| | Class 2 Device Recall MEDLINE |  |
| Date Initiated by Firm | September 30, 2025 |
| Date Posted | November 14, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0527-2026 |
| Recall Event ID |
97846 |
| Product Classification |
Hysterectomy kit - Product Code OJF
|
| Product | MEDLINE convenience kits labeled as:
1) GYN LAPAROTOMY HARPER PACK-LF, REF DYNJ0160684F;
2) MAJOR LITHOTOMY PACK-LF, REF DYNJ0425778O;
3) MINOR VAGINAL #76-RF, REF DYNJ27434R;
4) ABDOMINAL HYST PACK, REF DYNJ44848M;
5) DA VINCI HYSTERECTOMY, REF DYNJ44863Q;
6) VAGINAL HYSTERECTOMY PACK, REF DYNJ44900P;
7) URO GYN, REF DYNJ50047G;
8) ISS VAG HYST PACK, REF DYNJ50315F;
9) PK MAJOR GYN ST MICHAEL C, REF DYNJ60554B;
10) VAGINAL HYSTERECTOMY PACK, REF DYNJ67194B;
11) LAVH, REF DYNJ900298L;
12) LITHOTOMY-LF, REF DYNJ900721C;
13) GP-T A H PACK-LF, REF PHS396954006G;
14) GP-VAG HYSTERECTOMY PACK-LF, REF PHS396975006F. |
| Code Information |
1) REF DYNJ0160684F: UDI/DI 10195327304287 (each) 40195327304288 (case), Lot Numbers: 23DDC245, 23GDC219;
2) REF DYNJ0425778O: UDI/DI 10198459082290 (each) 40198459082291 (case), Lot Numbers: 24LMF192;
3) REF DYNJ27434R: UDI/DI 10195327107727 (each) 40195327107728 (case), Lot Numbers: 24BBR389, 24CBH709;
4) REF DYNJ44848M: UDI/DI 10195327415952 (each) 40195327415953 (case), Lot Numbers: 23EDB911, 23FDC315, 23HDA180;
5) REF DYNJ44863Q: UDI/DI 10195327298227 (each) 40195327298228 (case), Lot Numbers: 23FDB122, 23GDC100;
6) REF DYNJ44900P: UDI/DI 10195327298548 (each) 40195327298549 (case), Lot Numbers: 23DDB824, 23EDA215, 23FDB287;
7) REF DYNJ50047G: UDI/DI 10195327107734 (each) 40195327107735 (case), Lot Numbers: 24BBE163, 24DBO682;
8) REF DYNJ50315F: UDI/DI 10195327535353 (each) 40195327535354 (case), Lot Numbers: 23KMF190;
9) REF DYNJ60554B: UDI/DI 10195327361747 (each) 40195327361748 (case), Lot Numbers: 23EBD224, 23KBE886;
10) REF DYNJ67194B: UDI/DI 10195327299699 (each) 40195327299690 (case), Lot Numbers: 23HBN435;
11) REF DYNJ900298L: UDI/DI 10195327554170 (each) 40195327554171 (case), Lot Numbers: 23LBF269, 23LBT933, 24BBT248;
12) REF DYNJ900721C: UDI/DI 10193489951608 (each) 40193489951609 (case), Lot Numbers: 23HBW996;
13) REF PHS396954006G: UDI/DI 10195327454159 (each) 40195327454150 (case), Lot Numbers: 23HDB088;
14) REF PHS396975006F: UDI/DI 10195327454166 (each) 40195327454167 (case), Lot Numbers: 23JDB593.
|
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
Medline Industries, LP 3 Lakes Dr Northfield IL 60093-2753
|
| For Additional Information Contact | Haley Barclay 800-369-1704 |
Manufacturer Reason for Recall | Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Medline issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 09/30/2025 and 10/6/2025 via email and mail. The notice explained the issue with the component, potential risk, and requested the following actions:
1. Immediately check stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product.
2. Please use the link and the information found in the notice to complete the response form. Please list the quantity of affected product in inventory on the form. Even if no affected product is in inventory, please complete and submit the form.
3. Upon receipt of the submitted response form, the customer account will receive over-labels to place on affected inventory, with instructions for the customer staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable.
4. For distributors or those who have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product.
If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com. |
| Quantity in Commerce | 570 kits |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|