| Date Initiated by Firm | March 19, 2026 |
| Date Posted | April 29, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2006-2026 |
| Recall Event ID |
98612 |
| 510(K)Number | K190887 |
| Product Classification |
Neurological stereotaxic Instrument - Product Code HAW
|
| Product | Brand Name: Leksell Vantage Arc System
REF: 1053958 |
| Code Information |
Lot Code: Arc System: UDI:07340048304856
Parts of Arc System:
Left Z-Slide: Serial Numbers :UDI: 07340048304818 00001 00625
Right Z-Slide: UDI: 07340048304801/ Serial Numbers: 00001 00675 |
| FEI Number |
1037831
|
Recalling Firm/ Manufacturer |
Elekta, Inc. 400 Perimeter Center Ter NE Ste 50 Atlanta GA 30346-1227
|
| For Additional Information Contact | Elekta Care Support 001-855-6935358 |
Manufacturer Reason for Recall | Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures. |
FDA Determined Cause 2 | Device Design |
| Action | On March 19, 2026 Elekta issued a "Urgent: Medical Device Recall" notification via e-mail. Elekta asked consignees
1. Before each operation, products should be cleaned, sterilized (unless specified otherwise), and checked for proper functionality.
2. Make sure to examine all scales and angles to make sure that they are correctly set and securely fixed before proceeding.
3. Make sure to obtain a second person in the OR room to read the coordinate settings so as to ensure they are correct and check they are securely locked.
4. Before you continue, check that all parts are assembled correctly, firmly attached, and locked in place.
5. Inspect the equipment for visible signs of damage before clinical use. Do not use damaged equipment for any purpose. If you suspect that any part is damaged, contact your Elekta representative immediately.
6. Please post this notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed.
7. Advise the appropriate personnel working with this product, on the content of this letter. If the product has been transferred to another facility, please provide them with a copy of this communication.
|
| Quantity in Commerce | 768 (364Left/364Right) |
| Distribution | US: AR AZ CA CO DC FL GA ID IL KS MA MD ME MO MT NC NE NH NJ NM NV NY OH OK PA SC TX UT VA WA WI WV
OUS: Algeria Australia Austria Azerbaijan Belgium China Colombia Costa Rica Croatia Czech Republic Egypt Estonia Finland France Germany Greece Hong Kong Hungary India Israel Italy Japan Korea, Republic of Kuwait Lithuania Mexico Morocco Netherlands New Zealand Norway Pakistan Panama Poland Portugal Saudi Arabia Singapore Slovenia South Africa Spain Sweden Switzerland Taiwan Turkey United Kingdom
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = HAW
|