| | Class 2 Device Recall SIGMA Spectrum Infusion Pump |  |
| Date Initiated by Firm | November 28, 2025 |
| Date Posted | January 16, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1121-2026 |
| Recall Event ID |
98167 |
| 510(K)Number | K251640 |
| Product Classification |
Pump, infusion - Product Code FRN
|
| Product | SIGMA Spectrum Infusion Pump, Model/Catalog Number: 35700BAX2 |
| Code Information |
GTIN 00085412498683;
Serial Numbers: 2119054
2135735
2132190
2154666
2154656
2096922
2107882
2115820
2116236
2123750
2124467
2142165
2145693
2146570
2135601
2150159
2019614
2030571
2064474
2065875
2007842
2077077
2110000
2139163
2152387
2090011
2150576
2006741
2090877
2091360
2115605
2136756
2145367
2145911
2144366
2147107
2005818
2036444
2040404
2064488
2054317
2064576
2065380
2064711
2055127
2066486
2065852
2065982
2041955
2081026
2023286
2088106
2089908
2089953
2090012
2096271
2106362
2114760
2114275
2116368
2116054
2117183
2118056
2112866
2136545
2139711
2142076
2161326
2112386
2127356
2161111
2003644
2042815
2089901
2026384
2157334
2004602
2093902
2134816
2155520
2156545
2107252
2136765
2152712
2027178
2135072
2081009
2145203
2156837
2124901
2077354
2077117
2122684
2118643
2106472
2145820
2009374
2115158
2090446
2125551
2150237
2004798
2013799
2023938
2092704
2148207
2096246
2007553
2014364
2089297
2089770
2089790
2119778
2105983
2088857
2140961
2119801
2075390
2013353
2014754
2115703
2105569
2106297
2108307
2107928
2107024
2094331
2010527
2108679
2133529
2133609
2091040
2105718
2138210
2106084
2062517
2022721
2106575
2106357
2106667
2026959
2105921
2098876
2106666
2104788
2106441
2020080
2106646
2106571
2020358
2102437
2106458
2106975
2104537
2026270
2021619
2104636
2105332
2106348
2106213
2019141
2023340
2106243
2100062
2106694
2106468
2106652
2106644
2106687
2104780
2106781
2105011
2105920
2105399
2106661
2106657
2104225
2101214
2020660
2106401
2014906
2104935
2105761
2106381
2106706
2105316
2105830
2102959
2106658
2105682
2128181
2108278
2152865
2139013
2109590
2106397
2105502
2105517
2089502
2085260
2055930
2062659
2031944
2115307
2125787
2151876
2152606
2126978
2163477
2140825
2115351
2037060
2121404
2125165
2093359
2021479
2136589
2156064
2126079
2125854
2011851
2021129
2144675
2127091
2133796
2110399
2106566
2096065
2096542
2127985
2105537
2110489
2106110
2105686
2106565
2105353
2106522
2105857
2111300
2112189
2115454
2102374
2114689
2112238
2106823
2158498
2119695
2143990
2027436
2081647
2090142
2081766
2132604
2087392
2153913
2004327
2064121
2058398
2150299
2047477
2106466
2103333
2019580
2106117
2106656
2106067
2106564
2106420
2104885
2023160
2106359
2010593
2106374
2105913
2105545
2108691
2099348
2106681
2106037
2094532
2108580
2108975
2107820
2108302
2107139
2108206
2107786
2100031
2103850
2105121
2102524
2014666
2106748
2106462
2025115
2119762
2106454
2111211
2111618
2097976
2134941
2151686
2151046
2151529
2161273
2093135
2067637
2106771
2106614
2106346
2013293
2104641
2105103
2023265
2137011
2152466
2110351
2094316
2142410
2125118
2125085
2115234
2152576
2150286
2125804
2152646
2145056
2108479
2090054
2090316
2112969
2104951
2108944
2125978
2011314
2020697
2012966
2034912
2093055
2091740
2093133
2093092
2090317
2112541
2119848
2124307
2156543
2115819
2108948
2068764
2145666
2110655
2011614
2092341
2090701
2015546
2103211
2146805
2122281
2122218
2135223
2155758
2102590
2137864
2141435
2153759
2139827
2108327
2106034
2103539
2105952
2108585
2139489
2093034
2023785
2092572
2092354
2070160
2093511
2147125
2013737
2013659
2084512
2009532
2068987
2109945
2002918
2103807
2068705
2138965
2140482
2089287
2090523
2045838
2090315
2100132
2109849
2104914
2106514
2077283
2152864
2108485
2082000
2009588
2108918
2107006
2108867
2108294
2100633
2107504
2109006
2127015
2028322
2002118
2000826
2106388
2092019
2089942
2107986
2108369
2108019
2108688
2108811
2102241
2094972
2124776
2002102
2134188
2150087
2162168
2004449
2126814
2093171
2025696
2095150
2138751
2026387
2014190
2021933
2025864
2115704
2106410
2019020
2134722
2111304
2089811
2089981
2090078
2089744
2106019
2089179
2089812
2089616
2089384
2134543
2089473
2089864
2089877
2097144
2090760
2085331
2010134
2013108
2013774
2125890
2119667
2112393
2125523
2125871
2127049
2126139
2014015
2005033
2134062
2134353
2050981
2036573
2089880
2090176
2089602
2089828
2152302
2126042
2072755
2158186
2155422
2025404
2069258
2073635
2153030
2156830
2031662
2091629
2093090
2089628
2133878
2089746
2106085
2106759
2122820
2107251
2089873
2089693
2112799
2089872
2084362
2104952
2106505
2117825
2092700
2101923
2017574
2042947
2110079
2147263
2147574
2081129
2139273
2095033
2110320
2093751
2009829
2124379
2135047
2138291
2135425
2133258
2138885
2135929
2154930
2014167
2155581
2161548
2156157
2123007
2106120
2037298
2038167
2107080
2108841
2107636
2091940
2023865
2106347
2106680
2023936
2108240
2112587
2110496
2113253
2112819
2094431
2137887
2096352
2069112
2108202
2145334
2034454
2125684
2004185
2092220
2125289
2148111
2150145
2108600
2103033
2038168
2003675
2110072
2093195
2137800
2105521
2158615
2156191
2148254
2133833
2125655
2146547
2138868
2143606
2150140
2125142
2002778
2019010
2091261 |
Recalling Firm/ Manufacturer |
Baxter Healthcare Corporation 1 Baxter Pkwy Deerfield IL 60015-4625
|
| For Additional Information Contact | Center for One Baxter 1-847-9484770 |
Manufacturer Reason for Recall | Certain pumps have potentially been released from service with defective grease applied to the cam and motor gears. Due to its low viscosity, the defective grease may lead to the device having insufficient lubrication on the cam, which could lead to premature wear of the mechanism assembly, resulting in over-infusion or a free-flow situation. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | An URGENT MEDICAL DEVICE RECALL notification letter dated 11/28/25 was sent to customers.
Actions to be Taken by Customers
1. Immediately locate, isolate, and cease all use of the impacted product (see Attachment A for complete list of affected serial numbers). The product code and serial number can be found on the bottom of the infusion pump.
2. If you received this communication directly from Baxter, please acknowledge receipt of this notification by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory.
Acknowledging receipt will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from Accenture LLP on behalf of Baxter to confirm your receipt of this notification.
3. If you purchased this product from a distributor or wholesaler, please contact them to arrange for return of the affected product. Please note that responding to Baxter is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions.
4. Please forward a copy of this communication to the director of nursing, director of pharmacy, director of purchasing, facility risk manager, chief of medicine, medical director, office manager, and any other departments within your institution who use the affected product.
5. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please conduct a user-level recall of the affected product that you distributed to customers and check the associated box on the customer portal.
For general questions regarding this communication, or if you experience quality problems, please contact your Baxter sales representative, or Baxter Global Technical Services at 800-843-7867 (select option 1, then option 2) Monday through Friday, between 7:00 am and 7:00 pm Eastern Time. |
| Quantity in Commerce | 585 units |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = FRN
|
|
|
|