 Class 2 Device Recallsee related information
Date Initiated by FirmMay 29, 2025
Date PostedJuly 28, 2025
Recall Status1 Open3, Classified
Recall NumberZ-2188-2025
Recall Event ID 96915
510(K)NumberK150958 
Product Classification Device, Vascular, for Promoting Embolization - Product Code KRD
ProductLIFEPEARL Drug Elutable Microspheres: 200 +/- 50 micrometers, REF: 8LP2S200; 400 +/- 50 micrometers, REF: 8LP2S400
Code Information REF/GTIN/Lot: 8LP2S20/(01)00812636020105/0000177644, 0000180340, 0000223990, 0000235256, 0000254143, 0000254142, 0000272315, 0000272316, 0000281719, 0000294361, 0000295741, 0000296446, 0000751662, 0000751661, 0001026153; 8LP2S400/(01)00812636020112/0000199869, 0000225205, 0000708506, 0000863001, 0001026156
Recalling Firm/
Manufacturer
MICROVENTION INC.
35 Enterprise
Aliso Viejo CA 92656
For Additional Information ContactTerrence Callahan
949-899-7194
Manufacturer Reason
for Recall
Drug Elutable Microspheres have a smaller actual average diameter that is not within specification, which may lead to inability to reach the desired treatment location (non-targeted embolization), additional procedure/treatment required, blockage other than target vessel, increased procedure time, incomplete embolization, and/or inability to treat the patient.
FDA Determined
Cause 2
Device Design
ActionOn 5/29/2025, recall notices were mailed and emailed to customers who were asked to do the following: 1) Inform all individuals within your organization of the recall and forward the recall notice to any organizations that may have received the affected products. 2) Complete and return the customer acknowledgement form via email: recalls@microvention.com 3) If your institution has affected inventory call firm's Customer Care at 800-888-3786 or email tmccustomer.admin@terumomedical.com Please direct questions to the firm at MVEMEAQARA@microvention.com
Quantity in Commerce1451
DistributionOUS: Italy, Spain, Portugal, Poland, Belgium, Germany, Denmark, France, Kazakhstan, Slovakia, Norway, Greece, Serbia, Turkey, Czech Republic, United Kingdom, Netherlands, Russian Federation, Kuwait, Republic of North Macedonia, Switzerland, Saudi Arabia, Georgia, Lithuania, Romania, Croatia, Chile, Hong Kong , India, Mexico, Taiwan, Vietnam
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = KRD
-
-