| | Class 2 Device Recall JenaValve Trilogy Delivery Catheter, |  |
| Date Initiated by Firm | July 31, 2026 |
| Date Posted | October 09, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0134-2027 |
| Recall Event ID |
99697 |
| PMA Number | P250024 |
| Product Classification |
Aortic valve, prosthesis, percutaneously delivered - Product Code NPT
|
| Product | JenaValve Trilogy Delivery Catheter, Size 23MM
REF: TRILOGY-DC-S |
| Code Information |
GTIN - 00850000898134/ Lot: DS-040826-003
DS-042326-003,
DS-042726-006,
DS-042826-005,
DS-050426-007,
DS-050626-002,
DS-050626-006,
DS-050626-009,
DS-050726-005
|
| FEI Number |
3012148004
|
Recalling Firm/ Manufacturer |
Jenavalve Technology Inc.
4 Cromwell
Irvine
CA
92618-1816
|
| For Additional Information Contact | John Kilcoyne +1-949-767-2110 Ext. 1030 |
Manufacturer Reason for Recall | Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | On the dates of August 1, 4, 6, and 19th, JenaValve Technology Inc. issued a Urgent: Medical Device Recall notification via email. JenaValve asked consignees to take the following actions:
1. Review the information contained in this letter and ensure that anyone in your facility who needs to be aware of this notification reads this letter carefully.
2. Examine your inventory immediately to determine if you have Trilogy Delivery Catheters from the affected lots and quarantine the subject products.
3. JenaValve field team members will schedule a visit to retrieve the subject Trilogy Delivery Catheters. If you prefer to return the products yourself, please contact us at 866-588-8287 and we will provide you with an RGA number prior to you shipping them to:
JenaValve Technology Inc., 4 Cromwell, Irvine, California 92618, USA; Attn: Quality Department; Reference: FA26001
4. Please complete and sign the Business reply form (BRF) of this letter and email it to customersupport.us@jenavalve.com or give it to your JenaValve field representative.
5. Upon receipt of the completed BRF and returned product, JenaValve will provide replacement products. Incomplete BRFs cannot be processed.
6. If any of the subject Trilogy Delivery Catheters have been transferred to another facility, identify the receiving facility on the Business Reply Form, contact customersupport.us@jenavalve.com, and provide a copy of this letter to the relevant personnel.
|
| Quantity in Commerce | 111 units |
| Distribution | US Nationwide distribution in the states of AZ, CA, FL, GA, IL, MA, MD, MI, MN, MO, NY, OH, PA, SC, TX, UT, VA, WA and the countries of Germany, Denmark, Great Britain, Italy.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| PMA Database | PMAs with Product Code = NPT
|
|
|
|