| | Class 2 Device Recall CareLink Clinic Software Version 4.4A |  |
| Date Initiated by Firm | September 11, 2026 |
| Date Posted | October 21, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0203-2027 |
| Recall Event ID |
99765 |
| Product Classification |
Continuous glucose monitor retrospective data analysis software - Product Code PHV
|
| Product | CareLink Clinic, REF: MMT-7350 |
| Code Information |
UDI-DI: 0763000B00008748K.
Software Versions: 4.4A
Software Available:
US: Between June 4th, 2026, and June 8th, 2026. |
| FEI Number |
3003166194
|
Recalling Firm/ Manufacturer |
Medtronic MiniMed, Inc.
18000 Devonshire St
Northridge
CA
91325-1219
|
| For Additional Information Contact | Ashley Patterson 800-646-4633 |
Manufacturer Reason for Recall | CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acting insulin when calculating the Carb Ratio, Insulin Sensitivity Factor (ISF), and Max Recommended value for Rapid-acting insulin. Instead, the calculator continues to use the system-recommended values, resulting in incorrect rapid-acting insulin parameters. If an HCP or patient inputs these incorrect setting values into the patient's MiniMed Go app, the patient could deliver higher or lower insulin doses compared to their insulin needs, potentially leading to over/under delivery of insulin with a severity ranging from Catastrophic (death from severe hypoglycemia) to Minor (self-treatable hypoglycemia/hyperglycemia). |
FDA Determined Cause 2 | Software design |
| Action | On 9/11/2026, correction notices were mailed to the health care professional, and customer asking them to do the following:
The Customer was asked to do the following:
Immediately contact your Healthcare Provider (HCP) to confirm your insulin settings values.
Complete and return the confirmation form by visiting https://info.medtronicdiabetes.com/FA1587
The health care professional was asked to do the following: If you provided diabetes treatment recommendations to patients based on custom Total Daily Dose(TDD) or long-acting insulin values entered into the CareLin Clinic Settings Calculator between: June 04, 2026, to June 08, 2026, please revisit those recommendations utilizing patients' individual CareLink reports.
Continue to follow up with those patients pursuant to your standard of care.
If you have further questions or need assistance, please call the firm's 24-Hour Technical Support line at 1-800-646-4633, option 1. |
| Quantity in Commerce | 1 |
| Distribution | US Nationwide distribution in the state of KS. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|