 Class 2 Device Recall SureStep Foley Tray Systemsee related information
Date Initiated by FirmJuly 17, 2025
Date PostedSeptember 18, 2025
Recall Status1 Open3, Classified
Recall NumberZ-2590-2025
Recall Event ID 97273
510(K)NumberK040658 
Product Classification Catheter, urological (antimicrobial) and accessories - Product Code MJC
ProductSureStep Foley Tray System Lubri-Sil I.C. Complete Care Infection Control Foley Catheter Tray, REF: A303414A
Code Information Lot# NGJY3685
FEI Number 1018233
Recalling Firm/
Manufacturer
C.R. Bard Inc
8195 Industrial Blvd Ne
Covington GA 30014-1497
For Additional Information Contact
770-784-6100
Manufacturer Reason
for Recall
Foley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with infection control coating, when it's silicone hydrogel with antimicrobial coating, or 3)Natural latex with infection control, when it's silicone hydrogel coated without antimicrobial coating, which may cause infection and/or local and/or systemic allergic reaction
FDA Determined
Cause 2
Labeling mix-ups
ActionOn 7/17/2025, recall notices were mailed and emailed to Recall Coordinator, Director of Nursing, Director of Purchasing, Director of Risk Management, and Distributors who were asked to do the following: 1) It is recommended that clinicians discontinue the use of impacted devices and find alternative catheters with the appropriate characteristics required for their patients to continue therapy and avoid potential health risks related to this issue. 2) Quarantine and discard all devices within your facility s control per your facility s procedures. 3) Share this notice with any users of the product within your facilities to ensure they are also aware of this Medical Device Recall (Correction). 4) If you have purchased this product through a distributor, please contact this distributor to request appropriate credit or order alternate replacement product. 5) Complete and return the attached Customer Response Form via email to BDRC38@bd.com In addition, distributors were asked to: a) Identify all customers within your distribution network that purchased any affected product as defined in this notification. Provide a copy of the attached customer letter to all customers to advise them of this field action notification on the recalling firm's behalf. b) Contact recalling firm if you require assistance with credits to be issued for discard-ed products or for replacement product options. If you require assistance contact the firm: Medical Information Services: 1-800-555-7422, medical.information@bd.com; Technical Support: 1-844-823-5433, productcomplaints@bd.com
Quantity in Commerce4,300
DistributionUS Nationwide distribution including in the states of NC, FL, NY, MA, MN, TX, CA, OH, MI, WA, NJ, IL, ID, SC, IN, OK, OR, AL, VA, HI, MO, MS, MD, AZ, KY, GA, IA, CO, NM, PA, ME, UT, NE.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = MJC
-
-