• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Centurion

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Centurionsee related information
Date Initiated by FirmApril 27, 2026
Date PostedJune 12, 2026
Recall Status1 Open3, Classified
Recall NumberZ-2437-2026
Recall Event ID 98951
Product Classification Dressing change tray - Product Code OXQ
ProductCenturion convenience kits: CENTRAL LINE DRESSING SV43 FCP DYNDC2356B DIALYSIS CATHETER MAINTENANCE KIT DT17570 DIALYSIS DRESSING CHANGE KIT DT21370B DRESSING CHANGE TRAY - GOLD DT21115B PORT DRESSING CHANGE KIT DYNDC2815A
Code Information DYNDC2356B UDI-DI 10653160327271 Lot 2025102090 DT17570 UDI-DI 10653160266891 Lot 2025102290 DT21370B UDI-DI 10653160360643 Lot 2025102290 DT21115B UDI-DI 10653160360681 Lot 2025111490 DYNDC2815A UDI-DI 10653160357070 Lot 2025110590
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries, LP
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactHaley Barclay
800-633-5463
Manufacturer Reason
for Recall
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
FDA Determined
Cause 2
Process control
ActionFirm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.
Quantity in Commerce49,654 kits total
DistributionUS Nationwide distribution. OUS distribution pending.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-