 Class 2 Device Recall Medlinesee related information
Date Initiated by FirmApril 27, 2026
Date PostedJune 12, 2026
Recall Status1 Open3, Classified
Recall NumberZ-2439-2026
Recall Event ID 98951
510(K)NumberK213481 
Product Classification General surgery tray - Product Code LRO
ProductMedline convenience kits: 20CM CVC INSERTION KIT ECVC6805B 2L 8.5FR 16CM CVC INSERTION KIT ECVC8055A VANTEX PI CVC 7F, 3L, 20CM BUNDLE ECVC6915A ECVC6985B
Code Information ECVC6805B UDI-DI 10653160387428 Lot 2025120890 ECVC8055A UDI-DI 10653160387848 Lot 2025112690 ECVC6915A UDI-DI 10653160389439 Lot 2025102790 ECVC6985B UDI-DI 10653160388272 Lot 2025103190
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries, LP
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactHaley Barclay
800-633-5463
Manufacturer Reason
for Recall
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
FDA Determined
Cause 2
Process control
ActionFirm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.
Quantity in Commerce49,654 kits total
DistributionUS Nationwide distribution. OUS distribution pending.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = LRO
-
-