• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Medline

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medlinesee related information
Date Initiated by FirmNovember 26, 2025
Date PostedJanuary 12, 2026
Recall Status1 Open3, Classified
Recall NumberZ-1059-2026
Recall Event ID 98154
Product Classification Neurological tray - Product Code OJG
ProductMedline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: NEURO ANGIO PACK, Medline kit SKU DYNJ38935
Code Information UDI/DI 10884389725092 (EA) 40884389725093 (CS), Lot number 25IBC144
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries, LP
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactHaley Barclay
866-359-1704
Manufacturer Reason
for Recall
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionMedline issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 11/26/2025 via email and USPS first class mail. The notice explained the issue potential risk to the patient, and referred the consignee to the attached B Braun notice for further instructions. The B Braun notice provide "Interim Measures for Users" to provide for the continued use of the product. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity in Commerce12 units
DistributionUS Nationwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-