• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall EDAN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall EDANsee related information
Date Initiated by FirmOctober 24, 2025
Date PostedJanuary 22, 2026
Recall Status1 Open3, Classified
Recall NumberZ-1140-2026
Recall Event ID 97879
510(K)NumberK102140 K173042 
Product Classification System, monitoring, perinatal - Product Code HGM
ProductBrand Name: EDAN Product Name: Fetal Monitor Model/Catalog Number: F2, F3 Product Description: The devices are bedside fetal and maternal monitors, which are used to monitor the physiological parameters of pregnant women including the fetus from 28 weeks gestation. The devices may be used antepartum as well as during labor and delivery. Component: No
Code Information EDAN Model: F3 UDI-DI code: 06944413800847 Serial/Lot Code: 460085-M16105760001 460085-M16105760002 460085-M16105760003 460085-M16105760004 460085-M16105760005 460085-M16105760006 460085-M16105760007 460085-M16105760008 460085-M16105760009 460085-M16105760010 460085-M16105760011 460085-M16105760012 460085-M16105760013 460085-M16105760014 460085-M16105760015 460085-M16105760016 460085-M16105760017 460085-M16105760018 460085-M16105760019 460085-M16105760020 460085-M16105760021 460085-M16105760022 460085-M16105760023 460085-M16105760024 460085-M16105760025 460085-M16105760026 460085-M16105760027 460085-M16105760028 460085-M16105760029 460085-M16105760030 460085-M16105760031 460085-M16105760032 460085-M16105760033 460085-M16105760034 460085-M16105760035 460085-M16105760036 460085-M16105760037 460085-M16105760038 460085-M16105760039 460085-M16105760040 460085-M16105760041 460085-M16105760042 460085-M16105760043 460085-M16105760044 460085-M16105760045 460085-M16105760046 460085-M16105760047 460085-M16105760048 460085-M16105760049 460085-M16105760050 460085-M16105760051 460085-M16105760052 460085-M16105760053 460085-M16803010001 460085-M16803010002 460085-M16803010003 460085-M16803010004 460085-M16803010005 460085-M16803010006 460085-M16803010007 460085-M16803010008 460085-M16803010009 460085-M16803010010 460085-M16803010011 460085-M16803010012 460085-M16803010013 460085-M16803010014 460085-M16803010015 460085-M16803010016 460085-M16803010017 460085-M16803010018 460085-M16803010019 460085-M16803010020 460085-M16803010021 460085-M16803010022 460085-M16803010023 460085-M16803010024 460085-M16803010025 460085-M16803010026 460085-M16803010027 460085-M16803010028 460085-M16803010029 460085-M16803010030 460085-M16803010031 460085-M16803010032 460085-M16803010033 460085-M16803010034 460085-M16803010035 460085-M16803010036 460085-M16803010037 460085-M16803010038 460085-M16803010039 460085-M16803010040 460085-M16803010041 460085-M16803010042 460085-M16803010043 460085-M16803010044 460085-M16803010045 460085-M16803010046 460085-M16803010047 460085-M16803010048 460085-M16803010049 460085-M16803010050 460085-M16803010051 460085-M16803010052 460085-M16803010053 460085-M16803010054 460085-M16803010055 460085-M16803010056 460085-M16803010057 460085-M16803010058 460085-M16803010059 460085-M16803010060 460085-M16803010061 460085-M16803010062 460085-M16803010063 460085-M16803010064 460085-M16A04870001 460085-M16A04870002 460085-M16A04870003 460085-M16A04870004 460085-M16A04870005 460085-M16A04870006 460085-M16A04870007 460085-M16A04870008 460085-M16A04870009 460085-M16A04870010 460085-M16A04870011 460085-M16A04870012 460085-M16A04870013 460085-M16A04870014 460085-M16A04870015 460085-M16A04870016 460085-M16A04870017 460085-M16A04870018 460085-M16A04870019 460085-M16A04870020 460085-M16A04870021 460085-M16A04870022 460085-M16A04870023 460085-M16A04870024 460085-M16A04870025 460085-M16A04870026 460085-M16A04870027 460085-M16A04870028 460085-M16A04870029 460085-M16A04870030 460085-M16A04870031 460085-M16A04870032 460085-M16A04870033 460085-M16A04870034 460085-M16A04870035 460085-M16A04870036 460085-M16A04870037 460085-M16A04870038 460085-M16A04870039 460085-M16A04870040 460085-M16A04870041 460085-M16A04870042 460085-M16A04870043 460085-M16A04870044 460085-M16A04870045 460085-M16A04870046 460085-M16A04880001 460085-M16A04880002 460085-M16A04880003 460085-M16A04880004 460085-M16A04880005 460085-M16A04880006 460085-M16A04880007 460085-M16A04880008 460085-M16A04880009 460085-M16A04880010 460085-M16A04880011 460085-M16A04880012 460085-M16A04880013 460085-M16A04880014 460085-M16A04880015 460085-M16A04880016 460085-M16A04880017 460085-M16A04880018 460085-M16A04880019 460085-M16A04880020 460085-M16A04880021 460085-M16A04880022 460085-M16A04880023 460085-M16A04880024 460085-M16A04880025 460085-M16A04880026 460085-M16A04880027 460085-M16A04880028 460085-M16A04880029 460085-M16A04880030 460085-M16A04880031 460085-M16A04880032 460085-M16A04880033 460085-M16A04880034 460085-M16A04880035 460085-M16A04880036 460085-M16A04880037 460085-M16A04880038 460085-M16A04880039 460085-M16A04880040 460085-M16A04880041 460085-M16A04880042 460085-M16A04880043 460085-M16A04880044 460085-M16A04880045 460085-M16A04880046 460085-M16A04880047 460085-M16A04880048 460085-M16A04880049 460085-M16A04880050 460085-M17107980001 460085-M17107980002 460085-M17107980003 460085-M17107980004 460085-M17107980005 460085-M17107980006 460085-M17107980007 460085-M17107980008 460085-M17107980009 460085-M17107980010 460085-M17107980011 460085-M17107980012 460085-M17107980013 460085-M17107980014 460085-M17107980015 460085-M17602680001 460085-M17602680002 460085-M17602680003 460085-M17602680004 460085-M17602680005 460085-M17602680006 460085-M17602680007 460085-M17602680008 460085-M17602680009 460085-M17602680010 460085-M17602680011 460085-M17602680012 460085-M17602680013 460085-M17602680014 460085-M17602680015 460085-M17602680016 460085-M17602680017 460085-M17602680018 460085-M17602680019 460085-M17602680020 460085-M17602680021 460085-M17602680022 460085-M17602680023 460085-M17602680024 460085-M17602680025 460085-M17602680026 460085-M17602680027 460085-M17602680028 460085-M17602680029 460085-M17602680030 460085-M17602680031 460085-M17602680032 460085-M17602680033 460085-M17602680034 460085-M17602680035 460085-M17602680036 460085-M17602680037 460085-M17602680038 460085-M17602680039 460085-M17602680040 460085-M17602680041 460085-M17602680042 460085-M17602680043 460085-M17602680044 460085-M17602680045 460085-M17602680046 460085-M17602680047 460085-M17602680048 460085-M17602680049 460085-M17602680050 460085-M17602680051 460085-M17602680052 460085-M17602680053 460085-M17602680054 460085-M17602680055 460085-M17602680056 460085-M17602680057 460085-M17602680058 460085-M17602680059 460085-M17602680060 460085-M17602680061 460085-M17602680062 460085-M17602680063 460085-M17602680064 460085-M17602680065 460085-M17602680066 460085-M17602680067 460085-M17602680068 460085-M17602680069 460085-M17602680070 460085-M17602680071 460085-M17602680072 460085-M17602680073 460085-M17602680074 460085-M17602680075 460085-M17602680076 460085-M17602680077 460085-M17602680078 460085-M17602680079 460085-M17602680080 460085-M17602680081 460085-M17602680082 460085-M17602680083 460085-M17602680084 460105-M15B00960028 460105-M15C00330007 460105-M16A04700004 460105-M18111370008 460105-M18202190005 460105-M18406560001 460105-M18406560002 460314-M17409920009 460325-M16B07190001 460325-M17803390001 460403-M15B02730001 460403-M15B02730002 460403-M15B02730003 460403-M15B02730004 460403-M15B02730005 460403-M15B02730006 460403-M15B02730007 460403-M15B02730008 460403-M15B02730009 460403-M15B02730010 460403-M15B02730011 460403-M15B02730012 460403-M15B02730013 460403-M15B02730014 460403-M15B02730015 460403-M15B02730016 460403-M15B02730017 460403-M15B02730018 460403-M15B02730019 460403-M15B02730020 460403-M15B02730021 460403-M15B02730022 460403-M15B02730023 460403-M15B02730024 460403-M15B02730025 460403-M15B02730026 460403-M15B02730027 460403-M15B02730028 460403-M15B02730029 460403-M15B02730030 460403-M15B02730031 460403-M15B02730032 460403-M15B02730033 460403-M15B02730034 460403-M15B02730035 460403-M15B02730036 460403-M15B02730037 460403-M15B02730038 460403-M15B02730039 460403-M15B02730040 460403-M15B02730041 460403-M15B02730042 460403-M15B02730043 460403-M15B02730044 460403-M15B02730045 460403-M15B02730046 460403-M15B02730047 460403-M15B02730048 460403-M15B02730049 460403-M15B02730050 460403-M15B02730051 460403-M15B02730052 460403-M15B02730053 460403-M15B02730054 460403-M15B02730055 460403-M15B02730056 460403-M15B02730057 460403-M15B02730058 460403-M15B02730059 460403-M15B02730060 460403-M15B02730061 460403-M15B02730062 460403-M15B02730063 460403-M15B02730064 460403-M15B02730065 460403-M16204580001 460403-M16204580002 460403-M16204580003 460403-M16204580004 460403-M16204580005 460403-M16204580006 460403-M16204580007 460403-M16204580008 460403-M16204580009 460403-M16204580010 460403-M16204580011 460403-M16204580012 460403-M16204580013 460403-M16204580014 460403-M16204580015 460403-M16204580016 460403-M16204580017 460403-M16204580018 460403-M16204580019 460403-M16204580020 460403-M16204580021 460403-M16204580022 460403-M16204580023 460403-M16204580024 460403-M16204580025 460403-M16204580026 460403-M16204580027 460403-M16204580028 460403-M16204580029 460403-M16204580030 460403-M16204580031 460403-M16204580032 460403-M16204580033 460403-M16204580034 460403-M16204580035 460403-M16204580036 460403-M16204580037 460403-M16204580038 460403-M16204580039 460403-M16204580040 460403-M16204580041 460403-M16204580042 460403-M16204580043 460403-M16204580044 460403-M16204580045 460403-M16204580046 460403-M16204580047 460403-M16204580048 460403-M16204580049 460403-M16204580050 460403-M16204580051 460403-M16204580052 460403-M16204580053 460403-M16204580054 460403-M16204580055 460403-M16204580056 460403-M16204580057 460403-M16204580058 460403-M16204580059 460403-M16204580060 460403-M16410170001 460403-M16410170002 460403-M16410170003 460403-M16410170004 460403-M16410170005 460403-M16410170006 460403-M16410170007 460403-M16410170008 460403-M16410170009 460403-M16410170010 460403-M16410170011 460403-M16410170012 460403-M16410170013 460403-M16410170014 460403-M16410170015 460403-M16410170016 460403-M16410170017 460403-M16410170018 460403-M16410170019 460403-M16410170020 460403-M16410170021 460403-M16410170022 460403-M16410170023 460403-M16410170024 460403-M16410170025 460403-M16410170026 460403-M16410170027 460403-M16410170028 460403-M16410170029 460403-M16410170030 460403-M16410170031 460403-M16410170032 460403-M16410170033 460403-M16410170034 460403-M16410170035 460403-M16410170036 460403-M16811370001 460403-M16811370002 460403-M16811370003 460403-M16811370004 460403-M16811370005 460403-M16811370006 460403-M16811370007 460403-M16811370008 460403-M16811370009 460403-M16811370010 460403-M16811370011 460403-M16811370012 460403-M16811370013 460403-M16811370014 460403-M16811370015 460403-M16811370016 460403-M16811370017 460403-M16811370018 460403-M16811370019 460403-M16811370020 460403-M16811370021 460403-M16811370022 460403-M16811370023 460403-M16811370024 460403-M16811370025 460403-M16811370026 460403-M16811370027 460403-M16811370028 460403-M16811370029 460403-M16811370030 460403-M16811370031 460403-M16811370032 460403-M16811370033 460403-M16811370034 460403-M16811370035 460403-M16811370036 460403-M16811370037 460403-M16811370038 460403-M16811370039 460403-M16811370040 460403-M16811370041 460403-M16811370042 460403-M16811370043 460403-M16811370044 460403-M16811370045 460403-M16811370046 460403-M16811370047 460403-M16811370048 460403-M16811370049 460403-M16811370050 460403-M16811370051 460403-M16811370052 460403-M16811370053 460403-M16811370054 460403-M16811370055 460403-M16811370056 460403-M17205680001 460403-M17205680002 460403-M17205680003 460403-M17205680004 460403-M17205680005 460403-M17205680006 460403-M17205680007 460403-M17205680008 460403-M17205680009 460403-M17205680010 460403-M17205680011 460403-M17205680012 460403-M17205680013 460403-M17205680014 460403-M17205680015 460403-M17205680016 460403-M17205680017 460403-M17205680018 460403-M17205680019 460403-M17205680020 460403-M17205680021 460403-M17205680022 460403-M17205680023 460403-M17205680024 460403-M17205680025 460403-M17205680026 460403-M17205680027 460403-M17205680028 460403-M17205680029 460403-M17205680030 460403-M17205680031 460403-M17205680032 460403-M17205680033 460403-M17205680034 460403-M17205680035 460403-M17205680036 460403-M17205680037 460403-M17205680038 460403-M17205680039 460403-M17205680040 460403-M17205680041 460403-M17205680042 460403-M17205680043 460403-M17205680044 460403-M17205680045 460403-M17205680046 460403-M17205680047 460403-M17205680048 460403-M17205680049 460403-M17205680050 460403-M17205690001 460403-M17205690002 460403-M17205690003 460403-M17205690004 460403-M17205690005 460403-M17205690006 460403-M17205690007 460403-M17205690008 460403-M17205690009 460403-M17205690010 460403-M17205690011 460403-M17205690012 460403-M17205690013 460403-M17205690014 460403-M17205690015 460403-M17205690016 460403-M17205690017 460403-M17205690018 460403-M17205690019 460403-M17205690020 460403-M17205690021 460403-M17205690022 460403-M17205690023 460403-M17205690024 460403-M17205690025 460403-M17205690026 460403-M17205690027 460403-M17205690028 460403-M17205690029 460403-M17205690030 460403-M17205690031 460403-M17205690032 460403-M17205690033 460403-M17205690034 460403-M17205690035 460403-M17205690036 460403-M17205690037 460403-M17205690038 460403-M17205690039 460403-M17205690040 460403-M17205690041 460403-M17205690042 460403-M17902760001 460403-M17902760002 460403-M17902760003 460403-M17902760004 460403-M17902760005 460403-M17902760006 460403-M17902760007 460403-M17902760008 460403-M17902760009 460403-M17902760010 460403-M17902760011 460403-M17902760012 460403-M17902760013 460403-M17902760014 460403-M17902760015 460403-M17902760016 460403-M17902760017 460403-M17902760018 460403-M17902760019 460403-M17902760020 460403-M17B10620001 460403-M17B10620002 460403-M17B10620003 460403-M17B10620004 460403-M17B10620005 460403-M17B10620006 460403-M17B10620007 460403-M17B10620008 460403-M17B10620009 460403-M17B10620010 460403-M17B10620011 460403-M17B10620012 460403-M17B10620013 460403-M17B10620014 460403-M17B10620015 460403-M17B10620016 460403-M17B10620017 460403-M17B10620018 460403-M17B10620019 460403-M17B10620020 460403-M17B10620021 460403-M17B10620022 460403-M17B10620023 460403-M17B10620024 460403-M17B10620025 460403-M17B10620026 460403-M17B10620027 460403-M17B10620028 460403-M17B10620029 460403-M17B10620030 460403-M17B10620031 460403-M17B10620032 460403-M17B10620033 460403-M17B10620034 460403-M17B10620035 460403-M17B10620036 460403-M17B10620037 460403-M17B10620038 460403-M17B10620039 460403-M17B10620040 460403-M18200670001 460403-M18200670002 460403-M18200670003 460403-M18200670004 460403-M18200670005 460403-M18200670006 460403-M18200670007 460403-M18200670008 460403-M18200670009 460403-M18200670010 460403-M18200670011 460403-M18200670012 460403-M18200670013 460403-M18200670014 460403-M18200670015 460403-M18200670016 460403-M18200670017 460403-M18200670018 460403-M18200670019 460403-M18200670020 460403-M18200670021 460403-M18200670022 460403-M18200670023 460403-M18200670024 460403-M18200670025 460403-M18200670026 460403-M18200670027 460403-M18200670028 460403-M18200670029 460403-M18200670030 460403-M18200670031 460403-M18200670032 460511-M16910870001 460511-M16910870002 460511-M17311840001 460512-M16700300001 460512-M16700300002 460512-M16700300003 460957-M18504380001 460957-M18504380002 460957-M18504380003 460957-M18504380004 460957-M18504380005 460957-M18504380006 460957-M18504380007 460957-M18504380008 460957-M18504380009 460957-M18504380010 460957-M18504380011 460957-M18504380012 460957-M18504380013 460957-M18504380014 460957-M18504380015 460957-M18504380016 460957-M18504380017 460957-M18504380018 460957-M18504380019 460957-M18504380020 460957-M18504380021 460957-M18504380022 460957-M18504380023 460957-M18504380024 460957-M18504380025 460957-M18504380026 460957-M18504380027 460957-M18504380028 460957-M18504380029 460957-M18504380030 460957-M18504380031 460957-M18504380032 460957-M18504380033 460957-M18504380034 460957-M18504380035 460957-M18504380036 460957-M18700650001 460957-M18700650002 460957-M18700650003 460957-M18700650004 460957-M18700650005 460957-M18700650006 460957-M18700650007 460957-M18700650008 460957-M18700650009 460957-M18700650010 460957-M18700650011 460957-M18700650012 460957-M18700650013 460957-M18700650014 460957-M18700650015 460957-M18700650016 460957-M18700650017 460957-M18700650018 460957-M18700650019 460957-M18700650020 460957-M18700650021 460957-M18700650022 460957-M18700650023 460957-M18700650024 460957-M18700650025 460957-M18700650026 460957-M18700650027 460957-M18700650028 460957-M18700650029 460957-M18700650030 460957-M18700650031 460957-M18700650032 460957-M18700650033 460957-M18700650034 460957-M18700650035 460957-M18700650036 460957-M18700650037 460957-M18700650038 460957-M18700650039 460957-M18700650040 460957-M18802590001 460957-M18802590002 460957-M18802590003 460957-M18802590004 460957-M18802590005 460957-M18802590006 460957-M18802590007 460957-M18802590008 460957-M18802590009 460957-M18802590010 460957-M18802590011 460957-M18802590012 460957-M18802590013 460957-M18802590014 460957-M18802590015 460957-M18802590016 460957-M18802590017 460957-M18802590018 460957-M18802590019 460957-M18802590020 460957-M18802590021 460957-M18802590022 460957-M18802590023 460957-M18802590024 460957-M18802590025 460957-M18802590026 460957-M18802590027 460957-M18802590028 460957-M18802590029 460957-M18802590030 460957-M18802590031 460957-M18802590032 460957-M18802590033 460957-M18802590034 460957-M18802590035 460957-M18802590036 460957-M18B00440001 460957-M18B00440002 460957-M18B00440003 460957-M18B00440004 460957-M18B00440005 460957-M18B00440006 460957-M18B00440007 460957-M18B00440008 460957-M18B00440009 460957-M18B00440010 460957-M18B00440011 460957-M18B00440012 460957-M18B00440013 460957-M18B00440014 460957-M18B00440015 460957-M18B00440016 460957-M18B00440017 460957-M18B00440018 460957-M18B00440019 460957-M18B00440020 460957-M18B00440021 460957-M18B00440022 460957-M18B00440023 460957-M18B00440024 460957-M18B00440025 460957-M18B00440026 460957-M18B00440027 460957-M18B00440028 460957-M18B00440029 460957-M18B00440030 460957-M18B00440031 460957-M18B00440032 460957-M19313700001 460957-M19313700002 460957-M19313700003 460957-M19313700004 460957-M19313700005 460957-M19313700006 460957-M19313700007 460957-M19313700008 460957-M19313700009 460957-M19313700010 460957-M19313700011 460957-M19313700012 460957-M19313700013 460957-M19313700014 460957-M19313700015 460957-M19313700016 460957-M19313700017 460957-M19313700018 460957-M19313700019 460957-M19313700020 460957-M19313700021 460957-M19313700022 460957-M19313700023 460957-M19313700024 460957-M19313700025 460957-M19313700026 460957-M19313700027 460957-M19313700028 460957-M19313700029 460957-M19313700030 460957-M19313700031 460957-M19313700032 460957-M19313700033 460957-M19313700034 460957-M19313700035 460957-M19313700036 460957-M19313700037 460957-M19313700038 460957-M19313700039 460957-M19313700040 460957-M19313700041 460957-M19313700042 460957-M19313700043 460957-M19313700044 460957-M19313710001 460957-M19313710002 460957-M19313710003 460957-M19313710004 460957-M19313710005 460957-M19313710006 460957-M19313710007 460957-M19313710008 460957-M19313710009 460957-M19313710010 460957-M19313710011 460957-M19313710012 460957-M19313710013 460957-M19313710014 460957-M19313710015 460957-M19313710016 460957-M19703110001 460957-M19703110002 460957-M19703110003 460957-M19703110004 460957-M19703110005 460957-M19703110006 460957-M19703110007 460957-M19703110008 460957-M19703110009 460957-M19703110010 460957-M19703110011 460957-M19703110012 460957-M19703110013 460957-M19703110014 460957-M19703110015 460957-M19703110016 460957-M19703110017 460957-M19703110018 460957-M19703110019 460957-M19703110020 460957-M19703220001 460957-M19703220002 460957-M19703220003 460957-M19703220004 460957-M19703220005 460957-M19703220006 460957-M19703220007 460957-M19703220008 460957-M19703220009 460957-M19703220010 460957-M19703220011 460957-M19703220012 460957-M19703220013 460957-M19703220014 460957-M19703220015 460957-M19703220016 460957-M19703220017 460957-M19703220018 460957-M19703220019 460957-M19703220020 460957-M19703500001 460957-M19703500002 460957-M19703500003 460957-M19703500004 460957-M19703500005 460957-M19703500006 460957-M19703500007 460957-M19703500008 460957-M19703500009 460957-M19703500010 460957-M19703500011 460957-M19703500012 460957-M19703500013 460957-M19703500014 460957-M19703500015 460957-M19703500016 460957-M19703500017 460957-M19703500018 460957-M19703500019 460957-M19703500020 460957-M19703500021 460957-M19703500022 460957-M19703500023 460957-M19703500024 460957-M19703500025 460957-M19703500026 460957-M19703500027 460957-M19703500028 460957-M19703500029 460957-M19703500030 460957-M19703500031 460957-M19703500032 460957-M19703500033 460957-M19703500034 460957-M19703500035 460957-M19703500036 460957-M19703500037 460957-M19703500038 460957-M19703500039 460957-M19703500040 460957-M19703510001 460957-M19703510002 460957-M19703510003 460957-M19703510004 460957-M19703510005 460957-M19703510006 460957-M19703510007 460957-M19703510008 460957-M19703510009 460957-M19703510010 460957-M19703510011 460957-M19703510012 460957-M19703510013 460957-M19703510014 460957-M19703510015 460957-M19703510016 460957-M19704690001 460957-M19704690002 460957-M19704690003 460957-M19704690004 460957-M19704690005 460957-M19704690006 460957-M19704690007 460957-M19704690008 460957-M19704690009 460957-M19704690010 460957-M19704690011 460957-M19704690012 460957-M19704690013 460957-M19704690014 460957-M19704690015 460957-M19704690016 460957-M19704690017 460957-M19704690018 460957-M19704690019 460957-M19704690020 460957-M19704690021 460957-M19704690022 460957-M19704690023 460957-M19704690024 460957-M19B00980001 460957-M19B00980002 460957-M19B00980003 460957-M19B00980004 460957-M19B00980005 460957-M19B00980006 460957-M19B00980007 460957-M19B00980008 460957-M19B00980009 460957-M19B00980010 460957-M19B02180001 460957-M19B02180002 460957-M19B02180003 460957-M19B02180004 460957-M19B02180005 460957-M19B02180006 460957-M19B02180007 460957-M19B02180008 460957-M19B02180009 460957-M19B02180010 460957-M19B02180011 460957-M19B02180012 460957-M19B02180013 460957-M19B02180014 460957-M19B02180015 460957-M19B02180016 460957-M19B02180017 460957-M19B02180018 460957-M19B02180019 460957-M19B02180020 460957-M19B02180021 460957-M19B02180022 460957-M19B02180023 460957-M19B02180024 460957-M19B02180025 460957-M19B02180026 460957-M19B02180027 460957-M19B02180028 460957-M19B02180029 460957-M19B02180030 460957-M19B02180031 460957-M19B02180032 460957-M19B02180033 460957-M19B02180034 460957-M19B02180035 460957-M19B02180036 460957-M19B02180037 460957-M19B02180038 460957-M19B02180039 460957-M19B02180040 460957-M19B02190001 460957-M19B02190002 460957-M19B02190003 460957-M19B02190004 460957-M19B02190005 460957-M19B02190006 460957-M19B02190007 460957-M19B02190008 460957-M19B02190009 460957-M19B02190010 460957-M19B02190011 460957-M19B02190012 460957-M19B02190013 460957-M19B02190014 460957-M19B02190015 460957-M19B02190016 460957-M19B02190017 460957-M19B02190018 460957-M19B02190019 460957-M19B02190020 460957-M19B02190021 460957-M19B02190022 460957-M19B02190023 460957-M19B02190024 460957-M19B02190025 460957-M19B02190026 460957-M19B02190027 460957-M19B02190028 460957-M19B02190029 460957-M19B02190030 460957-M19B02190031 460957-M19B02190032 460957-M19B02190033 460957-M19B02190034 460957-M19B02190035 460957-M19B02190036 460957-M19B02190037 460957-M19B02190038 460957-M19B02190039 460957-M19B02190040 460957-M20308630001 460957-M20308630002 460957-M20308630003 460957-M20308630004 460957-M20308630005 460957-M20308630006 460957-M20308660001 460957-M20308660002 460957-M20308660003 460957-M20308660004 460957-M20308660005 460957-M20308660006 460957-M20308660007 460957-M20308660008 460957-M20308660009 460957-M20308660010 460957-M20308660011 460957-M20308660012 460957-M20308660013 460957-M20308660014 460957-M20308660015 460957-M20308660016 460957-M20308660017 460957-M20308660018 460957-M20308660019 460957-M20308660020 460957-M20308660021 460957-M20308660022 460957-M20308660023 460957-M20308660024 460957-M20308660025 460957-M20308660026 460957-M20308660027 460957-M20308660028 460957-M20308660029 460957-M20308660030 460957-M20308660031 460957-M20308660032 460957-M20308660033 460957-M20308660034 460957-M20308710001 460957-M20308710002 460957-M20308710003 460957-M20308710004 460957-M20308710005 460957-M20308710006 460957-M20308710007 460957-M20308710008 460957-M20308710009 460957-M20308710010 460957-M20308710011 460957-M20308710012 460957-M20308710013 460957-M20308710014 460957-M20308710015 460957-M20308710016 460957-M20308710017 460957-M20308710018 460957-M20308710019 460957-M20308710020 460957-M20308710021 460957-M20308710022 460957-M20308710023 460957-M20308710024 460957-M20308710025 460957-M20308710026 460957-M20308710027 460957-M20308710028 460957-M20308710029 460957-M20308710030 460957-M20308710031 460957-M20308710032 460985-M21500880001 460985-M21500880002 460985-M22208620001 460985-M22208620004 460985-M22708790001 460985-M22708790002 460985-M22708790003 460985-M22708790004 460985-M22708790005 460985-M22708790006 460985-M23110890001 460985-M23110890002 460985-M23110890004 460985-M23110890006 460985-M23110890007 460985-M23110890008 461053-M21203260001 461053-M21203260002 461053-M21203260003 461053-M21203260004 461053-M21203260005 461053-M21203260006 461053-M21203260007 461053-M21203260008 461053-M21203260009 461053-M21203260010 461053-M21203260011 461053-M21203260012 461053-M21203260013 461053-M21203260014 461053-M21203260015 461053-M21303100001 461053-M21303100002 461053-M21303100003 461053-M21303100004 461053-M21303100005 461053-M21303100006 461053-M21303100007 461053-M21303100008 461053-M21303100009 461053-M21303100010 461053-M21303100011 461053-M21303100012 461053-M21303100013 461053-M21303100014 461053-M21303100015 461053-M21303100016 461053-M21303100017 461053-M21303100018 461053-M21303100019 461053-M21303100020 461053-M21303100021 461053-M21303100022 461053-M21303100023 461053-M21303100024 461053-M21303100025 461053-M21303100026 461053-M21303100027 461053-M21303100028 461053-M21303100029 461053-M21303100030 461053-M21410340001 461053-M21410340002 461053-M21410340003 461053-M21410340004 461053-M21410340005 461053-M21410340006 461053-M21410340007 461053-M21410340008 461053-M21410340009 461053-M21410340010 461053-M21410340011 461053-M21410340012 461053-M21410340013 461053-M21410340014 461053-M21410340015 461053-M21410340016 461053-M21410340017 461053-M21410340018 461053-M21410340019 461053-M21410340020 461053-M21410340021 461053-M21410340022 461053-M21410340023 461053-M21410340024 461053-M21410340025 461053-M21410340026 461053-M21410340027 461053-M21410340028 461053-M21410340029 461053-M21410340030 461053-M21410340031 461053-M21410340032 461053-M21410340033 461053-M21410340034 461053-M21410340035 461053-M21800900001 461053-M21800900002 461053-M21800900003 461053-M21800900004 461053-M21800900005 461053-M21800900006 461053-M21800900007 461053-M21800900008 461053-M21800900009 461053-M21800900010 461053-M21800900011 461053-M21800900012 461053-M21800900013 461053-M21800900014 461053-M21800900015 461053-M21800900016 461053-M21800900017 461053-M21800900018 461053-M21800900019 461053-M21800900020 461053-M21800900021 461053-M21800900022 461053-M21808250001 461053-M21808250002 461053-M21808250003 461053-M21808250004 461053-M21808250005 461053-M21808250006 461053-M21808250007 461053-M21808250008 461053-M21808250009 461053-M21808250010 461053-M21808250011 461053-M21808250012 461053-M21808250013 461053-M21808250014 461053-M21808250015 461053-M21808250016 461053-M21808250017 461053-M21808250018 461053-M21808250019 461053-M21808250020 461053-M21808250021 461053-M21808250022 461053-M21808250023 461053-M21808250024 461053-M21808250025 461053-M21808250026 461053-M21808250027 461053-M21808250028 461053-M21808250029 461053-M21808250030 461053-M21808250031 461053-M21808250032 461053-M21808250033 461053-M21808250034 461053-M21808250035 461053-M21808250036 461053-M21808250037 461053-M21808250038 461053-M21808250039 461053-M21808250040 461053-M21808250041 461053-M21808250042 461053-M21808250043 461053-M21808250044 461053-M21808250045 461053-M21808250046 461053-M21808250047 461053-M21808250048 461053-M21808250049 461053-M21808250050 461053-M21808250051 461053-M21808250052 461053-M21808250053 461053-M21808250054 461053-M21808250055 461053-M21808250056 461053-M21808250057 461053-M21808250058 461053-M21808250059 461053-M21808250060 461053-M21808250061 461053-M21808250062 461053-M21808250063 461053-M21808250064 461053-M21808250065 461053-M21808280001 461053-M21808280002 461053-M21808280003 461053-M21808280004 461053-M21808280005 461053-M21808280006 461053-M21808280007 461053-M21808280008 461053-M21808280009 461053-M21808280010 461053-M21808280011 461053-M21808280012 461053-M21808280013 461053-M21808280014 461053-M21808280015 461053-M21808280016 461053-M21808280017 461053-M21808280018 461053-M21808280019 461053-M21808280020 461053-M21808280021 461053-M21808280022 461053-M21808280023 461053-M21808280024 461053-M21808280025 461053-M21808290001 461053-M21808290002 461053-M21808290003 461053-M21808290004 461053-M21808290005 461053-M21808290006 461053-M21808290007 461053-M21808290008 461053-M21808290009 461053-M21808290010 461053-M21808290011 461053-M21808290012 461053-M21808290013 461053-M21808290014 461053-M21808290015 461053-M21808290016 461053-M21808290017 461053-M21808290018 461053-M21808290019 461053-M21808290020 461053-M21808290021 461053-M21808290022 461053-M21808290023 461053-M21808290024 461053-M21808290025 461053-M21808290026 461053-M21808290027 461053-M21808290028 461053-M21808290029 461053-M21808290030 461053-M21808290031 461053-M21808290032 461053-M21808290033 461053-M21808290034 461053-M21808290035 461053-M21808290036 461053-M21808290037 461053-M21808290038 461053-M21808290039 461053-M21808290040 461053-M22204660001 461053-M22204660002 461053-M22204660003 461053-M22204660004 461053-M22204660005 461053-M22204660006 461053-M22204660007 461053-M22204660008 461053-M22204660009 461053-M22204660010 461053-M22204660011 461053-M22204660012 461053-M22204660013 461053-M22204660014 461053-M22204660015 461053-M22204660016 461053-M22204660017 461053-M22204660018 461053-M22204660019 461053-M22204660020 461053-M22204660021 461053-M22204660022 461053-M22204660023 461053-M22204660024 461053-M22204660025 461053-M22204660026 461053-M22204660027 461053-M22204660028 461053-M22204660029 461053-M22204660030 461053-M22204660031 461053-M22204660032 461053-M22204660033 461053-M22204660034 461053-M22204660035 461053-M22204660036 461053-M22204660037 461053-M22204660038 461053-M22204660039 461053-M22204660040 461053-M22204660041 461053-M22204660042 461053-M22204660043 461053-M22204660044 461053-M22204660045 461053-M22204660046 461053-M22204660047 461053-M22204660048 461053-M22204660049 461053-M22204660050 461053-M22204670001 461053-M22204670002 461053-M22204670003 461053-M22204670004 461053-M22204670005 461053-M22204670006 461053-M22204670007 461053-M22204670008 461053-M22204670009 461053-M22204670010 461053-M22204670011 461053-M22204670012 461053-M22204670013 461053-M22204670014 461053-M22204670015 461053-M22204670016 461053-M22204670017 461053-M22204670018 461053-M22204670019 461053-M22204670020 461053-M22204670021 461053-M22204670022 461053-M22204670023 461053-M22204670024 461053-M22204670025 461053-M22204670026 461053-M22204670027 461053-M22204670028 461053-M22204670029 461053-M22204670030 461053-M22204670031 461053-M22204670032 461053-M22204670033 461053-M22204670034 461053-M22204670035 461053-M22204670036 461053-M22204670037 461053-M22204670038 461053-M22204670039 461053-M22204670040 461053-M22204670041 461053-M22204670042 461053-M22204670043 461053-M22204670044 461053-M22204670045 461053-M22204670046 461053-M22204670047 461053-M22204670048 461053-M22204670049 461053-M22204670050 461053-M22204680001 461053-M22204680002 461053-M22204680003 461053-M22204680004 461053-M22204680005 461053-M22204680006 461053-M22204680007 461053-M22204680008 461053-M22204680009 461053-M22204680010 461053-M22204680011 461053-M22204680012 461053-M22204680013 461053-M22204680014 461053-M22204680015 461053-M22204680016 461053-M22204680017 461053-M22204680018 461053-M22204680019 461053-M22204680020 461053-M22204680021 461053-M22204680022 461053-M22204680023 461053-M22204680024 461053-M22204680025 461053-M22705590001 461053-M22705590002 461053-M22705590003 461053-M22705590004 461053-M22705590005 461053-M22705590006 461053-M22705590007 461053-M22705590008 461053-M22705590009 461053-M22705590010 461053-M22705590011 461053-M22705590012 461053-M22705590013 461053-M22705590014 461053-M22705590015 461053-M22705590016 461053-M22705590017 461053-M22705590018 461053-M22705590019 461053-M22705590020 461053-M22705590021 461053-M22705590022 461053-M22705590023 461053-M22705590024 461053-M22705590025 461053-M22705590026 461053-M22705590027 461053-M22705590028 461053-M22705590029 461053-M22705590030 461053-M22705590031 461053-M22705590032 461053-M22705590033 461053-M22705590034 461053-M22705590035 461053-M22705590036 461053-M22705590037 461053-M22705590038 461053-M22705590039 461053-M22705590040 461053-M22705590041 461053-M22705590042 461053-M22705590043 461053-M22705590044 461053-M22705590045 461053-M22705590046 461053-M22705590047 461053-M22705590048 461053-M22705590049 461053-M22705590050 461053-M22810900001 461053-M22810900002 461053-M22810900003 461053-M22810900004 461053-M22810900005 461053-M22810900006 461053-M22810900007 461053-M22810900008 461053-M22810900009 461053-M22810900010 461053-M22810900011 461053-M22810900012 461053-M22810900013 461053-M22810900014 461053-M22810900015 461053-M22810900016 461053-M22810900017 461053-M22810900018 461053-M22810900019 461053-M22810900020 461053-M22810900021 461053-M22810900022 461053-M22810900023 461053-M22810900024 461053-M22810900025 461053-M22810900026 461053-M22810900027 461053-M22810900028 461053-M22810900029 461053-M22810900030 461053-M22810900031 461053-M22810900032 461053-M22810900033 461053-M22810900034 461053-M22810900035 461053-M22810900036 461053-M22810900037 461053-M22810900038 461053-M22810900039 461053-M22810900040 461053-M22810900041 461053-M22810900042 461053-M22810900043 461053-M22810900044 461053-M22810900045 461053-M22810900046 461053-M22810900047 461053-M22810900048 461053-M22810900049 461053-M22810900050 461053-M22810910001 461053-M22810910002 461053-M22810910003 461053-M22810910004 461053-M22810910005 461053-M22810910006 461053-M22810910007 461053-M22810910008 461053-M22810910009 461053-M22810910010 461053-M22810910011 461053-M22810910012 461053-M22810910013 461053-M22810910014 461053-M22810910015 461053-M22810910016 461053-M22810910017 461053-M22810910018 461053-M22810910019 461053-M22810910020 461053-M22810910021 461053-M22810910022 461053-M22810910023 461053-M22810910024 461053-M22810910025 461053-M22810910026 461053-M22810910027 461053-M22810910028 461053-M22810910029 461053-M22810910030 461053-M22810910031 461053-M22810910032 461053-M22810910033 461053-M22810910034 461053-M22810910035 461053-M22810910036 461053-M22810910037 461053-M22810910038 461053-M22810910039 461053-M22810910040 461053-M22810910041 461053-M22810910042 461053-M22810910043 461053-M22810910044 461053-M22810910045 461053-M22810910046 461053-M22810910047 461053-M22810910048 461053-M22810910049 461053-M22810910050 461053-M22904340001 461053-M22904340002 461053-M22904340003 461053-M22904340004 461053-M22904340005 461053-M22904340006 461053-M22904340007 461053-M22904340008 461053-M22904340009 461053-M22904340010 461053-M22904340011 461053-M22904340012 461053-M22904340013 461053-M22904340014 461053-M22904340015 461053-M22904340016 461053-M22904340017 461053-M22904340018 461053-M22904340019 461053-M22904340020 461053-M22904340021 461053-M22904340022 461053-M22904340023 461053-M22904340024 461053-M22904340025 461053-M22904340026 461053-M22904340027 461053-M22904340028 461053-M22904340029 461053-M22904340030 461053-M22904340031 461053-M22904340032 461053-M22904340033 461053-M22904340034 461053-M22904340035 461053-M22904340036 461053-M22904340037 461053-M22904340038 461053-M22904340039 461053-M22904340040 461053-M22904340041 461053-M22904340042 461053-M22904340043 461053-M22904340044 461053-M22904340045 461053-M22904340046 461053-M22904340047 461053-M22904340048 461053-M22904340049 461053-M22904340050 461053-M23221740001 461053-M23221740002 461053-M23221740003 461053-M23221740004 461053-M23221740005 461053-M23221740006 461053-M23221740007 461053-M23221740008 461053-M23221740009 461053-M23221740010 461053-M23221740011 461053-M23221740012 461053-M23221740013 461053-M23221740014 461053-M23221740015 461053-M23221740016 461053-M23221740017 461053-M23221740018 461053-M23221740019 461053-M23221740020 461053-M23221740021 461053-M23221740022 461053-M23221740023 461053-M23221740024 461053-M23221740025 461053-M23221740026 461053-M23221740027 461053-M23221740028 461053-M23221740029 461053-M23221740030 461053-M23221740031 461053-M23221740032 461053-M23221740033 461053-M23221740034 461053-M23221740035 461053-M23221740036 461053-M23221740037 461053-M23221740038 461053-M23221740039 461053-M23221740040 461053-M23221740041 461053-M23221740042 461053-M23221740043 461053-M23221740044 461053-M23221740045 461053-M23221740046 461053-M23221740047 461053-M23221740048 461053-M23221740049 461053-M23221740050 461053-M23221750001 461053-M23221750002 461053-M23221750003 461053-M23221750004 461053-M23221750005 461053-M23221750006 461053-M23221750007 461053-M23221750008 461053-M23221750009 461053-M23221750010 461053-M23221750011 461053-M23221750012 461053-M23221750013 461053-M23221750014 461053-M23221750015 461053-M23221750016 461053-M23221750017 461053-M23221750018 461053-M23221750019 461053-M23221750020 461053-M23221750021 461053-M23221750022 461053-M23221750023 461053-M23221750024 461053-M23221750025 461053-M23221750026 461053-M23221750027 461053-M23221750028 461053-M23221750029 461053-M23221750030 461053-M23221750031 461053-M23221750032 461053-M23221750033 461053-M23221750034 461053-M23221750035 461053-M23221750036 461053-M23221750037 461053-M23221750038 461053-M23221750039 461053-M23221750040 461053-M23221750041 461053-M23221750042 461053-M23221750043 461053-M23221750044 461053-M23221750045 461053-M23221750046 461053-M23221750047 461053-M23221750048 461053-M23221750049 461053-M23221750050 461053-M23311860001 461053-M23311860002 461053-M23311860003 461053-M23311860004 461053-M23311860005 461053-M23311860006 461053-M23311860007 461053-M23311860008 461053-M23311860009 461053-M23311860010 461053-M23311860011 461053-M23311860012 461053-M23311860013 461053-M23311860014 461053-M23311860015 461053-M23311860016 461053-M23311860017 461053-M23311860018 461053-M23311860019 461053-M23311860020 461053-M23311860021 461053-M23311860022 461053-M23311860023 461053-M23311860024 461053-M23311860025 461053-M23311860026 461053-M23311860027 461053-M23311860028 461053-M23311860029 461053-M23311860030 461053-M23311860031 461053-M23311860032 461053-M23311860033 461053-M23311860034 461053-M23311860035 461053-M23311860036 461053-M23311860037 461053-M23311860038 461053-M23311860039 461053-M23311860040 461053-M23311860041 461053-M23311860042 461053-M23311860043 461053-M23311860044 461053-M23311860045 461053-M23311860046 461053-M23311860047 461053-M23311860048 461053-M23311860049 461053-M23311860050 461053-M23912690001 461053-M23912690002 461053-M23912690003 461053-M23912690004 461053-M23912690005 461053-M23912690006 461053-M23912690007 461053-M23912690008 461053-M23912690009 461053-M23912690010 461053-M23912690011 461053-M23912690012 461053-M23912690013 461053-M23912690014 461053-M23912690015 461053-M23912690016 461053-M23912690017 461053-M23912690018 461053-M23912690019 461053-M23912690020 461053-M23912690021 461053-M23912690022 461053-M23912690023 461053-M23912690024 461053-M23912690025 461053-M23912690026 461053-M23912690027 461053-M23912690028 461053-M23912690029 461053-M23912690030 461053-M23912690031 461053-M23912690032 461053-M23912690033 461053-M23912690034 461053-M23912690035 461053-M23912690036 461053-M23912690037 461053-M23912690038 461053-M23912690039 461053-M23912690040 461053-M23912690041 461053-M23912690042 461053-M23912690043 461053-M23912690044 461053-M23912690045 461053-M23912690046 461053-M23912690047 461053-M23912690048 461053-M23912690049 461053-M23912690050 461053-M23C17670001 461053-M23C17670002 461053-M23C17670003 461053-M23C17670004 461053-M23C17670005 461053-M23C17670006 461053-M23C17670007 461053-M23C17670008 461053-M23C17670009 461053-M23C17670010 461053-M23C17670011 461053-M23C17670012 461053-M23C17670013 461053-M23C17670014 461053-M23C17670015 461053-M23C17670016 461053-M23C17670017 461053-M23C17670018 461053-M23C17670019 461053-M23C17670020 461053-M23C17670021 461053-M23C17670022 461053-M23C17670023 461053-M23C17670024 461053-M23C17670025 461053-M23C17670026 461053-M23C17670027 461053-M23C17670028 461053-M23C17670029 461053-M23C17670030 461053-M23C17670031 461053-M23C17670032 461053-M23C17670033 461053-M23C17670034 461053-M23C17670035 461053-M23C17670036 461053-M23C17670037 461053-M23C17670038 461053-M23C17670039 461053-M23C17670040 461053-M23C17670041 461053-M23C17670042 461053-M23C17670043 461053-M23C17670044 461053-M23C17670045 461053-M23C17670046 461053-M23C17670047 461053-M23C17670048 461053-M23C17670049 461053-M23C17670050 461053-M23C17950001 461053-M23C17950002 461053-M23C17950003 461053-M23C17950004 461053-M23C17950005 461053-M23C17950006 461053-M23C17950007 461053-M23C17950008 461053-M23C17950009 461053-M23C17950010 461053-M23C17950011 461053-M23C17950012 461053-M23C17950013 461053-M23C17950014 461053-M23C17950015 461053-M23C17950016 461053-M23C17950017 461053-M23C17950018 461053-M23C17950019 461053-M23C17950020 461053-M23C17950021 461053-M23C17950022 461053-M23C17950023 461053-M23C17950024 461053-M23C17950025 461053-M23C17950026 461053-M23C17950027 461053-M23C17950028 461053-M23C17950029 461053-M23C17950030 461053-M23C17950031 461053-M23C17950032 461053-M23C17950033 461053-M23C17950034 461053-M23C17950035 461053-M23C17950036 461053-M23C17950037 461053-M23C17950038 461053-M23C17950039 461053-M23C17950040 461053-M23C17950041 461053-M23C17950042 461053-M23C17950043 461053-M23C17950044 461053-M23C17950045 461053-M23C17950046 461053-M23C17950047 461053-M23C17950048 461053-M23C17950049 461053-M23C17950050 461053-M23C17960001 461053-M23C17960002 461053-M23C17960003 461053-M23C17960004 461053-M23C17960005 461053-M23C17960006 461053-M23C17960007 461053-M23C17960008 461053-M23C17960009 461053-M23C17960010 461053-M23C17960011 461053-M23C17960012 461053-M23C17960013 461053-M23C17960014 461053-M23C17960015 461053-M23C17960016 461053-M23C17960017 461053-M23C17960018 461053-M23C17960019 461053-M23C17960020 461053-M23C17960021 461053-M23C17960022 461053-M23C17960023 461053-M23C17960024 461053-M23C17960025 461053-M23C17960026 461053-M23C17960027 461053-M23C17960028 461053-M23C17960029 461053-M23C17960030 461053-M23C17960031 461053-M23C17960032 461053-M23C17960033 461053-M23C17960034 461053-M23C17960035 461053-M23C17960036 461053-M23C17960037 461053-M23C17960038 461053-M23C17960039 461053-M23C17960040 461053-M23C17960041 461053-M23C17960042 461053-M23C17960043 461053-M23C17960044 461053-M23C17960045 461053-M23C17960046 461053-M23C17960047 461053-M23C17960048 461053-M23C17960049 461053-M23C17960050 461053-M25611990001 461053-M25611990002 461053-M25611990003 461053-M25611990004 461053-M25611990005 461053-M25611990006 461053-M25611990007 461053-M25611990008 461053-M25611990009 461053-M25611990010 461053-M25611990011 461053-M25611990012 461053-M25611990013 461053-M25611990014 461053-M25611990015 461053-M25611990016 461053-M25611990017 461053-M25611990018 461053-M25611990019 461053-M25611990020 461053-M25611990021 461053-M25611990022 461053-M25611990023 461053-M25611990024 461053-M25611990025 461053-M25611990026 461053-M25611990027 461053-M25611990028 461053-M25611990029 461053-M25611990030 461053-M25611990031 461053-M25611990032 461053-M25611990033 461053-M25611990034 461053-M25611990035 461053-M25611990036 461053-M25611990037 461053-M25611990038 461053-M25611990039 461053-M25611990040 461053-M25611990041 461053-M25611990042 461053-M25611990043 461053-M25611990044 461053-M25611990045 461053-M25611990046 461053-M25611990047 461053-M25611990048 461053-M25611990049 461053-M25611990050 461053-M25611990051 461053-M25611990052 461053-M25611990053 461053-M25611990054 461053-M25611990055 461053-M25611990056 461053-M25611990057 461053-M25611990058 461053-M25611990059 461234-M22603160001 560034-M17812060001 560034-M18700480001 560034-M18908300001 560034-M18908300002 560034-M20405450001 560034-M20809580001 560034-M20B12040001 560034-M20C16880001 560034-M22406140001 560034-M23401580001 EDAN Model: F2 UDI-DI code: 06944413800861 Serial/Lot Numbers: 460305-M15B02670001 460305-M15B02670002 460305-M15B02670003 460305-M15B02670004 460305-M15B02670005 460305-M16103460001 460305-M16103460002 460305-M16103460003 460305-M16103460004 460305-M16103460005 460540-M16900210001 460540-M16900210002 460540-M16900210003 460540-M16900210004 460540-M16900210005 460324-M16B07150001 460540-M16B07460001 460540-M16B07460002 460540-M16B07460003 460540-M16B07460004 460540-M16B07460005 460540-M16B07460006 460540-M16B07460007 460540-M16B07460008 460540-M16B07460009 460540-M16B07460010 460540-M16B07460011 460540-M16B07460012 460540-M16B07460013 460324-M16C00080001 460324-M16C00080002 460324-M16C00080003 460540-M16B07470001 460540-M16B07470002 460540-M16B07470003 460540-M16B07470004 460540-M16B07470005 460540-M16B07470006 460540-M16B07470007 460540-M16B07470008 460540-M16B07470009 460540-M16B07470010 460540-M16B07470011 460540-M16B07470012 460540-M17200670003 460540-M17200670004 460540-M17200670005 460540-M17200670008 460540-M17200670007 460540-M17200670001 460540-M17200670009 460540-M17200670002 460540-M17200670006 460540-M17200670010 460324-M17301870001 460324-M17303600003 460324-M17303600008 460324-M17303600006 460324-M17303600002 460324-M17303600004 460324-M17303600005 460324-M17303600010 460324-M17303600001 460324-M17303600007 460324-M17303600009 460324-M17306710001 460324-M17408130001 460324-M17600330004 460324-M17600330003 460324-M17600330001 460324-M17600330002 460540-M17605140001 460540-M17605140003 460540-M17605140002 460324-M17603720004 460324-M17603720006 460324-M17603720008 460324-K17600120007 460324-K17600120008 460324-M17603720007 460324-K17600120005 460324-M17603720003 460324-K17600120002 460324-K17600120006 460324-M17605420001 460324-M17603720005 460324-K17600120001 460324-K17600120009 460324-K17600120010 460324-K17600120003 460324-K17600120004 460324-M17603720001 460324-M17603720002 460324-M17605420002 460540-M17704960001 460540-M17704960003 460540-M17704960002 460540-M17708050001 460540-M17708060003 460540-M17708060005 460540-M17708060007 460540-M17708060001 460540-M17708060002 460540-M17708060004 460540-M17708060009 460540-M17708060006 460540-M17708060008 460540-M17708060010 460540-M17706340007 460540-M17706340001 460540-M17706340005 460540-M17706340010 460540-M17706340004 460540-M17706340009 460540-M17706340003 460700-M17910060001 460700-M17910060002 460700-M17A04230002 460700-M17A04230001 460700-M17B00410002 460700-M17B00410003 460700-M17B00410001 460700-M17B08330002 460700-M17B08330007 460700-M17B08330006 460700-M17B08330001 460700-M17B08330003 460700-M17B08330004 460700-M17B08330009 460700-M17B08330005 460700-M17B08330008 460700-M17B08330010 560051-M17C06600001 560051-M17C06600004 560051-M17C06600002 560051-M17C06600005 560051-M17C06600003 460700-M18107170014 460700-M18107170008 460700-M18107170010 460700-M18107170002 460700-M18107170012 460700-M18107170004 460700-M18107170009 460700-M18107170011 460700-M18107170013 460700-M18107170005 460700-M18107170003 460700-M18107170007 460700-M18107170015 460700-M18107170001 460700-M18107170006 560051-M18200770003 560051-M18200770001 560051-M18200770002 460324-M18313090001 460324-M18313090002 460324-M18313090003 460324-M18406420001 460324-M18406420002 460324-M18406420004 460324-M18406420008 460324-M18406420006 460324-M18406420010 460324-M18406420003 460324-M18406420007 460324-M18406420005 460324-M18406420009 560051-M18510300001 560051-M18510300002 560051-M18510300003 460324-M18508050003 460324-M18508050005 460324-M18508050010 460324-M18508050009 460324-M18508050002 460324-M18508050007 460324-M18508050001 460324-M18508050008 460324-M18508050004 460324-M18508050006 460324-M18607060002 460324-M18607060004 460324-M18607060005 460324-M18607060003 460324-M18607060006 460324-M18607060001 560051-M18700490004 560051-M18700490002 560051-M18700490003 560051-M18700490005 560051-M18700490001 560051-M18811370002 560051-M18811370005 560051-M18811370001 560051-M18811370003 560051-M18811370004 460324-M18811210001 460324-M18811210004 460324-M18811210002 460324-M18811210003 460972-M18903450015 460972-M18903450016 460972-M18903450019 460972-M18903450004 460972-M18903450008 460972-M18903450001 460972-M18903450005 460972-M18903450006 460972-M18903450012 460972-M18903450003 460972-M18903450010 460972-M18903450014 460972-M18903450017 460972-M18903450018 460972-M18903450002 460972-M18903450007 460972-M18903450009 460972-M18903450011 460972-M18903450013 460972-M18903450020 560051-M18912340002 560051-M18912340005 560051-M18912340003 560051-M18912340004 560051-M18912340001 560051-M18A09630004 560051-M18A09630001 560051-M18A09630003 560051-M18A09630002 560051-M18A09630005 460972-M18A09780003 460972-M18A09780020 460972-M18A09780021 460972-M18A09780028 460972-M18A09780009 460972-M18A09780012 460972-M18A09780015 460972-M18A09780027 460972-M18A09780014 460972-M18A09780016 460972-M18A09780030 460972-M18A09780017 460972-M18A09780019 460972-M18A09780002 460972-M18A09780005 460972-M18A09780013 460972-M18A09780024 460972-M18A09780026 460972-M18A09780007 460972-M18A09780011 460972-M18A09780018 460972-M18A09780029 460972-M18A09780001 460972-M18A09780008 460972-M18A09780010 460972-M18A09780023 460972-M18A09780004 460972-M18A09780006 460972-M18A09780022 460972-M18A09780025 560051-M18B12280004 560051-M18B12280005 560051-M18B12280003 560051-M18B12280001 560051-M18B12280002 560051-M18C04490003 560051-M18C04490002 560051-M18C04490001 560051-M19206290005 560051-M19206290007 560051-M19206290010 560051-M19206290011 560051-M19206290024 560051-M19206260015 560051-M19206290015 560051-M19206290017 560051-M19206290023 560051-M19206290002 560051-M19206290012 560051-M19206290016 560051-M19206290020 560051-M19206290025 560051-M19206290027 560051-M19206260005 560051-M19206290003 560051-M19206290026 560051-M19206290029 560051-M19206260004 560051-M19206290001 560051-M19206290008 560051-M19206290018 560051-M19206260011 560051-M19206290004 560051-M19206290014 560051-M19206290030 560051-M19206260002 560051-M19206260006 560051-M19206260009 560051-M19206260012 560051-M19206260014 560051-M19206290013 560051-M19206290022 560051-M19206290028 560051-M19206260001 560051-M19206260010 560051-M19206260013 560051-M19206290006 560051-M19206290009 560051-M19206290019 560051-M19206290021 560051-M19206260003 560051-M19206260007 560051-M19206260008 560051-M19303280003 560051-M19303280001 560051-M19303280002 560051-M19305070001 560051-M19305070002 560051-M19305070008 560051-M19305070012 560051-M19305070003 560051-M19305070004 560051-M19305070010 560051-M19305070011 560051-M19305070009 560051-M19305070015 560051-M19305070013 560051-M19305070014 560051-M19305070007 560051-M19305070005 560051-M19305070006 560051-M19402510006 560051-M19402510014 560051-M19402510024 560051-M19402510025 560051-M19402510030 560051-M19402510021 560051-M19402510022 560051-M19402510028 560051-M19402510029 560051-M19402510003 560051-M19402510007 560051-M19402510011 560051-M19402510012 560051-M19402510018 560051-M19402510019 560051-M19402510034 560051-M19402510004 560051-M19402510005 560051-M19402510016 560051-M19402510026 560051-M19402510032 560051-M19402510039 560051-M19402510013 560051-M19402510017 560051-M19402510031 560051-M19402510038 560051-M19402510035 560051-M19402510040 560051-M19402510010 560051-M19402510020 560051-M19402510023 560051-M19402510027 560051-M19402510036 560051-M19402510001 560051-M19402510002 560051-M19402510008 560051-M19402510009 560051-M19402510015 560051-M19402510033 560051-M19402510037 560051-M19606490001 560051-M19606490002 560051-M19610950005 560051-M19610950006 560051-M19610950013 560051-M19610950004 560051-M19610950009 560051-M19610950017 560051-M19610950011 560051-M19610950016 560051-M19610950018 560051-M19610950003 560051-M19610950015 560051-M19610950002 560051-M19610950007 560051-M19610950012 560051-M19610950014 560051-M19610950019 560051-M19610950010 560051-M19610950020 560051-M19610950001 560051-M19610950008 560051-M19809450004 560051-M19809450008 560051-M19809450014 560051-M19809450018 560051-M19809450002 560051-M19809450006 560051-M19809450010 560051-M19809450012 560051-M19809450016 560051-M19809450003 560051-M19809450005 560051-M19809450013 560051-M19809450020 560051-M19809450015 560051-M19809450007 560051-M19809450009 560051-M19809450017 560051-M19809450019 560051-M19809450001 560051-M19809450011 560051-M19812700002 560051-M19812700003 560051-M19812700001 560051-M19812700004 560051-M19812700005 560051-M19A01420001 560051-M19A01420002 560051-M19A01420003 560051-M19B06160002 560051-M19B06160008 560051-M19B06160010 560051-M19B06160020 560051-M19B06160003 560051-M19B06160013 560051-M19B06160015 560051-M19B06160006 560051-M19B06160009 560051-M19B06160001 560051-M19B06160011 560051-M19B06160012 560051-M19B06160018 560051-M19B06160005 560051-M19B06160014 560051-M19B06160017 560051-M19B06160016 560051-M19B06160019 560051-M19B06160004 560051-M19B06160007 560051-M19C02460001 560051-M19C02460002 560051-M19C02460003 560051-M19C00660001 560051-M19C00660004 560051-M19C00660018 560051-M19C00660002 560051-M19C00660006 560051-M19C00660011 560051-M19C00660003 560051-M19C00660010 560051-M19C00660008 560051-M19C00660014 560051-M19C00660012 560051-M19C00660016 560051-M19C00660013 560051-M19C00660015 560051-M19C00660017 560051-M19C00660019 560051-M19C00660020 560051-M19C00660005 560051-M19C00660007 560051-M19C00660009 560051-M19C16290003 560051-M19C16290017 560051-M19C16290004 560051-M19C16290008 560051-M19C16290010 560051-M19C16290019 560051-M19C16290001 560051-M19C16290002 560051-M19C16290006 560051-M19C16290011 560051-M19C16290015 560051-M19C16290005 560051-M19C16290007 560051-M19C16290013 560051-M19C16290020 560051-M19C16290009 560051-M19C16290012 560051-M19C16290014 560051-M19C16290016 560051-M19C16290018 560051-M20107550003 560051-M20107550004 560051-M20107550005 560051-M20107550001 560051-M20107550002 560051-M20211370001 560051-M20211370003 560051-M20211370002 560051-M20211370005 560051-M20211370004 560051-M20308620008 560051-M20308620007 560051-M20308620001 560051-M20308620006 560051-M20308620010 560051-M20308620002 560051-M20308620003 560051-M20308620004 560051-M20308620009 560051-M20308620005 560051-M20412810010 560051-M20412810006 560051-M20412810007 560051-M20412810008 560051-M20412810003 560051-M20412810004 560051-M20412810009 560051-M20412810002 560051-M20412810005 560051-M20412810001 560051-M20505800007 560051-M20505800010 560051-M20505800001 560051-M20505800002 560051-M20505800008 560051-M20505800009 560051-M20505800005 560051-M20505800006 560051-M20505800003 560051-M20505800004 560051-M20700790010 560051-M20700790003 560051-M20700790007 560051-M20700790001 560051-M20700790004 560051-M20700790009 560051-M20700790002 560051-M20700790005 560051-M20700790006 560051-M20700790008 560051-M20708490002 560051-M20708490001 560051-M20708490003 560051-M20800530001 560051-M20800530004 560051-M20800530005 560051-M20800530007 560051-M20800530008 560051-M20800530009 560051-M20800530002 560051-M20800530010 560051-M20800530003 560051-M20800530006 560051-M20807950002 560051-M20807950005 560051-M20807950010 560051-M20807950001 560051-M20807950008 560051-M20807950003 560051-M20807950007 560051-M20807950009 560051-M20807950006 560051-M20807950004 560051-M20904530002 560051-M20904530001 560051-M20904530005 560051-M20904530003 560051-M20904530004 560051-M20913700002 560051-M20913700006 560051-M20913700001 560051-M20913700003 560051-M20913700005 560051-M20913700010 560051-M20913700007 560051-M20913700008 560051-M20913700009 560051-M20913700004 560051-M20A04870004 560051-M20A04870002 560051-M20A04870005 560051-M20A04870003 560051-M20A04870001 560051-M20A09240002 560051-M20A09240013 560051-M20A09240014 560051-M20A09240015 560051-M20A09240007 560051-M20A09240011 560051-M20A09240012 560051-M20A09240004 560051-M20A09240005 560051-M20A09240010 560051-M20A09240006 560051-M20A09240001 560051-M20A09240008 560051-M20A09240009 560051-M20A09240003 560051-M20B09690001 560051-M20B09690005 560051-M20B09690008 560051-M20B09690003 560051-M20B09690004 560051-M20B09690006 560051-M20B09690002 560051-M20B09690007 560051-M20B12120007 560051-M20B12120013 560051-M20B12120001 560051-M20B12120004 560051-M20B12120010 560051-M20B12120011 560051-M20B12120014 560051-M20B12120005 560051-M20B12120006 560051-M20B12120012 560051-M20B12120015 560051-M20B12120002 560051-M20B12120009 560051-M20B12120003 560051-M20B12120008 560051-M20C14650004 560051-M20C14650005 560051-M20C14650001 560051-M20C14650002 560051-M20C14650003 560051-M20C12230003 560051-M20C12230007 560051-M20C12230009 560051-M20C12230010 560051-M20C12230006 560051-M20C12230002 560051-M20C12230004 560051-M20C12230005 560051-M20C12230008 560051-M20C12230001 560051-M21107600006 560051-M21107600009 560051-M21107600004 560051-M21107600005 560051-M21107600007 560051-M21107600008 560051-M21107600010 560051-M21107600001 560051-M21107600002 560051-M21107600003 560051-M21113970001 560051-M21113970002 560051-M21113970004 560051-M21113970003 560051-M21113880001 560051-M21113880004 560051-M21113880002 560051-M21113880003 560051-M21206930008 560051-M21206930015 560051-M21206930013 560051-M21206930014 560051-M21206930002 560051-M21206930001 560051-M21206930005 560051-M21206930007 560051-M21206930010 560051-M21206930003 560051-M21206930004 560051-M21206930006 560051-M21206930009 560051-M21206930011 560051-M21206930012 560051-M21307270006 560051-M21307270010 560051-M21307270001 560051-M21307270002 560051-M21307270005 560051-M21307270003 560051-M21307270007 560051-M21307270004 560051-M21307270008 560051-M21307270009 560051-M21302260018 560051-M21302260022 560051-M21302260030 560051-M21302260005 560051-M21302260009 560051-M21302260012 560051-M21302260015 560051-M21302260019 560051-M21302260023 560051-M21302260024 560051-M21302260026 560051-M21302260027 560051-M21302260008 560051-M21302260010 560051-M21302260011 560051-M21302260002 560051-M21302260028 560051-M21302260020 560051-M21302260021 560051-M21302260001 560051-M21302260013 560051-M21302260014 560051-M21302260016 560051-M21302260017 560051-M21302260003 560051-M21302260004 560051-M21302260006 560051-M21302260007 560051-M21302260025 560051-M21302260029 560051-M21501090018 560051-M21501090030 560051-M21501090010 560051-M21501090011 560051-M21501090001 560051-M21501090013 560051-M21501090015 560051-M21501090017 560051-M21501090019 560051-M21501090022 560051-M21501090023 560051-M21501090024 560051-M21501090025 560051-M21501090027 560051-M21501090002 560051-M21501090006 560051-M21501090026 560051-M21501090003 560051-M21501090004 560051-M21501090005 560051-M21501090007 560051-M21501090009 560051-M21501090012 560051-M21501090014 560051-M21501090016 560051-M21501090029 560051-M21501090008 560051-M21501090020 560051-M21501090021 560051-M21501090028 560051-M21607750020 560051-M21607750008 560051-M21607750018 560051-M21607750005 560051-M21607750006 560051-M21607750010 560051-M21607750016 560051-M21607750007 560051-M21607750009 560051-M21607750019 560051-M21607750002 560051-M21607750003 560051-M21607750012 560051-M21607750017 560051-M21607750015 560051-M21607750001 560051-M21607750011 560051-M21607750013 560051-M21607750004 560051-M21607750014 560051-M21713700008 560051-M21713700017 560051-M21713700018 560051-M21713700020 560051-M21713700002 560051-M21713700005 560051-M21713700015 560051-M21713700004 560051-M21713700010 560051-M21713700014 560051-M21713700006 560051-M21713700016 560051-M21713700001 560051-M21713700011 560051-M21713700009 560051-M21713700012 560051-M21713700019 560051-M21713700003 560051-M21713700007 560051-M21713700013 560051-M21815570003 560051-M21815570004 560051-M21815570008 560051-M21815570001 560051-M21815570010 560051-M21815570005 560051-M21815570002 560051-M21815570009 560051-M21815570006 560051-M21815570007 560051-M21A03270001 560051-M21A03270006 560051-M21A03270002 560051-M21A03270003 560051-M21A03270007 560051-M21A03270008 560051-M21A03270009 560051-M21A03270004 560051-M21A03270005 560051-M21A03270010 560051-M21B05410005 560051-M21B05410002 560051-M21B05410003 560051-M21B05410001 560051-M21B05410004 560051-M21B13600011 560051-M21B13600002 560051-M21B13600004 560051-M21B13600005 560051-M21B13600001 560051-M21B13600003 560051-M21B13600007 560051-M21B13600009 560051-M21B13600010 560051-M21B13600013 560051-M21B13600006 560051-M21B13600008 560051-M21B13600012 560051-M21B13600014 560051-M21B13600015 560051-M21C03010009 560051-M21C03010010 560051-M21C03010004 560051-M21C03010005 560051-M21C03010006 560051-M21C03010003 560051-M21C03010001 560051-M21C03010002 560051-M21C03010008 560051-M21C03010007 560051-M22104890003 560051-M22104890004 560051-M22104890005 560051-M22104890001 560051-M22104890002 560051-M22104890011 560051-M22104890012 560051-M22104890006 560051-M22104890010 560051-M22104890007 560051-M22104890008 560051-M22104890009 560051-M22104890015 560051-M22104890013 560051-M22104890014 560051-M22210000002 560051-M22210000010 560051-M22210000013 560051-M22210000012 560051-M22210000015 560051-M22210000001 560051-M22210000009 560051-M22210000014 560051-M22210000008 560051-M22210000003 560051-M22300730002 560051-M22300730007 560051-M22213190009 560051-M22210000004 560051-M22300730009 560051-M22300730003 560051-M22300730004 560051-M22300730006 560051-M22213190014 560051-M22213190022 560051-M22213190023 560051-M22213190025 560051-M22213190004 560051-M22213190012 560051-M22213190013 560051-M22213190015 560051-M22300730005 560051-M22213190002 560051-M22213190003 560051-M22213190005 560051-M22213190017 560051-M22213190020 560051-M22213190021 560051-M22213190006 560051-M22213190007 560051-M22213190010 560051-M22213190011 560051-M22213190018 560051-M22210000006 560051-M22300730010 560051-M22213190001 560051-M22213190019 560051-M22210000007 560051-M22210000011 560051-M22300730001 560051-M22213190008 560051-M22213190016 560051-M22213190024 560051-M22210000005 560051-M22300730008 560051-M22314370007 560051-M22314370009 560051-M22314370005 560051-M22314370002 560051-M22314370010 560051-M22314370001 560051-M22314370004 560051-M22314370006 560051-M22314370008 560051-M22314370003 560051-M22406120010 560051-M22406120001 560051-M22406120005 560051-M22406120007 560051-M22406120002 560051-M22406120003 560051-M22406120004 560051-M22406120006 560051-M22406120008 560051-M22406120009 560051-M22404660008 560051-M22404660005 560051-M22404660002 560051-M22404660004 560051-M22404660006 560051-M22404660001 560051-M22404660003 560051-M22404660009 560051-M22404660010 560051-M22404660013 560051-M22404660007 560051-M22404660012 560051-M22404660014 560051-M22404660011 560051-M22404660015 560051-M22510390007 560051-M22510390009 560051-M22510390014 560051-M22510390022 560051-M22510390003 560051-M22510390005 560051-M22510390008 560051-M22510390011 560051-M22510390021 560051-M22510390024 560051-M22510390002 560051-M22510390016 560051-M22510390026 560051-M22510390030 560051-M22510390001 560051-M22510390004 560051-M22510390018 560051-M22510390019 560051-M22510390027 560051-M22510390029 560051-M22510390006 560051-M22510390010 560051-M22510390012 560051-M22510390013 560051-M22510390015 560051-M22510390020 560051-M22510390023 560051-M22510390025 560051-M22510390028 560051-M22510390017 560051-M22607660001 560051-M22607660002 560051-M22607660006 560051-M22607660009 560051-M22607660004 560051-M22607660005 560051-M22607660010 560051-M22607660003 560051-M22607660008 560051-M22607660007 560051-M22615460003 560051-M22615460012 560051-M22615460016 560051-M22615460018 560051-M22615460007 560051-M22615460004 560051-M22615460015 560051-M22615460019 560051-M22615460006 560051-M22615460017 560051-M22615460001 560051-M22615460013 560051-M22615460005 560051-M22615460009 560051-M22615460010 560051-M22615460011 560051-M22615460020 560051-M22615460002 560051-M22615460008 560051-M22615460014 560051-M22806940002 560051-M22806940012 560051-M22806940004 560051-M22806940014 560051-M22806940011 560051-M22806940001 560051-M22806940008 560051-M22806940010 560051-M22806940006 560051-M22806940009 560051-M22806940013 560051-M22806940007 560051-M22806940003 560051-M22806940005 560051-M22806940015 560051-M22917260002 560051-M22917260003 560051-M22917260010 560051-M22917260011 560051-M22900490007 560051-M22917260004 560051-M22900490008 560051-M22917260006 560051-M22917260012 560051-M22917260005 560051-M22917260001 560051-M22917260008 560051-M22900490003 560051-M22900490010 560051-M22917260014 560051-M22917260009 560051-M22900490004 560051-M22917260013 560051-M22917260015 560051-M22900490002 560051-M22900490005 560051-M22900490006 560051-M22917260007 560051-M22900490001 560051-M22900490009 560051-M22B00690005 560051-M22B00690002 560051-M22B00690001 560051-M22B00690007 560051-M22B00690006 560051-M22B00690004 560051-M22B00690012 560051-M22B00690010 560051-M22B00690008 560051-M22B00690009 560051-M22B00690013 560051-M22B00690014 560051-M22B00690015 560051-M22B00690003 560051-M22B00690011 560051-M22B16450010 560051-M22B16450001 560051-M22B16450002 560051-M22B16450004 560051-M22B16450003 560051-M22B16450005 560051-M22B16450007 560051-M22B16450006 560051-M22B16450008 560051-M22B16450009 560051-M22C11660006 560051-M22C11660003 560051-M22C11660004 560051-M22C11660005 560051-M22C11660011 560051-M22C11660012 560051-M22C11660007 560051-M22C11660015 560051-M22C11660013 560051-M22C11660014 560051-M22C11660001 560051-M22C11660002 560051-M22C11660008 560051-M22C11660009 560051-M22C11660010 560051-M22C06180007 560051-M22C06180009 560051-M22C06180014 560051-M22C06180024 560051-M22C06180001 560051-M22C06180013 560051-M22C06180004 560051-M22C06180006 560051-M22C06180018 560051-M22C06180021 560051-M22C06180005 560051-M22C06180010 560051-M22C06180019 560051-M22C06180020 560051-M22C06180003 560051-M22C06180011 560051-M22C06180015 560051-M22C06180025 560051-M22C06180002 560051-M22C06180008 560051-M22C06180016 560051-M22C06180023 560051-M22C06180012 560051-M22C06180017 560051-M22C06180022 560051-M23205340006 560051-M23205340010 560051-M23205340007 560051-M23205340008 560051-M23205340009 560051-M23205340001 560051-M23205340002 560051-M23205340003 560051-M23205340004 560051-M23205340005 560051-M23603980005 560051-M23603980020 560051-M23603980002 560051-M23603980003 560051-M23603980004 560051-M23603980008 560051-M23603980009 560051-M23603980010 560051-M23603980001 560051-M23603980007 560051-M23603980025 560051-M23603980013 560051-M23603980014 560051-M23603980021 560051-M23603980016 560051-M23603980011 560051-M23603980015 560051-M23603980017 560051-M23603980022 560051-M23603980023 560051-M23603980024 560051-M23603980006 560051-M23603980012 560051-M23603980018 560051-M23603980019 560051-M23711800014 560051-M23711800015 560051-M23711800004 560051-M23711800005 560051-M23711800011 560051-M23711800012 560051-M23711800019 560051-M23711800016 560051-M23711800017 560051-M23711800018 560051-M23711800003 560051-M23711800001 560051-M23711800002 560051-M23711800006 560051-M23711800007 560051-M23711800008 560051-M23711800009 560051-M23711800010 560051-M23711800013 560051-M23711800020 560051-M23812200003 560051-M23812200005 560051-M23812200013 560051-M23812200007 560051-M23812200014 560051-M23812200001 560051-M23812200004 560051-M23812200006 560051-M23812200011 560051-M23812200010 560051-M23812200008 560051-M23812200015 560051-M23812200012 560051-M23812200002 560051-M23812200009 560051-M23914940005 560051-M23914940004 560051-M23914940007 560051-M23914940008 560051-M23914940009 560051-M23914940010 560051-M23914940006 560051-M23914940001 560051-M23914940002 560051-M23914940003 560051-M23B16930010 560051-M23B16930002 560051-M23B16930009 560051-M23B16930001 560051-M23B16930005 560051-M23B16930007 560051-M23B16930008 560051-M23B16930003 560051-M23B16930004 560051-M23B16930006 560051-M23C14980005 560051-M23C14980009 560051-M23C14980002 560051-M23C14980008 560051-M23C14980004 560051-M23C14980006 560051-M23C14980010 560051-M23C14980001 560051-M23C14980003 560051-M23C14980007 560051-K24100040002 560051-K24100040001 560051-K24100040003 560051-K24100040006 560051-K24100040008 560051-K24100040010 560051-K24100040007 560051-K24100040004 560051-K24100040005 560051-K24100040009 560051-M24410930004 560051-M24410930010 560051-M24410930003 560051-M24410930005 560051-M24410930006 560051-M24410930002 560051-M24410930008 560051-M24410930001 560051-M24410930007 560051-M24410930009 560051-M24415470001 560051-M24502790013 560051-M24502790003 560051-M24502790014 560051-M24502790015 560051-M24502790010 560051-M24502790002 560051-M24502790006 560051-M24502790001 560051-M24502790007 560051-M24502790008 560051-M24502790004 560051-M24502790005 560051-M24502790009 560051-M24502790011 560051-M24502790012 560051-K24500200012 560051-K24500200018 560051-K24500200022 560051-K24500200011 560051-K24500200021 560051-K24500200001 560051-K24500200014 560051-K24500200015 560051-K24500200003 560051-K24500200006 560051-K24500200007 560051-K24500200002 560051-K24500200008 560051-K24500200010 560051-K24500200016 560051-K24500200019 560051-K24500200024 560051-K24500200025 560051-K24500200004 560051-K24500200005 560051-K24500200009 560051-K24500200013 560051-K24500200017 560051-K24500200020 560051-K24500200023 560051-M24600970007 560051-M24600970003 560051-M24600970002 560051-M24600970004 560051-M24600970005 560051-M24600970006 560051-M24600970001 560051-K24600260007 560051-K24600260014 560051-K24600260015 560051-M24607070004 560051-K24600260004 560051-K24600260005 560051-M24607070005 560051-K24600260001 560051-K24600260002 560051-K24600260008 560051-K24600260009 560051-K24600260013 560051-M24607070002 560051-K24600260010 560051-M24607070001 560051-M24607070003 560051-K24600260011 560051-K24600260012 560051-K24600260006 560051-K24600260003 560051-M24616380001 560051-M24620270002 560051-M24620270008 560051-M24620270004 560051-M24620270006 560051-M24620270003 560051-M24620270005 560051-M24620270001 560051-M24620270007 560051-M24706880005 560051-M24706880010 560051-M24706880001 560051-M24706880006 560051-M24706880015 560051-M24706880003 560051-M24706880012 560051-M24706880002 560051-M24706880008 560051-M24706880014 560051-M24706880009 560051-M24706880011 560051-M24706880013 560051-M24706880004 560051-M24706880007 560051-M24715650004 560051-M24715650006 560051-M24715650014 560051-M24715650001 560051-M24715650010 560051-M24715650009 560051-M24715650013 560051-M24715650008 560051-M24715650011 560051-M24715650002 560051-M24715650003 560051-M24715650005 560051-M24715650007 560051-M24715650012 560051-M24715650015 560051-M24815220007 560051-M24815220010 560051-M24815220001 560051-M24815220003 560051-M24815220009 560051-M24815220004 560051-M24815220014 560051-M24815220002 560051-M24815220006 560051-M24815220008 560051-M24815220012 560051-M24815220005 560051-M24815220011 560051-M24815220013 560051-M24815220015 560051-M24A07570019 560051-M24A07570012 560051-M24A07570017 560051-M24A07570006 560051-M24A07570001 560051-M24A07570003 560051-M24A07570002 560051-M24A07570005 560051-M24A07570010 560051-M24A07570016 560051-M24A07570007 560051-M24A07570009 560051-M24A07570014 560051-M24A07570018 560051-M24A07570020 560051-M24A07570004 560051-M24A07570008 560051-M24A07570011 560051-M24A07570013 560051-M24A07570015 560051-M24B07480005 560051-M24B07480009 560051-M24B07480002 560051-M24B07480004 560051-M24B07480008 560051-M24B07480001 560051-M24B07480006 560051-M24B07480007 560051-M24B07480010 560051-M24B07480003 560051-M24A04910001 560051-M24A04910002 560051-M24A17720010 560051-M24A17720016 560051-M24A17720018 560051-M24A17720019 560051-M24A17720023 560051-M24A17720030 560051-M24A17720036 560051-M24A17720038 560051-M24A17720039 560051-M24A17720042 560051-M24A17720047 560051-M24A17720056 560051-M24A17720058 560051-M24A17720059 560051-M24A17720007 560051-M24A17720035 560051-M24A17720055 560051-M24B18920010 560051-M24A17720024 560051-M24A17720033 560051-M24A17720034 560051-M24A17720046 560051-M24A17720048 560051-M24A17720049 560051-M24A17720053 560051-M24A17720054 560051-M24B18920009 560051-M24A17720009 560051-M24A17720041 560051-M24B18920002 560051-M24B18920003 560051-M24B18920004 560051-M24B18920006 560051-M24B18920007 560051-M24B18920008 560051-M24A17720020 560051-M24A17720026 560051-M24A17720028 560051-M24A17720045 560051-M24A17720050 560051-M24B18920001 560051-M24A17720005 560051-M24A17720013 560051-M24A17720021 560051-M24A17720022 560051-M24A17720027 560051-M24A17720052 560051-M24A17720008 560051-M24A17720011 560051-M24A17720012 560051-M24A17720017 560051-M24A17720029 560051-M24A17720031 560051-M24A17720032 560051-M24A17720037 560051-M24A17720043 560051-M24A17720044 560051-M24A17720051 560051-M24A17720057 560051-M24A17720060 560051-M24A17720025 560051-M24A17720040 560051-M24B18920005 560051-M24A17720004 560051-M24A17720006 560051-M24A17720014 560051-M24A17720063 560051-M24A17720003 560051-M24A17720001 560051-M24A17720002 560051-M24A17720015 560051-M24A17720061 560051-M24A17720062 560051-M24C18240003 560051-M24C18240005 560051-M24C18240011 560051-M24C18240018 560051-M24C18240020 560051-M24C18240021 560051-M24C18240015 560051-M24C18240006 560051-M24C18240012 560051-M24C18240009 560051-M24C18240014 560051-M24C18240016 560051-M24C18240024 560051-M24C18240002 560051-M24C18240010 560051-M24C18240001 560051-M24C18240008 560051-M24C18240013 560051-M24C18240023 560051-M24C18240025 560051-M24C18240007 560051-M24C18240022 560051-M24C18240004 560051-M24C18240017 560051-M24C18240019 560051-M25200890001 560051-M25200890002 560051-M25200890004 560051-M25200890003 560051-M25200890005 560051-M25300770002 560051-M25300770001 560051-M25300770003 560051-M25300770005 560051-M25300770004 560051-M25216200004 560051-M25216200006 560051-M25216200008 560051-M25216200005 560051-M25216200009 560051-M25216200010 560051-M25216200011 560051-M25216200024 560051-M25216200002 560051-M25216200021 560051-M25216200023 560051-M25216200013 560051-M25216200017 560051-M25216200001 560051-M25216200003 560051-M25216200007 560051-M25216200015 560051-M25216200019 560051-M25216200022 560051-M25216200014 560051-M25216200016 560051-M25216200018 560051-M25923530001 560051-M25923530005 560051-M25923530010 560051-M25923530011 560051-M25923530016 560051-M25923530020 560051-M25923530003 560051-M25923530007 560051-M25923530012 560051-M25923530019 560051-M25923530002 560051-M25923530009 560051-M25923530015 560051-M25923530013 560051-M25923530017 560051-M25923530004 560051-M25923530006 560051-M25923530008 560051-M25923530014 560051-M25923530018
FEI Number 3009499478
Recalling Firm/
Manufacturer
Edan Diagnostics
9833 Pacific Heights Blvd
San Diego CA 92121-4707
For Additional Information ContactChao Lin
858-7503066
Manufacturer Reason
for Recall
As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential cybersecurity issues with various of the firm device.
FDA Determined
Cause 2
Device Design
ActionOn 10/24/2025, the firm emailed notifying customers of a product recall. On 11/06/2025, the firm emailed an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them of labeling issues involving cybersecurity which arose from a communication from the U.S. Food & Drug Administration. -First issue involves an inappropriate publicly routable IP address that was used as an example in the user manual for demonstrating user-configurable IP settings and Second issue involves an inappropriate publicly IP address that was used as the factory default configuration. Customer are instructed to: 1. Immediately examine their inventory and quarantine product subject to correction. 2. If customers may have further distributed affected product, identify their customers and notify them at once of this product correction. The notification to their customers may be enhanced by including the enclosed URGENT MEDICAL DEVICE CORRECTION. 3. For affected product still in their inventory, place copy of the enclosed Notice Before Use and Corrigenda of User Manual into its package before distribution. 4. For affected product that has been distributed or is in use: (1) Confirm that they have received and acknowledged Corrigenda of User Manual. Print out a copy, attach to the User Manual. (2) Confirm the publicly routable IP address 202.114.4.* is NOT used in the User Maintain in the device. If 202.114.4.* is being used, change to appropriate IP settings of device in the Network Setup according to Instruction Instructions for IP Settings Configuration. 5. Complete the attached Acknowledgement and Receipt Form, sign it, and return a scanned copy to the sender of this notification by email. For questions - Contact: EDAN Instruments, Inc. Monday to Friday 09:00-17:00 (UTC+8) Tel:+86-0755-26898326 EDAN Diagnostics, Inc. Monday to Friday 09:00-17:00 (PST) Tel: 1-(858)-750-3066
Quantity in Commerce3721
DistributionWorldwide - U.S. Nationwide distribution in the states of CA, KY, PR, FL, CT, IL, PA, RI, NC, NY, LA, TX, MI and the country of Mexico.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = HGM
-
-