| Date Initiated by Firm | April 27, 2026 |
| Date Posted | June 12, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2445-2026 |
| Recall Event ID |
98951 |
| Product Classification |
Central venous catheter tray - Product Code OFF
|
| Product | Medline convenience kits:
ADD-A-CATH ARTERIAL INSERTION
ARTNC185
PED SCHOOL AGE PICC DRSG TRAY
DYNDC3291 |
| Code Information |
ARTNC185
UDI-DI 10195327640729
Lot 25JBU463
DYNDC3291
UDI-DI 10195327334741
Lot 25KBH082
|
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
Medline Industries, LP 3 Lakes Dr Northfield IL 60093-2753
|
| For Additional Information Contact | Haley Barclay 800-633-5463 |
Manufacturer Reason for Recall | Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application. |
FDA Determined Cause 2 | Process control |
| Action | Firm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned. |
| Quantity in Commerce | 49,654 kits total |
| Distribution | US Nationwide distribution. OUS distribution pending. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|