| | Class 2 Device Recall Presource Packs |  |
| Date Initiated by Firm | July 29, 2026 |
| Date Posted | September 03, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-3098-2026 |
| Recall Event ID |
99553 |
| Product Classification |
Cardiovascular procedure kit - Product Code OEZ
|
| Product | Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays:
1) KIT, GVI OPEN HEART A B , Catalog Number: PCVNCG540;
2) KIT, GVI OPEN HEART A B , Catalog Number: PCVNCG542;
3) KIT, GVI OPEN HEART A B , Catalog Number: PCVNCG55I;
4) KIT, BLO BASIC OPEN HEART , Catalog Number: PVCGBOB10;
5) KIT, BLO BASIC OPEN HEART , Catalog Number: PVCGBOB10;
6) KIT, BLO BASIC OPEN HEART , Catalog Number: PVCGBOB12;
7) KIT, BLO BASIC OPEN HEART , Catalog Number: PVCGBOB17;
8) KIT, OPEN HEART CON , Catalog Number: PVCGOHCOH;
9) KIT,OPEN HEART ADULT MDST , Catalog Number: PVCGOHCPA;
10) KIT, CABG , Catalog Number: PVHDCBAGK;
11) KIT, ORP CABG , Catalog Number: PVHDCGVCP;
12) KIT, ORP CABG , Catalog Number: PVHDCGV40;
13) KIT, VALVE , Catalog Number: PVHDVAAGH;
14) KIT, ORP VALVE , Catalog Number: PVHDVPVCQ;
15) KIT, ORP VALVE , Catalog Number: PVHDVPV43;
16) KIT, MAJOR CARDIAC , Catalog Number: PVOCMGJ36;
17) KIT, OPEN HEART DH , Catalog Number: PVOCOHNHE;
18) KIT, OPEN HEART , Catalog Number: PVOCOH12O;
19) KIT, CAROTID ENDARTERECTOMY , Catalog Number: PVVMCV07E;
20) KIT, FEM POP BARNES JEW , Catalog Number: PVVMFP22R;
21) KIT, FEM POP BARNES JEW , Catalog Number: PVVMFP245;
22) KIT, THORACIC BARNES JEWISH , Catalog Number: PVVMTH22J;
23) KIT, THORACIC OUTLET SYNDROME , Catalog Number: PVVMTO22R;
24) KIT, THORACIC OUTLET SYNDROME , Catalog Number: PVVMTO241;
25) KIT, THORACIC OUTLET SYNDROME , Catalog Number: PVVMTO242;
26) KIT, THORACIC , Catalog Number: PVVXTHLCG;
27) KIT,OPEN HEART BASIC PRESBYTER, Catalog Number: PVV2CC10R;
28) KIT,OPEN HEART BASIC PRESBYTER, Catalog Number: PVV2CC10S;
29) KIT, MAJOR VASC, PRESBYTERIAN , Catalog Number: PVV2MJ01M;
30) KIT, CARDIAC VALVE , Catalog Number: PVV207962;
31) KIT, CARDIAC VALVE , Catalog Number: PVV207963;
32) KIT, CARDIAC VALVE , Catalog Number: PVV207964;
33) KIT, CARDIAC CABG , Catalog Number: PVV293340;
34) KIT, CARDIAC CABG , Catalog Number: PVV293362;
35) KIT, CARDIAC CABG , Catalog Number: PVV293363;
36) KIT, PUH CARDIAC CABG , Catalog Number: PVX4CAP14;
37) KIT, PUH CARDIAC , Catalog Number: PVX4CRPR7;
38) KIT, EAS THORACIC ABDOMINAL , Catalog Number: PVX4THEAA;
39) KIT, CV , Catalog Number: PV15CGSPF;
40) KIT, CV , Catalog Number: PV15CGSPG;
41) KIT,MAJOR VASCULAR , Catalog Number: PV15MJSPD;
42) KIT, OPEN HEART SET UP , Catalog Number: PV24OHN29;
43) KIT, OPEN HEART , Catalog Number: PV29OHMME;
44) KIT, OPEN HEART , Catalog Number: PV3LOHSGJ;
45) KIT, MER THORACOTOMY CHEST , Catalog Number: PV33TCMYB;
46) KIT, MER THORACOTOMY CHEST , Catalog Number: PV33TCMY3;
47) KIT, BASIC OPEN HEART , Catalog Number: PV40OHAR47;
48) KIT, BASIC OPEN HEART , Catalog Number: PV40OHAR48;
49) KIT, BASIC OPEN HEART , Catalog Number: PV40OHAR49;
50) KIT, OPEN HEART , Catalog Number: PV40OHJUD;
51) KIT, MODULE OPEN HEART , Catalog Number: PV60OH15F;
52) CV W AUX PACK 2 OF 2 CMC , Catalog Number: SCVCGPAACD;
53) THORACIC ROBOTIC PACK CMC , Catalog Number: SCVCGTRACI;
54) CV PACK 2 OF 2 GSH , Catalog Number: SCVCGVCG17;
55) CV PACK 2 OF 2 GSH , Catalog Number: SCVCGVCG18;
56) CV PACK 2 OF 2 SHER , Catalog Number: SCVCGVCS16 |
| Code Information |
1) PCVNCG540, UDI-DI: 10198956230057(each), 50198956230055(case), Lot Number: NGKT3319 ;
2) PCVNCG542, UDI-DI: 10198956312463(each), 50198956312461(case), Lot Number: NGKU0951 ;
3) PCVNCG55I, UDI-DI: 10197106283059(each), 50197106283057(case), Lot Number: NGKU0977 ;
4) PVCGBOB10, UDI-DI: 10198956219885(each), 50198956219883(case), Lot Number: NGKT2391 ;
5) PVCGBOB11, UDI-DI: 10198956220676(each), 50198956220674(case), Lot Number: NGKT2391 ;
6) PVCGBOB12, UDI-DI: 10198956256477(each), 50198956256475(case), Lot Number: NGKT2391 ;
7) PVCGBOB17, UDI-DI: 10198956393387(each), 50198956393385(case), Lot Number: NGKT2391 ;
8) PVCGOHCOH, UDI-DI: 10198956022270(each), 50198956022278(case), Lot Number: NGKT2391 ;
9) PVCGOHCPA, UDI-DI: 10197106666777(each), 50197106666775(case), Lot Number: NGKT3319 ;
10) PVCGOHCPA, UDI-DI: 10197106666777(each), 50197106666775(case), Lot Number: NGKT3319 ;
11) PVCGOHCPA, UDI-DI: 10197106666777(each), 50197106666775(case), Lot Number: NGKU0951 ;
12) PVHDCBAGK, UDI-DI: 10197106946473(each), 50197106946471(case), Lot Number: NGKT2391 ;
13) PVHDCBAGK, UDI-DI: 10197106946473(each), 50197106946471(case), Lot Number: NGKT2391 ;
14) PVHDCBAGK, UDI-DI: 10197106946473(each), 50197106946471(case), Lot Number: NGKT2391 ;
15) PVHDCGVCP, UDI-DI: 10198956254596(each), 50198956254594(case), Lot Number: NGKV2393 ;
16) PVHDCGVCP, UDI-DI: 10198956254596(each), 50198956254594(case), Lot Number: NGKV2393 ;
17) PVHDCGVCP, UDI-DI: 10198956254596(each), 50198956254594(case), Lot Number: NGKW2687 ;
18) PVHDCGVCP, UDI-DI: 10198956254596(each), 50198956254594(case), Lot Number: NGKV2393 ;
19) PVHDCGVCP, UDI-DI: 10198956254596(each), 50198956254594(case), Lot Number: NGKV2393 ;
20) PVHDCGV40, UDI-DI: 10198956326941(each), 50198956326949(case), Lot Number: NGKV2393 ;
21) PVHDVAAGH, UDI-DI: 10197106946459(each), 50197106946457(case), Lot Number: NGKT2391 ;
22) PVHDVPVCQ, UDI-DI: 10198956254619(each), 50198956254617(case), Lot Number: NGKV2393 ;
23) PVHDVPVCQ, UDI-DI: 10198956254619(each), 50198956254617(case), Lot Number: NGKV2393 ;
24) PVHDVPVCQ, UDI-DI: 10198956254619(each), 50198956254617(case), Lot Number: NGKW2687 ;
25) PVHDVPVCQ, UDI-DI: 10198956254619(each), 50198956254617(case), Lot Number: NGKW2687 ;
26) PVHDVPV43, UDI-DI: 10198956354838(each), 50198956354836(case), Lot Number: NGKW2687 ;
27) PVOCMGJ36, UDI-DI: 10198956248663(each), 50198956248661(case), Lot Number: NGKT2391 ;
28) PVOCOHNHE, UDI-DI: 10198956071940(each), 50198956071948(case), Lot Number: NGKT2391 ;
29) PVOCOHNHE, UDI-DI: 10198956071940(each), 50198956071948(case), Lot Number: NGKT2391 ;
30) PVOCOHNHE, UDI-DI: 10198956071940(each), 50198956071948(case), Lot Number: NGKT2455 ;
31) PVOCOH12O, UDI-DI: 10198956202054(each), 50198956202052(case), Lot Number: NGKT2391 ;
32) PVVMCV07E, UDI-DI: 10198956179615(each), 50198956179613(case), Lot Number: NGKT2394 ;
33) PVVMFP22R, UDI-DI: 10198956179653(each), 50198956179651(case), Lot Number: NGKT2394 ;
34) PVVMFP245, UDI-DI: 10198956249141(each), 50198956249149(case), Lot Number: NGKT2394 ;
35) PVVMTH22J, UDI-DI: 10198956119314(each), 50198956119312(case), Lot Number: NGKU0856 ;
36) PVVMTH22J, UDI-DI: 10198956119314(each), 50198956119312(case), Lot Number: NGKU0957 ;
37) PVVMTH22J, UDI-DI: 10198956119314(each), 50198956119312(case), Lot Number: NGKU0856 ;
38) PVVMTO22R, UDI-DI: 10198956161016(each), 50198956161014(case), Lot Number: NGKT2394 ;
39) PVVMTO241, UDI-DI: 10198956221741(each), 50198956221749(case), Lot Number: NGKT2394 ;
40) PVVMTO242, UDI-DI: 10198956300521(each), 50198956300529(case), Lot Number: NGKT2394 ;
41) PVVXTHLCG, UDI-DI: 10198956170759(each), 50198956170757(case), Lot Number: NGKT1789 ;
42) PVV2CC10R, UDI-DI: 10198956285262(each), 50198956285260(case), Lot Number: NGKV3390 ;
43) PVV2CC10S, UDI-DI: 10198956346611(each), 50198956346619(case), Lot Number: NGKV3390 ;
44) PVV2CC10S, UDI-DI: 10198956346611(each), 50198956346619(case), Lot Number: NGKV2390 ;
45) PVV2MJ01M, UDI-DI: 10198956285224(each), 50198956285222(case), Lot Number: NGKV3390 ;
46) PVV207962, UDI-DI: 10198956212503(each), 50198956212501(case), Lot Number: NGKT2391 ;
47) PVV207963, UDI-DI: 10198956244245(each), 50198956244243(case), Lot Number: NGKT2455 ;
48) PVV207964, UDI-DI: 10198956255142(each), 50198956255140(case), Lot Number: NGKT2391 ;
49) PVV293340, UDI-DI: 10198956218451(each), 50198956218459(case), Lot Number: NGKT2391 ;
50) PVV293362, UDI-DI: 10198956240605(each), 50198956240603(case), Lot Number: NGKT2391 ;
51) PVV293363, UDI-DI: 10198956248007(each), 50198956248005(case), Lot Number: NGKT2391 ;
52) PVX4CAP14, UDI-DI: 10198956258921(each), 50198956258929(case), Lot Number: NGKU1994 ;
53) PVX4CRPR7, UDI-DI: 10198956244894(each), 50198956244892(case), Lot Number: NGKU1994 ;
54) PVX4THEAA, UDI-DI: 10197106795033(each), 50197106795031(case), Lot Number: NGKU4043 ;
55) PVX4THEAA, UDI-DI: 10197106795033(each), 50197106795031(case), Lot Number: NGKU4716 ;
56) PV15CGSPF, UDI-DI: 10197106847046(each), 50197106847044(case), Lot Number: NGKU0977 ;
57) PV15CGSPF, UDI-DI: 10197106847046(each), 50197106847044(case), Lot Number: NGKU0951 ;
58) PV15CGSPG, UDI-DI: 10198956324992(each), 50198956324990(case), Lot Number: NGKT3319 ;
59) PV15CGSPG, UDI-DI: 10198956324992(each), 50198956324990(case), Lot Number: NGKU0951 ;
60) PV15MJSPD, UDI-DI: 10197106847190(each), 50197106847198(case), Lot Number: NGKT3319 ;
61) PV15MJSPD, UDI-DI: 10197106847190(each), 50197106847198(case), Lot Number: NGKU0977 ;
62) PV24OHN29, UDI-DI: 10198956266315(each), 50198956266313(case), Lot Number: NGKU0977 ;
63) PV29OHMME, UDI-DI: 10198956042827(each), 50198956042825(case), Lot Number: NGKT3319 ;
64) PV29OHMME, UDI-DI: 10198956042827(each), 50198956042825(case), Lot Number: NGKU0977 ;
65) PV3LOHSGJ, UDI-DI: 10198956127968(each), 50198956127966(case), Lot Number: NGKU1994 ;
66) PV33TCMYB, UDI-DI: 10197106892442(each), 50197106892440(case), Lot Number: NGKU4716 ;
67) PV33TCMY3, UDI-DI: 10198956263994(each), 50198956263992(case), Lot Number: NGKU2887 ;
68) PV40OHAR47, UDI-DI: 10198956247086(each), 50198956247084(case), Lot Number: NGKT2391 ;
69) PV40OHAR48, UDI-DI: 10198956249165(each), 50198956249163(case), Lot Number: NGKT2391 ;
70) PV40OHAR48, UDI-DI: 10198956249165(each), 50198956249163(case), Lot Number: NGKT2455 ;
71) PV40OHAR49, UDI-DI: 10198956257924(each), 50198956257922(case), Lot Number: NGKT2391 ;
72) PV40OHJUD, UDI-DI: 10197106874844(each), 50197106874842(case), Lot Number: NGKT2455 ;
73) PV40OHJUD, UDI-DI: 10197106874844(each), 50197106874842(case), Lot Number: NGKT2391 ;
74) PV40OHJUD, UDI-DI: 10197106874844(each), 50197106874842(case), Lot Number: NGKT2455 ;
75) PV60OH15F, UDI-DI: 10197106634905(each), 50197106634903(case), Lot Number: NGKT2391 ;
76) SCVCGPAACD, UDI-DI: 10197106748169(each), 50197106748167(case), Lot Number: NGKU0977 ;
77) SCVCGPAACD, UDI-DI: 10197106748169(each), 50197106748167(case), Lot Number: NGKU0951 ;
78) SCVCGPAACD, UDI-DI: 10197106748169(each), 50197106748167(case), Lot Number: NGKT3319 ;
79) SCVCGTRACI, UDI-DI: 10197106453735(each), 50197106453733(case), Lot Number: N/A ;
80) SCVCGVCG17, UDI-DI: NGKU0957 (each), 10198956229624(case), Lot Number: 50198956229622;
81) SCVCGVCG18, UDI-DI: NGKU0856 (each), 10198956230651(case), Lot Number: 50198956230659;
82) SCVCGVCS16, UDI-DI: 10198956408548(each), 50198956408546(case), Lot Number: N/A ;
83) SCVCGVCS16, UDI-DI: 10198956408548(each), 50198956408546(case), Lot Number: NGKV5078 |
| FEI Number |
3001236905
|
Recalling Firm/ Manufacturer |
Cardinal Health 200, LLC
3651 Birchwood Dr
Waukegan
IL
60085-8337
|
| For Additional Information Contact | Cardinal Health Quality Systems 800-292-9332 |
Manufacturer Reason for Recall | The Urinary Catheter Trays contain CliniClean Pericare Wipes which may potentially be contaminated with a lubricant used during the manufacturing process. If a contaminated wipe is used on a patient, localized skin effects may occur. These may include irritation, erythema, or mild inflammation. |
FDA Determined Cause 2 | Process change control |
| Action | An URGENT: PRODUCT CORRECTION notification letter dated 7/29/26 was sent to customers.
If you locate affected product within your facility, please do the following:
1. QUARANTINE affected kits.
2. ALERT your clinicians of this action.
3. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to
clinicians, instructing them to remove and discard the recalled component(s).
4. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this
action and the completion of labeling product via fax to 614.652.9648.
Additional labels will be sent to your facility upon request. Please send an email to:
gmb-FieldCorrectiveAction@cardinalhealth.com
For component credit or replacement, please contact the following:
" Hospitals Cardinal Customer Service at 866.551.0533
" Distributors Distributor Management at 800.635.6021
" Federal Government facilities Government Customer Service at 800.444.1166
If you are a Distributor and have further distributed affected kits please notify your customers of this recall.
We sincerely apologize for any inconvenience this notice may have caused you and your staff.
Should you have any questions, require additional labels or desire special assistance relating to this recall, please feel free to contact Cardinal Health Quality Systems at 800.292.9332. |
| Quantity in Commerce | 2448 units |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|