• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Precision Pointer Electromagnetic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Precision Pointer Electromagneticsee related information
Date Initiated by FirmSeptember 24, 2026
Date PostedOctober 29, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0314-2027
Recall Event ID 99855
510(K)NumberK212194 
Product Classification Neurological stereotaxic Instrument - Product Code HAW
ProductPrecision Pointer Electromagnetic (8000-050-001)
Code Information Lot 7000413361 UDI-DI 07613327413205
FEI Number 3002821504
Recalling Firm/
Manufacturer
Stryker Leibinger GmbH & Co. KG
Botzinger Str. 39-41
Freiburg Im Breisgau Germany
For Additional Information ContactKatharina Feige-Goetz
+497614512-0 Ext. 0
Manufacturer Reason
for Recall
Device may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the device is inside the patient.
FDA Determined
Cause 2
Process control
ActionThe firm began notifying customers on September 24, 2026 via Urgent Medical Device Recall letters. Customers were instructed to immediately locate and quarantine any affected products at their facility. Customers may either destroy affected products onsite with their Stryker representative and send a completed Certificate of Destruction to the firm, or return the products to Stryker. The firm will provide a replacement upon receipt of the recalled product or Certificate of Destruction.
Quantity in Commerce513
DistributionWorldwide - US Nationwide distribution in the states of Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Hawaii, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Michigan, Missouri, Montana, New Hampshire, New York, Ohio, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Washington and the countries of Austria, Canada, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Norway, Portugal, South Korea, Spain, Switzerland, UK.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = HAW
-
-