• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall EDAN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall EDANsee related information
Date Initiated by FirmOctober 24, 2025
Date PostedJanuary 22, 2026
Recall Status1 Open3, Classified
Recall NumberZ-1148-2026
Recall Event ID 97879
510(K)NumberK192514 
Product Classification Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
ProductBrand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: X8, X10, X12 Product Description: The X8 X10 X12 Patient Monitor (hereinafter called X series) can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormity so that doctors and nurses can deal with them in time. Component: No
Code Information EDAN Model: X8 UDI-DI code: 06944413810907 Serial/Lot Numbers: 261586-M21116620010 261586-M21116620002 261586-M21116620007 261586-M21116620006 261586-M21116620009 261586-M21116630001 261586-M21116620008 261586-M21116620003 261586-M21116630005 261586-M21116630003 261586-M21116620004 261586-M21116620001 261586-M21116630004 261586-M21116630002 261586-M21116620005 261586-M21301970003 261586-M21301970004 261586-M21301960001 261586-M21301970005 261586-M21301960003 261586-M21301970002 261586-M21301970001 261586-M21301960002 261586-M21714500002 261586-M21714500001 261586-M21714500003 261586-M21816750005 261586-M21816750003 261586-M21816750002 261586-M21816750004 261586-M21816750001 261586-M22105750001 261586-M22105750007 261586-M22105750008 261586-M22105750010 261586-M22105750002 261586-M22105750005 261586-M22105750009 261586-M22105750006 261586-M22105750003 261586-M22105750004 261586-M22405430007 261586-M22405430001 261586-M22405430004 261586-M22405430010 261586-M22405430006 261586-M22405430009 261586-M22405430008 261586-M22405430003 261586-M22405430002 261586-M22405430005 261586-M22615270004 261586-M22615270005 261586-M22615270001 261586-M22615270002 261586-M22615270003 261586-M23226950003 261586-M23226950002 261586-M23226950004 261586-M23226950005 261586-M23226950001 261586-M23B18320003 261586-M23B18320004 261586-M23B18330001 261586-M23B18330002 261586-M23B18330003 261586-M23B18330004 261586-M23B18320001 261586-M23B18320005 261586-M23B18330005 261586-M23B18320002 261586-M23C15910003 261586-M23C15920001 261586-M23C15920005 261586-M23C15920004 261586-M23C15920009 261586-M23C15910002 261586-M23C15920010 261586-M23C15920002 261586-M23C15910001 261586-M23C15910005 261586-M23C15920006 261586-M23C15920003 261586-M23C15910004 261586-M23C15920008 261586-M23C15920007 261586-M24104080002 261586-M24104080010 261586-M24104080003 261586-M24104080004 261586-M24104080006 261586-M24104080001 261586-M24104080005 261586-M24104080008 261586-M24104080009 261586-M24104080007 261586-M20611590012 261586-M20611590007 261586-M20611570016 261586-M20611590017 261586-M20611590018 261586-M20611580019 261586-M20611590004 261586-M20611590006 261586-M20611580022 261586-M20611590011 261586-M20611580002 261586-M20611580007 261586-M20611580017 261586-M20611580009 261586-M20611590009 261586-M20611590008 261586-M20611590019 261586-M20611590003 261586-M20611570006 261586-M20611580024 261586-M20611560003 261586-M20611570017 261586-M20611570015 261586-M20611560017 261586-M20611560007 261586-M20611560025 261586-M20611560020 261586-M20611560014 261586-M20611560013 261586-M20611560004 261586-M20611560022 261586-M20611560021 261586-M20611560018 261586-M20611560009 261586-M20611580012 261586-M20611560008 261586-M20611560011 261586-M20611560010 261586-M20611560002 261586-M20611560001 261586-M20611560019 261586-M20611560015 261586-M20611560006 261586-M20611560005 261586-M20611560024 261586-M20611560023 261586-M20611560012 261586-M20611590005 261586-M20611580015 261586-M20611580025 261586-M20611560016 261586-M20611580014 261586-M20611570002 261586-M20611580004 261586-M20611580005 261586-M20611590010 261586-M20611590013 261586-M20611590014 261586-M20611590015 261586-M20611590016 261586-M20611570001 261586-M20611570005 261586-M20611570007 261586-M20611570009 261586-M20611570019 261586-M20611580003 261586-M20611580013 261586-M20611580018 261586-M20611570010 261586-M20611570012 261586-M20611570018 261586-M20611570020 261586-M20611580006 261586-M20611580016 261586-M20611580023 261586-M20611590002 261586-M20611590001 261586-M20611590020 261586-M20611580020 261586-M20611580021 261586-M20611570004 261586-M20611580001 261586-M20611580010 261586-M20611570008 261586-M20611570003 261586-M20611570011 261586-M20611570013 261586-M20611580008 261586-M20611580011 261586-M20611570014 261586-M21313600009 261586-M21313600020 261586-M21313600017 261586-M21313600013 261586-M21313600010 261586-M21313600007 261586-M21313600024 261586-M21313600006 261586-M21313600023 261586-M21313600021 261586-M21313600015 261586-M21313600005 261586-M21313600025 261586-M21313600022 261586-M21313600008 261586-M21313600018 261586-M21313600016 261586-M21313600012 261586-M21313600011 261586-M21313600004 261586-M21313600003 261586-M21313600002 261586-M21313600001 261586-M21313600014 261586-M21313600019 261586-M21409420001 261586-M21409420015 261586-M21409420014 261586-M21409420005 261586-M21409420004 261586-M21409420008 261586-M21409420007 261586-M21409420009 261586-M21409420003 261586-M21409420013 261586-M21409420010 261586-M21409420012 261586-M21409420011 261586-M21409420006 261586-M21409420002 261586-M21808840003 261586-M21808840004 261586-M21808840005 261586-M21808840002 261586-M21808840009 261586-M21808840006 261586-M21808840010 261586-M21808840016 261586-M21808840013 261586-M21808840014 261586-M21808840015 261586-M21808840001 261586-M21808840007 261586-M21808840008 261586-M21808840011 261586-M21808840012 261586-M21808840018 261586-M21808840019 261586-M21808840020 261586-M21808840017 261586-M21903040001 261586-M21903040011 261586-M21903040004 261586-M21903040008 261586-M21903040005 261586-M21903040003 261586-M21903040012 261586-M21903040009 261586-M21903040007 261586-M21903040006 261586-M21903040002 261586-M21903040010 261586-M21903520014 261586-M21903520015 261586-M21903520019 261586-M21903520003 261586-M21903520004 261586-M21903520005 261586-M21903520009 261586-M21903520012 261586-M21903520020 261586-M21903520010 261586-M21903520017 261586-M21903520002 261586-M21903520008 261586-M21903520007 261586-M21903520006 261586-M21903520011 261586-M21903520016 261586-M21903520018 261586-M21903520013 261586-M21903520001 261586-M22102040002 261586-M22102040001 261586-M22102040006 261586-M22102040003 261586-M22102040020 261586-M22102040016 261586-M22102040015 261586-M22102040013 261586-M22102040010 261586-M22102040008 261586-M22102040007 261586-M22102040017 261586-M22102040009 261586-M22102040018 261586-M22102040014 261586-M22102040011 261586-M22102040019 261586-M22102040004 261586-M22102040005 261586-M22102030014 261586-M22102040012 261586-M22102030009 261586-M22102030007 261586-M22102030006 261586-M22102030005 261586-M22102030003 261586-M22102030002 261586-M22102030001 261586-M22102030008 261586-M22102030011 261586-M22102030010 261586-M22102030015 261586-M22102030004 261586-M22102030012 261586-M22102030013 261586-M22610800010 261586-M22610800001 261586-M22610800020 261586-M22610800015 261586-M22610800013 261586-M22610800006 261586-M22610800012 261586-M22610800018 261586-M22610800019 261586-M22610800002 261586-M22610800011 261586-M22610800007 261586-M22610800004 261586-M22610800003 261586-M22610800017 261586-M22610800014 261586-M22610800005 261586-M22610800016 261586-M22610800009 261586-M22610800008 261586-M22604160020 261586-M22604160019 261586-M22604160018 261586-M22604160017 261586-M22604160006 261586-M22604160001 261586-M22604160014 261586-M22604160015 261586-M22604160003 261586-M22604160016 261586-M22604160005 261586-M22604160004 261586-M22604160023 261586-M22604160013 261586-M22604160008 261586-M22604160007 261586-M22604160025 261586-M22604160024 261586-M22604160002 261586-M22604160022 261586-M22604160010 261586-M22604160009 261586-M22604160021 261586-M22604160012 261586-M22604160011 261586-M22903010024 261586-M22903010002 261586-M22903010013 261586-M22903010016 261586-M22903010021 261586-M22903010023 261586-M22903010026 261586-M22903010028 261586-M22903010009 261586-M22903010015 261586-M22903010025 261586-M22903010005 261586-M22903010004 261586-M22903010030 261586-M22903010017 261586-M22903010008 261586-M22903010003 261586-M22903010001 261586-M22903010007 261586-M22903010006 261586-M22903010011 261586-M22903010012 261586-M22903010014 261586-M22903010019 261586-M22903010022 261586-M22903010029 261586-M22903010010 261586-M22903010018 261586-M22903010020 261586-M22903010027 261586-M22B13630003 261586-M22B13630010 261586-M22B13630007 261586-M22B13630008 261586-M22B13630004 261586-M22B13630001 261586-M22B13630002 261586-M22B13630005 261586-M22B13630006 261586-M22B13630009 261586-M23222150005 261586-M23222150003 261586-M23222150004 261586-M23222150002 261586-M23222150001 261586-M23515090004 261586-M23515090017 261586-M23515090002 261586-M23515090001 261586-M23515090005 261586-M23515090010 261586-M23515090016 261586-M23515090003 261586-M23515090013 261586-M23515090009 261586-M23515090019 261586-M23515090018 261586-M23515090014 261586-M23515090008 261586-M23515090020 261586-M23515090007 261586-M23515090015 261586-M23515090012 261586-M23515090011 261586-M23515090006 261586-M23619570008 261586-M23619570007 261586-M23619570003 261586-M23619580010 261586-M23619580007 261586-M23619570009 261586-M23619580006 261586-M23619570001 261586-M23619580004 261586-M23619580005 261586-M23619580002 261586-M23619570010 261586-M23619580008 261586-M23619580009 261586-M23619580001 261586-M23619570005 261586-M23619570004 261586-M23619570002 261586-M23619570006 261586-M23619580003 261586-M23817650005 261586-M23817650008 261586-M23817650001 261586-M23817650010 261586-M23817650003 261586-M23817650006 261586-M23817650009 261586-M23817650002 261586-M23817650007 261586-M23817650004 261586-M23B04130004 261586-M23B04130002 261586-M23B04130005 261586-M23B04130007 261586-M23B04130001 261586-M23B04130009 261586-M23B04130003 261586-M23B04130008 261586-M23B04130010 261586-M23B04130006 261586-M23C20830011 261586-M23C20830015 261586-M23C20830013 261586-M23C20830008 261586-M23C20830012 261586-M23C20830002 261586-M23C20830006 261586-M23C20830010 261586-M23C20830007 261586-M23C20830004 261586-M23C20830009 261586-M23C20830014 261586-M23C20830001 261586-M23C20830003 261586-M23C20830005 261586-M23C18750005 261586-M23C18750003 261586-M23C18750006 261586-M23C18750002 261586-M23C18750007 261586-M23C18750001 261586-M23C18750008 261586-M23C18750004 261586-M24208270003 261586-M24208270007 261586-M24208270015 261586-M24208270009 261586-M24208270006 261586-M24208270005 261586-M24208270004 261586-M24208270020 261586-M24208270017 261586-M24208270010 261586-M24208270002 261586-M24208270001 261586-M24208270019 261586-M24208270018 261586-M24208270016 261586-M24208270014 261586-M24208270013 261586-M24208270012 261586-M24208270011 261586-M24208270008 261586-M24308410001 261586-M24610250010 261586-M24610250004 261586-M24610250002 261586-M24610250012 261586-M24610250003 261586-M24610250005 261586-M24610250001 261586-M24610250007 261586-M24610250006 261586-M24610250011 261586-M24610250009 261586-M24610250008 261586-M24707600021 261586-M24707600007 261586-M24707600023 261586-M24707590008 261586-M24707590009 261586-M24707600011 261586-M24707600017 261586-M24707600020 261586-M24707600024 261586-M24707590010 261586-M24707590014 261586-M24707600002 261586-M24707590004 261586-M24707590003 261586-M24707590006 261586-M24707600001 261586-M24707600012 261586-M24707590005 261586-M24707600022 261586-M24707600016 261586-M24707590011 261586-M24707590007 261586-M24707590012 261586-M24707590015 261586-M24707590013 261586-M24707600005 261586-M24707600010 261586-M24707600013 261586-M24707600014 261586-M24707600019 261586-M24707600009 261586-M24707600015 261586-M24707590001 261586-M24707590002 261586-M24707600025 261586-M24707600008 261586-M24707600006 261586-M24707600004 261586-M24707600003 261586-M24707600018 261586-M24A19120008 261586-M24A19040002 261586-M24A19040006 261586-M24A19040016 261586-M24A19120002 261586-M24A19120003 261586-M24A19120012 261586-M24A19040012 261586-M24A19120005 261586-M24A19040004 261586-M24A19040003 261586-M24A19120011 261586-M24A19040014 261586-M24A19040001 261586-M24A19040007 261586-M24A19120006 261586-M24A19120007 261586-M24A19040008 261586-M24A19040005 261586-M24A19040015 261586-M24A19120009 261586-M24A19040009 261586-M24A19040013 261586-M24A19120001 261586-M24A19120010 261586-M24A19120004 261586-M24A19040010 261586-M24A19040011 261586-M24A19040017 261586-M24C02200005 261586-M24C02200004 261586-M24C02200003 261586-M24C02200010 261586-M24C02200002 261586-M24C02200007 261586-M24C02200009 261586-M24C02200008 261586-M24C02200001 261586-M24C02200006 EDAN Model: X10 UDI-DI code: 06944413810921 Serial/Lot Numbers: 261590-M21116650004 261590-M21116640010 261590-M21116650001 261590-M21116640009 261590-M21116640002 261590-M21116640001 261590-M21116640008 261590-M21116640006 261590-M21116650003 261590-M21116640004 261590-M21116650005 261590-M21116640007 261590-M21116640003 261590-M21116640005 261590-M21116650002 261590-M21301930004 261590-M21301930003 261590-M21301930002 261590-M21301940001 261590-M21301940003 261590-M21301930005 261590-M21301930001 261590-M21301940002 261590-M21608900002 261590-M21608900001 261590-M21608900003 261590-M21608900004 261590-M21714510002 261590-M21714510005 261590-M21714510001 261590-M21714510003 261590-M21714510004 261590-M21816760004 261590-M21816760003 261590-M21816760005 261590-M21816760002 261590-M21816760001 261590-M21C04540001 261590-M21C04540005 261590-M21C04540009 261590-M21C04540003 261590-M21C04540004 261590-M21C04540007 261590-M21C04540015 261590-M21C04540006 261590-M21C04540010 261590-M21C04540013 261590-M21C04540014 261590-M21C04540008 261590-M21C04540012 261590-M21C04540011 261590-M21C04540002 261590-M22105760002 261590-M22105760001 261590-M22105760005 261590-M22105760004 261590-M22105760003 261590-M22209740001 261590-M22209740004 261590-M22209750005 261590-M22209750004 261590-M22209740007 261590-M22209740002 261590-M22209740009 261590-M22209740005 261590-M22209740006 261590-M22209750003 261590-M22209740010 261590-M22209750002 261590-M22209750001 261590-M22209740008 261590-M22209740003 261590-M22405450002 261590-M22405450004 261590-M22405450001 261590-M22405450010 261590-M22405440002 261590-M22405450003 261590-M22405440001 261590-M22405440005 261590-M22405450006 261590-M22405440003 261590-M22405450009 261590-M22405440004 261590-M22405450008 261590-M22405450007 261590-M22405450005 261590-M22700990001 261590-M23813090008 261590-M23813090005 261590-M23813090001 261590-M23813090003 261590-M23813090015 261590-M23813090009 261590-M23813090002 261590-M23813090013 261590-M23813090012 261590-M23813090014 261590-M23813090011 261590-M23813090007 261590-M23813090010 261590-M23813090006 261590-M23813090004 261590-M23B18340003 261590-M23B18340004 261590-M23B18340005 261590-M23B18340001 261590-M23B18340002 261590-M23C15890007 261590-M23C15900004 261590-M23C15900002 261590-M23C15900003 261590-M23C15890003 261590-M23C15890005 261590-M23C15890010 261590-M23C15900005 261590-M23C15900010 261590-M23C15890002 261590-M23C15890009 261590-M23C15900001 261590-M23C15900008 261590-M23C15900007 261590-M23C15900009 261590-M23C15890008 261590-M23C15900006 261590-M23C15890004 261590-M23C15890001 261590-M23C15890006 261590-M24104070008 261590-M24104070010 261590-M24104070009 261590-M24104070004 261590-M24104070003 261590-M24104070005 261590-M24104070015 261590-M24104070013 261590-M24104070007 261590-M24104070014 261590-M24104070011 261590-M24104070006 261590-M24104070002 261590-M24104070001 261590-M24104070012 261590-M24511440005 261590-M24511440004 261590-M24511440002 261590-M24511440001 261590-M24511440003 261590-M24608030001 261590-M20A08050028 261590-M20A08050001 261590-M20A08050013 261590-M20A08050019 261590-M20A08050022 261590-M20A08050025 261590-M20A08050011 261590-M20A08050004 261590-M20A08050015 261590-M20A08050026 261590-M20A08050003 261590-M20A08050009 261590-M20A08050012 261590-M20A08050027 261590-M20A08050010 261590-M20A08050030 261590-M20A08050018 261590-M20A08050014 261590-M20A08050024 261590-M20A08050007 261590-M20A08050021 261590-M20A08050016 261590-M20A08050006 261590-M20A08050008 261590-M20A08050020 261590-M20A08050023 261590-M20A08050029 261590-M20A08050005 261590-M20A08050017 261590-M20A08050002 261590-M20B18320017 261590-M20B18310005 261590-M20B18310015 261590-M20B18320004 261590-M20B18320005 261590-M20B18310009 261590-M20B18320003 261590-M20B18320019 261590-M20B18310004 261590-M20B18310013 261590-M20B18320012 261590-M20B18320016 261590-M20B18310017 261590-M20B18320002 261590-M20B18320010 261590-M20B18320013 261590-M20B18320018 261590-M20B18310001 261590-M20B18310010 261590-M20B18320008 261590-M20B18320006 261590-M20B18310007 261590-M20B18310003 261590-M20B18320007 261590-M20B18310018 261590-M20B18310019 261590-M20B18320014 261590-M20B18320015 261590-M20B18320020 261590-M20B18310002 261590-M20B18310006 261590-M20B18310008 261590-M20B18310011 261590-M20B18310016 261590-M20B18310020 261590-M20B18320001 261590-M20B18320009 261590-M20B18310014 261590-M20B18320011 261590-M20B18310012 261590-M21108880023 261590-M21108880015 261590-M21108880022 261590-M21108880026 261590-M21108880003 261590-M21108880001 261590-M21108880029 261590-M21108880027 261590-M21108880004 261590-M21108880002 261590-M21108880017 261590-M21108880016 261590-M21108880014 261590-M21108880013 261590-M21108880012 261590-M21108880010 261590-M21108880005 261590-M21108880024 261590-M21108880007 261590-M21108880006 261590-M21108880028 261590-M21108880030 261590-M21108880020 261590-M21108880025 261590-M21108880011 261590-M21108880009 261590-M21108880008 261590-M21108880021 261590-M21108880019 261590-M21108880018 261590-M21113720013 261590-M21113720038 261590-M21113720018 261590-M21113720017 261590-M21113710033 261590-M21113710003 261590-M21113720020 261590-M21113720014 261590-M21113720015 261590-M21113710020 261590-M21113720009 261590-M21113710039 261590-M21113710022 261590-M21113720039 261590-M21113720019 261590-M21113710024 261590-M21113710025 261590-M21113710040 261590-M21113720040 261590-M21113710029 261590-M21113720016 261590-M21113720030 261590-M21113710015 261590-M21113710023 261590-M21113720028 261590-M21113710035 261590-M21113710011 261590-M21113710016 261590-M21113710017 261590-M21113710018 261590-M21113710037 261590-M21113720007 261590-M21113720008 261590-M21113720029 261590-M21113720032 261590-M21113720031 261590-M21113710038 261590-M21113710026 261590-M21113720035 261590-M21113710021 261590-M21113720021 261590-M21113720034 261590-M21113710027 261590-M21113710032 261590-M21113710028 261590-M21113710019 261590-M21113720022 261590-M21113720011 261590-M21113710013 261590-M21113710005 261590-M21113710012 261590-M21113710009 261590-M21113720027 261590-M21113720012 261590-M21113720001 261590-M21113710008 261590-M21113710002 261590-M21113710001 261590-M21113720036 261590-M21113720010 261590-M21113720004 261590-M21113710007 261590-M21113710006 261590-M21113720026 261590-M21113720005 261590-M21113720023 261590-M21113720024 261590-M21113710010 261590-M21113710030 261590-M21113710004 261590-M21113720025 261590-M21113720003 261590-M21113710034 261590-M21113710031 261590-M21113710014 261590-M21113720037 261590-M21113720033 261590-M21113720002 261590-M21113710036 261590-M21113720006 261590-M21205590001 261590-M21205590050 261590-M21205590047 261590-M21205590033 261590-M21205590031 261590-M21205590020 261590-M21205590040 261590-M21205590010 261590-M21205590007 261590-M21205590034 261590-M21205590028 261590-M21205590018 261590-M21205590003 261590-M21205590037 261590-M21205590012 261590-M21205590009 261590-M21205590026 261590-M21205590027 261590-M21205590035 261590-M21205590048 261590-M21205590032 261590-M21205590030 261590-M21205590024 261590-M21205590016 261590-M21205590008 261590-M21205590049 261590-M21205590043 261590-M21205590041 261590-M21205590014 261590-M21205590029 261590-M21205590021 261590-M21205590004 261590-M21205590006 261590-M21205590019 261590-M21205590022 261590-M21205590038 261590-M21205590044 261590-M21205590046 261590-M21205590002 261590-M21205590015 261590-M21205590017 261590-M21205590025 261590-M21205590036 261590-M21205590042 261590-M21205590005 261590-M21205590011 261590-M21205590013 261590-M21205590039 261590-M21205590045 261590-M21205590023 261590-M21400950035 261590-M21400950022 261590-M21400950034 261590-M21400950037 261590-M21400950008 261590-M21400950010 261590-M21400950013 261590-M21400950023 261590-M21400950043 261590-M21400950003 261590-M21400950028 261590-M21400950030 261590-M21400950032 261590-M21400950048 261590-M21400950002 261590-M21400950011 261590-M21400950016 261590-M21400950019 261590-M21400950021 261590-M21400950026 261590-M21400950041 261590-M21400950046 261590-M21400950049 261590-M21400950009 261590-M21400950006 261590-M21400950029 261590-M21400950039 261590-M21400950001 261590-M21400950012 261590-M21400950038 261590-M21400950036 261590-M21400950020 261590-M21400950025 261590-M21400950007 261590-M21400950047 261590-M21400950045 261590-M21400950044 261590-M21400950027 261590-M21400950024 261590-M21400950017 261590-M21400950031 261590-M21400950014 261590-M21400950040 261590-M21400950042 261590-M21400950050 261590-M21400950015 261590-M21400950033 261590-M21400950004 261590-M21400950005 261590-M21400950018 261590-M21409440004 261590-M21409440033 261590-M21409440005 261590-M21409440010 261590-M21409440017 261590-M21409440022 261590-M21409440024 261590-M21409440009 261590-M21409440035 261590-M21409440031 261590-M21409440011 261590-M21409440015 261590-M21409440020 261590-M21409440002 261590-M21409440028 261590-M21409440025 261590-M21409440007 261590-M21409440001 261590-M21409440003 261590-M21409440013 261590-M21409440029 261590-M21409440026 261590-M21409440008 261590-M21409440032 261590-M21409440034 261590-M21409440006 261590-M21409440019 261590-M21409440021 261590-M21409440023 261590-M21409440012 261590-M21409440014 261590-M21409440018 261590-M21409440027 261590-M21409440030 261590-M21409440016 261590-M21409450015 261590-M21409450017 261590-M21409450013 261590-M21409450010 261590-M21409450006 261590-M21409450001 261590-M21409450016 261590-M21409450002 261590-M21409450020 261590-M21409450019 261590-M21409450007 261590-M21409450004 261590-M21409450005 261590-M21409450018 261590-M21409450014 261590-M21409450012 261590-M21409450009 261590-M21409450003 261590-M21409450008 261590-M21409450011 261590-M21701960040 261590-M21701960050 261590-M21701960051 261590-M21701960058 261590-M21701960004 261590-M21701960011 261590-M21701960048 261590-M21701960022 261590-M21701960041 261590-M21701960032 261590-M21701960065 261590-M21701960074 261590-M21701960015 261590-M21701960018 261590-M21701960023 261590-M21701960010 261590-M21701960016 261590-M21701960034 261590-M21701960003 261590-M21701960072 261590-M21701960062 261590-M21701960053 261590-M21701960039 261590-M21701960037 261590-M21701960036 261590-M21701960024 261590-M21701960008 261590-M21701960035 261590-M21701960014 261590-M21701960019 261590-M21701960045 261590-M21701960063 261590-M21701960068 261590-M21701960070 261590-M21701960071 261590-M21701960006 261590-M21701960007 261590-M21701960009 261590-M21701960013 261590-M21701960061 261590-M21701960026 261590-M21701960029 261590-M21701960030 261590-M21701960031 261590-M21701960038 261590-M21701960043 261590-M21701960052 261590-M21701960064 261590-M21701960027 261590-M21701960055 261590-M21701960060 261590-M21701960056 261590-M21701960005 261590-M21701960012 261590-M21701960020 261590-M21701960021 261590-M21701960028 261590-M21701960042 261590-M21701960054 261590-M21701960059 261590-M21701960067 261590-M21701960002 261590-M21701960017 261590-M21701960025 261590-M21701960033 261590-M21701960046 261590-M21701960047 261590-M21701960049 261590-M21701960075 261590-M21701960073 261590-M21701960044 261590-M21701960057 261590-M21701960066 261590-M21701960069 261590-M21701960001 261590-M21903300036 261590-M21903300016 261590-M21903300045 261590-M21903300044 261590-M21903300043 261590-M21903300030 261590-M21903300049 261590-M21903300038 261590-M21903300037 261590-M21903300035 261590-M21903300033 261590-M21903300025 261590-M21903300015 261590-M21903300014 261590-M21903300007 261590-M21903300005 261590-M21903300003 261590-M21903300046 261590-M21903300040 261590-M21903300032 261590-M21903300029 261590-M21903300019 261590-M21903300002 261590-M21903300048 261590-M21903300047 261590-M21903300024 261590-M21903300010 261590-M21903300011 261590-M21903300009 261590-M21903300021 261590-M21903300039 261590-M21903300008 261590-M21903300006 261590-M21903300050 261590-M21903300041 261590-M21903300034 261590-M21903300026 261590-M21903300004 261590-M21903300022 261590-M21903300001 261590-M21903300012 261590-M21903300031 261590-M21903300028 261590-M21903300027 261590-M21903300023 261590-M21903300018 261590-M21903300017 261590-M21903300013 261590-M21903300042 261590-M21903300020 261590-M21903530045 261590-M21903530025 261590-M21903530018 261590-M21903530050 261590-M21903530038 261590-M21903530019 261590-M21903530016 261590-M21903530047 261590-M21903530005 261590-M21903530002 261590-M21903530035 261590-M21903530033 261590-M21903530027 261590-M21903530022 261590-M21903530010 261590-M21903530007 261590-M21903530014 261590-M21903530012 261590-M21903530026 261590-M21903530020 261590-M21903530004 261590-M21903530039 261590-M21903530037 261590-M21903530036 261590-M21903530048 261590-M21903530044 261590-M21903530031 261590-M21903530028 261590-M21903530024 261590-M21903530003 261590-M21903530011 261590-M21903530001 261590-M21903530009 261590-M21903530008 261590-M21903530030 261590-M21903530017 261590-M21903530041 261590-M21903530043 261590-M21903530029 261590-M21903530032 261590-M21903530034 261590-M21903530040 261590-M21903530046 261590-M21903530049 261590-M21903530006 261590-M21903530013 261590-M21903530015 261590-M21903530021 261590-M21903530023 261590-M21903530042 261590-M21B01800018 261590-M21B01800005 261590-M21B01800023 261590-M21B01800024 261590-M21B01800002 261590-M21B01800008 261590-M21B01800021 261590-M21B01800026 261590-M21B01800029 261590-M21B01800035 261590-M21B01800010 261590-M21B01800015 261590-M21B01800020 261590-M21B01800033 261590-M21B01800034 261590-M21B01800001 261590-M21B01800006 261590-M21B01800009 261590-M21B01800011 261590-M21B01800016 261590-M21B01800017 261590-M21B01800019 261590-M21B01800022 261590-M21B01800028 261590-M21B01800037 261590-M21B01800032 261590-M21B01800030 261590-M21B01800038 261590-M21B01800014 261590-M21B01800025 261590-M21B01800013 261590-M21B01800004 261590-M21B01800003 261590-M21B01800039 261590-M21B01800036 261590-M21B01800040 261590-M21B01800027 261590-M21B01800012 261590-M21B01800007 261590-M21B01800031 261590-M21C15310028 261590-M21C15310015 261590-M21C15310002 261590-M21C15310007 261590-M21C15310008 261590-M21C15310010 261590-M21C15310016 261590-M21C15310021 261590-M21C15310023 261590-M21C15310024 261590-M21C15310013 261590-M21C15310017 261590-M21C15310018 261590-M21C15310029 261590-M21C15310006 261590-M21C15310019 261590-M21C15310020 261590-M21C15310022 261590-M21C15310026 261590-M21C15310014 261590-M21C15310030 261590-M21C15310009 261590-M21C15310027 261590-M21C15310001 261590-M21C15310003 261590-M21C15310004 261590-M21C15310005 261590-M21C15310012 261590-M21C15310011 261590-M21C15310025 261590-M22212420003 261590-M22212420002 261590-M22212420001 261590-M22212420024 261590-M22212420013 261590-M22212420011 261590-M22212420006 261590-M22212420014 261590-M22212420018 261590-M22212420026 261590-M22212420005 261590-M22212420012 261590-M22212420027 261590-M22212420029 261590-M22212420025 261590-M22212420017 261590-M22212420004 261590-M22212420019 261590-M22212420015 261590-M22212420009 261590-M22212420023 261590-M22212420022 261590-M22212420021 261590-M22212420010 261590-M22212420007 261590-M22212420030 261590-M22212420008 261590-M22212420016 261590-M22212420028 261590-M22212420020 261590-M22301860028 261590-M22301860020 261590-M22301860010 261590-M22301860025 261590-M22301860003 261590-M22301860008 261590-M22301860004 261590-M22301860013 261590-M22301860015 261590-M22301860030 261590-M22301860024 261590-M22301860018 261590-M22301860027 261590-M22301860029 261590-M22301860021 261590-M22301860026 261590-M22301860001 261590-M22301860006 261590-M22301860007 261590-M22301860009 261590-M22301860022 261590-M22301860005 261590-M22301860012 261590-M22301860016 261590-M22301860017 261590-M22301860019 261590-M22301860023 261590-M22301860002 261590-M22301860014 261590-M22301860011 261590-M22407370011 261590-M22407370005 261590-M22407370001 261590-M22407370023 261590-M22407370015 261590-M22407370006 261590-M22407370002 261590-M22407370008 261590-M22407370022 261590-M22407370003 261590-M22407370010 261590-M22407370016 261590-M22407370014 261590-M22407370017 261590-M22407370019 261590-M22407370025 261590-M22407370012 261590-M22407370018 261590-M22407370020 261590-M22407370013 261590-M22407370021 261590-M22407370004 261590-M22407370007 261590-M22407370009 261590-M22407370024 261590-M22501290013 261590-M22501290014 261590-M22501290029 261590-M22501290002 261590-M22501290020 261590-M22501290026 261590-M22501290015 261590-M22501290028 261590-M22501290033 261590-M22501290018 261590-M22501290005 261590-M22501290036 261590-M22501290016 261590-M22501290021 261590-M22501290025 261590-M22501290012 261590-M22501290009 261590-M22501290037 261590-M22501290031 261590-M22501290019 261590-M22501290003 261590-M22501290040 261590-M22501290034 261590-M22501290023 261590-M22501290035 261590-M22501290007 261590-M22501290032 261590-M22501290024 261590-M22501290017 261590-M22501290011 261590-M22501290008 261590-M22501290006 261590-M22501290030 261590-M22501290027 261590-M22501290004 261590-M22501290039 261590-M22501290022 261590-M22501290010 261590-M22501290001 261590-M22501290038 261590-M22B07360035 261590-M22B07360001 261590-M22B07360005 261590-M22B07360012 261590-M22B07360003 261590-M22B07360046 261590-M22B07360008 261590-M22B07360004 261590-M22B07360007 261590-M22B07360049 261590-M22B07360044 261590-M22B07360040 261590-M22B07360039 261590-M22B07360030 261590-M22B07360026 261590-M22B07360023 261590-M22B07360002 261590-M22B07360038 261590-M22B07360043 261590-M22B07360009 261590-M22B07360022 261590-M22B07360042 261590-M22B07360032 261590-M22B07360028 261590-M22B07360027 261590-M22B07360015 261590-M22B07360006 261590-M22B07360050 261590-M22B07360045 261590-M22B07360041 261590-M22B07360010 261590-M22B07360029 261590-M22B07360033 261590-M22B07360034 261590-M22B07360036 261590-M22B07360017 261590-M22B07360011 261590-M22B07360018 261590-M22B07360020 261590-M22B07360024 261590-M22B07360019 261590-M22B07360021 261590-M22B07360013 261590-M22B07360014 261590-M22B07360016 261590-M22B07360031 261590-M22B07360047 261590-M22B07360048 261590-M22B07360037 261590-M22B07360025 261590-M23222350007 261590-M23222350017 261590-M23222350023 261590-M23222350035 261590-M23222350029 261590-M23222350028 261590-M23222350019 261590-M23222350018 261590-M23222350009 261590-M23222350008 261590-M23222350014 261590-M23222350032 261590-M23222350031 261590-M23222350030 261590-M23222350004 261590-M23222350025 261590-M23222350033 261590-M23222350026 261590-M23222350024 261590-M23222350016 261590-M23222350039 261590-M23222350038 261590-M23222350015 261590-M23222350005 261590-M23222350037 261590-M23222350022 261590-M23222350021 261590-M23222350020 261590-M23222350040 261590-M23222350012 261590-M23222350011 261590-M23222350010 261590-M23222350002 261590-M23222350001 261590-M23222350036 261590-M23222350034 261590-M23222350027 261590-M23222350013 261590-M23222350006 261590-M23222350003 261590-M23614470039 261590-M23614470033 261590-M23614470024 261590-M23614470021 261590-M23614470019 261590-M23614470002 261590-M23614470011 261590-M23614470034 261590-M23614470031 261590-M23614470029 261590-M23614470023 261590-M23614470018 261590-M23614470017 261590-M23614470014 261590-M23614470035 261590-M23614470015 261590-M23614470005 261590-M23614470001 261590-M23614470003 261590-M23614470022 261590-M23614470016 261590-M23614470010 261590-M23614470009 261590-M23614470030 261590-M23614470036 261590-M23614470037 261590-M23614470028 261590-M23614470027 261590-M23614470026 261590-M23614470020 261590-M23614470013 261590-M23614470007 261590-M23614470006 261590-M23614470012 261590-M23614470040 261590-M23614470038 261590-M23614470032 261590-M23614470008 261590-M23614470004 261590-M23614470025 261590-M23818720033 261590-M23818720029 261590-M23818720003 261590-M23818720005 261590-M23818720011 261590-M23818720038 261590-M23818720030 261590-M23818720028 261590-M23818720024 261590-M23818720019 261590-M23818720017 261590-M23818720006 261590-M23818720034 261590-M23818720009 261590-M23818720016 261590-M23818720022 261590-M23818720027 261590-M23818720035 261590-M23818720032 261590-M23818720021 261590-M23818720040 261590-M23818720039 261590-M23818720037 261590-M23818720025 261590-M23818720018 261590-M23818720014 261590-M23818720031 261590-M23818720026 261590-M23818720010 261590-M23818720007 261590-M23818720002 261590-M23818720036 261590-M23818720023 261590-M23818720015 261590-M23818720012 261590-M23818720013 261590-M23818720001 261590-M23818720020 261590-M23818720008 261590-M23818720004 261590-M23911440017 261590-M23911440018 261590-M23911440006 261590-M23911440027 261590-M23911440038 261590-M23911440039 261590-M23911440040 261590-M23911440007 261590-M23911440024 261590-M23911440025 261590-M23911440031 261590-M23911440001 261590-M23911440033 261590-M23911440030 261590-M23911440026 261590-M23911440022 261590-M23911440021 261590-M23911440015 261590-M23911440014 261590-M23911440037 261590-M23911440002 261590-M23911440023 261590-M23911440005 261590-M23911440003 261590-M23911440011 261590-M23911440012 261590-M23911440016 261590-M23911440034 261590-M23911440035 261590-M23911440020 261590-M23911440008 261590-M23911440010 261590-M23911440013 261590-M23911440028 261590-M23911440029 261590-M23911440032 261590-M23911440036 261590-M23911440004 261590-M23911440009 261590-M23911440019 261590-M23910690005 261590-M23910690009 261590-M23910690003 261590-M23910690007 261590-M23910690004 261590-M23910690011 261590-M23910690014 261590-M23910690006 261590-M23910690002 261590-M23910690012 261590-M23910690010 261590-M23910690015 261590-M23910690013 261590-M23910690001 261590-M23910690008 261590-M23B05760004 261590-M23B05760048 261590-M23B05760017 261590-M23B05760018 261590-M23B05760026 261590-M23B05760034 261590-M23B05760038 261590-M23B05760040 261590-M23B05760001 261590-M23B05760028 261590-M23B05760047 261590-M23B05760045 261590-M23B05760006 261590-M23B05760014 261590-M23B05760016 261590-M23B05760022 261590-M23B05760023 261590-M23B05760025 261590-M23B05760036 261590-M23B05760039 261590-M23B05760008 261590-M23B05760042 261590-M23B05760035 261590-M23B05760007 261590-M23B05760015 261590-M23B05760031 261590-M23B05760011 261590-M23B05760019 261590-M23B05760027 261590-M23B05760030 261590-M23B05760044 261590-M23B05760046 261590-M23B05760010 261590-M23B05760021 261590-M23B05760037 261590-M23B05760043 261590-M23B05760002 261590-M23B05760003 261590-M23B05760005 261590-M23B05760013 261590-M23B05760029 261590-M23B05760032 261590-M23B05760033 261590-M23B05760041 261590-M23B05760009 261590-M23B05760012 261590-M23B05760020 261590-M23B05760024 262935-M24106050015 262935-M24106050012 262935-M24106050017 262935-M24106050057 262935-M24106050021 262935-M24106050022 262935-M24106050027 262935-M24106050031 262935-M24106050060 262935-M24106050052 262935-M24106050051 262935-M24106050049 262935-M24106050048 262935-M24106050005 262935-M24106050006 262935-M24106050016 262935-M24106050013 262935-M24106050039 262935-M24106050034 262935-M24106050023 262935-M24106050056 262935-M24106050055 262935-M24106050053 262935-M24106050047 262935-M24106050042 262935-M24106050041 262935-M24106050037 262935-M24106050032 262935-M24106050030 262935-M24106050024 262935-M24106050020 262935-M24106050014 262935-M24106050010 262935-M24106050004 262935-M24106050036 262935-M24106050026 262935-M24106050025 262935-M24106050011 262935-M24106050007 262935-M24106050002 262935-M24106050059 262935-M24106050001 262935-M24106050009 262935-M24106050019 262935-M24106050028 262935-M24106050029 262935-M24106050038 262935-M24106050018 262935-M24106050033 262935-M24106050035 262935-M24106050008 262935-M24106050045 262935-M24106050043 262935-M24106050058 262935-M24106050040 262935-M24106050003 262935-M24106050044 262935-M24106050050 262935-M24106050046 262935-M24106050054 262935-M24115330001 262935-M24115330014 262935-M24115330031 262935-M24115330040 262935-M24115330004 262935-M24115330027 262935-M24115330039 262935-M24115330037 262935-M24115330006 262935-M24115330057 262935-M24115330030 262935-M24115330053 262935-M24115330033 262935-M24115330051 262935-M24115330007 262935-M24115330012 262935-M24115330013 262935-M24115330018 262935-M24115330041 262935-M24115330042 262935-M24115330043 262935-M24115330005 262935-M24115330008 262935-M24115330052 262935-M24115330016 262935-M24115330034 262935-M24115330046 262935-M24115330050 262935-M24115330021 262935-M24115330022 262935-M24115330023 262935-M24115330032 262935-M24115330038 262935-M24115330044 262935-M24115330020 262935-M24115330009 262935-M24115330055 262935-M24115330035 262935-M24115330058 262935-M24115330059 262935-M24115330015 262935-M24115330019 262935-M24115330026 262935-M24115330036 262935-M24115330045 262935-M24115330029 262935-M24115330028 262935-M24115330011 262935-M24115330025 262935-M24115330047 262935-M24115330024 262935-M24115330017 262935-M24115330010 262935-M24115330003 262935-M24115330002 262935-M24115330048 262935-M24115330049 262935-M24115330056 262935-M24115330060 262935-M24115330054 262935-M24305320004 262935-M24305320009 262935-M24305320001 262935-M24305320008 262935-M24305320003 262935-M24305320002 262935-M24305320010 262935-M24305320006 262935-M24305320005 262935-M24305320007 262935-M24404340018 262935-M24404340019 262935-M24404340035 262935-M24404340014 262935-M24404340020 262935-M24404340037 262935-M24404340006 262935-M24404340011 262935-M24404340040 262935-M24404340013 262935-M24404340045 262935-M24404340036 262935-M24404340030 262935-M24404340024 262935-M24404340015 262935-M24404340009 262935-M24404340008 262935-M24404340004 262935-M24404340003 262935-M24404340002 262935-M24404340038 262935-M24404340012 262935-M24404340041 262935-M24404340032 262935-M24404340044 262935-M24404340050 262935-M24404340047 262935-M24404340046 262935-M24404340029 262935-M24404340028 262935-M24404340023 262935-M24404340022 262935-M24404340017 262935-M24404340016 262935-M24404340039 262935-M24404340043 262935-M24404340034 262935-M24404340031 262935-M24404340025 262935-M24404340010 262935-M24404340049 262935-M24404340048 262935-M24404340042 262935-M24404340027 262935-M24404340026 262935-M24404340005 262935-M24404340007 262935-M24404340033 262935-M24404340001 262935-M24404340021 262935-M24404350036 262935-M24404350040 262935-M24404350050 262935-M24404350002 262935-M24404350005 262935-M24404350027 262935-M24404350034 262935-M24404350038 262935-M24404350047 262935-M24404350057 262935-M24404350007 262935-M24404350014 262935-M24404350022 262935-M24404350030 262935-M24404350025 262935-M24404350045 262935-M24404350049 262935-M24404350052 262935-M24404350055 262935-M24404350016 262935-M24404350018 262935-M24404350041 262935-M24404350046 262935-M24404350051 262935-M24404350053 262935-M24404350056 262935-M24404350059 262935-M24404350043 262935-M24404350031 262935-M24404350012 262935-M24404350003 262935-M24404350039 262935-M24404350004 262935-M24404350006 262935-M24404350011 262935-M24404350013 262935-M24404350015 262935-M24404350019 262935-M24404350037 262935-M24404350044 262935-M24404350048 262935-M24404350054 262935-M24404350058 262935-M24404350009 262935-M24404350060 262935-M24404350026 262935-M24404350028 262935-M24404350032 262935-M24404350033 262935-M24404350035 262935-M24404350010 262935-M24404350017 262935-M24404350021 262935-M24404350023 262935-M24404350029 262935-M24404350042 262935-M24404350001 262935-M24404350024 262935-M24404350020 262935-M24404350008 261590-M24511200002 261590-M24511200006 261590-M24511200020 261590-M24511200015 261590-M24511200010 261590-M24511200019 261590-M24511200005 261590-M24511200011 261590-M24511200012 261590-M24511200004 261590-M24511200007 261590-M24511200014 261590-M24511200018 261590-M24511200016 261590-M24511200013 261590-M24511200009 261590-M24511200001 261590-M24511200003 261590-M24511200008 261590-M24511200017 262935-M24516320001 262935-M24516320006 262935-M24516320014 262935-M24516330052 262935-M24516330038 262935-M24516320056 262935-M24516320046 262935-M24516320042 262935-M24516320010 262935-M24516320013 262935-M24516320041 262935-M24516320002 262935-M24516330054 262935-M24516320029 262935-M24516320035 262935-M24516330039 262935-M24516330042 262935-M24516330044 262935-M24516330053 262935-M24516330057 262935-M24516320020 262935-M24516320039 262935-M24516320023 262935-M24516320027 262935-M24516320028 262935-M24516320015 262935-M24516320030 262935-M24516320033 262935-M24516320034 262935-M24516320024 262935-M24516320019 262935-M24516320031 262935-M24516320036 262935-M24516320026 262935-M24516330035 262935-M24516330055 262935-M24516330050 262935-M24516330018 262935-M24516330008 262935-M24516330006 262935-M24516320032 262935-M24516320022 262935-M24516320021 262935-M24516330058 262935-M24516330048 262935-M24516330046 262935-M24516330037 262935-M24516330016 262935-M24516330011 262935-M24516330059 262935-M24516330030 262935-M24516330033 262935-M24516330025 262935-M24516320016 262935-M24516320040 262935-M24516320043 262935-M24516320047 262935-M24516320048 262935-M24516320049 262935-M24516320005 262935-M24516320057 262935-M24516330002 262935-M24516330014 262935-M24516330043 262935-M24516330047 262935-M24516320017 262935-M24516320018 262935-M24516320060 262935-M24516320055 262935-M24516320004 262935-M24516330032 262935-M24516330029 262935-M24516320011 262935-M24516320012 262935-M24516320044 262935-M24516320045 262935-M24516320050 262935-M24516330034 262935-M24516320053 262935-M24516320058 262935-M24516320059 262935-M24516330013 262935-M24516330017 262935-M24516330022 262935-M24516330024 262935-M24516320054 262935-M24516330005 262935-M24516330004 262935-M24516330023 262935-M24516330026 262935-M24516330028 262935-M24516330020 262935-M24516330012 262935-M24516330009 262935-M24516330007 262935-M24516330021 262935-M24516330003 262935-M24516320051 262935-M24516320025 262935-M24516330031 262935-M24516330036 262935-M24516330040 262935-M24516330045 262935-M24516320052 262935-M24516330049 262935-M24516330019 262935-M24516330010 262935-M24516330027 262935-M24516330041 262935-M24516330051 262935-M24516330056 262935-M24516320003 262935-M24516330015 262935-M24516320007 262935-M24516320009 262935-M24516320037 262935-M24516320038 262935-M24516330001 262935-M24516320008 262935-M24516330060 262935-M24701880023 262935-M24701880003 262935-M24701880060 262935-M24701880032 262935-M24701880033 262935-M24701880026 262935-M24701880058 262935-M24701880047 262935-M24701880011 262935-M24701880057 262935-M24701880006 262935-M24701880013 262935-M24701880053 262935-M24701880039 262935-M24701880036 262935-M24701880028 262935-M24701880027 262935-M24701880009 262935-M24701880021 262935-M24701880038 262935-M24701880043 262935-M24701880041 262935-M24701880029 262935-M24701880052 262935-M24701880025 262935-M24701880024 262935-M24701880020 262935-M24701880008 262935-M24701880007 262935-M24701880059 262935-M24701880056 262935-M24701880050 262935-M24701880045 262935-M24701880031 262935-M24701880015 262935-M24701880012 262935-M24701880001 262935-M24701880048 262935-M24701880044 262935-M24701880040 262935-M24701880042 262935-M24701880018 262935-M24701880037 262935-M24701880014 262935-M24701880054 262935-M24701880055 262935-M24701880002 262935-M24701880004 262935-M24701880005 262935-M24701880017 262935-M24701880019 262935-M24701880016 262935-M24701880030 262935-M24701880034 262935-M24701880035 262935-M24701880046 262935-M24701880049 262935-M24701880051 262935-M24701880010 262935-M24701880022 262935-M24803620018 262935-M24803620052 262935-M24803620056 262935-M24803620007 262935-M24803620001 262935-M24803620003 262935-M24803620017 262935-M24803620019 262935-M24803620048 262935-M24803620008 262935-M24803620039 262935-M24803620047 262935-M24803620049 262935-M24803620057 262935-M24803620058 262935-M24803620033 262935-M24803620028 262935-M24803620031 262935-M24803620027 262935-M24803620024 262935-M24803620040 262935-M24803620060 262935-M24803620011 262935-M24803620012 262935-M24803620016 262935-M24803620021 262935-M24803620023 262935-M24803620032 262935-M24803620035 262935-M24803620036 262935-M24803620042 262935-M24803620046 262935-M24803620025 262935-M24803620051 262935-M24803620053 262935-M24803620002 262935-M24803620006 262935-M24803620029 262935-M24803620041 262935-M24803620030 262935-M24803620034 262935-M24803620037 262935-M24803620014 262935-M24803620013 262935-M24803620055 262935-M24803620054 262935-M24803620059 262935-M24803620038 262935-M24803620010 262935-M24803620015 262935-M24803620026 262935-M24803620005 262935-M24803620004 262935-M24803620050 262935-M24803620045 262935-M24803620044 262935-M24803620043 262935-M24803620009 262935-M24803620022 262935-M24803620020 262935-M24903910004 262935-M24903910005 262935-M24903910048 262935-M24903910049 262935-M24903910057 262935-M24903910020 262935-M24903910023 262935-M24903910025 262935-M24903910039 262935-M24903910007 262935-M24903910017 262935-M24903910022 262935-M24903910034 262935-M24903910050 262935-M24903910051 262935-M24903910058 262935-M24903910059 262935-M24903910003 262935-M24903910010 262935-M24903910011 262935-M24903910013 262935-M24903910015 262935-M24903910018 262935-M24903910024 262935-M24903910035 262935-M24903910037 262935-M24903910053 262935-M24903910055 262935-M24903910041 262935-M24903910012 262935-M24903910046 262935-M24903910052 262935-M24903910014 262935-M24903910026 262935-M24903910032 262935-M24903910040 262935-M24903910043 262935-M24903910045 262935-M24903910002 262935-M24903910006 262935-M24903910016 262935-M24903910036 262935-M24903910044 262935-M24903910056 262935-M24903910060 262935-M24903910001 262935-M24903910054 262935-M24903910009 262935-M24903910008 262935-M24903910047 262935-M24903910029 262935-M24903910028 262935-M24903910042 262935-M24903910038 262935-M24903910021 262935-M24903910031 262935-M24903910030 262935-M24903910027 262935-M24903910019 262935-M24903910033 261590-M24A06740009 261590-M24A06740014 261590-M24A06740040 261590-M24A06740007 261590-M24A06740031 261590-M24A06740028 261590-M24A06740023 261590-M24A06740016 261590-M24A06740015 261590-M24A06740002 261590-M24A06740038 261590-M24A06740024 261590-M24A06740019 261590-M24A06740017 261590-M24A06740010 261590-M24A06740008 261590-M24A06740036 261590-M24A06740035 261590-M24A06740034 261590-M24A06740026 261590-M24A06740001 261590-M24A06740029 261590-M24A06740027 261590-M24A06740022 261590-M24A06740020 261590-M24A06740012 261590-M24A06740003 261590-M24A06740013 261590-M24A06740032 261590-M24A06740005 261590-M24A06740018 261590-M24A06740025 261590-M24A06740033 261590-M24A06740004 261590-M24A06740021 261590-M24A06740030 261590-M24A06740006 261590-M24A06740039 261590-M24A06740037 261590-M24A06740011 262935-M24B03740024 262935-M24B03740041 262935-M24B03740047 262935-M24B03740004 262935-M24B03740038 262935-M24B03740045 262935-M24B03740006 262935-M24B03740003 262935-M24B03740039 262935-M24B03740012 262935-M24B03740060 262935-M24B03740052 262935-M24B03740037 262935-M24B03740022 262935-M24B03740027 262935-M24B03740055 262935-M24B03740019 262935-M24B03740017 262935-M24B03740005 262935-M24B03740059 262935-M24B03740057 262935-M24B03740050 262935-M24B03740043 262935-M24B03740035 262935-M24B03740021 262935-M24B03740011 262935-M24B03740029 262935-M24B03740046 262935-M24B03740018 262935-M24B03740036 262935-M24B03740031 262935-M24B03740034 262935-M24B03740044 262935-M24B03740048 262935-M24B03740056 262935-M24B03740015 262935-M24B03740025 262935-M24B03740030 262935-M24B03740032 262935-M24B03740040 262935-M24B03740049 262935-M24B03740053 262935-M24B03740042 262935-M24B03740001 262935-M24B03740016 262935-M24B03740023 262935-M24B03740026 262935-M24B03740033 262935-M24B03740051 262935-M24B03740054 262935-M24B03740002 262935-M24B03740007 262935-M24B03740008 262935-M24B03740058 262935-M24B03740010 262935-M24B03740028 262935-M24B03740013 262935-M24B03740009 262935-M24B03740014 262935-M24B03740020 261590-M24A19020016 261590-M24A19020007 261590-M24A19020008 261590-M24A19020014 261590-M24A19020015 261590-M24A19020011 261590-M24A19020012 261590-M24A19020002 261590-M24A19020020 261590-M24A19020019 261590-M24A19020018 261590-M24A19020017 261590-M24A19020013 261590-M24A19020009 261590-M24A19020005 261590-M24A19020003 261590-M24A19020004 261590-M24A19020006 261590-M24A19020001 261590-M24A19020010 261590-M24A19030033 261590-M24A19030030 261590-M24A19030018 261590-M24A19030006 261590-M24A19030040 261590-M24A19030002 261590-M24A19030007 261590-M24A19030028 261590-M24A19030022 261590-M24A19030011 261590-M24A19030026 261590-M24A19030038 261590-M24A19030024 261590-M24A19030037 261590-M24A19030009 261590-M24A19030001 261590-M24A19030027 261590-M24A19030025 261590-M24A19030013 261590-M24A19030020 261590-M24A19030023 261590-M24A19030032 261590-M24A19030035 261590-M24A19030003 261590-M24A19030008 261590-M24A19030014 261590-M24A19030016 261590-M24A19030021 261590-M24A19030029 261590-M24A19030010 261590-M24A19030005 261590-M24A19030031 261590-M24A19030015 261590-M24A19030012 261590-M24A19030004 261590-M24A19030039 261590-M24A19030017 261590-M24A19030036 261590-M24A19030034 261590-M24A19030019 261590-M24C13490013 261590-M24C13490001 261590-M24C13490007 261590-M24C13490011 261590-M24C13490018 261590-M24C13490002 261590-M24C13490034 261590-M24C13490023 261590-M24C13490016 261590-M24C13490015 261590-M24C13490012 261590-M24C13490010 261590-M24C13490008 261590-M24C13490005 261590-M24C13490032 261590-M24C13490035 261590-M24C13490003 261590-M24C13490021 261590-M24C13490027 261590-M24C13490006 261590-M24C13490017 261590-M24C13490030 261590-M24C13490024 261590-M24C13490031 261590-M24C13490033 261590-M24C13490004 261590-M24C13490009 261590-M24C13490014 261590-M24C13490019 261590-M24C13490020 261590-M24C13490026 261590-M24C13490028 261590-M24C13490029 261590-M24C13490022 261590-M24C13490025 262935-M25510510077 262935-M25510510072 262935-M25510510057 262935-M25510510055 262935-M25510510013 262935-M25510510010 262935-M25510510007 262935-M25510510002 262935-M25510510052 262935-M25510510051 262935-M25510510046 262935-M25510510044 262935-M25510510040 262935-M25510510039 262935-M25510510038 262935-M25510510029 262935-M25510510019 262935-M25510510018 262935-M25510520074 262935-M25510520073 262935-M25510520049 262935-M25510520040 262935-M25510520022 262935-M25510520012 262935-M25510520008 262935-M25510520009 262935-M25510520011 262935-M25510510016 262935-M25510520031 262935-M25510510080 262935-M25510510079 262935-M25510510078 262935-M25510510066 262935-M25510510064 262935-M25510510063 262935-M25510510060 262935-M25510510059 262935-M25510510058 262935-M25510510026 262935-M25510510024 262935-M25510510023 262935-M25510510020 262935-M25510520013 262935-M25510520004 262935-M25510520003 262935-M25510510071 262935-M25510510075 262935-M25510510014 262935-M25510520099 262935-M25510520098 262935-M25510520089 262935-M25510520088 262935-M25510520080 262935-M25510520069 262935-M25510520068 262935-M25510520042 262935-M25510520032 262935-M25510520030 262935-M25510520029 262935-M25510520028 262935-M25510520020 262935-M25510520019 262935-M25510520018 262935-M25510520010 262935-M25510520001 262935-M25510510076 262935-M25510510074 262935-M25510510073 262935-M25510510070 262935-M25510510009 262935-M25510510008 262935-M25510510028 262935-M25510510030 262935-M25510510033 262935-M25510510034 262935-M25510510069 262935-M25510510068 262935-M25510510043 262935-M25510510005 262935-M25510510001 262935-M25510520094 262935-M25510520093 262935-M25510520091 262935-M25510520084 262935-M25510520083 262935-M25510520077 262935-M25510520075 262935-M25510520064 262935-M25510520063 262935-M25510520057 262935-M25510520037 262935-M25510520027 262935-M25510520025 262935-M25510520017 262935-M25510520015 262935-M25510510065 262935-M25510510062 262935-M25510510061 262935-M25510510049 262935-M25510510048 262935-M25510510036 262935-M25510520061 262935-M25510510027 262935-M25510520055 262935-M25510520044 262935-M25510520034 262935-M25510520033 262935-M25510520056 262935-M25510520016 262935-M25510520026 262935-M25510520035 262935-M25510520097 262935-M25510520081 262935-M25510520076 262935-M25510520067 262935-M25510520071 262935-M25510520054 262935-M25510520053 262935-M25510520051 262935-M25510520048 262935-M25510520036 262935-M25510510015 262935-M25510520066 262935-M25510520065 262935-M25510520043 262935-M25510520041 262935-M25510520024 262935-M25510520023 262935-M25510520021 262935-M25510520014 262935-M25510510047 262935-M25510510045 262935-M25510510042 262935-M25510510041 262935-M25510510025 262935-M25510510022 262935-M25510510021 262935-M25510520002 262935-M25510520072 262935-M25510520085 262935-M25510520087 262935-M25510520095 262935-M25510510035 262935-M25510510032 262935-M25510510031 262935-M25510510017 262935-M25510510012 262935-M25510510011 262935-M25510510006 262935-M25510510067 262935-M25510510004 262935-M25510520096 262935-M25510510003 262935-M25510520006 262935-M25510520050 262935-M25510520047 262935-M25510520100 262935-M25510520086 262935-M25510520079 262935-M25510520078 262935-M25510520070 262935-M25510520052 262935-M25510520046 262935-M25510520045 262935-M25510520039 262935-M25510520038 262935-M25510520007 262935-M25510520005 262935-M25510510056 262935-M25510510054 262935-M25510510053 262935-M25510520058 262935-M25510510050 262935-M25510510037 262935-M25510520059 262935-M25510520062 262935-M25510520060 262935-M25510520092 262935-M25510520090 262935-M25510520082 262935-M25510550019 262935-M25510550016 262935-M25510550012 262935-M25510550003 262935-M25510550067 262935-M25510550062 262935-M25510550058 262935-M25510550051 262935-M25510550004 262935-M25510550022 262935-M25510550043 262935-M25510550008 262935-M25510550001 262935-M25510550045 262935-M25510550070 262935-M25510550076 262935-M25510550100 262935-M25510550108 262935-M25510550027 262935-M25510550030 262935-M25510550065 262935-M25510550063 262935-M25510550105 262935-M25510550104 262935-M25510550099 262935-M25510550096 262935-M25510550092 262935-M25510550090 262935-M25510550086 262935-M25510550080 262935-M25510550078 262935-M25510550071 262935-M25510550066 262935-M25510550060 262935-M25510550057 262935-M25510550026 262935-M25510550020 262935-M25510550010 262935-M25510550005 262935-M25510550049 262935-M25510550046 262935-M25510550042 262935-M25510550085 262935-M25510550093 262935-M25510550095 262935-M25510550101 262935-M25510550091 262935-M25510550088 262935-M25510550081 262935-M25510550075 262935-M25510550068 262935-M25510550061 262935-M25510550054 262935-M25510550103 262935-M25510550079 262935-M25510550077 262935-M25510550072 262935-M25510550040 262935-M25510550059 262935-M25510550052 262935-M25510550050 262935-M25510550035 262935-M25510550033 262935-M25510550023 262935-M25510550013 262935-M25510550009 262935-M25510550007 262935-M25510550002 262935-M25510550102 262935-M25510550056 262935-M25510550029 262935-M25510550039 262935-M25510550032 262935-M25510550015 262935-M25510550006 262935-M25510550097 262935-M25510550089 262935-M25510550087 262935-M25510550082 262935-M25510550074 262935-M25510550069 262935-M25510550109 262935-M25510550106 262935-M25510550094 262935-M25510550084 262935-M25510550073 262935-M25510550064 262935-M25510550055 262935-M25510550053 262935-M25510550047 262935-M25510550036 262935-M25510550025 262935-M25510550034 262935-M25510550041 262935-M25510550044 262935-M25510550037 262935-M25510550028 262935-M25510550017 262935-M25510550048 262935-M25510550021 262935-M25510550107 262935-M25510550083 262935-M25510550038 262935-M25510550031 262935-M25510550024 262935-M25510550018 262935-M25510550014 262935-M25510550011 262935-M25510550098 262935-M25510550110 262935-M25621160043 262935-M25621160044 262935-M25621160045 262935-M25621160055 262935-M25621160057 262935-M25621160076 262935-M25621160099 262935-M25621160111 262935-M25621160127 262935-M25621160129 262935-M25621160147 262935-M25621170006 262935-M25621170016 262935-M25621170046 262935-M25621170054 262935-M25621170055 262935-M25621170060 262935-M25621170084 262935-M25621800021 262935-M25621800040 262935-M25621800043 262935-M25621800046 262935-M25621800048 262935-M25621800052 262935-M25704350012 262935-M25704350014 262935-M25704350015 262935-M25621160040 262935-M25621160027 262935-M25621160018 262935-M25621160009 262935-M25621160019 262935-M25621160062 262935-M25621160088 262935-M25621160004 262935-M25621160005 262935-M25621160015 262935-M25704350088 262935-M25621160113 262935-M25621160125 262935-M25621160132 262935-M25621160139 262935-M25621170004 262935-M25621170028 262935-M25621170036 262935-M25621170044 262935-M25621170045 262935-M25621170056 262935-M25621170061 262935-M25621170077 262935-M25621160017 262935-M25621160038 262935-M25621160039 262935-M25621160047 262935-M25621160064 262935-M25621160070 262935-M25621170078 262935-M25621170079 262935-M25621170090 262935-M25621160037 262935-M25621800003 262935-M25621160035 262935-M25704350086 262935-M25621800024 262935-M25621800037 262935-M25621800042 262935-M25704350006 262935-M25704350007 262935-M25704350008 262935-M25704350022 262935-M25704350024 262935-M25621800034 262935-M25621800045 262935-M25621800051 262935-M25704350003 262935-M25704350043 262935-M25704350046 262935-M25704350050 262935-M25704350025 262935-M25704350030 262935-M25704350031 262935-M25704350039 262935-M25704350057 262935-M25704350083 262935-M25621160007 262935-M25704350051 262935-M25704350059 262935-M25704350062 262935-M25704350064 262935-M25704350065 262935-M25704350087 262935-M25621800006 262935-M25621800013 262935-M25621800016 262935-M25621800031 262935-M25621800049 262935-M25621800055 262935-M25704350033 262935-M25704350044 262935-M25704350053 262935-M25704350067 262935-M25704350092 262935-M25704350095 262935-M25704350040 262935-M25704350049 262935-M25704350052 262935-M25704350054 262935-M25704350055 262935-M25704350076 262935-M25704350078 262935-M25704350080 262935-M25704350081 262935-M25704350089 262935-M25704350097 262935-M25704350098 262935-M25621800022 262935-M25621800011 262935-M25621800001 262935-M25704350035 262935-M25621800028 262935-M25621800029 262935-M25621800030 262935-M25621800033 262935-M25621800036 262935-M25621800057 262935-M25621160074 262935-M25621160107 262935-M25621160119 262935-M25621160120 262935-M25621160123 262935-M25621160135 262935-M25621160149 262935-M25704350020 262935-M25621160150 262935-M25621170033 262935-M25621170034 262935-M25621170035 262935-M25621170042 262935-M25621170053 262935-M25621170076 262935-M25621170083 262935-M25704350001 262935-M25704350009 262935-M25704350016 262935-M25704350018 262935-M25704350032 262935-M25704350034 262935-M25621170026 262935-M25704350021 262935-M25704350029 262935-M25704350036 262935-M25621160028 262935-M25621160059 262935-M25621160095 262935-M25621170024 262935-M25621800018 262935-M25621800041 262935-M25621160014 262935-M25621160022 262935-M25621160031 262935-M25621160104 262935-M25621160108 262935-M25621160126 262935-M25621160134 262935-M25621160138 262935-M25621160143 262935-M25621160011 262935-M25621160016 262935-M25621160023 262935-M25621160024 262935-M25621160061 262935-M25621160069 262935-M25621160091 262935-M25621160002 262935-M25621160026 262935-M25621160056 262935-M25621160068 262935-M25621160101 262935-M25621160114 262935-M25621160118 262935-M25621160008 262935-M25621160060 262935-M25621160063 262935-M25621160066 262935-M25621160071 262935-M25621160087 262935-M25621160098 262935-M25621160105 262935-M25621160115 262935-M25621160142 262935-M25621160033 262935-M25621160034 262935-M25621160100 262935-M25621160103 262935-M25621160110 262935-M25621160116 262935-M25621170005 262935-M25621160144 262935-M25621160093 262935-M25621160106 262935-M25621160133 262935-M25621170037 262935-M25621170058 262935-M25621170012 262935-M25621170087 262935-M25621170057 262935-M25621170059 262935-M25621170075 262935-M25621170089 262935-M25621160072 262935-M25621160089 262935-M25621160137 262935-M25621170069 262935-M25621160065 262935-M25621170086 262935-M25621160073 262935-M25621160102 262935-M25621160112 262935-M25621160117 262935-M25621170085 262935-M25621160109 262935-M25621800017 262935-M25621160012 262935-M25621160030 262935-M25621160042 262935-M25621160092 262935-M25621160146 262935-M25621170003 262935-M25621170013 262935-M25621170014 262935-M25621170015 262935-M25621170021 262935-M25621170027 262935-M25621170029 262935-M25704350037 262935-M25704350038 262935-M25704350047 262935-M25704350048 262935-M25704350073 262935-M25704350093 262935-M25704350096 262935-M25621170047 262935-M25621170048 262935-M25621170062 262935-M25621170071 262935-M25621170072 262935-M25621800005 262935-M25621800012 262935-M25621800015 262935-M25621800020 262935-M25621800023 262935-M25621800026 262935-M25621800035 262935-M25621800044 262935-M25621800058 262935-M25621800059 262935-M25704350017 262935-M25704350023 262935-M25704350045 262935-M25704350066 262935-M25704350068 262935-M25704350070 262935-M25704350071 262935-M25704350079 262935-M25704350082 262935-M25704350084 262935-M25704350085 262935-M25704350042 262935-M25621800007 262935-M25621170074 262935-M25621170081 262935-M25704350074 262935-M25704350075 262935-M25704350094 262935-M25704350100 262935-M25621160003 262935-M25621160006 262935-M25621160010 262935-M25621160013 262935-M25621160036 262935-M25621160052 262935-M25621160067 262935-M25621160075 262935-M25621160077 262935-M25704350072 262935-M25621160082 262935-M25621160140 262935-M25621170010 262935-M25621170022 262935-M25621170030 262935-M25621170041 262935-M25621160001 262935-M25621160020 262935-M25621160025 262935-M25621160046 262935-M25621160050 262935-M25621160053 262935-M25621160054 262935-M25621160078 262935-M25621160131 262935-M25704350069 262935-M25704350061 262935-M25704350060 262935-M25621160048 262935-M25621160058 262935-M25621160081 262935-M25621160097 262935-M25621160121 262935-M25621160145 262935-M25621170007 262935-M25621170008 262935-M25621170017 262935-M25621170018 262935-M25621170019 262935-M25621170025 262935-M25621170043 262935-M25621170052 262935-M25621170067 262935-M25621170068 262935-M25621170082 262935-M25621800008 262935-M25621800009 262935-M25621800010 262935-M25621800019 262935-M25621800027 262935-M25621800038 262935-M25621800039 262935-M25621800054 262935-M25621800060 262935-M25704350013 262935-M25704350056 262935-M25704350058 262935-M25621160079 262935-M25621160084 262935-M25621160085 262935-M25621170049 262935-M25704350010 262935-M25704350011 262935-M25704350019 262935-M25704350026 262935-M25704350027 262935-M25704350028 262935-M25704350041 262935-M25621160136 262935-M25621160141 262935-M25621170001 262935-M25621170009 262935-M25621170011 262935-M25621170023 262935-M25704350063 262935-M25704350077 262935-M25704350090 262935-M25704350091 262935-M25704350099 262935-M25621170065 262935-M25621170070 262935-M25621170073 262935-M25621170088 262935-M25621800004 262935-M25621800014 262935-M25621170031 262935-M25621170032 262935-M25621170051 262935-M25621170064 262935-M25621170063 262935-M25621160029 262935-M25621170039 262935-M25621170020 262935-M25621170050 262935-M25621170066 262935-M25621170080 262935-M25621800002 262935-M25621800025 262935-M25621800032 262935-M25621160032 262935-M25621160041 262935-M25621160049 262935-M25621160051 262935-M25621160080 262935-M25621160083 262935-M25621160086 262935-M25621160090 262935-M25621800047 262935-M25621800050 262935-M25621800053 262935-M25621800056 262935-M25704350002 262935-M25704350004 262935-M25704350005 262935-M25621160096 262935-M25621160122 262935-M25621160148 262935-M25621160094 262935-M25621160124 262935-M25621160128 262935-M25621160130 262935-M25621170002 262935-M25621170038 262935-M25621160021 262935-M25621170040 EDAN Model: X12 UDI-DI code: 06944413810945 Serial/Numbers: 261594-M21116660002 261594-M21116670003 261594-M21116670002 261594-M21116660003 261594-M21116660007 261594-M21116660006 261594-M21116660001 261594-M21116670005 261594-M21116660010 261594-M21116660004 261594-M21116670001 261594-M21116660008 261594-M21116670004 261594-M21116660009 261594-M21116660005 261594-K21200340002 261594-M21208600001 261594-K21200340001 261594-M21208610001 261594-M21303640004 261594-M21303640002 261594-M21301850003 261594-M21303640001 261594-M21303640003 261594-M21301850002 261594-M21303640005 261594-M21301850001 261594-M21714520004 261594-M21714520003 261594-M21714520001 261594-M21714520005 261594-M21714520002 261594-M21C04550004 261594-M21C04550014 261594-M21C04550008 261594-M21C04550003 261594-M21C04550005 261594-M21C04550012 261594-M21C04550002 261594-M21C04550011 261594-M21C04550009 261594-M21C04550001 261594-M21C04550015 261594-M21C04550013 261594-M21C04550006 261594-M21C04550010 261594-M21C04550007 261594-M22105770003 261594-M22105770004 261594-M22105770005 261594-M22105770001 261594-M22105770002 261594-M22209760004 261594-M22209760002 261594-M22209760005 261594-M22209760003 261594-M22209760001 261594-M22405460003 261594-M22405460001 261594-M22405460004 261594-M22405460005 261594-M22405460002 261594-M22507880001 261594-M22511340008 261594-M22511340009 261594-M22511340010 261594-M22511330023 261594-M22511330024 261594-M22511340005 261594-M22511340007 261594-M22511330006 261594-M22511330022 261594-M22511330025 261594-M22511340006 261594-M22511340002 261594-M22511330021 261594-M22511330020 261594-M22511330016 261594-M22511340001 261594-M22511330014 261594-M22511330017 261594-M22511330015 261594-M22511330012 261594-M22511330010 261594-M22511330005 261594-M22511330002 261594-M22511330004 261594-M22511340004 261594-M22511340003 261594-M22511330019 261594-M22511330013 261594-M22511330011 261594-M22511330009 261594-M22511330007 261594-M22511330003 261594-M22511330001 261594-M22511330018 261594-M22511330008 261594-M22B11580001 261594-M23710510001 261594-M23710510002 261594-M23813080006 261594-M23813080001 261594-M23813080014 261594-M23813080012 261594-M23813080013 261594-M23813080010 261594-M23813080009 261594-M23813080008 261594-M23813080007 261594-M23813080002 261594-M23813080015 261594-M23813080004 261594-M23813080003 261594-M23813080005 261594-M23813080011 261594-M24306040004 261594-M24306040005 261594-M24306040003 261594-M24306040002 261594-M24306040001 261594-M24306030001 261594-M24412190001 261594-M24412190002 261594-M24412190003 261594-M24412190004 261594-M24412190005 261594-M24511450005 261594-M24511450003 261594-M24511450004 261594-M24511460002 261594-M24511460001 261594-M24511460005 261594-M24511460004 261594-M24511460003 261594-M24511450002 261594-M24511450001 261594-M20610040006 261594-M20610040010 261594-M20610040002 261594-M20610040004 261594-M20610040001 261594-M20610040009 261594-M20610040003 261594-M20610040008 261594-M20610040005 261594-M20610040007 261594-M20611390019 261594-M20611380012 261594-M20611370008 261594-M20611360011 261594-M20611390005 261594-M20611390017 261594-M20611380001 261594-M20611380004 261594-M20611370001 261594-M20611370010 261594-M20611360008 261594-M20611390009 261594-M20611390010 261594-M20611390011 261594-M20611380009 261594-M20611380010 261594-M20611360002 261594-M20611360003 261594-M20611360007 261594-M20611360009 261594-M20611390012 261594-M20611390001 261594-M20611390008 261594-M20611360015 261594-M20611390002 261594-M20611390015 261594-M20611390020 261594-M20611380002 261594-M20611380011 261594-M20611380013 261594-M20611380014 261594-M20611360005 261594-M20611390004 261594-M20611380015 261594-M20611360001 261594-M20611360012 261594-M20611360013 261594-M20611390007 261594-M20611390013 261594-M20611370005 261594-M20611370006 261594-M20611370009 261594-M20611360004 261594-M20611360006 261594-M20611360010 261594-M20611390006 261594-M20611390014 261594-M20611370007 261594-M20611380007 261594-M20611380006 261594-M20611370004 261594-M20611370002 261594-M20611380008 261594-M20611390016 261594-M20611390018 261594-M20611360014 261594-M20611370003 261594-M20611380005 261594-M20611380003 261594-M20611390003 261594-M21108850020 261594-M21108850019 261594-M21108850018 261594-M21108850015 261594-M21108850017 261594-M21108850013 261594-M21108850009 261594-M21108850008 261594-M21108850010 261594-M21108850006 261594-M21108850007 261594-M21108850014 261594-M21108850012 261594-M21108850011 261594-M21108850016 261594-M21108850004 261594-M21108850002 261594-M21108850001 261594-M21108850005 261594-M21108850003 261594-M21113770018 261594-M21113760011 261594-M21113760009 261594-M21113760031 261594-M21113760027 261594-M21113760019 261594-M21113760010 261594-M21113760039 261594-M21113760024 261594-M21113760038 261594-M21113760036 261594-M21113760035 261594-M21113760033 261594-M21113760021 261594-M21113760006 261594-M21113760040 261594-M21113760015 261594-M21113760037 261594-M21113760030 261594-M21113760032 261594-M21113760034 261594-M21113770016 261594-M21113770012 261594-M21113770010 261594-M21113770001 261594-M21113770038 261594-M21113770017 261594-M21113770013 261594-M21113770015 261594-M21113770011 261594-M21113770003 261594-M21113770014 261594-M21113770008 261594-M21113770006 261594-M21113770002 261594-M21113770040 261594-M21113770004 261594-M21113770009 261594-M21113770005 261594-M21113770007 261594-M21113770022 261594-M21113770026 261594-M21113760018 261594-M21113760013 261594-M21113760016 261594-M21113760014 261594-M21113760012 261594-M21113760028 261594-M21113760025 261594-M21113760023 261594-M21113760029 261594-M21113760020 261594-M21113760017 261594-M21113760004 261594-M21113760002 261594-M21113760001 261594-M21113760003 261594-M21113760008 261594-M21113770020 261594-M21113770036 261594-M21113770032 261594-M21113770030 261594-M21113770029 261594-M21113770025 261594-M21113770039 261594-M21113770021 261594-M21113770027 261594-M21113770023 261594-M21113770019 261594-M21113770034 261594-M21113760026 261594-M21113760022 261594-M21113760007 261594-M21113760005 261594-M21113770028 261594-M21113770024 261594-M21113770033 261594-M21113770037 261594-M21113770035 261594-M21113770031 261594-M21400960007 261594-M21400960014 261594-M21400960008 261594-M21400960002 261594-M21400960003 261594-M21400960016 261594-M21400960011 261594-M21400960010 261594-M21400960015 261594-M21400960020 261594-M21400960012 261594-M21400960017 261594-M21400960004 261594-M21400960018 261594-M21400960001 261594-M21400960006 261594-M21400960019 261594-M21400960005 261594-M21400960013 261594-M21400960009 261594-M21409430005 261594-M21409430013 261594-M21409430018 261594-M21409430008 261594-M21409430017 261594-M21409430014 261594-M21409430022 261594-M21409430010 261594-M21409430024 261594-M21409430004 261594-M21409430006 261594-M21409430011 261594-M21409430019 261594-M21409430020 261594-M21409430002 261594-M21409430009 261594-M21409430025 261594-M21409430007 261594-M21409430001 261594-M21409430003 261594-M21409430012 261594-M21409430016 261594-M21409430015 261594-M21409430021 261594-M21409430023 261594-M21610140023 261594-M21610140006 261594-M21610140020 261594-M21610140003 261594-M21610140004 261594-M21610140005 261594-M21610140010 261594-M21610140007 261594-M21610140008 261594-M21610140009 261594-M21610140012 261594-M21610140024 261594-M21610140025 261594-M21610140001 261594-M21610140002 261594-M21610140019 261594-M21610140017 261594-M21610140018 261594-M21610140021 261594-M21610140022 261594-M21610140016 261594-M21610140013 261594-M21610140014 261594-M21610140015 261594-M21610140011 261594-M21903070019 261594-M21903070009 261594-M21903070021 261594-M21903070022 261594-M21903070011 261594-M21903070002 261594-M21903070006 261594-M21903070010 261594-M21903070016 261594-M21903070020 261594-M21903070003 261594-M21903070015 261594-M21903070007 261594-M21903070017 261594-M21903070004 261594-M21903070014 261594-M21903070018 261594-M21903070001 261594-M21903070005 261594-M21903070013 261594-M21903070012 261594-M21903070008 261594-M21903320001 261594-M21903320007 261594-M21903320003 261594-M21903320016 261594-M21903320002 261594-M21903320011 261594-M21903320012 261594-M21903320006 261594-M21903320020 261594-M21903320019 261594-M21903320018 261594-M21903320017 261594-M21903320015 261594-M21903320014 261594-M21903320013 261594-M21903320010 261594-M21903320009 261594-M21903320008 261594-M21903320005 261594-M21903320004 261594-M21B01790017 261594-M21B01790007 261594-M21B01790006 261594-M21B01790012 261594-M21B01790001 261594-M21B01790003 261594-M21B01790002 261594-M21B01790011 261594-M21B01790004 261594-M21B01790016 261594-M21B01790021 261594-M21B01790022 261594-M21B01790023 261594-M21B01790024 261594-M21B01790013 261594-M21B01790014 261594-M21B01790018 261594-M21B01790019 261594-M21B01790020 261594-M21B01790005 261594-M21B01790025 261594-M21B01790015 261594-M21B01790010 261594-M21B01790009 261594-M21B01790008 261594-M22610810008 261594-M22610810002 261594-M22610810001 261594-M22610810003 261594-M22610810007 261594-M22610810004 261594-M22610810009 261594-M22610810006 261594-M22610810005 261594-M22610810010 261594-M22903020010 261594-M22903020007 261594-M22903020009 261594-M22903020004 261594-M22903020005 261594-M22903020008 261594-M22903020002 261594-M22903020001 261594-M22903020003 261594-M22903020006 261594-K22A00440001 261594-M22B07210001 261594-M22B07210015 261594-M22B07210016 261594-M22B07210019 261594-M22B07210017 261594-M22B07210010 261594-M22B07210005 261594-M22B07210014 261594-M22B07210008 261594-M22B07210004 261594-M22B07210018 261594-M22B07210002 261594-M22B07210003 261594-M22B07210006 261594-M22B07210012 261594-M22B07210013 261594-M22B07210007 261594-M22B07210009 261594-M22B07210011 261594-M22B07210020 261594-M22B13640001 261594-M22B13640004 261594-M22B13640008 261594-M22B13640009 261594-M22B13640025 261594-M22B13650013 261594-M22B13650025 261594-M22B13650012 261594-M22B13640023 261594-M22B13640012 261594-M22B13640013 261594-M22B13640017 261594-M22B13640020 261594-M22B13650005 261594-M22B13650017 261594-M22B13650019 261594-M22B13650022 261594-M22B13640019 261594-M22B13640018 261594-M22B13650024 261594-M22B13640005 261594-M22B13640010 261594-M22B13640016 261594-M22B13650018 261594-M22B13650023 261594-M22B13650016 261594-M22B13640003 261594-M22B13650002 261594-M22B13650015 261594-M22B13650003 261594-M22B13650006 261594-M22B13650021 261594-M22B13650004 261594-M22B13650008 261594-M22B13650011 261594-M22B13650014 261594-M22B13650020 261594-M22B13650001 261594-M22B13650007 261594-M22B13650009 261594-M22B13640002 261594-M22B13640015 261594-M22B13640021 261594-M22B13640024 261594-M22B13640014 261594-M22B13640006 261594-M22B13640011 261594-M22B13640007 261594-M22B13650010 261594-M22B13640022 261594-M23222250003 261594-M23222250002 261594-M23222250001 261594-M23222250006 261594-M23222250005 261594-M23222250004 261594-M23222250010 261594-M23222250007 261594-M23222250009 261594-M23222250008 261594-M23408260002 261594-M23408260012 261594-M23408260006 261594-M23408260003 261594-M23408260007 261594-M23408260013 261594-M23408260014 261594-M23408260004 261594-M23408260008 261594-M23408260015 261594-M23408260009 261594-M23408260010 261594-M23408260011 261594-M23408260001 261594-M23408260005 261594-M23514810004 261594-M23514810012 261594-M23514810008 261594-M23514810017 261594-M23514810020 261594-M23514810011 261594-M23514810006 261594-M23514810002 261594-M23514810010 261594-M23514810016 261594-M23514810018 261594-M23514810007 261594-M23514810001 261594-M23514810014 261594-M23514810015 261594-M23514810009 261594-M23514810003 261594-M23514810013 261594-M23514810005 261594-M23514810019 261594-M23818740001 261594-M23818740004 261594-M23818740005 261594-M23818740003 261594-M23818740002 261594-M23A07310015 261594-M23A07310006 261594-M23A07310019 261594-M23A07310025 261594-M23A07310005 261594-M23A07310001 261594-M23A07310012 261594-M23A07310018 261594-M23A07310022 261594-M23A07310028 261594-M23A07310013 261594-M23A07310026 261594-M23A07310024 261594-M23A07310014 261594-M23A07310017 261594-M23A07310021 261594-M23A07310003 261594-M23A07310016 261594-M23A07310008 261594-M23A07310030 261594-M23A07310023 261594-M23A07310004 261594-M23A07310007 261594-M23A07310010 261594-M23A07310011 261594-M23A07310020 261594-M23A07310027 261594-M23A07310029 261594-M23A07310002 261594-M23A07310009 261594-M24808210006 261594-M24808210001 261594-M24808210005 261594-M24808210009 261594-M24808210004 261594-M24808210003 261594-M24808210002 261594-M24808210008 261594-M24808210007 261594-M24808210010 261594-M24A00740001 261594-M24A00730001 261594-M24A01770001 261594-M24A01770004 261594-M24A01770002 261594-M24A19050016 261594-M24A19050012 261594-M24A19050010 261594-M24A19050006 261594-M24A19060003 261594-M24A19050002 261594-M24A19050003 261594-M24A19050019 261594-M24A19050018 261594-M24A19050009 261594-M24A19050015 261594-M24A19050013 261594-M24A19050004 261594-M24A19050020 261594-M24A19050014 261594-M24A19050008 261594-M24A19060001 261594-M24A19060002 261594-M24A19050017 261594-M24A19050011 261594-M24A19050005 261594-M24A19050007 261594-M24A19050001 261594-M24C12870001
FEI Number 3009499478
Recalling Firm/
Manufacturer
Edan Diagnostics
9833 Pacific Heights Blvd
San Diego CA 92121-4707
For Additional Information ContactChao Lin
858-7503066
Manufacturer Reason
for Recall
As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential cybersecurity issues with various of the firm device.
FDA Determined
Cause 2
Device Design
ActionOn 10/24/2025, the firm emailed notifying customers of a product recall. On 11/06/2025, the firm emailed an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them of labeling issues involving cybersecurity which arose from a communication from the U.S. Food & Drug Administration. -First issue involves an inappropriate publicly routable IP address that was used as an example in the user manual for demonstrating user-configurable IP settings and Second issue involves an inappropriate publicly IP address that was used as the factory default configuration. Customer are instructed to: 1. Immediately examine their inventory and quarantine product subject to correction. 2. If customers may have further distributed affected product, identify their customers and notify them at once of this product correction. The notification to their customers may be enhanced by including the enclosed URGENT MEDICAL DEVICE CORRECTION. 3. For affected product still in their inventory, place copy of the enclosed Notice Before Use and Corrigenda of User Manual into its package before distribution. 4. For affected product that has been distributed or is in use: (1) Confirm that they have received and acknowledged Corrigenda of User Manual. Print out a copy, attach to the User Manual. (2) Confirm the publicly routable IP address 202.114.4.* is NOT used in the User Maintain in the device. If 202.114.4.* is being used, change to appropriate IP settings of device in the Network Setup according to Instruction Instructions for IP Settings Configuration. 5. Complete the attached Acknowledgement and Receipt Form, sign it, and return a scanned copy to the sender of this notification by email. For questions - Contact: EDAN Instruments, Inc. Monday to Friday 09:00-17:00 (UTC+8) Tel:+86-0755-26898326 EDAN Diagnostics, Inc. Monday to Friday 09:00-17:00 (PST) Tel: 1-(858)-750-3066
Quantity in Commerce3824
DistributionWorldwide - U.S. Nationwide distribution in the states of CA, KY, PR, FL, CT, IL, PA, RI, NC, NY, LA, TX, MI and the country of Mexico.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = MHX
-
-