| | Class 3 Device Recall LeadCare II |  |
| Date Initiated by Firm | September 04, 2026 |
| Date Posted | October 15, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0162-2027 |
| Recall Event ID |
99776 |
| 510(K)Number | K052549 |
| Product Classification |
Lead, atomic absorption - Product Code DOF
|
| Product | LeadCare II Blood Lead Test Kit. Model/Catalog Number: 70-6762. |
| Code Information |
Model/Catalog Number: 70-6762. UD-DI: 00850355006000. Lot Number. 2530M-01.
|
| FEI Number |
1218996
|
Recalling Firm/ Manufacturer |
Magellan Diagnostics, Inc.
101 Billerica Ave Bldg 4
North Billerica
MA
01862-1271
|
| For Additional Information Contact | Ryan Spradling 1-800-2750102 |
Manufacturer Reason for Recall | In kit sub-lot 2530M-01 the date of expiration printed on the outer labeling may indicate the wrong expiration date of July 22, 2026 while the correct expiration date is July 13, 2026. If used the system will display an error message and prevent the operator from running the test. Resulting in delayed results. |
FDA Determined Cause 2 | Labeling Change Control |
| Action | Magellan notified consignees on about September 4, 2026, via telephone call indicating a written notification of the recall would follow. On 09/17/2026, a notification letter was emailed to consignees with the following instructions:
1. Complete, sign, and return the CONFIRMATION OF NOTIFICATION FORM.
2. Review your inventory for the affected LeadCare II Blood Lead Test Kit lot identified.
3. Immediately discontinue use of any remaining affected kits.
4. Return any recalled product still in your possession to Magellan Diagnostics, Inc. for destruction. |
| Quantity in Commerce | 321 units |
| Distribution | US Nationwide distribution in the states of CA, IN, PA, TX, NE, IL, MA, MO, KY. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|