• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Melker Universal Cuffed Emergency Cricothyrotomy Catheter Setsee related information
Date Initiated by FirmMarch 05, 2026
Date PostedApril 06, 2026
Recall Status1 Open3, Classified
Recall NumberZ-1785-2026
Recall Event ID 98495
510(K)NumberK010016 K160200 
Product Classification Needle, emergency airway - Product Code BWC
ProductMelker Universal Cuffed Emergency Cricothyrotomy Catheter Set: Reference Part Number C-TCCSB-500-UNI, Order Number G32193; Reference Part Number C-TCCSB-500-UNI-01, Order Number G47665.
Code Information Reference Part Number C-TCCSB-500-UNI, UDI (01)00827002321939(17)270803(10)NS16189671, Lot Number NS16189671; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270805(10)16191668, Lot Number 16191668; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270807(10)16195472, Lot Number 16195472; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270807(10)16195467, Lot Number 16195467; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270805(10)16191666, Lot Number 16191666; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270805(10)16191677, Lot Number 16191677; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270805(10)16191661, Lot Number 16191661; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270812(10)16200360, Lot Number 16200360; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270731(10)16185929, Lot Number 16185929; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270904(10)16237917, Lot Number 16237917; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270805(10)16191674, Lot Number 16191674; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270805(10)16191671, Lot Number 16191671; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270806(10)16193388, Lot Number 16193388; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270806(10)16193387, Lot Number 16193387; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270805(10)16191667, Lot Number 16191667; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270807(10)16195473, Lot Number 16195473; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270805(10)16191662, Lot Number 16191662; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270805(10)16191689, Lot Number 16191689; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270807(10)16195468, Lot Number 16195468; Reference Part Number C-TCCSB-500-UNI-01, UDI (01)00827002476653(17)270927(10)16271019, Lot Number 16271019.
FEI Number 1820334
Recalling Firm/
Manufacturer
Cook Incorporated
750 N Daniels Way
Bloomington IN 47404-9120
For Additional Information ContactCook Medical Customer Relations Department
812-457-4500
Manufacturer Reason
for Recall
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
FDA Determined
Cause 2
Incorrect or no expiration date
ActionCook Medical issued an URGENT: MEDICAL DEVICE RECALL to its consignee on 3/5 2026 via courier or email. The notice explained the issue, potential risk to health, and requested the following: Actions to be Taken by the Customer 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all further distribution and use of the affected products. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected product(s) on hand. You may return the Form via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the affected devices have been transferred. 5. Please immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.
Quantity in Commerce476 units
DistributionWorldwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = BWC