| | Class 1 Device Recall VARIPULSE BiDirectional Ablation Catheter |  |
| Date Initiated by Firm | January 05, 2025 |
| Date Posted | February 28, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1205-2025 |
| Recall Event ID |
96170 |
| PMA Number | P240006 |
| Product Classification |
Percutaneous Cardiac Ablation Catheter for Treatment of Atrial Fibrillation with Irreversible Electroporation - Product Code QZI
|
| Product | VARIPULSE Bi-Directional Ablation Catheter REF D141201.
The Field Catheter is indicated for use in catheter based cardiac electrophysiological mapping (stimulating and recording) and, when used for TRUPULSE Generator, for the treatment of drug refractory recurrent symptomatic paroxysmal atrial fibrillation. The catheter provides location information when used with the CARTO 3 System. |
| Code Information |
Product Number: D141201-12
UDI-DI code: 10846835025460
Batch Numbers:
31483574L
31483560L
31483581L
31483542L
31487313L
31483552L
31483544L
31487311L
31483550L
31483558L
31483575L
31483556L
31487312L
31483553L
31483565L
31483576L
31483547L
31487320L
31483549L
31483561L
31483548L
31483564L
31483541L
31483546L
31483562L
31483566L
31483583L
31483567L
|
Recalling Firm/ Manufacturer |
Biosense Webster, Inc. 31 Technology Dr Ste 200 Irvine CA 92618-2302
|
| For Additional Information Contact | 909-839-8500 |
Manufacturer Reason for Recall | Due to an observed trend of neurovascular events |
FDA Determined Cause 2 | Labeling design |
| Action | On 01/05/2025, the firm called and emailed customers to inform them to pause the US External Evaluations and all Varipulse cases due an observed trend in neurovascular events associated with the use of the VariPulse Bi-Directional Ablation Catheters distributed between December 2024 and January 2025.
For U.S. Customers, Biosense Webster/JNJ Medtech EP is pausing the US External Evaluation and all US Varipulse cases. Healthcare providers who have treated patients using the products prior to the pause should continue to follow those patients according to their standard of care.
For any questions, contact Biosense Webster/JNJ MEDTECH EP representatives
On 01/08/2025, the firm released the information on the firm's parent company's website https://www.jnj.com/media-center/press-releases/johnson-johnson-statement-on-varipulse-u-s-external-evaluation |
| Quantity in Commerce | 497 total catheter - 378 US and 119 OUS |
| Distribution | Worldwide distribution; US states of AR, CA, CO, KS, LA, MA, MO, NY, OH and TX.
EMEA, New Zealand, Saudi Arabia, Oman, United Ecuador, Panama, South Africa, China
Austria
Belgium
Czech Republic
France
Germany
Hungary
Ireland
Israel
Italy
UK
Sweden
Norway
Denmark
Finland
Luxembourg
Netherlands
Poland
Portugal
Slovenia
Spain
Switzerland
Serbia
United Arab Emirates
Croatia
Canada
South Korea
Japan
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| PMA Database | PMAs with Product Code = QZI
|
|
|
|