| | Class 2 Device Recall Medline Convenience Kits |  |
| Date Initiated by Firm | June 12, 2026 |
| Date Posted | August 31, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-3055-2026 |
| Recall Event ID |
99334 |
| Product Classification |
Tray, surgical - Product Code LRP
|
| Product | Medline Convenience Kits:
1) ARTERIAL LINE INSERTION KIT, Model Number: ART1295 |
| Code Information |
1) ART1295, UDI-DI: 10653160380733(each), 00653160380736(case), Lot Number: 2024092790;
2) ART1295, UDI-DI: 10653160380733(each), 00653160380736(case), Lot Number: 2025042590;
3) ART1295, UDI-DI: 10653160380733(each), 00653160380736(case), Lot Number: 2025070290 |
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
Medline Industries, LP
3 Lakes Dr
Northfield
IL
60093-2753
|
| For Additional Information Contact | Haley Barclay 866-359-1704 |
Manufacturer Reason for Recall | The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | An URGENT MEDICAL DEVICE RECALL notification letter dated 6/12/25 was sent to customers.
REQUIRED ACTION:
1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product.
2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form.
Website link: https://recalls.medline.com Recall Reference #: R-26-097-FGX1
Recall Code:
3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable.
4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product.
If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com. |
| Quantity in Commerce | 1840 kits |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|