Class 2 Device Recall Artisan Link Pro Special Staining System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Artisan Link Pro Special Staining Systemsee related information
Date Initiated by FirmAugust 11, 2026
Date PostedNovember 27, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0564-2027
Recall Event ID 99775
Product Classification Slide stainer, automated - Product Code KPA
ProductGeneric Name: Artisan Link Pro Special Staining System Generic Name: Artisan Link Pro Model/Catalog Number: AR31027 Software Version: Not Applicable Product Description: SLIDE STAINER, AUTOMATED Component: No
Code Information Lot Code: UDI-DI 05700571112786 All product serial numbers.
FEI Number 3003423869
Recalling Firm/
Manufacturer
Agilent Technologies Singapore (International) Pte Ltd
1 Yishun Avenue 7
Agilent Technologies
Singapore Singapore
Manufacturer Reason
for Recall
The analyzer may experience reagent overflow, resulting in heater unit malfunction or inoperability. This could impact instrument performance and/or cause minor burns to users.
FDA Determined
Cause 2
Device Design
ActionOn 09/02/2026, an "URGENT: MEDICAL DEVICE CORRECTION" letter was emailed to customers stating that, under certain operating conditions, Artisan Link Pro instruments may experience reagent liquid overflow that can expose heater boards to liquid. This occurs when the slide clip fails to form a proper seal, allowing reagent to leak from the reagent bath onto the heater electronics. If reagent contacts the heater board, the affected board may short circuit or malfunction, causing the heater plate to operate at an incorrect temperature or cease heating. Minor burns may result from touching an overheated plate, and a short circuit could cause self-limiting ignition of ethanol-containing reagents. These conditions may also interrupt or compromise staining runs. Conditions that may compromise the slide clip seal include (see Appendix A for images): Customers are instructed to: Follow all guidance in the Basic User Guide (Instructions For Use | Agilent), including: Immediately power off the instrument and contact Agilent service if liquid overflow or heating malfunctions are observed. Do not touch the heater plate; use designated tabs to rotate the carousel. Ensure all waste bottle caps are securely tightened before operation. Verify the "Waste Valve Rinse" option is enabled prior to use. Use small flap labels, securely attached; do not use double labels. Check the drip tray after each run. Use slide clips once only and store per packaging instructions. Perform the "Prime All" function before each run. Ensure waste bottle caps are tightened and all recommended maintenance procedures are followed. Contact local technical support for issues or questions. If a control slide is affected, follow the applicable reagent Instructions For Use. For questions - contact firm at fieldactions@agilent.com.
Quantity in Commerce23
DistributionWorldwide - U.S. Nationwide distribution including the states and territory of AK, AL AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. The countries of Australia, Austria, Bahrain, Belgium, Brazil, Canada, Canary Islands, Chile, China, Costa Rica, Croatia, Cyprus, Czechia, Denmark, Ecuador, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Morocco, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Poland, Portugal, Qatar, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey, United Arab Emirates, United Kingdom.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-