| | Class 2 Device Recall Medline |  |
| Date Initiated by Firm | May 28, 2026 |
| Date Posted | June 26, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2606-2026 |
| Recall Event ID |
99209 |
| Product Classification |
Cardiac catheterization kit - Product Code OES
|
| Product | Medline Convenience kits containing recalled Swan-Ganz Catheters
RIGHT HEART CATH PACK CTX
DYNJ68530B |
| Code Information |
UDI-DI 10195327345044
Lots
23DBI807
23FBG708
23GBK642
23KBO477
24BBG198
24FBN889
24GBJ879
24HBR045
24LBT492
25ABL957
25BBN572
25GBT734
25HBQ817
25LBM617
26BBK766
26DBE720
|
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
Medline Industries, LP 3 Lakes Dr Northfield IL 60093-2753
|
| For Additional Information Contact | Haley Barclay 800-633-5463 |
Manufacturer Reason for Recall | Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss. |
FDA Determined Cause 2 | Process change control |
| Action | On May 28, 2026, the firm notified customers via letters titled "URGENT MEDICAL DEVICE CORRECTION."
Customers were referred to BD's recall notice, which provides advice for users and instructions to mitigate hub breakage issues. Product return is not required.
If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com. |
| Quantity in Commerce | 9,272 kits |
| Distribution | US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|