Class 2 Device Recall Philips Azurion system

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Philips Azurion systemsee related information
Date Initiated by FirmDecember 12, 2025
Date PostedJanuary 16, 2026
Recall Status1 Open3, Classified
Recall NumberZ-1113-2026
Recall Event ID 98108
510(K)NumberK200917 
Product Classification Interventional fluoroscopic x-ray system - Product Code OWB
ProductPhilips Azurion system; Software Version Number: R3.1;
Code Information Software Version Number: R3.1; (1) Model Number: 722221; System Product Name: Azurion 3 M12; UDI-DI: 00884838099203; Serial Numbers: (2) Model Number: 722222; System Product Name: Azurion 3 M15; UDI-DI: 00884838099210; Serial Numbers: (3) Model Number: 722223; System Product Name: Azurion 7 M12; UDI-DI: 00884838099241; Serial Numbers: (4) Model Number: 722224; System Product Name: Azurion 7 M20; UDI-DI: 00884838099258; Serial Numbers: (5) Model Number: 722225; System Product Name: Azurion 7 B12; UDI-DI: 00884838099265; Serial Numbers: (6) Model Number: 722226; System Product Name: Azurion 7 B20; UDI-DI: 00884838099272; Serial Numbers: (7) Model Number: 722227; System Product Name: Azurion 5 M12; UDI-DI: 00884838099227; Serial Numbers: (8) Model Number: 722228; System Product Name: Azurion 5 M20; UDI-DI: 00884838099234; Serial Numbers: (9) Model Number: 722229; System Product Name: Azurion 3 M12; UDI-DI: 00884838116726; Serial Numbers: 5, 9, 32, 40, 35, 31, 33; (10) Model Number: 722230; System Product Name: Azurion 3 M15; UDI-DI: 00884838116733; Serial Numbers: 25, 54, 50, 52, 55, 30, 63, 60, 67; (11) Model Number: 722231; System Product Name: Azurion 5 M12; UDI-DI: 00884838116740; Serial Numbers: 62, 7, 8, 14, 17, 63, 6, 66, 59, 69, 56, 67, 48, 68, 80, 93, 87, 94, 81, 92, 18; (12) Model Number: 722232; System Product Name: Azurion 5 M20; UDI-DI: 00884838116757; Serial Numbers: 97, 39, 31, 100, 104, 93, 115, 113, 118, 119, 89, 54, 67, 7, 128, 86, 110, 132, 130, 131, 123, 137, 139, 117, 260; (13) Model Number: 722233; System Product Name: Azurion 7 M12; UDI-DI: 00884838116764; Serial Numbers: 239, 100, 233, 240, 176, 227, 223, 130, 161, 194, 243, 56, 193, 280, 75, 253, 156, 288, 72, 281, 304, 300, 105, 295, 312, 310, 285, 292, 290, 298, 297; (14) Model Number: 722234; System Product Name: Azurion 7 M20; UDI-DI: 00884838116771; Serial Numbers: 731, 727, 261, 473, 683, 722, 562, 750, 745, 766, 265, 701, 400, 310, 421, 314, 182, 579, 96, 763, 272, 612, 15, 844, 105, 702, 751, 384, 851, 869, 743, 337, 651, 455, 706, 312, 563, 709, 764, 871, 846, 278, 713, 802, 868, 277, 742, 184, 819, 759, 910, 928, 415, 414, 945, 961, 355, 933, 692, 920, 569, 939, 909, 553, 837, 856, 571, 756, 941, 570, 761, 981, 353, 888, 964, 820, 746, 736, 876, 599, 1017, 908, 487, 983, 495; (15) Model Number: 722235; System Product Name: Azurion 7 B12; UDI-DI: 00884838116788; Serial Numbers: 96, 79, 111, 100, 101, 38, 27, 92, 78, 118, 48, 131, 135, 22, 125, 103, 138, 139, 137; (16) Model Number: 722236; System Product Name: Azurion 7 B20; UDI-DI: 00884838116801; Serial Numbers: 241, 297, 198, 215, 305, 274, 299, 78, 279, 287, 318, 353, 324, 331, 293, 245, 314, 310, 317, 312, 77, 358, 206, 383, 380, 370, 377, 385, 405, 379, 328;
FEI Number 3002807880
Recalling Firm/
Manufacturer
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Veenpluis 4-6
Best Netherlands
Manufacturer Reason
for Recall
Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.
FDA Determined
Cause 2
Software design
ActionOn December 12, 2025, URGENT Medical Device Correction letters were sent to customers. Actions to be taken: - Circulate this Urgent Medical Device Correction letter to all users of the system so that they are aware of the issues. " In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. " Keep this Urgent Medical Device Correction letter with the documentation of the system until Philips corrects your system. Ensure that the letter is in a place likely to be seen/viewed. " Complete and return the response form included in this Urgent Medical Device Correction letter to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction letter and understanding of the issues and required actions to be taken. " If you experience an issue described in this letter, please report it to your local Philips representative. Actions planned by Philips IGT Systems to correct the issues Philips will address the identified issues by implementing software updates in all affected systems. Philips expects to have software update 3.1.5 (FCO72200635, FC072200671), that resolves issues 1, released by Q1 2026 (subject to regulatory clearance). The Longitudinal Position Error (issue 2) will not be resolved with software update 3.1.5. The solution for this issue is planned for Q4 2026 (subject to regulatory clearance) with software update 3.1.15 (FCO72200684). Your local Philips representative will contact you to schedule visits to install the software updates once available.
Quantity in Commerce228 units (all OUS)
DistributionInternational distribution in the countries of AE, AR, AT, AZ, BE, BG, BH, BR, BS, CH, CL, CO, CR, CY, CZ, DE, DK, DZ, EC, EG, ES, GB, GR, GT, ID, IE, IL, IN, IQ, IT, JE, JP, LB, MA, MO, NI, NL, NO, PE, PL, RO, SA, SE, SV, TR, UA, VE, VN, ZA;
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = OWB
-
-