| | Class 2 Device Recall Olympus High Flow Insufflation Unit UHI4 |  |
| Date Initiated by Firm | June 16, 2026 |
| Date Posted | July 22, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2776-2026 |
| Recall Event ID |
99109 |
| 510(K)Number | K243527 |
| Product Classification |
Insufflator, laparoscopic - Product Code HIF
|
| Product | Olympus High Flow Insufflation Unit - UHI-4. Model Number: UHI-4 |
| Code Information |
Model Number: UHI-4. UDI: 04953170324147, 04953170467219. All Serial Numbers.
|
| FEI Number |
2429304
|
Recalling Firm/ Manufacturer |
Olympus Corporation of the Americas
3500 Corporate Pkwy
Center Valley
PA
18034-8229
|
| For Additional Information Contact | Ashley Mitch 484-5531029 |
Manufacturer Reason for Recall | Potential for unexpected instances where the front panel display turns off unexpectedly, and air delivery (insufflation) ceases. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Olympus notified consignees on about 06/16/2026 via letter. Consignees were informed that they many continue to use their UHI-4 until the offset adjustment is performed. However, as noted in the Instructions for Use (IFU), prepare a spare UHI-4 unit as a backup to ensure uninterrupted operation can be completed without complication in the case of a malfunction. Actions to take included to carefully read the notification letter and ensure all personnel are completely knowledgeable on the content of the notification, an Olympus field service representative will contact you to schedule a mutually convenient time to perform the offset adjustment, acknowledge receipt of the notification through the Olympus recall web portal, and forward the notification to other users who may have the affected products if you have further distributed it. |
| Quantity in Commerce | 3,087 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, El Salvador, Germany, Honduras, Hong Kong, India, Ireland, Japan, Korea, Malaysia, Mexico, Myanmar, New Zealand, Paraguay, Peru, Russia, Singapore, Sri Lanka, Taiwan, Thailand, United Kingdom, Uruguay, Vietnam.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = HIF
|
|
|
|