| | Class 2 Device Recall Halyard |  |
| Date Initiated by Firm | May 19, 2026 |
| Date Posted | August 28, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-3045-2026 |
| Recall Event ID |
99469 |
| Product Classification |
Neurological tray - Product Code OJG
|
| Product | 1) NEURO, (kit) ACNE73-01;
2) NEURO, (kit) ACNE73-03;
3) CRANI, (kit) AKCR19-01;
4) ORTHO SPINE PACK, (kit) ARMC017;
5) BMHM NEURO PACK, (kit) BMHT016-05;
6) LUMBAR PACK, (kit) BMHT112-03;
7) NEURO LAMINECTOMY PACK, (kit) DGNE75;
8) NEURO LAMINECTOMY PACK, (kit) DGNE75-02;
9) DRAH NEURO LAMINECTOMY SNEDHDN,(kit) DRNE75;
10) DRAH NEURO LAMINECTOMY SNEDHDN, (kit) DRNE75-01;
11) CRANIOSYNOSTOSIS CUSTOM SNE23C, (kit) DUCR36-01;
12) CRANIOSYNOSTOSIS CUSTOM SNE23C, (kit) DUCR36-02;
13) NEURO CRANIOTOMY PK SNE23C1116, (kit) DUNC42-01;
14) PEDS NEURO PACK SNE2301147, (kit) DUNE88-01;
15) NEURO SPINE PACK SNE23LM110, (kit) DUSP98-01;
16) NEURO SPINE PACK SNE23LM110, (kit) DUSP98-02;
17) CRANI PACK 712638, (kit) EFCR86-01;
18) SPINE PACK 712040, (kit) EFSP30-01;
19) SPINE PACK 712040, (kit) EFSP30-02;
20) CRANIOTOMY TRAY, (kit) EURO009-09;
21) CRANI PACK, (kit) IHCR36-01;
22) NEURO PACK, (kit) IHNE35-01;
23) CRANIOTOMY PACK, (kit) JACK410-11;
24) BASIC NEURO PACK, (kit) MAMC275-06;
25) BASIC NEURO PACK, (kit) MAMC275-07;
26) SPINE CRANI PACK, (kit) MCSP92-01;
27) CRANIOTOMY PACK, (kit) MECP05R;
28) SPINE TRAY GLOBALMICROSURG, (kit) MEDF217-02;
29) NEURO BASIC PACK, (kit) MENE06T;
30) CRANIAL PACK - 205992, (kit) MHCP75-01;
31) NEURO PACK, (kit) MMOK007-03;
32) NEURO PACK, (kit) MMTN012-04;
33) CRANIOTOMY PACK SNECGCRM11, (kit) MSMC108-01;
34) NEURO ANT CERV LAMI ADD A PACK, (kit) NAMC413-07;
35) STRL NEURO LUMB LAMI ADD A PK, (kit) NAMC433-05;
36) NEURO CRANI ADD A PACK, (kit) NAMC434-08;
37) NEURO PACK, (kit) NWHR010-03;
38) CRANIOTOMY PACK OSBORN 176518, (kit) OSCR20-01;
39) NEURO PACK, (kit) RDLD001-14;
40) CRANI PACK, (kit) RDLD008-14;
41) NEURO LUMBAR CERVICAL PACK,(kit) SAMM031-18;
42) CRANIOTOMY PACK, (kit) SELF111-02;
43) PELHAM ACDF, (kit) SGAC31-02;
44) PELHAM ACDF, (kit) SGAC31-03;
45) PELHAM MINA SPINE, (kit) SGSP11;
46) PELHAM MINA SPINE, (kit) SGSP11-03;
47) PELHAM SPINE, (kit) SGSP44-02;
48) PELHAM SPINE, (kit) SGSP44-03;
49) SPINE, (kit) SGSP91;
50) SPINE, (kit) SGSP91-02;
51) SPINE, (kit) SGSP91-03;
52) NEURO SPINAL TRAY SHANDS JACKS, (kit) SHND416-02;
53) DR. DELASOTTA NEURO NECK, (kit) SHOR003-15;
54) CRANI PACK, (kit) SLCN72-01;
55) CRANI PACK, (kit) SLCN72-02;
56) CRANI PACK, (kit) SLCN72-03;
57) CRANI PACK - 208400, (kit) UDCP57-01;
58) NEURO TRAY-LF, (kit) UMIA104-11;
59) CRANIOTOMY PACK, (kit) USMA005;
60) CERVICAL NEURO PACK PGYBK, (kit) VAML012-07;
61) LUMBAR NEURO PACK PGYBK, (kit) VAML028-06;
62) NASAL SINUS, (kit) VAPX009-02;
63) LUMBAR LAMINECTOMY TRAY, (kit) VMCC024-14;
64) VA LUMBAR LAMINECTOMY, (kit) VMED009-05;
65) ANTERIOR CERVICAL PACK, (kit) WEAC53-01;
66) CRANI PACK, (kit) WECR41-01;
67) LUMBAR LAMINECTOMY PACK,67) LUMBAR LAMINECTOMY PACK, Model/Item Number (kit) WELL73-01;
68) LUMBAR LAMINECTOMY, (kit) WKMC114-03;
69) CRANIOTOMY PK, (kit) WKMC125-03;
70) DISKECTOMY PACK, (kit) WRNM030-21. |
| Code Information |
Model/Item Number (kit) ACNE73-01, UDI/DI 10809160458532, Kit Lot Number 1649716;
Model/Item Number (kit) ACNE73-01, UDI/DI 10809160458532, Kit Lot Number 1666559;
Model/Item Number (kit) ACNE73-01, UDI/DI 10809160458532, Kit Lot Number 1649717;
Model/Item Number (kit) ACNE73-03, UDI/DI 10809160513996, Kit Lot Number 1696563;
Model/Item Number (kit) AKCR19-01, UDI/DI 10809160459515, Kit Lot Number 1650945;
Model/Item Number (kit) AKCR19-01, UDI/DI 10809160459515, Kit Lot Number 1666564;
Model/Item Number (kit) AKCR19-01, UDI/DI 10809160459515, Kit Lot Number 1664544;
Model/Item Number (kit) AKCR19-01, UDI/DI 10809160459515, Kit Lot Number 1666978;
Model/Item Number (kit) ARMC017, UDI/DI 10809160506455, Kit Lot Number 1685117;
Model/Item Number (kit) ARMC017, UDI/DI 10809160506455, Kit Lot Number 1688043;
Model/Item Number (kit) ARMC017, UDI/DI 10809160506455, Kit Lot Number 1688044;
Model/Item Number (kit) BMHT016-05, UDI/DI 10809160468821, Kit Lot Number 1666733;
Model/Item Number (kit) BMHT016-05, UDI/DI 10809160468821, Kit Lot Number 1659950;
Model/Item Number (kit) BMHT112-03, UDI/DI 10809160457634, Kit Lot Number 1649141;
Model/Item Number (kit) BMHT112-03, UDI/DI 10809160404430, Kit Lot Number 1606455;
Model/Item Number (kit) DGNE75, UDI/DI 10809160468104, Kit Lot Number 1659330;
Model/Item Number (kit) DGNE75-02, UDI/DI 10809160510001, Kit Lot Number 1688056;
Model/Item Number (kit) DRNE75, UDI/DI 10809160467190, Kit Lot Number 1658535;
Model/Item Number (kit) DRNE75-01, UDI/DI 10809160503843, Kit Lot Number 1681499;
Model/Item Number (kit) DRNE75-01, UDI/DI 10809160503843, Kit Lot Number 1691081;
Model/Item Number (kit) DUCR36-01, UDI/DI 10809160497388, Kit Lot Number 1675499;
Model/Item Number (kit) DUCR36-02, UDI/DI 10809160513538, Kit Lot Number 1691089;
Model/Item Number (kit) DUNC42-01, UDI/DI 10809160497524, Kit Lot Number 1675518;
Model/Item Number (kit) DUNC42-01, UDI/DI 10809160497524, Kit Lot Number 1686203;
Model/Item Number (kit) DUNC42-01, UDI/DI 10809160497524, Kit Lot Number 1686204;
Model/Item Number (kit) DUNE88-01, UDI/DI 10809160498217, Kit Lot Number 1675520;
Model/Item Number (kit) DUSP98-01, UDI/DI 10809160497708, Kit Lot Number 1680111;
Model/Item Number (kit) DUSP98-02, UDI/DI 10809160513392, Kit Lot Number 1691105;
Model/Item Number (kit) EFCR86-01, UDI/DI 10809160442210, Kit Lot Number 1633353;
Model/Item Number (kit) EFCR86-01, UDI/DI 10809160456026, Kit Lot Number 1694281;
Model/Item Number (kit) EFCR86-01, UDI/DI 10809160456026, Kit Lot Number 1691158;
Model/Item Number (kit) EFCR86-01, UDI/DI 10809160442210, Kit Lot Number 374076;
Model/Item Number (kit) EFSP30-01, UDI/DI 10809160456187, Kit Lot Number 1648983;
Model/Item Number (kit) EFSP30-01, UDI/DI 10809160456187, Kit Lot Number 1646886;
Model/Item Number (kit) EFSP30-01, UDI/DI 10809160456187, Kit Lot Number 1649731;
Model/Item Number (kit) EFSP30-01, UDI/DI 10809160456187, Kit Lot Number 1633403;
Model/Item Number (kit) EFSP30-02, UDI/DI 10809160492024, Kit Lot Number 1669417;
Model/Item Number (kit) EFSP30-02, UDI/DI 10809160492024, Kit Lot Number 1678530;
Model/Item Number (kit) EFSP30-02, UDI/DI 10809160492024, Kit Lot Number 1670304;
Model/Item Number (kit) EFSP30-02, UDI/DI 10809160492024, Kit Lot Number 1691551;
Model/Item Number (kit) EFSP30-02, UDI/DI 10809160492024, Kit Lot Number 1669415;
Model/Item Number (kit) EFSP30-02, UDI/DI 10809160492024, Kit Lot Number 1680385;
Model/Item Number (kit) EURO009-09, UDI/DI 10809160414217, Kit Lot Number 1649363;
Model/Item Number (kit) IHCR36-01, UDI/DI 10809160477557, Kit Lot Number 344863;
Model/Item Number (kit) IHNE35-01, UDI/DI 10809160477724, Kit Lot Number 343178;
Model/Item Number (kit) IHNE35-01, UDI/DI 10809160477724, Kit Lot Number 343377;
Model/Item Number (kit) IHNE35-01, UDI/DI 10809160477724, Kit Lot Number 343179;
Model/Item Number (kit) IHNE35-01, UDI/DI 10809160477724, Kit Lot Number 348826;
Model/Item Number (kit) IHNE35-01, UDI/DI 10809160477724, Kit Lot Number 348828;
Model/Item Number (kit) IHNE35-01, UDI/DI 10809160477724, Kit Lot Number 348829;
Model/Item Number (kit) IHNE35-01, UDI/DI 10809160477724, Kit Lot Number 348830;
Model/Item Number (kit) IHNE35-01, UDI/DI 10809160477724, Kit Lot Number 348834;
Model/Item Number (kit) IHNE35-01, UDI/DI 10809160477724, Kit Lot Number 348835;
Model/Item Number (kit) JACK410-11, UDI/DI 10809160464991, Kit Lot Number 1664733;
Model/Item Number (kit) JACK410-11, UDI/DI 10809160464991, Kit Lot Number 1656356;
Model/Item Number (kit) JACK410-11, UDI/DI 10809160464991, Kit Lot Number 1667191;
Model/Item Number (kit) JACK410-11, UDI/DI 10809160464991, Kit Lot Number 838447;
Model/Item Number (kit) JACK410-11, UDI/DI 10809160464991, Kit Lot Number 1664206;
Model/Item Number (kit) JACK410-11, UDI/DI 10809160464991, Kit Lot Number 1664732;
Model/Item Number (kit) MAMC275-06, UDI/DI 10809160421666, Kit Lot Number 1644895;
Model/Item Number (kit) MAMC275-07, UDI/DI 10809160495902, Kit Lot Number 1675387;
Model/Item Number (kit) MCSP92-01, UDI/DI 10809160453124, Kit Lot Number 1652837;
Model/Item Number (kit) MCSP92-01, UDI/DI 10809160442999, Kit Lot Number 1633247;
Model/Item Number (kit) MCSP92-01, UDI/DI 10809160442999, Kit Lot Number 1633983;
Model/Item Number (kit) MCSP92-01, UDI/DI 10809160453124, Kit Lot Number 1670241;
Model/Item Number (kit) MCSP92-01, UDI/DI 10809160442999, Kit Lot Number 1636102;
Model/Item Number (kit) MCSP92-01, UDI/DI 10809160442999, Kit Lot Number 1633985;
Model/Item Number (kit) MCSP92-01, UDI/DI 10809160453124, Kit Lot Number 1670240;
Model/Item Number (kit) MCSP92-01, UDI/DI 10809160514719, Kit Lot Number 1697302;
Model/Item Number (kit) MCSP92-01, UDI/DI 10809160453124, Kit Lot Number 1674543;
Model/Item Number (kit) MCSP92-01, UDI/DI 10809160453124, Kit Lot Number 1670238;
Model/Item Number (kit) MCSP92-01, UDI/DI 10809160453124, Kit Lot Number 1680358;
Model/Item Number (kit) MCSP92-01, UDI/DI 10809160453124, Kit Lot Number 1691128;
Model/Item Number (kit) MCSP92-01, UDI/DI 10809160514719, Kit Lot Number 1700133;
Model/Item Number (kit) MECP05R, UDI/DI 10809160418109, Kit Lot Number 1658933;
Model/Item Number (kit) MECP05R, UDI/DI 10809160418109, Kit Lot Number 1694755;
Model/Item Number (kit) MECP05R, UDI/DI 10809160417331, Kit Lot Number 1671174;
Model/Item Number (kit) MEDF217-02, UDI/DI 10809160382912, Kit Lot Number 1632022;
Model/Item Number (kit) MEDF217-02, UDI/DI 10809160382912, Kit Lot Number 1628605;
Model/Item Number (kit) MEDF217-02, UDI/DI 10809160459706, Kit Lot Number 1651140;
Model/Item Number (kit) MEDF217-02, UDI/DI 10809160459706, Kit Lot Number 1651139;
Model/Item Number (kit) MEDF217-02, UDI/DI 10809160459706, Kit Lot Number 1651141;
Model/Item Number (kit) MEDF217-02, UDI/DI 10809160459706, Kit Lot Number 1651142;
Model/Item Number (kit) MENE06T, UDI/DI 10809160461327, Kit Lot Number 1671525;
Model/Item Number (kit) MENE06T, UDI/DI 10809160461327, Kit Lot Number 1652669;
Model/Item Number (kit) MENE06T, UDI/DI 10809160461327, Kit Lot Number 1654982;
Model/Item Number (kit) MENE06T, UDI/DI 10809160461327, Kit Lot Number 1652667;
Model/Item Number (kit) MENE06T, UDI/DI 10809160461327, Kit Lot Number 1652668;
Model/Item Number (kit) MENE06T, UDI/DI 10809160461327, Kit Lot Number 1646454;
Model/Item Number (kit) MHCP75-01, UDI/DI 10809160479568, Kit Lot Number 345849;
Model/Item Number (kit) MMOK007-03, UDI/DI 10809160460764, Kit Lot Number 1696427;
Model/Item Number (kit) MMTN012-04, UDI/DI 10809160444085, Kit Lot Number 1644099;
Model/Item Number (kit) MSMC108-01, UDI/DI 10809160427972, Kit Lot Number 1623100;
Model/Item Number (kit) MSMC108-01, UDI/DI 10809160427972, Kit Lot Number 1650608;
Model/Item Number (kit) MSMC108-01, UDI/DI 10809160427972, Kit Lot Number 1623102;
Model/Item Number (kit) MSMC108-01, UDI/DI 10809160427972, Kit Lot Number 1639185;
Model/Item Number (kit) MSMC108-01, UDI/DI 10809160427972, Kit Lot Number 1657124;
Model/Item Number (kit) NAMC413-07, UDI/DI 10809160374306, Kit Lot Number 1652550;
Model/Item Number (kit) NAMC433-05, UDI/DI 10809160417584, Kit Lot Number 1629049;
Model/Item Number (kit) NAMC433-05, UDI/DI 10809160417584, Kit Lot Number 1656868;
Model/Item Number (kit) NAMC434-08, UDI/DI 10809160417584, Kit Lot Number 1649106;
Model/Item Number (kit) NWHR010-03, UDI/DI 10809160456637, Kit Lot Number 1681967;
Model/Item Number (kit) NWHR010-03, UDI/DI 10809160456637, Kit Lot Number 1656161;
Model/Item Number (kit) NWHR010-03, UDI/DI 10809160456637, Kit Lot Number 1688028;
Model/Item Number (kit) NWHR010-03, UDI/DI 10809160456637, Kit Lot Number 1675548;
Model/Item Number (kit) NWHR010-03, UDI/DI 10809160456637, Kit Lot Number 1664880;
Model/Item Number (kit) NWHR010-03, UDI/DI 10809160456637, Kit Lot Number 1684189;
Model/Item Number (kit) OSCR20-01, UDI/DI 10809160436370, Kit Lot Number 1664480;
Model/Item Number (kit) RDLD001-14, UDI/DI 10809160467374, Kit Lot Number 1658194;
Model/Item Number (kit) RDLD001-14, UDI/DI 10809160467374, Kit Lot Number 1658423;
Model/Item Number (kit) RDLD001-14, UDI/DI 10809160467374, Kit Lot Number 1658192;
Model/Item Number (kit) RDLD001-14, UDI/DI 10809160467374, Kit Lot Number 1691893;
Model/Item Number (kit) RDLD008-14, UDI/DI 10809160467367, Kit Lot Number 1658206;
Model/Item Number (kit) SAMM031-18, UDI/DI 10809160494363, Kit Lot Number 1672515;
Model/Item Number (kit) SELF111-02, UDI/DI 10809160458808, Kit Lot Number 1651950;
Model/Item Number (kit) SELF111-02, UDI/DI 10809160458808, Kit Lot Number 1652543;
Model/Item Number (kit) SGAC31-02, UDI/DI 10809160469491, Kit Lot Number 1661052;
Model/Item Number (kit) SGAC31-03, UDI/DI 10809160489437, Kit Lot Number 1667346;
Model/Item Number (kit) SGAC31-03, UDI/DI 10809160489437, Kit Lot Number 1673274;
Model/Item Number (kit) SGAC31-03, UDI/DI 10809160489437, Kit Lot Number 1667347;
Model/Item Number (kit) SGAC31-03, UDI/DI 10809160489437, Kit Lot Number 1668002;
Model/Item Number (kit) SGAC31-03, UDI/DI 10809160489437, Kit Lot Number 1691061;
Model/Item Number (kit) SGAC31-03, UDI/DI 10809160489437, Kit Lot Number 1700807;
Model/Item Number (kit) SGSP11, UDI/DI 10809160461730, Kit Lot Number 1656113;
Model/Item Number (kit) SGSP11-03, UDI/DI 10809160511145, Kit Lot Number 1668007;
Model/Item Number (kit) SGSP11-03, UDI/DI 10809160511145, Kit Lot Number 1691439;
Model/Item Number (kit) SGSP11-03, UDI/DI 10809160511145, Kit Lot Number 1668009;
Model/Item Number (kit) SGSP11-03, UDI/DI 10809160511145, Kit Lot Number 1668008;
Model/Item Number (kit) SGSP44-02, UDI/DI 10809160469620, Kit Lot Number 1661236;
Model/Item Number (kit) SGSP44-03, UDI/DI 10809160490037, Kit Lot Number 1668013;
Model/Item Number (kit) SGSP44-03, UDI/DI 10809160490037, Kit Lot Number 1676432;
Model/Item Number (kit) SGSP44-03, UDI/DI 10809160490037, Kit Lot Number 1668011;
Model/Item Number (kit) SGSP44-03, UDI/DI 10809160490037, Kit Lot Number 1668014;
Model/Item Number (kit) SGSP44-03, UDI/DI 10809160490037, Kit Lot Number 1667304;
Model/Item Number (kit) SGSP44-03, UDI/DI 10809160490037, Kit Lot Number 1668012;
Model/Item Number (kit) SGSP91, UDI/DI 10809160462102, Kit Lot Number 1656747;
Model/Item Number (kit) SGSP91, UDI/DI 10809160462102, Kit Lot Number 1653016;
Model/Item Number (kit) SGSP91-02, UDI/DI 10809160469644, Kit Lot Number 1661240;
Model/Item Number (kit) SGSP91-03, UDI/DI 10809160511558, Kit Lot Number 1671768;
Model/Item Number (kit) SGSP91-03, UDI/DI 10809160511558, Kit Lot Number 1671769;
Model/Item Number (kit) SGSP91-03, UDI/DI 10809160511558, Kit Lot Number 1667616;
Model/Item Number (kit) SGSP91-03, UDI/DI 10809160511558, Kit Lot Number 1673361;
Model/Item Number (kit) SGSP91-03, UDI/DI 10809160511558, Kit Lot Number 1673362;
Model/Item Number (kit) SGSP91-03, UDI/DI 10809160511558, Kit Lot Number 1691066;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160363539, Kit Lot Number 1574836;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160363539, Kit Lot Number 1579673;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160363539, Kit Lot Number 1582963;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160363539, Kit Lot Number 1588397;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160363539, Kit Lot Number 1589834;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160363539, Kit Lot Number 1590905;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160363539, Kit Lot Number 1603715;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160363539, Kit Lot Number 1610818;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160363539, Kit Lot Number 1655825;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160363539, Kit Lot Number 1609281;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160363539, Kit Lot Number 1603716;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160363539, Kit Lot Number 1648191;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160363539, Kit Lot Number 1653849;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160363539, Kit Lot Number 1611976;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160363539, Kit Lot Number 1622358;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160363539, Kit Lot Number 1617996;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160363539, Kit Lot Number 1660322;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160363539, Kit Lot Number 1665018;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160502297, Kit Lot Number 1680232;
Model/Item Number (kit) SHND416-02, UDI/DI 10809160502297, Kit Lot Number 1691224;
Model/Item Number (kit) SHOR003-15, UDI/DI 10809160390498, Kit Lot Number 1566522;
Model/Item Number (kit) SHOR003-15, UDI/DI 10809160390498, Kit Lot Number 1568213;
Model/Item Number (kit) SHOR003-15, UDI/DI 10809160411513, Kit Lot Number 1612999;
Model/Item Number (kit) SHOR003-15, UDI/DI 10809160411513, Kit Lot Number 1613422;
Model/Item Number (kit) SHOR003-15, UDI/DI 10809160411513, Kit Lot Number 1644462;
Model/Item Number (kit) SLCN72-01, UDI/DI 10809160446416, Kit Lot Number 1636631;
Model/Item Number (kit) SLCN72-02, UDI/DI 10809160457030, Kit Lot Number 1647141;
Model/Item Number (kit) SLCN72-03, UDI/DI 10809160470701, Kit Lot Number 1667104;
Model/Item Number (kit) SLCN72-03, UDI/DI 10809160470701, Kit Lot Number 1662160;
Model/Item Number (kit) SLCN72-03, UDI/DI 10809160470701, Kit Lot Number 1662162;
Model/Item Number (kit) SLCN72-03, UDI/DI 10809160470701, Kit Lot Number 1668135;
Model/Item Number (kit) SLCN72-03, UDI/DI 10809160470701, Kit Lot Number 1691204;
Model/Item Number (kit) UDCP57-01, UDI/DI 10809160483404, Kit Lot Number 345399;
Model/Item Number (kit) UMIA104-11, UDI/DI 10809160465189, Kit Lot Number 1664831;
Model/Item Number (kit) UMIA104-11, UDI/DI 10809160465189, Kit Lot Number 1664832;
Model/Item Number (kit) USMA005, UDI/DI 10809160388297, Kit Lot Number 1669219;
Model/Item Number (kit) USMA005, UDI/DI 10809160388297, Kit Lot Number 1680592;
Model/Item Number (kit) VAML012-07, UDI/DI 10809160379011, Kit Lot Number 1612231;
Model/Item Number (kit) VAML028-06, UDI/DI 10809160379127, Kit Lot Number 1626107;
Model/Item Number (kit) VAPX009-02, UDI/DI 10809160302491, Kit Lot Number 1694165;
Model/Item Number (kit) VMCC024-14, UDI/DI 10809160458891, Kit Lot Number 1653475;
Model/Item Number (kit) VMED009-05, UDI/DI 10809160487808, Kit Lot Number 1663908;
Model/Item Number (kit) VMED009-05, UDI/DI 10809160487808, Kit Lot Number 1663979;
Model/Item Number (kit) WEAC53-01, UDI/DI 10809160485347, Kit Lot Number 349078;
Model/Item Number (kit) WECR41-01, UDI/DI 10809160485392, Kit Lot Number 348573;
Model/Item Number (kit) WECR41-01, UDI/DI 10809160485392, Kit Lot Number 348575;
Model/Item Number (kit) WECR41-01, UDI/DI 10809160485392, Kit Lot Number 348673;
Model/Item Number (kit) WELL73-01, UDI/DI 10809160485545, Kit Lot Number 344499;
Model/Item Number (kit) WELL73-01, UDI/DI 10809160485545, Kit Lot Number 348358;
Model/Item Number (kit) WELL73-01, UDI/DI 10809160485545, Kit Lot Number 348360;
Model/Item Number (kit) WELL73-01, UDI/DI 10809160485545, Kit Lot Number 348364;
Model/Item Number (kit) WELL73-01, UDI/DI 10809160485545, Kit Lot Number 348402;
Model/Item Number (kit) WELL73-01, UDI/DI 10809160485545, Kit Lot Number 348409;
Model/Item Number (kit) WKMC114-03, UDI/DI 10809160448069, Kit Lot Number 1652912;
Model/Item Number (kit) WKMC114-03, UDI/DI 10809160448069, Kit Lot Number 1660282;
Model/Item Number (kit) WKMC125-03, UDI/DI 10809160448113, Kit Lot Number 1641373;
Model/Item Number (kit) WRNM030-21, UDI/DI 10809160441961, Kit Lot Number 1661529. |
| FEI Number |
1047429
|
Recalling Firm/ Manufacturer |
AVID Medical, Inc.
9000 Westmont Dr
Toano
VA
23168-9351
|
| For Additional Information Contact | Marlene Jones 470-280-4277 |
Manufacturer Reason for Recall | The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential for unexpected variability of endotoxin levels on the products. |
FDA Determined Cause 2 | Material/Component Contamination |
| Action | Owens & Minor issued an URGENT: MEDICAL DEVICE RECALL NOTIFICATION to its consignees on 5/19/2026 via email. The notice explained the issue, potential risk to health and requested the following:
"ACTIONS TO TAKE:
" Identify, segregate, and quarantine all impacted product.
" Add recall labels to affected Avid kits to ensure all users are aware of the actions to discard the impacted component. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. Please contact your Sales Representative if pre-printed labels are needed. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling.
" Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have.
" Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness." |
| Quantity in Commerce | 16223 kits |
| Distribution | US Nationwide distribution in the states of AZ, CA, CO, WA, WY, TX, MI, FL, MD, UT, NC, MS, PA, MO, IA, NE, OH, IL, SD, MN, GA. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|