| | Class 2 Device Recall Medline medical convenience kits |  |
| Date Initiated by Firm | June 11, 2026 |
| Date Posted | July 23, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2788-2026 |
| Recall Event ID |
99329 |
| Product Classification |
Cardiovascular procedure kit - Product Code OEZ
|
| Product | Medline medical convenience kits labeled as:
1) KIT PEDIATRIC OPEN HEART MOSES, Medline Kit SKU DYNJ910103B;
2) BASIC HEART CDS, Medline Kit SKU DYNJ910961A. |
| Code Information |
Medline Kit SKU DYNJ910103B, UDI/DI 10198459346873 (each) 40198459346874 (case), Lot Number 25FBR237;
Medline Kit SKU DYNJ910961A, UDI/DI 10198459213342 (each) 40198459213343 (case), Lot Number 25ABM761;
Medline Kit SKU DYNJ910961A, UDI/DI 10198459213342 (each) 40198459213343 (case), Lot Number 25BBD107;
Medline Kit SKU DYNJ910961A, UDI/DI 10198459213342 (each) 40198459213343 (case), Lot Number 25FBD534;
Medline Kit SKU DYNJ910961A, UDI/DI 10198459213342 (each) 40198459213343 (case), Lot Number 25FBL772.
|
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
Medline Industries, LP
3 Lakes Dr
Northfield
IL
60093-2753
|
| For Additional Information Contact | Haley Barclay 866-359-1704 |
Manufacturer Reason for Recall | Medline Industries, LP is issuing a recall for specific Medline convenience kits containing BD PurPrep 26 mL which may be contaminated with low levels of Bacillus species. |
FDA Determined Cause 2 | Process change control |
| Action | Medline Industries, LP issued an URGENT MEDICAL DEVICE RECALL notice to its consignees via email and USPS first class mail. The notice explained the issue with the device, potential harms, and requested the following:
REQUIRED ACTION:
1. Immediately check stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product.
2. Please use the link and the information below to complete the response form. Please list the quantity of affected product in inventory on the form. Even if there is no affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com
3. Upon receipt of the response form, the consignee account will receive over-labels to place on affected inventory, with instructions for the staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable.
Distributors and those who transferred the affected product were directed to notify those to which the product was distributed or transferred.
For contact the Recall Department at 866-359-1704 or recalls@medline.com. |
| Quantity in Commerce | 155 kits |
| Distribution | US: CT, FL, KS, OH, OR, NY, MD, MT, VA, MA. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|