| | Class 1 Device Recall Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set |  |
| Date Initiated by Firm | June 26, 2025 |
| Date Posted | July 30, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2167-2025 |
| Recall Event ID |
97222 |
| Product Classification |
Set, tubing and support, ventilator (w harness) - Product Code BZO
|
| Product | Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators. |
| Code Information |
UDI-DI: 07630002802963, Lots: 199675, 199676
|
| FEI Number |
3014766741
|
Recalling Firm/ Manufacturer |
Hamilton Medical AG Parc Industrial Vial 10 Domat/Ems Switzerland
|
Manufacturer Reason for Recall | Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lead to acute hypercapnia, respiratory acidosis, and if unrecognized organ dysfunction. |
FDA Determined Cause 2 | Process control |
| Action | Starting on 6/26/2025, recall notices were emailed to customers who were asked to do the following:
Return affected devices.
If one of the affected breathing circuits would be already in use with a patient, it is requested to exchange it and discard the product after replacement.
If you are a distributor, please forward to any organization where potentially affected product has been transferred.
Complete and return the acknowledgement form via email to complaints@hamiltonmedical.com
If you have any questions contact firm at: complaints@hamiltonmedical.com and 800-426-6331
|
| Quantity in Commerce | 2,560 |
| Distribution | US Nationwide distribution in the states of NH, NY, CT, PA, IL, ME, CO, NC, NJ, UT, TX, GA, SC, VA, IN, SD, MO, OR, TN, MA, AZ, NM.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|