 Class 1 Device Recall HemoslL AcuStar HITlgG(PF4)see related information
Date Initiated by FirmSeptember 03, 2026
Date PostedOctober 10, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0001-2027
Recall Event ID 99843
510(K)NumberK170854 
Product Classification Platelet factor 4 radioimmunoassay - Product Code LCO
ProductHemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009800127. in vitro diagnostic
Code Information Part Number: 0009800127. UDI-DI: 08426951126545. Lot Number (Expiration Date): B40005 (01/16/2027), B38062 (11/01/2026).
FEI Number 1217183
Recalling Firm/
Manufacturer
Instrumentation Laboratory
180 Hartwell Rd
Bedford MA 01730-2443
For Additional Information ContactKeith Kelly
732-3792123
Manufacturer Reason
for Recall
Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
FDA Determined
Cause 2
Under Investigation by firm
ActionInstrumentation Laboratory notified consignees on about 09/03/2026 via emailed and mailed URGENT MEDICAL DEVICE REALL letter. Consignees were made aware of the issue and hazard involved. They were instructed to check inventory for affected lots, destroy any remaining product following each facility's standard procedures, and document the destruction on the enclosed Urgent Mandatory Response Tracking Form, including quantity destroyed and date of destruction. Consignees were asked to identify an alternative method for HIT antibody testing as a supply interruption is expected and work local Werfen representatives for guidance and transition support. Consignees were also requested to inform all laboratory personnel of the notification, forward the notification to all affected departments and locations within each facility, retain a copy of the notification for records and complete and return the Urgent Mandatory Response Tracking Form within 10 days of receipt even if there is no remaining inventory. On 09/29/2026, an expanded URGENT MEDICAL DEVICE RECALL Notification was sent to consignees through registered mail or email. The updated notification informed consignees of additional lot numbers affected. Instructions remained the same.
Quantity in Commerce434 units
DistributionWorldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, FL, GA, MA, MD, NJ, NY, PA, RI, TN, WA, WV and the countries of Austria, Belgium, Switzerland, Costa Rica, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, India, Italy, Jordan, Japan, South Korea, Luxembourg, Latvia, Netherlands, Saudi Arabia, Sweden, Singapore.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.