306104 306106 306107 306108 306109 306110 306111 306113 306114 306115 306116 306117 306118 306120 306122 306123 306124 306125 306126 306127 306128 306129 306130 306131 306132 306133 306134 306135 306136 306137 306138 306139 306140 306141 306142 306144 306145 306146 306147 306148 306149 306150 306151 306152 306153 306154 306155 306157 306158 306159 306160 306161 306162 306163 306164 306165 306166 306167 306168 306169 306170 306171 306172 306174 306175 306176 306177 306178 306179 306180 306182 306183 306184 306185 306186 306187 306188 306189 306190 306191 306192 306193 306194 306196 306197 306198 306199 306200 306201 306202 306203 306204 306205 306206 306207 306208 306209 306210 306211 306212 306213 306214 306215 306216 306217 306219 306220 306221 306222 306223 306224 306225 306226 306227 306228 306229 306230 306231 306232 306233 306234 306235 306236 306237 306238 306240 306241 306242 306243 306244 306245 306246 306247 306248 306249 306250 306251 306252 306253 306254 306255 306256 306257 306258 306261 306264 306265 306266 306267 306269 306270 306271 306272 306273 306275 306277 306278 306279 306280 306281 306282 306283 306284 306285 306286 306287 306289 306290 306291 306292 306293 306294 306295 306296 306297 306298 306300 306301 306302 306303 306304 306306 306307 306308 306310 306311 306312 306313 306314 306315 306316 306319 306320 306321 306322 306323 306325 306326 306327 306328 306329 306330 306331 306332 323456 323457 323458 323459 323460 323461 323462 323463 323465 323466 323467 323468 323469 323470 323472 323473 323474 323475 323476 323477 323478 323479 323480 323481 323483 323485 323486 323487 323488 323489 323490 323491 323493 323494 323495 323496 323497 323498 323499 323500 323501 323502 323503 323504 323505 323506 323507 323508 323510 323511 323512 323513 323514 323515 323516 323517 323518 323519 323520 323521 323522 323523 323524 323525 323526 323528 323529 323530 323531 323533 323535 323536 323537 323538 323539 323540 323542 323543 323544 323545 323548 323549 323550 323551 323552 323553 323554 323555 323556 323558 323559 323560 323561 323562 323563 323564 323565 323566 323567 323569 323570 323571 323572 323573 323574 323575 323576 323577 323578 323579 323580 323581 323582 323583 323584 323585 323586 323587 323588 323589 323590 323591 323593 323594 323595 323596 323597 323598 323599 323600 323601 323602 323603 323604 323605 323606 323607 323608 323609 323610 323611 323613 323614 323615 323616 323617 323618 323619 323620 323621 323622 323623 323624 323625 323626 323627 323628 323629 323631 323632 323633 323634 323635 323636 323637 323638 323639 323640 323641 323642 323643 323644 323645 323646 323647 323648 323649 323650 323651 323652 323653 323654 323655 323656 323657 323659 323661 323662 323663 323664 323666 323667 323668 323669 323670 323671 323672 323673 323674 323676 323677 323679 323681 323682 323683 323684 323685 323686 323687 323688 323689 323690 323691 323692 323693 323695 323696 323697 323700 323701 323702 323703 323704 323705 323706 323707 323708 323709 323710 323711 323712 323713 323714 323715 323716 323717 323718 323720 323721 323722 323723 323724 323725 323726 323727 323728 323729 323730 323731 323732 323733 323734 323735 323736 323737 323738 323739 323740 323741 323742 323743 323744 323745 323746 323747 323748 323749 323750 323751 323754 323755 323756 323757 323758 323759 323760 323762 323763 323764 323765 323766 323767 323768 323769 323770 323771 323772 323773 323774 323775 323776 323777 323778 323779 323780 323781 323782 323783 323784 323785 323786 323787 323789 323791 323792 323793 323794 323795 323796 323797 323798 323799 323800 323801 323802 323803 323804 323805 323806 323807 323808 323810 323811 323812 323813 323814 323815 323816 323817 323820 323821 323823 323824 323826 323827 323828 323829 323830 323831 323832 323833 323834 323835 323836 323837 323838 323839 323840 323841 323842 323843 323844 323845 323846 323847 323848 323849 323851 323852 323853 323854 323855 323856 323857 323858 323859 323860 323861 323862 323863 323864 323865 323866 323867 323868 323869 323870 323871 323872 323873 323874 323875 323876 323877 323878 323879 323881 323883 323884 323885 323886 323887 323888 323889 323890 323891 323892 323893 323894 323895 323896 323897 323898 323899 323900 323901 323902 323903 323904 323905 323906 323907 323908 323909 323910 323911 323912 323913 323914 323915 323916 323917 323918 323919 323920 323921 323924 323925 323926 323927 323928 323929 323930 323931 323933 323934 323935 323936 323937 323938 323939 323940 323942 323945 323946 323947 323948 323949 323950 323951 323952 323953 323954 323955 323956 323957 323958 323959 323960 323961 323962 323963 323964 323965 323966 323967 323968 323970 323971 323972 323973 323974 323976 323977 323978 323979 323980 323981 323982 323983 323985 323986 323987 323988 323989 323990 323991 323992 323993 323994 323995 323996 323997 323998 323999 324000 324001 324002 324003 324004 324006 324007 324008 324009 324010 324011 324012 324013 324014 324015 324016 324017 324018 324019 324020 324022 324023 324024 324025 324026 324027 324028 324029 324030 324031 324032 324033 324035 324037 324038 324039 324040 324041 324042 324044 324045 324046 324047 324048 324049 324050 324053 324055 324056 324057 324058 324062 324063 324064 324065 324066 324067 324069 324070 324071 324072 324073 324074 324075 324076 324077 324078 324079 324080 324081 324082 324083 324084 324085 324086 324087 324088 324089 324090 324091 324092 324093 324094 324095 324096 324097 324098 324099 324100 324101 324102 324103 324104 324105 324106 324107 324108 324109 324110 324111 324112 324113 324114 324115 324116 324117 324118 324119 324120 324121 324122 324123 324124 324125 324126 324127 324128 324129 324130 324131 324132 324133 324134 324135 324136 324137 324138 324139 324140 324141 324142 324143 324145 324146 324147 324148 324149 324150 324151 324152 324153 324154 324155 324156 324157 324158 324159 324160 324161 324162 324163 324165 324166 324167 324168 324169 324170 324171 324172 324173 324174 324176 324177 324178 324179 324181 324182 324183 324184 324185 324186 324187 324188 324189 324191 324192 324193 324194 324195 324196 324197 324198 324199 324200 324201 324202 324203 324204 324205 324206 324207 324208 324209 324210 324211 324212 324213 324214 324215 324216 324217 324218 324220 324221 324222 324223 324224 324225 324226 324227 324228 324229 324230 324231 324232 324233 324235 324236 324237 324238 324241 324242 324243 324244 324245 324246 324247 324248 324249 324250 324251 324252 324253 324255 324256 324257 324258 324259 324260 324261 324263 324264 324265 324267 324268 324269 324270 324272 324273 324274 324275 324276 324277 324279 324280 324281 324282 324283 324284 324285 324286 324287 324288 324289 324290 324291 324292 324293 324294 324296 324297 324298 324299 324300 324301 324302 324303 324304 324305 324306 324307 324308 324310 324311 324312 324313 324314 324315 324316 324317 324318 324320 324321 324322 324323 324324 324325 324326 324327 324328 324329 324330 324331 324332 324333 324334 324335 324336 324337 324338 324339 324340 324341 324344 324345 324346 324347 324348 324349 324350 324351 324353 324354 324355 324356 324357 324358 324359 324360 324361 324362 324363 324365 324366 324367 324368 324369 324370 324371 324372 324373 324375 324376 324377 324378 324379 324380 324381 324382 324383 324384 324385 324386 324387 324388 324389 324390 324391 324392 324393 324394 324395 324396 324397 324398 324399 324401 324402 324403 324404 324405 324406 324407 324408 324409 324410 324411 324412 324413 324414 324416 324417 324418 324419 324420 324421 324422 324423 324425 324427 324428 324429 324430 324432 324433 324434 324435 324436 324437 324438 324439 324440 324441 324442 324444 324445 324446 324447 324448 324449 324450 324451 324452 324453 324455 324456 324457 324458 324459 324460 324461 324462 324463 324464 324465 324466 324468 324469 324470 324471 324472 324473 324474 324475 324476 324477 324478 324479 324480 324481 324482 324484 324485 324486 324487 324488 324489 324490 324491 324492 324493 324494 324495 324496 324497 324498 324499 324500 324501 324502 324503 324504 324505 324506 324507 324508 324509 324510 324511 324512 324513 324514 324515 324516 324517 324518 324519 324520 324521 324522 324523 324524 324525 324527 324528 324529 324530 324531 324532 324533 324534 324535 324536 324537 324538 324539 324540 324541 324543 324544 324545 324546 324547 324548 324549 324551 324552 324553 324554 324555 324556 324557 324558 324559 324560 324561 324563 324564 324565 324566 324567 324570 324571 324572 324573 324574 324575 324576 324577 324578 324579 324580 324581 324582 324584 324585 324586 324587 324588 324652 324653 324654 324655 324657 324658 324659 324661 324662 324663 324664 324666 324667 324668 324669 324670 324671 324672 324674 324675 324676 324678 324679 324680 324681 324682 324683 324684 324685 324686 324687 324688 324689 324690 324691 324692 324693 324694 324695 324696 324697 324698 324699 324700 324701 324702 324703 324704 324705 324706 324707 324708 324709 324710 324711 324712 324714 324715 324716 324717 324718 324719 324720 324721 324722 324723 324724 324725 324726 324727 324728 324729 324730 324732 324733 324734 324735 324736 324737 324738 324740 324741 324743 324744 324745 324746 324747 324748 324749 324750 324751 324752 324754 324755 324756 324757 324759 324760 324761 324762 324763 324764 324765 324766 324767 324768 324770 324771 324772 324773 324774 324775 324776 324777 324778 324779 324780 324781 324782 324784 324785 324786 324787 324788 324789 324790 324792 324793 324794 324795 324796 324797 324798 324799 324800 324801 324802 324803 324804 324805 324806 324807 324808 324809 324810 324811 324812 324813 324814 324815 324816 324817 324818 324819 324820 324821 324822 324823 324824 324825 324826 324827 324828 324829 324830 324831 324832 324833 324834 324835 324836 324839 324840 324853 324854 324855 324856 324857 324858 324859 324860 324861 324862 324863 324864 324865 324866 324867 324868 324869 324870 324871 324872 324873 324874 324875 324876 324877 324878 324879 324880 324881 324882 324883 324884 324885 324886 324887 324888 324889 324890 324891 324892 324893 324894 324895 324896 324897 324898 324899 324900 324902 324903 324904 324905 324906 324907 324908 324909 324910 324911 324912 324913 324914 324915 324916 324917 324918 324919 324920 324921 324922 324923 324924 324925 324926 324927 324928 324929 324930 324931 324932 324933 324934 324935 324936 324937 324938 324939 324940 324941 324942 324943 324944 324945 324946 324947 324948 324949 324950 324951 324952 324953 324954 324955 324956 324958 324959 324960 324961 324962 324963 324964 324965 324966 324967 324968 324969 324971 324972 324973 324974 324975 324976 324977 324978 324979 324980 324981 324982 324983 324984 324985 324986 324987 324988 324989 324990 324991 324992 324993 324995 324996 324998 324999 325001 325002 325003 325004 325005 325006 325007 325008 325009 325010 325011 325012 325013 325015 325016 325017 325018 325019 325020 325022 325023 325024 325025 325026 325027 325028 325029 325030 325031 325032 325033 325034 325035 325036 325037 325038 325039 325040 325041 325042 325043 325044 325045 325047 325048 325049 325050 325051 325052 325053 325054 325055 325056 325058 325059 325060 325061 325062 325063 325064 325065 325066 325067 325068 325069 325070 325071 325072 325073 325074 325075 325076 325077 325078 325079 325080 325081 325082 325083 325084 325085 325086 325087 325088 325089 325090 325091 325092 325093 325094 325095 325096 325097 325099 325101 325102 325103 325104 325105 325106 325107 325108 325109 325110 325111 325112 325113 325114 325116 325117 325118 325119 325120 325121 325122 325123 325124 325125 325126 325127 325128 325129 325130 325131 325132 325134 325135 325137 325138 325139 325140 325141 325142 325143 325144 325145 325146 325147 325148 325149 325150 325151 325152 325153 325154 325155 325156 325157 325158 325159 325160 325161 325162 325163 325164 325165 325166 325167 325168 325170 325171 325172 325175 325176 325178 325179 325180 325181 325182 325183 325184 325185 325186 325187 325188 325189 325190 325191 325192 325193 325194 325195 325197 325199 325200 325201 325202 325203 325204 325205 325206 325207 325208 325209 325210 325211 325212 325213 325214 325217 325218 325219 325220 325221 325222 325224 325225 325226 325227 325229 325230 325231 325232 325233 325234 325235 325236 325237 325238 325239 325240 325241 325242 325243 325244 325245 325246 325247 325248 325249 325250 325251 325252 325253 325254 325256 325257 325258 325259 325260 325262 325263 325264 325265 325266 325267 325269 325271 325272 325273 325274 325275 325276 325277 325278 325279 325280 325281 325282 325283 325284 325285 325286 325287 325288 325289 325290 325291 325292 325293 325294 325295 325296 325297 325298 325299 325300 325302 325303 325304 325305 325306 325307 325308 325309 325310 325311 325312 325313 325314 325315 325316 325317 325318 325319 325320 325321 325322 325323 325324 325325 325326 325327 325328 325329 325330 325331 325332 325333 325334 325335 325336 325337 325338 325340 325341 325342 325343 325344 325345 325346 325347 325348 325349 325350 325351 325352 325353 325354 325355 325356 325357 325358 325359 325360 325361 325362 325363 325364 325365 325367 325368 325369 325370 325371 325372 325373 325374 325375 325376 325377 325378 325379 325380 325381 325382 325383 325384 325385 325386 325387 325388 325389 325390 325391 325392 325394 325396 325397 325398 325399 325401 325402 325403 325404 325405 325406 325407 325408 325409 325410 325411 325412 325413 325414 325415 325418 325419 325420 325421 325422 325424 325425 325427 325428 325429 325430 325431 325432 325433 325434 325435 325436 325437 325438 325440 325441 325442 325443 325445 325446 325447 325448 325449 325450 325451 325452 325453 325454 325455 325456 325457 325458 325459 325460 325461 325462 325463 325464 325465 325466 325467 325468 325469 325470 331543 331544 331545 331546 331548 331549 331550 331551 331552 331553 331554 331555 331556 331557 331558 331559 331560 331561 331562 331563 331564 331565 331566 331567 331568 331569 331570 331571 331572 331573 331574 331575 331576 331577 331578 331579 331581 331582 331583 331584 331585 331586 331587 331588 331589 331590 331591 331592 331593 331594 331595 331596 331597 331598 331599 331600 331601 331602 331603 331604 331605 331606 331607 331608 331609 331610 331611 331612 331613 331614 331615 331616 331617 331618 331619 331621 331622 331623 331624 331625 331626 331627 331628 331629 331630 331631 331632 331633 331635 331636 331637 331638 331639 331640 331641 331642 331643 331644 331645 331646 331648 331649 331650 331651 331654 331656 331657 331658 331659 331660 331662 331663 331664 331665 331666 331667 331668 331670 331671 331672 331674 331675 331676 331677 331678 331679 331680 331681 331682 331683 331684 331685 331686 331687 331688 331689 331691 331692 331693 331694 331695 331696 331697 331698 331699 331700 331701 331702 331703 331704 331705 331706 331707 331708 331709 331710 331711 331712 331713 331714 331715 331716 331717 331718 331719 331720 331721 331722 331723 331724 331725 331726 331727 331728 331729 331730 331731 331732 331733 331734 331736 331738 331741 331742 331743 331744 331745 331748 331749 331750 331752 331753 331754 331755 331756 331757 331758 331759 331760 331761 331762 331763 331764 331765 331766 331767 331768 331769 331771 331772 331773 331774 331775 331776 331777 331778 331779 331780 331781 331782 331783 331784 331785 331786 331787 331788 331789 331790 331791 331792 331793 331795 331796 331797 331798 331799 331801 331802 331803 331804 331805 331806 331807 331808 331809 331810 331811 331812 331813 331814 331815 331816 331817 331818 331819 331820 331821 331822 331824 331825 331826 331827 331828 331829 331830 331831 331832 331833 331834 331835 331836 331837 331838 331839 331840 331841 331842 331843 331844 331845 331846 331847 331848 331851 331852 331854 331855 331856 331857 331858 331859 331860 331861 331862 331863 331864 331865 331866 331867 331868 331869 331870 331871 331872 331873 331874 331875 331876 331877 331878 331879 331880 331881 331882 331883 331884 331885 331886 331887 331888 331890 331891 331892 331893 331895 331896 331897 331898 331899 331900 331901 331902 331903 331904 331905 331906 331907 331908 331909 331910 331911 331912 331913 331914 331915 331916 331917 331918 331919 331920 331922 331923 331924 331925 331926 331927 331928 331929 331930 331931 331933 331934 331935 331936 331937 331938 331940 331942 331943 331944 331945 331946 331947 331948 331949 331950 331951 331952 331953 331954 331955 331956 331957 331958 331959 331960 331961 331962 331963 331964 331965 331966 331967 331968 331969 331970 331971 331972 331973 331974 331976 331977 331978 331979 331980 331981 331982 331983 331984 331985 331986 331987 331989 331990 331991 331992 331993 331994 331995 331996 331997 331998 331999 332000 332001 332002 332003 332004 332005 332006 332007 332008 332009 332010 332011 332012 332013 332014 332015 332016 332017 332018 332019 332020 332021 332022 332023 332024 332025 332026 332027 332028 332029 332030 332031 332032 332033 332034 332035 332036 332037 332038 332039 332040 332041 332043 332044 332045 332046 332047 332048 332049 332050 332051 332052 332053 332054 332055 332056 332058 332059 332060 332061 332064 332065 332066 332067 332069 332070 332071 332073 332074 332075 332076 332078 332079 332080 332081 332082 332083 332084 332085 332086 332087 332088 332091 332092 332093 332094 332095 332096 332097 332098 332099 332100 332101 332103 332104 332105 332106 332107 332108 332110 332113 332114 332115 332116 332117 332120 332121 332122 332123 332124 332126 332127 332128 332129 332130 332131 332132 332133 332134 332135 332136 332137 332138 332139 332140 332141 332142 332143 332145 332146 332147 332148 332149 332150 332151 332152 332153 332154 332155 332156 332157 332158 332159 332160 332161 332162 332163 332164 332165 332166 332167 332168 332169 332170 332171 332172 332174 332176 332177 332178 332179 332180 332182 332183 332186 332187 332188 332189 332190 332192 332193 332194 332195 332196 332197 332198 332199 332200 332201 332202 332203 332204 332206 332207 332208 332209 332210 332211 332212 332213 332214 332215 332216 332217 332218 332219 332221 332222 332223 332224 332225 332227 332228 332229 332230 332231 332232 332233 332234 332235 332237 332238 332239 332240 332241 332242 332243 332244 332245 332246 332247 332248 332249 332250 332251 332252 332253 332254 332255 332256 332257 332259 332260 332261 332262 332263 332264 332265 332266 332267 332268 332269 332270 332271 332272 332273 332274 332275 332276 332277 332278 332279 332280 332281 332282 332283 332284 332285 332286 332287 332288 332289 332290 332291 332292 332293 332294 332295 332297 332298 332299 332300 332301 332302 332303 332304 332305 332306 332307 332308 332309 332310 332311 332312 332314 332315 332316 332317 332318 332319 332320 332321 332322 332323 332324 332325 332326 332327 332328 332329 332330 332331 332332 332333 332335 332336 332337 332338 332339 332340 332343 332345 332346 332347 332348 332349 332350 332351 332352 332353 332354 332355 332356 332357 332358 332359 332360 332361 332362 332363 332365 332366 332367 332368 332369 332370 332371 332372 332373 332375 332376 332377 332378 332379 332380 332381 332382 332383 332384 332385 332387 332388 332389 332390 332391 332392 332393 332394 332395 332396 332397 332400 332401 332402 332403 332404 332405 332406 332407 332410 332411 332412 332413 332414 332415 332416 332417 332418 332419 332420 332421 332422 332423 332424 332425 332426 332427 332428 332429 332430 332431 332432 332434 332435 332436 332440 332441 332442 332443 332444 332446 332447 332449 332450 332451 332452 332453 332454 332455 332456 332457 332458 332459 332460 332462 332463 332464 332465 332466 332467 332468 332469 332470 332471 332472 332473 332474 332475 332476 332477 332478 332479 332480 332481 332482 332483 332484 332485 332486 332487 332489 332490 332491 332492 332493 332494 332495 332497 332498 332499 332500 332501 332502 332503 332504 332505 332506 332507 332508 332509 332510 332511 332512 332513 332515 332516 332517 332518 332519 332520 332522 332523 332524 332525 332526 332527 332528 332529 332530 332531 332532 332533 332534 332535 332537 332538 332539 332540 332541 332542 332543 332544 332545 332546 332547 332548 332549 332550 332551 332552 332553 332555 332556 332557 332558 332559 332560 332561 332562 332563 332564 332565 332566 332567 332568 332569 332570 332571 332572 332573 332574 332577 332578 332579 332580 332581 332582 332583 332584 332585 332586 332587 332588 332589 332590 332591 332592 332593 332594 332595 332596 332597 332598 332599 332600 332601 332602 332603 332604 332605 332606 332607 332608 332609 332610 332611 332612 332613 332614 332615 332616 332617 332618 332620 332621 332622 332623 332624 332625 332626 332627 332628 332629 332630 332631 332632 332633 332635 332636 332637 332638 332639 332640 332641 332642 332643 332644 332646 332647 332648 332649 332650 332651 332652 332653 332656 332657 332658 332659 332661 332662 332663 332664 332665 332666 332667 332668 332670 332671 332672 332674 332675 332676 332680 332711 332712 332713 332714 332715 332716 332717 332719 332720 332722 332723 332724 332725 332727 332728 332730 332731 332732 332733 332734 332735 332736 332737 332738 332739 332740 332741 332742 332743 332744 332745 332746 332748 332749 332750 332751 332752 332753 332754 332755 332756 332757 332758 332759 332760 332761 332762 332763 332764 332765 332766 332767 332768 332769 332770 332771 332772 332773 332774 332775 332776 332777 332778 332779 332780 332781 332782 332783 332784 332785 332786 332787 332788 332789 332790 332791 332792 332793 332794 332795 332796 332797 332798 332799 332800 332803 332804 332805 332806 332807 332808 332809 332834 332835 332836 332837 332838 332839 332840 332841 332842 332843 332844 332845 332846 332847 332848 332849 332850 332852 332853 332854 332855 332856 332857 332858 332859 332861 332862 332863 332864 332865 332866 332867 332868 332869 332871 332872 332873 332874 332875 332876 332877 332878 332879 332880 332881 332882 332883 332884 332885 332886 332887 332888 332889 332890 332891 332894 332895 332896 332897 332898 332899 332901 332902 332903 332904 332905 332906 332907 332908 332909 332910 332911 332912 332913 332914 332915 332916 332917 332918 332919 332921 332923 332924 332925 332926 332927 332928 332929 332930 332931 332932 332933 332934 332935 332937 332938 332939 332940 332941 332942 332943 332944 332945 332946 332947 332948 332949 332950 332951 332952 332953 332954 332955 332956 332957 332958 332959 332960 332961 332963 332964 332965 332966 332968 332969 332970 332971 332972 332973 332974 332975 332976 332977 332978 332979 332980 332981 332982 332983 332985 332986 332987 332988 332989 332990 332991 332992 332993 332994 332995 332996 332998 332999 333000 333002 333003 333004 333005 333006 333007 333008 333009 333010 333011 333012 333013 333015 333016 333017 333018 333019 333020 333021 333022 333023 333024 333025 333027 333028 333029 333030 333031 333035 333036 333037 333038 333039 333040 333041 333042 333043 333044 333045 333047 333048 333049 333050 333051 333052 333054 333055 333056 333057 333060 333061 333062 333063 333064 333065 333066 333068 333069 333070 333071 333072 333073 333074 333075 333077 333078 333079 333080 333081 333082 333083 333084 333085 333086 333088 333089 333090 333091 333092 333093 333094 333095 333096 333097 333098 333099 333100 333101 333102 333103 333105 333106 333107 333108 333109 333110 333111 333112 333113 333114 333115 333116 333117 333118 333120 333121 333122 333123 333124 333125 333126 333127 333128 333129 333130 333132 333133 333134 333135 333136 333137 333138 333139 333141 333142 333143 333144 333145 333146 333147 333148 333149 333150 333151 333152 333153 333154 333155 333156 333157 333158 333160 333161 333163 333164 333165 333166 333167 333168 333169 333170 333171 333172 333173 333174 333175 333176 333177 333178 333179 333180 333181 333183 333184 333185 333186 333187 333188 333189 333190 333191 333192 333193 333194 333196 333197 333198 333199 333201 333202 333203 333204 333205 333206 333208 333209 333211 333212 333214 333215 333216 333217 333218 333219 333220 333221 333222 333223 333224 333225 333226 333228 333229 333230 333231 333232 333233 333234 333236 333237 333238 333239 333240 333241 333242 333243 333244 333246 333247 333248 333249 333250 333251 333252 333253 333254 333255 333256 333257 333258 333259 333260 333261 333262 333264 333265 333266 333267 333268 333269 333270 333271 333272 333273 333274 333275 333276 333277 333278 333279 333280 333281 333282 333283 333284 333286 333287 333288 333289 333290 333291 333292 333293 333294 333295 333296 333297 333298 333299 333300 333301 333302 333303 333304 333305 333306 333307 333308 333309 333310 333311 333312 333314 333315 333316 333317 333318 333319 333320 333321 333323 333324 333325 333326 333327 333328 333329 333330 333331 333332 333333 333334 333336 333337 333338 333339 333340 333341 333342 333343 333344 333346 333347 333348 333349 333350 333351 333352 333353 333354 333355 333356 333357 333358 333359 333360 333362 333363 333364 333365 333366 333367 333368 333370 333372 333373 333375 333377 333380 333381 333382 333383 333384 333385 333386 333387 333388 333389 333390 333391 333392 333393 333394 333397 333398 333399 333400 333401 333403 333404 333405 333406 333407 333408 333409 333411 333412 333413 333414 333416 333417 333418 333419 333420 333421 333422 333423 333424 333425 333426 333427 333428 333429 333430 333431 333432 333433 333434 333435 333436 333438 333439 333440 333441 333442 333443 333444 333446 333447 333448 333449 333450 333451 333452 333453 333454 333455 333456 333457 333458 333459 333460 333461 333462 333463 333464 333465 333466 333468 333469 333470 333471 333472 333473 333474 333475 333476 333478 333479 333480 333481 333482 333483 333484 333485 333486 333487 333488 333489 333490 333491 333492 333493 333494 333495 333497 333498 333499 333500 333501 333502 333503 333505 333506 333507 333508 333509 333510 333511 333512 333513 333514 333515 333516 333517 333518 333519 333521 333522 333523 333524 333525 333526 333528 333530 333531 333532 333533 333534 333535 333536 333537 333538 333539 333540 333541 333542 333544 333545 333546 333547 333548 333549 333550 333551 333552 333554 333555 333556 333557 333558 333559 333560 333561 333562 333563 333564 333565 333566 333567 333568 333569 333570 333571 333572 333573 333574 333575 333576 333577 333578 333579 333580 333581 333582 333583 333584 333585 333586 333587 333588 333589 333590 333591 333592 333593 333594 333595 333596 333597 333598 333599 333600 333601 333603 333604 333605 333606 333608 333609 333610 333611 333613 333614 333615 333616 333617 333618 333619 333620 333621 333623 333625 333626 333627 333628 333629 333630 333631 333632 333633 333634 333635 333637 333638 333639 333640 333641 333643 333644 333645 333646 333647 333648 333649 333650 333651 333652 333653 333654 333655 333656 333657 333658 333659 333660 333661 333662 333663 333664 333665 333666 333667 333668 333669 333670 333671 333672 333673 333674 333675 333676 333677 333678 333679 333680 333681 333682 333684 333685 333686 333687 333688 333689 333690 333692 333693 333694 333695 333696 333697 333698 333699 333700 333701 333702 333705 333706 333707 333709 333710 333711 333713 333714 333715 333716 333717 333718 333719 333724 333725 333726 333727 333728 333729 333730 333731 333732 333734 333735 333736 333737 333738 333739 333740 333741 333742 333743 333744 333745 333746 333747 333748 333750 333751 333752 333753 333754 333755 333756 333757 333759 333760 333761 333762 333763 333764 333765 333766 333767 333768 333769 333771 333772 333773 333777 333779 333780 333781 333782 333783 333784 333785 333786 333787 333788 333790 333791 333792 333793 333795 333796 333797 333798 333799 333803 333804 333805 333806 333807 333808 333809 333810 333811 333812 333813 333814 333824 333825 333826 333828 333829 333830 333831 333832 333833 333834 333835 333836 333837 333839 333840 333841 333842 333843 333844 333845 333846 333847 333848 333850 333851 333852 333853 333854 333855 333857 333858 333859 333860 333861 333862 333863 333864 333865 333866 333867 333868 333869 333870 333871 333872 333873 333874 333875 333876 333877 333878 333879 333880 333881 333882 333883 333884 333885 333886 333887 333888 333889 333890 333891 333894 333895 333896 333897 333898 333899 333900 333902 333903 333904 333905 333906 333907 333908 333909 333910 333911 333912 333913 333914 333916 333917 333918 333919 333920 333921 333922 333923 333925 333926 333927 333928 333929 333930 333931 333933 333934 333935 333936 333937 333938 333939 333940 333941 333942 333943 333944 333945 333946 333947 333948 333949 333950 333951 333952 333953 333954 333955 333956 333957 333959 333960 333961 333962 333963 333964 333965 333966 333967 333968 333969 333970 333971 333972 333973 333975 333978 333979 333980 333981 333982 333983 333984 333985 333986 333987 333988 333989 333991 333992 333993 333994 333995 333996 333997 333998 333999 334000 334002 334003 334004 334005 334006 334007 334008 334009 334011 334012 334013 334014 334015 334016 334017 334018 334019 334020 334021 334022 334023 334024 334025 334026 334027 334028 334029 334030 334031 334033 334034 334035 334036 334037 334039 334040 334042 334043 334044 334045 334046 334047 334048 334049 334050 334052 334053 334054 334055 334056 334057 334058 334059 334060 334062 334063 334064 334065 334066 334067 334068 334069 334070 334071 334072 334073 334074 334075 334076 334077 334079 334080 334081 334082 334083 334084 334085 334086 334087 334088 334089 334090 334091 334093 334094 334095 334096 334097 334098 334099 334100 334101 334102 334103 334104 334105 334106 334107 334108 334109 334110 334111 334112 334113 334114 334115 334116 334117 334118 334119 334120 334121 334123 334124 334125 334126 334127 334128 334129 334130 334132 334133 334134 334135 334136 334137 334138 334139 334140 334141 334142 334143 334145 334146 334147 334148 334149 334150 334151 334153 334154 334155 334156 334157 334158 334159 334160 334161 334162 334164 334165 334166 334167 334168 334169 334170 334171 334172 334173 334174 334175 334176 334177 334178 334179 334180 334181 334182 334183 334184 334185 334186 334187 334188 334189 334190 334191 334192 334193 334194 334196 334197 334198 334199 334200 334201 334202 334203 334205 334206 334207 334208 334209 334210 334211 334212 334213 334214 334215 334216 334217 334218 334221 334222 334223 334224 334225 334226 334227 334228 334230 334231 334232 334234 334235 334236 334237 334238 334240 334241 334242 334243 334244 334245 334247 334249 334251 334252 334253 334254 334255 334256 334257 334258 334259 334260 334261 334262 334263 334264 334265 334266 334267 334268 334269 334270 334272 334273 334274 334275 334276 334277 334278 334279 334280 334281 334282 334284 334285 334286 334287 334288 334289 334290 334291 334292 334294 334295 334296 334297 334298 334299 334300 334301 334302 334303 334304 334305 334306 334307 334308 334309 334310 334312 334313 334315 334316 334317 334318 334319 334320 334321 334322 334323 334324 334325 334326 334328 334329 334330 334331 334332 334333 334334 334335 334336 334337 334338 334340 334342 334343 334344 334345 334346 334347 334348 334349 334350 334351 334352 334354 334355 334356 334357 334358 334359 334360 334361 334362 334363 334364 334365 334366 334367 334368 334369 334370 334371 334372 334373 334374 334375 334376 334378 334379 334380 334381 334382 334383 334384 334385 334386 334387 334388 334389 334390 334391 334392 334393 334394 334395 334396 334397 334398 334399 334400 334401 334403 334404 334405 334406 334407 334408 334409 334410 334412 334413 334414 334415 334416 334417 334418 334419 334420 334421 334422 334423 334424 334425 334429 334430 334431 334432 334435 334436 334437 334438 334439 334440 334442 334443 334444 334445 334446 334447 334448 334449 334450 334451 334452 334453 334454 334455 334456 334458 334459 334460 334461 334462 334463 334464 334465 334466 334467 334468 334469 334470 334471 334472 334473 334474 334475 334476 334477 334478 334479 334480 334481 334482 334484 334485 334487 334488 334489 334490 334491 334492 334493 334494 334495 334496 334497 334498 334499 334500 334501 334502 334504 334505 334506 334507 334508 334510 334511 334512 334513 334514 334515 334516 334517 334518 334519 334520 334521 334522 334523 334525 334527 334528 334529 334530 334531 334533 334534 334535 334536 334537 334538 334539 334540 334541 334542 334543 334544 334545 334546 334547 334548 334549 334550 334551 334552 334553 334554 334555 334556 334558 334559 334560 334561 334562 334563 334564 334565 334566 334567 334568 334569 334570 334571 334572 334574 334575 334576 334577 334578 334579 334580 334581 334582 334583 334584 334585 334586 334587 334588 334589 334590 334591 334592 334593 334594 334595 334596 334597 334599 334600 334601 334602 334603 334604 334605 334606 334607 334608 334609 334610 334611 334612 334613 334614 334615 334616 334617 334618 334620 334621 334622 334623 334624 334625 334626 334628 334629 334630 334631 334633 334634 334635 334636 334637 334638 334639 334640 334641 334642 334643 334644 334646 334647 334648 334650 334651 334652 334653 334654 334655 334657 334659 334660 334661 334662 334663 334664 334665 334666 334667 334669 334670 334671 334673 334674 334676 334677 334679 334680 334681 334682 334683 334684 334685 334686 334687 334688 334689 334690 334691 334692 334693 334695 334696 334697 334698 334699 334700 334701 334702 334703 334704 334705 334706 334707 334709 334710 334711 334712 334713 334714 334715 334716 334717 334718 334719 334720 334722 334723 334724 334725 334726 334729 334730 334731 334732 334733 334734 334735 334736 334738 334739 334740 334741 334742 334743 334744 334745 334746 334748 334749 334750 334751 334752 334753 334754 334756 334757 334758 334759 334760 334761 334762 334763 334764 334766 334767 334768 334769 334770 334771 334772 334773 334774 334775 334776 334777 334779 334780 334781 334782 334783 334784 334785 334786 334787 334788 334789 334790 334791 334792 334793 334795 334796 334797 334798 334799 334800 334801 334802 334803 334806 334807 334808 334809 334810 334811 334812 334813 334814 334815 334816 334817 334818 334819 334820 334821 334822 334823 334825 334826 334828 334829 334830 334831 334832 334833 334834 334835 334836 334838 334839 334840 334841 334842 334844 334845 334846 334847 334848 334849 334850 334851 334853 334854 334856 334857 334858 334859 334861 334862 334863 334864 334865 334867 334868 334869 334870 334871 334872 334873 334874 334875 334876 334878 334879 334880 334881 334882 334884 334886 334887 334888 334889 334890 334891 334893 334894 334895 334896 334897 334899 334900 334901 334902 334903 334904 334905 334906 334907 334909 334911 334913 334914 334915 334916 334917 334918 334919 334922 334923 334924 334925 334927 334928 334930 334931 334932 334933 334935 334936 334937 334938 334939 334940 334941 334942 334943 334944 334945 334946 334947 334948 334949 334950 334951 334952 334953 334955 334956 334957 334958 334960 334961 334962 334963 334964 334965 334966 334967 334968 334969 334970 334971 334972 334973 334974 334975 334976 334977 334978 334979 334980 334981 334982 334983 334984 334985 334986 334987 334988 334989 334990 334991 334992 334993 334994 334995 334996 334997 334998 334999 335000 335001 335002 335003 335004 335005 335006 335008 335009 335010 335011 335013 335014 335015 335016 335017 335018 335019 335020 335021 335022 335023 335024 335026 335027 335028 335029 335030 335031 335032 335033 335034 335035 335036 335037 335038 335039 335040 335041 335042 335043 335044 335045 335046 335047 335048 335049 335050 335051 335052 335053 335054 335055 335057 335058 335059 335060 335061 335062 335063 335064 335065 335066 335067 335068 335070 335071 335072 335073 335075 335076 335077 335078 335079 335080 335081 335082 335084 335085 335086 335087 335088 335089 335090 335091 335092 335093 335094 335095 335096 335097 335098 335099 335100 335102 335103 335105 335106 335107 335108 335109 335110 335111 335112 335113 335114 335115 335116 335117 335118 335119 335120 335121 335122 335123 335124 335125 335126 335127 335128 335129 335130 335131 335132 335133 335134 335135 335136 335137 335138 335139 335140 335141 335142 335143 335144 335145 335146 335147 335148 335149 335150 335151 335152 335153 335154 335155 335156 335157 335158 335159 335160 335161 335162 335163 335164 335165 335166 335167 335169 335171 335173 335174 335175 335176 335177 335178 335179 335180 335182 335183 335184 335185 335186 335187 335188 335189 335190 335191 335192 335193 335194 335195 335196 335197 335198 335199 335200 335201 335202 335203 335204 335205 335206 335207 335208 335209 335210 335211 335213 335214 335215 335216 335217 335218 335219 335220 335221 335222 335223 335224 335225 335226 335227 335228 335229 335230 335231 335232 335233 335234 335235 335236 335237 335238 335239 335240 335241 335242 335243 335244 335245 335246 335248 335249 335250 335252 335253 335254 335255 335256 335257 335259 335260 335261 335262 335263 335264 335265 335266 335267 335268 335269 335270 335271 335272 335273 335274 335275 335276 335277 335278 335279 335280 335281 335282 335283 335284 335285 335286 335287 335288 335289 335290 335291 335292 335293 335294 335295 335296 335297 335298 338657 338659 338660 338661 338662 338663 338664 338665 338666 338667 338668 338669 338670 338671 338672 338673 338674 338675 338676 338678 338679 338680 338681 338682 338683 338684 338685 338686 338687 338688 338690 338691 338692 338693 338694 338695 338696 338698 338699 338700 338701 338702 338703 338705 338706 338707 338708 338709 338710 338711 338712 338713 338714 338715 338717 338718 338720 338721 338722 338723 338724 338725 338726 338727 338728 338729 338730 338731 338732 338733 338734 338735 338737 338738 338739 338740 338741 338742 338743 338744 338745 338746 338747 338748 338749 338750 338751 338752 338753 338755 338756 338757 338758 338759 338760 338762 338763 338764 338766 338767 338768 338769 338770 338771 338772 338773 338774 338775 338776 338777 338778 338780 338781 338783 338784 338785 338786 338787 338788 338789 338790 338791 338792 338793 338794 338795 338796 338798 338799 338800 338801 338802 338803 338804 338805 338806 338807 338808 338809 338810 338812 338813 338814 338816 338817 338818 338819 338820 338821 338822 338823 338825 338826 338827 338828 338829 338830 338831 338832 338834 338835 338836 338837 338838 338839 338840 338841 338844 338845 338846 338847 338848 338849 338850 338851 338852 338853 338854 338855 338856 338857 338858 338859 338862 338863 338864 338865 338866 338867 338868 338869 338870 338871 338872 338873 338874 338875 338876 338877 338878 338879 338880 338881 338882 338883 338884 338885 338886 338887 338888 338889 338890 338891 338892 338893 338894 338895 338896 338897 338898 338899 338900 338901 338902 338903 338904 338905 338906 338907 338908 338910 338911 338912 338913 338914 338915 338916 338917 338918 338919 338920 338921 338922 338924 338925 338926 338927 338928 338930 338931 338934 338935 338938 338939 338940 338941 338942 338943 338944 338945 338946 338947 338948 338949 338950 338951 338952 338954 338955 338956 338957 338958 338959 338961 338962 338963 338964 338965 338966 338967 338968 338969 338970 338971 338972 338973 338974 338976 338977 338978 338980 338981 338983 338984 338985 338986 338987 338988 338989 338990 338991 338992 338993 338994 338995 338996 338997 338998 339000 339004 339005 339006 339007 339008 339010 339011 339012 339013 339014 339015 339016 339017 339018 339019 339020 339021 339023 339024 339025 339026 339027 339028 339029 339031 339032 339033 339034 339036 339037 339038 339040 339041 339042 339043 339044 339045 339046 339047 339048 339049 339050 339051 339052 339054 339055 339056 339057 339058 339059 339060 339061 339062 339063 339064 339065 339066 339067 339068 339069 339070 339071 339072 339073 339074 339075 339076 339077 339078 339079 339080 339081 339082 339083 339084 339085 339087 339088 339089 339090 339091 339092 339093 339094 339095 339096 339097 339099 339100 339102 339103 339104 339105 339106 339107 339108 339109 339110 339111 339112 339113 339114 339115 339116 339117 339118 339119 339120 339121 339122 339123 339124 339125 339127 339128 339129 339130 339131 339132 339133 339134 339135 339136 339137 339138 339139 339140 339141 339142 339143 339144 339145 339146 339147 339148 339149 339150 339151 339152 339153 339154 339155 339156 339157 339158 339159 339164 339165 339166 339167 339168 339169 339170 339171 339173 339174 339176 339177 339178 339180 339182 339183 339185 339186 339188 339190 339195 339197 339198 339199 339201 339202 339203 339204 339205 339206 339207 339208 339209 339211 339212 339214 339215 339216 339217 339218 339219 339220 339221 339223 339224 339225 339226 339227 339228 339229 339230 339231 339232 339233 339234 339236 339238 339239 339240 339241 339242 339243 339244 339245 339247 339248 339250 339251 339252 339253 339254 339255 339256 339257 339258 339259 339260 339261 339262 339263 339264 339265 339266 339267 339269 339270 339271 339272 339273 339274 339275 339276 339277 339278 339279 339280 339281 339282 339283 339284 339285 339286 339287 339288 339289 339291 339293 339294 339295 339297 339298 339299 339300 339301 339302 339303 339304 339306 339307 339308 339310 339311 339312 339313 339314 339315 339316 339317 339318 339319 339320 339321 339322 339323 339324 339325 339326 339327 339328 339329 339330 339331 339332 339333 339334 339335 339336 339338 339339 339340 339342 339343 339344 339345 339346 339347 339348 339349 339350 339351 339352 339353 339355 339356 339359 339360 339361 339362 339363 339364 339365 339366 339367 339368 339433 339434 339435 339436 339437 339439 339440 339441 339442 339443 339444 339445 339446 339447 339448 339449 339450 339451 339452 339453 339454 339455 339456 339457 339458 339459 339460 339461 339462 339463 339464 339465 339466 339467 339468 339470 339471 339472 339473 339474 339475 339476 339477 339478 339479 339480 339481 339483 339484 339485 339486 339487 339489 339490 339491 339492 339493 339494 339495 339496 339497 339498 339499 339500 339501 339502 339503 339504 339505 339506 339507 339508 339510 339511 339512 339513 339515 339516 339517 339518 339519 339520 339521 339522 339523 339524 339525 339526 339527 339528 339529 339530 339531 339532 339533 339534 339536 339537 339538 339539 339540 339543 339544 339545 339546 339547 339548 339549 339550 339551 339553 339554 339555 339556 339557 339558 339560 339561 339562 339563 339565 339566 339567 339568 339569 339570 339571 339572 339573 339574 339575 339576 339577 339578 339579 339580 339581 339582 339584 339585 339586 339587 339588 339589 339590 339591 339592 339593 339594 339595 339596 339597 339598 339599 339600 339601 339602 339603 339604 339606 339607 339608 339609 339610 339611 339612 339613 339614 339615 339617 339618 339619 339620 339621 339622 339623 339624 339625 339626 339627 339628 339629 339630 339631 339632 339633 339634 339635 339636 339637 339638 339639 339640 339641 339642 339643 339644 339645 339646 339647 339648 339649 339650 339651 339652 339653 339654 339655 339657 339658 339659 339660 339661 339662 339663 339664 339665 339667 339668 339670 339671 339672 339673 339674 339675 339676 339677 339678 339679 339680 339681 339682 339683 339684 339685 339686 339687 339688 339689 339690 339691 339692 339693 339694 339695 339696 339697 339698 339699 339700 339701 339702 339703 339704 339705 339706 339707 339708 339709 339710 339711 339712 339713 339714 339715 339716 339717 339718 339719 339720 339721 339722 339723 339724 339725 339726 339727 339728 339729 339731 339732 339733 339734 339735 339736 339737 339738 339739 339740 339741 339742 339743 339744 339745 339746 339747 339748 339749 339750 339751 339752 339753 339754 339755 339756 339757 339758 339759 339762 339763 339764 339765 339766 339767 339768 339769 339770 339771 339773 339774 339775 339776 339777 339778 339779 339780 339781 339782 339783 339784 339785 339786 339787 339788 339789 339791 339792 339793 339794 339795 339797 339798 339800 339801 339803 339804 339805 339806 339807 339809 339810 339811 339812 339813 339814 339815 339816 339817 339818 339819 339820 339821 339822 339823 339824 339825 339826 339829 339831 339832 339833 339834 339835 339836 339837 339838 339839 339840 339841 339842 339843 339844 339845 339846 339847 339848 339849 339850 339851 339852 339853 339854 339855 339856 339857 339858 339859 339861 339862 339863 339864 339865 339866 339867 339868 339869 339870 339871 339872 339873 339874 339875 339876 339877 339878 339879 339881 339882 339883 339884 339885 339886 339887 339888 339889 339890 339891 339892 339893 339894 339895 339897 339898 339899 339900 339901 339902 339903 339905 339906 339907 339909 339910 339911 339912 339913 339914 339917 339918 339919 339920 339921 339922 339923 339924 339925 339926 339927 339928 339929 339930 339931 339932 339933 339934 339935 339936 339937 339938 339939 339940 339941 339942 339944 339945 339946 339947 339948 339949 339951 339952 339953 339954 339955 339956 339957 339958 339960 339961 339962 339963 339964 339965 339966 339968 339969 339971 339972 339973 339974 339975 339976 339977 339978 339979 339980 339981 339982 339983 339984 339985 339986 339987 339988 339990 339991 339992 339993 339994 339995 339996 339997 339998 340002 340003 340004 340006 340007 340008 340009 340010 340011 340012 340013 340014 340015 340016 340017 340018 340019 340020 340021 340022 340023 340024 340025 340026 340027 340028 340029 340030 340031 340032 340033 340034 340035 340036 340037 340038 340039 340040 340041 340042 340043 340044 340046 340047 340048 340049 340050 340051 340052 340053 340054 340057 340058 340059 340060 340061 340062 340063 340064 340065 340066 340067 340068 340069 340070 340071 340074 340077 340078 340079 340080 340081 340083 340084 340085 340087 340088 340090 340091 340092 340093 340094 340095 340096 340097 340099 340100 340101 340102 340103 340104 340105 340107 340123 340124 340125 340126 340127 340128 340130 340131 340132 340133 340134 340136 340137 340138 340139 340140 340141 340143 340144 340145 340146 340147 340148 340149 340150 340151 340152 340153 340154 340155 340156 340158 340159 340161 340162 340163 340164 340165 340166 340167 340168 340169 340170 340171 340172 340173 340174 340175 340176 340177 340178 340179 340180 340182 340184 340185 340186 340187 340188 340189 340190 340191 340192 340193 340194 340195 340196 340197 340198 340199 340200 340201 340202 340203 340205 340206 340209 340210 340211 340212 340213 340215 340216 340217 340218 340219 340220 340221 340222 340223 340224 340225 340226 340227 340228 340229 340230 340231 340232 340233 340234 340235 340236 340237 340238 340239 340240 340241 340242 340244 340245 340246 340247 340249 340250 340251 340252 340253 340254 340255 340257 340258 340259 340265 340266 340267 340268 340270 340272 340273 340274 340275 340276 340277 340278 340279 340281 340282 340283 340284 340285 340286 340287 340288 340289 340290 340291 340292 340296 340297 340298 340299 340300 340301 340302 340303 340304 340305 340307 340308 340309 340310 340311 340312 340313 340314 340315 340316 340317 340318 340319 340321 340322 340324 340325 340328 340330 340331 340332 340334 340335 340336 340337 340338 340339 340340 340341 340342 340343 340344 340345 340346 340347 340348 340349 340350 340351 340352 340353 340354 340355 340356 340357 340358 340359 340360 340361 340362 340363 340365 340366 340367 340368 340369 340370 340371 340372 340373 340375 340377 340378 340381 340382 340383 340384 340386 340387 340388 340389 340390 340391 340392 340393 340394 340395 340396 340397 340398 340399 340400 340401 340402 340403 340404 340405 340407 340408 340409 340410 340411 340413 340414 340415 340416 340417 340418 340420 340421 340422 340423 340424 340428 340429 340430 340431 340432 340433 340434 340435 340436 340437 340438 340439 340440 340441 340444 340445 340446 340449 340451 340452 340455 340456 340457 340458 340459 340460 340461 340462 340463 340464 340465 340467 340468 340469 340470 340472 340473 340474 340475 340476 340477 340478 340479 340480 340482 340483 340485 340486 340488 340489 340491 340492 340493 340494 340495 340496 340497 340499 340500 340501 340502 340503 340504 340506 340507 340508 340509 340510 340511 340512 340513 340514 340515 340516 340517 340518 340519 340520 340521 340522 340523 340524 340525 340526 340527 340528 340531 340532 340533 340534 340535 340536 340537 340538 340539 340540 340541 340542 340543 340544 340545 340546 340547 340548 340549 340550 340552 340553 340554 340555 340556 340557 340558 340559 340560 340561 340562 340563 340564 340565 340566 340568 340590 340591 340592 340593 340594 340595 340596 340597 340599 340600 340601 340602 340603 340604 340605 340606 340607 340608 340609 340610 340611 340612 340613 340614 340615 340616 340617 340618 340619 340620 340621 340622 340624 340625 340627 340628 340629 340630 340631 340632 340633 340634 340635 340636 340637 340638 340639 340640 340641 340642 340643 340644 340645 340646 340647 340648 340649 340650 340651 340652 340653 340654 340656 340657 340658 340659 340660 340661 340662 340664 340665 340666 340668 340669 340671 340672 340674 340675 340676 340677 340678 340679 340680 340681 340682 340683 340684 340685 340686 340687 340688 340689 340690 340691 340692 340693 340694 340695 340696 340697 340698 340699 340700 340701 340702 340704 340705
Recalling Firm/
Manufacturer
LivaNova USA, Inc.
100 Cyberonics Blvd
Houston TX 77058-2069
For Additional Information ContactLivaNova Customer Quality
+1-866-8828804
Manufacturer Reason
for Recall
Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.
FDA Determined
Cause 2
Device Design
ActionOn January 3, 2025, LivaNova issued a "Urgent Medical Device Removal" via Mail to affected consignees. LivaNova ask consignees to take the following actions: 1. Monitor the patient for changes in clinical symptoms or if the patient loses perception of stimulation (e.g., the typical cadence for patient visits range from every 3-12 months). 2. Perform system diagnostic testing at each patient office visit to monitor generator battery status. 3. Counsel patients to do the following: i. Notify you if there is a change in perceived clinical symptoms ii. Notify you if they no longer perceive any form of stimulation. iii. Patients with epilepsy who have Magnet Mode enabled should be reminded to use the patient magnet daily to check for the sensation of stimulation to confirm proper generator function, in alignment with recommendations in product labeling for all VNS generator models. Notification Acknowledgement Follow one of the options below to acknowledge receipt of this notification: Option 1 1. Using the QR Code or website address and access code on page 1 of this letter, access your customer page via our secure customer portal ( TRIEVR ) to provide the required information to LivaNova USA, Inc. via a secure, transmission-encrypted system. 2. Answer the question in the Response section indicating your receipt and acknowledgement of the notification. 3. Submit the response by selecting the Send Response button at the bottom of the page.
Quantity in Commerce8223 units
DistributionWorldwide - US Nationwide distribution including in the states of AK, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR,PA, PN, SC, SD, TN, TX, VA, VT, WA, WI, WV, UT and the countries of AE, AR, AT, BE, BG, BR, CA, CL, CO, CR, CZ, DE, DK, GR, HR, IL, IN,IT, JO, KW, LB, MK, NL, OM, PA, PE, PL, QA, RO, RS, SA, SE, SG, SI, TR, TT, UK, ZA, Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Iceland, Ireland, Israel, Italy, Japan, Kuwait, Lithuania, Luxembourg, New Zealand, Norway, Panama, Peru, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, The Netherlands, Turkey, UAE, United Kingdom.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA DatabasePMAs with Product Code = LYJ
-
-