• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall ARROW RADIAL ARTERY CATHETERIZATION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall ARROW RADIAL ARTERY CATHETERIZATION KITsee related information
Date Initiated by FirmMarch 11, 2026
Date PostedApril 22, 2026
Recall Status1 Open3, Classified
Recall NumberZ-1911-2026
Recall Event ID 98618
Product Classification Catheter guide wire kit - Product Code OFB
ProductARROW RADIAL ARTERY CATHETERIZATION KIT REF ASK-04220-UCL1 UDI code: (01)10801902159127(17)270331(11)251203(10)33F25J0352 REF ASK-04020-EUH2 UDI code: (01)10801902213379(17)270731(11)251218(10)33F25J0869 REF ASK-04020-FMC UDI code: (01)10801902212099(17)270731(11)251218(10)33F25J0870 REF ASK-04020-MMC UDI code: (01)10801902117769(17)271130(11)251211(10)33F25J0880 The Arrow Arterial Catheterization Device permits access to the peripheral arterial circulation or to other small vessels.
Code Information Product code:ASK-04220-UCL1 UDI code: (01)10801902159127(17)270331(11)251203(10)33F25J0352 Batch Number: 33F25J0352 Product code: ASK-04020-EUH2 UDI code: (01)10801902213379(17)270731(11)251218(10)33F25J0869 Batch Number: 33F25J0869 Product code: ASK-04020-FMC UDI code: (01)10801902212099(17)270731(11)251218(10)33F25J0870 Batch Number: 33F25J0870 Product code: ASK-04020-MMC UDI code: (01)10801902117769(17)271130(11)251211(10)33F25J0880 Batch Number: 33F25J0880
FEI Number 3013162291
Recalling Firm/
Manufacturer
ARROW INTERNATIONAL, LLC
3015 Carrington Mill Blvd
Morrisville NC 27560-5437
Manufacturer Reason
for Recall
Due to receiving a notice from a supplier that the liquid adhesive was incorrectly manufactured
FDA Determined
Cause 2
Process control
ActionOn 03/18/2026, the firm sent via FedEx-2-day an "URGENT MEDICAL DEVICE RECALL" Letter informing customers that Ferndale Laboratories issued a Field Safety Notice (FSN) to inform users of multiple complaints reporting the butyrate tubing of the Mastisol Liquid Adhesive vial cracked during actuation. According to the Ferndale Laboratories FSN, one lot of MASTISOL Liquid Adhesive was exposed to a double sterilization cycle which may have contributed to the butyrate tubing becoming brittle and susceptible to cracking during actuation. Medical Facilities should: 1. Immediately check inventory for product within the scope of this recall and cease use and distribution of affected product and immediately quarantine the affected product. 2. For impacted product, mark the applicable checkbox on the Acknowledgement Form (Appendix 1) and contact Teleflex Customer Service utilizing one of the following methods: a. Fax, including Attn: Customer Service , or email customer service using the point of contact details outlined on Page 3 with a copy of Appendix 1 OR b. Using the online FSN Portal, complete the acknowledgement form virtually using the following link: https://fca.myteleflex.com/en/recall/000606 Note: Option B is solely applicable to customers who received product direct from Teleflex. 3. For impacted product, mark the applicable checkbox on the Acknowledgement Form (Appendix 1) and return the form to Teleflex at the contact details provided. 4. Teleflex (or your local dealer) will issue a credit note upon receipt of the returned affected product. Distributor should: 1. Provide a copy of this recall notice to all customers who have received impacted product. Each customers is then required to complete the Acknowledgement Form and return it to Distributor. 2. Request immediate check of inventory, cease use and distribution of, and quarantine, affected products and returned to Teleflex. For questions-contact customer service at 1-866-396-2111 or email Recalls@telefle
Quantity in Commerce675
DistributionU.S. Nationwide distribution in the states of AZ, CA, GA, MA, ME, MI, OR, PA, and VA.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-