| | Class 2 Device Recall EOPA |  |
| Date Initiated by Firm | June 24, 2026 |
| Date Posted | July 27, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2807-2026 |
| Recall Event ID |
99320 |
| 510(K)Number | K031037 |
| Product Classification |
Catheter, cannula and tubing, vascular, cardiopulmonary bypass - Product Code DWF
|
| Product | EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77522 |
| Code Information |
UDI-DI: 00199150023288,
Lot Numbers: 0233286436;
UDI-DI: 20199150023282,
Lot Numbers: 0232709428
0232744924
0232745061
0232745092
0232798556
0232862763
0232862778
0232862794
0232862798
0232903216
0232903291
0232903383
0232990983
0233286318
0233286414
0233286436
0233427120
0233528730
0233553623
0233556386
0233585113
0233585114
0232670802 |
| FEI Number |
1000116158
|
Recalling Firm/ Manufacturer |
Medtronic Perfusion Systems
7611 Northland Dr N
Brooklyn Park
MN
55428-1088
|
| For Additional Information Contact | Medtronic Customer Service 391-9001 |
Manufacturer Reason for Recall | Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | An URGENT: MEDICAL DEVICE RECALL notification letter dated June 2026 was sent to customers beginning 6/24/26.
Customer Actions:
Medtronic requests that you take the following actions:
" Review your inventory for listed products using Attachment A.
" Immediately identify and quarantine all unused, listed product in your inventory.
" Return unused, listed product in your inventory to Medtronic by contacting Customer Service at 1-800-854-3570, Option 1 then Option 4, and referencing this communication to initiate return for credit or replacement product. Your Medtronic sales representative can assist you in the return of affected product as necessary.
" Complete the enclosed Customer Confirmation Form and email to RS.CFQFCA@medtronic.com. This form must be returned even if you do not have any affected product in your possession.
" Please share this notification with others in your organization as appropriate. If the product listed above has been forwarded to another facility, please notify the facility of this Medtronic Urgent Medical Device Recall.
" Please maintain a copy of this communication in your records.
If you have any questions regarding this communication, please contact 1-800-854-3570, Option 1, then Option 4. |
| Quantity in Commerce | 11,720 units |
| Distribution | Worldwide - US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|